Professional Documents
Culture Documents
D5XXX - XX
Input unit
MODEL FD5000 SERIES 01: DC voltage measuring unit (range 11: ± 99.99 mV)
INSTRUCTION MANUAL 02: DC voltage measuring unit
(range 12: ± 999.9 mV ; range 13: ± 9.999 V)
(range 14: ± 99.99 V ; range 15: ± 600 V)
03: DC current measuring unit
(range 23: ± 9.999 mA; range 24: ± 99.99 mA)
(range 25: ± 999.9 mA)
04: AC voltage measuring unit (average rms)
(range 11: 99.99 mV; range 12: 999.9 mV)
(range 13: 9.999 V)
05: AC voltage measuring unit (average rms)
(range 14: 99.99 V; range 15: 600 V)
06: AC voltage measuring unit (true rms)
! Caution (range 11: 99.99 mV; range 12: 999.9 mV)
(range 13: 9.999 V)
(1) Do not apply a voltage or current exceeding the maximum allowable 07: AC voltage measuring unit (true rms)
value; otherwise, it may damage the equipment. (range 14: 99.99 V; range 15: 600 V)
(2) Use a power voltage within the operation range; otherwise, it may 08: AC current measuring unit (average rms)
result in a fire, electrical shock, or malfunction. (range 23: 9.999 mA; range 24: 99.99 mA)
(3) The contents of this manual are subject to change without notice. (range 25: 999.9 mA)
(4) Although the contents of this manual have been prepared with extra 09: AC current measuring unit (average rms)
care, if you have any questions, or find errors or missing (range 26: 5 A)
information, contact the sales agent from which you purchased the 10: AC current measuring unit (true rms)
product or Fuji Electric Instruments Co.,Ltd. (range 23: 9.999 mA; range 24: 99.99 mA)
(5) After reading this manual thoroughly, keep it in a convenient place (range 25: 999.9 mA)
for future reference. 11: AC current measuring unit (true rms)
(range 26: 5 A)
12: Resistance measuring unit
13: Temperature measuring unit (TC)
1. Before Using the Product 14: Temperature measuring unit (RTD)
Thank you for purchasing the FD5000 series. This manual should be passed 15: Frequency measuring unit
on to the person who operates the product. Examine the product for dam- (inputs: open collector, logic, and magnet)
16: Frequency measuring unit (input: 50 to 500 Vrms)
age caused by transportation or any other defects. If you find any damage or
17: Strain gauze input unit (load cell)
defects, contact the sales agent from which you purchased the product or
18: Process signal measuring unit
Fuji Electric Instruments Co., Ltd. (4 to 20 mA or 1 to 5 V)
1.1 Model Codes
The model lineup of the FD5000 series is shown below. Check that the 2. Mounting the Product
model code and specifications of your product match those you specified 2.1 Dimensions for Cutting Panel
when ordering.
Cut the panel for mounting according to the following dimensions.
FD5XXX - XX 92 +0.8
-0
Power unit
1: 100 to 240 V AC ± 10% 45
2: 9 to 60 V DC
-0 +0.6
Display unit
1: Single display
2: Multi display
70 mm
Output unit min.
0: None
1: Comparison
2: Analog
3: RS-232C
4: RS-485
5: Comparison and analog
6: Comparison, analog, and RS-232C
7: Comparison, analog, and RS-485 120 mm min.
3.3.4 AC Voltage Measuring Unit (Ranges 11 to 13) Conversion system : PWM conversion
3. Terminal Arrangement Resoluton : Equivalent to 13 bits
No. Name Description
Scaling : Digital scaling
Comparison Serial Positive input terminal for ranges
Analog output 1 11-12
output communication 11 (99.99 mV) and 12(999.9 mV)
Positive input terminal for range
Response speed : About 0.5 second
1 2 3 2 13
13 (9.999 V)
3 LO Common input terminal 8.3.3 Communicating Function
12 13 14 15 16 17 18 19 20 22 24 25 26
8.1.14 Temperature Measuring Unit (RTD) 8.2 Common Specifications 3.3.10 Temperature Measuring Unit (RTD) 3.5 Analog Output
8.1.4 AC Voltage Measuring Unit (average value detection: ranges 8.1.9 AC Current Measuring Unit (average value detection: range
Pressing the Mode and Enter keys Pressing the Shift and Enter keys
together changes to the parameter Selects the item to be set. together changes to the parameter 11 to 13) 26)
setting mode. checking mode. (Comparator data Changes the digit to be set.
Pressing the Mode and Shift keys can be set.) Maximum
Maximum Highest
Pressing the Shift and Mode keys Highest Measurement Input
(7) Mode key together changes to the shift function Range Measurement Indication Input
permissible Accuracy Range
range
Indication
resolution impedance
permissible Accuracy
setup mode. together changes to the shift function range resolution impedance input
input
Pressing the Mode and Incremental (8) Shift key setup mode. Offset :± 9999 ± (0.5% of FS
keys together turns on/off the"Digital Holding down the Shift key for about 11 99.99mV 10µV 26 5A 1mA (CT) 8A
Full scale : 0 to± 9999 +10digit)
zero" indicator. one second moves to the HI judgment ± 100V
12 999.9mV Offset :± 9999 100µV 1MΩ ± (0.2% of FS
Pressing the Shift and Enter keys value indicator. Full scale : 0 to± 9999 or more +10digit) Input circuit : CT isoration type
together changes to the parameter Selects from items in the maximum/ Operating system : ∆ Σ conversion
Changes the digit to be set. minimum/(maximum-minimum)/input 13 9.999V 1mV ± 250V Maximum sampling rate : 12.5 times per second
checking mode. (Comparator data
can be set.) value monitoring mode. (Hold down Frequency range : 50 Hz or 60Hz
Pressing the Shift and Mode keys the key for about one second.) Input circuit : Single ended type Response speed : About 1 second
(8) Shift key together changes to the shift function Pressing the Increment and Mode Changes the value or content Operating system : ∆ Σ conversion Crest factor : 4:1 at full scale
setup mode. keys together turns on/off the"Digital of a selected digit. Maximum sampling rate : 12.5 times per second Dead zone : 0 to 99 digits
Selects from items in the maximum/ zero" indicator. (Increments the value) Frequency range : 40 Hz to 1 kHz
minimum/(maximum-minimum)/input Holding down the Increment key for Response speed : About 1 second
value monitoring mode. (Hold down about one second moves to the LO Dead zone : 0 to 99 digits 8.1.10 AC Current Measuring Unit (true rms value: ranges 23 to 25)
the key for about one second.) judgment value indicator.
