You are on page 1of 13

Analytica Chimica Acta 445 (2001) 255–267

The analysis of alkylphenol ethoxysulphonate surfactants by


high-performance liquid chromatography, liquid
chromatography–electrospray ionisation–mass spectrometry and
matrix-assisted laser desorption ionisation–mass spectrometry
Salim H. Benomar, Malcolm R. Clench∗ , David W. Allen
Division of Chemistry, School of Science and Mathematics, Sheffield Hallam University, Howard Street, Sheffield S1 1WB, UK
Received 25 January 2001; received in revised form 12 July 2001; accepted 12 July 2001

Abstract
Commercial nonylphenol ethoxysulphonate and octylphenol ethoxysulphonate surfactant formulations, used for enhanced
oil recovery, have been analysed by high-performance liquid chromatography (LC), liquid chromatography–electrospray
ionisation–mass spectrometry (LC–ESI–MS) and matrix-assisted laser desorption ionisation–mass spectrometry (MALDI–MS).
Mixed-mode C18/SAX and C8/SAX columns were used for both liquid chromatography and liquid chromatography–mass
spectrometry analyses. Matrix-assisted laser desorption ionisation–mass spectra were obtained using either ␣-cyano-4-
hydroxycinnamic acid or 2,5-dihydroxybenzoic acid as matrix with the addition of lithium chloride to simplify the mass spec-
tra obtained. Data obtained from each method indicate that the nonylphenol ethoxysulphonate formulation has an ethoxymer
(EO) chain length ranging from 2 to 13 EO units with an average of 6.26. This is in broad agreement with earlier studies,
although the range was reported as 2–15. However, the data obtained suggest that the octylphenol ethoxysulphonate formu-
lation has an ethoxymer chain length ranging from 1 to 8 EO units with an average chain length of 3.67. This is in contrast to
earlier studies carried out by LC only, which suggested that the chain length ranged from 2 to 6 EO units with an average of
3.6. A method for the extraction of alkylphenol ethoxysulphonates from sea water using graphitised carbon black solid-phase
extraction cartridges has also been developed. © 2001 Elsevier Science B.V. All rights reserved.
Keywords: Enhanced oil recovery; Surfactants; High-performance liquid chromatography; Mass spectrometry; MALDI

1. Introduction high-salinity brine, but their surface-active proper-


ties (in particular the reduction of interfacial tension
Anionic and non-ionic surfactants have been widely (IFT)) are not generally as good as those of anionic
used in enhanced oil recovery (EOR) processes [1]. surfactants. Anionic surfactants have been the most
Non-ionic surfactants have been used primarily as extensively used type of surfactant in EOR since they
co-surfactants in order to improve the behaviour of have good surfactant properties, are relatively stable,
surfactant systems. They are much more tolerant of exhibit relatively low adsorption on reservoir rock,
and can be manufactured economically. However, the
∗ Corresponding author. Tel.: +44-114-225-3054;
very low cost anionic surfactants such as petroleum
fax: +44-114-225-3066. sulphonates cannot be used for off-shore reservoirs
E-mail address: m.r.clench@shu.ac.uk (M.R. Clench). where sea water is the only source of injection

0003-2670/01/$ – see front matter © 2001 Elsevier Science B.V. All rights reserved.
PII: S 0 0 0 3 - 2 6 7 0 ( 0 1 ) 0 1 2 8 0 - 6
256 S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267

