Tomógrafo LightSpeed16
Tomógrafo LightSpeed16
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 2 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
Clinical Performance that focuses on image quality, coverage, • Built-in patient breathing lights and
and exam speed. digital counter provides a goal-oriented
approach to coach the patients in holding
Chest / Abdomen / Pelvis
their breath during an exam.
Main Features
Coverage 600 mm • SmartStart gantry-mounted start scan
• Full 360° rotation speed of 0.5 sec.
Rotation 0.5 sec. button and countdown display, facilitates
provides for improved temporal resolution
single-technologist operation by allowing
Mode 16 x 0.625 mm and image quality in areas such as cardiac
start of scan at the gantry, with a visual
Pitch 1.75:1 imaging;
reminder of time until X-ray initiation.
mA 420 • Faster scan times enable shorter breath
mAs 210 • SmartHelical GE proprietary, non-linear
holds and more comfortable exams for
Speed 35 mm/s interpolation algorithms, balance slice
patients;
profile, helical pitch, image noise and
Scan Time 17.1 seconds
• Faster scan times reduce the occurrence required technique.
of re-scans by lowering the possibility of
Peripheral Run-Off • SmartTrack advanced hardware and
patient motion during the exam;
software for X-ray beam tracking
• Routine scanning with a slice thickness of minimizes patient dose.
Coverage 1,400 mm
0.625mm - optimizes lesion detection and
Rotation 0.5 sec. • SmartTools software automates every
facilitates the use of thinner images for
Mode 16 x 1.25 mm exam task in order to increase throughput.
volume presentations.
Pitch 1.75:1 • Smar tBeam TM hardware and software
mA 380 • Reduced partial volume artifacts:
optimizes X-ray beam filtration
mAs 190 • Prospective or retrospective image independently for body and head
Speed 70 mm/s reconstruction of 0.625 mm, 1.25 mm, applications.
2.5 mm, 3.75 mm, 5 mm, 7.5 mm or 10
Scan Time 20 seconds • HiRes Chest software provides a
mm images from inherently thinner
single-slice, high-resolution scan mode that
acquisition data sets;
minimizes dose for this critical application.
• For both helical and axial acquisitions.
• SmartPrep , standard on LightSpeed16,
High Resolution Chest • Image decomposition to: provides software for real-time monitoring
• Retrospective reconstruction of thin of contrast enhancement.
Coverage 200 mm images from data sets where thicker
Rotation 0.5 sec. • * Direct3D , makes 3D routine and instant
images were initially reconstructed;
Mode 16 x 0.625 mm by building the 3D model during axial image
• Facilitates more detailed image analysis reconstruction, reducing the need for user
Pitch 1.75:1 without need for re-scans; interaction.
mA 440
• Improves 3D visualization;
mAs 220 • * SmartScorePro , provides EKG-gated
• For both helical and axial acquisitions hardware and software on the Advantage
Speed 35 mm/s
Scan Time 5.7 seconds • Up to 2,881 multi-slice 1-sec scan rotations Workstation for coronary artery calcium
stored on disk. Facilitates retrospective scoring.
The net result is that in many cases, helical image decomposition during off hours.
scans on the Lightspeed16 are up to 12 times • * CardiacSnapShot, provides
• Eight fundamental scan modes to software and hardware to perform
faster than 4-slice CT systems. With the simplify the inherently complex nature of
Lightspeed16, users can routinely use a 0.5 EKG-gated reconstructions of the heart
multi-slice helical scanning. in SnapShot Segment mode.
second scan speed and 0.5625:1, 0.9375:1,
1.375:1, and 1.75:1 helical pitches. This • Standard set of clinically- proven - SnapShot Segment is a single sector
added performance, at the same image protocols derived from leading luminary protocol using information from one heart
quality, may allow you to: reduce contrast sites around the world. Up to 8,460 cycle to generate an image with temporal
media, perform better thin-slice CT protocols can be edited, modified and resolution of 250ms.
angiography exams, use thinner slices for stored on the system. - SnapShot Burst is a multi-sector
most exams, and perform longer helical • Remote Gantry Tilt from the Operator’s protocol using up to two sectors from two
exams without tube cooling delays. Console to increase exam speed, including different heart cycles to produce an image
built-in safety features to prevent with temporal resolution of 125ms.
