You are on page 1of 9

La radiologia medica (2020) 125:365–373

https://doi.org/10.1007/s11547-020-01179-x

CHEST RADIOLOGY

Proposal of a low‑dose, long‑pitch, dual‑source chest CT protocol


on third‑generation dual‑source CT using a tin filter for spectral
shaping at 100 kVp for CoronaVirus Disease 2019 (COVID‑19) patients:
a feasibility study
Andrea Agostini1,2   · Chiara Floridi1,2   · Alessandra Borgheresi2   · Myriam Badaloni2 · Paolo Esposto Pirani2 ·
Filippo Terilli2 · Letizia Ottaviani2 · Andrea Giovagnoni1,2

Received: 9 March 2020 / Accepted: 18 March 2020 / Published online: 1 April 2020
© Italian Society of Medical Radiology 2020

Abstract
Aim  To subjectively and objectively evaluate the feasibility and diagnostic reliability of a low-dose, long-pitch dual-source
chest CT protocol on third-generation dual-source CT (DSCT) with spectral shaping at 100Sn kVp for COVID-19 patients.
Materials and methods  Patients with COVID-19 and positive swab-test undergoing to a chest CT on third-generation
DSCT were included. The imaging protocol included a dual-energy acquisition (HD-DECT, 90/150Sn kVp) and fast, low-
dose, long-pitch CT, dual-source scan at 100Sn kVp (LDCT). Subjective (Likert Scales) and objective (signal-to-noise
and contrast-to-noise ratios, SNR and CNR) analyses were performed; radiation dose and acquisition times were recorded.
Nonparametric tests were used.
Results  The median radiation dose was lower for LDCT than HD-DECT (Effective dose, ED: 0.28 mSv vs. 3.28 mSv,
p = 0.016). LDCT had median acquisition time of 0.62 s (vs 2.02 s, p = 0.016). SNR and CNR were significantly different
in several thoracic structures between HD-DECT and LDCT, with exception of lung parenchyma. Qualitative analysis dem-
onstrated significant reduction in motion artifacts (p = 0.031) with comparable diagnostic reliability between HD-DECT
and LDCT.
Conclusions  Ultra-low-dose, dual-source, fast CT protocol provides highly diagnostic images for COVID-19 with potential
for reduction in dose and motion artifacts.

Keywords  COVID-19 · 2019-nCOV · Dual-source CT · Spectral shaping · Low-dose CT · Chest radiology

Introduction clinical symptoms are fever and cough in addition to other


unspecific symptoms including, fatigue, dyspnea, muscle
In December 2019, a pneumonia of unknown origin out- soreness and headache [3]. Intriguingly, a small percentage
break in Wuhan, Hubei province (China); the responsible (5%) case is asymptomatic (i.e., with normal body tempera-
pathogen was identified as the novel coronavirus (2019- ture or minor discomfort) [4, 5], while reverse-transcription
nCOV), and the related pulmonary syndrome was named polymerase chain reaction (RT-PCR) from swab samples
as COVID-19 (CoronaVirus Disease 2019) by the World has demonstrated high specificity but relatively low sen-
Health Organization (WHO) [1, 2]. Common presenting sitivity (60–70%) [6, 7]. Moreover, when comparing RT-
PCR to chest CT, the latter demonstrated a better sensitivity
in particular in early stages [6, 7]. Therefore, the RT-PCR
* Alessandra Borgheresi
from swab samples is still the standard of reference in the
alessandra.borgheresi@gmail.com
diagnosis of COVID-19, while unenhanced, high-resolution
1
Department of Clinical, Special and Dental Sciences, chest computed tomography (CT) has a central role in detec-
University Politecnica delle Marche, Ancona, AN, Italy tion, diagnosis and follow-up of the disease [8–10].
2
Division of Special and Pediatric Radiology, Department CT is a widely available technique allowing for high-
of Radiology, University Hospital “Umberto I – Lancisi – quality and standardized evaluation of the lung parenchyma.
Salesi”, Via Conca 71, 60126 Ancona, AN, Italy

