Professional Documents
Culture Documents
Donald S. Prough
The University of Texas Medical Branch
Department of Anesthesiology
Galveston, Texas 77555-0456
Fig. 2 OCT signal slope and blood glucose concentration versus time
Fig. 1 OCT signal slope and blood glucose concentration versus time when low pressure 共⬍1 kPa兲 was applied on the skin by the OCT
when high pressure 共⬎10 kPa兲 was applied on the skin by the 20-mm probe with the 50-mm adaptor.
OCT probe.
In the third set of experiments, we used the same probe as Table 1 OCT signal slope drift measured before glucose injection as
in the second set of experiments and applied two levels of shown in Figs. 1 and 2 when variation in glucose concentration did
not exceed 15 mg/ dL. The time interval used for the fit is shown in the
pressure: high ⬎10 kPa 共positive pressure was applied to the
third column. Depth refers to the distance from the skin surface at
probe兲 and low ⬍1 kPa 共negative pressure was applied to the which the correlation between blood glucose concentration and OCT
probe兲. In the first, second, and third sets of experiments, we signal slope was greatest.
used 12, 5, and 2 pigs, respectively.
The pressure applied to the pig skin was measured using Pig # Drift 共a.u./min兲 Time 共min兲 Depth 共m兲 Pressure 共kPa兲
the following procedure. An ex vivo pig skin sample was
placed between the OCT probe and a balance plate. The net 1 −0.06 25 515 ⬎10
force applied by the OCT probe was Ft = mef g = mg + Fadd,
where mef 共kg兲 is the reading of the balance display, m is the 2 0.006 20 185 ⬎10
net mass of the probe, Fadd is the external force applied to the 3 −0.005 30 505 ⬎10
probe to achieve the desired level of pressure on the tissue,
and g ⬇ 10 m · s−2 is the free-fall acceleration. The mass of 4a 0.009 35 155 ⬎10
skin sample 共0.5 to 2 g兲 was negligible compared to the mass
of the OCT probe. Thus, the total pressure was P = Ft / S, 5 0.009 30 190 ⬎10
where S = d2 / 4 is the area of the probe 共or adaptor兲 surface, 6 −0.009 30 415 ⬎10
and d is the probe 共or adaptor兲 diameter.
The OCT probe was attached with a spring system to an 7 −0.004 25 450 ⬎10
arm that could be moved up and down on a stable pole. The
moving arm permitted control of the pressure applied to the 8 0.010 30 485 ⬎10
skin. The vertical position of the arm was marked when the 9 −0.010 30 170 ⬎10
OCT probe was placed on the object without additional pres-
sure 共only the probe weight兲. The shift of the arm from the 10 0.006 30 425 ⬎10
recorded position in the in vivo experiments was used to cal-
culate the actual, total pressure applied to the skin in vivo. The 11 0.006 30 140 ⬎10
local and inter-subject variability in abdominal pig skin elas-
12 −0.006 30 200 ⬎10
ticity and in internal pressure 共opposite to the external pres-
sure applied by the probe兲 resulted in the large variability 13 −0.004 25 235 ⬍1
共about threefold兲 in the pressure levels needed to achieve the
same low level of motion artifacts in OCT glucose sensing in 14 −0.001 30 190 ⬍1
the different sets of experiments.
15b 0.004 45 160 ⬍1
The results of the first two sets of experiments are summa- 17 0.0005 35 130 ⬍1
rized in Table 1. For each pig we calculated the drift at the a
The results for this pig are shown in Fig. 1.
depth where the maximal absolute value of correlation coef- b
The results for this pig are shown in Fig. 2.
ficient 共R兲 between the OCT signal slope and blood glucose
concentration was observed 共兩R 兩 ⬎ 0.8, p ⬍ 0.03兲. The depth
at which the OCT signal slope was calculated in each pig is When we lifted the arm attached to the 20-mm OCT probe,
shown in column 4 of the table. The drift of the OCT signal the motion artifacts increased and precluded accurate OCT
slope can be positive or negative depending on the layer signal measurement.
