You are on page 1of 8

FEATURE

An Introduction to Risk Analysis


for Medical Devices
By Fredric Mattsson, SEMKO, +46 8 750 03 98

Medical Devices: Risk mentary methods. To ensure that all

L
ike all other "Ne w
Approach" directives, risks relating to a particular product
Analysis-prEN 1441 have been considered, several differ­
the Medical Devices
Directive (93/42/EEC), ent techniques may have to be used in
The European Committee for
hereafter referred to as parallel. In such instances, some kind
Standardization (CEN), under the
MDD, calls for the products it covers of compilation will be needed, along
mandate of the EC Commission, has
to meet a number of essential require­ with a cross-reference document that
worked to develop a standard for risk
ments. They are formulated in very will allow the analysis of the various
analysis. Still a prelirninary European
general terms and specify no Iimit val­ risks to be traced backward.
standard as of this writing, it bears the
ues or test methods; it is the manufac­ designation prEN 1441 and is entitled Step 1
turer's responsibility to ensure that "Medical Devices: Risk Analysis."
th�se essential requirements are met. prEN 1441 requires that the method
prEN 1441 uses and defines a num­
and results of the risk-analysis proc-
ber of important terms, some of which
Why Risk Analysis? are derived from ISOIIEC Guide 51
guidelines for the inclusion of safety
The MDD explicitly states that the aspects in standards. For example:
results of any risk analysis conducted
during the design stage must be docu­ Harm: Physical injury and/or darn­
ts age to health or property.
mented and included in the technical
file. This statement may be found in Hazard: A potential source of
Annex II (Full Quality Assurance harm.
System), Annex 111 (EC Type Risk: The probable rate of occur­
�Examination), and Annex VII (EC rence of a hazard causing harm and
Declaration of Conformity). the degree of severity of the harm.
Where no harmonized standards Risk analysis: The investigation of
exist, risk analysis is an important available information to identify haz­
means of ascertaining that all risks ards and to estimate risks.
relating to the product in question Risk assessment: An analysis and
have been considered. lt additionally evaluation of identified risks.
serves to outline the procedures Safety: Freedom from unacceptable
applied to reduce any possible risks. risk of harm. [The notion of "accept­
For those products that are covered able" risk of harm may seem strange
by harmonized standards, risk analy­ to many people.]
sis can have other benefits. Few stan­
The flow chart shown in Figure 1 is
dards (if any) can cover all essential
presented to facilitate an overview of
requirements for a particular product,
the standard. The various steps are
and most tend to lag behind technical
examined in some detail below.
developments (generally the case in
In addition to the r isk-analysis
all standardization work).
methods explained and discussed in
the standard itself, this article will FIGURE 1: Flow chart according to
briefly address two further, comple- prEN 1441.

Compliance Engineering 47
' :'''' A,,alysis for Medical Devices

dure be documented. The documenta­ to reduce the risk to acceptable Iev­ Iimitation could mean, for example,
·=,n must include: els, where considered necessary. that the risk analysis is restricted to

An unambiguous definition of the apply only to part of a larger medical­


• A Iist of names of the persons who
product or the accessory as weil as device system, while a system defini­
carried out the risk analysis.
a complete description of the prod­ tion (or product definition, where the
uct and its characteristics. risk analysis refers to a product) must
Step 2
• A Iist of identified risks. include a description of its character­
Step 2 calls for both a system Iimita­
istics, chiefly those parameters relat­
• A description of any measures taken tion and a system definition. A system
ing to safety. The standard provides a
Iist of questions that may be used as
an aide in verifying that no informa­
tion has been forgotten. Among these
questions (which number about 20 in
all) are the following:
• What is the intended use?

• Is the device intended to contact


the patient or other persons?
• Is energy or [any other substance]
delivered to or extracted from the
patient?

• Is the product sterile?

• Is the device interpretative?

• Does the device contain software?


Several of these questions require
further explication.

What is the intended use?


Defining the intended use of a prod­
uct is very important. "Intended use"
refers, according to the MDD, to the
use for which a product is intended, as
stated by the manufacturer in the
product's marking, directions for use,
andlor advertising material.
Intended use is the starting point in
identifying which essential require­
ments will apply to a particular prod­
uct. The major factors to be consid­
ered are the user' s competence and
training, environmental influences,
whether or not the patient can adjust
the device, and any special needs of
weaker patient groups (such as elderly

�-
and handicapperl persons). Through
restrictions, the intended use can also
serve as an instrument for risk reduc­
800.467.no5 Voice: 516.467.8400
tion.

