Professional Documents
Culture Documents
Sistema de Fuerza Tridimensional
Sistema de Fuerza Tridimensional
When both vertical alignment and first-order rotation of teeth are to occur simultaneously, a 3-dimensional
force system is required. This numerical study evaluated several appliances (rectangular loops and L-loops)
used to vertically align teeth. Consideration was given to how these designs might be modified to produce the
appropriate force system to allow both movements to occur simultaneously. It was found that the rectangular
loop was the most appropriate choice for first-order corrections. For the rectangular loops studied, the in-plane
force system was shown to be essentially independent of the out-of-plane effects, which allowed the 2
corrections to be controlled separately. (Am J Orthod Dentofacial Orthop 2001;119:21-9)
V
ertical alignment of teeth presents challenges in these studies investigated the effects of out-of-plane
terms of designing an appliance that will pro- (OOP) designs and/or activation on the force and
vide appropriate force systems and will readily moment systems created.
fit into the limited space available. This usually requires More recently, Menghi et al3 studied R, L, and T-
force application through the long axis of the tooth (sec- loops experimentally with a view to establishing a 3-
ond order). In addition to vertical alignment, it is often dimensional description of the force systems delivered
desirable to rotate these teeth through the application of by these 3 loop configurations for both stainless steel
a moment (first order) or to move them buccolingually. and titanium molybdenum alloy (TMA) for first-order
If both corrections are to be performed simultaneously, corrections. All the specimens that were tested had
then a 3-dimensional force system is necessary. nominal cross-sections of .017 × .025 in (.432 × .635
Vanderby et al1 performed an extensive experimental mm) and were configured for a 21-mm IBD. In addi-
study of vertically activated orthodontic loops. They tion to the measurement of the forces in the plane of
considered T, L, and rectangular configurations designed the appliance, the 3-dimensional force system was
for an interbracket distance (IBD) of 7 mm. The mater- measured during buccolingual movements and for
ial used was .010 × .021-in (.254 × .533-mm) stainless ±10° rotations of the end of the loop around the long
steel wire. For vertical activations of 3 mm and an IBD axis of the tooth. The authors concluded that correc-
of 7 mm, these authors found a considerable nonlinear tion of first-order discrepancies was best accom-
effect (ie, the amount of activation was not linearly plished using rectangular loops.
related to the vertical force and moment generated). In The present study is a numerical simulation of the
addition, they suggested that the forces and moments 3-dimensional force system produced by modifications
generated by the stainless steel loops were sensitive to of existing 2-dimensional designs and is to provide a
small inaccuracies during loop fabrication. Koenig et al2 more thorough understanding of the appliance design
used an analytic approach to determine the force systems parameters and of the effect of alteration of the planar
produced by appliance activations but documented lim- designs. The simulation also allows the study of these
ited results with regard to rectangular loops. Neither of 3-dimensional effects so that design guidelines can be
established, thereby allowing rotational movements to
Supported by the Natural Sciences and Engineering Research Council of occur simultaneously with the vertical displacements.
Canada, grants RGP 7514 and RGP 6926.
aAssistant Professor, Department of Mechanical Engineering, University of MATERIAL AND METHODS
Alberta, Edmonton, Alberta, Canada.
b,cProfessor, Department of Mechanical Engineering, University of Alberta, Simulation of the force systems that were generated
Edmonton, Alberta, Canada. and the effect of parametric changes to the appliance
dPrivate Practice, Edmonton, Alberta, Canada.
design was performed with the use of a previously
Reprint requests to: M.G. Faulkner, PhD, Department of Mechanical Engineer-
ing, University of Alberta, Edmonton, Alberta, Canada T6G 2G8; e-mail, described numerical method.4,5 This method uses a com-
gary.faulkner@ualberta.ca. plete nonlinear formulation in which the appliance is bro-
Submitted, August 1999; revised and accepted, May 2000. ken into several segments. These segments are analyzed
Copyright © 2001 by the American Association of Orthodontists.
0889-5406/2001/$35.00 + 0 8/1/110810 sequentially, starting at one end of the appliance. Force
doi:10.1067/mod.2001.110810 and displacement compatibility between the segments
21
22 Raboud et al American Journal of Orthodontics and Dentofacial Orthopedics
January 2001
Fig 1. Rectangular loop dimensions in mm; varying the horizontal position of the posterior vertical segment.