Pressing the Increment and Mode Changes the value or content (9) Increment key Pressing the Increment and Enter
keys changes to the maximum/ Maximum
keys together turns on/off the"Digital of a selected digit. Highest
8.1.5 AC Voltage Measuring Unit (average value detection: ranges Range Measurement Indication Input
permissible Accuracy
zero" indicator. (Increments the value) minimum/(maximum-minimum)/input range resolution impedance
input
Pressing the Increment and Enter value monitoring mode. 14 and 15) About
keys changes to the maximum/ Resets the maximum/minimum/ 23 9.999mA 1µA 100mA
10Ω
(9) Increment key minimum/(maximum-minimum)/input (maximum-minimum)/input value Offset :± 9999 About ± (0.5% of FS
value monitoring mode. monitoring mode. (Hold down the key Maximum 24 99.99mA 10µA 500mA
Measurement Highest Input Full scale : 0 to± 9999 1Ω +20digit)
Resets the maximum/minimum/ for about one second.) Range Indication permissible Accuracy
range resolution impedance About
(maximum-minimum)/input value input 25 999.9mA 100µA 3A
± (0.2% of FS 0.1Ω
monitoring mode. (Hold down the key 14 99.99V 10mV 250V
Offset :± 9999 1MΩ +10digit)
for about one second.) Full scale : 0 to± 9999 or more ± (0.3% of FS
Input circuit : Single ended type
5. Parameter Setup 15 600V 100mV 600V +10digit) Operating system : ∆ Σ conversion
Maximum sampling rate : 12.5 times per second
5.1 Differences between Display Units Input circuit : Single ended type Frequency range : 40 Hz to 1 kHz
4.2 Single Display Unit Operating system : ∆ Σ conversion Response speed : About 1 second
Maximum sampling rate : 12.5 times per second Crest factor : 4:1 at full scale
Frequency range : 40 Hz to 1 kHz Dead zone : 0 to 99 digits
(2) (1) (3) (1) (3) Response speed : About 1 second
Dead zone : 0 to 99 digits
8.1.11 AC Current Measuring Unit (true rms value: range 26)
Maximum
Measurement Highest Input
8.1.6 AC Voltage Measuring Unit (true rms value: ranges 11 to 13) Range
range
Indication
resolution impedance
permissible Accuracy
input
Offset :± 9999 ± (0.5% of FS
26 5A 1mA (CT) 8A
Maximum Full scale : 0 to± 9999 +20digit)
Highest
Setup contents Parameter name Range Measurement
range
Indication Input
permissible Accuracy
resolution impedance
input Input circuit : CT isoration type
(6) (7) (8) (9) (6) (7) (8) (9) Operating system : ∆ Σ conversion
11 99.99mV 10µV
Main Functions 5.1.1 Multi-display Unit ± 100V
Maximum sampling rate : 12.5 times per second
No. Name 12 999.9mV Offset :± 9999 100µV 1MΩ ± (0.2% of FS Frequency range : 50 Hz or 60Hz
During measurement During parameter setup Full scale : 0 to± 9999 or more +20digit) Response speed : About 1 second
IIndicates information on the Crest factor : 4:1 at full scale
Indicates the measured value. 13 9.999V 1mV ± 250V
parameter to be set. Dead zone : 0 to 99 digits
(1) Main display Indicates information on the item in the
maximum/minimum/(maximum- Input circuit : Single ended type
minimum)/input value monitoring mode. Operating system : ∆ Σ conversion
Maximum sampling rate : 12.5 times per second 8.1.12 Resistance Measuring Unit
Indicates the result of judgment and
HI turns on if the measured value > HI Frequency range : 40 Hz to 1 kHz
judgment value.
Response speed : About 1 second
Crest factor : 4:1 at full scale Measurement Highest Circuit
Indicates the result of judgment and Range Indication Accuracy
Judgment Setup contents and parameter name Dead zone : 0 to 99 digits range resolution current
(2) GO turns on if LO judgment value≤ the
indicators 11 99.99Ω 10mΩ
About
measured value ≤ HI judgment value. 5mA
Indicates the result of judgment and Note: Pressing the Mode key displays the next parameter. 8.1.7 AC Voltage Measuring Unit (true rms value: ranges 14 and About
LO turns on if the measured value < LO 12 999.9Ω Offset :± 9999 100mΩ
5.1.2 Single Display Unit 15) 500µA ± (0.2% of FS)
judgment value. Full scale : 0 to± 9999 About
13 9.999kΩ 1Ω
Turns on if "peak hold/valley hold/peak - Note 1: Pressing the mode key with the parameter name shown changes the 50µA
PH Maximum About
valley hold" is on. Measurement Highest Input 14 99.99kΩ 10Ω
display to the parameter information indication. If there is no key Range Indication permissible Accuracy 5µA
Turns on if "digital zero" is on. range resolution impedance
input
DZ Flashes when linearization data output
operation for about one second when the parameter name is shown, ± (0.2% of FS
values are set. 14 99.99V 10mV 250V Input circuit : Single ended type
Function Offset :± 9999 1MΩ +20digit) Operating system : ∆ Σ conversion
(3)
indicators
ME Turns on if "digital zero backup" is on. the display automatically changes to the parameter information Full scale : 0 to± 9999 or more ± (0.3% of FS Maximum sampling rate : 12.5 times per second
15 600V 100mV 600V
Turns on if remote control is being +20digit) Measuring system : Two-wire system or four-wire system
performed through RS-232C or RS-485 indication (however, this change does not automatically occur for (internal socket change-over)
RE interface. parameters PH/S-HI/FSC, etc., right after COND/COM/MET is in- Input circuit : Single ended type Open-circuit voltage : About 5 V
Flashes when linearization data input Operating system : ∆ Σ conversion
values are set. Maximum sampling rate : 12.5 times per second
dicated). Frequency range : 40 Hz to 1 kHz
Pressing the Mode and Enter keys Response speed : About 1 second 8.1.13 Temperature Measuring Unit (TC)
Returns to the measurement
together changes to the parameter
mode.
Note 2: Pressing the Mode key when the parameter information indication Crest factor : 4:1 at full scale
setting mode. Dead zone : 0 to 99 digits
Pressing the Enter and Increment keys is shown results in the next parameter being displayed. Input Highest
Range Indication Accuracy
together changes to the maximum/ sensor resolution
Note 3: If there is no key operation for about 8 seconds with the parameter
minimum/(maximum-minimum)/input 8.1.8 AC Current Measuring Unit (average value detection: ranges -50.0 to 199.9°C 0.1°C
(6) Enter key value monitoring mode. information indication shown, the display returns to the parameter KA K ± (0.5% of FS)
(-58.0 to 391.8°F) (0.1°F)
Switches from the maximum/minimum/ 23 to 25) -50 to 1200°C
name indication. KB K
(-58 to 2192°F)
(maximum-maximum/minimum/ ± (0.2% of FS)
(maximum-minimum)/input value Maximum -50 to 1000°C
Highest J J
monitoring mode to the comparative Range Measurement
range
Indication Input
permissible Accuracy (-58 to 1832°F)
resolution impedance -50 to 400°C
judgment reading mode. input T T ± (0.6% of FS)
About (-58 to 752°F) 1°C
Pressing the Mode and Enter keys 23 9.999mA 1µA 100mA 0 to 1700°C
10Ω S S (1°F)
together changes to the parameter Selects the item to be set. Offset :± 9999 About ± (0.5% of FS (32 to 3092°F)
setting mode. 24 99.99mA 10µA 500mA ± (0.4% of FS)
Full scale : 0 to± 9999 1Ω +10digit) -10 to 1700°C
R R
Pressing the Mode and Shift keys About (14 to 3092°F)
25 999.9mA 100µA 3A ± (0.4% of FS)
(7) Mode key together changes to the shift function 0.1Ω 100 to 1800°C
setup mode. B B Note : The accuracy of range B is applicable
(212 to 3272°F)
Pressing the Mode and Incremental Input circuit : Single ended type to temperatures of 500°C or more.
keys together turns on/off the"Digital Operating system : ∆ Σ conversion
Maximum sampling rate : 12.5 times per second Input circuit : Single ended type
zero" indicator. Frequency range : 40 Hz to 1 kHz Operating system : ∆ Σ conversion
Response speed : About 1 second Maximum sampling rate : 6.25 times per second
Crest factor : 4:1 at full scale Cold junction compensation error : ± 2°C(at 10 through 40°C)
Dead zone : 0 to 99 digits Internal resistance of sensor : 50Ω or less
Linearizer : Digital linearizer
Burnout alarm : It blinks by ---- display.
Instruction Manual for FD5000 Series 8/12 Instruction Manual for FD5000 Series 5/12
5.4.5 Method of Setting Linearization Data 5.2 Moving to the Parameter Setup Mode
6.3 Peak Hold Function
The linearization function means a function that changes the slope of straight Calibration data
The Peak Hold function retains one of the maximum (peak hold)/minimum Measurement
lines in the relationship between the input and indication by correcting the rela-
(valley hold)/maximum - minimum (peak-valley hold) values and provides
tions at arbitrary points. Linearization data are set using the input value (indicated
output for that value. Selection from these values is made using the condition
value before correction) and the output value (indicated value after correction) at
data. The peak hold function is enabled by shortcircuiting the PH and COM Shift data
each arbitrary point.
terminals or setting both terminals to the same voltage level.
Multi-display unit Single display unit 7. Output Function
7.1 Comparison Output Function
(1) Press the Mode key and the Enter key during The FD5000 series of unit meters is designed so that the two judgment values
measurement.
HI and LO can be set for the measured (indicated) value to provide the results
of judgment as relay contact output. (This function is effective when the
(2) Press the Shift key a few times to move to the
linearization data menu. meter is equipped with a comparison output unit.) For details on the contact
ratings and other specifications, refer to the section “Output Specifications.”
(3) Press the Mode key to move to the setup for the 7.2 Analog Output Function
number of data to be corrected.
The FD5000 series of unit meters can output an analog signal for an indicated
(4) Press the Shift key (change digit) and the value (when the meter is equipped with an analog output unit). There are four
Increment key (change numeric value) and then press
& output ranges, 0 to 1 V/0 to 10 V/1 to 5 V/4 to 20 mA, from which a selection
the Mode Key to after the number of data to be
corrected has been set.