13
Vol.:(0123456789)

366 La radiologia medica (2020) 125:365–373

However, radiation exposure and motion artifacts are major Infectious Disease, referred to the Department of Radiol-
issues in uncooperative, young patients undergoing to ogy for a chest high-resolution CT between Feb 24, 2020
repeated CT examinations [11]. To reduce motion artifacts and March 4, 2020, with no basal chest X-ray performed,
in uncooperative patients (e.g., pediatric patients or patients were prospectively included to be scanned on the third-
with dyspnea), fast acquisitions are obtained by lowering the generation DSCT (Somatom Force, Siemens Health-
rotation time of the tube-detector system with high pitch and ineers). Anamnestic and clinical information were col-
wide collimation values [12]. Conversely, lower radiation lected. Patients with pneumonia other than COVID-19, or
doses are achieved at low kV with the implementation of with contraindication to CT were excluded.
iterative reconstructions for noise reduction [13]. The dual-
source CT scanners (DSCT, Siemens Healthineers, Erlan-
gen, Germany) are equipped with two asymmetrical tube- Image acquisition and reconstruction
detector systems (i.e., different scan Field of View, FOV),
mounted in the gantry with an offset of ~ 90°. The two tube- The imaging protocol on the third-generation DSCT
detector systems work at different kVp settings for dual- (Somatom Force, Siemens Healthineers, Erlangen) was
energy acquisitions or at the same kVp setting for ultrafast composed by a spiral high-resolution dual-energy acqui-
acquisitions at long pitch (Pitch 1.5–3, Flash or Turbo Flash sition (HD-DECT) and by an ultra-low-dose acquisition
mode, Siemens Healthineers, Erlangen) [14, 15]. Moreover, (LDCT) in deep inspiration when possible; no contrast
the X-ray tubes in the second- and third-generation scan- material was administered. Since COVID-19 pathologi-
ners (respectively, the Somatom Flash and Force, Siemens cal mechanism is not completely understood, and other
Healthineers, Erlangen) have additional tin filtration [16, strains of coronavirus family demonstrated similar pul-
17]. In particular, the more aggressive tin filtration in the monary syndromes with development of fibrosis (i.e., air
Somatom Force provides the best spectral separation in trapping); the first three patients were evaluated also with
dual-energy (DECT) acquisitions by increasing the mean an ultra-low-dose, fast scan in deep expiration [20, 21].
energy of high-kVp spectrum (i.e., 150Sn kVp) [17]. More- A relatively fast DECT, spiral, caudocranial acquisi-
over, the spectral shaping with tin filter (i.e., 100Sn kVp) tion (HD-DECT) was set with the following parameters:
allows for reduction in the low-energy component of the 90/150Sn kVp, modulated mA (CareDose 4D, Siemens
X-ray spectrum, leading to significant dose reduction [18, Healthineers, Erlangen, Germany) with reference 85 mAs,
19]. Coupling an ultra-low-dose, fast, long-pitch dual-source rotation time 0.25 s, a relatively long pitch of 1.1 and a
acquisition with spectral shaping may be of relevant value wide collimation (2 × 192 × 0.6  mm). In the ultra-low-
in serial evaluations in dyspneic or coughing patients with dose, fast acquisition (LDCT), both the tubes worked at
COVID-19. 100 kVp with 0.6-mm tin filter (100Sn kVp), with a wide
The aim of this work is to test the feasibility, with subjec- collimation (2 × 192 × 0.6 mm), a rotation time of 0.25 s,
tive and objective analysis, of an ultra-low-dose, fast, long- an ultra-long pitch (pitch = 3, TurboFlash mode, Siemens
pitch, dual-source acquisition on third-generation DSCT Healthineers), with modulated mA at 180 mAs reference.
(Somatom Force, Siemens Healthineers, Erlangen) for the Both the HD-DECT and LDCT datasets were recon-
lung evaluation in patients affected by COVID-19 related structed with the available iterative reconstruction algo-
pneumonia. rithm ADMIRE, Strength 4 (Advanced Modeled Iterative
Reconstruction, Siemens Healthineers, Erlangen). The
lung parenchyma was reconstructed with sharp kernels
(HD-DECT: Bl64; LDCT: Bl57), with linear blending of
Materials and methods 0.7 at slice thickness/spacing of 1.5/1 mm and a window/
level of 1200/-600 HU (named Lung Images, LUNG). The
Ethical standards mediastinal structures were evaluated with softer kernel
(Br40 for both acquisitions) with linear blending of 0.7,
This study was approved by the local Ethical Board, and the slice thickness/spacing of 3/1.5 mm and a window/level of
informed consent was not collected in written form because 350/50 HU (named Mediastinal Images, MED). Sagittal
of the specific disease. and coronal reconstructions were obtained.
DECT datasets were reconstructed with different linear
blending ratios (Blending Ratio 0.2) in order to reduce
Patient population beam-hardening artifacts in uncooperative patients on a
dedicated workstation (Syngo.via VA20, Dual Energy,
Inpatients > 18  years old, positive for COVID-19 of Siemens Healthineers, Erlangen).
the upper respiratory tract swab from the Division of