The LightSpeed16 delivers this level of accidental contact of the gantry with the - SnapShot Burst Plus is a multi-sector
performance through a balanced design patient. protocol using up to four sectors of data
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 3 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
from four different heart cycles to • Completely protocol-driven scan control configuration, scan mode, scan rotation
produce images with temporal resolution with a dramatic reduction in number of speed, table speed, or image thickness
to 65ms. screens; changes. It takes the guesswork out of
• CTDI w, DLP (Dose Length Product), and • Highly flexible editing tools that allow easy setting scan technique when changing
Dose Efficiency display during scan tailoring of the exam to the patient; parameters (note: user must select initial
prescription provides patient dose Noise Index as well as maximum mA
• Large, 1024 color display; setting). Noise index enable 100% Image
informatbon to the operator.
• Leading edge, real-time image Quality reproducibity from one patient to
processing (MPR, MPVR, Volume another and from one user to another.
Interactive CT Technology Viewer*). • Protocol Pro™ protocol manager -
The LightSpeed16 CT Scanner supplies provides operator control of automated
In summary, primary benefits of Interactive
features (like AutoFilm, AutoStore, and
exceptional computer and image processing CT on the LightSpeed16 Scanner:
power that significantly enhances clinical AutoTransfer) on per patient basis.
• A natural scan control user interface
productivity, building upon the strength of the • Patient demographics and exam protocols
LightSpeed family - true interactive CT • Dramatic reduction in the number of can be preprogrammed in advance of
technology. screens; only 2 screens to set up first patient arrival through the Schedule Patient
scan and 1 screen for real-time feature.
Interactive CT embodies a variety of design monitoring while scanning;
• A preprogrammed selection of AutoView
choices all striving to enhance operator and • Easier and more flexible protocols
and Image Review Layouts allow simple
department productivity. A truly interactive CT • Much more flexible and intuitive customization of the image presentation to
system will: graphic prescription process with a match the anatomical area of interest -
• Provide a user interface beyond “intuitive” 1024 Localizer; without the complexity of free-form
to become purely “natural” - from the • View/Edit Wizard™ intuitively adjusts “windows”.
screens to the console hardware itself; dependent parameters automatically in • ImageWorks™ provides instant access
• Allow two users to review cases side - response to operator-initiated changes to advanced image processing features
by-side , with minimal interference; and highlights them for quick review; such as MIP, MPR, MPVR, Volume Viewer*,
also alerts the operator to incompatible
• Supply a truly multi-tasking environment Advanced Vessel Analysis*, CT Perfusion
dependencies requiring operator 2* and DentaScan*.
where even advanced image processing
intervention;
can take place quickly and simultaneously
with other processes underway; • DynaPlan Plus™ full screen display • No waiting for network transfer or
illustrates scan status pictorially, with image retrieval from archive media.
• Operate with a very high degree of real-time feedback.
automation, yet allow patient-specific • Background filming allows use of the full
changes to be easily made, with virtually • Large color screen screen for AutoView and image
no restrictions; • Extensive use of picture icons and color review/processing without interruption
cues enhance ease of use; when auto or batch filming. Special “one
• Be as self-teaching as possible, and have touch” controls provide on-screen viewing
on-line access to tutorials as needed. • Large on-screen controls and attractive
of camera progress during AutoFilm
color palette provide comfortable
without disrupting other image processes
One key element of this design is to combine viewing over extended periods.
in progress.
some of the best features from several • Enhanced multi-tasking allows operators
product families into one state-of-the art CT • ProView™ visualization algorithms
to review more than one exam
system. For example, the LightSpeed16 CT simultaneously, independently - even with
available to enhance anatomical structures
Scanner combines: without additional image reconstruction
AutoView and AutoFilm on/
time.
• SmartTools productivity software to • BrightBox™ dedicated controls for
automate every step of the examination, image next, prior, manual paging and • Operator console convenient to locate in
critical for ensuring the highest productivity trackball W/L helps make two person suite.
and throughput possible with the image review practical; • Computer, image processor and image
LightSpeed16 CT Scanner; reconstruction hardware completely
• Up to four 512x512 images from four
• Large screen interface for controlling different exams can be viewed on a integrated in base of console - no
scan acquisition easily, with virtually large 1024 color display. separate computer cabinet to site;
everything at a single glance; • Split table top allows unrestricted patient
• SmartmA™ User Interface - automatically
viewing while still supporting 2 large
• Excellent simultaneity and multi-tasking optimizes mA to maintain constant image
color monitors;
performance; noise when collimation/detector
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 4 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
• Front and back work surfaces can be and physician flexibility to tailor technique Helical Multi-slice Modes:
set during installation within a range of to specific needs, optimizing patient dose,
vertical heights that help accommodate a and providing the power needed to perform
Simplified scan prescriptions and easy-to-
variety of siting requirements - a broad spectrum of axial and helical
use default protocols make the LightSpeed16
especially surrounding the height of the examinations.
fast and efficient in patient set up.
console relative to the window into the
scanning suite.
