Overview
User interface
Communication ports
On/Off switch
Expiratory Inspiratory filter
Fan filter
8 |
Touch Screen Navigation
• Swipe
• Double-tap
• Drag
• Touch and hold
• Drag and drop
• “Touch, turn, accept”
9 |
Setting Up the Puritan Bennett™
980 Ventilator Prior to Use
*U.S. only
*only disposable inspiratory and expiratory filters may be used in the U.S.
*in the U.S. the EVQ has to be cleaned and disinfected in between patients
Power on
using the
ON/OFF
switch
• You can choose to 1. Select PBW or Gender and Height 1. Touch New
set up a new patient 2. Select ventilation type—Invasive or
Patient.
Same patient The ventilator will be ready to start ventilating at the settings in
place at power down.
• To enable/disable the
Proximal Flow Option
– Touch the Configure/Wrench icon
– Touch the Options tab
– Touch the Prox tab
– Touch the Enabled or Disabled
button.
• To complete a Manual Purge,
push the Start button
How the Puritan Bennett™ Proximal Flow
Sensor Impacts Monitoring
When the Proximal Flow Sensor is enabled, new and replacement data
values appear, representing data measured with the Proximal Flow Sensor.
The monitored volume labels have a Y added to indicate that the
measurement comes from the Proximal Flow Sensor.
Inspired tidal volume (mandatory or spontaneous)
VTI Y at patient circuit wye
Exhaled spontaneous/mandatory tidal volume at patient
VTE Y circuit wye
Exhaled spontaneous tidal volume at patient circuit wye
VTE SPONT Y
Exhaled mandatory tidal volume at patient circuit wye
VTE MAND Y
Exhaled total minute volume at patient circuit wye
VE TOT Y
Touch a parameter in
the lower margin of the
touch screen.
• Touch the Vent Setup button to access the Apnea settings tab.
• Apnea alarm triggers apnea backup:
• Current apnea ventilation settings are displayed.
• Non-apnea ventilation settings may be changed during apnea backup.
• Apnea timer resets with every breath.
• Autoreset of apnea backup occurs when the patient triggers two consecutive
inspirations and the exhaled volume is equal to or greater than 50% of the
delivered volume. To manually reset it, touch the Alarm Reset key.
• Apnea ventilation setting for inspiratory pressure or tidal volume is also used for
manual inflations; it is displayed in the Vent Setup button.
Normal
• Steadily lit green
mode
• LED flashes—color
Alarm
corresponds to alarm
condition priority
Concurrent • LED displays
alarms highest-priority color
• Latched (sides)
display highest
Alarm
priority
de-escalates • Center displays
current alarm priority
The Alarm Log contains the last 1000 alarms that have occurred,
whether they have been reset or autoreset, the priority level, and
their analysis messages.
• Accessible during Normal and • If one or more alarms have
Service states occurred since the last time
the alarm log was viewed,
• Records:
a yellow triangle appears on
• Date/time-stamped entry when
an alarm is detected, escalated, the touch screen indicating
reset or autoreset there are unread items.
• Date/time-stamped entry if the
• Alarm log is cleared when
ventilator enters backup ventilation
• Priority level
New Patient is selected during
• Analysis message the ventilator startup process.
• Alarm silence interval begins,
ends or is cancelled
Status display
• Additional patient
data screen
• Large-font patient
data screen
• More settings:
– Leak Sync: Enabled/
Disabled
– O2 sensor: Enabled/
Disabled/Calibrate
– Humidification type: Heated
Exp Tube/Non-heated Exp
Tube/HME (impacts
expiratory spirometry)
More Settings
Setup:
1. Touch the Menu tab on the left side of the touch screen.
2. Touch the Stand-by button. A stand-by state pending dialog appears instructing
the clinician to disconnect the patient circuit. An on-screen timer allows 30 seconds
for disconnect.
3. After you disconnect, you must reconfirm your intent by touching the Confirm
button within 30 seconds.
4. Flow sensors are monitored to detect patient reconnection.
5. To exit stand-by—reconnect the patient to the ventilator.
• The Inspiratory Pause There are two ways to access this function, which
maneuver is used to can be done in two ways, automatically or manually:
determine plateau 1. Access via Menu tab:
pressure, static compliance
a) Automaticà touch RM key à touch
and static resistance. Inspiratory Pause à touch Start à touch
• It takes place at the end of Accept or Reject
the inspiratory phase of a b) Manual à touch RM key à touch Inspiratory
breath. Pause à touch and hold Start for up to 7
seconds à touch Accept or Reject
• Breathing efforts could
skew the measurement so 2. Access via bezel key:
it is important to ensure a) Automatic à touch and release Inspiratory
that the patient is not Pause bezel key à touch Start à touch
Accept or Reject
actively breathing when
you perform the maneuver. b) Manual à touch and hold Inspiratory Pause
bezel key for up to 7 seconds à touch Start
à touch Accept or Reject
Home icon (house) Dismisses all open dialogs on the touch screen.
Resumes showing ventilator waveforms screen.
Help icon (question Drag the help (question mark) icon to the item
mark) in question and release. A tooltip will appear
describing the item’s function.
1. Display Brightness
2. Display Lock
3. Alarm Volume
4. Manual Inspiration
5. Adjustment Knob
6. Inspiratory Pause
Easy to Access 7. Expiratory Pause
8. Alarm Reset
9. Alarm Silence
71
| Covidien Respiratory & Monitoring Solutions | April 14, 2014
Additional Safety Features