You are on page 1of 11

Int. J. Radiation Oncology Biol. Phys., Vol. 59, No. 3, pp.

861– 871, 2004


Copyright © 2004 Elsevier Inc.
Printed in the USA. All rights reserved
0360-3016/04/$–see front matter

doi:10.1016/j.ijrobp.2004.02.043

PHYSICS CONTRIBUTION

DOSIMETRIC EFFECTS WITHIN TARGET AND ORGANS AT RISK OF


INTERFRACTIONAL PATIENT MISPOSITIONING IN LEFT BREAST CANCER
RADIOTHERAPY

GUIDO BARONI, PH.D.,* CRISTINA GARIBALDI, M.S.,† MARCO SCABINI, B.S.,* MARCO RIBOLDI, M.S.,*
GIANPIERO CATALANO, M.D.,‡ GIANPIERO TOSI, M.S.,† ROBERTO ORECCHIA, M.D.,‡§ AND
ANTONIO PEDOTTI, M.S.*
*TBM Lab, Department of Bioengineering, Politecnico di Milano University, Milan, Italy; †Medical Physics Department and

Radiotherapy Division, Istituto Europeo Oncologico-I.R.C.C.S., Milan, Italy; §Istituto di Scienze Radiologiche, Facoltà di
Medicina e Chirurgia, Università degli Studi di Milano, Milan, Italy

Purpose: To investigate the effects of interfraction setup uncertainties on the dose distribution within the clinical
target volume (CTV) and the organs at risk (OAR) of left-sided breast cancer patients undergoing external
radiotherapy.
Methods and Materials: Interfractional setup errors were assessed by measuring surface control points displace-
ments during 89 irradiation sessions in 4 patients, by means of opto-electronic localization. The measured
position deviations were fed back to the treatment planning system for the evaluation of the corresponding
dosimetric effects within CTV and OARs (lung, heart).
Results: Results revealed errors above 5 mm on some of the control points, but corresponding volumetric
variations were on average below 2% for both the CTV within the 95–105% dose range and the OARs receiving
more than 50% and 90% of the prescribed dose. A specific sensitivity to the setup errors was found as a function
of the treatment plan design, leading to isolated cases exhibiting volumetric variations of CTV and OARs
exceeding 2%.
Conclusions: This study confirms the potential increase of treatment quality provided by the systematic patient
position verification and highlights the role of opto-electronic position detection systems for the real-time check
of patient setup errors and the evaluation of the corresponding dosimetric consequences, as a way to achieve
consistent dose delivery. © 2004 Elsevier Inc.

Radiotherapy, Breast cancer, Treatment planning, Dosimetry, Quality assurance.

INTRODUCTION for the irradiation geometrical setup and on the means used
for patient alignment (1, 9). In recent studies, the experi-
Patient setup errors in the irradiation of breast cancer may
mental measurement of patient setup inaccuracies and organ
significantly affect the treatment quality. This applies par-
motion was used for the quantification of their effects on the
ticularly when tissue compensators are used or when inten-
radiation dose distribution (1, 2, 10). Hector et al. (1–3)
sity-modulated beams are generated with the use of mul-
used a method for the three-dimensional (3D) anatomic
tileaf collimators, as a way to reduce dose inhomogeneities
within the target (1). In these cases, patient misalignments structure reconstruction from portal images and quantified
and movements may alter the planned dose distribution and the dosimetric consequences of interfractional patient
potentially lead to complications or cause late radiation movements within the target volume, which they related to
effects. These include edema, skin fibrosis (1–3), pneumo- different irradiation techniques.
nitis (4, 5), and heart diseases; the latter is associated with An alternative approach to the quantitative assessment of
left breast irradiation (6 – 8). patient position repeatability is based on opto-electronic
The frequency and magnitude of patient localization er- localization. A basic feature of these systems is the real-time
rors and movements in breast cancer radiotherapy have been 3D localization of landmarks on the patient, identified by
investigated by means of portal imaging. As a whole, results infrared light-reflecting (passive) or light-emitting (active)
reveal that errors strongly depend on the specific protocol markers. The applicability and reliability of opto-electronic

Reprint requests to: Guido Baroni, Ph.D., Dipartimento di Bio- This work is partly supported by the Associazione Italiana per la
ingegneria, Politecnico di Milano, Piazza Leonardo da Vinci, 32, Ricerca sul Cancro (A.I.R.C.).
I-20133, Milan, Italy. Tel.: (⫹39) 02-2399-3346; Fax: (⫹39) Received Jul 28, 2003, and in revised form Feb 13, 2004.
02-2399-3360; E-mail: baroni@biomed.polimi.it Accepted for publication Feb 18, 2004.

