UT511 Insulation Tester User Manual
UT511 Insulation Tester User Manual
OPERATING MANUAL
Model UT511: OPERATING MANUAL
TITLE PAGE
Introduction 5
Unpacking the Meter 5
Safety Information 5
International Electrical Symbols 7
Battery Saver (Sleep Mode) 8
Battery Indication 8
The Meter Structure 9
Display 10
Key Functions 12
Measurement Operation 15
A.Measuring Voltages 15
B.Measuring Insulation Resistance 16
a) Continuous Measurement 17
b) Timed Measurement 18
c) Polarization Index (PI) Measurement 18
d) Compare Function 19
C. Low Resistance Measurement 20
1
Model UT511: OPERATING MANUAL
TITLE PAGE
2
Model UT511: OPERATING MANUAL
1. Unpacking Inspection 5
2. International Electrical Symbols 7
3. Battery Indication 8
4. Meter Front Description 9
5. Meter Side Description 10
6. Display Description 11
7. Key Description 12
3
Model UT511: OPERATING MANUAL
4
Model UT511: OPERATING MANUAL
6
Model UT511: OPERATING MANUAL
7
Model UT511: OPERATING MANUAL
The Meter enters the Sleep Mode and blanks the display There is a battery indicator shows on the display upper
if there is no button press for 15 minutes. This is done left hand corner. Below Table 3 is the explanation:
to conserve battery power. The Meter comes out of Table 3. Battery Indication
Sleep Mode when ON/OFF button is pressed two times.
Battery Battery Voltage
The 15 minutes timer is disabled during any insulation Indicator
resistance measurement. The time period starts
immediately following any measurement. 8.5V or less. It means the battery is
empty, don’t use the Meter as it cannot
guarantee accuracy.
8.6V~9.0V. It means the battery is nearly
empty, replacing battery is necessary.
Accuracy will not be affected.
9.1V~10.2V
10.3V or more
8
Model UT511: OPERATING MANUAL
9
Model UT511: OPERATING MANUAL
1 Safety Shutter
2 Power adaptor Input Terminal
10
Model UT511: OPERATING MANUAL
11
Model UT511: OPERATING MANUAL
Key Functions
Table 7. Key Description
ON/OFF Turn on or off the Meter. Press and hold l Under insulation resistance measurement
the button for 1 second to turn the Meter mode: press to select one voltage
on. range up.
CLEAR/ Press to clear the stored data,Push 1 l Under load mode: press to recall
SEC to turn on and off the display the previous stored value.
backlight. l Under insulation resistance measurement
SAVE Press to store the current measurement mode: press to select one voltage
value. The maximum number of stored range down.
reading is 18. When the stored readings l Under load mode: press to recall
memory is full, the Meter shows FULL the next stored value.
and stop storing. Press CLEAR to clear l When set the timer duration for the
the stored value in order to store the measurement of insulation resistance
next measurement value. or polarization index, press to
LOAD l Press once to recall the first stored decrement the time. The maximum
value. length of time is 30 minutes, the
l Press again to exit Load feature. Meter will automatically carry out
l Load feature can only be used when measurement.
there is no high voltage output.
12
Model UT511: OPERATING MANUAL
13
Model UT511: OPERATING MANUAL
STEP l When the Meter is under compare Lo Press to initiate low resistance
mode: measurement
S1 means increment of 1, then DVC Pres to initiate voltages measurement
each press of increase 1 or /ACV
decrease 1.
E-STOP Emergency stop button. Press this
S2 means increment of 10, then
button when the Meter is hang and
each press of increase 10 or
cannot turn off the power.
decrease 10.
S3 means increment of 100, then
each press of increase 100 or
decrease 100.
COMP Set a pass / fail limit for insulation tests.
The default value is 100M
TIME Pres to step through continuous
measurement, timed measurement and
polarization index measurement in
sequence.
TEST Press to stop or start an insulation
resistance test
Ho Press to initiate insulation resistance
measurement
14
Model UT511: OPERATING MANUAL
Measurement Operation
Caution
Below section explains how to make measurements.
A. Measuring Voltages l Special care should be taken when measuring
high voltage.
Warning
15
Model UT511: OPERATING MANUAL
Black
16
Model UT511: OPERATING MANUAL
17
Model UT511: OPERATING MANUAL
18
Model UT511: OPERATING MANUAL
19
Model UT511: OPERATING MANUAL
20
Model UT511: OPERATING MANUAL
The Use of Power Adaptor 1. Open the side safey shutter, then you will see there
The use of power adaptor, see figure 7 is a power adaptor input terminal.
2. Make sure the Meter is power off and Insert the
UT511 power adaptor to the input terminal.
3. It is highly recommed to take out all the batteries
when you are using the power adaptor.
4. Make sure the Meter is power off when you
disconnect the UT511 power adaptor from the Meter.
(Input voltage 230VAC, Frequency 50/60Hz, Input
current 50mA, Output voltage DC 15V, MAX current
600mA)
Caution
If you want to choose power adaptor for power supply,
please use special power adaptor SA48-150060EU
which supported by our company, otherwise it will be
dangerous.
21
Model UT511: OPERATING MANUAL
22
Model UT511: OPERATING MANUAL
23
Model UT511: OPERATING MANUAL
Specifications
Certification
Compliances IEC 61010 CAT.II 1000V, CAT.III 600V overvoltage and double insulation standard
Physical Specifications
Display (LCD) Digital: 9999 counts Analog bar graph.
Operating Temperature -10 ~40 (14 ~104 )
Storage Temperature -20 ~60 (-4 ~140 )
Relative Humidity 85% @ -10 ~40 below;
90% @ -20 ~60 :
Battery Type 8pcs of 1.5V (R14 or LR14) batteries or DC15V power adaptor. DC15V
power adaptor is optionally at extra cost.
Dimensions (HxW xL) 202 x 155 x 94 mm
Weight Approx. 2kg (including battery)
24
Model UT511: OPERATING MANUAL
General Specifications
Range Auto
Overloading Display OL on insulation resistance range
Battery Indicator Display
Icon Display Equips with function and battery indicator icons.
Current Consumption Maximum: around 90mA
Average: around 20mA
Feature Summary
Display Backlight Bright backlight for clear readings in poorly lighted areas.
Autorange The Meter automatically selects best range
Warning and red light will on.
Voltage Auto release voltage
COMP Measurement Use the Compare function to set a pass/fail compare level for the insulation
measurements.
PI Measurement Polarization Index is the ratio of insulation resistance. You can pre-set two
point of times and automatically carry out the measurement.
25
Model UT511: OPERATING MANUAL
26
Model UT511: OPERATING MANUAL
Caution
At any output voltage, when the tested resistance is les than 5MΩ, the testing time cannot exceed 10 seconds.
C. Low Resistance Measurement
Function Resistance
Measurement Range 0.1Ω~999.9Ω
Resolution 0.1Ω
Accuracy (1%+3)
Maximum open circuit voltage Around 2.8V
Buzzer Open at less than 30Ω
Overload Protection 220V rms/10 seconds
27
Model UT511: OPERATING MANUAL
*END*
28
Model UT511: OPERATING MANUAL
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China
Postal Code: 523 925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
Fax: (852) 2950 9303
Email: info@uni-trend.com
http://www.uni-trend.com
29