13
La radiologia medica (2020) 125:365–373 367

Radiation Dose evaluation and Acquisition Time 1. Major anatomical structures (fissures and small vessels);
2. Small anatomical structures (bronchi < 2 mm and sep-
The CT dose index (CTDIvol) and dose length product tae);
(DLP) for each scanned patient were recorded. The effec-
tive dose (ED) was calculated using a conversion factor In both cases, the Likert scale was defined as follows:
(k) of 0.0145 mSv × mGy−1×cm−1 for 120 kV according to 1 = unacceptable (landmarks not visible); 2 = poor (< 25%
Deak et al. [22] based on the following formula: landmarks visible); 3 = fair (25–75% landmarks visible);
4 = good (> 75% landmarks visible); 5 = excellent (all
ED = DLP × k landmarks visible).
Finally, the acquisition (exposure) times for HD-DECT and Finally, a third radiologist (10  years of experience)
LDCT were recorded. detected main lung pathological findings related to
COVID-19 on HD-DECTLUNG and HD-DECTMED images
and the other two radiologists were asked to evaluate
Subjective image analysis the same finding on the ultra-low-dose acquisitions, in
particular:
Two independent radiologists with 15 and 10  years of
experience evaluated the quality of the CT images on a 1. Ground glass opacities (GGO) (< 20 mm);
PACS workstations (Picture Archiving and Communica- 2. COVID-19 signs, e.g., reverse halo/crazy paving;
tion System; Centricity Radiology, GE Healthcare, Mil- 3. Centrolobular nodules;
waukee), after removing personal and technical data from 4. Consolidations (segmental–subsegmental);
the images. The lung parenchyma and airways were eval- 5. Lymph nodes (on HD-DECTMED);
uated on Lung Images, while the mediastinal structures
(e.g., lymph nodes) were evaluated on the Mediastinal The relative Likert scale was set as follows: 1 = finding
Images, in HD-DECT and LDCT. not detected; 2 = barely detected, unreliable interpretation;
The subjective analysis was performed by using 3 = visible finding with marked blurring and uncertain
5-points Likert scale and included three sections: general interpretation; 4 = visible finding, blurred, with no influ-
quality of the image (mediastinum and lung images), ana- ence on diagnosis; 5 = finding clearly visible with good
tomical structures (lung images), pathological findings demarcation.
(lung or mediastinal images).
The evaluation of general quality included the following
parameters and scales: Objective image analysis

