You are on page 1of 6

Bodily Fluid Analysis of Non-Serum Samples using Point-of-Care

Testing with iSTAT and Piccolo Analyzers Versus a Fixed Hospital


Chemistry Analytical Platform

William Londeree MD; Konrad Davis MD; Donald Helman MD; and Jude Abadie PhD

Abstract the POCT devices have a potential role in the analysis of non-
Introduction: Forward deployed military medical units can provide sophisti- serum samples such as pleural, peritoneal, and cerebrospinal
cated medical care with limited resources. Point-of-Care Testing (POCT) may fluid.
facilitate care and expedite diagnosis. This study assessed the accuracy of
results for POCT for non-serum samples (pleural, peritoneal, and cerebrospi-
nal fluid) using iSTAT and Piccolo hand-held devices compared with results Methods and Materials
obtained using a hospital chemistry analyzer. Existing pleural, peritoneal, and cerebrospinal fluids collected
Methods: Pleural, peritoneal, and cerebrospinal fluids obtained during rou- for standard of care analysis were stored at -20°C for further
tine care were simultaneously analyzed on a Vitros 5600 automated clinical POCT analysis. The samples were thawed and analyzed on
chemistry hospital analyzer, iSTAT, and Piccolo POCT devices. an Ortho Clinical Diagnostics Vitros 5600 automated clinical
Results: POCT results were highly correlated with the Vitros 5600 for pleural chemistry analytical platform, considered be the analytical ref-
fluid LDH, glucose, and triglycerides (TG); for peritoneal fluid bilirubin, TG,
erence standard in this study. All samples were simultaneously
glucose, albumin, and protein; and glucose for cerebrospinal fluid.
Conclusion: POCT results for non-serum samples from pleural, peritoneal, analyzed using an iSTAT and/or Piccolo POCT hand-held device,
and cerebrospinal fluid correlate with standard hospital chemistry analysis. employing test-specific cartridges for individual analytes. The
The results of this study demonstrate potential for possible new diagnostic iSTAT used EC8+ cartridge, and the Piccolo used Lipid Panel,
roles for POCT in resource-limited environments. Basic Metabolic Plus, and General Chemistry 13 cartridges.
Pleural fluid analysis measured glucose and pH on the iSTAT
Introduction using an EC8+ cartridge. The Piccolo POCT device measured
The delivery of quality healthcare in developing countries LDH on the Basic Metabolic Panel Plus cartridge. Amylase,
has dramatically increased during the last 10-15 years.1 This protein, and albumin were measured using the General Chem-
growth is related to global health initiatives and advances in sitry13 cartridge, and the Lipid Panel cartridge was used to
medical technology.2 The advent and expansion of point-of- quantify triglycerides (TG) and cholesterol levels in pleural
care technology illustrates how diagnostic acumen in austere fluid.
or low-income environments has changed. Point-of-care testing Peritoneal fluid analytes were only measured with the Pic-
(POCT) offers rapid and accurate biochemical testing on serum colo POCT device. The Piccolo measured LDH with the Basic
samples. POCT has well established clinical utility for serum Metabolic Panel Plus cartridge and amylase, protein, albumin,
chemistry and blood gas analysis, as well as for the assessment bilirubin; however, the glucose was measured with the General
of diabetes, pregnancy, HIV, and malaria.3 Chemistry13 cartridge while the Lipid Panel cartridge was used
Most point-of-care technologies have targeted blood tests, to measure triglycerides (TG).
yet comprehensive laboratory testing (to include the evalua- The EC8+ cartridge was used to measure cerebrospinal fluid
tion of non-serum samples) continues to require conventional glucose using the iSTAT’s platform.
bench top analytic platforms, whose cost and size limit their The null hypothesis was, “The differences of the means
practicality in resource constrained environments. Only two between POCT and bench top analysis is 0.” Bland-Altman
studies have compared POCT to traditional biochemical analysis charts were constructed to assess accuracy and precision of the
of non-serum samples in humans. Kohn, et al, demonstrated POCT analyzer compared to the Vitros 5600. The Bland-Altman
that pleural fluid pH can be accurately performed on an iSTAT charts plot the difference of the paired measurements for each
POCT device.4 Wockenfus, et al, demonstrated that the iSTAT is sample against the average, and also display 95% confidence
comparable to a validated blood gas analyzer for pleural fluid pH intervals.7 A two-sided paired t-test assessed differences between
analysis.5 An animal study on equine synovial fluid accurately the two measurements. A mean difference significantly different
measured lactate and glucose but was limited in the number of from zero indicated the average bias for the POCT analyzer.
samples (N = 8).6 Overall, there is scant evidence-based literature For analytes with a large number of measurements below the
regarding the use of POCT on non-serum samples, and those limit of detection, results were dichotomized as detected or
uses are not approved by the Food and Drug Administration non-detected, and a kappa statistic assessed concordance. A
(FDA). significance of P ≤ .05 was used for all statistical tests. All
The purpose of this study was to compare the accuracy of analyses were performed using Statistical Analysis Software
iSTAT and Piccolo devices to a standard hospital bench top (SAS) v 9.2 software, SAS Institute, Cary, NC.
analyzer using analyte measurement, in order to determine if