column is the most suitable for the separation of alkyl


sulphate and alkane sulphonate surfactants due to
short analysis times and high peak efficiencies.
Austad and Fjelde [3] have reported a LC
method for the separation of nonylphenol ethoxy-
sulphonate (NPEOS) and octylphenol ethoxy-
sulphonate (OPEOS) surfactants at sea water salinity
using two different mixed-mode reversed-phase/ion-
exchange columns (C8/SAX and C18/SAX columns).
They were able to separate the different APEOS
oligomers up to a degree of ethoxylation of at least
15. Furthermore, the separate oligomers of uncon-
verted ethoxylated alcohols could be qualitatively
detected and impurities of the type R–(EO)–R could
be analysed in one chromatographic determination.
They concluded that LC was an excellent quality test
for commercial APEOS.
Liquid chromatography–mass spectrometry (LC–
MS) is a widely used method for trace-level quan-
tification of charged, non-volatile compounds [7]. It
Fig. 1. Structures of (a) nonylphenol ethoxysulphonate, (b) has been widely applied to the analysis of trace or-
octylphenol ethoxysulphonate and (c) a typical alkyl aryl ganic compounds in water [8–11] and many papers
sulphonate. reporting its use for the determination of surfactants
in water have appeared. The application of LC–MS to
water, since they are not sufficiently salt tolerant. A the characterisation of surfactants and their biodegra-
recent extensive publication has described the use of dation intermediates has been extensively reviewed
surfactants in the petroleum industry [2]. by Di Corcia [12]. However, at present there appear
Alkylphenol ethoxylated sulphonates (APEOS) to be no reports of the analysis of either APEOS
(Fig. 1(a) and (b)) are a class of anionic surfactants. formulations or their determination in environmental
When the ethoxymer (EO) chain length of an APEOS matrices by LC–MS.
is greater than three it is completely soluble in sea In contrast to the widespread use of LC–MS there
water [3]. The development of APEOS and also alkyl have been only a few papers published on the analy-
aryl sulphonates (Fig. 1(c)) that are able to tolerate sis of surfactants by matrix-assisted laser desorption
both hard water and high salinities is therefore of ionisation–mass spectrometry (MALDI–MS). Just
great interest to the oil industry. Mixtures of these et al. [13] have compared the use of MALDI–MS with
surfactants can be used for the chemical flooding of supercritical fluid chromatography (SFC) for molar
offshore oil reservoirs with the use of sea water as the mass determination of some alkylphenol ethoxylate
injection fluid. Mixing the expensive APEOS with the (APEO) surfactants. Thomson et al. [14] have re-
less expensive alkyl aryl sulphonates leads to a reduc- ported the acquisition of MALDI–MS spectra for
tion in the cost of chemical flooding in comparison to three classes of surfactant — non-ionic, anionic and
the use of APEOS alone [4]. cationic. Parees et al. [15] have briefly compared
High-performance liquid chromatography (LC) is electrospray, FAB and MALDI ionisation techniques
the most widely used technique in surfactant anal- for the analysis of some commercial nonylphenol
ysis. Shamsi and Danielson [5,6] have compared ethoxylate (NPEO) surfactants. Their findings seemed
the separation of long-chain alkyl sulphates and to indicate that the data obtained from the three
alkane sulphonates on three mixed-mode (RP8, RP4 methods are comparable for surfactant formulations
and RPphenyl/anion-exchange) stationary phases. with lower average molecular distributions. However,
They suggested that the RPphenyl/anion mixed-mode at higher average molecular weights, although the
S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267 257