J
731 Union Parkway • FAX: 516.467.8558
Ronkonkoma, Long l sland
New York 11n9-7414
r.J Internet address: ifi@ifi.com
America Online: IFI net
ls the device intended to contact the
patient or other persons?
Is the product i nvasive-that is,
Aostralla: 61-2-415-3944 Canado: 90-613-226-2365 Eaglalld: 44-0-462-421234 Fnace: 33-1-M-46-21-10
Ge1'11W17: 49�109-2788 Hoa1 Kaag: 852-763-5123 ladia: 91-842-847924 Israel: 972-3-7526333 lt.aly: 39-11-4551388 does it penetrate into the body, either
Japaa: 81-03-3435-4814 Korea: 82-02-501-4271 Netherlands: 3145206-41214 New Zealaad: 64-4-237-8392
SiDgapore: 6.5-7477234 Spaln: 34-1-3589048 Swedea: 46-8-930280 Tlilwu: 886-2-6864758 via an orifice or through the skin? For
how long is the device, or part of it, in
Reader Service #42
48 Compliance Engineering NoVEMBERIDECEMBER1995
Risk Analysis for Medical Devices

We've Bot Your


PowerMains
contact with the patient? The MDD sure that the intended use specified at
specifies three different durations: the initial stage has been achieved.

transient (continuous use for less

�rotected!

Step3
than 60 minutes);
In Step 3, all hazards associated
• short-term (continuous use for not
with the device must be identified and
more than 30 days); and
recorded. The goal at this stage is to
• long-term (continuous use for more draw up to a rough Iist of risks that
than 30 days). may then be gradually refined. Such a
Finally, is the device intended for structured document will serve t o
implantation? facilitate the identification o f different
hazards and decrease the risk of inad­
Does the device contain software? vertent omission of significant haz­
A proposal is currently being devel­
ards. The standard divides possible
oped for a collateral standard govem­
hazards into five different categories:
ing validation of software for medical
devices. This standard carries the des­
1. Hazards related to some kind of
ignation IEC 601-1-4 and is entitled
energy
"Medical Electrical Equipment, Part
2. Biological hazards
1: General Requirements for Safety, 4.
3. Environmental hazards
Collateral Standard: Safety
4. Hazards related to the use of the
Requirements for Programmahle
device
Electronic Medical Systems." The
5. Hazards arising from functional
proposal at present enjoys CDV
failure, maintenance, or aging
(Committee Draft for Vote) status.
In many respects, this standard dif­ By complying with the requirements
fers from its predecessors. Applying of the IEC 601 series, manufacturers
�" '\449 Lts'f:

��<)*
pass/fail criteria to the final product, of many (though not all) electrica1

....... �.
as per the usual test procedure, could medical devices may be able to verify
cause critical safety faults to be over­ that their products do not constitute a
looked; to avoid that possibility, IEC hazard within certain specific areas.
...,. Fast <25ns - Full Leads 601-1-4 requires t h e preparation, For devices covered by IEC 601-1,
before the development phase even compliance with the requirements
...,. Breakered lnterna/ly
begins, of documents strictly control­ specified in Section 3, "Protection
...,. Loca/ and Remote Status ling the device's design and develop­ against Electric Shock Hazards," and
...,. Independent of User Loads ment. Section 9, "Abnormal Operation and
The standard comprises two basic Fault Conditions," can demoostrate
...,. Surges Shunted to
sections, one dealing with risk analy­ that the device itself does not consti­
Entrance Ground
sis and the other with the "life-cycle" tute an electric hazard.
...,. *NRTL 1449 Approved concept. The risk-analysis part repris­ Fu rther, by complying w i t h the "'
p
...,. 10- Year Warranty es the prEN 1441 flow chart shown in requirements contained in the collater­
al standard for electromagnetic com­
d
Figure 1, and consequently needs no
VI
2500 models of coax, further elaboration in this context. The patibility, IEC 601-1-2, manufacturers
p
life-cycle concept is concemed with can also show that their products nei­ tt
power and twistedpair
the division into discrete phases of the ther influence nor are influenced by d
protectors ... plus their environment.
e n tire life of a p rod uct, from the D
lightning/EMP and preparation of the frrst specifications In developing new medical devices
grounding solutions to the product's entry into the market. (perhaps within new spheres), manu­ 1:
facturers may need to rely on scientific
V
As each phase is completed, the man­

lfl'olyraf10"A�!�
a
ufacturer must verify that the require­ data, experience with equivalent kinds
c
ments set up prior to the beginning of of products, or clinical evaluation.
1
the phase have been fulfilled; after the I
(800) 325·7170. (702) 782-2511 Steps 4 and 5
last phase, the manufacturer must val­
FAX: (702) 782-4476 In Step 4, the risk of each identified
idate the system and check to make
25 Park Place • P.O. Box 9000 hazard must be estimated. "Risk" is
Minden. NV 89423-9000