where the OCT signal slope was measured. As we showed In the second set of experiments, in which we used the
previously,6,13 the correlation of the OCT signal slope with adaptor and lifted the support arm, the level of motion arti-
blood glucose concentration varies between +1 and −1 with facts was acceptable because the amplitude of motion was
depth due to the layered structure of the skin. The long-term inversely proportional to the force needed to move the heavier
drift 共for about 4 hours兲 in the OCT signal slope produced by system 共the probe with the adaptor兲 relative to the skin. When
the high pressure level with the 20-mm probe was noted in the arm was lifted, the estimated residual level of pressure
our earlier studies.6 The absolute value of the drift in the OCT applied to the skin surface was in the range of 0.25 to
signal slope was measured in this study, and for the high 0.60 kPa. Results for one of the pigs in this set of experi-
共⬎10 kPa兲 pressure was 0.012⫾ 0.015 arbitrary units 共a.u.兲/ ments are shown in Fig. 2. No significant pressure-induced
min 共Table 1兲. This was greater 共4.8 times兲 than that for the drift of the glucose-induced changes in OCT signal slope was
low 共⬍1 kPa兲 pressure: 0.0025⫾ 0.0017 a.u./ min. observed for 3 to 4 hours in this set of experiments. However,
Figure 1 shows typical OCT signal slopes and blood glu- in this particular pig, before the glucose injection 共first
cose concentrations over time when the 20-mm OCT probe 70 min. of the experiment兲, the drift in the OCT signal slope
was used. The estimated pressure applied to the skin was in was 0.004 a.u./ min. This value was higher compared with
the range of 13 to 38 kPa. The second peak in the OCT signal the average drift value of 0.0025 a.u./ min for this set of ex-
slope is shifted up due to the pressure-induced effects. periments.
To reduce but not eliminate the pressure of the OCT probe In the third set of experiments, we applied two levels of
on the skin, we lifted the arm attached to the OCT probe. pressure 共⬍1 kPa and ⬎10 kPa兲 at the same tissue site at
different times 共Fig. 3兲. During intervals 1 to 4, the glucose five-fold by increasing the contact area between the OCT
level was stable 共no glucose injection兲. Starting from interval probe and tissue, then applying optimal negative external
4, the pressure was ⬍1 kPa. Glucose was injected during in- pressure to the probe. This results in more accurate and repro-
terval 5 while the pressure was ⬍1 kPa. We estimated the ducible glucose monitoring with OCT at minimal long-term
drift in the OCT signal slope with time at these pressure levels drift and motion artifacts.
using linear fit of the data. Figure 3 shows that the OCT signal
slope drift during application of high pressure on the skin was Acknowledgments
0.013 a.u./ min and 0.008 a.u./ min 共intervals 1 and 3, re- This work was supported by the National Institutes of Health
spectively兲 and was four- to five-fold greater than drift at low 共grant #R01 EB001467 from the National Institute for Bio-
pressure: 0.003 a.u./ min and 0.002 a.u./ min 共intervals 2 and medical Imaging and Bioengineering兲. The authors thank Dr.
4, respectively兲. In the other pig in this set of experiments, the Donald J. Deyo for assistance with the animal experiments.
drift at high pressure 共0.05 a.u./ min and 0.01 a.u./ min for
intervals 1 and 3, respectively兲 was greater than that at low References
pressure 共0.001 a.u./ min and 0.003 a.u./ min for intervals 2 1. E. K. Chan, B. Sorg, D. Protsenko, M. O‘Neil, M. Motamedi, and A.
and 4, respectively兲. In the third set of experiments, the OCT J. Welch, “Effects of compression on soft tissue optical properties,”
IEEE J. Sel. Top. Quantum Electron. 2, 943–950 共1996兲.
signal slopes were calculated at depths of 120 m and 2. V. V. Tuchin, Optical Clearing of Tissue and Blood, SPIE Press,
125 m for the first and second pigs, respectively. The results Bellingham, Washington 共2006兲.
of the third set of experiments are consistent with results from 3. R. O. Esenaliev, K. V. Larin, I. V. Larina, and M. Motamedi, “Non-
the first and second sets of experiments. invasive monitoring of glucose concentration with optical coherence
tomography,” Opt. Lett. 26, 992–994 共2001兲.
