You are on page 1of 10

Strabismus, 18(2), 48–57, 2010

Copyright © 2010 Informa UK Ltd.


ISSN: 0927-3972 print/ 1744-5132 online
DOI: 10.3109/09273971003758412

Fusional Vergence Measures and Their


Significance in Clinical Assessment
Fiona J. Rowe, PhD, DBO
Directorate of Orthoptics and Vision Science, University of Liverpool, UK
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

Abstract
Purpose: The aims of this study have been to evaluate differences between ranges in eso or exo
deviances and with varying target size.
Methods: Evaluation of sensory status in compensated deviations included measures of fusional
vergence (blur, break and recovery values) with randomization of target size and order of prism
base. Angle of deviation measures were assessed for each case.
Results: Twenty-two subjects, 7 esophoric and 15 exophoric, were recruited. Median prism fusional
vergence range at near fixation was 25BO to 10BI (central target); 35BO to 12BI (peripheral target).
At distance fixation median range was 16BO to 6BI (central); 25BO to 6BI (peripheral). Esophorics
For personal use only.

had a range shift towards the base out range. Exophorics had a shift towards the base in range.
Conclusions: Smaller positive fusional vergence values were obtained with small compared with
large target sizes. Additionally, esophoric subjects had a trend towards larger base out ranges while
exophoric subjects had a trend towards larger base in ranges.
Keywords:  fusion; vergence; strabismus; eso; exo

Introduction have varied, but generally the subjects measured have


had small horizontal latent deviations of less than 10
Motor fusional vergences are the ability to maintain prism diopters (PD) and ranges have been quoted
binocular vision through a range of induced ver- for either near or distance fixation. Other parameters
gence (often prism-induced) until a point is reached included method of prism fusion range measurement
where binocular vision is interrupted and the patient (prism bars, loose prisms, or Risley prisms), ocular
commonly appreciates diplopia. Induced vergence dominance (prism before dominant or non-dominant
results in a “stress” situation for vergence and an eye), and size of deviation.
initial response is made by fast fusional vergence to On the basis of these studies the normal fusional
realign the visual axes following a retinal disparity vergence range is generally quoted as 35 base-out
error message. Slow fusional responses then take (BO: convergent range) to 16 base-in (BI: divergent
over to further adapt to the ocular position.1 Normal range) at near fixation and 20 BO to 8 BI at distance
and abnormal fast and/or slow vergence systems are fixation.17 However, there are considerable differences
well reported in both asymptomatic and symptomatic reported in the various normative studies on fusional
subjects.2-4 vergence and in particular for positive fusional ver-
Data pertaining to the normal fusional vergence gence when measured with step fusion (prism bar).
range have been documented since 1948.5-16 The meth- The parameters that have been reported as not
ods of measurement of normal fusional vergences impacting on fusional vergence include age and gen-
Received 17 August 2009; Revised 21 January 2010; der.13 There is some controversy about the effect of
Accepted 27 January 2010 ocular dominance, as some studies show effect and
Correspondence: Dr Fiona Rowe, Division of Orthoptics, ­Thompson
others do not.8,13,14 However, these studies also state
Yates Building, University of Liverpool, Brownlow Hill, ­Liverpool that where effects are seen, these do not constitute
L69 3GB, UK. E-mail: rowef@liverpool.ac.uk clinically significant effect. Another confounder to
48
Significance of Fusion Measurements    49

ocular dominance is the way in which it is assessed. Inclusion criteria included a best corrected visual
Ocular dominance relates to where we see ourselves in acuity of 6/6 Snellens or 0.0 logMAR in either eye,
space. However, some studies have used preferred eye refractive error within ± 5DS/±2DC, esophoria or
viewing as an indicator of ocular dominance13,14 when exophoria within ± 10 prism diopters (measured by
in fact, this assessment is largely dependent on level prism cover test at near and distance fixation) with no
of acuity and thus does not always accurately indicate decompensation to intermittent strabismus, full ocular
ocular dominance. Measurement of fusional vergence rotations, presence of fusional vergence and stereop-
at near versus distance fixation also shows differences sis (TNO response of 60 seconds or arc or better), and
in values and again, this is notable for positive fusional visually asymptomatic.
vergence.5-9,13-16 There is no apparent information in the Exclusion criteria included diagnosis other than
literature regarding mid distance fusional vergence (ie, orthophoria, esophoria, or exophoria such as microtro-
3–meter distance). In view of changes in many testing pia, hyperphoria, and manifest strabismus. Individuals
environments where long distance rooms are often with abnormal ocular motility other than physiological
unavailable, there is a need to understand what dif- V pattern were also excluded.
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