electrospray and MALDI results were similar, the USA). ␣-Cyano-4-hydroxycinnamic acid and 2,5-
FAB data exhibited distinctly lower molecular weight dihydroxybenzoic were purchased from Aldrich
distributions than electrospray and MALDI, possibly (Poole, Dorset, UK). All solvents used and ammo-
due to fragmentation. nium acetate and acetic acid were LC grade and were
Another comparison has been made between purchased from Fisher Scientific (Loughborough,
MALDI, reversed-phase LC and thin-layer chro- UK).
matography (TLC) for the analysis of non-ionic sur-
factants. In this work, Cumme et al. [16] compared 2.2. Sample pretreatment
the molecular weight information given by MALDI,
TLC (using mass spectrometry to analyse the frac- In order to simplify the mass spectra and chro-
tions) and LC. The results from these experiments did matograms obtained from the surfactant formulations,
not show any major discrepancy between the average residual non-ionic starting material was selectively
molecular weight found by LC and TLC with that removed by the use of SAX solid-phase extraction
found by MALDI. However, the authors state that for cartridges. The surfactant solutions were acidified
MALDI, ethoxymers with masses below 405 Da were to pH 3–4 with glacial acetic acid. The cartridge
not included in the results because of interferences was washed with methanol (5 ml) and distilled water
from matrix ions. (10 ml). The acidified surfactant solution was passed
We have recently reported a method for the through the cartridge which was subsequently washed
determination of nonylphenol ethoxylates in sur- with 2% acetic acid in methanol (5 ml) to selectively
face water by MALDI–MS [17]. Following extrac- elute the non-ionic surfactants. The cartridge was
tion/preconcentration by C18 solid-phase extraction, then washed with methanol (5 ml) before elution of
the sample was mixed with a standard MALDI ma- the anionic fraction in 2 M HCl in methanol:water
trix (2,5-dihydroxy benzoic acid) with the addition of (50:50) (15 ml). This was evaporated to dryness and
LiCl. The limit of detection was 40 ␮g/l based on a the now “pure” NPEOS and OPEOS were dissolved
250 ml sample. in the appropriate solvent for further analysis.
In this paper we present the application of simi-
lar MALDI–MS methodology to the analysis of com- 2.3. Extraction and preconcentration
mercial nonylphenol ethoxysulphonate and octylphe-
nol ethoxysulphonate formulations and critically com- NPEOS and OPEOS were dissolved in synthetic
pare the data obtained with that obtained from both sea water or reagent water (as appropriate) prior to
LC with fluorescence detection and LC–MS data ob- extraction/preconcentration by solid-phase extraction
tained via the use of an electrospray interface. (SPE). SPE was performed using a GCB cartridge
method developed by Di Corcia et al. [18]. The
cartridges were first washed with dichloromethane
2. Experimental (DCM) (50 ml), methanol (MeOH) (50 ml) and ace-
tonitrile (ACN) (50 ml) and left to stand in ACN
2.1. Reagents and materials overnight. The ACN was removed and the washing
was continued with 5 mmol/l of TMAOH in 90:10
A commercial NPEOS formulation manufactured (v/v) DCM:MeOH (7 ml), MeOH (3 ml) and acidic
by Hoescht and an OPEOS formulation manufac- water (pH 2, conc. HCl) (30 ml) before the sea water
tured by PPG industries were gifts from Dr. Tor samples were introduced slowly. Prior to introduc-
Austad (Rogaland University Centre, Stavanger, tion the pH of the sample was adjusted to 3. After
Norway). Synthetic sea water was purchased from the samples had been passed through, the cartridges
BDH (Poole, Dorset, UK). Graphitised carbon were eluted with distilled water (7 ml), MeOH (2 ml),
black (GCB) cartridges (500 mg) were purchased (70:30 (v/v) DCM:MeOH) eluent A (7 ml), (25 ml/l
from Supelco (Poole, Dorset, UK). Strong anion- formic acid in 90:10 DCM:MeOH) eluent B (7 ml)
exchange (SAX) cartridges (200 mg) were pur- and finally (5 mmol/l of TMAOH in 90:10 (v/v)
chased from Varian (Frampton Avenue, Harbor City, DCM:MeOH) eluent C (7 ml). The fractions obtained
258 S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267

Table 1
Gradient elution program for the separation of commercial alkylphenol ethoxysulphonate surfactants using mixed-mode C8/SAX and
C18/SAX columns
Time (min) Flow rate (ml/min) Solvent Aa (vol.%) Solvent Bb (vol.%) Gradient curve

0 1 60 40 6
20 1 70 30 6
30 1 80 20 6
40 1 60 40 6
a Acetonitrile.
b Distilled water containing 0.1 M ammonium acetate (pH 4.7).

