AIHA
Laboratory Quality
Assurance
Manual
Ten
Eric L. Botnick, CIH
Ram $, Suga, SC, CIH, CHMM
Kenneth 7, White, CIH, CSP
A Publication by
American Industrial Hygiene AssociationLaboratory
Quality Assurance
Manual
Third Edition
Edited by Eric L. Botnck, CH, Ram S, Suga, SC, CI, CHMM,
fd Kenneth T Whit, CNT, CSP‘This manval was developed by experts with backgrounds, taining, and experience in laboratory quality
assurance, storking with information and conditions existing atthe time of publication, The American
Industrial Hygiene Association (AIHA), as publisher, and the authors and editors have been diligent in
‘ensuring tht the material and methods adessed in this book reflec prevailing oseupational health and
safty and industrial hygiene practices. It is possible, however, that eemain procedures discussed will
requis modification because of changing federal, state, and local regulations, oF heretofore unknown
developments in research,
AIHA and the authors and editors disclaim any Habiliy, loss o risk resulting directly or ininectly from
use of the practices andiortheries discussed in this book. Moreover it isthe users responsibility to say
informed of any changing federal, state, of lea regulations that might affect the material eantained herein,
and the polices adopted specially in the user's workplace
Specific mention of manufacturers and products in this book doesnot represent an endorsement by AINA
Please send all comments regarding this Lab Quality Assurance Manual to
‘rie botnick@monsanio.com
‘Copyright 2005 by the American Industrial iygiene Association.
All rights reserved. No part ofthis publication may be reproduced in any form, by photostat, mero,
retrieval system, or any other means, without pri written permission ofthe publisher
ISBN 1-931504.63.6
2700 Prospsity Avemue, Suite 250
Fairfax, VA 22031
Stock No, LOAMOS-699Laboratory Quality Assurance Manual ‘Third Edition
Contents
Page
Figures, Tables, and Forms ¥
Preface si
1, Introduction 1
LL Introduction, 1
12 Purpose. 1
13 Paliies.
14 Definitions 1
1.5. Elements ofa Quality Assuranoe Program, ei a 2
16 List of Abbreviation oun eats i a 2
17. References 4
2._Personnel and Training 5
ion, and Control of Errors eee ees
34 Sources and Types of Ero. 9
3.5 Sunifcan! Figures ond Rounding 9
Sle rao ace a
+ Refrenoe
38 Furthor Reading 12
Sampling Procedures... zs a sas 13
42 Quality Control (QC) Samples 14
“43. Field Fquipment Calibration Considerations iS
44 Diflision Monitor Considerations Sl
45 Sampling ldetifieation, 11
{46_ Sampling Data Sheets. vo a 17
‘42 Shipping, Field Preservation, and Procauion assum sa
Aa tra TE
5. Sample Receipt and Handling. ar
5.1 Introduction, en
52. Chain of Custody 2
53 Information Reporting 22
6. Intradaboratory and Intertaboratory Quality Testing. 2
6.1 Introduction 29
62. Ints-lboratory Considerations 29
63 Inter aborwry Considerations boos cn bos 30
7. Analytical Methods at
7.1 Intraduotion, 3
72. Method Planing. 3
73 Method Validation Docu 33
74. Method Format 34
American Indusval Hygiene Association iitLaboratory Quality Assurance Manual Third Edition
Figures, Tables, and Forms
Figures
Chaprer 3
Figure 3.1 Additive Eeror, 4
Figure 3.2 Proportional Eror 5
Chapter 4
Figure 41 System Flow Chart 13
Chapter §
Figure $.1 Sample Seal. 23
Chapter 9
Figure 9.1 Duplicates-Range Variation. en na en “4
Figure 9.2Youden Tow — Sample Plt 45
Figure 9.3 Reference Contol Chast 46
Figure 9.4 Duplicate Control Chart Wrong Method a
Figure 9-8 Duplicate Control Chart — Britton’s Method, a7
Figure 0.6 Duplicate Contol Chart — Taylor's Method a8
Figure 97 Duplicate Conttol Chart — NYSDOH Method. a8
Tables
Chapter 3
‘Table 3.1 Sources of Potential Ero rw se bow 10
Chapter 7
Table 7.1 QC Table 39
Chapter 9
‘Table 9.1 Example Data 46
Forms
Chapter 4
‘tachment A: Industrial Hygiene Sampling Data Sheet 20
Chapter §
Adachment Bi: Air Sampling Worksheet 24
Atachment C: Air Sampling Report. 36
Aachment D: Chain-of Custody Record 27
American Indusval Hygiene Association vLaboratory Quality Assurance Manual
“Third Edition
75 Analytcal Teems and Limits.
7.6 Quality Contol Sample Aceeptnce Crit
17 References
Equipment Calibration and Maintenance.