was used to move from one segment to the next. The advan- (1.00 × 107 psi) and a yield stress of 1240 MPa (180 ×
tage of this approach (which differs from others, such as the 103 psi). These designs were assumed to be prefabri-
finite element method) is that the large 3-dimensional cated to the rectangular and L-loop shapes shown in
deformations that occur can be readily incorporated into the Figs 1 and 2, respectively. To facilitate comparisons
procedure. In addition, changes in material and/or wire size among the different designs, all the loops that were
can be easily accommodated. This numerical approach has studied were fit into brackets with a 5-mm IBD. This
been previously verified by a comparison of the results with was believed to closely resemble the clinical geometry
both experimental measurements and the results of alterna- of a maxillary permanent first molar and a second pre-
tive numerical simulations, such as the finite element molar. (Note that the configuration of the L-loop dic-
method applied to specific T-loop appliances.6 tates that both vertical segments are situated within the
The appliance designs that were simulated were 5.0-mm IBD.) The rectangular loop design places each
assumed to be fabricated from .017 × .025-in (.432 × of the vertical segments anterior to the brackets, thus the
.635-mm) TMA with a Young’s modulus of 69 GPa distance between the points at which the wires enter the
American Journal of Orthodontics and Dentofacial Orthopedics Raboud et al 23
Volume 119, Number 1
Planar appliances
Fig 3 A, shows the typical placement of a planar rec-
tangular loop into a pair of maxillary brackets. For the
purpose of discussion, the right end of each appliance
was considered the anterior end. The appliances were
activated by the insertion of the anterior end fixed and
activated vertically with respect to the posterior end
(also fixed). For the anterior end, gingival activation
was considered positive, and occlusal activation was
considered negative. The positive directions of the force
and moment components on the teeth are shown in Fig
3 B. As seen in Fig 3 C, the directions of the positive
A
forces and moments on the appliance were then equal
and opposite to those shown on the teeth. For these pla-
nar appliances force and moment equilibrium leads to:
FA = FP (1)
VA = VP (2)
MP + MA – FA (a) + VA (b) = 0 (3)
where the subscripts A and P correspond to anterior
and posterior, respectively. This serves as verification
of the results for any given activation.
For each activation, the forces and moments neces-
sary for equilibrium were used to determine the
stresses within the appliance. When the activation for B
the specific design resulted in any of these stresses
reaching the elastic limit of the material, this was
defined as the maximum activation because any further
activation would cause the appliance to become perma-
nently deformed compared with the original shape. No
consideration was given to deformations into the plas-
tic range beyond this limit.
3-Dimensional appliances
To appreciate the effect that OOP designs had on the
force system produced, the rectangular loop design with
maximum spacing between the vertical segments (Fig 1,
#1) is modified by pretwisting (first-order activation)
the vertical anterior segment through angles of over 30°
in both the labial and lingual directions. This modified C
appliance is then installed in the same bracket configu-
ration as the original planar appliance and activated to
investigate the new force systems that are developed. Fig 3. Vertical alignment with the use of rectangular
loop. A, Activation of appliance. B, Force system on
An equilibrium analysis, similar to that shown in
teeth. FP, Posterior horizontal forces; MP, in-plane
equations 1 to 3 can be performed for the 3-dimen- moment acting on posterior tooth; VP,intrusive force act-
sional systems; however, the equations are more com- ing on posterior tooth; VA, extrusive force acting on ante-
plex because the analysis involves 3 moments and 3 rior tooth; MA, in-plane moment on anterior tooth; FA,
forces at each end. As a result, these equations are not anterior horizontal forces. C, Force system on appliance.
shown here but were used as a check on the results a, Activation; b, horizontal distance between points at
delivered by the numerical procedure. which wires enter brackets.
24 Raboud et al American Journal of Orthodontics and Dentofacial Orthopedics
January 2001
A B
C D
Fig 4. Force systems produced by rectangular loops.
A B
C D
A B
C D
A B
C D
Fig 9. Vertical activation of rectangular loops. A, Horizontal forces. B, Vertical forces. C, Labial-lingual
forces. D, In-plane moment (M) on posterior tooth.
and the anterior and posterior moments should be at a times those that occur for the rectangular loops that are
minimum. This force system will also lead to minimal shown in Fig 4. However, the moment at the posterior
root movement. This ideal system is most closely tooth is lower than the moment for the rectangular loop
approximated by the rectangular loop in which the ver- at any given activation. Note that for a given activation,
tical segments are spaced as far apart as is possible. placing the vertical section of the L-loop closer to the
The large spacing means that the upper portion of the anterior end results in a larger load-deflection rate and
appliance (between the 2 vertical segments) is more also reduces the range of activation. The spacing between
flexible and therefore does not develop as large a the vertical segments for the L-loops was also found to
moment when activated vertically. have little effect on the response (Fig 6). These results
Although the L-loop design incorporates more wire suggest that, to create intrusive/extrusive forces alone, the
than the rectangular loop, its distribution does not result rectangular loop design is the more appropriate choice.
in a lower load-deflection rate and does not produce a Of the rectangular loops evaluated, the one with the max-
larger range of activations. Relatively large horizontal imum spacing between the vertical segments resulted in
forces accompany the desirable vertical forces (Figs 5, A, the least anterior moments and also developed relatively
and 6, A). In addition, anterior moments, which would low horizontal forces when activated.
not normally be desired, arise. The magnitudes of both The results for the rectangular loops cannot be
the horizontal forces and the anterior moments are 2 to 4 directly compared with the previous work of Vanderby
28 Raboud et al American Journal of Orthodontics and Dentofacial Orthopedics
January 2001
3-Dimensional effects
The introduction of OOP modifications to the pla-
nar rectangular loop does introduce a torquing
moment (Fig 8, B and C) to the anterior tooth.