with eluent C were evaporated to dryness under a 2.6. MALDI–MS


steady steam of N2 gas and redissolved in 1 ml LC
mobile phase (eluents A and B were discarded). All MALDI analyses were performed on a Finni-
gan Vision 2000 reflectron-based time-of-flight
2.4. High-performance liquid chromatography mass spectrometer. The NPEOS and OPEOS
(6.8 g/l) were dissolved in methanol: water (50:50).
␣-Cyano-4-hydroxycinnamic acid and 2,5-dihydroxy-
All LC analyses were performed on a Waters
benzoic acid (90 mg/ml) were dissolved in 0.1%
600-MS gradient LC system. The columns used were
TFA in methanol. Lithium chloride (10 mg/ml) was
Alltech (Carnforth, Lancashire, UK) mixed-mode
dissolved in methanol. An amount of 200 ␮l of the
reversed-phase/ion-exchange columns, mixed-mode
NPEOS and OPEOS solutions, 40 ␮l of matrix and
RP8/anion 100A, 7 ␮m (150 mm × 4.6 mm) and
10 ␮l of LiCl were mixed and 1 ␮l of the resulting
mixed-mode RPC18/anion 100A, 5 ␮m (150 mm ×
solution was placed on the stainless steel target of the
4.6 mm). Fluorescence detection was carried out us-
instrument for analysis.
ing a Jasco FP-920 (Cheltenham, UK) fluorescence
detector, λex = 225 nm and λem = 295 nm. Data
were output to Hewlett-Packard HP 3396A integrator.
3. Results and discussion
Injections (100 ␮l) were made using a Perkin-Elmer
ISS-101 autosampler. The mobile-phase gradient used Austad and Fjelde [3], and Fjelde and Austad [4]
is shown in Table 1. All analyses of NPEOS were com- describe three LC methods based on mixed-mode
pleted using the C8/SAX column described above and reversed-phase/ion-exchange column of C4, C8 and
all analyses of OPEOS using the C18/SAX column. C18 types for the analysis of APEOS and alkyl aryl
sulphonates. In the present work in order make the
2.5. Liquid chromatography–mass spectrometry mixed-mode separation compatible with the electro-
spray interface of the Quattro I mass spectrometer
All analyses were performed on a Quattro I (Micro- (which is not compatible with the use of phosphate
mass, Manchester, UK), mass spectrometer, equipped buffers), potassium dihydrogen phosphate (KH2 PO4 )
with an electrospray ionisation (ESI) source. A Jasco (as employed by Austad and Fjelde), was replaced
Pu-980 intelligent LC pump system was used in this by ammonium acetate as buffer solution. Ammo-
case, with all other LC conditions as described above. nium acetate is a good choice as a buffer system for
The mass spectrometer was operated in positive ion LC–MS since it is (i) chemically stable, non-toxic,
mode. A capillary voltage of 3.5 kV and skimmer cone inexpensive and readily available; (ii) highly soluble
voltage of 30 V were used. The source temperature in methanol and acetonitrile; (iii) relatively volatile
was maintained at 80◦ C. Nitrogen was used as the dry- and easily removed when used in conjunction with
ing and nebulising gas. Mass spectra were collected a mass spectrometer; and (iv) an excellent masking
in full-scan mode, scanning over the range m/z = agent for residual silanol groups in chromatographic
100–1200 at 3 s per scan. columns [19].
S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267 259

chromatograms for NPEOS and OPEOS obtained us-


ing the ammonium acetate buffer system and C8/SAX
and C18/SAX mixed-mode columns, respectively.
The method developed by Di Corcia et al. [18]
for the extraction of aromatic surfactants and their
biodegradation intermediates from raw and treated
sewage worked successfully when applied to the ex-
traction of APEOS from both reagent and sea water.
The recovery data for the extraction of NPEOS (at
two different concentrations) and OPEOS from each
of these matrices are shown in Table 2. As can be seen
excellent and reproducible recoveries were obtained
for both surfactants from distilled and sea water. The
reasons for the slightly poorer reproducibility of the
surfactants from distilled water is unknown but may
be due to the possibility of the surfactants existing
in differing ionic forms in distilled water. These are
the first data reported showing the extraction of these
surfactants from sea water by SPE and the good sensi-
tivity and recoveries do suggest that SPE followed by
LC with fluorescence detection would be the method
of choice for the determination of these compounds
in surface water samples.
Fig. 2. Separation of a commercial NPEOS formulation manufac- The MALDI–MS of NPEOS acquired using
tured by Hoescht achieved on a C8/SAX mixed-mode LC column ␣-cyano-4-hydroxycinnamic acid is shown in Fig. 4.
using the gradient program described in Table 1, fluorescence de- As can be seen, and as we reported previously
tection λex = 225 nm and λem = 295 nm.
for nonylphenol ethoxylates [17], the addition of
lithium chloride to surfactant samples results in rela-
A secondary (and unexpected) benefit of the change tively simple MALDI–MS. In this case the spectrum
of buffer system was a reduction in the retention times for NPEOS exhibits predominantly an envelope of
for the individual ethoxylate oligomers, in compar- [M − Na + 2Li]+ adduct ions at m/z = 429.2, 473.3,
ison to the published separations [3,4], without any 517.7, 561.4, 605.9, 649.3, 692.5, 737.3, 781.6, 825.1,
adverse effect on their chromatographic resolution. 869.0 and 912.7, with a minor contribution from
Thus, the analyses of nonylphenol ethoxysulphonate [M + Na]+ ions at, e.g. m/z = 577. Hence, these
and octylphenol ethoxysulphonate were achieved with data suggest that the surfactant formulation contained
the simple ammonium acetate buffer system and this molecules with between 2 and 13 EO units.
was employed for all subsequent LC and LC–MS The MALDI–MS spectrum of OPEOS acquired us-
analyses. Figs. 2 and 3 show typical LC fluorescence ing 2,5-dihydroxybenzoic acid as matrix, Fig. 5, shows