861
862 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004

Fig. 1. (Left panel) Photograph of a breast cancer patient in treatment position with vacuum-formed cushion for
immobilization and supporting device. (Right panel) Three-dimensional graphic representation of the treatment
geometry; letters T, L, and H indicate target, lung, and heart, respectively.

localizers for the real-time detection of patient misalign- scription of 50 Gy at the isocenter delivered in 25 fractions,
ments in radiation therapy was experimentally investigated followed by 10 Gy boost to the tumor bed. The patient was
(11, 12). Their clinical application in extracranial radiother- lying in supine position on an inclined plane, immobilized
apy as patient repositioning control devices was reported for in a personalized, vacuum-formed cushion, with arms ele-
breast (13), prostate (14), and gynecologic cancer irradia- vated (Fig. 1). For each patient, a computed tomography
tion (15). (CT)-based 3D treatment plan (Cadplan, Varian Medical
In the present study, we report the results obtained by the Systems, Palo Alto, CA) was performed, built on about 50
real-time monitoring of patient position during the treatment CT slices acquired contiguously with 0.5-cm spacing and
of left-sided breast cancer by means of opto-electronic 0.5-cm thickness.
position detection, including a respiratory gating protocol. The breast clinical target volume (CTV) was outlined on
The detected misalignments were used to quantify the re- each CT slice using the visible breast parenchyma; the
lated dosimetric effects within the target volume and the superficial CTV limit was defined to extend to 5 mm be-
organs at risk (OAR) (lung, heart). This was obtained by neath the skin. In field definition, safety margins were added
calculating a spatial transformation, describing the mea- in all directions to the CTV to take into account breathing
sured errors and feeding this information back to the treat-
movements, setup errors, and beam penumbra. Particularly,
ment planning system (TPS). The clinical aim of the study
in cranial and caudal directions field edges extended 1.5 cm
was to check quantitatively whether the size and occurrence
from the CTV; superficially, the border of the field extended
of patient setup errors might alter significantly the planned
in air with 1-cm margin. Adequate CTV coverage, central
radiation dose distribution in patients undergoing radiother-
lung distance not exceeding 2.5 cm, and avoidance of con-
apy for breast cancer.
tralateral breast irradiation were the criteria followed for the
deep field margin definition (Fig. 1).
METHODS AND MATERIALS The two considered OAR were heart and left lung. The
The analysis was performed on a set of 89 acquisitions cranial limit of the heart included the infundibulum of the
from a group of 4 patients (age, 58.7 ⫾ 5.2 years), who gave right ventricle, the right atrium, and the right atrium auricle
informed consent to be included in this study. Patient se- and excluded the pulmonary trunk, the ascending aorta, and
lection criteria were mainly related to the specific clinical the superior vena cava. The lowest external contour of the
approach (quadrantectomy followed by postoperative radio- heart was the caudal border of the myocardium. The peri-
therapy) and to the side of the pathology (for the left breast, cardium was not excluded from the heart volume. The
heart is an OAR). contour of the lung was automatically outlined.
Dose–volume histograms (DVHs) were computed for the
Irradiation technique CTV and for the OARs and constituted the baseline data for
Treatment technique consisted of whole breast irradiation the quantification of the dosimetric deviations due to patient
with opposed tangential 6-MV photon beams and a pre- localization errors detected at each daily fraction.
Setup errors dosimetric effects in left breast irradiation ● G. BARONI et al. 863

Fig. 2. Control points selection procedure. Anatomic skin landmarks were highlighted by means of radiopaque markers
(upper left panel). The corresponding three-dimensional reference positions were manually extracted from the virtual
environment of the treatment planning software (upper right panel). For each patient, the resulting configuration
guaranteed the presence of: one control point in correspondence to the center of irradiation field (C); two control points
in correspondence to the tattoos on the mediosternal line (Tup, Tdown); two control points in correspondence to the upper
margin of the irradiation field (FupM, FupL); one control point in correspondence to the lower margin of the irradiation
field (Fdown). Two additional control points (R1, R2) placed on the transversal umbilical line were used for patient
breathing phase detection.