ferences are seen in fusional vergence at mid distances Each subject had a full orthoptic evaluation in an
in comparison with the traditional values noted at near emmetropic state (wearing habitual refractive correc-
and distance fixation. tion, if required, to achieve inclusion criteria for visual
One aspect that does not appear to have been fully acuity). This included uniocular visual acuity (Snellens
addressed in the literature is the size of target that is or logMAR), cover test, ocular motility, retinal cor-
utilized when measuring fusional vergence. Many respondence (Bagolini glasses), stereoacuity (TNO),
studies state the type of target used.6,7,9,13,16 However, fusional vergence, accommodation near point, and
these target sizes vary considerably from one study to prism cover test. Measures of fusional vergence were
the next ranging from small, detailed targets (central undertaken by step prism fusion range using horizon-
stimulus) to large targets (peripheral stimulus). This tal (range 1 to 45 PD) and vertical (range 1 to 25 PD)
For personal use only.

leads to the question as to whether target size may prism bars (Haag Streit, UK) including: blur, break
influence the measurements obtained for fusional and recovery points to central (6/6:0.0 level), para-
vergence. foveal (6/9:0.2 level) and peripheral (6/60:1.0 level)
Type of deviation versus measured fusional ver- targets at near and distance fixation and at 3 meters.
gence does not receive much attention in the literature. The uncrowded target was a single Snellen letter and
This may be because inclusion criteria for studies limits the crowded target was a horizontal line of Snellen
the angle measurement to within ±10 prism dioptres. letters. The order of testing for these measures was
One early study stated that greater amounts of exopho- randomized in terms of BI versus BO range first, near
ria tend to accompany greater amounts of BI range and versus distance fixation first, horizontal versus vertical
the same applied to esophoria and BO ranges.5 That fusional vergence and choice of target. Previous studies
study’s data on normal vergence ranges however did have shown some variation in measures according to
not separate the deviation types. This raises the ques- these variables.9,18-20 Thus, we sought to minimize order
tion as to what differences can be seen when directly effect plus fatigue effect by randomization. Measure-
comparing fusional vergence of eso deviations to those ments of fusional vergence were undertaken during
of exo deviations. one testing session with short breaks of 2 to 3 minutes
The purpose of this prospective pilot study has been allowed between measures.
to (1) measure fusional vergence for blur, break, and From this study, one group with esophoria and one
recovery values to determine overall ranges with dif- group with exophoria were gender and age matched
fering target sizes; (2) evaluate differences in fusional to form subgroups for comparison.
vergence between esophoria and exophoria; and
(3) evaluate the fusional vergence values at a distance
of 3 meters for comparison with values obtained at Statistical Analysis
near and distance fixation.
As measurements taken with prism bars are not a
uniform step-wise assessment, non-parametric statis-
Methods tical evaluation is often recommended to take median
values into consideration. In this study the fusional
The study design was that of a prospective cross sec- vergence measures were assessed for goodness of fit
tion pilot study. Subjects were recruited from a general (Smirnov-Kolmogorov test) and for skewness. Results
university population and were naive to the study were not found to be standardly distributed with a
aims and measurements. skew of values to the lower prism strengths. Mean

© 2010 Informa UK Ltd.


50    F. J. Rowe

values were consistently higher than median values 23 years). Seven subjects had esophoria on cover test
ranging from a 0.5 to 5.5 prism diopter difference. and 15 had exophoria. The median prism cover test in
Thus non-parametric analysis was undertaken using prism diopters (PD) was −1 PD (-8 to +10 PD) at near
Wilcoxon test for related measures and Mann-Whitney fixation and 0 PD (−10 to +2 PD) at distance fixation
test for independent measures. Bonferroni adjustment (Figure 1).
was undertaken for multiple tests.
This research study adhered to the tenets of the Dec-
laration of Helsinki. Informed consent was obtained Varying Target Size
from the subjects and the research was approved by
the Local Research Ethics Committee. Table 1 shows the fusional vergence median and
ranges. The median horizontal prism fusion break
range at near fixation was BO 25 to BI 10 PD to a cen-
RESULTS tral (6/6) target and BO 35 to BI 12 PD to a peripheral
(6/60) target. At distance fixation this was BO 16 to
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

Twenty-two subjects were recruited; 7 males and 15 BI 6 PD to a central target and BO 25 to BI 6 PD to a


females. The mean age was 20 years (range was 19 to peripheral target.

