Changing Oil Mist Filter in 8800 ICP-MS/MS
 Replacing the cartridge and o-rings in oil
mist filter used in rough pump DS402 of
Agilent’s 8800 ICP-MS/MS.
 Cartridge and o-rings part number:
9499342M002.

George L. Donati
Changing Oil Mist Filter in 8800 ICP-MS/MS

1. Remove all 4 screws on top


of the oil mist filter.

2. Remove the spring on


top of the filter cartridge.

3. Replace the o-ring.


Changing Oil Mist Filter in 8800 ICP-MS/MS

4. Remove the filter cartridge.

5. Remove the cylinder


after removing all 4
screws at the bottom of
the oil mist filter.
Changing Oil Mist Filter in 8800 ICP-MS/MS

6. Replace the second o-ring.

7. Now you can reverse


the process and put it all
back together.