SPECIFICATIONS
Method: Spectrophotometric (DR 2800)
Range: 0.1 to 5.0 mg/L as PO4
Testing Time: 20 minutes
DESCRIPTION
The Total Inorganic Phosphate DR 2800 Procedure
was designed to help control inorganic phosphate
levels in most boilers and cooling water systems
without having to submit samples for laboratory
analysis. Inorganic phosphate-based products are
used to help control both scaling and corrosion in
boilers and cooling water systems. It is important
to monitor phosphate levels to ensure that the
proper level and type of phosphate are maintained
in the system. The test does not measure organic
phosphate. Nalco DR 2800 Spectrophotometer
ORDERING INFORMATION
Order all replacement parts and reagents by their Description Part No.
part numbers. To be sure this product is compatible 25 mL glass graduated cylinder
with your treatment program, contact your local with plastic cap 500-P1204.88
Nalco Sales Engineer. To place your order, please 125 mL glass Erlenmeyer flask
contact your local customer service department. (2 required) 500-P1503.88
3-1/2", 325 watt round hot plate
(2 required) 500-P1813.88
REPLACEMENT PARTS AND REAGENTS Teflon boiling chips, 2 oz
Required Parts (60 mL) bottle 500-P1725.88
5 mL serological pipet 500-P2171.88
Description Part No. 10 mL Pi-Pump Pipetter 500-P2721.88
Nalco DR 2800 Spectrophotometer 6017050
Reagent set, Ortho and Total
Inorganic Phosphate 460-P3109.88
Nalco Pressure Filter
(with filter membranes 420-C0091.88
Accessories INTERFERENCES
Description Part No. The species listed below do not interfere under the
Sample Cell, 10-mL, 1-inch glass stated values.
square matched pair 500-P2500.88
Sample Cell, 25-mL, 1-inch glass Species mg/L
square matched pair 500-P2502.88 Fe3+ 150
Pour through cell assembly kit 500-P2525.88 Ba2+ 50
47 mm, Filter Membrane, Cl– 5000
0.45 micron, 100/pkg 500-P1405.88 Ca2+ 400
1 mL plastic dropper pipet Pb2+ 20
with bulb 500-P0015.88 Cu2+ 20
2 oz (60 mL) dropper bottle 500-P0359.88 COD 500
1:1 HCl, 1.7 oz/50 mL 460-S0726.71 Al3+ 20
1:1 HCl, 3.4 oz/100 mL* 460-S0726.72 Mg2+ 1000
1:1 HCl, 16.9 oz/500 mL 460-S0726.74 CrO42– 60
1:1 HCl, 34.8 oz/1 L 460-S0726.75
1:1 HCl, 1.06 gal/4 L 460-S0726.77
TP-1, 3.4 oz/100 mL* 460-SH357.72 SAFETY PRECAUTIONS
TP-1, 16.9 oz/500 mL 460-SH357.74 Refer to Material Safety Data Sheet (MSDS) for
TP-1, 34.8 oz/1 L 460-SH357.75 specific reagent personal safety and disposal
XP-2, 0.9 oz/25 g 460-S5091.80 information.
XP-2, 3.5 oz/100 g 460-S5091.82
XP-2, 8.8 oz/250 g 460-S5091.83
DI Water, 1.06 gal/4 L 460-S0470.77 SUPPORT
Pressure Filter Unit Only 500-C0092.88 If you have any questions regarding this procedure,
47 mm glass fiber filter pad please contact your Nalco Sales Engineer. In North
(100/pkg) 500-P1413.88 America you can also contact the Nalco Equipment
47 mm, 0.45 micron filter Solutions Help Desk at 1-800-323-8483.
membrane (100/pkg) 500-P1405.88
4 oz. (120 mL) plastic bottle
with 0.5–1.0 mL pipet 500-P0346.88
0.5–1.0 mL pipet, standard size 500-P2120.88
0.5–1.0 mL pipet/cap/bulb
assembly 500-P2179.88
Blue pipet bulb for 0.5–1.0 mL pipet 500-P2403.88
Orange pipet bulb for 1 mL plastic
dropper pipet 500-P2404.88
1 oz (30 mL) vial with
0.5 gram scoop 500-P0350.88
0.1 and 0.5 gram scoops
(3 of each size) 500-P0354A.88
Dial thermometer, 50 to 300°F 500-P2019.88
Digital thermometer, –40 to 450°F 500-P2098.88
Lab thermometer, –10 to 260°C,
glass 500-P2020.88
Lab thermometer, 0 to 230°F, glass 500-P2028.88
TOTAL INORGANIC PHOSPHATE DR 2800 PROCEDURE
Zero Read
11 12 13 14
When the timer beeps, place Touch ZERO. Place the sample cell Touch READ
the cell containing the blank containing the prepared
solution into the cell holder. The display will show: sample into the cell holder. The results in mg/L PO4 will
3– be displayed.
0.0 mg/L PO4