Note: The decimal point in the selected digit
can be made using the condition data. In addition, the analog output of the
flashes.
FD5000 series allows for arbitrary output scaling. This scaling can be achieved
(5) For a single display unit, press the Mode key to
change to the parameter information indication. by setting the indication value for an output of the maximum scale value (20
Note: For a single display unit, the unit
automatically returns to N-01 indication if there is mA for 4–20 mA output range) in the AOHI parameter of the scaling data.
no key operation for about 8 seconds. In this case,
RE
press the Mode key to return to the numeric value 7.3 RS-485 Interface Function Condition data Comparator data Scaling data Linearization data
indication.
(6) Press the Shift key (change digit) and Increment The FD5000 series can be equipped with an RS-485 interface (when the
& key (change numeric value) and then press the Mode
key after the N-01 input value has been set. meter is provided with an RS-485 unit). For details on the RS-485 function,
DZ
Note: For a single display unit, RE flashes when see the separate manual on communication functions.
the input value is set and DZ flashes when the
output value is set.
(7) Press the Shift key (change digit) and Increment
7.4 RS-232C Interface Function Pressing the ENTER key saves the data and returns to the measurement mode.
& key (change numeric value) and then press the Mode
key after the N-01 output value has been set. The FD5000 series can be equipped with an RS-232C interface (when the
(Data are backed up with EEPROM even when the power is turned off.)
meter is provided with an RS-232C unit). For details on the RS-232C
(8) Press the Mode key to display the setup for output function, see the separate manual on communication functions.
values of the data to be corrected.
5.3 Data Lists and Default Settings
Note: For a single display unit, the unit is
automatically returned to N-02 indication if there is
8. Specifications and External Dimensions
no key operation for about 8 seconds. In this case, Default Equipped Input unit number
RE
press the Mode key to return to the numeric value 8.1 Input Specifications Indication Name value as 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 0 1 2 3 4 5
indication. Condition data
(9) Press the Shift key (change digit) and Increment 8.1.1DC Voltage Measuring Unit (range 11) PVH Peak hold setup PH
& key (change numeric value) and then press the Mode RANG Measurement range setup *1 X X X X
key after the N-02 input value has been set. 11 15 25 13 15 13 15 25 26 25 26 14 B JPB 14 14 2A
Maximum
Measurement Highest Input AVG Number of averaging 1 X X
Range Indication permissible Accuracy
Repeat steps (5) to (9) until all the settings have range resolution impedance MAV Number of moving averaging setup OFF
input
been made. S.UD Step wide setup 1
Offset :± 9999 About
11 ± 99.99mV 10µV ±100V ± (0.1% of FS) BLNK Indication blank setup OFF
Full scale : 0 to± 9999 100MΩ UNIT Unit setup C X X X X X X X X X X X X X X X X
(10) When all the settings have been made, press the BAUD Baud rate setup 9600 X X X X
Enter key to return to the measurement mode.
Input circuit : Single ended type DATA Data length setup 7 X X X X
Operating system : ∆ Σ conversion P.BIT Parity bit setup E X X X X
Maximum sampling rate : 12.5 times per second S.BIT Stop bit setup 2 X X X X
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz) T- Delimiter setup CR.LF X X X X
ADR Equipment ID setup 00 X X X X X
Note: The setup conditions are N-1 < N-2 … N-15 < N-16. 8.1.2DC Voltage Measuring Unit (ranges 12 to 15)
A.OUT Analog output setup OFF X X X X
B.UP Digital zero backup setup OFF
6. Control Functions LINE Linearization setup CLR
I.SEL Input selection OC X X X X X X X X X X X X X X X X X
Maximum TR T Tracking zeroing time setup 00 X X
Highest
6.1 Hold Function Range Measurement Indication Input
permissible Accuracy TR V Tracking zeroing width setup *2 01 X X
range resolution impedance
input SNSR Sensor power setup 10 X X X X X X X X X X X X X X X X X
The Hold function temporarily retains the indication. The hold function is enabled About PÇnÇm Power-on delay setup OFF
12 ± 999.9mV 100µV ± 100V
100MΩ PRO Protect setup OFF
by shortcircuiting the HOLD and COM terminals or setting both terminals to the same About
13 ± 9.999V 1mV ± 250V ± (0.1% of FS) U-NO. Unit number Indication setup ON
voltage level. As a result the display unit retains the indication given at that moment. Offset :± 9999 1MΩ Comparator data
Full scale : 0 to± 9999 About S-HI HI side judgment value setup 1000 X X X X
14 ± 99.99V 10mV ± 250V
6.2 Digital Zero Function 10MΩ S-LO LO side judgment value setup 500 X X X X
About H-HI HI side hysteresis setup 0 X X X X
15 ± 600V 100mV ± 600V ± (0.15% of FS)
The Digital Zero function zeros the indication given at an arbitrary timing. 10MΩ H-LO LO side hysteresis setup 0 X X X X
Scaling data
Thereafter, the function shows the amount of change from the point of zero- Input circuit : Single ended type FSC Full scale Indication value setup *1 X X X X X
Operating system : ∆ Σ conversion 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999
ing. However, this function serves as an indication resetting function for a Maximum sampling rate : 12.5 times per second FIN Full scale input value setup *1 X X X X X
frequency measuring unit. Thus, the Digital Zero function can be used to reset Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz) 9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999 *3
OFS Offset indication value setup *1 X X X X X
the indication when there is no input signal at all. 0 0 0 0 0 0 0 0 0 0 0 0 0
OIN Offset input value setup *1 X X X X X
Note that, the on/off control of the Digital Zero function can be achieved by 8.1.3DC Current Measuring Unit 0 0 0 0 0 0 0 0 0 0 0 0 *4
means of terminal control or front panel keys. In the case of terminal control, PS Pre-scaling value setup 1 X X X X X X X X X X X X X X X X
PPR Frequency division setup 1 X X X X X X X X X X X X X X X X
the Digital Zero function is turned on by shortcircuiting the DZ and COM termi- Maximum DLHI Digital limiter HI value setup 9999 X X X X X X X X
Measurement Highest Input
Range Indication permissible Accuracy DLLO Digital limiter LO value setup -9999 X X X X X X X X
range resolution impedance
nals or setting both terminals to the same voltage level. The indication at that input AOHI Analog output HI indication setup 9999 X X X X
About AOLO Analog output LO indication setup 0 X X X X
moment is zeroed. In the case of control with the front panel keys, hold down 23 ± 9.999mA 1µA
10Ω
± 100mA
± (0.2% of FS) DEP Decimal point position setup None
the Mode key and press the Increment key for about 1 second to zero the indi- Offset :± 9999 About *5
24 ± 99.99mA Full scale : 0 to± 9999 10µA 1Ω ± 500mA
cation at that moment. 25 ± 999.9mA 100µA
About
± 3A ± (0.3% of FS) ZERO Zero input value *6 0 X X X X X X X X X X X X X X X X X
0.1Ω SPIN Span input value *6 2000 X X X X X X X X X X X X X X X X X
Note: Operation with the control terminals takes priority over operation SPAN Span indication 9000 X X X X X X X X X X X X X X X X X
with the front panel keys. The Digital Zero function is disabled if Input circuit : Single ended type Shift data
Operating system : ∆ Σ conversion SHF Shift data setup 0
the control terminals are made to go through the off-on-off se- Maximum sampling rate : 12.5 times per second *1 Each value in the lower part of a cell in the columns on the right is the default value.
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz) *2 Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
quence with the function enabled by means of the front panel keys. *3 5000 for 1 V range and 2000 for 2 A range
*4 1000 for 1 V range and 400 for 2 A range
*5 Linearization data are not set up for the default values.
*6 This value is not indicated if calibration is done using an actual load.
The shaded parts show the parameters that must be set for each unit.
Instruction Manual for FD5000 Series 6/12 Instruction Manual for FD5000 Series 7/12
5.4 Information on Each Parameter 5.4.2 Method of Setting Comparator Data Multi-display unit Single display unit 5.4.4Method of Setting Calibration Data
This section explains comparator data and shows a typical example of setting
Default 5.4.4.1 Actual Load Calibration
Indication Name Setup options
value the HI side judgment value. The same method applies to all other parameters.