8.1 Introduction,
82 Instruments
83 Glassware and Volumetric Apparatus
S4 Lleetncel Power Sourve
Data Validation and Interpretation
9.1 Introduction
9.2. Data Review Procedure,
93. Quality Control Chart
9.4 Occasional Samples
95. Statistical Analysis for Quality Contol of Precision
9.6. Rejoeton of Results.
9.7 Comctive Action Pan
98 System Aus
9.9) Referenees.
10, Reporting and Record Keeping
LT Introduction,
02 Elements of the Analsial Report
103 Recond Retention Program
14 Sample Retention Program
11, Analysis of Fiber Containing Samples.
HL Introduction,
11.2 Sample Shipment
13 Contamination Checks.
11.4 Fiber Counting Using Phase Contrast Microscopy (PCN)
1.5 Aishorne Asbestos Fibers by Transmission Electron Miroscopy
11.6 Asbestos Bulk Identification Using Polarized Light Microscopy
1137 References
12, Quality Assurance Considerations for Blologleal Monitoring
121 Introduction,
122 Definition .
123 Regulations and Exposure Indie
124 Project Initiation
125 Sample Collection
12.6 Sample Analysis
122 Quality Assurance Quality Control
128 Reporting Results
129 References,
36
39
30
41
al
al
2
2
4
8
a3
m3
45
46
49
SI
si
52
33
53
53
55
"39
a
s
65
aS
“6S
66
70
7
2
75
76
MOB a al pearance memmeancamaatt
American Indusval Hygiene AssociationLaboratory Quality Assurance Manual Third Edition
Chapter 2
Personnel and Training
21 Introduc
Qualified, properly trained personnel are a key aspect of satisfactory laboratory performance.
Although good fciities and equipment are essential, they are of litle value without well-trained
petsonnel. A laboratory, therefore, must establish adequate qualifications for personnel and
Provide an ongoing training program to develop and maintain the necessary abilities. Ako,
Petsonnel should be knowledgeable of the sampling and analysis processes, inorder to facilitate
{ener communication with field personne! and clients. Effective eoordinston is essential to
provide laboratory clients with saisictory analytical services,
22 Personnel Qualifications
‘The minimum qualifications for laboratory personnel should be those curently sot Forth in AIA
acerectation and proficieney testing programs, of other relevant accreditation qualification
Programs, Some laboratories may want to establish supplementary qualifications, such as
adkitional experience in specifi areas or on specifi topics, oF requiring that some personnel have
‘advanced scientific degrees and/or additional camistry education,
23 Personnel Trai
ig
2.3.1 General Considerations
‘QA program must include and provide forthe ongoing taining of beratory persomn. This
training program must be in print, adaptable, and include demonstrations of proficiency. A rscord
‘of qualifications and training must be kept foreach person,
This the responsibilty ofmanagem
otto establish and implement a traning program that ensures
that methods and procedures ste carried out propery. The program should include, a minim:
sampling techniques, analytes! methods, specifiy. accuracy and precision monitoring, limits oF
detection, potential interferences, and regulatory’authortaive His
2.3.2 Specific Training
It's the responsibility of laboratory management (however tiled) to ensure that all aboratory
petsonne! are introduced tothe QAM and ttsined peopetly in laboratory techniques, methods oF
analysis, and QA QC procedures. The laboratory should have a writen training program or
‘checklist to ensure that ll Iboratory personnel rseive adequate training in policies, procedures,
methods, instrumentation, caleulations, reporting, quality coatrl, and safety, with special
‘emphasis on th safe handling of toxie and carcinogenic materials. Training of new personne!
should be done on the job under the guidance of qualified, experienced analysts, This taining
should be supplemented with attendance at seminars and courses, such as those presented by
instrament manufactrers. so, eoutses in the fundamentals of industrial hygiene, industrial
hygiene chemistry, ahoratory accreditation, quality conto, ssbestos analysis, lead sampling ond
analysis, ete, are presented by various college and university educational resource centers, in
Addition to the National Institute for Occupational Safety and Heath (NIOSH), the US.