Acquisition of CT reference position Patient position detection system


The CT reference position was defined by identifying the At each daily fraction, the patient was aligned at the
3D coordinates of a set of surface control points on the CT therapy unit by means of conventional manual alignment
slices used for treatment planning (Fig. 2). These were techniques (laser centering), and the accuracy of patient
selected in correspondence to the tattooed references on the repositioning was monitored by means of an opto-electronic
sternum used for manual patient alignment and to additional position detection system (ELITE, B.T.S. Spa, Milan, Italy)
natural skin marks identified on each patient. During CT equipped with specific software for radiotherapy application
scan acquisition, the landmarks were highlighted by means (Fig. 3) and breathing detection.
of radiopaque markers (3 mm in diameter). Their 3D posi- The system provided real-time detection and 3D recon-
tions were measured on the CT slice, which best displayed struction of the position of reflective markers (radius, 5 mm;
the entire marker section. This procedure provided the CT- fractions of gram in weight) included in the field of view of
based reference dataset. Interfractional patient setup errors, a couple of infrared (750 – 820 nm) TV cameras (100 Hz
measured at the therapy unit by means of the opto-electronic sample rate) installed in the therapy room. The calibration
patient position detection system (see “Acquisition of treat- procedure for the 3D marker localization was performed by
ment patient position”), were superimposed on the CT ref- means of a 7 ⫻ 7 grid of markers, which was aligned to the
erence position to quantify the related dosimetric effects isocentric reference laser lines and acquired at five different
(see “Quantification of the altered dosimetry”). vertical positions of the treatment couch. The resulting
864 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004

Fig. 3. Schematic representation of hardware (left) and software (right) components of the opto-electronic localization
system. TVC1 and TVC 2 are the CCD infrared (IR)-light sensitive TV cameras equipped with IR LEDs (light emitting
diodes) for scene illumination. Markers on subjects appear as bright spots on the video images which are elaborated in
real time for marker 2D localization by means of shape cross-correlation. Markers are automatically labeled via software
exploiting the memorized 2D reference configuration of the control points. The application of stereophotogrammetric
techniques leads to the real-time three-dimensional reconstructions of markers position, the calculation of the control
points displacements, and the estimation of patient position correction parameters by means of a least-squares
minimization procedure.

reference system axes (x, y, z) of the opto-electronic local- along the sagittal laser reference line; the cranial tattoo was
izer coincided with the latero-lateral, craniocaudal, and an- positioned at the intersection between the sagittal and the
teroposterior directions of the CT/linac units. In the frame of lateral laser lines at the prescribed source–skin distance; the
this specific motion capture system setup (the calibrated final patient setup was obtained by shifting laterally the
volume centered in correspondence to the linac isocenter treatment couch according to the indication provided by the
measured 480 ⫻ 480 ⫻ 160 mm3), errors in 3D marker treatment plan. Portal images of the two tangential fields
localization were assessed to be lower than 0.5 mm (stan- were acquired at the first irradiation session to check the
dard deviation superimposed on the length of a marked stick irradiation geometry. Patients were fitted with a set of
freely moved within the calibrated volume) (13). light-reflecting markers, applied in correspondence to the
skin landmarks selected during CT scanning. Markers’ 3D
Acquisition of treatment patient position positions were recorded for 15 s (1500 frames) immediately
At each irradiation, the patient was positioned in the before the irradiation (localization acquisition).
vacuum cushion and arm-supporting device (Fig. 1) and According to Baroni et al. (11, 13), the recorded markers’
was manually centered at the therapy unit according to the coordinates underwent a specific data processing for breath-
following procedure: the two sternal tattoos were aligned ing movement compensation, consisting of:
Setup errors dosimetric effects in left breast irradiation ● G. BARONI et al. 865