Table 2
Recoveries of NPEOS and OPEOS from sea water and reagent water using GCB SPE cartridges and extraction method of Di Corcia [18]a
Type of surfactant Type of water Added (ppm) Found (ppm) Recovery (%) R.S.D. (%)

NPEOS Distilled water (three replicates) 1.50, 0.61 1.52, 0.60 101, 98 6.9, 1.25
NPEOS Sea water (three replicates) 1.51, 0.61 1.62, 0.63 106, 103 0.96, 0.08
OPEOS Distilled water (three replicates) 1.56 1.55 99 8.55
OPEOS Sea water (three replicates) 1.51 1.56 103 1.86
a For full experimental conditions, see text.
260 S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267

Fig. 3. Separation of a commercial OPEOS formulation manufactured by PPG Industries achieved on a C18/SAX mixed-mode LC column
using the gradient program described in Table 1, fluorescence detection λex = 225 nm and λem = 295 nm.

an envelope of intense [M − Na + 2Li]+ adduct ions observed at m/z = 478, 522, 566, 610, 654, 698, 742,
at m/z = 371, 415, 459, 503, 548, 592, 636 and 680. 786, 830 and 874.
Hence, these data suggest that the surfactant formula- In the electrospray LC–MS analysis of OPEOS
tion contained molecules with 1–8 EO units. the ethoxymers eluted from t r = 9.70–19.95 min
The electrospray positive ion LC–MS base peak in- (Fig. 8(a)–(i)). The mass spectra obtained (Fig. 9(a)–
tensity chromatogram of NPEOS is shown, along with (h)) also exhibit ions corresponding to [M − Na +
selected mass chromatograms, in Fig. 6(a)–(k). Using H + NH4 ]+ adducts and these were observable at
the chromatographic conditions employing the ammo- m/z = 377, 421, 465, 509, 553, 597, 641 and 685.
nium acetate buffer system the NPEOS ethoxymers Comparisons of the results obtained from the three
eluted from retention time 18.39 to retention time different techniques (LC, MALDI–MS, LC–MS) for
31.63 min. Fig. 7(a)–(j) shows the mass spectra ob- both surfactants are given in Tables 3 and 4. An ap-
tained for each of the ethoxymers. Interestingly, the proximation of the mole fraction of each ethoxymer
spectra for the NPEOS oligomers in Fig. 7 show that indicated by each technique was calculated measur-
the peaks produced have m/z values corresponding to ing the peak height of individual ethoxymers in the
NH4 + adducts, the sodium presumably being com- LC, LC–MS and MALDI data and quoting it as a
pletely displaced by the use of the ammonium ac- percentage of the total height. The average number
etate in the mobile phase. Ions corresponding to [M − of ethoxymers was then calculated as a weighted av-
Na+H+NH4 ]+ adducts for NPEOS ethoxymers were erage based on these data. This is obviously a very
S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267 261

Fig. 4. Positive ion MALDI–MS of a commercial NPEOS formulation from Hoescht showing the range of EO units from n = 2 (m/z = 429)
to n = 13 (m/z = 913) with a weighted average chain length of 6.32.

Fig. 5. Positive ion MALDI–MS of a commercial OPEOS formulation from PPG Industries showing the range of EO units from n = 1
(m/z = 371) to n = 8 (m/z = 680) with a weighted average chain length of 3.42.
262 S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267

Fig. 6. (b)–(k) Individual mass chromatograms for the [M − Na + H + NH4 ]+ ions obtained from each oligomer and (a) the base peak
intensity mass chromatogram for the positive ion electrospray LC–MS analysis of a commercial NPEOS formulation from Hoescht.
Separation was achieved on a mixed-mode C8/SAX column using the mobile-phase gradient described in Table 1. The peaks corresponding
to the 3 EO (t r = 31.32 min) to 12 EO (t r = 18.32 min) containing oligomers are clearly visible in this chromatogram.
S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267 263

Fig. 7. (a)–(j) Electrospray positive ion mass spectra for the individual oligomers of a commercial NPEOS formulation from Hoescht
separated using a mixed-mode C8/SAX column using the mobile-phase gradient described in Table 1. The spectra exhibit predominantly
[M − Na + H + NH4 ]+ ions and range from the 3 EO units containing oligomer (a) to the 12 EO units containing oligomer (j).
264 S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267

