HF 525Plus Installation, Operation and Service Manual

9.6 Collimator

Adjustable with manual control knobs, exclusive use for


Application
radiography
Power supply 24V AC
Maximum voltage 150kVp
Lamp LED lamp 24V AC 150W
Inherent Filtration 1.5mm Al
Added Filtration 0.8mm Al
Maximum X-ray Field size Below 5cm x 5cm at SID 100cm
Minimum X-ray Field size Below 35cm x 35cm at SID 65cm
Light projection More than 160lux at SID 100Cm
Lamp timer Approx. 30sec (Push button type)
Dimension Approx. 185 x 220 x 180mm
Weight Approx. 5Kg

9.7 Table Bucky (4-Way)

Type Wide 4 way floating table top

Compatible cassette sizes inch 8"x 10"∼17"x17"

Longitudinal table top movement Approx. ±150mm

Transverse table top movement Approx. ±400mm

Movement of Bucky Device Approx. 400mm

Maximum load of tabletop Approx. 250kg

Inherent Filtration 0.8mm Al

Grid specification Ratio: 10:1, 103lines/inch

X-ray IMAGE RECEPTOR Film cassette OR DR Detector

Cassette loading From front side

Tabletop locking Electromagnetic lock by optical sensor

Dimension Approx. 2,200 x 800 x 690mm

Weight 140Kg

Power supply 24V AC

SID Under 80mm (Distance between detector and table top)

184
Technical Publication Document NO: SYSTEM-007-KOR (Ver.1.1)