HAWAI‘I JOURNAL OF MEDICINE & PUBLIC HEALTH, SEPTEMBER 2014, VOL 73, NO 9, SUPPLEMENT 1
3
Correlation was determined using a correlation coefficient min, glucose, and cholesterol based on the null hypothesis;
(r), reflecting a linear relationship of POCT (y-axis) versus however, protein, TG, and pH did not demonstrate a statistical
control (x-axis) values with increasing correlation as 1.00 is difference, with P values of .96, .08, and .11, respectively. The
approached. Because the sensitivity of the control and POCT mean values represented positive biases for protein, albumin,
analyzers varied, a concordance (k) value was calculated that glucose, and pH at 0.01, 0.27, 11, and 0.01. A negative bias was
represented the reliability of POCT to categorize samples as calculated for LDH (-343), cholesterol (-16), and TG (-2.7).
detectable or non-detectable when compared to the control, Correlation (r) values were determined for glucose (0.99), pH
with increasing k as 1.00 is approached. (0.98), LDH (0.97), protein (0.90), albumin (0.81), TG (0.91),
and cholesterol (0.91).
Results The Bland Altman chart for pH is an example of how each
Pleural Fluid analyte was plotted to demonstrate the difference of the paired
Table 1 summarizes the statistical results for pleural fluid. The measurements for each sample versus the average. Figure 1 is
number of samples for the analytes was 6-23 (n) with range of a representation of pH. Correlation is reflected in Figure 2 for
data values referenced in table 1. Statistical difference (P ≤ .05) pH as linear relationship of POCT versus Vitros 5600 values,
between testing modalities was demonstrated for LDH, albu- with increasing correlation as 1.00 is approached.

Table 1. Pleural Fluid Results


n Mean Std Mean P-value Precision Correlation Concordance Data Range
95% CI 95% CI (r) (k)
LDH (U/L) 12 -343 243 -498/ -188 < .01 -820/134 0.97 1.00 2018
Protein (g/dL) 17 0.01 0.52 -0.26/0.27 .96 -1.01/1.03 0.90 1.00 3.7
Albumin (g/dL) 17 0.27 0.35 0.09/0.45 < .01 -0.41/0.95 0.81 0.16 1.37
Glucose (mg/dL) 23 11 8 8/15 < .01 -5/27 0.99 1.00 200
Cholesterol (mg/dL) 6 -16 11 -28/-4.6 .02 -38/5.9 0.91 0.87 64
TG (mg/dL) 17 -2.7 5.9 -5.7/0.32 .08 -14/8.8 0.91 0.61 42
pH 20 0.01 0.04 -0.003/0.028 .11 -0.06/0.08 0.98 0.74 1.05