System Specifications Multi-slice acquisitions and short intergroup
• Work surface accommodates a large delays significantly reduce potential mis-
variety of working conditions and individual Scan Modes: registration between scans by increasing the
operator preferences. number of scans possible in a patient breath
The LightSpeed16 CT system can perform hold. Contrast agents may be better utilized
• Wide work surface has room for
virtually any clinical application due to its wide as well due to significantly faster scan
papers, patient charts, logbooks or other
variety of scan modes. With the LightSpeed16 acquisition.
peripheral devices;
, body CT studies are easier to perform and
• Wide area also helps make two person more productive than ever before. Helical protocols are almost identical to
operation more comfortable and “classical” axial scan protocols. At the
practical; beginning of a study, the operator selects the
Helical:
• All console components (monitors, type of exam with the anatomical
keyboard, mouse and BrightBox) are • Continuous 360° scanning with table programmer, and indicates the desired scan
free-standing and can be easily moved incrementation and no interscan delay. range - either manually, or from a Scout.
on the front work surface; • Scans can be acquired in a wide variety of
• All surfaces made of high-impact resin speeds. After completing the prescribed exam, the
that hides marks and are easy to clean. system remains ready to continue with
additional helical scans or a set of axial
• Direct network connection means a Axial:
scans.
multi-suite Ethernet card is not required for • Up to 16 contiguous axial slices acquired
a gateway out of the suite - saving costs simultaneously with each 360° rotation, The operator may reconstruct helical scans
and simplifying installation. with the time between scans set by the prospectively with up to 90% overlap, and
• Learning Solutions provides an on-line user-selected interscan delay (ISD) or retrospectively, at any arbitrary table location
Operator manual detailing system operation intergroup delay (IGD). in 0.1 mm increments.
via a multi-media CD-ROM player integrated • Scans may be easily clustered in groups to
into the front of the operator console. allow multiple scans in a single breath hold. Helical Multi-slice Modes
Learning Solutions can also be accessed
on a stand alone PC providing flexibility and • Minimum scan-to-scan cycle time of only
1.5 sec with table moves of ≤ 10 mm (any The complex nature of helical multi-slice
productivity for on demand learning of scanning has been simplified by grouping
system operation scan time).
all critical acquisition parameters within 8
basic scan modes, all optimized for image
Scout™:
Advanced Tube & Generator quality and speed. For eight-slice
Technology • Single radiographic plane for scan acquisition: 0.625: 1; 0.875: 1; 1.35: 1 and
localization and graphical prescription of 1.675: 1. For 16-slice acquisition: 0.5625:
prospective reconstruction; 1; 0.9375: 1; 1.375: 1 and 1.75: 1. These
• The Performix X-ray tube, with 6.3 MHU of • Extended range matches helical scannable
clinically derived multi-slice scan modes
storage and 53.2 kW operation, provides offer a wide range of selections that
range.
increased helical performance with greater carefully balance acquisition speed,
patient throughput and virtually no tube image thickness, artifact level and
cooling. Advanced technology in the Helical Scans retrospective image reconstruction
Performix tube includes a metal-ceramic flexibility.
frame for long life, a high-speed bearing for Slip ring technology has advanced axial
scanning by enabling scans with zero This simplified user interface guides the
sub-second scanning, a high-efficiency
interscan delay and simultaneous table user in the choice of scan parameters.
motor to accelerate the large anode and
movement. The user selects a pitch mode, a desired
efficient cooling for high throughput and
image slice thickness and table travel per
superior helical performance.
rotation. The user interface also displays
• Wide range of technique factors (10 to 440 the resulting choice of retrospective
mA, in 5 mA increments) gives the operator
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 5 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
The 16 slice helical acquisition modes 8-SLICE HELICAL MODES Scan Speeds: Full 360° rotational scans
provide table speeds from 5.625 in 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 seconds.