Condition data Actual load calibration means that calibration is carried out by applying actu-
PVH Peak hold setup PH (peak hold)/VH (valley hold)/PVH (peak-valley hold) PH (1) Press the Mode and Enter keys together
HI side judgment value : 900 during measurement.
RANG Measurement range setup *1 *1
HI side hysteresis value : 200
ally measured pressure to a sensor such as a load cell connected to the meter.
Number of averaging operations
AVG 1/2/4/8/10/20/40/80 1 LO side judgment value : 300
setup
Number of moving average LO side hysteresis value : 150
MAV OFF/2/4/8/16/32 OFF Multi-display unit Single display unit
operations setup (2) Press the Shift key a few times to change to
S.UD Step width setup 1(1digit)/2(2digit)/5(5digit)/0(10digit) 1 the scaling data menu.
Indicated value
BLNK Indication blank setup OFF/B-3/B-2/B-1/ON OFF
UNIT Unit setup C/F C HI side judgment value
1000
BAUD Baud rate setup 9600/4800/2400/384(38400)/192(19200) 9600
DATA Data length setup 7(7bit)/8(8bit) 7bit
900
HI side hysteresis (1) Press the Shift, Increment, and Enter keys
P.BIT Parity bit setup E (even number), O (odd number), N (none) E 200 range (3) Press the Mode key a few times to display the together during measurement.
S.BIT Stop bit setup 2(2bit)/1(1bit) 2 parameter to be set.
T- Delimiter setup CR.LF(CR+LF)/CR CRLF
ADR Equipment ID setup 01 to 99 00
500
A.OUT Analog output setup OFF/0-1(0 to 1V)/0-10(0 to 10V)/1-5(1 to 5V)/4-20(4 to 20mA) OFF (4) For a single display unit, press the Mode key
LO side hysteresis to change to the parameter information indication. (2) Press the Mode key to change to the actual
B.UP Digital zero backup setup OFF/ON OFF 150 range
LINE Linearization setup OFF/ON CLR 300 (The display automatically changes to this load calibration mode.
I.SEL Input selection OC (open collector)/LGC (logic)/MAG (magnet) O.C indication in about 1 second, except for parameter
TR T Tracking zeroing time setup 00 to 99 00 LO side judgment value FSC right after MET is indicated.)
TR V Tracking zeroing width setup *2 00 to 99 01
SNSR Sensor power setup 10(10V)/05(5V) 10 0 (3) Press the Mode key while applying pressure
(5) Press the Shift key (change digit) and press
PÇnÇm Power on delay time setup OFF/ON OFF that will cause the display to show zero.
& the Increment key (change numeric value) to set
PRO Protect setup OFF/1 to 30 OFF to 10.
U-NO. Unit number indication setup OFF/ON ON Judgment GO HI GO LO GO Note:The decimal point in the selected digit
Comparator data
flashes.
S-HI HI side judgment value setup -9999 to 9999 1000
S-LO LO side judgment value setup -9999 to 9999 500 (6) Press the Enter key to return to the (4) For a single display unit, press the Mode key
H-HI HI side hysteresis setup -9999 to 9999 0 or measurement mode (Pressing the Mode key to change to the parameter information indication.
Multi-display unit Single display unit
H-LO LO side hysteresis setup -9999 to 9999 0 changes to the next parameter).
Scaling data
FSC Full scale indication value setup -9999 to 9999 *1
FIN Full scale input value setup -9999 to 9999 *1 (5) Press the Shift key (change digit) and
OFS Offset indication value setup -9999 to 9999 *1 & Increment key (change numeric value) to set
OIN Offset input value setup -9999 to 9999 *1 (1) Press the Mode and Enter keys together Note: For the process signal measuring unit, set the full scale input value 8000.
PS Pre-scaling value setup 0.001 to 5.000 1.000 during measurement. Note:The decimal point in the selected digit
PPR Frequency division setup 1 to 100 1 to 5.000 for the 1 V range and to 20.00 for the 2 A range, and set the flashes.
DLHI Digital limiter HI value setup -9999 to 9999 9999
DLLO Digital limiter LO value setup -9999 to 9999 -9999 offset input value to 1.000 for the 1 V range and to 4.00 for the 2 or
(6) Press the Enter or Mode key to return to the
Analog output HI indication measurement mode.
AOHI -9999 to 9999 9999 (2) Press the Shift key a few times to display the
setup
comparator data menu. A range.
Analog output LO indication
AOLO -9999 to 9999 0
setup
Decimal point indication position
DEP None/place of 100 /place of 101/place of 102/place of 103 None
setup The following explains the frequency measuring unit. (The same method 5.4.4.2 Equivalent Calibration
*2 *2
(3) Press the Mode key a few times to display the
parameter to be set. applies to the full scale indication parameter.) Equivalent calibration means that calibration is carried out according to the
ZERO Zero input value -0.300 to 2.000 0.000
SPIN Span input value 1.000 to 3.000 2.000
SPAN Span indication 0 to 9999 9000 (4) For a single display unit, press the Mode key ratings (specifications) of such a sensor as a load cell. It is not necessary to
to change to the parameter information indication.
(The display automatically changes to this Determining the revolution speed (rpm) using the rotary encoder set to 30 pulses per minut connect the sensor or to apply pressure to the sensor.
indication in about 1 second, except for parameter
5.4.1Method of Setting Condition Data S-HI right after COM is indicated.) (1)Determine the measurement range by calculating the maximum frequency.
Multi-display unit Single display unit
The figure below shows an example where the revolution rises to a maximum speed o
This section shows a typical example of setting the peak hold parameter. about 100 rpm.
(5) Press the Shift key (change digit) and press
The same method applies to other parameters. & the Increment key (change numeric value) to set
to 10. 30 × 100 ÷ 60 = 50
Note:The decimal point in the selected digit (1) Press the Shift, Increment, and Enter key together
flashes. Number of pulses per second during measurement.
Multi-display unit Single display unit (6) Press the Enter key to return to the
or measurement mode (Pressing the Mode key Revolution speed per second
changes to the next parameter).
Number of pulses per revolution at the (2) Press the Increment key to select the equivalent
rotary encoder calibration mode.
(1) Press the Mode and Enter keys together (2)Since the number of pulses determined in (1) is 50 per second (50 Hz), set the range
during measurement. range 11(for how to set the range, see the section on setting condition data). (3) Press the Mode key to move to the equivalent
Note: The setup conditions are HI side judgment value > LO side judgment calibration mode.
(3)The display unit shows 500 if 50 Hz pulse input is measured under range 11 (when
value, HI side judgment value ⊕ LO side judgment value + LO side PS=1 and PPR=1 by default). Therefore, the parameters should be set as PS=2 and
1 digit(100.0 is is indicated 50 Hz
PPR=1 so that the decimal point is positioned in the 10
(2) Press the Mode key to change to the peak
hold setup mode. hysteresis, and LO side judgment value ≤ HI side judgment value-HI input). (4) Press the Shift key to display the zero-input setup
mode.
side hysteresis. If these conditions are not satisfied, an error indica-
(3) For a single display unit, press the Mode key Note:For a single display unit, the unit
to change to the parameter information indication. tion appears and the display returns to the HI side judgment value Note: For the frequency measuring unit, set the relationship between the automatically returns to ZERO indication if there is
(The display automatically changes to this no key operation for about 8 seconds. In such a
indication in about 1 second, except right after setup. input and indication using the PS and PPR parameters (parameters case, press the Mode key to return to the numerical
COND is indicated.) value indication.
5.4.3 Method of Setting Scaling Data of FSC, FIN, OFS, and OIN are not indicated). (5) Press the Shift key (change digit) and the
&
Increment key (change numeric value) to set 0.004.
(4) Press the Increment key a few times to set to This section explains comparator data and shows a typical example of
Valley Hold. Note:The decimal point in the selected digit flashes.
setting the full scale indication parameter. The same method applies to all The following explains the scaling of analog output (The same method
(6) Press the Mode key to change to the span input
(5) Press the Enter key to return to measurement other parameters. applies to the full scale indication parameter.) value setup mode.
or mode. (Pressing the Mode key changes to the Output
next parameter).
Indication
20mA (7) For a single display unit, press the mode key to
Indication : 0.0 to 500.0 display the parameter information indication.