1. Detection of patient breathing movements (opto-elec- points displacements. Among the six parameters (three ro-
tronic respiratory gating). The 3D coordinates of two tations and three translations) provided by the least-squares
additional markers placed on the transverse umbilical algorithm, we were forced to discard the rotation around the
line (R1 and R2 in Fig. 2) were used to reconstruct a latero-lateral axis, due to an intrinsic limitation of TPS. The
kinematic signal related to the thoracoabdominal move- least-square optimization procedure was therefore per-
ments due to respiration. formed on the remaining five parameters. Residual values of
2. The averaging of control points position only on the the cost function (the squared sum of the marker displace-
frames corresponding to the minima of the obtained ments) greater than zero were obtained in all cases. This
breathing phase signal (EE, end of spontaneous indicated, as expected, that the measured localization errors
expiration). could not be completely interpreted by a five-parameter
rigid roto-translation. The size of cost function residuals
This procedure was guaranteed to calculate the treatment
was used to estimate the intrinsic accuracy of the proposed
position with no influence of patient breathing or random
rigid body approach. The five roto-translation parameters
movements (within each acquisition, control points position
were used for two purposes:
intrasession variability at the EE frames was checked to be
lower than 1.0 mm). To avoid biases in data comparability 1. To simulate the improvement of patient position, as if the
between CT and the therapy unit and to valorize the mask- radiographers would have acted on treatment couch and
ing of the influence of breathing on the measurement of gantry servo-controlled movements, according to the es-
setup errors, the markers’ positions measured at the first timated roto-translation parameter
radiation session defined the treatment reference position. 2. To alter the treatment geometry within the TPS accord-
Interfractional patient mispositioning was calculated by ing to the following criteria:
comparing treatment reference and treatment current posi-
tions, which were obtained according to the above-de- —rotation around the craniocaudal axis was set by acting
scribed data processing. This allowed us to quantify the pure on the gantry out-plane orientation;
repositioning inaccuracies related to the intrinsic efficacy of —rotation around the anteroposterior direction was set
the conventional means of patient manual alignment. by acting on the table in-plane isocentric orientation;
Control points positions were again recorded during ir- —the three translations were set by acting on the 3D
radiation (verification acquisition). In this case, data analy- position of the isocenter.
sis was finalized to quantify the peak position deviation, To ensure the dosimetric comparability with the original
which occurred during the irradiation. Among the set of plan, the new dose distribution was renormalized on the
recorded frames, the one exhibiting the highest markers original isocenter position and the field weights were even-
displacement with respect to the reference position was tually changed to preserve the monitor units delivered by
identified. This arose from a combination of setup inaccu- each field.
racies, breathing, and random patient movements. Dosimetric deviations with respect to the original treat-
ment plan were assessed on the differential DVHs for the
Quantification of the altered dosimetry CTV and the OARs (lung and heart) in terms of percentage
The difference between reference and current markers variation of the volume included in predefined dose level
coordinates was the first level of information related to the intervals. For the CTV, we computed the percentage varia-
accuracy of patient repositioning. These data were used to tion of the volume receiving less than 95% (⌬V⬍95%) and
evaluate the corresponding deviations in the absorbed dose more than 105% (⌬V⬎105%) of the daily dose; for the
distribution, by means of the following procedure OARs, the percentage variations of the volume receiving
more than 50% (⌬V⬎50%) and 90% (⌬V⬎90%) of the daily
1. Marker displacements measured at each irradiation were dose were calculated.
superimposed on the corresponding CT reference data-
set, to define an irradiation-specific CT marker configu-
ration. RESULTS
2. Specific software in MATLAB (The MathWorks Inc.,
Setup errors
Natick, MA) was developed to estimate the rigid spatial
Figure 4 reports the linear displacements of the control
transformation (roto-translation), which best interpreted
points with breathing compensation, with respect to the
marker displacements between reference and current
reference configuration. Distribution fitting (Kolgomorov-
configurations.
Smirnov statistical test) revealed that the displacements
3. The irradiation geometry was altered in the treatment
were normally distributed around the mean values. The
plan system and a new dose distribution was calculated.
analysis of variance (ANOVA) with displacement direction
Marker redundancy (only three points are necessary to as independent variable, followed by post hoc comparison
describe a rigid spatial transformation) required the imple- of means showed that the observed significant differences (p
mentation of a least-squares algorithm, to estimate the best ⬍ 10⫺4) were imputable to dorsoventral repositioning er-
transformation, which described the set of measured control rors along the anteroposterior direction. ANOVA, applied
866 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004