Fig. 8. (b)–(i) Individual mass chromatograms for the [M − Na + H + NH4 ]+ ions obtained from each oligomer and (a) the base peak
intensity mass chromatogram for the positive ion electrospray LC–MS analysis of a commercial OPEOS formulation from PPG Industries.
Separation was achieved on a mixed-mode C18/SAX column using the mobile-phase gradient described in Table 1. The peaks corresponding
to the 1 EO (t r = 19.88 min) to 8 EO (t r = 9.70 min) containing oligomers are visible in this chromatogram.
S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267 265

Fig. 9. (a)–(h) Electrospray positive ion mass spectra for the individual oligomers of a commercial OPEOS formulation from PPG
separated using a mixed-mode C18/SAX column using the mobile-phase gradient described in Table 1. The spectra exhibit predominantly
[M − Na + H + NH4 ]+ ions and range from the 8 EO units containing oligomer (a) to the 1 EO unit containing oligomer (h).
266 S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267

Table 3
Comparison of the approximate mole fractions of individual ethoxymers of a NPEOS formulation manufactured by Hoescht calculated
from LC, LC–MS and MALDI–MS data
Number of EO units LC (%) LC–MS (%) MALDI–MS (%)

1 0.00 0.00 0.00


2 0.77 0.00 3.43
3 7.51 4.1 5.65
4 15.25 12.80 9.70
5 18.78 17.49 16.56
6 18.34 18.65 20.80
7 15.45 16.58 17.58
8 11.05 15.02 11.52
9 8.18 8.42 7.07
10 2.87 5.18 3.64
11 1.10 1.55 2.22
12 0.44 0.13 1.21
13 0.22 0.00 0.61
Weighted average number of ethoxymers 6.00 6.46 6.32

Table 4
Comparison of the approximate mole fractions of individual ethoxymers of a commercial OPEOS formulation manufactured by PPG
industries calculated from LC, LC–MS and MALDI–MS data
Number of EO units LC (%) LC–MS (%) MALDI–MS (%)

1 1.66 5.79 8.31


2 6.60 13.51 29.87
3 35.16 29.54 28.57
4 30.77 22.23 16.36
5 15.93 16.23 9.35
6 6.59 10.23 5.19
7 2.19 0.96 1.82
8 1.10 1.50 0.50
Weighted average number of ethoxymers 3.87 3.72 3.14

approximate way of handling the data in terms of cal- cussed by Parees et al. [20] who found that the agree-
culating accurate mole ratios of individual ethoxymers ment between molecular weight distributions obtained
but does enable a simple comparison between the data from mass spectral data and size exclusion experi-
sets to be made. This method does not take into ac- ments for a range of surfactants was very compound
count any differences in the relative fluorescence emis- dependent.
sion, electrospray or MALDI ion yields of individual As can be seen from Table 3, for NPEOS, the data
ethoxymers. This could only be obtained by use of obtained for average length of the ethylene oxide
pure standards of each and may account for the differ- chain (EO number) are in reasonable agreement with
ences in the mole ratios of individual ethoxymers ob- values of 6.00 being obtained from LC, 6.46 from
tained from each technique. The differences in mole LC–MS and 6.32 from MALDI. In earlier studies
percentages of ethoxymers obtained by the use of the conducted by LC only the average EO number was
differing analytical techniques in the absence of pure determined as 6.19 [3]. However, in our studies the
standards does beg the question as to what the “true” range of EO units found by LC, i.e. from 2 to 13 and
values actually are and which technique gives the best also those found by LC–MS (3–12) and MALDI–MS
representation. These issues have been extensively dis- (2–13) are narrower than that reported in the earlier
S.H. Benomar et al. / Analytica Chimica Acta 445 (2001) 255–267 267