Input voltage : 0 to 10V Output : 4 to 20mA
Indication : 0 to 5.000
5.4 Information on Each Parameter 5.4.2 Method of Setting Comparator Data Multi-display unit Single display unit 5.4.4Method of Setting Calibration Data
This section explains comparator data and shows a typical example of setting
Default 5.4.4.1 Actual Load Calibration
Indication Name Setup options
value the HI side judgment value. The same method applies to all other parameters.
Condition data Actual load calibration means that calibration is carried out by applying actu-
PVH Peak hold setup PH (peak hold)/VH (valley hold)/PVH (peak-valley hold) PH (1) Press the Mode and Enter keys together
HI side judgment value : 900 during measurement.
RANG Measurement range setup *1 *1
HI side hysteresis value : 200
ally measured pressure to a sensor such as a load cell connected to the meter.
Number of averaging operations
AVG 1/2/4/8/10/20/40/80 1 LO side judgment value : 300
setup
Number of moving average LO side hysteresis value : 150
MAV OFF/2/4/8/16/32 OFF Multi-display unit Single display unit
operations setup (2) Press the Shift key a few times to change to
S.UD Step width setup 1(1digit)/2(2digit)/5(5digit)/0(10digit) 1 the scaling data menu.
Indicated value
BLNK Indication blank setup OFF/B-3/B-2/B-1/ON OFF
UNIT Unit setup C/F C HI side judgment value
1000
BAUD Baud rate setup 9600/4800/2400/384(38400)/192(19200) 9600
DATA Data length setup 7(7bit)/8(8bit) 7bit
900
HI side hysteresis (1) Press the Shift, Increment, and Enter keys
P.BIT Parity bit setup E (even number), O (odd number), N (none) E 200 range (3) Press the Mode key a few times to display the together during measurement.
S.BIT Stop bit setup 2(2bit)/1(1bit) 2 parameter to be set.
T- Delimiter setup CR.LF(CR+LF)/CR CRLF
ADR Equipment ID setup 01 to 99 00
500
A.OUT Analog output setup OFF/0-1(0 to 1V)/0-10(0 to 10V)/1-5(1 to 5V)/4-20(4 to 20mA) OFF (4) For a single display unit, press the Mode key
LO side hysteresis to change to the parameter information indication. (2) Press the Mode key to change to the actual
B.UP Digital zero backup setup OFF/ON OFF 150 range
LINE Linearization setup OFF/ON CLR 300 (The display automatically changes to this load calibration mode.
I.SEL Input selection OC (open collector)/LGC (logic)/MAG (magnet) O.C indication in about 1 second, except for parameter
TR T Tracking zeroing time setup 00 to 99 00 LO side judgment value FSC right after MET is indicated.)
TR V Tracking zeroing width setup *2 00 to 99 01
SNSR Sensor power setup 10(10V)/05(5V) 10 0 (3) Press the Mode key while applying pressure
(5) Press the Shift key (change digit) and press
PÇnÇm Power on delay time setup OFF/ON OFF that will cause the display to show zero.
& the Increment key (change numeric value) to set
PRO Protect setup OFF/1 to 30 OFF to 10.
U-NO. Unit number indication setup OFF/ON ON Judgment GO HI GO LO GO Note:The decimal point in the selected digit
Comparator data
flashes.
S-HI HI side judgment value setup -9999 to 9999 1000
S-LO LO side judgment value setup -9999 to 9999 500 (6) Press the Enter key to return to the (4) For a single display unit, press the Mode key
H-HI HI side hysteresis setup -9999 to 9999 0 or measurement mode (Pressing the Mode key to change to the parameter information indication.
Multi-display unit Single display unit
H-LO LO side hysteresis setup -9999 to 9999 0 changes to the next parameter).
Scaling data
FSC Full scale indication value setup -9999 to 9999 *1
FIN Full scale input value setup -9999 to 9999 *1 (5) Press the Shift key (change digit) and
OFS Offset indication value setup -9999 to 9999 *1 & Increment key (change numeric value) to set
OIN Offset input value setup -9999 to 9999 *1 (1) Press the Mode and Enter keys together Note: For the process signal measuring unit, set the full scale input value 8000.
PS Pre-scaling value setup 0.001 to 5.000 1.000 during measurement. Note:The decimal point in the selected digit
PPR Frequency division setup 1 to 100 1 to 5.000 for the 1 V range and to 20.00 for the 2 A range, and set the flashes.
DLHI Digital limiter HI value setup -9999 to 9999 9999
DLLO Digital limiter LO value setup -9999 to 9999 -9999 offset input value to 1.000 for the 1 V range and to 4.00 for the 2 or
(6) Press the Enter or Mode key to return to the
Analog output HI indication measurement mode.
AOHI -9999 to 9999 9999 (2) Press the Shift key a few times to display the
setup
comparator data menu. A range.
Analog output LO indication
AOLO -9999 to 9999 0
setup
Decimal point indication position
DEP None/place of 100 /place of 101/place of 102/place of 103 None
setup The following explains the frequency measuring unit. (The same method 5.4.4.2 Equivalent Calibration
*2 *2
(3) Press the Mode key a few times to display the
parameter to be set. applies to the full scale indication parameter.) Equivalent calibration means that calibration is carried out according to the
ZERO Zero input value -0.300 to 2.000 0.000
SPIN Span input value 1.000 to 3.000 2.000
SPAN Span indication 0 to 9999 9000 (4) For a single display unit, press the Mode key ratings (specifications) of such a sensor as a load cell. It is not necessary to
to change to the parameter information indication.
(The display automatically changes to this Determining the revolution speed (rpm) using the rotary encoder set to 30 pulses per minut connect the sensor or to apply pressure to the sensor.
indication in about 1 second, except for parameter
5.4.1Method of Setting Condition Data S-HI right after COM is indicated.) (1)Determine the measurement range by calculating the maximum frequency.
Multi-display unit Single display unit
The figure below shows an example where the revolution rises to a maximum speed o
This section shows a typical example of setting the peak hold parameter. about 100 rpm.
(5) Press the Shift key (change digit) and press
The same method applies to other parameters. & the Increment key (change numeric value) to set
to 10. 30 × 100 ÷ 60 = 50
Note:The decimal point in the selected digit (1) Press the Shift, Increment, and Enter key together
flashes. Number of pulses per second during measurement.
Multi-display unit Single display unit (6) Press the Enter key to return to the
or measurement mode (Pressing the Mode key Revolution speed per second
changes to the next parameter).
Number of pulses per revolution at the (2) Press the Increment key to select the equivalent
rotary encoder calibration mode.
(1) Press the Mode and Enter keys together (2)Since the number of pulses determined in (1) is 50 per second (50 Hz), set the range
during measurement. range 11(for how to set the range, see the section on setting condition data). (3) Press the Mode key to move to the equivalent
Note: The setup conditions are HI side judgment value > LO side judgment calibration mode.
(3)The display unit shows 500 if 50 Hz pulse input is measured under range 11 (when
value, HI side judgment value ⊕ LO side judgment value + LO side PS=1 and PPR=1 by default). Therefore, the parameters should be set as PS=2 and
1 digit(100.0 is is indicated 50 Hz
PPR=1 so that the decimal point is positioned in the 10
(2) Press the Mode key to change to the peak
hold setup mode. hysteresis, and LO side judgment value ≤ HI side judgment value-HI input). (4) Press the Shift key to display the zero-input setup
mode.
side hysteresis. If these conditions are not satisfied, an error indica-
(3) For a single display unit, press the Mode key Note:For a single display unit, the unit
to change to the parameter information indication. tion appears and the display returns to the HI side judgment value Note: For the frequency measuring unit, set the relationship between the automatically returns to ZERO indication if there is
(The display automatically changes to this no key operation for about 8 seconds. In such a
indication in about 1 second, except right after setup. input and indication using the PS and PPR parameters (parameters case, press the Mode key to return to the numerical
COND is indicated.) value indication.
5.4.3 Method of Setting Scaling Data of FSC, FIN, OFS, and OIN are not indicated). (5) Press the Shift key (change digit) and the
&
Increment key (change numeric value) to set 0.004.
(4) Press the Increment key a few times to set to This section explains comparator data and shows a typical example of
Valley Hold. Note:The decimal point in the selected digit flashes.
setting the full scale indication parameter. The same method applies to all The following explains the scaling of analog output (The same method
(6) Press the Mode key to change to the span input
(5) Press the Enter key to return to measurement other parameters. applies to the full scale indication parameter.) value setup mode.
or mode. (Pressing the Mode key changes to the Output
next parameter).