Fig. 4. Linear marker displacements (mean ⫾ standard deviation, and highest-lowest values are reported as whisker
plots) along the three main anatomic directions. Data were averaged on a total of 89 irradiation sessions for the 4
patients. Marker positions are reported on the right figure panels.
Setup errors dosimetric effects in left breast irradiation ● G. BARONI et al. 867

Fig. 5. Three-dimensional marker displacements (mean ⫾ quartile, and highest-lowest values are reported as whisker plots)
averaged on a total of 89 irradiation sessions for the 4 patients. Upper panel: 3D peak marker displacements identified on the
set of verification acquisitions. Middle panel: Marker displacements measured in correspondence to the end of spontaneous
expiration (EE) and used to evaluate the corresponding dosimetric deviations. Lower panel: 3D residual errors after the
hypothetical patient position correction by means of the least-squares estimated roto-translation parameters.
868 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004

Fig. 6. Frequency plot of volumetric variations of the clinical target volume (CTV) receiving less than 95% (⌬V⬍95%)
and more than 105% (⌬V⬎105%) of the prescribed dose (2 Gy) with respect to the original plan.

with the patient as an independent variable, did not show ⌬V⬍95% shows the occurrence of nine isolated cases, ex-
significance differences. hibiting volumetric variations of the CTV receiving less
Figure 5 depicts the corresponding 3D localization errors. than 95% of the prescribed dose greater than 2% and up to
In the upper panel, the peak markers’ displacements de- 12%. Statistical analysis (ANOVA, with patient as indepen-
tected during the irradiation (verification acquisition) are dent variable, and post hoc contrast analysis Scheffé test)
reported. According to the data analysis procedure, these revealed that the overall significant differences between the
data represent the interfractional average of control points examined patients (⌬V⬍95%: F (3, 85) ⫽ 4.91, p ⫽ 0.003)
displacements, measured in correspondence to the frame at were due to larger volumetric variations for one specific
which the highest position deviation was detected. patient (Patient 3) in comparison to the others (⌬V⬍95%:
When respiratory compensation was added in the data lowest significance p ⫽ 0.025). Only four observations
analysis, the corresponding 3D patient misalignments (Fig. exhibiting variations exceeding 2% in the ⌬V⬎105% distri-
5, middle panel) show that eliminating the influence of bution were observed.
breathing and random movements led only to a slight de-
crease of the median values and variability of setup errors. OAR dosimetric variations
Only the control points placed on the sternal tattoos (Tup, Figures 7 and 8 report the frequency distribution of the
Tdown in Fig. 2) were affected by 3D errors below 5 mm. OARs (lung and heart) percentage volumetric variations
Noteworthy is the presence of control point displacements receiving more than 50% (⌬V⬎50%) and 90% (⌬V⬎90%) of
up to 1.2 cm, which might represent the effects of isolated the daily dose. Lung volumetric variations (Fig. 7) exhibited
macroscopic operator-dependent misalignments. mean values and data variability of ⫺0.7 ⫾ 2.1% and ⫺0.4
Figure 5 is completed by the residual 3D displacements ⫾ 1.6%, respectively. ⌬V⬎50% and ⌬V⬎90% showed volu-
affecting the control points after the simulated application metric variations exceeding 2% in 11 and 7 cases, respec-
of the position correction, described by the five-parameter tively. Results account for a generalized decrease of
rigid transformation (lower panel). They represent the re- ⌬V⬎50% and ⌬V⬎90% for 3 patients (mean ⫾ SD: ⌬V⬎50%
sidual errors, if the patient position would have been cor- ⫽ ⫺1.6 ⫾ 1.5%; ⌬V⬎90% ⫽ ⫺1.2 ⫾ 1.1%). The largest
rected by acting on the treatment couch and gantry. positive variations were found for 1 patient (Patient 4), for
whom ⌬V⬎50% and ⌬V⬎90% were 1.5 ⫾ 1.6% and 1.3 ⫾
CTV dosimetric variations 1.4%, respectively. Statistical analysis confirmed significant
Figure 6 shows the dosimetric effects within the planned differences for this patient in comparison to the others (p ⬍
target volume of the patients’ setup errors when breathing 10⫺6).
movement effects were compensated. The results are re- The percentage variations of the heart volume receiving
ported in terms of frequency distribution of the percentage more than 50% and 90% of the daily dose (Fig. 8) were
CTV volumetric variations outside the 95–105% dose range found to vary between ⫺3.0 and ⫹ 3.0%, with notably low
with respect to the original treatment plan. average values and data variability (mean ⫾ SD: ⌬V⬎50% ⫽
Outcomes revealed that mean values and data variability ⫺0.005 ⫾ 1.1%; ⌬V⬎90% ⫽ ⫺0.02 ⫾ 0.5%). Only three
(standard deviation, SD) were 1.0 ⫾ 2.1% and 0.4 ⫾ 0.7% cases with volumetric variations greater than 2% were ob-
for ⌬V⬍95% and ⌬V⬎105%, respectively. The distribution of served.
Setup errors dosimetric effects in left breast irradiation ● G. BARONI et al. 869