Reader Service #63 NOVEMBERnDECBMBER1995


Risk .�nalysis for Medical Devices

defined here as both the likelihood of An error in any element is called


;· ')Ccurrence and the gravity of its failure mode, and the effect of that
Frequent

.equences. error-that is, what the error means to

gi
An easy way of illustrating this is to elements on higher levels-is called
Occasional
use a graph as shown in Figure 2. The failure effect. A failure effec t on a
.s::;
X-axis corresponds to the conse­ � lower Ievel thus corresponds to a fail­
quence of the harm and the Y-axis to ure mode on a higher Ievel.
::J

its probability. Demarealion of three lncrecllble Starting from the lowest level-for
different areas allows the hazard to be example, the component level-the
J l 1 I
classified as acceptable (i.e., falling Minor Moderale Major Cataslrophic analysis works its way through the
w i t h i n the area corresponding t o system by asking at each new Ievel,
minor probability and minor conse­ . "What happens if . . . ?" The failure
quence), unacceptable (great probabil­ effect is then recorded for all compo­
FIGURE 2: Risk gra ph showing risk
i t y a n d serious conseque n c e ) , or nents and possible failure modes.
regions.
ambiguous, necessitating a further FMEA is an efficient means of ana­
estimation o f whether or not risk lyzing failure modes in elements that
reduction is necessary (this area is affect the performance characteristics
number of systems involved depends
usually called ALARP, for "as low as of a whole system, though it can get
not only on the complexity of the sys­
reasonably possible"). very complex as the number of func­
tem but also on how detailed an analy­
Because some uncertainty is tions and components increases. For
sis is required. The lowest subsystem
inevitable in estimating not only prob­ cer tain s t r u c t u r e s , i t m a y a l s o b e
Ievel may be equivalent to component
ability but also degree of harm, the essential t o determine exactly when
Ievel or lowest spare-part Ievel, or it
risk graph will always contain an the failure mode occurs; different fail­
may be somewhat higher.
ambiguous zone. This area can be ure modes may be obtained at differ­
lf the analysis is to be feasible, each
reduced, however, as more informa­ ent times and consequently may have
element in the system must be clearly
tion is received-for example, from different effects on the total system.
defined as to its characteristics, func­
f<>1Jlt reports. A risk analysis must be A useful extension of the F M EA
tion, and connection with other ele­
ated continually and must never be method is failure modes, effects, and
ments. Any redundancy (information
static. criticality analysis, or FMECA, which
that is not strictly necessary but can
There are a number of different risk­ additionally calls for an investigation
help to detect and correct errors) must
analysis me thods, most emanating of b o t h the consequences a n d the -.

also be shown. Very often block dia­


from some specific application in the probability of a failure mode.
grams are used to show the structure
industrial sector. Five methods are FMEA makes it possible to delin­
of a system, with the various blocks
described below. Two o f these, eate the relationship between failure
(elements) connected by straight lines
FMEA and FTA, have been published modes in basic individual elements
representing input and output. For a
in IEC documents and are referred to and the total system. It is an essential
medical device, a number of different
in the standard. A third, HAZOP, has tool, but it is often insufficient for
structures may be necessary to cover
not been published in any IEC docu­ analyzing complex systems. In the ini­
mul tiple functions and functional
ment but is both referred to and dis­ tial d e v e l o p m e n t p h a s e of a n e w
modes.
cussed in the standard as weil as in device, when little more than the func­
other extensive literature. The final tion and a general system structure
two methods, A E A and E T A , are have been established, a complemen­
never mentioned in the standard, but a Deviation tary analysis method, such as fault­
great deal of Iiterature exists about tree analysis, or FTA, will likely be
them. Figure 3 illustrates some of the FMEA Failura FMEA
'C required (see below).
mode •
differences and similarities among the c A worksheet like the one shown in

ETA FTA &.
various methods. ()
c Unwan1ed E Figure 4 may be used to document an
• event 0
::s u
c:r FME(C)A analysis.