Indication
20mA (7) For a single display unit, press the mode key to
Indication : 0.0 to 500.0 display the parameter information indication.
Input voltage : 0 to 10V Output : 4 to 20mA
Indication : 0 to 5.000
5.4.5 Method of Setting Linearization Data 5.2 Moving to the Parameter Setup Mode
6.3 Peak Hold Function
The linearization function means a function that changes the slope of straight Calibration data
The Peak Hold function retains one of the maximum (peak hold)/minimum Measurement
lines in the relationship between the input and indication by correcting the rela-
(valley hold)/maximum - minimum (peak-valley hold) values and provides
tions at arbitrary points. Linearization data are set using the input value (indicated
output for that value. Selection from these values is made using the condition
value before correction) and the output value (indicated value after correction) at
data. The peak hold function is enabled by shortcircuiting the PH and COM Shift data
each arbitrary point.
terminals or setting both terminals to the same voltage level.
Multi-display unit Single display unit 7. Output Function
7.1 Comparison Output Function
(1) Press the Mode key and the Enter key during The FD5000 series of unit meters is designed so that the two judgment values
measurement.
HI and LO can be set for the measured (indicated) value to provide the results
of judgment as relay contact output. (This function is effective when the
(2) Press the Shift key a few times to move to the
linearization data menu. meter is equipped with a comparison output unit.) For details on the contact
ratings and other specifications, refer to the section “Output Specifications.”
(3) Press the Mode key to move to the setup for the 7.2 Analog Output Function
number of data to be corrected.
The FD5000 series of unit meters can output an analog signal for an indicated
(4) Press the Shift key (change digit) and the value (when the meter is equipped with an analog output unit). There are four
Increment key (change numeric value) and then press
& output ranges, 0 to 1 V/0 to 10 V/1 to 5 V/4 to 20 mA, from which a selection
the Mode Key to after the number of data to be
corrected has been set.
Note: The decimal point in the selected digit
can be made using the condition data. In addition, the analog output of the
flashes.
FD5000 series allows for arbitrary output scaling. This scaling can be achieved
(5) For a single display unit, press the Mode key to
change to the parameter information indication. by setting the indication value for an output of the maximum scale value (20
Note: For a single display unit, the unit
automatically returns to N-01 indication if there is mA for 4–20 mA output range) in the AOHI parameter of the scaling data.
no key operation for about 8 seconds. In this case,
RE
press the Mode key to return to the numeric value 7.3 RS-485 Interface Function Condition data Comparator data Scaling data Linearization data
indication.
(6) Press the Shift key (change digit) and Increment The FD5000 series can be equipped with an RS-485 interface (when the
& key (change numeric value) and then press the Mode
key after the N-01 input value has been set. meter is provided with an RS-485 unit). For details on the RS-485 function,
DZ
Note: For a single display unit, RE flashes when see the separate manual on communication functions.
the input value is set and DZ flashes when the
output value is set.
(7) Press the Shift key (change digit) and Increment
7.4 RS-232C Interface Function Pressing the ENTER key saves the data and returns to the measurement mode.
& key (change numeric value) and then press the Mode
key after the N-01 output value has been set. The FD5000 series can be equipped with an RS-232C interface (when the
(Data are backed up with EEPROM even when the power is turned off.)
meter is provided with an RS-232C unit). For details on the RS-232C
(8) Press the Mode key to display the setup for output function, see the separate manual on communication functions.
values of the data to be corrected.
5.3 Data Lists and Default Settings
Note: For a single display unit, the unit is
automatically returned to N-02 indication if there is
8. Specifications and External Dimensions
no key operation for about 8 seconds. In this case, Default Equipped Input unit number
RE
press the Mode key to return to the numeric value 8.1 Input Specifications Indication Name value as 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 0 1 2 3 4 5
indication. Condition data
(9) Press the Shift key (change digit) and Increment 8.1.1DC Voltage Measuring Unit (range 11) PVH Peak hold setup PH
& key (change numeric value) and then press the Mode RANG Measurement range setup *1 X X X X
key after the N-02 input value has been set. 11 15 25 13 15 13 15 25 26 25 26 14 B JPB 14 14 2A
Maximum
Measurement Highest Input AVG Number of averaging 1 X X
Range Indication permissible Accuracy
Repeat steps (5) to (9) until all the settings have range resolution impedance MAV Number of moving averaging setup OFF
input
been made. S.UD Step wide setup 1
Offset :± 9999 About
11 ± 99.99mV 10µV ±100V ± (0.1% of FS) BLNK Indication blank setup OFF
Full scale : 0 to± 9999 100MΩ UNIT Unit setup C X X X X X X X X X X X X X X X X
(10) When all the settings have been made, press the BAUD Baud rate setup 9600 X X X X
Enter key to return to the measurement mode.
Input circuit : Single ended type DATA Data length setup 7 X X X X
Operating system : ∆ Σ conversion P.BIT Parity bit setup E X X X X
Maximum sampling rate : 12.5 times per second S.BIT Stop bit setup 2 X X X X
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz) T- Delimiter setup CR.LF X X X X
ADR Equipment ID setup 00 X X X X X
Note: The setup conditions are N-1 < N-2 … N-15 < N-16. 8.1.2DC Voltage Measuring Unit (ranges 12 to 15)
A.OUT Analog output setup OFF X X X X
B.UP Digital zero backup setup OFF
6. Control Functions LINE Linearization setup CLR
I.SEL Input selection OC X X X X X X X X X X X X X X X X X
Maximum TR T Tracking zeroing time setup 00 X X
Highest
6.1 Hold Function Range Measurement Indication Input
permissible Accuracy TR V Tracking zeroing width setup *2 01 X X
range resolution impedance
input SNSR Sensor power setup 10 X X X X X X X X X X X X X X X X X
The Hold function temporarily retains the indication. The hold function is enabled About PÇnÇm Power-on delay setup OFF
12 ± 999.9mV 100µV ± 100V
100MΩ PRO Protect setup OFF
by shortcircuiting the HOLD and COM terminals or setting both terminals to the same About
13 ± 9.999V 1mV ± 250V ± (0.1% of FS) U-NO. Unit number Indication setup ON
voltage level. As a result the display unit retains the indication given at that moment. Offset :± 9999 1MΩ Comparator data
Full scale : 0 to± 9999 About S-HI HI side judgment value setup 1000 X X X X
14 ± 99.99V 10mV ± 250V
6.2 Digital Zero Function 10MΩ S-LO LO side judgment value setup 500 X X X X
About H-HI HI side hysteresis setup 0 X X X X
15 ± 600V 100mV ± 600V ± (0.15% of FS)
The Digital Zero function zeros the indication given at an arbitrary timing. 10MΩ H-LO LO side hysteresis setup 0 X X X X
Scaling data
Thereafter, the function shows the amount of change from the point of zero- Input circuit : Single ended type FSC Full scale Indication value setup *1 X X X X X
Operating system : ∆ Σ conversion 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999
ing. However, this function serves as an indication resetting function for a Maximum sampling rate : 12.5 times per second FIN Full scale input value setup *1 X X X X X
frequency measuring unit. Thus, the Digital Zero function can be used to reset Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz) 9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999 *3
OFS Offset indication value setup *1 X X X X X
the indication when there is no input signal at all. 0 0 0 0 0 0 0 0 0 0 0 0 0
OIN Offset input value setup *1 X X X X X
Note that, the on/off control of the Digital Zero function can be achieved by 8.1.3DC Current Measuring Unit 0 0 0 0 0 0 0 0 0 0 0 0 *4
means of terminal control or front panel keys. In the case of terminal control, PS Pre-scaling value setup 1 X X X X X X X X X X X X X X X X
PPR Frequency division setup 1 X X X X X X X X X X X X X X X X
the Digital Zero function is turned on by shortcircuiting the DZ and COM termi- Maximum DLHI Digital limiter HI value setup 9999 X X X X X X X X
Measurement Highest Input
Range Indication permissible Accuracy DLLO Digital limiter LO value setup -9999 X X X X X X X X
range resolution impedance
nals or setting both terminals to the same voltage level. The indication at that input AOHI Analog output HI indication setup 9999 X X X X
About AOLO Analog output LO indication setup 0 X X X X
moment is zeroed. In the case of control with the front panel keys, hold down 23 ± 9.999mA 1µA
10Ω
± 100mA
± (0.2% of FS) DEP Decimal point position setup None
the Mode key and press the Increment key for about 1 second to zero the indi- Offset :± 9999 About *5
24 ± 99.99mA Full scale : 0 to± 9999 10µA 1Ω ± 500mA
cation at that moment. 25 ± 999.9mA 100µA
About
± 3A ± (0.3% of FS) ZERO Zero input value *6 0 X X X X X X X X X X X X X X X X X
0.1Ω SPIN Span input value *6 2000 X X X X X X X X X X X X X X X X X
Note: Operation with the control terminals takes priority over operation SPAN Span indication 9000 X X X X X X X X X X X X X X X X X
with the front panel keys. The Digital Zero function is disabled if Input circuit : Single ended type Shift data
Operating system : ∆ Σ conversion SHF Shift data setup 0
the control terminals are made to go through the off-on-off se- Maximum sampling rate : 12.5 times per second *1 Each value in the lower part of a cell in the columns on the right is the default value.