A Near fixation

3
Number of subjects
For personal use only.

0
−8 −6 −4 −2 −1 2 4 6 10
Near fixation angle of deviation

B Distance fixation

5
Number of subjects

0
−10 −6 −4 −2 −1 0 1 2
Distance fixation angle of deviation

FIGURE 1  Distribution of angles of deviation.

Strabismus
Significance of Fusion Measurements    51

TABLE 1  Fusional vergence values are provided for the overall group including BO and BI ranges according to target size (central,
parafoveal, and peripheral) and for blur, break, and recovery points. The amplitudes (BO and BI ranges combined) are provided for
central and peripheral target sizes
NEAR Base out Base in
Target size Central Parafoveal Peripheral Central Parafoveal Peripheral
Blur 25[P]=0.002 25[P]=0.002 35 10[NS] 10[NS] 11
Break 25[P]=0.006 27.5[P]=0.002 35 10[NS] 10[NS] 12
Recovery 20[P]=0.007 22.5[P]=0.003 30 8[NS] 8[NS] 10
Amplitude of break point: 42.5 central; 50 peripheral
Vertical: Base down 5 to base up 4 PD
The p values relate to non-parametric assessment of fusional vergence measures with either a central or parafoveal target in
comparison with the measures obtained with a peripheral target (Wilcoxon test). NS denotes non-significant results

DISTANCE Base out Base in


Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

Target size Central Parafoveal Peripheral Central Parafoveal Peripheral


Blur 16[P]=0.029 16[P]=0.029 25 6[NS] 6[NS] 6
Break 16[NS] 17[NS] 25 6[NS] 6[NS] 6
Recovery 12[P]=0.036 14[NS] 16 4[NS] 4[NS] 4
Amplitude of break point: 28 central; 30 peripheral
Vertical: Base down 4 to base up 3 PD
The p values relate to non-parametric assessment of fusional vergence measures with either a central or parafoveal target in
comparison with the measures obtained with a peripheral target (Wilcoxon test). NS denotes non-significant results

MID 3m Base out Base in


Target size Central Peripheral Central Peripheral
Blur 18 25 6 8
For personal use only.

Break 20 25 6 8
Recovery 15 16 4 6

When comparing fusional vergence measures distance. There was also a significant difference com-
obtained with different sized targets at near and dis- paring BI values at near fixation to those at mid dis-
tance fixation there was no significant difference in the tance (p = 0.0001 central stimulus; p = 0.003 peripheral
measures obtained for many BI ranges using varying stimulus: Wilcoxon) but with little difference between
target sizes at the same testing difference. There was measures for distance fixation and mid distance.
a significant difference when comparing BI values at
near and at distance fixation distance using a central
target size (p = 0.0001) and using a peripheral target Esophoria Versus Exophoria Values
size (p = 0.0001): Wilcoxon. However for BO ranges, a
significantly higher measure was obtained when using From the overall group of 22 subjects, 7 subjects with
a peripheral target in comparison to a central or para- esophoria and 7 subjects with exophoria were taken to
foveal sized target at near fixation, p = 0.003 (Wilcoxon). form age and gender matched groups (4 females and
Additionally, there was a significant difference between 3 males in each group). The median angle of deviation
BO values from near to distance fixation with both a was 2 PD (2 to 10 PD) at near fixation and 1 PD (0 to 2
central target size (p = 0.004) and peripheral target size PD) at distance for esophoric subjects and -2 PD (-1 to
(p = 0.0001): Wilcoxon. -8 PD) at near fixation and -1 PD (0 to 6 PD) at distance
for exophoric subjects.
Table 2 shows the fusional vergence median and
Values at 3 Meters ranges for esophoria subjects. The median prism hori-
zontal fusion break range at near fixation was BO 30 to
The median prism fusion break range at mid fixation BI 10 PD to a central (6/6) target and BO 40 to BI 12 PD
was BO 18 to BI 6 PD to a central target and BO 25 to a peripheral (6/60) target. At distance fixation this
to BI 8PD to a peripheral target. A significant differ- was BO 25 to BI 6 PD to both central and peripheral
ence existed for comparison of near BO values to mid targets.
distance values with varying target sizes (p = 0.007 for When comparing fusional vergence measures
central stimulus and p = 0.0001 for peripheral stimu- obtained with different sized targets at near and dis-
lus: Wilcoxon) but not for distance fixation versus mid tance fixation there was no significant difference in