Figure 1. Bland Altman Chart for pH

HAWAI‘I JOURNAL OF MEDICINE & PUBLIC HEALTH, SEPTEMBER 2014, VOL 73, NO 9, SUPPLEMENT 1
4
Peritoneal Fluid Cerebrospinal Fluid
Table 2 summarizes the statistical data for peritoneal fluid. Table 3 summarizes the data for cerebrospinal fluid. Fifty-five
Protein and TG were the only two analytes where statistical samples were obtained in total and glucose was the single analyte
difference (P ≤ .05) between the POCT and the control was not measured for cerebrospinal fluid. It did demonstrate statistical
significant with P values of .23 and .17. Glucose, bilirubin, and difference with a P < .01 but notably had a positive bias with a
TG had correlation values of 1.00, 1.00, and 0.98. Albumin mean of 6 mg/dL and a strong correlation (r = 0.97).
did not demonstrate a statistical difference, and only 3 of 16
samples were detectable on the testing modalities.

Figure 2. Correlation for pH. X axis is control measurement of pH. Y axis is POCT measurement of pH

Table 2. Peritoneal Fluid Results


n Mean Std Mean P-value Precision Correlation Concordance Data Range
95% CI 95% CI (r) (k)
LDH (U/L) 11 -301 398 -598/- 33.5 .03 -1114/466 0.88 0.11 1829
Protein (g/dL) 16 0.09 0.30 -0.07/0.25 .23 -0.50/0.69 0.92 1.00 2.6
Albumin (g/dL) 16 0.20 0.30 0.04/0.36 .02 -0.40/0.80 0.90 0.46 1.4
Bilirubin (mg/dL) 16 0.23 0.25 0.09/0.36 < .01 -0.27/0.72 1.00 -0.12 9.0
Glucose (mg/dL) 20 19 24 8/30 < .01 -28/65 1.00 0.64 598
TG (mg/dL) 12 18.5 43 -9/46 .17 -66/103 0.98 1.00 190

Table 3. Cerebrospinal Fluid Results


n Mean Std Mean P-value Precision Correlation Concordance Data Range
95% CI 95% CI (r) (k)
Glucose (mg/dL) 55 6 4 5/7 < .01 -1/14 0.97 1.00 53