mm/rotation up to 35 mm per rotation, Table Speed (mm/rotation)
Slice
Scan Technique:
enabling scan speeds that are up to 12 Pitch
Thick-
times faster than 4-slice helical scanners. 0.625:1 0.875:1 1.35:1 1.675 — kVp: 80, 100, 120, 140
ness
(mm) :1
— mA: 10 to 440 in 5 mA increments
Prospective Multiple-
Thickness Reconstruction: 1.25 6.25 8.75 13.5 16.75 — Power: 0.8 to 53.2 kW
In addition to the initial reconstructed slice 5 6.25 8.75 13.5 16.75 Single Acquisition: 120 second scan
thickness, the operator has the option to 12.5 17.5 27.0 33.5 maximum.
prospectively specify additional images to be
7.5 6.25 8.75 13.5 16.75 Helical Tilt: helical acquisition is possible
reconstructed from a single raw data set.
These images can be reconstructed at any of 12.5 17.5 27.0 33.5 with the gantry tilted to a maximum of 30
the defined nominal slice thicknesses 10 6.25 8.75 13.5 16.75 degree, in half degree increments.
available for a given table speed and scan
12.5 17.5 27.0 33.5 Multiple Acquisition Maximum Scan Time:
mode.
Multiple scans may be acquired in one
This effectively facilitates later, more detailed series to produce up to 1,500 contiguous
image analysis without additional patient helical images. Up to 2,881 seconds
scans and subsequent dose and image helical coverage is possible in multiple
registration concerns. 16-SLICE HELICAL MODES series.
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 6 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 7 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 8 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
CTDI: • Nominal Thickness: 0.625, 2.5, 5, 7.5, 10 • Minimum scan-to-scan cycle of 1.5
mm seconds possible for 0.5 sec. scan speed
On CTDI Head and Body Dose Reference with minimum ISD's.
Phantoms: 4i Mode:
• Produces 4 images per rotation Scan Fields of View:
CTDIvol expressed in mGy/100 mAs for
IEC pitch 1: (normalized to a pitch of 1). • Nominal Thickness: 2.5, 3.75, 5 mm • 25 cm for adult head
• 25, 50 cm for body
Head 20.9 mGy/100 mAs 8i Mode:
• 25 cm for pediatric head
Body 10.6 mGy/100 mAs • Produces 8 images per rotation
• Nominal Thickness: 1.25, 2.5 mm Scan with no table incrementation,
Axial Scans contiguous image location, or skipped
16i Mode: image location are possible. Overlapped
Multi-slice acquisitions and short interscan axial scans are not possible.
• Produces 16 images per rotation
delays significantly reduce potential missed
registration between scans by increasing the • Nominal Thickness: 0.625, 1.25 mm
Axial Image Reconstruction
number of scans possible in a patient breath
hold. Contrast agents may be better utilized Axial Scan Parameters Reconstruction Algorithms: Soft Tissue,
as well due to significantly faster scans.
Standard, Detail, Bone, Bone Plus, Lung and
Scan Time: Edge.
Axial Multi-slice Prescription • 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 2.0, 3.0 and 4.0
Reconstruction Matrix: 512
second full scans (360° acquisition)
Simplified scan prescriptions and easy-to- Display Matrix: 1024.
use default protocols make the Lightspeed16 Scan Technique:
CT Scanner fast and efficient in patient set- • kVp: 80, 100, 120, 140 Display FOV: Freely variable center/off-
up. Axial protocols are nearly identical to center, prospective/retrospective target
• mA: 10 to 440, in 5 mA increments
helical scan protocols. selection.
• Power: 0.8 to 53.2 kW
• Focal Spot Selection: CT Number Scale: -1024 to 3071 HU
Axial Multi-slice Modes
• Small spot for up to 24 kW Axial Image Reconstruction:
The LightSpeed16 acquires axial scans in • Large spot for greater than 24 kW
• Reconstruction time as fast as 6 images
sets of 8 or 16 contiguous images in one
Scan Plane Geometry: per second.
360° rotation.
• +/- 30° gantry tilt, in 0.5° increments • Typical 0.17 sec. image-to-image recon in
For each rotation of the gantry, the 16 slice recon mode.
