Professional Documents
Culture Documents
The manufacturer's reference number of the instrument manual(s) and its location
A reference to the keeping of a record of any changes to the above and its location
Apart from the SOP itself ( but nothing to do with an SOP as such) it might be a local
requirement for anyone using the instrument be a trained person and a record held of that
training.
Regards
1. Purpose: To describe a procedure for the operation, calibration and maintenance of Gas
Chromatograph/Head Space (GC-HS).
2. Scope: This SOP applies to the Gas Chromatograph, GC-HS (Make: Agilent Technologies; Model:
6850/7694E located in instruments section of analytical laboratory.
4. Procedure
The primary sources of information for operation, maintenance and calibration
are the manufacturer's manuals. This SOP, as written, highlights the important sections of the manuals.
4.1 Operation
4.1.1 Ensure that the instrument is clean, free from dust and vibration.
4.1.2 Ensure that environment condition are in suitable (Temp 15-35°C) Humidity 40-60%.
4.1.3 Ensure that part of instrument and system in configuration.
4.1.4 Turn on the cylinders of nitrogen (grey with a black top), air (grey) and hydrogen (red) kept
in the gas station on the ground floor. Always check the pressure level on the regulators.
4.3.2 The frequency selected to carry out the calibration is once a year for instrument performance. Test solutions
and STD used or recommended by vendors are acceptable if Certificates of Analysis (COAs) of these
are available. Use test solutions or STDs traceable to NIST or International Reference if they
are available.
4.3.3 Service Qualification: An analyst or a qualified service representative is qualified for calibration. Training
certifications must be provided prior to commencing work.
4.3.4 Column: Capillary column would be used for calibration of split and split less injector with
appropriate standards.
4.4.5 Signal Noise and Drift: The base signal is recorded at the beginning of the test, signal noise is
calculated as ASTM noise which is average peak to peak noise in a number of signal segments and
signal drift is calculated as the slope of the linear regression of the signal. The acceptance limits are as
follows:
If the test fails, arrange a service and place an “Out of Service” sticker on the instrument.
5. Responsibilities: It is the responsibility of the section in charge to ensure that staff are trained in all
aspects of this SOP or deemed competent in its application. It is the responsibility of the operator to
ensure compliance with this document.
6. References
6.1 Agilent service manuals, Part # G
6.2 Agilent 7694E Automatic Head Space Sampler Installation Guide, Part #
7. Appendices
8. Revision History
SOP No. Supersedes No. & Date Changes made (in brief)
Effective Date:
Supersedes:
Author(s):
Signature
Date
1. Purpose: To describe a procedure for the operation, calibration and maintenance of Gas
Chromatograph/Mass spectroscopy (GC/MS).
2. Scope: This SOP applies to the Gas Chromatograph/Mass spectroscopy, GC/MS (Make: Agilent
Technologies; Model: 7890A/5975C) located in instruments section of analytical laboratory.
4. Procedure
The primary sources of information for operation, maintenance and calibration
are the manufacturer's manuals. This SOP, as written, highlights the important sections of the manuals.
4.1 Operation
4.1.1 Ensure that Instrument is clean , free from dust and vibration
4.1.2 Ensure that Environment conditions are in suitable (Temp. 15 -35 °C) Hum. 40 – 60 %
4.1.5 ‘ON’ the GC & ‘ON’ the MS, wait for some time till turbo pump speed is 100% of full.
4.1.9 For run the sample :- Click on method , Click on run method, a window will display , gave
the sample information and, Click OK & RUN method
4.2.0 After complete the run time & post run time , the data file of unknown sample is compare
with std curve data file ( on data analysis) , & take print.
4.2.3 Wait for some time till , turbo pump speed >50%, MS Source temp > 100°C, MS Quad
temp > 100 °C
4.3.2 The frequency selected to carry out the calibration is once a year for instrument performance. Test solutions
and STD used or recommended by vendors are acceptable if Certificates of Analysis (COAs) of these
are available. Use test solutions or STDs traceable to NIST or International Reference if they
are available.
4.3.3 Service Qualification: An analyst or a qualified service representative is qualified for calibration. Training
certifications must be provided prior to commencing work.
4.3.4 Column: Capillary column would be used for calibration of split and splitless injector with
appropriate standards.
4.4.1 Temperature accuracy and stability: Herein compare selected detector set point temperature
versus actual system reading (refer to Agilent’s user manual for set temperature recommendations).
Temperature accuracy at two set temp is measured at 230° C and 100 ° C. Actual temperature reading
should be within ±1% of the set point. The temperature stability should be ≤ 0.5 ° C.
4.4.2 MS Mass Calibration: for this purpose we have used PFTBA calibrant as a automatic calibration
system by the auto tuning for mass 69, 219 and 512.
4.4.3 Inlet pressure decay: This test demonstrates the pressure integrity of the GC inlet (with a valve
controlled injection system, if applicable) and all flows controlled by GC inlet pneumatics. The acceptable
limit for pressure change should be ≥ -2.0 and <0.5 psi/5min.
4.4.4 Inlet pressure accuracy: This test demonstrates the ability of the system to provide accurate
pressure to the head of the column. The acceptance limit should be ≤ 1.2.
4.4.5 Signal Noise and Drift: The base signal is recorded at the beginning of the test, signal noise is
calculated as ASTM noise which is average peak to peak noise in a number of signal segments and
signal drift is calculated as the slope of the linear regression of the signal. The acceptance limits are as
follows:
If the test fails, arrange a service and place an “Out of Service” sticker on the instrument.
5. Responsibilities: It is the responsibility of the section in charge to ensure that staff are trained in all
aspects of this SOP or deemed competent in its application. It is the responsibility of the operator to
ensure compliance with this document.
6. References
6.1 Agilent service manuals, Part # G 1530-90220
6.2 Agilent 6890 Automatic Liquid Sampler Installation Guide, Part # G2612- 90107
7. Appendices
8. Revision History
SOP No. Supersedes No. & Date Changes made (in brief)
END OF THE DOCUMENT