Fig. 7. Frequency plot of volumetric variations of the left lung receiving more than 50% (⌬V⬎50%) and 90% (⌬V⬎90%)
of the prescribed dose (2 Gy) with respect to the original plan.

No variations with respect to the original plan were found With respect to the displacements of the external control
for 1 patient (Patient 2), and one isolated case (Patient 1) points, results agreed with a previous clinical study on 5
showing noteworthy variations (⌬V⬎50% ⫽ 5.8%; ⌬V⬎90% breast-cancer patients (13). When breathing movements
⫽ 3.2%) was observed. Statistical analysis revealed signif- were compensated during data analysis (control points dis-
icant differences (⌬V⬎50%: p ⫽ 0.0005; ⌬V⬎90%: p ⫽ 0.03) placements were measured at patients’ end of expiration),
between data from Patient 1 and Patient 4. only markers placed on the sternal tattoos used for patient
manual alignment exhibited on average an acceptable posi-
tion reproducibility (Figs. 4 and 5, middle panel). In partic-
DISCUSSION
ular, Tup turned out to be the best repositioned marker, as the
In this work, the interfractional patient setup errors were only reference that was directly checked by the radiogra-
measured and their effects on the dose distribution were phers to be at the point of intersection of the lateral and
quantified in patients undergoing external radiotherapy at sagittal reference laser lines and at the prescribed skin–
the left breast. With respect to previous studies relying on source distance. Considerably higher errors affected the
portal imaging (1, 2, 10), the method was in this case based control points, which were relatively far from the directly
on opto-electronic patient position verification, providing at inspected skin landmarks, but still within or close to the
each irradiation session the displacements of a set of surface irradiation field. When the peak localization errors (includ-
control points with respect to a corresponding reference ing the effects of breathing and random movements) were
configuration. The detected mispositioning was fed back measured, even higher median values and variability of the
into the treatment planning system for the evaluation of the displacements were found (Fig. 5, upper panel). The occur-
resulting dosimetric alteration within the target and the rence and size of the observed setup errors suggest that the
OARs (lung, heart). couple of skin tattoos (Tup, Tdown, see Fig. 2) used for

Fig. 8. Frequency plot of volumetric variations of the heart receiving more than 50% (⌬V⬎50%) and 90% (⌬V⬎90%) of
the prescribed dose (2 Gy) with respect to the original plan.
870 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004