LightSpeed16 collects 16 rows of scan • Longitudinal positioning in 0.01 mm per slice
data. There are five reconstruction increment. Gantry display in 0.5 mm • Maximum image-to-image cycle time is ±
modes available for creating images from increments. 10% for prospective and retrospective
the multi-slice scan data (1i, 2i, 4i, 8i, and image-to-image display. This applies for
Interscan Delay (ISD): 512 matrix; any display FOV; in
16i). By using 1i, 2i, 4i, and 8i
reconstruction modes, scan data can be • Minimum ISD with table movements of AutoView (all layouts); with concurrent
combined prior to image reconstruction to 0 - 10 mm: 1.0 sec. filming and image archival for all scan
modes.
create slices with reduced partial-volume • Minimum ISD with table movements of more
artifacts. This is particularly useful for than 10 mm and up to 20 mm: 1.3 sec • Iterative bone processing increases time by
posterior-fossa imaging. 250 milliseconds.
• User-selectable.
Prospective Multiple Reconstruction
Inter Group Delay (IGD):
1i Mode: (PMR): Up to 3 sets of reconstructions
• Minimum IGD is the same as minimum ISD; can be pre-programmed as part of the
• Produces 1 image per rotation also user-selectable. scan protocol prior to acquisition. The
• Nominal Thickness: 1.25, 5, 10 mm operator can select different
Scan-to-Scan Cycle:
reconstruction algorithms and display
2i Mode: fields of view for each reconstruction.
• Produces 2 images per rotation This frees the operator from sitting at the
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 9 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 10 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
Body 5.4 mGy/100 mAs center • Image monitor for image display, ImageWorks:
11.2 mGy/100 mAs surface analysis, processing, and management
• Each monitor provides a 1280 x 1024 ImageWorks is a desktop environment
CTDIvol expressed in mGy/100 mAs: designed to take advantage of the
high resolution, flicker-free display
Lightspeed16 CT Scanner System computer
Head 18.4 mGy/100 mAs • Optional 18” LCD Monitors.
and image processor.
Body 9.3 mGy/100 mAs • Scan control keyboard assembly with
intercom speaker, microphone and volume Standard features include archive, network
Scout Scans controls and manual film control, as well as some
• Three button mouse with mouse pad advanced image processing such as multi-
ScoutView™ scans provide excellent planar reformatting (MPR), multi-projection
• BrightBox (trackball assembly) volume rendering (MPVR), and MR image
detail for anatomical localization in
conjunction with scan prescription. • Two wide work surfaces display. It also has optional add-on packages
for Volume Viewer*, CT Perfusion2* and
Scan locations may be prescribed at the All these devices are free-standing and can DentaScan*.
operator console either graphically (via be easily moved to accommodate a large
mouse), or explicitly (keyboard entry) variety of working conditions and individual The ImageWorks desktop also provides a
from a Scout scan. operator preferences. gateway for DICOM 3.0 image transactions,
either through a local area network, or via
Prescription of scans with multiple gantry Split table top allows unrestricted patient DICOM-formatted MOD media.
angles are also available on a single viewing while still supporting 2 monitors.
Scout. Each work surface can be adjusted at Learning Solutions:
installation to help accommodate a variety of
siting requirements. The LightSpeed16 provides an on-screen,
Scout Scan Parameters on-line operator manual via a multi-media CD-
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 11 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
Patient information can be easily recalled to display the 1024 Localizer each time a new Scan Data Acquisition:
set up an immediate exam via List/Select series is requested.
Scheduled Patient on the scan/recon monitor. • Default protocols also include preset scan • Full-screen DynaPlan Plus illustrates scan
Pre-programmed patient exams can also be time, kVp, mA, slice thickness, scan mode, status graphically, with real-time feedback
recalled from the New Patient screen table speed, image interval, gantry tilt, scan while the exam is underway. Scans,
automatically by entering the patient ID field-of-view, display field-of-view and programmed delays (prep, breathing, inter-
number. center, recon types, and breath timing group), and even AutoVoice
parameters. announcements are clearly shown before
and during scanning.