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz) *2 Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
quence with the function enabled by means of the front panel keys. *3 5000 for 1 V range and 2000 for 2 A range
*4 1000 for 1 V range and 400 for 2 A range
*5 Linearization data are not set up for the default values.
*6 This value is not indicated if calibration is done using an actual load.
The shaded parts show the parameters that must be set for each unit.
Instruction Manual for FD5000 Series 4/12 Instruction Manual for FD5000 Series 9/12
8.1.4 AC Voltage Measuring Unit (average value detection: ranges 8.1.9 AC Current Measuring Unit (average value detection: range
Pressing the Mode and Enter keys Pressing the Shift and Enter keys
together changes to the parameter Selects the item to be set. together changes to the parameter 11 to 13) 26)
setting mode. checking mode. (Comparator data Changes the digit to be set.
Pressing the Mode and Shift keys can be set.) Maximum
Maximum Highest
Pressing the Shift and Mode keys Highest Measurement Input
(7) Mode key together changes to the shift function Range Measurement Indication Input
permissible Accuracy Range
range
Indication
resolution impedance
permissible Accuracy
setup mode. together changes to the shift function range resolution impedance input
input
Pressing the Mode and Incremental (8) Shift key setup mode. Offset :± 9999 ± (0.5% of FS
keys together turns on/off the"Digital Holding down the Shift key for about 11 99.99mV 10µV 26 5A 1mA (CT) 8A
Full scale : 0 to± 9999 +10digit)
zero" indicator. one second moves to the HI judgment ± 100V
12 999.9mV Offset :± 9999 100µV 1MΩ ± (0.2% of FS
Pressing the Shift and Enter keys value indicator. Full scale : 0 to± 9999 or more +10digit) Input circuit : CT isoration type
together changes to the parameter Selects from items in the maximum/ Operating system : ∆ Σ conversion
Changes the digit to be set. minimum/(maximum-minimum)/input 13 9.999V 1mV ± 250V Maximum sampling rate : 12.5 times per second
checking mode. (Comparator data
can be set.) value monitoring mode. (Hold down Frequency range : 50 Hz or 60Hz
Pressing the Shift and Mode keys the key for about one second.) Input circuit : Single ended type Response speed : About 1 second
(8) Shift key together changes to the shift function Pressing the Increment and Mode Changes the value or content Operating system : ∆ Σ conversion Crest factor : 4:1 at full scale
setup mode. keys together turns on/off the"Digital of a selected digit. Maximum sampling rate : 12.5 times per second Dead zone : 0 to 99 digits
Selects from items in the maximum/ zero" indicator. (Increments the value) Frequency range : 40 Hz to 1 kHz
minimum/(maximum-minimum)/input Holding down the Increment key for Response speed : About 1 second
value monitoring mode. (Hold down about one second moves to the LO Dead zone : 0 to 99 digits 8.1.10 AC Current Measuring Unit (true rms value: ranges 23 to 25)
the key for about one second.) judgment value indicator.
Pressing the Increment and Mode Changes the value or content (9) Increment key Pressing the Increment and Enter
keys changes to the maximum/ Maximum
keys together turns on/off the"Digital of a selected digit. Highest
8.1.5 AC Voltage Measuring Unit (average value detection: ranges Range Measurement Indication Input
permissible Accuracy
zero" indicator. (Increments the value) minimum/(maximum-minimum)/input range resolution impedance
input
Pressing the Increment and Enter value monitoring mode. 14 and 15) About
keys changes to the maximum/ Resets the maximum/minimum/ 23 9.999mA 1µA 100mA
10Ω
(9) Increment key minimum/(maximum-minimum)/input (maximum-minimum)/input value Offset :± 9999 About ± (0.5% of FS
value monitoring mode. monitoring mode. (Hold down the key Maximum 24 99.99mA 10µA 500mA
Measurement Highest Input Full scale : 0 to± 9999 1Ω +20digit)
Resets the maximum/minimum/ for about one second.) Range Indication permissible Accuracy
range resolution impedance About
(maximum-minimum)/input value input 25 999.9mA 100µA 3A
± (0.2% of FS 0.1Ω
monitoring mode. (Hold down the key 14 99.99V 10mV 250V
Offset :± 9999 1MΩ +10digit)
for about one second.) Full scale : 0 to± 9999 or more ± (0.3% of FS
Input circuit : Single ended type
5. Parameter Setup 15 600V 100mV 600V +10digit) Operating system : ∆ Σ conversion
Maximum sampling rate : 12.5 times per second
5.1 Differences between Display Units Input circuit : Single ended type Frequency range : 40 Hz to 1 kHz
4.2 Single Display Unit Operating system : ∆ Σ conversion Response speed : About 1 second
Maximum sampling rate : 12.5 times per second Crest factor : 4:1 at full scale
Frequency range : 40 Hz to 1 kHz Dead zone : 0 to 99 digits
(2) (1) (3) (1) (3) Response speed : About 1 second
Dead zone : 0 to 99 digits
8.1.11 AC Current Measuring Unit (true rms value: range 26)
Maximum
Measurement Highest Input
8.1.6 AC Voltage Measuring Unit (true rms value: ranges 11 to 13) Range
range
Indication
resolution impedance
permissible Accuracy
input
Offset :± 9999 ± (0.5% of FS
26 5A 1mA (CT) 8A
Maximum Full scale : 0 to± 9999 +20digit)
Highest
Setup contents Parameter name Range Measurement
range
Indication Input
permissible Accuracy
resolution impedance
input Input circuit : CT isoration type
(6) (7) (8) (9) (6) (7) (8) (9) Operating system : ∆ Σ conversion
11 99.99mV 10µV
Main Functions 5.1.1 Multi-display Unit ± 100V
Maximum sampling rate : 12.5 times per second
No. Name 12 999.9mV Offset :± 9999 100µV 1MΩ ± (0.2% of FS Frequency range : 50 Hz or 60Hz
During measurement During parameter setup Full scale : 0 to± 9999 or more +20digit) Response speed : About 1 second
IIndicates information on the Crest factor : 4:1 at full scale
Indicates the measured value. 13 9.999V 1mV ± 250V
parameter to be set. Dead zone : 0 to 99 digits
(1) Main display Indicates information on the item in the
maximum/minimum/(maximum- Input circuit : Single ended type
minimum)/input value monitoring mode. Operating system : ∆ Σ conversion
Maximum sampling rate : 12.5 times per second 8.1.12 Resistance Measuring Unit
Indicates the result of judgment and
HI turns on if the measured value > HI Frequency range : 40 Hz to 1 kHz
judgment value.