Environmental Protcction Agency (EPA), and AIHA. Finally there should be opportunities for
Petsonnel to ksep curent in their fick of interest through attendance at appropriate technical
meetings, such asthe annual American Industral Hygiene Conferenee & Exposition (Hee).
An analyst must be required to demonstrate profcisney in the performance of new methods or
techniques as the last stp in the training process before beginning to analyze actual fd samples.
‘This ean be done with the sttisfactory analysis of Known samples, such as certified reference
American Indusval Hygiene Association 5Laboratory Quality Assurance Manual Third Edition
Chapter 3
Error and Sources of Error
31 Introduction
An essential tsk in a QA program is the detection of error ara its elimination of contol, This
‘chapter discusses eror in sampling and analysis. Error and is various types are defined, and ways
to detcot and contral error aze presented briefly, Also, specific sources of error are given for
‘consideration during sampling and analysis,
As the first step inthe detsction of err, the analyst must be familiar with the definitions ofthe
lypes of cror that might Be encountered, an the relationship of error Io aecuraey and precision.
An understanding of these definitions increases awareness ofthe potential for eror aa allows
sources and types of eror to be descted nd classified, Once the sures and types of error are
idomitid, th elimination or contol of err should follow:
3.2 Definitions
Error cannot be defined without frst defining accuracy. Therefore, accuracy is defined as the
degree of agreement between the measured value and he tre value. rors defined as the degree
‘of disagreement ofa measured value with an accepted rferenee value (ie the “tue” value) Error
may have both systematic and random compononts!""
Systematic error is synonymous with dotsminate error, assignable crpor, oF “bias and is @
‘comsstont deviation in the results of @ measurement fom an aecepted reference value, The
‘of bias may or may not be known, but is considered to be assignable. Systmatiectrors ean be
classified as “additive” o¢ “proportional” The errors additive if it is # constant vale, regard
‘of the amount or concentration of the analyte (See Figure 3.1). It is possible to sce the diference
between additive and proportional error by plotting the observed values asa function of theoretical
values. Additive err would produce a linear plot with slope of one (just like ta that had no
‘tematic ero) but the y-interoept would not be zero it would be the same a the eer value
ANALYTICAL VALUE
THEORETICAL VALUE
Figure 3.1 — Additive error.
American Indusval Hygiene Association 7Laboratory Quality Assurance Manual “Third Edition
24
materials (CRM), extra sts of industrial hygiene profiieney analytical testing (PAT) program
samples from AIHA, extra sets of environmental lad proficiency analytical testing (ELPAT)
samples ftom AIA, or blind samples prepared by somone else ithe laboratory. The use of
Sondard reference materials (SRM) for Such purposes is discouraged, as it depletes the supply of
the materials, and other sources for samples are readily available. The laboratory’s routine QC
program show serve a8 a continuing check on the analysts performance
cori must he kept to document the qualifications and taining ofall laboratory personnel. The
‘minimum information in these records should ince formal education, prior experienc, taining
receivod on the job, et. Those records should also ineluds an outins, a Sasription of the course
an overview of the techaigue oF procedure, the dates trining took place, and the person of
btganization providing it. There also needs Yo bs a record indicating the qualifications of the
course instructor. A designated management representative should sign the record, indicating
satisfactory completion ofeach phase of traning and authorization to operate specific equipment
‘or perform specific analytical methods! Accreditation recognition programs may have specie
‘and more extensive recordkeeping requirements,
References
Anna
1. American Industrial Hyglene Association (ATHA): “AIA Laboratory Programs
Foye aa one/LaboratoryServieesthmblgap htm,
2. US. Environmental Protection Agency (EPA): Principles. In Quality Assurance Handbook
for sr Pollution Measurement Systoms, vol. 1 (EPA Pub. Ne. 600-9-76-008}. Cincinnati
EPA, 1976,
6
American Indusval Hygiene AssociationLaboratory Quality Assurance Manual “Third Edition
A systematic errr is sid to be proportional ifthe magnitude ofthe ero is proportional tothe
concentration of the analyte. 1f here isa systematic erro, thon a plot of observed values as a
function of theoretical values would demonstrate a slope diferent fiom unity. Systematic ror
ray be a more complex funetion of concentration and could result i a euviinear slope for sel
a plot (se Figure 3.2).