Patient Data Entry: • Any scan parameter can be edited for
each scan or all scans either before or • AutoScan: Fully automates longitudinal
Patient data can be entered as part of New during an exam. Scans can be easily table movement and the start of each scan
Patient set-up, or can be recalled from the list added or removed from the prescription. • AutoVoice: Preset (English) and user-
of pre-scheduled patients. recorded messages automatically deliver
• Scan/recon control uses only 2 screens to
set up first scan - New Patient and Protocol patient breathing instructions, especially
Trauma Patient ID allows patient scans and useful for multiple or multi-pass helical
image display/analysis without entering View/Edit.
scans
patient data before scanning.
• Full Simultaneity allows scan and recon to
Protocol View/Edit: work concurrently with image display,
Exam Protocol Selection: • A single, full screen View/Edit table allows processing and analysis (including
fast and easy examination and modification computationally intensive features such as
One of the main contributions of the of exam parameters before scanning MPR, MPVR and 3D*/MIP) while still running
LightSpeed16 CT Scanner System to begins image archival, filming and networking
department productivity is its simplified exam processes.
• Exam parameters can be changed for just
set-up.
one scan, or for all scans in a series
• Exam parameter set-up has been greatly Dose Computation & Display
• When used in conjunction with the 1024
simplified through the exclusive use of
Localizer, changes made in the View/Edit
protocols CTDIvol (CTDI volume), DLP (Dose Length
table that affect the number of scans,
• Protocols can be easily selected in one of image interval, starting/ending locations, tilt,
Product), and Dose Efficiency computation
three convenient ways: and display during scan prescription provides
or display FOV are automatically shown on
patient dose information to the operator.
• A large, graphical Anatomical the 1024 Localizer
Programmer located on the New Patient • Any changes made directly on the 1024 CTDI vol is a dose index defined by IEC 601-2-
screen Localizer display using the mouse and the 44. This index is computed automatically by
• A default list of the “top 10” most on-screen cursor controls are also the Lightspeed16 CT System and reported on
commonly used protocols located near reflected automatically in the View/Edit the Exam Rx screen. CTDIvol is a single
the anatomical programmer table number consisting of 2/3 of the CTDI100
• A numerical entry peripheral dose plus 1/3 of the CTDI100 central
• View/Edit Wizard intuitively adjusts
dose that is divided by the helical or axial
• Two Anatomical Programmers - one for dependent parameters automatically in
pitch factor.
adults and one for pediatrics - provide response to operator-initiated changes and
quick and easy access to 8,460 user- highlights them for quick review. It also CTDI100 is a dose index based upon CTDI
programmable protocols (total). Each alerts the operator to incompatible dose measurements over a 100 mm volume,
programmer has ten anatomical regions dependencies requiring operator as defined in IEC 601-2-44.
with 90 protocols for each region intervention.
Dose Length Product (DLP) is given in
• Default protocols have been expanded • Tab card groupings for Timing, Recon and mGy*cm and is computed and displayed for
through Protocol Pro - a “behind the Filming help organize the large number of each group prior to the scan. Additionally, an
scenes” protocol manager - that allows parameters available within each protocol. accumulated DLP is displayed for the entire
preselection of automated features like • As many as 8,460 protocols can be stored exam, as the exam prescription progresses.
AutoVoice, AutoFilm, AutoStore and on the Operator Console. The final exam accumulated DLP provides a
AutoTransfer on a per-exam basis. convenient measure for maintaining patient or
• Protocol Pro also provides preselection of procedure dose management statistics.
two different window/level settings per
Dose Efficiency is automatically
image for AutoFilm and can automatically
computed and displayed on the Exam Rx
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 12 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 13 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
Auto Image Management: Because some of the image analysis and • Pseudo Surface Shading Mode
display features of ImageWorks replicate • Predefined Cut Planes
those in Exam Rx, the next section describes
The Exam Rx work environment conveniently
only features that are incremental or • Volume Rendered Navigator views
provides for selection of AutoFilm, AutoStore
significantly different. • VR Preset save/recall
(to local or remote MOD), and AutoTransfer
(across a network). • 3D Rendered Lumen View
Image Analysis • Automatic Path Tracking
An AutoFilm Composer provides a simple
programming interface for automated filming • Path Bridging (in case of occlusions)
• Multi-Projection Volume
set-up. Reconstruction (MPVR): Quick and • SmartCursor™ for Easy Navigation
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 14 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
• Window/Level values may be: • Batch film can typically enable you to Industry Standards
• Preset to provide six on-screen instant reduce filming images by 50%, thickening
the slab from 1.25 to 2.5mm with no • The LightSpeed16 CT Scanner System
window/level settings
information loss. complies with a wide variety of industry
• Set independently for up to 16 images standards to facilitate more rapid adoption
on the screen of features and performance improvements
• User-modified in discrete or variable Image Networking: as the computing and medical imaging
steps industry evolves.
