ElectroSurgical Unit DT-300S/400S SPECIFICATION

No. Factor DAIWHA PRODUCT SPECIFICATION


120(60Hz) or 230VAC(±10%) (50 or 60Hz)/switchable
1 Mains and Line Frequency
See the label on rear side for correct information

Two of F5.0A, 250V for 230VAC


2 Fuses
Two of F10.0A, 250V for 115 VAC

3 Power Consumption 950VA ±10%


4 Protection Class Class 1, Type CF
5 IPX Foot Switch: IPx6
6 Carrier Frequency 400, 482KHz
7 Repeat Frequency 33KHz
8 Low Freq. Leakage Current EN60601-1 [2012] Requirement
9 High Freq. Leakage Current Less than 150mA
10 Operation Conditions 10℃~40℃, 30%~ 85% noncondensing, 700mbar~1060mbar
11 Storage Conditions -10℃~60℃, 30%~ 95% noncondensing, 700mbar~1060mbar
-10℃~60℃, 30%~ 95% noncondensing, 700mbar~1060mbar
12 Transportation Conditions Do not drop the unit from higher than 50cm
Do not use the hook
13 Installation Conditions 5cm clearance required on each side unit for cooling
14 Cooling 1 fan mounted
15 Recommended Duty Cycle 10 sec ON, 30 sec, OFF(Output for 10 seconds, pause for 30 seconds )
16 Size (H, W, D) 147mm x 330mm x 417mm
17 Weight About 16Kg
Not AP/APG device
18 APG Equipment not suitable for use in the presence of a flammable anesthetic mixture
with or with oxygen or nitro oxide
19 Operation Sound Difference in operation sounds between cutting and coagulation.
20 Operation Display Visible difference in operations between cutting and coagulation.
21 Output Setting Displayed in [W] unit from Seven Segment (LED number displayer)