1. Sharpness, where sharpness was rated as: 1 = unaccep- Signal-to-noise ratio (SNR) and contrast-to-noise ratio
table, 2 = significantly reduced with blurring of adja- (CNR) were evaluated by the third radiologist (10 years of
cent structures, 3 = minimally reduced sharpness with experience) by placing 1 ± 0.05 cm2 circular Region of Inter-
blurring of adjacent structures, 4 = sharpness minimally est (ROI) for the following structures:
reduced, 5 = excellent sharpness;
2. Perceived image noise: 1 = unacceptable, 2 = above aver- 1. SNR:
age, 3 = average image, 4 = less than average, 5 = mini- a. HD-DECTLUNG and L ­ DCTLUNG: lung parenchyma,
mal; trachea and muscle and subcutaneous fat;
3. Motion artifacts: 1 = unacceptable, non-diagnostic; b. HD-DECTMED and ­LDCTMED: trachea, descending
2 = marked; 3 = mild; 4 = minimal; 5 = no motion arti- aorta, paraspinal muscle, subcutaneous fat;
facts
4. Subjective diagnostic reliability: 1 = no diagnostic sig- 2. CNR (the selected reference was the subcutaneous fat):
nificance; 2 = poor; 3 = fair; 4 = good; 5 = optimal diag-
nostic reliability; a. HD-DECTLUNG and L ­ DCTLUNG: lung parenchyma,
5. Overall image quality, including other artifacts—beam- trachea, paraspinal muscle;
hardening artifacts: 1 = unacceptable, 2 = fair, 3 = mod- b. HD-DECTMED and L ­ DCTMED: descending aorta,
erate, 4 = good, 5 = excellent. trachea, paraspinal muscle.

The evaluation of anatomic structures included the SNR and CNR were calculated considering average and
parameters: standard deviation (SD) of the HU from the placed ROI as
follows [13]:

13

368 La radiologia medica (2020) 125:365–373

|Average HUAnatomical Structure | Results


SNR = | |
SDAnatomical Structure
Patient Population and radiological findings on CT
|Average HUAnatomical Structure − Average HUfat |
CNR = | | In this study were included 10 patients (M/F = 7/3) with a
SDfat median age of 53 (IQR: 47–83) and a median BMI of 28
(IQR: 26–30). Two of 10 patients (20%) had a BMI > 30
and 3/10 (39%) had severe symptoms (Temperature > 38 °C,
dyspnea and respiratory failure) and were not able to main-
Statistical analysis
tain the arms raised during the CT examination. The radi-
ological findings of each patient at presentation are sum-
Quantitative parameters were expressed as median and
marized in Table 1 and are comparable to those previously
interquartile ranges (25–75 p, IQR). Qualitative and quan-
reported [8].
titative parameters were compared with nonparametric
tests (Wilcoxon for paired samples). A dedicated statistical
Radiation dose and acquisition time
software was used (MedCalc v19.1.6, MedCalc Software,
Ostend, Belgium).
Table 2 summarizes the applied radiation dose and acquisi-
tion time. Statistically significant differences were found in
radiation dose and acquisition times. The use of spectral
shaping at 100 kV (100Sn kVp) allowed for a significant
reduction in ED, with a median dose reduction of 90.6%
(− 88.35 to − 91.12%).

Objective image analysis


Table 1  Main radiological findings in 10 COVID-19 positive patients
Radiological finding Number of The results of the objective image quality analysis are shown
patients: 10 in Table 3. Statistically significant differences were found in
SNR and CNR, with higher values in HD-DECT images, for
Ground glass opacities 10 (100%)
different anatomical structure with exception of lung paren-
Multiple and Bilateral 10 (100%)
chyma (p > 0.05).
with consolidation 7 (70%)
Linear opacities 7 (70%)
Subjective image analysis
Rounded opacities 2 (20%)
Peripheral involvement 10 (100%)
The results of the subjective image evaluation assessment
Central parenchyma involvement 1 (10%)
are summarized in Table 4. On the basis of the average
“Crazy-Paving” pattern 3 (30%)
Likert scale, the two radiologists assessed for the over-
“Reverse Halo” Sign 2 (20%)
all subjective image quality a median point value of 4
Bronchial Wall Thickening 5 (50%)
for both the protocols. As expected, the subjective image
Bronchiectasis 4 (40%)
noise had a trend to be worse in the LDCT. However,
Lymphadenopathy 0
the LDCT demonstrated a significant reduction in the
Bilateral lung involvement 9 (90%)
motion artifacts when compared to the standard DECT
> 2 lobes affected 8 (8%)
(4 vs. 5) (Fig. 1a, b). Nevertheless, in terms of diagnostic
Air trapping 0