© 2010 Informa UK Ltd.


52    F. J. Rowe

TABLE 2  Fusional vergence values are provided for the eso group including BO and BI ranges according to target size (central,
parafoveal, and peripheral) and for blur, break, and recovery points. The amplitudes (BO and BI ranges combined) are provided for
central and peripheral target sizes
NEAR Base out Base in
Target size Central Parafoveal Peripheral Central Parafoveal Peripheral
Blur 30[P]=0.042 25[NS] 40 10[NS] 8[NS] 12
Break 30[P]=0.042 30[P]=0.043 40 10[NS] 10[NS] 12
Recovery 20[P]=0.039 25[P]=0.043 30 8[NS] 8[NS] 10
Amplitude of break point: 42 central; 52 peripheral
Vertical: Base down 5 to base up 3 PD
The p values relate to non-parametric assessment of fusional vergence measures with either a central or parafoveal target in
comparison with the measures obtained with a peripheral target (Wilcoxon test). NS denotes non-significant results

DISTANCE Base out Base in


Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

Target size Central Parafoveal Peripheral Central Parafoveal Peripheral


Blur 25[NS] 25[NS] 25 6[NS] 6[NS] 6
Break 25[NS] 25[NS] 25 6[NS] 6[NS] 6
Recovery 20[NS] 20[NS] 20 2[NS] 4[NS] 4
Amplitude of break point: 29 central; 33 peripheral
Vertical: Base down 4 to base up 3 PD
The p values relate to non-parametric assessment of fusional vergence measures with either a central or parafoveal target in
comparison with the measures obtained with a peripheral target (Wilcoxon test). NS denotes non-significant results

TABLE 3  Fusional vergence values are provided for the exo group including BO and BI ranges according to target size (central,
For personal use only.

parafoveal, and peripheral) and for blur, break, and recovery points. The amplitudes (BO and BI ranges combined) are provided for
central and peripheral target sizes
NEAR Base out Base in
Target size Central Parafoveal Peripheral Central Parafoveal Peripheral
Blur 20[P]=0.016 25[NS] 35 10[NS] 10[NS] 10
Break 25[P]=0.018 25[P]=0.046 35 10[NS] 10[NS] 14
Recovery 18[P]=0.028 16[P]=0.046 30 8[NS] 8[NS] 10
Amplitude of break point: 35 central; 49 peripheral
Vertical: Base down 5 to base up 4.5 PD
The p values relate to non-parametric assessment of fusional vergence measures with either a central or parafoveal target in
comparison with the measures obtained with a peripheral target (Wilcoxon test). NS denotes non-significant results

DISTANCE Base out Base in


Target size Central Parafoveal Peripheral Central Parafoveal Peripheral
Blur 14[NS] 12[NS] 16 8[NS] 8[NS] 8
Break 14[NS] 12[NS] 16 8[NS] 8[NS] 8
Recovery 10[NS] 10[NS] 12 6[NS] 6[NS] 6
Amplitude of break point: 28 central; 28 peripheral
Vertical: Base down 3 to base up 3 PD
The p values relate to non-parametric assessment of fusional vergence measures with either a central or parafoveal target in
comparison with the measures obtained with a peripheral target (Wilcoxon test). NS denotes non-significant results