HAWAI‘I JOURNAL OF MEDICINE & PUBLIC HEALTH, SEPTEMBER 2014, VOL 73, NO 9, SUPPLEMENT 1
5
Discussion Protein had a positive bias at 0.01g/dL with no significant
Pleural Fluid statistical difference (P = .96). Five of the 17 samples were < 2.0g/
Prior studies have demonstrated utility for pH measurement in dL on the POCT and control. The other 12 were detectable on
pleural fluid with POCT.4, 5 In this study, no significant statistical both devices. Protein is another clinically important analyte in
difference was observed (P = .11) with slightly positive mean distinguishing exudate from a transudate with Light’s criteria,
bias. POCT testing could be used at bedside for diagnostic but a serum sample is needed to calculate a ratio.12,16 Future
purposes of diagnosing a parapneumonic effusion which can studies should investigate if the POCT device can categorize
have a pH < 7.2.8 All of the other samples had a pH > 7.4 with the pleural fluid as transudative or exudative.
10 yielding a pH > 8.2 on both devices. Alkaloid sampling might TG had no significant statistical difference (P = .08) with a
be from the delay in measuring pH since some of the samples slightly negative bias (2.7mg/dL). The highest value measured
were in the core laboratory freezer for a week. pH can become was 60 mg/dL on the control. TG are routinely measured in
more alkaloid if there is a delay > 4hrs between obtaining the pleural fluid if a chylothorax is suspected and confirmed with
pleural fluid and analysis.9,10 A future improvement on this TG level > 110 mg/dL.15 The value range needs to be expanded
study would be to obtain a timely bedside measurement of to assess whether the POCT is clinically useful for diagnosing
pH versus a bench top control and assess the accuracy and the a chylothorax. The low concordance was due to the control
ability of the POCT device to correctly classify the effusion detecting TG levels below 20, while POCT could not.
as a possible complicated parapneumonic effusion based on a Albumin levels were analyzed because they have been used in
pH < 7.2 given the correct clinical scenario. A timely specimen studies to categorize transudative versus exudative effusions.16
collection and analysis would be classified as < 2hrs since this Albumin demonstrated a statistical difference observed with a
should not affect the pleural fluid pH.11 (P-value < .01) with a positive mean value at 0.27mg/dL with a
Glucose demonstrated a statistical difference (P < .01) and low concordance value leading to the conclusion that albumin
a bias of 11mg/dL on the POCT when compared bench top has limited if no potential future use on POCT.
analysis. The clinical utility of glucose is when it is ≤ 60mg/dL, Cholesterol had a negative bias at 16 mg/dL with statistically
usually in the setting of a parapneumonic effusion or malignant significant difference (P = .02), however, only 6 samples were
effusion.16 Glucose measurements on POCT correctly identi- measurable on the control. Cholesterol can categorize transuda-
fied 6 of 7 effusions with glucose ≤ 60 mg/dL. It would be 7 of tive and exudative pleural effusions. An exudative effusion has
7 but one sample on POCT would be calculated at 62mg/dL cholesterol level > 60 mg/dL for higher specificity or > 43 mg/