Table 2  Radiation dose and HD-DECT median (25 - 75 p) LDCT median (25 - 75 p) P*


exposure times
CTDIvol (mGy) 6.38 (3.91–7.51) 0.64 (0.47–1.12) 0.016
DLP (mGy × cm) 226.21 (176.01–322.03) 19.5 (17.5–29.02) 0.016
ED (mSv) 3.28 (2.55–4.67) 0.28 (0.25–0.42) 0.016
Exposure time (s) 2.02 (1.81–2.36) 0.62 (0.54–0.72) 0.016

CTDI CT dose index; DLP dose length product; ED effective dose


*Wilcoxon test for paired samples

13
La radiologia medica (2020) 125:365–373 369

Table 3  Objective image analysis: signal-to-noise (SNR) and contrast-to-noise (CNR) ratios


HD-DECTLUNG LDCTLUNG P* HD-DECTMED LDCTMED P*
Median Median Median Median
(25–75 p) (25–75 p) (25–75 p) (25–75 p)

SNR
Descending aorta // // – 4.06 (3.44–5.51) 1.84 (1.72–2.22) 0.016
Lung parenchyma 13.49 (12.76–17.48) 12.2 (9.83–14.21) 0.578 // // –
Trachea 24.52 (16.67–40.29) 19.9 (14.33–20.84) 0.219 29.38 (12.64–94.91) 41.93 (19.34–56.52) 0.038
Subcutaneous fat 2.37 (1.34–2.67) 0.98 (0.93–1.22) 0.016 10.53 (9.85–12.15) 4.82 (4.43–6.03) 0.016
Muscle 0.65 (0.46–0.77) 0.44 (0.25–0.52) 0.016 3.83 (1.13–4.14) 1.7 (0.82–2.25) 0.047
CNR
Descending aorta // // – 14.35 (12.93–16.29) 6.66 (6.11–8.61) 0.016
Lung parenchyma 15.63 (8.48–18.89) 7.54 (76.82–9.68) 0.057 // // –
Trachea 17.4 (9.35–20.21) 8.21 (7.56–10.29) 0.047 78.68 (73.25–93.24) 40.70 (35.17–53.28) 0.016
Muscle 3.00 (1.72–3.87) 1.48 (1.33–1.74) 0.016 15.21 (11.81–15.42) 7.18 (6.56–7.26) 0.016

Table 4  Subjective image HD-DECT LDCT P*


analysis with Likert scales (see Median (25–75 p) Median (25–75 p)
text)
General evaluation
Image sharpness 4 (4–4) 4 (4–4) –
Additional image noise 4 (4–5) 4 (3–4) 0.063
The presence of motion artifacts 4 (2–4) 5 (5–5) 0.031
Subjective diagnostic reliability 5 (4–5) 5 (4–5) –
Overall diagnostic image quality 4 (4–5) 4 (4–5) –
Anatomy
normal lung structures (major fissures and small vessels) 5 (5–5) 5 (4–5) –
bronchi (< 2 mm diameter) 5 (4–5) 5 (4–5) –
Parenchymal findings
GGO < 2 cm Reference standard 5 (4–5) //
COVID signs (Crazy paving—reverse Halo sign) Reference standard 5 (4–5) //
Centrolobular nodules Reference standard 3 (3–4) //
lobar/subsegmental consolidations Reference standard 5 (4–5) //
Lymphadenopathy Reference standard 4 (4 –5) //

Median values on the Likert scale (from 1 unacceptable to 5 excellent) for the two readers
* Wilcoxon test for paired samples
– Unable to calculate p
// Test not performed

performance the radiologist did not experience a signifi- Discussion


cant reduction in diagnostic reliability and evaluation
of the pathological findings between the two protocols In the present study, we included a small sample of
(Fig. 2a–c). patients affected by COVD-19-related pneumonia with
Finally, post-processing of DECT datasets at different variable clinical and radiological manifestation of the
blending ratio allowed for reduction in beam-hardening disease, and comparable findings described in the litera-
artifacts in 1/10 (10%) patients who was unable to main- ture [8, 9]. Given the etiological and clinical similarities
tain the arms raised (Fig. 1c–f). The beam-hardening arti- between COVID-19 and other pulmonary syndromes such
facts minorly affected the 100Sn kV.