the measures obtained for BI ranges. However for BO 14 PD to a peripheral (6/60) target. At distance fixation
ranges, a higher measure was obtained when using a this was BO 14 to BI 8 PD to a central target and BO 16
peripheral target in comparison with a central or para- to BI 8 PD to a peripheral target.
foveal sized target for near fixation only, p = 0.039 to When comparing fusional vergence measures
p = 0.043 (Wilcoxon). These higher measures were not obtained with different sized targets at near and dis-
found to be significant after Bonferroni correction. tance fixation there was no significant difference in
Table 3 shows the fusional vergence median and the measures obtained for BI ranges. However for BO
ranges for the exophoria group. The median prism ranges, a higher measure was obtained when using a
horizontal fusion break range at near fixation was BO peripheral target in comparison to a central or para-
25 to BI 10 PD to a central (6/6) target and BO 35 to BI foveal sized target for near fixation only, p = 0.016 to

Strabismus
Significance of Fusion Measurements    53

p = 0.046 (Wilcoxon). Again, these higher measures to BI 6 using a central target and BO 25 to BI 8 using
were not found to be significant after Bonferroni cor- a peripheral target. These were comparable to the
rection. distance ranges. Three-meter distance is not a routine
Figures 2 and 3 detail the prism fusion range at near clinical assessment distance for fusional vergence
and distance fixation to central and peripheral targets ranges. However where room constraints allow a
for the overall group plus the separate esophoria and 3-meter testing distance only, it is feasible to use these
exophoria groups. A shift towards the base-out range vergence ranges for comparison.
in seen in the esophoria group while a shift towards the A limitation of this study was the small number
base-in range is seen in the exophoria group. However, of subjects recruited. However, the study was under-
this shift was not found to be statistically significant. taken under repeatable and consistent environment
conditions using naive subjects. All measurements
were taken by the same examiner (FR). Measure-
DISCUSSION ments were taken of fusional vergence responses in
a randomized order to take into consideration both
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

The normal horizontal vergence range for all normal fatigue and practice effects. A larger-scale study is
subjects in this study was BO 25 to BI 10 using a central recommended to evaluate further the differences
target and BO 35 to BI 12 using a peripheral target at found in fusional vergence values in esophoria ver-
near fixation. At distance the normal vergence range sus exophoria and taking into account varying fixa-
was BO 16 to BI 6 using a central target and BO 25 tion distances and target sizes.
to BI 6 using a peripheral target. The normal vertical Table 4 outlines the step prism bar fusional ver-
vergence range was BD 5 to BU 4 for near fixation and gence values reported in the literature against the
BD 4 to BU 3 at distance fixation. values obtained in this current study. Scobee and
Additionally, horizontal vergence ranges were Green measured prism fusion vergence at both near
measured at 3-meter distance. The ranges were BO 18 (28 BO to 20 BI) and distance fixation (19 BO to 7 BI)
For personal use only.

A
A
PFR: Distance fixation (central stimulus)
PFR: Near fixation (central stimulus) 50
45 45
50 45 40
45
40 30
35 25 25
30 30
25 25 20
20 16 14
12 10
10 10 10 6 6 10
PD 0 PD 0 −4
−1 −4
−10 −6 −6 −6 −10 −6 −6 −8
−10 −10 −10
−20 −14 −20 −20 −12 −20
−30 −30
-30 −30
−40 Overall Eso Exo
Overall Eso Exo
B
B

PFR: Distance fixation (peripheral stimulus)


PFR: Near fixation (peripheral stimulus)

50 50
45 45 45 45 45
40 40
35 40 35 35
30 30
25
20 20 22
14 14 16 16
10 10 10
PD 0 PD 6 6
−6 −6 −6 0 −4 −4 −4
−10
−12 −12 −14 −10 −6 −6 −8
−20
−30 −20 −8 −20
−30 −30 −20
−40 −30
Overall Eso Exo Overall Eso Exo

FIGURE 2  Prism fusion range at near fixation for overall FIGURE 3  Prism fusion range at distance fixation for over-
normal range, eso and exo ranges. all normal range, eso, and exo ranges.

© 2010 Informa UK Ltd.


Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14
For personal use only.