.. HAZOP e
Failure-Mode and Effects Analysis c Construcled .!
0 faiJure ..
u >- Fault-Tree Analysis (IEC 1025)
(IEC 812) ..
Sl
Failure-mode and effects analysis AEA Critical
::s
(/) A fault-tree analysis ( F T A) is a
(FMEA) builds on the principle of
action
structured graphic presentation in the
J.. �eaking down a system-for exam­ shape of a tree of circumstances con­
. , a medical device-into a series of tributing to or causing a defined event,
subsystems or elements. The precise
FIGURE 3: Comparison of risk­ called a top even t. Thi s an aly sis
analysis methods.

52 Compliance Engineering NOVEMBERnDECEMBER1995


Risk Analysis for Medical Devices Ri

FME(C)A
method is particularly weil suited to
r�"lplex systems comprising a num­ System!Equipment: _______________ Date: ------
Jf subsystems with different func­ Person(s) who carried out the analysis: ------- Page: (

tional purposes.
The top event, which serves as the d'
starting point for the whole analysis, tl
corresponds to an unwanted result e

such as a deterioration of the perfor­ 1


mance characteristics of the system or tJ
a change in function entailing a haz­ a
ard to a patient, a user, or the environ­ i·
ment. E a c h d iffer e n t top e v e n t
requires its own specific tree. f
An FfA that does not describe the
right top events is, of course, useless. FIGURE 4: Worksheet for use in documenting an FME(C)A analysis.
The graphic presentation of an FfA
may employ a number of different
same event is found in more than one most common gate types are AND,
symbols. The standard calls out some
place, whether in the same tree or in OR, and Exclusive- O R , defined as
15 of these, but in some cases the
different trees, the same designation follows:
analysis may be made easier if new,
should be used. The boxes are then o AND: The event out occurs only if
case-specific symbols are defined.
linked together with the aid of logical all input events occur.
In an FTA, each event corresponds
gates. Each gate may have an unlimit­
to a box. It is very important that each o OR: The event out occurs if any
ed number of inputs but may corre­
box be given a precise designation so input event occurs, either by itself
spond to only a single output. The
that it can be readily identified. If the or in combination with other events.

INSTANT
A DATA BY •••

f:l\Okaya Electric America lnc. �:: aF���;��-8561


� 1509001
�=-�������������=-����

--:.�
. � CER TI F IE D
------��
�---�
�· -'��----��----------�
Jt
AC NOISE SUPPRESSION CAPACITORS 'M@@@(S)@®®<®@

XYE
DELTA NETWORK

e -40'C to +lOO'C Operoffng Temp. Range e Ale suppresslng R+C network e Best pertormonce serleS tn most e INiatled cost sovlngs wHh slngte pockage
e CapacHance votues from O.OlUFO e UL. CSA. VDE approved populor apptlcotlons deslgn and smatl PCB rootprlnl
to 2.2 UFO ot 250 VAC 50/60Hz e 250 VAC roted e Drop ln replocement for most e 8oth normal ! common mode ottenuotlon
e New X2 closs copocHor serles ls pelfect lor e AC/DC apptlcotlons populor conllgurotlons uslng one X ! two V nolse supp. networks
high votume. tow cost appllcotlon s e PC8 mount or nex lnsuloled Ieads e Dtelectrlc wHhstond voHoge e Deslgned wHh seH-heaßng metatltzed potyester
e Cose material and Internot polltng e Hfgh pulse/non-lnductlve twlce sotety agency requlrennents ntm, double wound constructlon and oll
are roted UL 94-VO reslstor used e High DV/DT surge reslslonce and 1/11 rotlngs tnnpregnoted lor high relabUity
e Sotety ogency approvols lnclude, e Fo;xbock Documem. 1102 e Xl and V closs roted e Meels elghl sotety agency requlrements
UL. CSA. VDE and SEV e Double wound constructlon melalllzed ot 250 VAC 50/60Hz
e Fo;xbock DocvrMnlt 1101 polyester11m 1or high DV/DT • The XYE serles conslsts ol75 models ln
e Faxbock DocumenN 110J three dlsllnct pockeging stytes
e Faxbock DocvmenN 1106

For a Complete Nolse Suppression Or Call/ Fax Us At... ..