13

370 La radiologia medica (2020) 125:365–373

Fig. 1  Technical advantages of DECT and LDCT. Axial images, a, c, ening artifacts. c, e have a linear blending of 0.7 while D, F have lin-
d, e, f: HD-DECT; b: LDCT, lung kernel. a, b Comparison of HD- ear blending of 0.2 resulting in reduction in artifacts in mediastinum
DECT and LDCT, demonstrating the reduction in the motion artifact (d) and left lung parenchyma (f)
in the LDCT. c–f Shows the role of DECT in reduction in beam-hard-

as SARS and MERS (which demonstrated the development results and thus not investigated any more. Further studies
of air trapping and fibrosis [21]), the presence of air trap- on follow-up of COVID-19 should be performed to clarify
ping was searched in the first three cases, with negative possible chronic lung injuries.

13
La radiologia medica (2020) 125:365–373 371

Fig. 2  COVID-19 Pathological Findings: HD-DECT versus LDCT. Axial Images. a, c HD-DECT; b, d LDCT. a, b Good demonstration of
crazy-paving both in a and b. c, d demonstration of Reverse Halo and linear opacities in HD-DECT and LDCT

In these 10 patients with COVID-19, we evaluated the comparable to the values reported in previous studies and
feasibility of an ultra-low-dose, long-pitch, dual-source, close to a chest X-ray examination. In these patients, chest
fast CT acquisition to be implemented for serial follow-up X-ray examination at baseline was not performed to avoid
examinations in symptomatic patients with COVID-19. At unnecessary radiation exposure.
baseline, a DECT acquisition was chosen as internal refer- The rationale behind the DECT choice was to couple a rela-
ence standard. tively fast acquisition (2–2.5 s) with the possibility of post-
Previous studies in the literature demonstrated significant processing for eventual artifact reduction with different blend-
dose reduction with spectral shaping in chest CT, achiev- ing combinations in uncooperative patients unable to maintain
ing a dose comparable to a chest x-ray examination [12, 13, arms raised. The post-processing of DECT datasets reduced
18, 19, 23, 24]. In our population, the effective dose of the beam-hardening artifacts in 1/10 patients (10%). These arti-
ultra-low-dose, long-pitch, dual-source fast acquisition was facts minorly affected the LDCT. An explanation of this can