54    F. J. Rowe

Table 4  Comparison of fusional vergence with varying target size


Rowe Scobee & Green Tait Wesson Rosenbaum et al Narbheram & Firth Antona et al
2009 1948 1949 1982 1995 1997 2008
6/6 target size Unknown target 1.4° target size at Parafoveal 6/6 target size 0.8 target size
size near target size
6/12 target size at
distance
Near Distance Near Distance Near Distance Near Distance Near Distance Near Distance
25 BO– 16 BO– 28 BO– 19 B– 19 BO– 11 BO– 27 BO– 57 BO– 33 BO– 28.91 BO– 23.25 BO–
10 BI 6 BI 20 BI O7 BI 13 BI 7 BI 8 BI 18.5 BI 9.7 BI 12.14 BI 8.63 BI
6/36 and 6/60 target 6/30 target size
size
Near Distance Near Distance
35 BO– 25 BO– 30 BO–
12 BI 6 BI 12 BI

Strabismus
Significance of Fusion Measurements    55

but with unknown target size.5 The near and distance target and overcome the prism effect; too much, and
ranges are midway between the ranges for this study the results can be overestimated.
for central and peripheral target sizes. One possible Target size appears to impact on the extent of
explanation is the use of a parafoveal target size. Tait fusional vergence values obtained and particularly
measured prism fusion vergence with a 6/30 periph- for positive fusional vergence. This finding warrants
eral target at distance fixation and obtained an average further study in a larger study series. Amplitudes
range of 18/30 BO to 10/12 BI.6 This is comparable of fusional vergence depend on the amount of fus-
with the range in this study with a similar target size. ible material in the subject’s field of view. Therefore
In this study prism fusion range was measured with the greater the amount of fusible material, the larger
a 6/36 sized target as well as a 6/60 target and found the amplitudes.21 This is not surprising considering
both target sizes gave the same measures as detailed the size of the fusible stimulus plus accommodative
in Table 1. convergence will play a greater role in fusional con-
Wesson et al. used a 1.4° target at near fixation and vergence at near than at distance fixation. There was
a 6/12 target at distance fixation and reported fusional a far greater difference in target size results for exo-
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

vergence of 19 BO to 13 BI for near and 11 BO to 7 phoria compared to esophoria, which would suggest
BI for distance.13 The distance values in particular are that exophoric subjects make far use of accommoda-
not that dissimilar to those found in our study with tive convergence in the control of their underlying
smaller target sizes. Rosenfield et al.. documented a deviation.
range of 27 BO to 8 BI using a parafoveal target at From a clinical viewpoint, certain aspects can
distance fixation.9 In 1997, Narbheram and Firth used be controlled. It is evident that the order of testing
a central target at both near (57 BO to 18.5 BI) and can impact on fusion vergence. The BO range in
distance fixation (33 BO to 9.7 BI).7 Their BO range particular can produce vergence/prism adaptation
was particularly high, which may relate to the subjects resulting in a transient fusional after-effect.18 This
used in the study (orthoptic undergraduate students). response may not have worn off by the time the BI
For personal use only.

The convergence range can be improved with long- range is subsequently measured.19 The result is that
term adaptation14,21 and this is likely to have influ- the fusional vergence range is biased in the direc-
enced and enhanced the results of their study. Hainey tion of the first measure. The authors recommend an
et al. used a peripheral target at distance fixation but order of testing such as BO, base-up, BI, and base-
placed before the dominant and non-dominant eye.8 down. For many patients it should be the procedure
They found a slightly increased BO range when placed adopted to prevent vergence adaptation. However
before the non-dominant eye but results were gener- where the deviation is known, i.e., eso or exo, a fur-
ally comparable with our distance peripheral target ther recommendation is that the compensating range
range. A more recent study in 2008 utilized a 0.8 deci- for the heterophoria should be measured first as it is
mal letter size and reported fusional vergence values this range that determines the quality of control for
of 28.91 BO to 12.14 BI for near fixation and 23.25 BO the deviation, e.g., BO range for an exo deviation.
to 8.63 BI for distance fixation.16 The BI values are very This would ensure that the critical vergence range is
similar to those reported in the literature for similar assessed first without introducing bias from a pre-
target sizes. The positive fusional vergence is slightly ceding vergence measure.9,18-20
larger, however, for distance fixation and more com- In addition to testing the critical range first the
parable to that of peripheral sized target values. true vergence range for horizontal deviations is of
It would appear from the current study and much importance. Scobee and Green reported their ranges
of the literature that the early prism fusional vergence of fusional vergence.5 At a near testing distance they
values from 1948 generally remain consistent across found the amount of exo deviation to be significantly
various studies. Where there are some discrepancies, correlated with the prism divergence break point and
these undoubtedly relate to varying methodology commented that greater amounts of exo deviation
across studies, namely biased/unbiased subjects, accompanied greater amounts of prism divergence
underlying angle of deviation, target size, prisms while greater amounts of eso deviation accompanied
placed before one or both eyes, and placement of prism lower amounts of prism divergence. This was also
(prentice versus frontal plane). In addition the time seen at distance testing with greater measurements at
allowed to fixate the target, use of dominant or non- distance fixation having greater prism vergence at near
dominant eye, and order of testing (BO before or after fixation.
BI range) will impact on fusional vergence measures The current study found a clear difference between
and typically these factors are specified in individual fusional vergence for eso versus exo deviations with
studies. This study allowed 2 seconds per prism: too a greater BO range for esos and greater BI range for
little, and the subject has insufficient time to fixate the exos. Amplitudes were also compared and found to