Componen1s Product Selector OKAYA ELECTRIC AMERICA1 INC.
Call.... 80�755-8561 503 Wall Street, Valparaiso, Indiane 46383
And Select Document #1001 Ph: 219-477-4488/ Fax: 219-477-4856

Reader Service #58

54 Compliance Engineering NOVEMBER/DECEMBER1995


Risk Analysis for Medical Devices·

• Exclusive-OR: The event out


occurs if one and only one input
event occurs.
Construction of a tree starts with a
determination of what it is that makes
the top event occur; one event or sev­
eral may be necessary to generate it.
These generating events are in turn
treated as top events themselves, with
a corresponding determination of what
it is that makes them occur. Figure 5
shows an example of a highly simpli­
fied tree.
When the tree has been completed,
careful examination is required. Many
systems yield trees that are too large to
survey without the aid of suitable soft­
ware; the upper Iimit for feasible visu­
al analysis is about 70 events. Because
events link:ed to the top event only by
Catch rf emissions problems at
OR-gates will generate the top event
every time they occur, no further
board Ievel, where compliance
analysis is required for a tree that con­
sists solely of OR-gates. If other gate fixes are least costly.
types exist, it means that the system

N ow you can quickly get a color intensive spatial examination.


contains a redundancy-that is, it is
image of the electromagnetic After the development stage,
error-tolerant.
performance of your printed-circuit you can use EMSCAN as a quality­
Certain events may be found in sev­ board or subassembly before final control tool, checking completed
eral different parts of a tree. Estab­ compliance testing. Spatial and boards against a "good" scan before
lishing that such events do not remove spectral displays generated by the they go into assembly. This is the
the redundancy in the system requires EMSCAN PCB emissions scanner point where production compliance
recourse to B o o l e a n reduction o r show you which frequencies and becomes virtually assured.
Boolean algebra. which areas of the board under test The software operates under
When the tree has been reduced, the are guilty. These scans are stored "Windows" to make early diagnosis
for later comparison after design easy, even for those who are new to
so-called minimal cut set-corre­
alterations, to check whether offend­ compliance testing. It can run on
sponding to the smallest group of
ing emissions are now down to several PCs and workstations, and
events required for the top event to
acceptable levels. is readily ported to other environ­
occur-may be isolated.
Just plug your receiver or spec­ ments for analysis.
trum analyzer, and your computer You should learn all about this
Hazard and Operability Study
with IEEE-488 interface, into the qualitative and quantitative measure
(HAZOP)
EMSCAN scanner, and a matrix of of emissions for use during product
The HAZOP method was originally
1280 H-field probes maps the area development-where design correc­
developed to study the effects of vari­ tions are least costly. To start, call
of your test board (up to 9" x 12") for
ous deviations or disturbances on a high, medium, and low-emissions toll-free (1-800-933-8181) to speak
factory production line by following spots within the 10 MHz-1.5 GHz with an applications engineer and
the natural flow through the line. The frequency range. Or you can see a arrange to see·
same principle could also be used to spectral display showing the overall a demonstra­
follow a natural flow other than the condition of the board across the tion in your
one in a manufacturing process-for spectrum. You may then choose a office or
example, the route taken by blood in frequency of particular interest for plant.

hemodialysis equipment.
According to this method, faults or 160 School House Road
disturbances are fabricated without Souderton, PA 18964-9990 USA
215-723-8181 Fax 215-723-5688

regard to their feasibility. Examples of


guide words and possible deviations For engineering assistance, sales, and seruice throughout Europe, call
EMV • Munich, 89-612-8054 • London, 908-566-556 • Paris, 1-64-61-<>3-29 12140

Reader Service #9

I
Compliance Engineering 55
Risk Analysis for Medical Devices Ri
TRANSIEHT SURGE
PROTECTION

FIGURE 5: Event A will occur only if events B and C both


occur. Event C will occur if either Event D or Event E
occurs. The terminating input symbol at Event D indicates
c
that it cannot be subdivided, while the input symbol at ..
Event B indicates that it is further developed on another
tree. The transfer-out symbol at Event B indicates that B r
reappears elsewhere.

include none (flow failure), more (higher pressure), and less


(lower pressure).
HAZOP requires documentation of a !arge nurober of
deviations and disturbances, many of which can be quickly
discounted, with no additional measures necessary. Other,
TECCOR ELECTRONICS, INC. more serious deviations and disturbances may be processed
1801 Hurd Drive PH: 214-580·1515 further, most likely using a supplementary risk-analysis
lrving, Texas 75038 FAX: 214-550-1309
' method.
Reader Service #84 One advantage of HAZOP is its unbiased approach, which
tends to reveal faults that might otherwise be missed. Then,
too, unlike FMEA and others, HAZOP is a time-efficient

�t lntegrity Design & Test Services... method.