while on the control analysis it would be 60mg/dL. Glucose has dL for higher sensitivity but neither is more sensitive nor more
demonstrated the potential to be used for clinical assessment specific than Light’s criteria.16 POCT correctly identified 2 of
at the bedside in a future study. 4 exudates with a cholesterol cutoff > 60mg/dL. If the cutoff is
LDH had statistically different values (P-value < .01) when dropped to a cholesterol level > 43 mg/dl then 6 of 6 exudates
comparing testing modalities with large negative mean bias are identified. Future testing should be conducted to assess the
(-343U/L); however, it had a strong r at 0.97. LDH had statisti- accuracy of cholesterol measurement on POCT due to the low
cal differences in measurements on the POCT compared to the number of measurable samples.
control due to measurement of different reaction products on
each separate device. LDH is reported in international units on Peritoneal Fluid
both devices but has different references ranges with a 99-192 LDH had 11 samples with a wide data range; however, it did
U/L on the POCT and 313-618 U/L on the control. The control not have a strong correlation. This test used the same cartridge
uses LDH as a catalyst for the oxidation of NADH to NAD+ as pleural fluid on the POCT device, and there was a large
monitored by reflectance spectrophotometry, and the activity of difference in values due to different analyte measurements on
LDH corresponds to the LDH concentration.13 The POCT test each device.13,14 This yielded a statistical difference between
for LDH was the Piccolo Basic Cartridge which uses the same the values with a weaker correlation than pleural fluid. LDH
reaction as above but carries out a second reaction measuring measurement would not be particularly useful for peritoneal
the formation of formazan by spectrophotometry.14 LDH is fluid in diagnosing spontaneous bacterial peritonitis or second-
utilized diagnostically to characterize pleural fluid into exuda- ary bacterial peritonitis. A future study should be repeated with
tive or transudative classes based on Light’s criteria.12 If using more samples.17
Light’s Criteria with LDH being 2/3 the upper limit of normal Protein was measured on 16 samples but only detectable on
an exudate on the POCT device will be a value ≥ 161 U/L while 5 samples with a positive bias and no statistical significance.
on the control the value will be ≥ 516 U/L. The POCT identifies Since 11 of the samples read less than 2.0g/dL on both testing
7 exudates and the control identifies 6. It would have identified modalities and 5 detected samples > 2.0 g/dL the concordance
the same seven but one of the value was only 503 U/L while was 1.00. Only 5 samples had elevated protein levels so further
165 U/I on the POCT which is borderline exudative/transuda- testing should be conducted to assess potential use for protein
tive on both devices. LDH demonstrated potential for clinical measurement. If a sample is read as less than 2.0g/dL of protein
use on POCT testing but further testing should be conducted on POCT testing it is most likely correct. It has demonstrated
to broaden the range of values and increase the sample size. utility for use at bedside if it is negative or < 2.0g/dl but if the