Although specific soursos of determinate (systematic) ctror ate listed for various aspoets of
sampling and analysis in a fate postion of this chapter the following ae luseative examples!
+ Poe recovery ofthe analyte from the sample media
+ Incorrect preparation of standards
+ Improper instrument calibration
+ Coataminated or outdated reagents
+ Human visual acuity deficiencies
+ man carelessness or personal bias
+ Improper selection of stmpling or analytic
+ Improper shipment and storage of samples
+ Interferences fiom other elements or ompourds
method
indeterminate e
ments, Ra
Random error is synonymous wit or o¢ non-assignable error an often is
Aofined in toms of the pcesion of meas ndom err i characterized by ieguat
‘aration in ropated observations of acasurements, Because the variations ae random in natu
they are unpredictable individually, The eause of this type of error generally cannot be asigne
Random error statistics usually are represented by & normal dstebution—a distribution having a
Frequency vs. size curve tat is belFshapedl and symmetrical about the arithmetic mean size
However, measurements of small entities with possibly large exors in the upward but not
dovsnward dieetion (Zero-fimited, non-negative measures) give skewed distribution curve that
is offen represented by a lognormal distribution (ic, the logarithms of the measurements ane
‘normally distributed) Random ctor canbe estimated statistically fom the proesion of replicate
‘determinations,
Sources of random erro ae identified typically a8 potential soures af variation inherent withthe
spestie steps of an established sampling and analytical method, Examples ince variations in
ANALYTICAL VALUE
THEORETICAL VALUE
Figure 3.2 — Proportional error.
8 American Indusval Hygiene AssociationLaboratory Quality Assurance Manual Third Edition
33
34
35
physical measurements of volume and mass, variations in instrament response, electrical line
Volage transients and sample heterogeneity
For a quantity or concentration that is derived from the measurement of several variables, an
estimate of th overall uncertainty in the devived quantity caused by random error can be made
using the law ofthe propagation of erors, This is dane by combining the contnbution of each
‘variable. Implicit im the law is the assumption that cach contnbuting variable is independent. For
linear relationships, the overall variance is the weighted average of the individual variances, with
degre of frecdom used a8 weighing factors Por prodicts and ratios, he square ofthe relative
‘rors (ic. the souares of the coeffcints of variation) are adkitive. The overall variance isthe
Square root ofthe sum ofthe squares of the coe cients of variation for each measured variable
‘Nore thorough presenation ofthis mathematical approach i given by Mandel snd Nanni!
An example of the use of the law of the propagation of errs is provided by NIOSH in the
validation of personal air sampling and analytical methods!) ‘The concentiation of an air
contaminant is the ratio of @ measured mass of the contaminant to a measured air volume
‘Consequently, NIOSH defines the overall coefficient of variation asthe square root ofthe sum of
the squares ofthe cocficicnts of variation for sampling (CVs), analysis and desorption efficieney
(CV gp and flow (CV
Detection, Elimination, and Control of Errors
‘The QA measures discussed in other chapters of this manual are useful in the detection
Laboratory Quality Assurance Manual
Third Edition
3.6
INOTE: Sce vendor's troubleshooting guide for typical
instrumentel problems or contact vendor's application
specialist for unusual mechanical problems |
Matrix effects
Methodology Incorrect recoveries of QC samples
Excessively high blanks
Deviations from documented procedures
Improper calculations of mass concentration
Inadequate extraction efficiency data over the
range of analysis
Data Interpretation Improper evaluation of reported results
Discounting large breakthrough results on solid
sorbent samples
Erroneous —
Spa = Beer Te eR
8 Co
BShoerne aad
83 | sw scm — =
e
Cane PP
roe ExeSETEM [anes Tao
zt 1 Loa ne
3 | pasbscines trolserarorhes
i
§ TWEE i SRT
S| pump "ERE ge a
Gime T poromeren
S| iiss | ee
ST oie
5 | Séilng
orale sal i » ow uel SB
‘asanecTER dtl SERRE eo, VOLUME SAUPLED CONCENTRATION. ‘Oita STANDARE
“ a Hs aty sey
3
5 - -
5h. ee
:
gl _