13

372 La radiologia medica (2020) 125:365–373

be found in the relatively high mean energy of the 100Sn kVp Conclusions


spectrum [19].
Conversely, the LDCT acquisition was conceived to avoid The study demonstrated the feasibility of an ultra-low-
artifacts from heart beats, breathing and coughing in patients dose, fast chest CT acquisition with spectral shaping at
with severe symptoms, which was significantly demonstrated 100 kVp (100Sn kV) and dual-source acquisition with
in the subjective analysis (Table 4, Fig. 1). The feasibility, the ultra-long pitch (Turbo Flash, Siemens Healthineers) in
quality and the diagnostic performance of ultra-low-dose chest patients affected by COVID-19 with good diagnostic reli-
CT with spectral shaping, and other technical solutions have ability and potential for reduction in radiation dose and
already been demonstrated [12, 13, 18, 19, 24] sometimes with motion artifacts.
questionable results about the pathological findings [25]. To
our knowledge, this is the first study evaluating a long-pitch,
dual-source acquisition with spectral shaping in acute setting Funding  This study was not supported by any funding.
in patients with COVID-19. As expected from a low-dose
acquisition, the LDCT images had trend, though poorly signifi- Compliance with ethical standards 
cant, to be evaluated as more noisy than the HD-DECT images
(Table 4). However, it did not significantly influence the detec- Conflict of interest  A.A. is a speaker for Siemens Helthineers. The oth-
er authors declare that they have no conflict of interests.
tion and characterization of main anatomical and pathological
pulmonary findings in this subset of patients, as demonstrated Ethical approval  All procedures in studies involving human partici-
by the high rankings (Likert ≥ 3 for the pathological findings pants were in accordance with the ethical standards of the institutional
in Table 4) (Fig. 2). and/or national research committee and with the 1964 Declaration of
Helsinki and its later amendments or comparable ethical standards.
The quantitative analysis found significant differences in
SNR and CNR for several anatomical structures, as expected Informed consent  For this type of study, formal consent is not required.
when comparing a standard-dose and an ultra-low-dose acqui-
sition with an ultra-long pitch. However, these differences had Consent for publication  For this type of study, consent for publication
is not required.
poor significance in the evaluation of lung SNR and CNR,
in agreement to the subjective analysis. Again, this can be
explained by considering the relatively high mean energy of
the 100Sn kVp spectrum contributing to lowering the noise
together with the radiation dose (Table 2, Fig. 2) [19]. More- References
over, in this protocol, the reconstructions have been kept as
similar as possible, in particular the slice thickness, the slice 1. Chen N, Zhou M, Dong X et al (2020) Epidemiological and clini-
cal characteristics of 99 cases of 2019 novel coronavirus pneumo-
spacing and the iterative reconstructions while for LDCT a nia in Wuhan, China: a descriptive study. Lancet 395:507–513
slightly softer kernel was used. Again, in Table 4, the image 2. Zu ZY, Jiang MD, Xu PP et al (2020) Coronavirus Disease 2019
sharpness of LDCT was comparable with maintained diag- (COVID-19): a Perspective from China. Radiology. https​://doi.
nostic reliability. Further studies are necessary to fully explore org/10.1148/radio​l.20202​00490​:20049​0
3. Wang W, Tang J, Wei F (2020) Updated understanding of the out-
the role of iterative reconstructions at lower doses in this acute break of 2019 novel coronavirus (2019-nCoV) in Wuhan, China.
setting [18]. J Med Virol 92:441–447
The study has several limitations. It is a single-center study 4. Tian S, Hu N, Lou J et al (2020) Characteristics of COVID-
with a small population to evaluate the feasibility of a low-dose 19 infection in Beijing. J Infect. https ​ : //doi.org/10.1016/j.
jinf.2020.02.018
CT protocol in acute setting. The small size of the sample did 5. Huang P, Liu T, Huang L et al (2020) Use of chest CT in combina-
not allow for more sophisticated statistical analysis, such as tion with negative RT-PCR assay for the 2019 novel coronavirus
inter-reader agreement or the relation between BMI, radiation but high clinical suspicion. Radiology. https​://doi.org/10.1148/
dose and image quality. radio​l.20202​00330​:20033​0
6. Ai T, Yang Z, Hou H et al (2020) Correlation of chest CT and RT-
Concluding, a compromise including preserved image PCR testing in coronavirus disease 2019 (COVID-19) in China:
sharpness within lung parenchyma, with minimally blurred a report of 1014 cases. Radiology. https​://doi.org/10.1148/radio​
mediastinal structures in the face of a median dose reduction l.20202​00642​:20064​2
of more than 90% may be acceptable when performing serial 7. Fang Y, Zhang H, Xie J et al (2020) Sensitivity of chest CT for
COVID-19: comparison to RT-PCR. Radiology. https​://doi.
CT controls in severely ill young patients [26]. org/10.1148/radio​l.20202​00432​:20043​2
8. Bernheim A, Mei X, Huang M et al (2020) Chest CT findings in
coronavirus disease-19 (COVID-19): relationship to duration of
infection. Radiology. https​://doi.org/10.1148/radio​l.20202​00463​
:20046​3
9. Pan F, Ye T, Sun P et al (2020) Time course of lung changes
on chest ct during recovery from 2019 novel coronavirus