manual patient alignment might not be enough to guarantee affected by negligible variations of the volume receiving
an adequate patient repositioning, when patient position more than 50% and 90% of the prescribed dose (see Fig. 7,
verification methods (portal imaging or opto-electronic lo- Fig. 8, and “Results” section). However, among the 11 cases
calization) are not applied systematically. of lung ⌬V⬎50% higher than 2%, 9 were found for one
A least-square minimization procedure was applied to specific patient (Patient 4) on a total of 27 (33%). This same
condense the measured local displacements of the control patient was the one for whom the two cases of heart
points into a five-parameter (two rotations and three ⌬V⬎50% over 2% were observed.
translations) spatial transformation. The simulation of the Deterministic relationships between surface control
corresponding patient position correction caused a dra- points displacements and dosimetric deviations could not
matic reduction of the initial control points displacements be established. Control points displacements turned out
(Fig. 5, lower panel). Although recent studies have ques- to exhibit negligible differences between patients, thus
tioned the effective advantages of using opto-electronic confirming that their size and occurrence are mainly
devices with respect to manual repositioning procedures related to the specific means and procedures used for
(15), this result agrees with previous clinical experience manual patient alignment (9). On the contrary, a patient-
(13, 14) and confirms that opto-electronic position detec- specific categorization of the corresponding dosimetric
tion systems might serve efficiently as patient reposition- effects was observed. This mainly depended upon the
ing means, leading to a significant improvement of the modalities with which the CTV and the OARs were
accuracy in patient setup. drawn on the CT sets. As an example, it was clear that the
The five-parameter spatial transformation was also esti- largest reduction of the CTV within the dose range 95–
mated from the CT-based dataset, and was used to alter the 105% for 1 patient was caused by the particular BEV
treatment geometry within the plan, to quantify the corre- design, which was very close to the CTV internal tan-
sponding dosimetric alteration (see “Quantification of the gential border. In this case, even small patient misalign-
altered dosimetry”). From the methodological point of view, ments along specific directions could cause large percent-
the reliability of this approach is supported by the redundant age volume variations.
number of control points, by the application of a least- In conclusion, although it is difficult to give a clinical
squares procedure, and by the opto-electronic breathing interpretation of the described dosimetric deviations, the
movement detection and compensation. This latter excluded reported results highlight the role of opto-electronic patient
biases due to organ motion of CTV and OARs induced by localization systems for the quantification of size and oc-
respiration. Furthermore, the least-squares estimation of the currence of interfractional patient setup errors, as a function
roto-translation operator from multiple control points en- of the efficacy of specific patient alignment and immobili-
sured that most of the interfractional soft tissue deforma- zation devices. This information is considered essential for
tions and volumetric variations of the CTV were taken into the optimization of the CTV safety margins design, ac-
account. The intrinsic inaccuracies of the reported approach counting for patient-specific effects of mispositionings on
are represented by the residual values of the cost function of dose distribution (17). The availability of a quantitative
the least-squares minimization procedure. Actually, they description in 3D of experimentally collected setup errors
coincide with the residual errors reported in Fig. 5 (lower can be used to assess a priori the sensitivity of the specific
panel). Their size is on average lower than the uncertainties PTV design to dosimetric deviations due to patients’ mis-
of the pseudo-CT set reconstruction from portal images (16) alignments (3). Conversely, the use of patient position ver-
and is put forward to represent an acceptable range of ification methods during patient irradiation provides data for
confidence, within which the measured dosimetric devia- the evaluation a posteriori of mispositioning dosimetric
tions were expressed. effects, thus permitting one to undertake corrective proce-
The evaluation of the altered dose distribution within dures for the remaining set of irradiations. In this frame, the
CTV and OARs revealed that, although the observed patient opto-electronic real-time detection of the 3D displacements
mispositionings were considerable, they did not generally of surface control points emerges as an alternative, prom-
lead to critical underdosages or overdosages. This testifies ising technique, alongside conventional portal imaging and
that the trade-off between the intrinsic efficacy of the con- CT reconstruction methods, implying low time cost of op-
ventional methods of patient alignment and the criteria for erator-dependent procedures for the verification of the do-
treatment plan design (beam’s eye view [BEV]) turned out simetric effects of setup errors. The noteworthy advantages
to guarantee sufficient dose homogeneity in the CTV and are related to the real-time detection of localization errors,
OARs. Volumetric variations of the CTV outside the dose to the possibility of differentiating the factors contributing
range 95–105% were on average 0.7 ⫾ 1.6% (mean ⫾ SD). to the measured mispositioning (operator-dependant mis-
However, among the 13 CTV volumetric variations outside alignments, breathing movements), and to the potential ap-
the 95–105% dose range, which overcame the action level plication of opto-electronic body surface detection and reg-
threshold (2%), 8 were observed in one single patient (Pa- istration methods (18) for the measurement of changes in
tient 3) in a total of 23 cases (35%). The same reasoning breast morphology and volume over the course of treatment
applies to the OARs, which on average turned out to be (19).
Setup errors dosimetric effects in left breast irradiation ● G. BARONI et al. 871