LC work of 2–15. This accounts for the slightly lower developed which gives excellent recoveries from both
weighted average EO number determined by LC in our reagent and sea water at ppm levels.
experiments.
For OPEOS the average number of EO units cal-
culated by each technique is in broad agreement; LC Acknowledgements
(3.87), LC–MS (3.72) and MALDI (3.14) and these
data are also in agreement with the earlier studies car- The authors thank Dr. Tor Austad and Dr. Ingebret
ried out by LC only which gave an average EO num- Fjelde for the kind gift of the surfactants studied in
ber of 3.6 [3]. However, in our studies the range of this work. This work was supported by a grant from
EO units for the OPEOS was found to be from 1 to the Petroleum Research Centre, Tripoli, Libya.
8 and this is not in agreement with the earlier work
carried out by LC only where the range was reported References
to be from 2 to 6 EO units. The discrepancy does ap-
pear to arise from the assignment of peaks in the LC [1] D.W. Green, G.P. Willhite, Enhanced Oil Recovery, Textbook
chromatogram with Austad and Fjelde [3] appearing Series, Vol. 6, Society of Petroleum Engineers, Dallas, 1998.
to have misidentified the peaks in their tentative as- [2] L.L. Schramm (Ed.), Surfactants Fundamentals and Applica-
tions in the Petroleum Industry, Cambridge University Press,
signment. The misidentification results from the lack
Cambridge, 2000.
of pure ethoxymer standards (Austad and Fjelde are [3] T. Austad, I. Fjelde, Anal. Lett. 25 (1992) 957.
in fact cautious in their peak assignment in the paper) [4] I. Fjelde, T. Austad, Colloids Surf. A: Physicochem. Eng.
and it is a clear demonstration of the advantage of the Aspects 82 (1994) 85.
use of mass spectrometry either as MALDI or LC–MS [5] S.A. Shamsi, N.D. Danielson, J. Chromatogr. Sci. 33 (1995)
505.
in the analysis of polymers.
[6] S.A. Shamsi, N.D. Danielson, Chromatographia 40 (1995)
237.
[7] D.D. Popanoe, S.J. Morris, P.S. Horn, K.T. Norwood, Anal.
4. Conclusions Chem. 66 (1994) 1620.
[8] C.J. Miles, D.R. Doerge, S. Bajic, Arch. Environ. Contam.
Toxicol. 22 (1992) 247.
LC, LC–ESI–MS and MALDI–MS techniques have
[9] L.B. Clark, R.T. Rosen, T.G. Hartmann, L.H. Alaimo, J.B.
all been shown to be useful tools for the analysis of Louis, C. Hertz, C.T. Ho, J.D. Rosen, Res. J. Water Pollut.
commercial NPEOS and OPEOS surfactant formula- Control. Fed. 63 (1991) 104.
tions. Data obtained from each method indicate that [10] I.A. Kim, F.I. Sasinos, D.K. Rishi, R.D. Stephens, M.A.
the NPEOS formulation has an ethoxymer chain length Brown, J. Chromatogr. 589 (1992) 177.
[11] M.A. Brown, R.D. Stephens, I.S. Kim, Trends Anal. Chem.
ranging from 2 to 13 EO units with an average of ap-
10 (1991) 330.
proximately 6.26 (calculating an average of the val- [12] A. Di Corcia, J. Chromatogr. A 794 (1998) 165.
ues obtained from each technique). This is in broad [13] U. Just, H.R. Holzbauer, M. Resch, J. Chromatogr. 667 (1994)
agreement with earlier studies although the range of 354.
ethoxymer units was reported in these to be from 2 [14] B. Thomson, Z. Wang, A. Paine, A. Rudin, G. Lajoie, JOACS
72 (1995) 11.
to 15 EO units. The data for the OPEOS formula-
[15] D.M. Parees, S.D. Hanton, D.A. Willcox, P.A. Cornelic-Clark,
tion suggests that it has an ethoxymer chain length Polym. Prep. 37 (1992) 321.
ranging from 1 to 8 EO units with an average chain [16] G.A. Cumme, E. Blume, R. Blublitz, H. Happe, A. Horn, J.
length of 3.58. This is in contrast to earlier studies car- Chromatogr. 791 (1997) 245.
ried out by LC only, which suggested that the chain [17] M. Willetts, M.R. Clench, R. Greenwood, G. Mills, V.A.
Carolan, Rapid Commun. Mass Spectrom. 13 (1999) 251.
length ranged from 2 to 6 EO units with an average
[18] A. Di Corcia, R. Samperi, A. Marcomini, Environ. Sci.
of 3.6. Technol. 28 (1994) 850.
A method for the extraction of alkylphenol ethoxy- [19] C.K. Lim, T.J. Peters, J. Chromatogr. 316 (1984) 397.
sulphonates from water using graphitised carbon [20] D.M. Parees, S.D. Hanton, P.A. Cornelio-Clark, D.A. Willocx,
black solid-phase extraction cartridges has also been J. Am. Soc. Mass Spectrom. 9 (1998) 282.

You might also like