13
La radiologia medica (2020) 125:365–373 373

(COVID-19) pneumonia. Radiology. https:​ //doi.org/10.1148/radio​ performance evaluation of single energy scanning with spectral
l.20202​00370​:20037​0 shaping. Invest Radiol 49:465–473
10. Kanne JP, Little BP, Chung JH, Elicker BM, Ketai LH (2020) 19. Haubenreisser H, Meyer M, Sudarski S, Allmendinger T, Schoen-
Essentials for radiologists on COVID-19: an update-radiology berg SO, Henzler T (2015) Unenhanced third-generation dual-
scientific expert panel. Radiology. https​://doi.org/10.1148/radio​ source chest CT using a tin filter for spectral shaping at 100 kVp.
l.20202​00527​:20052​7 Eur J Radiol 84:1608–1613
11. Rawat U, Cohen SL, Levsky JM, Haramati LB (2015) ACR white 20. Chang YC, Yu CJ, Chang SC et al (2005) Pulmonary sequelae
paper-based comprehensive dose reduction initiative is associated in convalescent patients after severe acute respiratory syndrome:
with a reversal of the upward trend in radiation dose for chest CT. evaluation with thin-section CT. Radiology 236:1067–1075
J Am Coll Radiol 12:1251–1256 21. Hosseiny M, Kooraki S, Gholamrezanezhad A, Reddy S, Myers
12. Lim HK, Ha HI, Hwang HJ, Lee K (2019) High-pitch, 120 kVp/30 L (2020) Radiology perspective of coronavirus disease 2019
mAs, low-dose dual-source chest CT with iterative reconstruction: (COVID-19): lessons from severe acute respiratory syndrome and
prospective evaluation of radiation dose reduction and image qual- middle east respiratory syndrome. AJR Am J Roentgenol. https​://
ity compared with those of standard-pitch low-dose chest CT in doi.org/10.2214/AJR.20.22969​:1-5
healthy adult volunteers. PLoS One 14:e0211097 22. Deak PD, Smal Y, Kalender WA (2010) Multisection CT proto-
13. Bodelle B, Fischbach C, Booz C et al (2017) Free-breathing high- cols: sex- and age-specific conversion factors used to determine
pitch 80kVp dual-source computed tomography of the pediatric effective dose from dose-length product. Radiology 257:158–166
chest: image quality, presence of motion artifacts and radiation 23. Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effec-
dose. Eur J Radiol 89:208–214 tive doses in radiology and diagnostic nuclear medicine: a catalog.
14. Hagelstein C, Henzler T, Haubenreisser H et al (2016) Ultra-high Radiology 248:254–263
pitch chest computed tomography at 70 kVp tube voltage in an 24. Macri F, Greffier J, Pereira FR et al (2016) Ultra-low-dose chest
anthropomorphic pediatric phantom and non-sedated pediatric CT with iterative reconstruction does not alter anatomical image
patients: initial experience with 3(rd) generation dual-source CT. quality. Diagn Interv Imaging 97:1131–1140
Z Med Phys 26:349–361 25. Ludes C, Labani A, Severac F et al (2019) Ultra-low-dose unen-
15. Johnson TR, Krauss B, Sedlmair M et al (2007) Material dif- hanced chest CT: prospective comparison of high kV/low mA ver-
ferentiation by dual energy CT: initial experience. Eur Radiol sus low kV/high mA protocols. Diagn Interv Imaging 100:85–93
17:1510–1517 26. Agostini A, Mari A, Lanza C et al (2019) Trends in radiation dose
16. Marin D, Pratts-Emanuelli JJ, Mileto A et al (2015) Interdepend- and image quality for pediatric patients with a multidetector CT
encies of acquisition, detection, and reconstruction techniques and a third-generation dual-source dual-energy CT. Radiol Med
on the accuracy of iodine quantification in varying patient sizes 124:745–752
employing dual-energy CT. Eur Radiol 25:679–686
17. Krauss B, Grant KL, Schmidt BT, Flohr TG (2015) The impor- Publisher’s Note Springer Nature remains neutral with regard to
tance of spectral separation: an assessment of dual-energy spec- jurisdictional claims in published maps and institutional affiliations.
tral separation for quantitative ability and dose efficiency. Invest
Radiol 50:114–118
18. Gordic S, Morsbach F, Schmidt B et al (2014) Ultralow-dose
chest computed tomography for pulmonary nodule detection: first

13

You might also like