© 2010 Informa UK Ltd.


56    F. J. Rowe

be skewed according to the type of deviation present. attain deviations at 6 m within +/−5 prism diopters
There appears to be a skewed vergence range with eso- and at 1/3 m within 10 prism diopters of exophoria
phoria having a bias towards the convergent range and and 4 prism diopters of esophoria to have comfort-
exophorias having a bias towards the divergent range. able binocular control. Tait discussed this fact in his
The differences found in this study did not reach sig- paper, stating that subjects use about one quarter to
nificance, which may reflect the small numbers in each one third of their vergence amplitude to control their
group and the variance across groups. This warrants deviation successfully.6 Criteria exist that can aid the
further investigation in a larger cohort of esophoric prediction of angle of deviation with presence/absence
versus exophoric subjects. of symptoms. Sheard’s criteria have been reported to
It is proposed that these findings have potential be a good discriminatory tool with exo deviations
clinical significance. Patients who have small angle whereas eso deviations were better predicted by a
latent deviations with fusional vergence measures fall- variant of Percival’s criteria or degree of deviation
ing within these normal ranges are considered most measurement.23,24
likely to be asymptomatic with regard to control of the
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

heterophoria.6 Clinically however we often see patients


with poor fusional vergences who are asymptomatic CONCLUSIONS
or patients with normal fusional vergences who are
symptomatic.15,22 In addition it is also evident that eso The results of this study demonstrate that the differ-
deviations can have different vergence responses to ences between fusional vergence values reported in
exo deviations. the literature appear less when target size is taken
Numerous parameters contribute to the success or into consideration indicating a potential influence of
failure of treatment for eso and exo deviations. For central versus peripheral stimulus on fusional ampli-
example, refractive error, presence or absence of sen- tude. Subjects with eso deviations show larger BO and
sory and motor binocular function, angle of deviation, smaller BI vergences than exo deviations although a
For personal use only.

and level of visual acuity. Binocular function is of par- study with a larger sample size is required to confirm
ticular importance and specifically the motor fusional whether this shift is of statistical significance.
vergences. It is important to fully assess binocular
vision in symptomatic patients22 despite the presence
of apparently normal fusional vergence values. Atten- ACKNOWLEDGMENT
tion must be paid to the target size being used to obtain
the fusional vergence values so that accurate compari-
son of vergence measures can be made to the angle of
deviation, particularly when considering the extent of Declaration of Interest:  No funding was allocated to
binocular control of the deviation. this paper. The author reports no conflicts of interest.
This information is important when considering The author alone is responsible for the content and
what values are being used when determining the writing of the article.
patient’s use of fusional vergence in control of the
patient’s angle of deviation. It is recommended that
the patient fixate a target of size commensurate with REFERENCES
his or her visual acuity when measuring their angle of
deviation. The same target size should then be used 1. Schor CM. The relationship between fusional vergence eye
when measuring fusional vergence. Subsequently like movements and fixation disparity. Vision Res. 1979;19:1359–
1367.
with like can be compared. If not, erroneous interpreta- 2. Carter DB. Fixation disparity and heterophoria following
tion of ability to control a deviation may occur. prolonged wearing of prisms. Am J Optom Arch Am Acad
In addition, it may be important to take angle Optom. 1965;42:141–151.
of deviation into consideration when considering 3. Pickwell LD, Kurtz BH. Lateral short-term prism adaptation
whether fusional vergence measures are normal for in clinical evaluation. Ophthalmic Physiol Opt. 1986;6:67–73.
4. North RV, Henson DB. The effect of orthoptic treatment
the individual. Considering so much of the literature upon the vergence adaptation mechanism. Optom Vis Sci.
relates binocular outcome success as to within +/− 10 1992;69:294–299.
prism diopters of ortho it is imperative that we ques- 5. Scobee RG, Green EL. Relationships between lateral hetero-
tion our outcome measures for future studies for those phoria, prism vergence and the near point of convergence.
deviations with binocular potential. Am J Ophthalmol. 1948;31:427−441.
6. Tait EF. Fusional vergence. Am J Ophthalmol. 1949; 32:
Scobee and Green stated that patients with large 1223−1230.
esophoria and poor BI range would be symptomat- 7. Narbheram J, Firth AY. Prism fusion range: blur point, break
ic.5 They recommended that the aim should be to point and recovery point. Br Orthopt J. 1997;54:2−6.