• Adjusted real-time on-image by holding Exams can be selected and moved from
down the middle mouse button and MR Signa Advantage, CT HiSpeed
Advantage, Advantage Independent DICOM Conformance
moving the mouse
Console, Advantage Windows Standards:
• Cine mode provides paging in up to 4 view
Workstation to the Lightspeed16 CT
ports of up to 128 previously-stored CT or
Scanner’s Operator Console. This is • DICOM 3.0 Storage Service Class
MR images at full selected display frame • Service Class User (SCU) for image
performed via the AdvantageNet
rate. For more than 128 images, display send
(GenesisNet) point-to-point protocol. This
frame rate may be reduced.
protocol supports send, receive and • Service Class Provider (SCP) for image
• Cine mode also provides temporal, spatial pull/query (no broadcast). NOTE: receive
or manual playback loops. Because the Lightspeed16 creates
• DICOM 3.0 Query/Retrieve Service Class
• Text Page images in pure DICOM format, exams can
not be moved from the Lightspeed16 CT • DICOM 3.0 MOD Media Service Class on
Scanner System to MR Signa Advantage, 1.2- and 2.3-GB MOD media
Image Annotation CT HiSpeed Advantage or the Advantage • DICOM 3.0 Storage Commitment Class Push
Independent Console systems.
• Image annotation and cursor are • DICOM 3.0 Modality Worklist*
shadowed to permit ease in reading. Exams can also be selected and moved • DICOM 3.0 Modality Performed Procedure
between the Lightspeed16 CT Scanner Step*
Image Management System and any imaging system
• DICOM 3.0 Print
supporting the DICOM 3.0 protocol for
• Images may be stored and retrieved via network send, receive and pull/query. GE
Magnetic Optical Drive (MOD) media using systems that support this interoperability Filming Protocol:
DICOM 3.0 format. This allows interchange include CT HiSpeed CT/i, X/i, NX/i, QX/i,
with other imaging systems supporting • 3M-952 Standard
MR Signa LX, and Advantage Windows
DICOM 3.0 MOD media. Not all vendors 1.2/2.x/3.x/4.x.
implementation of DICOM 3.0 are identical, System Components
so please check with the manufacturer for Image transfer time using either
compatibility. AdvantageNet or DICOM 3.0 protocols is
approximately 0.1 second per 512 image Gantry:
• Off-line retrieval of all image files. Images
on a 100baseT Network.
may be viewed as soon as they are
Advanced slip ring design continuously
restored from MOD.
rotates generator, tube, detector and data
• Image retrieval time is approximately 30 Computer Based Training acquisition system around the patient.
images 512 per second. This innovative feature provides on-
• Aperture: 70 cm
screen, on-line support of Exam Rx and
ImageWorks functions via a multi-media • Tilt: ± 30°
VariViewerTM: CDROM player, integrated into a tower on • Tilt Speed: 1°/sec.
• VariViewer is an interactive axial review operator console for easy access. Use
• Focus to Detector: 95 cm
mode that can change the slice thickness of rich graphics and text to provide truly
reconstruction instantaneously. useful system help. • Focus to Isocenter: 54 cm
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 15 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
• Integrated breathing lights and countdown • 205 kg (450 lb) maximum allowed with • mA: 10 to 440 mA, 5 mA increments
timer. normal operation and +/- 1 mm positional
accuracy Maximum mA for each kVp selection:
• Integrated start scan button with
countdown to X-ray on. • Controls on gantry for elevation and cradle
incrementation. Foot pedals on both sides
Scan plane toward front of gantry for of table for fast elevation. Cradle position KVp Max mA
improved positioning access. controlled from OC for prescribed scans. 80 400
Biopsy and interventional studies have been • IV Pole integrated at the foot-end of the 100 420
facilitated through a more streamlined gantry table prevents IV lines from becoming 120 440
shroud, and bilateral table/gantry controls crossed and tangled, and ensures that the 140 380
and gantry display that maximize lines stay securely in place on the patient.