E Although this moment is not exceptionally large, the
clinical effect will be mild labial crown torque while
the tooth is tipping lingually and rotating distal-
in/mesial-out. Conversely, there will be mild labial
root torque while the tooth is tipping labially and
rotating distal-in/mesial-out.
As shown in Fig 9, the introduction of the OOP
modifications has almost no effect on the planar force
system developed by the original planar rectangular
loop. This means that the OOP activation and the
forces and moments created by it are essentially inde-
pendent of the in-plane effects discussed earlier for the
purely planar designs. This has the consequence of
allowing the additional OOP preactivation to be
included without fear that the in-plane effects will be
altered. It should be noted that these results apply when
there is no first-order rotation of the tooth and that,
when rotation occurs, the moments and the labial/lin-
F gual forces would diminish.
The recent results of Menghi et al3 were determined
over ± 1 mm buccal/lingual activations for a 21-mm
IBD and, as a result, are not comparable with the cur-
rent study. However, Menghi et al3 also comment that
the rectangular loop appears to be the most appropriate
appliance to activate OOP because the forces and
moments associated with the OOP activations were
essentially independent of the force system that was
generated in the plane of the appliance.
CONCLUSIONS
The simulations evaluated in the current study
confirm the validity of the use of rectangular and L-
loops for vertical adjustments of adjacent teeth. The
designs, when used with TMA, can effectively pro-
vide the force systems that are necessary to create
G intrusive/extrusive forces and, with the OOP pretwist-
ing, also assist in rotational movement if necessary.
Fig 9. (cont’d) E, In-plane moment (M) on anterior tooth. The design of the rectangular loops allowed the 2
F, First order moment (M*). G, Torquing moment (MT). effects to be independently controlled because the in-
American Journal of Orthodontics and Dentofacial Orthopedics Raboud et al 29
Volume 119, Number 1
plane force system was not altered when the OOP pre- REFERENCES
activation was included. 1. Vanderby R, Burstone CJ, Solonche DJ, Ratches JA. Experi-
One disadvantage of these force systems is that the mentally determined force systems from vertically activated
intrusive/extrusive forces are nearly linearly related to orthodontic loops. Angle Orthod 1977;47:272-9.
the amount of vertical activation, which suggests that 2. Koenig HA, Vanderby R, Solonche DJ, Burstone CJ. Force sys-
tems from orthodontic appliances: an analytic and experimental
the force level will decrease as vertical movement
comparison. J Biomech Eng 1980;102:294-300.
occurs. This is a characteristic of the elastic nature of 3. Menghi C, Planert J, Mensen B. 3-D experimental identification
the appliance; however, different materials may pro- of force systems from orthodontic loops activated for first order
vide different effects. The use of shape memory alloy corrections. Angle Orthod 1999;69:49-57.
materials could produce force levels that are much 4. Raboud D, Faulkner MG, Lipsett AW. A segmental approach for
large three-dimensional rod deformations. International Journal
more uniform as the activation changes. Simulations
of Solids and Structures 1996;33:1137-56.
that use shape memory alloy wire in a rectangular loop 5. Raboud D, Faulkner MG, Lipsett AW, Haberstock DL. Three-
planar design7 showed that intrusive/extrusive forces dimensional effects in retraction appliance design. Am J Orthod
can be maintained at nearly constant levels over an acti- Dentofacial Orthop 1997;112:378-92.
vation range of approximately 3 mm although the total 6. Lipsett AW, Faulkner MG, El-Rayes K. Large deformation
analysis of orthodontic appliances. J Biomech Eng 1990;
range of activations was on the order of 6 mm, com-
112:29-37.
pared with the 4 mm for the TMA discussed earlier.7 7. Raboud D, Faulkner MG, Lipsett AW. Superelastic response of
The possible use of these materials for vertical align- NiTi sma wires for orthodontic applications. International Jour-
ment must be investigated further. nal of Smart Materials and Structures 2000;9:684-92.
O N THE MOVE?
Send us your new address at least six weeks ahead
Don’t miss a single issue of the journal! To ensure prompt service when you change your address, please
photocopy and complete the form below.
Please send your change of address notification at least six weeks before your move to ensure continued ser-
vice. We regret we cannot guarantee replacement of issues missed due to late notification.
JOURNAL TITLE:
Fill in the title of the journal here.
Name
Address
City/State/ZIP
INDIVIDUAL SUBSCRIBERS
COPY AND MAIL THIS FORM TO: OR FAX TO: OR PHONE:
Mosby 407-363-9661 800-654-2452
Subscription Customer Service Outside the U.S., call
6277 Sea Harbor Dr 407-345-4000
Orlando, FL 32887