Action-Error Analysis
Action-error analysis, or AEA, analyzes interactions
between machine and man. lt can be used to analyze such
things as starting procedures of devices and to answer ques­
tions such as "What will happen if the right thing is done at
the wrong time, the wrong thing is done at the right time, or
nothing is done at all?" Where complex patient connections
are involved, AEA can disclose deficiencies in instructions
or directions.
o A Wlde range of EMIIEMC lest servlces
o CE marklng expertlse Event Tree Analysis
Event-tree analysis, or ETA, is again based on the ques­
o Helpful, frtendly, professlonal slaff
tion "What happens if . . . ?" But whereas an FMEA applies
o Fully accredlted faciJIUes
that question to subsystems or components, an ETA Iooks
o Fully equlpped Iab areas at the unwanted event itself.
o Independent 3rd party lest Iab This method can help determine whether the event needs
o Aggressive pr1clng further analysis-as, for example, a top event in an FfA

o Rapid servlce analysis.

o Convenlent JocaUon

Step 6

IIMifltl
Reader Service #43 If the risk is deemed unacceptable, a risk reduction is

ß Design & Test


r��(508)486�32 mandated. The MDD's essential requirements call for each
,.. of the following measures to be taken in turn:
.4: (508)486-0592

37-7 1rt« Road - l.itiEioo, MA 01460 \T" Services, !nc. • Inherent safe design and construction to be incorporated

56 Compliance Engineering NOVEMBERnDECEMBER1995


'
Risk Analys s for Medical Devices

• Where risks cannot be eliminated, i n fact the risk-analysis concepts • IEC 601-1, "Medical Electrical
adequate protection measures to be should be part of the natural product Equipment, Part 1 : General
implemented, including alarms development from the start. In this Requirements for Safety" (1988)

w ay, structuring the production IEC 601-1-2, "Medical Electrical


• Users to be informed of any resid­ •

process with the aid of a standard may Equipment, Part 1: General


ual risks attributable to shortcom­
be regarded as a boon rather than a Requirements for Safety, 2. Collateral
ings of the protection measures
chore. Standard: Electromagnetic
adopted. (Residual risks are dini­
Compatibility: Requirements and
cally acceptable if the advantages Reference Literature Tests" (1993)
associated with the device out­
prEN 1441, "Medical Devices: Risk • EN 30993-1, "Biological Evaluation of
weigh the possible disadvantages.) •

Analysis" (1994) Medical Devices, Part 1: Guidance on


Selection of Tests" (1994)
Steps 7 and 8 • CD IEC 601-1-4, "Medical Electrical
Equipment, Part 1: General • Copious Iiterature is available on the
Any new hazards introduced by the
Requirements for Safety, 4. Collateral hazard and Operability study (HAZOP)
risk reduction shall be treated in the
Standard: Safety Requirements for method, event-tree analysis (ETA), and
same way as other hazards.
Programmabte Electronic Medical action error analysis (AEA).

Step9 Systems" (1993)


Fredric Mattssan works at
A decision must be made as to the • IEC 812, "Analysis Techniques for
SEMKO's Centre for Medical
System Reliability-Procedure for
intended use of the device, and decision Engineering and Physics in Sweden,
Failure Mode and Effects Analysis"
criteria set up before the analysis is appointed as a Notified Body accord­
(FMEA) (1985)
started, to determine whether or not the ing to the Medical Devices Directive.
risks of the identified hazards for the • IEC 1025, "Fault Tree Analysis" (FTA)
He holds an M.Sc. from Chalmers
product can be considered acceptable. (1990)
University ofTechnology.
Informative Annexes
The standard also contains a nurober
of informative annexes dealing with
such matters as biological hazards and
hazards associated with in vitro diag­
nostic devices. One of these, EN
30993-l, "Biological Evaluation of
Medical Devices, Part 1: Guidance on
Selection of Tests," addresses the
question of which tests should be con­
sidered for which applications. The
need for testing should be reviewed
on a case-by-case basis so that unnec­
essary testing may be avoided.
Since in vitro diagnostic devices do
n ot come into direct contact with
patients, they do not represent a direct
hazard; however, risks of indirect haz­
ard do exist, as for example through
incorrect analysis of test samples. A
thorough evaluation of such risks may
require the use of criteria different
from those employed for medical
devices.

Conclusion
Introducing a standardized proce­
dure for risk analysis may seem at
first to constitute an extra burden, but
Reader Service #26

Compliance Engineering 57

You might also like