Response speed : About 1 second
Crest factor : 4:1 at full scale Measurement Highest Circuit
Indicates the result of judgment and Range Indication Accuracy
Judgment Setup contents and parameter name Dead zone : 0 to 99 digits range resolution current
(2) GO turns on if LO judgment value≤ the
indicators 11 99.99Ω 10mΩ
About
measured value ≤ HI judgment value. 5mA
Indicates the result of judgment and Note: Pressing the Mode key displays the next parameter. 8.1.7 AC Voltage Measuring Unit (true rms value: ranges 14 and About
LO turns on if the measured value < LO 12 999.9Ω Offset :± 9999 100mΩ
5.1.2 Single Display Unit 15) 500µA ± (0.2% of FS)
judgment value. Full scale : 0 to± 9999 About
13 9.999kΩ 1Ω
Turns on if "peak hold/valley hold/peak - Note 1: Pressing the mode key with the parameter name shown changes the 50µA
PH Maximum About
valley hold" is on. Measurement Highest Input 14 99.99kΩ 10Ω
display to the parameter information indication. If there is no key Range Indication permissible Accuracy 5µA
Turns on if "digital zero" is on. range resolution impedance
input
DZ Flashes when linearization data output
operation for about one second when the parameter name is shown, ± (0.2% of FS
values are set. 14 99.99V 10mV 250V Input circuit : Single ended type
Function Offset :± 9999 1MΩ +20digit) Operating system : ∆ Σ conversion
(3)
indicators
ME Turns on if "digital zero backup" is on. the display automatically changes to the parameter information Full scale : 0 to± 9999 or more ± (0.3% of FS Maximum sampling rate : 12.5 times per second
15 600V 100mV 600V
Turns on if remote control is being +20digit) Measuring system : Two-wire system or four-wire system
performed through RS-232C or RS-485 indication (however, this change does not automatically occur for (internal socket change-over)
RE interface. parameters PH/S-HI/FSC, etc., right after COND/COM/MET is in- Input circuit : Single ended type Open-circuit voltage : About 5 V
Flashes when linearization data input Operating system : ∆ Σ conversion
values are set. Maximum sampling rate : 12.5 times per second
dicated). Frequency range : 40 Hz to 1 kHz
Pressing the Mode and Enter keys Response speed : About 1 second 8.1.13 Temperature Measuring Unit (TC)
Returns to the measurement
together changes to the parameter
mode.
Note 2: Pressing the Mode key when the parameter information indication Crest factor : 4:1 at full scale
setting mode. Dead zone : 0 to 99 digits
Pressing the Enter and Increment keys is shown results in the next parameter being displayed. Input Highest
Range Indication Accuracy
together changes to the maximum/ sensor resolution
Note 3: If there is no key operation for about 8 seconds with the parameter
minimum/(maximum-minimum)/input 8.1.8 AC Current Measuring Unit (average value detection: ranges -50.0 to 199.9°C 0.1°C
(6) Enter key value monitoring mode. information indication shown, the display returns to the parameter KA K ± (0.5% of FS)
(-58.0 to 391.8°F) (0.1°F)
Switches from the maximum/minimum/ 23 to 25) -50 to 1200°C
name indication. KB K
(-58 to 2192°F)
(maximum-maximum/minimum/ ± (0.2% of FS)
(maximum-minimum)/input value Maximum -50 to 1000°C
Highest J J
monitoring mode to the comparative Range Measurement
range
Indication Input
permissible Accuracy (-58 to 1832°F)
resolution impedance -50 to 400°C
judgment reading mode. input T T ± (0.6% of FS)
About (-58 to 752°F) 1°C
Pressing the Mode and Enter keys 23 9.999mA 1µA 100mA 0 to 1700°C
10Ω S S (1°F)
together changes to the parameter Selects the item to be set. Offset :± 9999 About ± (0.5% of FS (32 to 3092°F)
setting mode. 24 99.99mA 10µA 500mA ± (0.4% of FS)
Full scale : 0 to± 9999 1Ω +10digit) -10 to 1700°C
R R
Pressing the Mode and Shift keys About (14 to 3092°F)
25 999.9mA 100µA 3A ± (0.4% of FS)
(7) Mode key together changes to the shift function 0.1Ω 100 to 1800°C
setup mode. B B Note : The accuracy of range B is applicable
(212 to 3272°F)
Pressing the Mode and Incremental Input circuit : Single ended type to temperatures of 500°C or more.
keys together turns on/off the"Digital Operating system : ∆ Σ conversion
Maximum sampling rate : 12.5 times per second Input circuit : Single ended type
zero" indicator. Frequency range : 40 Hz to 1 kHz Operating system : ∆ Σ conversion
Response speed : About 1 second Maximum sampling rate : 6.25 times per second
Crest factor : 4:1 at full scale Cold junction compensation error : ± 2°C(at 10 through 40°C)
Dead zone : 0 to 99 digits Internal resistance of sensor : 50Ω or less
Linearizer : Digital linearizer
Burnout alarm : It blinks by ---- display.
Instruction Manual for FD5000 Series 10/12 Instruction Manual for FD5000 Series 3/12
8.1.14 Temperature Measuring Unit (RTD) 8.2 Common Specifications 3.3.10 Temperature Measuring Unit (RTD) 3.5 Analog Output
3.3.4 AC Voltage Measuring Unit (Ranges 11 to 13) Conversion system : PWM conversion
3. Terminal Arrangement Resoluton : Equivalent to 13 bits
No. Name Description
Scaling : Digital scaling
Comparison Serial Positive input terminal for ranges
Analog output 1 11-12
output communication 11 (99.99 mV) and 12(999.9 mV)
Positive input terminal for range
Response speed : About 0.5 second
1 2 3 2 13
13 (9.999 V)
3 LO Common input terminal 8.3.3 Communicating Function
12 13 14 15 16 17 18 19 20 22 24 25 26
D5XXX - XX
Input unit
MODEL FD5000 SERIES 01: DC voltage measuring unit (range 11: ± 99.99 mV)
INSTRUCTION MANUAL 02: DC voltage measuring unit
(range 12: ± 999.9 mV ; range 13: ± 9.999 V)
(range 14: ± 99.99 V ; range 15: ± 600 V)
03: DC current measuring unit
(range 23: ± 9.999 mA; range 24: ± 99.99 mA)
(range 25: ± 999.9 mA)
04: AC voltage measuring unit (average rms)
(range 11: 99.99 mV; range 12: 999.9 mV)
(range 13: 9.999 V)
05: AC voltage measuring unit (average rms)
(range 14: 99.99 V; range 15: 600 V)
06: AC voltage measuring unit (true rms)
! Caution (range 11: 99.99 mV; range 12: 999.9 mV)
(range 13: 9.999 V)
(1) Do not apply a voltage or current exceeding the maximum allowable 07: AC voltage measuring unit (true rms)
value; otherwise, it may damage the equipment. (range 14: 99.99 V; range 15: 600 V)
(2) Use a power voltage within the operation range; otherwise, it may 08: AC current measuring unit (average rms)
result in a fire, electrical shock, or malfunction. (range 23: 9.999 mA; range 24: 99.99 mA)
(3) The contents of this manual are subject to change without notice. (range 25: 999.9 mA)
(4) Although the contents of this manual have been prepared with extra 09: AC current measuring unit (average rms)
care, if you have any questions, or find errors or missing (range 26: 5 A)
information, contact the sales agent from which you purchased the 10: AC current measuring unit (true rms)
product or Fuji Electric Instruments Co.,Ltd. (range 23: 9.999 mA; range 24: 99.99 mA)
(5) After reading this manual thoroughly, keep it in a convenient place (range 25: 999.9 mA)
for future reference. 11: AC current measuring unit (true rms)
(range 26: 5 A)
12: Resistance measuring unit
13: Temperature measuring unit (TC)
1. Before Using the Product 14: Temperature measuring unit (RTD)
Thank you for purchasing the FD5000 series. This manual should be passed 15: Frequency measuring unit
on to the person who operates the product. Examine the product for dam- (inputs: open collector, logic, and magnet)
16: Frequency measuring unit (input: 50 to 500 Vrms)
age caused by transportation or any other defects. If you find any damage or
17: Strain gauze input unit (load cell)
defects, contact the sales agent from which you purchased the product or
18: Process signal measuring unit
Fuji Electric Instruments Co., Ltd. (4 to 20 mA or 1 to 5 V)
1.1 Model Codes
The model lineup of the FD5000 series is shown below. Check that the 2. Mounting the Product
model code and specifications of your product match those you specified 2.1 Dimensions for Cutting Panel
when ordering.
Cut the panel for mounting according to the following dimensions.
FD5XXX - XX 92 +0.8
-0
Power unit
1: 100 to 240 V AC ± 10% 45
2: 9 to 60 V DC
-0 +0.6
Display unit
1: Single display
2: Multi display
70 mm
Output unit min.
0: None
1: Comparison
2: Analog
3: RS-232C
4: RS-485
5: Comparison and analog
6: Comparison, analog, and RS-232C
7: Comparison, analog, and RS-485 120 mm min.