HAWAI‘I JOURNAL OF MEDICINE & PUBLIC HEALTH, SEPTEMBER 2014, VOL 73, NO 9, SUPPLEMENT 1
6
level is positive there is not data to clinically utilize this testing negative samples in agreement in both testing modalities, mean
modality. bias goes from 18.5 mg/dL to 6.2 mg/dl with an STD of 7.2
Albumin demonstrated a statistically significant difference mg/dL (P = .017). It seems the testing could be used for lower
between the POCT device and the control with a mean bias of TG levels but it remains uncertain of its clinical utility with
positive 0.2 mg/dl due to POCT reporting elevated albumin levels > 109 mg/dL since there was only 1 sample which was
levels compared to the control. Of the 16 samples analyzed elevated.
only3 had a detectable albumin levels on the control and POCT
device; however, the POCT device had 3 other samples with Cerebrospinal Fluid
falsely high elevations when compared to the control. Albumin 55 samples demonstrated a mean positive bias of 6mg/dl with
is diagnostically utilized to calculate a serum-ascites albumin statistical significance (P = <.01) but it may not be clinically
gradient to aid in diagnosing portal hypertension.20, 21 A future relevant. The POCT device ranged higher on every sample
study could assess a serum-ascites albumin gradient calculated except for one in which the POCT measured a low CSF glucose
from a control versus the POCT testing to account for the posi- at 20 mg/dL with a control of 22 mg/dL. Both values still fall
tive bias of the device but currently the POCT device should into the low range. Normal values of blood glucose typically
not be used to measure albumin in ascitic fluid. range from 40-80 mg/dL or a CSF to serum glucose ratio of
Glucose demonstrated a large mean bias of 19mg/dL with 0.6. Bacterial meningitis typically causes a ratio of CSF/serum
statistical significance (P <  .01). However, there were 2 samples glucose of < 0.4 .24-26 Forty-eight of the samples were within
with larger values recorded on POCT at > 700 mg/dL and 699 the range of 40-80 mg/dL and three samples were > 80 mg/dL.
mg/dl while control recorded values of 618 mg/dl and 611 mg/ Four samples were less than 40 mg/dL. Future studies should
dL. These two results may have skewed the results and elevated sample glucose at bedside along with serum glucose on POCT
the positive mean bias. The overall r was 1.00 demonstrating a and control devices. Further investigation needs to be conducted
relationship between testing modalities. Glucose is measured in values outside of 40-80mg/dL.
in ascites typically if there is a concern for secondary bacterial
peritonitis with a level of < 50 mg/dL in this setting.22 Only 3 Conclusion
of the samples from the study had a reading on < 50 mg/dL Pleural fluid protein, pH, glucose, and TG, and peritoneal fluid
and another 3 samples had a value of 50-70 mg/dL with the protein, bilirubin, glucose, and TG correlated and had similar
POCT and control agreeing for each set of 3 samples. Further clinical results as the control, demonstrating potential for clini-
testing should be conducted to assess the accuracy of glucose cal use. Pleural fluid LDH had a strong correlation between
since it may be able to be utilized for aiding in the diagnosis the fixed platform and the POCT device; however, differences
of secondary bacterial peritonitis. were noted between reference ranges and methodology. Further
Bilirubin demonstrated a slightly positive mean on the 16 studies will need to be conducted to confirm the utility of LDH
samples but only 3 of these samples had elevated bilirubin levels POCT. Cholesterol POCT requires further investigation, CSF
at > 1 mg/dl. The r was excellent at 1.00 but concordance was glucose POCT results were promising, and future studies should
poor due to the slightly positive bias on the POCT testing which study more samples with glucose < 40 mg/dL.
lead to detectable tests when the control read a negative result
on 7 samples. Bilirubin is tested in ascitic fluid to investigate Conflict of Interest
a possible choleperitoneum from a gallbladder perforation None of the authors identify a conflict of interest.
which typically causes a bilirubin concentration greater than 6
mg/dL with an ascitic fluid/serum bilirubin ratio > 1.23 Only 1 Acknowledgements
sample in the study had a bilirubin > 6 mg/dL and at 8.7 mg/dL Raymond Manalo and Michael Lustik for their contributions
and 9.1 mg/dL on the POCT and control, respectively. POCT to this project.
did demonstrate utility and even with a slightly positive mean
bias it was able to detect elevated bilirubin levels. It should be Authors’ Affiliations:
- Department of Medicine, Tripler Army Medical Center, Honolulu, HI (WL)
assessed in further studies with a bedside assessment versus a - Pulmonary and Critical Care Service, Naval Medical Center San Diego, San Diego,
control because it could impact patient care immediately. CA (KD)
TG testing demonstrated no statistical significance in the - Pulmonary and Critical Care Service, Tripler Army Medical Center, Honolulu, HI
(DH)
results between the two testing modalities but there was positive - Department of Pathology, Tripler Army Medical Center, Honolulu, HI (JA)
mean bias which is greatly affected by one sample where the
POCT recorded 373 mg/dL and the control recorded 219 mg/ Correspondence to: William Londeree MD; Tripler Army Medical Center,
dL. If this measurement is excluded since all the other mea- 1 Jarrett White Road, Honolulu, HI, 96859;
Ph: (561) 376-3515; Email: wlonderee@gmail.com
surements were less than a 109 mg/dL on both devices with 3