Strabismus
Significance of Fusion Measurements    57

8. Hainey J, Cleary M, Wright L. Does ocular dominance horizontal fusional vergences. Ophthalmic Physiol Opt.
influence the clinical measurement of fusional amplitude? 2008;28:475−491.
Br Orthopt J. 1999;56:72−76. 17. Von Noorden GK, Campos EC. Binocular vision and ocular
9. Rosenfield M, Ciuffreda KJ, Ong E, Super S. Vergence motility. Theory and Management of Strabismus. 6th edition.
adaptation and the order of clinical vergence range testing. Mosby, 2002.
Optom Vis Sci. 1995;72:219−223. 18. Sethi B. Vergence adaptation: a review. Documenta Ophthal-
10. Jimenez R, Perez MA, Garcia JA, Gonzalez MD. Statisti- mologica. 1986;63:247−263.
cal normal values of visual parameters that characterize 19. Morgan MW. The clinical aspects of accommodation
binocular function in children. Ophthalmic Physiol Opt. and convergence. Am J Optom Arch Am Acad Optom.
2004;24:528−542. 1944;21:301−313.
11. Sethi B, North RV. Vergence adaptive changes with 20. Tuff LC, Firth AY, Griffiths HJ. Prism vergence measure-
varying magnitudes of prism induced disparities and ments following adaptation to a base out prism. Br Orthopt
fusional amplitudes. Am J Optom Physiol Opt. 1987;6: J. 2000;57:42−44.
151−156. 21. Goldstein JH, Hornblass A, Clahane AC. Effect of peripheral
12. Carter DB. Effects of prolonged wearing of prisms. Am J vision on the binocular relationship: a further study. Am J
Optom Physiol Opt. 1963;40:265−273. Ophthalmol. 1968;66:86.
Strabismus Downloaded from informahealthcare.com by The University of Manchester on 12/01/14

13. Wesson MD. Normalization of prism bar vergences. Am J 22. Gall R, Wick B. The symptomatic patient with normal pho-
Optom Physiol Opt. 1982;59:628−634. rias at distance and near: what tests detect a binocular vision
14. Wesson MD, Amos JF. Norms for hand-held rotary prism problem? Optometry. 2003;74:309.
vergences. Am J Optom Physiol Opt. 1985;62:88−94. 23. Sheedy JE, Saladin JJ. Association of symptoms with mea-
15. Scheiman M, Herzberg H, Frantz K, Margolies M. A norma- sures of oculomotor deficiencies. Am J Optom Physiol Opt.
tive study of step vergence in elementary school children. 1978;55:670−676.
J Am Optom Assoc. 1989;60:276−280. 24. Sheedy JE, Saladin JJ. Phoria, vergence and fixation dis-
16. Antona B, Barrio A, Barra F, Gonzalez E, Sanchez I. parity in oculomotor problems. Am J Optom Physiol Opt.
Repeatability and agreement in the measurement of 1977;54:474−478.
For personal use only.

© 2010 Informa UK Ltd.

You might also like