Technical Bulletin Issued by: Randox Laboratories Ltd.

55 Diamond Road, Crumlin, County Antrim, BT29 4QY.

Daytona Plus – Main Unit - Software Upgrade May 2016

Technical Bulletin No.: RXTB-0052 Issue Date: 16 May 2016


Affected Analysers: Daytona Plus
Importance: HIGH
Items Required: None
Specialist Tools/Software: Unit Main Ver.2550535115.zip, FFFTP, Telnet.
Software Update Required: Yes

Purpose:
To notify End Users of Daytona Plus analysers of a software upgrade to the analyser main unit,
introduced to correct issues and optimise performance.

Updates:
1. Correction of work Wash Pump Unit (WPP) and Sample Pump Unit (SPP) hour counters to
start at 300H.
2. Correction of the liquid level detection sequence to enable detection overflow at WU3
during the wash cycle.
3. Correction of the software to switch OFF the lamp after analyser software is closed using the
‘Shift+Ctrl+.’ key combination.
4. Wash programs have been optimized to improve washing performance.
5. Operating range of Wash Unit (WU) height has been extended from ±40 to ± 50.

Implementation:
1. End Users should contact a Local Service Provider to arrange for this upgrade to be done.
2. Local Service provider should contact Randox Technical Services to obtain the software and
instructions for this upgrade.

Technical Bulletin Number: RXTB-0052 FORM NO: 8133-EN


Daytona Plus – Main Unit - Software Upgrade May 2016 REVISION (1)
Page 1 of 1 20th January 2016