It should be noted that the signal-to-noise ratio for glucose 4. K. V. Larin, M. Motamedi, T. V. Ashitkov, and R. O. Esenaliev,
monitoring in the second 共Fig. 2兲 and third 共Fig. 3兲 sets of “Specificity of noninvasive blood glucose sensing using optical co-
experiments was higher than that in the first set of experi- herence tomography technique: A pilot study,” Phys. Med. Biol. 48,
ments 共Fig. 1兲. This difference is associated with more effi- 1371–1390 共2003兲.
5. M. Kinnunen, R. Myllyla, T. Jokela, and S. Vainio, “In vitro studies
cient removal of the motion artifacts with the 50-mm probe toward noninvasive glucose monitoring with optical coherence to-
compared to the 20-mm probe. mography,” Appl. Opt. 45, 2251–2260 共2006兲.
The OCT signal slope in Fig. 1 was much smaller than 6. R. V. Kuranov, V. V. Sapozhnikova, D. S. Prough, I. Cicenaite, and R.
those in the other figures due to local variations of the OCT O. Esenaliev, “In vivo study of glucose-induced changes in skin
properties assessed with optical coherence tomography,” Phys. Med.
signal with depth that were observed in this and previous Biol. 51, 3885–3900 共2006兲.
studies.6,13 The OCT signal slope in pig skin is sharply posi- 7. A. Vogel, C. Dlugos, R. Nuffer, and R. Birngruber, “Optical-
tive, then negative from the skin surface to 20 to 50 m. The properties of human sclera, and their consequences for transscleral
slope becomes positive at the papillary layer at approximately laser applications,” Lasers Surg. Med. 11, 331–340 共1991兲.
8. B. Fromy, P. Abraham, C. Bouvet, B. Bouhanick, P. Frasinaud, and J.
100 to 250 m, then negative at the papillary-reticular junc- L. Saumet, “Early decrease of skin blood flow in response to locally
tion 共250 to 400 m兲 and the reticular skin 共400 to 650 m兲. applied pressure in diabetic subjects,” Diabetes 51, 1214–1217
At the dermis-hypodermis junction, it becomes sharply 共2002兲.
negative. 9. N. G. Klebtsov, I. L. Maksimova, V. V. Tuchin, and L. Wang, Hand-
book of Optical Biomedical Diagnostics, PM 107, V. V. Tuchin, Ed.,
One can see from Table 1 and our previous SPIE Press, Bellingham, Washington 共2002兲.
investigations6,13,14 that the maximum correlation between the 10. P. Abraham, B. Fromy, S. Merzeau, A. Jardel, and J.-L. Saumet,
OCT signal slope and glucose concentration is observed near “Dynamics of local pressure-induced cutaneous vasodilation in the
specific morphological regions, namely the papillary-reticular human hand,” Microvasc. Res. 61, 122–129 共2001兲.
11. M. Nitzan, A. Babchenko, B. Khanokh, and H. Taitelbaum, “Mea-
and dermis-hypodermis junctions. The depth of these junc- surement of oxygen saturation in venous blood by dynamic near in-
tions has inter-subject variability. To predict glucose levels, frared spectroscopy,” J. Biomed. Opt. 5, 155–162 共2000兲.
the OCT system first should be calibrated for each subject 12. S. A. Carp, T. Kauffman, Q. Q. Fang, E. Rafferty, R. Moore, D.
using standard invasive glucose measurement methods. Kopans, and D. Boas, “Compression-induced changes in the physi-
ological state of the breast as observed through frequency domain
photon migration measurements,” J. Biomed. Opt. 11, 064106 共2006兲.
13. V. V. Sapozhnikova, D. S. Prough, R. V. Kuranov, I. Cicenaite, and R.
Conclusion O. Esenaliev, “Influence of osmolytes on in vivo glucose monitoring
In conclusion, using two substantially different pressure lev- using optical coherence tomography,” Exp. Biol. Med. 231, 1323–
els, we showed in vivo that pressure applied to the skin by an 1332 共2006兲.
14. R. V. Kuranov, V. V. Sapozhnikova, D. S. Prough, I. Cicenaite, and R.
OCT probe reduces motion artifacts but may generate drift of O. Esenaliev, “Prediction capability of optical coherence tomography
the OCT signal slope during continuous noninvasive blood for blood glucose concentration monitoring,” J. Diabet. Sci. Technol.
glucose monitoring. Moreover, we reduced the drift four- to 1, 164–171 共2007兲.