HAWAI‘I JOURNAL OF MEDICINE & PUBLIC HEALTH, SEPTEMBER 2014, VOL 73, NO 9, SUPPLEMENT 1
7
References 12. Light RW, MacGregor MI, Luchsinger PC, et al. Pleural effusions: the diagnostic separation of
1. McCoy D, Chand S, Sridhar. Global health funding: how much, where it comes from and where transudates and exudates. Ann Intern Med 1972;77:507–14
it goes. Health Policy Plan 2009;24:407-17. 13. VITROS Chemistry Products LDH Slides Lactate Dehydrogenase. Ref 838 4489. Ortho-Clinical
2. Jani I, Peter T. How Point-of-Care Testing Could Drive Innovation in Global Health. New England Diagnostics, Inc. Rochester, NY 14626
Journal of Medicine. 2013;368(24):2319-2324. 14. Piccolo® Basic Metabolic Panel Plus Disc July 2011 PN : 400-7152-1 Rev: G © 2005, Abaxis,
3. Nichols JH, Christenson RH, Clarke W, Gronowski A, Hammett-Stabler CA, Jacobs E, Ka- Inc., Union City, CA 94587
zmeirczak S, Lewandrowski K, Price C, Sacks DB, Sautter RL, Shipp G, Sokoll L, Watson 15. Huggins JT. Chylothorax and cholesterol pleural effusion. Semin Respir Crit Care Med. 2010
ID, Winter W, Zucker ML. National Academy of Clinical Biochemistry Laboratory Medicine Dec;31(6):743-50.
Practice Guideline: evidence-based practice for point-of-care testing . Clin Chim Acta. 2007 16. Light, RW. Pleural Effusion. N Engl J Med 2002; 346:1971-1977June 20, 2002
Apr;379(1-2):14-28. 17. Press AG, Meyer zum Büschenfelde KH, Ramadori G. Spontaneous bacterial peritonitis.
4. Kohn GL, Hardie WD. Measuring Pleural Fluid pH: High Correlation of a Handheld Unit to a Gastroenterol. 1992 Aug;30(8):543-52.
Traditional Tabletop Blood Gas Analyzer. Gary L. Kohn and William D. Hardie Chest. 2000; 18. Runyon BA. Low-protein-concentration ascitic fluid is predisposed to spontaneous bacterial
118; 1626-1629. peritonitis. Gastroenterology 1986; 91:1343.
5. Wockenfus AM, Koch CD, Karon BS. Analytical performance of three point of care methods 19. Runyon BA. Patients with deficient ascitic fluid opsonic activity are predisposed to spontaneous
for pleural fluid pH analysis. Clin Biochem. 2013 Aug;46(12):1139-41. bacterial peritonitis. Hepatology 1988; 8:632.
6. Dechant JE, Symm WA, Nieto JE. Comparison of pH, Lactate, and Glucose Analysis of Equine 20. Runyon BA, Montano AA, Akriviadis EA, et al. The serum-ascites albumin gradient is superior
Synovial Fluid using a Portable Clinical Analyzer with a Bench-Top Blood Gas Analyzer. Veterinary to the exudate-transudate concept in the differential diagnosis of ascites. Ann Intern Med. 1992;
Surgery 2011;Jul19:1-6. 117:215.
7. Bland MJ, Altman DG. Statistical methods for assessing agreement between two methods of 21. Wong CL, Holroyd-Leduc J, Thorpe K, Straus S. Does This Patient Have Bacterial Peritonitis
clinical measurement. The Lancet 327, no. 8476 (1986):307-310. or Portal Hypertension? How Do I Perform a Paracentesis and Analyze the Results? JAMA.
8. Colice GL, Curtis A, Deslauriers J, Heffner J, Light R, Littenberg B, Sahn S, Weinstein RA, 2008;299(10):1166-1178.
Yusen RD. Medical and surgical treatment of parapneumonic effusions : an evidence-based 22. Soriano G, Castellote J, Alvarez C, et al. Secondary bacterial peritonitis in cirrhosis: a retro-
guideline. Chest. 2000 Oct;118(4):1158-71. spective study of clinical and analytical characteristics, diagnosis and management. J Hepatol
9. Mishra EK, Rahman NM. Factors influencing the measurement of pleural fluid pH. Curr Opin 2010; 52:39.
Pulm Med. 2009 Jul;15(4):353-7. 23. Runyon BA. Ascitic fluid bilirubin concentration as a key to choleperitoneum. J Clin Gastroenterol
10. Rahman NM, Mishra EK, Davies HE, Davies RJ, Lee YC. Clinically important factors influencing 1987; 9:543.
the diagnostic measurement of pleural fluid pH and glucose. Am J Respir Crit Care Med. 2008 24. Agamanolis, D. THE NORMAL CSF - Cerebrospinal Fluid. Neuropathology. Northeast Ohio
Sep 1;178(5):483-90. Medical University. Chapter 14. Jan 2013
11. Haro-Estarriol M, Baldó-Padró X, Lora-Díez M, Rubio-Garay M, Rubio-Goday M, Sebastián- 25. Fishman RA. Studies of the Transport of Sugars between Blood and Cerebrospinal Fluid in
Quetglás F. Changes in the acid-base equilibrium of pleural fluid during the first 2 hours after the Normal States and in Meningeal Carcinomatosis. Trans Am Neurol Assoc. 1963;88:114
thoracentesis. Arch Bronconeumol. 2005 Nov;41(11):612-7. 26. van de Beek D, de Gans J, Tunkel AR, Wijdicks EF. Community-acquired bacterial meningitis
in adults. N Engl J Med. 2006 Jan 5;354(1):44-53.

HAWAI‘I JOURNAL OF MEDICINE & PUBLIC HEALTH, SEPTEMBER 2014, VOL 73, NO 9, SUPPLEMENT 1
8

You might also like