maneuverability while working next to the HiLight Matrix II Detector:
gantry. X-Ray Tube
Laser Alignment Lights: 21,888 individual elements composed by: 8
Performix ADV Metal-Ceramic Tube Unit. rows of 1.25 mm thickness and 16 rows of
• Define both internal and external scan Design optimized for exams requiring a 0.625 mm thickness, each containing 888
planes to ± 1 mm accuracy. large number of scans without tube active patient elements; 24 reference
• Operate over full range of gantry tilt; cooling. elements.
activated any time during exam (with tube • Heat Storage Capacity: 6.3 MHU 4 typical modes of data output:
stationary). • 8 x 1.25 mm (uses center 16 rows)
• Heat Dissipation:
• Coronal light remains perpendicular to axial • 8 x 2.5 mm (uses all 24 rows)
• Anode (max) 840 KHU/min
light as gantry tilts.
• Casing (cont) 300 KHU/min • 16 x 1.25 mm (uses all 24 rows)
Visual readout is easy to read from the table • Tube Unit: 6.9 kW continuous for 10 • 16 x 0.625 mm (uses center 16 rows )
side or from the operator console. minutes
70% geometric efficiency; 98% absorption
Gantry tilt controls are located on the side of • Dual Focal Spots: efficiency.
the gantry. • Small Focal Spot:
0.7 mm (W) x 0.6 mm (L) nominal
value
Data Acquisition System
Table (IEC 336/93)
12,288 available input channels.
• Single table, cantilever design with wide 0.9 mm (W) x 0.7 mm (L) traditional
height range methodology 1,640 Hz maximum sample rate.
• Vertical Range: 51.6 cm to 99.1 cm • Large Focal Spot: Effective analog to digital conversion range
• Vertical Scannable Range: 80.6 cm to 99.1 0.9 mm (W) x 0.9 mm (L) nominal greater than two million to one.
cm value
(IEC 336/93)
• Elevation Speeds: 5 mm/sec and 40 Scan/Control Unit
mm/sec 1.2 mm (W) x 1.2 mm (L)
traditional methodology
• Horizontal Range: 170 cm Located in base of Operator
• Horizontal Scannable Range: 170 cm • Maximum Power: 53.2 kW Console.
metal-free (axial) and 160 cm metal-free • Beam collimated to 55° fan angle.
(helical & Scout)
Average time to replace tube: < 10 hours
• Horizontal Speed: Up to 100 mm/sec Host Computer
• Table automatically recenters on scan
High Voltage Generation • Dual SMP 2.66.GHz Intel Xeon processors
plane with changes in vertical position
with 512KB L2 cache.
(after setting internal landmark with • High-frequency on-board generator.
alignment lights on) • Intel Hyper-threading technology.
Continuous operation during scan.
• Table Load Capacity: • 2GB DDR266 Dual Channel Memory with a
• 53.2 kW output power.
• 180 kg (400 lb) with +/- 0.25 mm throughput of 4.2GB/sec
• kVp: 80, 100, 120, 140
positional accuracy
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 16 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 17 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
The only facility input to the system is a 380 Table 200 700
to 480 V nominal, 3 phase Delta or Wye, PDU 1,000 3,400
50/60 Hz, 90 kVA service, 20 kVA average
Operator 2,165 7,400 Warranty
power; main disconnect to be located within
Console
5 feet (1.5 m) of the PDU. The facility must
also provide a protective disconnect device Optional Laser 800 2,730 The published Company warranty in effect
with low voltage, low energy local and multi- Camera on the date of shipment shall apply. The
point remote capability, in the line feeders to Company reserves the right to make
Recommended cooling values should not be changes.
the PDU.
used for calculating system input power
Complete, detailed specifications of all power requirements. All specifications are subject to change.
requirements are available upon request. For
** Recommended Allowance: 7,150/24,400
most installations, the Lightspeed16 CT Regulatory Compliance
Scanner System does not require any power Growth Allowance: 9,200/31,400
conditioning equipment to be used in This product is designed to comply with
conjunction with the PDU. Regulators are not applicable standards under the Radiation
recommended for use with this system. For Control for Health and Safety Act of 1968.
those sites with known large power line
transients, a suppresser filter for the system Laser alignment devices contained within this
computer and peripherals may be useful. In product are appropriately labeled according
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 18 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 19 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 20 With Xtream Hong Kong - Fax: 852 2559 3588
CT Scanner System
* Option