REFERENCES
1. Hector CL, Evans PM, Webb S. The dosimetric consequences electronic portal imaging to assess cardiac irradiation in breast
of inter-fractional patient movement on conventional and in- radiotherapy. Clin Oncol (R Coll Radiol) 1997;9:259–261.
tensity modulated breast radiotherapy treatments. Radiother 11. Baroni G, Ferrigno G, Orecchia R, Pedotti A. Real-time three-
Oncol 2000;54:57–64. dimensional motion analysis for patient positioning verifica-
2. Hector CL, Evans PM, Webb S. The dosimetric consequences tion. Radiother Oncol 2000;54:21–27.
of inter-fractional patient movement on three classes of inten- 12. Rogus RD, Stern RL, Kubo HD. Accuracy of a photogram-
sity-modulated delivery techniques in breast radiotherapy. Ra- metry-based patient positioning and monitoring system for
diother Oncol 2001;59:281–291. radiation therapy. Med Phys 1999;26:721–728.
3. Hector C, Webb S, Evans PM. A simulation of the effects of 13. Baroni G, Ferrigno G, Orecchia R, Pedotti A. Real-time opto-
setup error and changes in breast volume on conventional and electronic verification of patient position in breast cancer
intensity-modulated treatments in breast radiotherapy. Phys radiotherapy. Comput Aided Surg 2000;5:296–306.
Med Biol 2001;46:1451–1471. 14. Soete G, Van de Steene J, Verellen D, et al. Initial clinical
4. Hurkmans CW, Borger JH, Bos LJ, et al. Cardiac and lung experience with infrared-reflecting skin markers in the posi-
complication probabilities after breast cancer irradiation. Ra- tioning of patients treated by conformal radiotherapy for pros-
diother Oncol 2000;55:145–151. tate cancer. Int J Radiat Oncol Biol Phys 2002;52:694–698.
5. Gagliardi G, Bjohle J, Lax I, et al. Radiation pneumonitis after 15. Weiss E, Vorwerk H, Richter S, Hess CF. Interfractional and
breast cancer irradiation: Analysis of the complication prob- intrafractional accuracy during radiotherapy of gynecologic
ability using the relative seriality model. Int J Radiat Oncol carcinomas: A comprehensive evaluation using the ExacTrac
Biol Phys 2000;46:373–381. system. Int J Radiat Oncol Biol Phys 2003;56:69–79.
6. Gagliardi G, Lax I, Ottolenghi A, Rutqvist LE. Long-term 16. Evans PM, Hansen VN, Mayles WPM, Swindell W, Torr M,
cardiac mortality after radiotherapy of breast cancer: Appli- Yarnold JR. Design of compensators for breast radiotherapy
cation of the relative seriality model. Br J Radiol 1996;69: using electronic portal imaging. Radiother Oncol 1995;37:43–
839–846. 54.
7. Canney PA, Deehan C, Glegg M, Dickson J. Reducing cardiac 17. Baroni G, Pedotti A, Orecchia R. Radiotherapy and tamoxifen
dose in post-operative irradiation of breast cancer patients: in women with completely excised ductal carcinoma in situ.
The relative importance of patient positioning and CT scan Lancet 2003;362:1155.
planning. Br J Radiol 1999;72:986–993. 18. Moore C, Lilley F, Sauret V, Lalor M, Burton D. Opto-
8. Canney PA, Sanderson R, Deehan C, Wheldon T. Variation in electronic sensing of body surface topology changes during
the probability of cardiac complications with radiation tech- radiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys
nique in early breast cancer. Br J Radiol 2001;74:262–265. 2003;56:248–258.
9. Lirette A, Pouliot J, Aubin M, Larochelle M. The role of 19. Baroni G, Troia A, Riboldi M, Orecchia R, Ferrigno G,
electronic portal imaging in tangential breast irradiation: A Pedotti A. Evaluation of methods for opto-electronic body
prospective study. Radiother Oncol 1995;37:241–245. surface sensing applied to patient position control in breast
10. Magee B, Coyle C, Kirby MC, Kane B, Williams PC. Use of radiation therapy. Med Biol Eng Comput 2003;41:679–688.

You might also like