GE Medical Systems-America

Milwaukee, USA - Fax: 1 414 544


3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 1 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

• CrossBeam T M reconstruction algorithm


solves the technical challenge of cone
beam artifacts.
• Large breadth of Advanced
Applications: Advanced Lung Analysis,
CT Colonography, Advanced Vessel
Analysis, Brain and Body Perfusion,
Cardiac Analysis, Cardiac Function,
SmartScore Pro and Advantage Sim.
• Compact system design of the
LightSpeed16 allows for installation in only
28m² (302 square feet).
• Productivity features designed for the
CT Technologist: In-Room Start, Remote
Gantry Tilt, Breathing Lights with
countdown timer, Gantry Controls mounted
on all four corners of the gantry and an
Integrated IV Pole at the foot of the table.
• 1700 mm scannable range for full body
scans.
• 254 GB Disk (system, image, scan disks)
standard with a one-year, non pro-rated stores up to 250,000 5122 images and 2881
Introduction
warranty. scan data files.
The LightSpeed16 CT Scanner with • microVoxelTM imaging improves 3D and • Cardiac acquisition with Segment,
Xtream TM technology is the next step in reformatted 2D resolution through the Burst and Burst Plus enable you to
GE’s CT Continuum and its award- optimum choice of sub-millimeter slice scan patients with heart rates from 40 to
winning LightSpeed CT platform. The thickness and reconstructed voxel size. 110 BPM, and provide for temporal
LightSpeed16 CT Scanner routinely resolution up to 65 ms.
• Unmatched Volume Image Quality
acquires 16 sub-millimeter slices in one
specification in all planes (sagittal, coronal • Color Coding for Kids, winner of a
rotation, at a scan speed of 0.5 seconds.
and axial): 15 lp/cm @ 0% MTF @ 0.5 sec. National Heroes Award from the
As with the very first LightSpeed product
Rotation Emergency Medical Services for Children,
introduction, the LightSpeed16 utilizes
provides pediatric scan protocols based on
Designed for Six Sigma (DFSST M) • HiLightTM Matrix II Detector is designed
the Broselow -LutenT M Pediatric System.
methodology in all aspects of to deliver consistent image quality with its
This Color Coding system is incorporated
manufacturing and engineering. 21,888 individual elements: 16 rows of
into the protocol selection on the operator’s
0.625mm thickness and 8 rows of 1.25mm
console and is designed to facilitate
thickness
pediatric emergency care and reduce
• Xtream TM technology, a revolutionary medical errors.
Primary Benefits new workflow design, breaks through
• VariViewer TM: provides an interactive
existing limits on speed, image quality and
• Routine use of sub-millimeter slices axial review mode that can change the
flexibility to provide an optimized workflow
without image noise or coverage slice thickness reconstruction
solution from acquisition to final report.
compromise. instantaneously, resulting in the ability to
• Lowest dose acquisition using “scan thin and view thick.”
• Small lesion and small vessels
SmartmA, TM an automatic modulation
assessments – pancreas, liver or
technique and DLP display that tells the
circle of willis, renal arteries, coronary
operator the dose before the scan starts.
arteries and peripheral vascular
arteries. • HyperPlane TM reconstruction algorithm
tackles the inherent issues with fast, high
• The 6.3 MHU Performix X-Ray Tube
pitch helical scanning by reducing helical
demonstrates the highest reliability and
artifacts and dose, and optimizing the slice
delivers the power you need for image
profile.
quality. The Performix Tube now comes

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 2 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Clinical Performance that focuses on image quality, coverage, • Built-in patient breathing lights and
and exam speed. digital counter provides a goal-oriented
approach to coach the patients in holding
Chest / Abdomen / Pelvis
their breath during an exam.
Main Features
Coverage 600 mm • SmartStart gantry-mounted start scan
• Full 360° rotation speed of 0.5 sec.
Rotation 0.5 sec. button and countdown display, facilitates
provides for improved temporal resolution
single-technologist operation by allowing
Mode 16 x 0.625 mm and image quality in areas such as cardiac
start of scan at the gantry, with a visual
Pitch 1.75:1 imaging;
reminder of time until X-ray initiation.
mA 420 • Faster scan times enable shorter breath
mAs 210 • SmartHelical GE proprietary, non-linear
holds and more comfortable exams for
Speed 35 mm/s interpolation algorithms, balance slice
patients;
profile, helical pitch, image noise and
Scan Time 17.1 seconds
• Faster scan times reduce the occurrence required technique.
of re-scans by lowering the possibility of
Peripheral Run-Off • SmartTrack  advanced hardware and
patient motion during the exam;
software for X-ray beam tracking
• Routine scanning with a slice thickness of minimizes patient dose.
Coverage 1,400 mm
0.625mm - optimizes lesion detection and
Rotation 0.5 sec. • SmartTools  software automates every
facilitates the use of thinner images for
Mode 16 x 1.25 mm exam task in order to increase throughput.
volume presentations.
Pitch 1.75:1 • Smar tBeam TM hardware and software
mA 380 • Reduced partial volume artifacts:
optimizes X-ray beam filtration
mAs 190 • Prospective or retrospective image independently for body and head
Speed 70 mm/s reconstruction of 0.625 mm, 1.25 mm, applications.
2.5 mm, 3.75 mm, 5 mm, 7.5 mm or 10
Scan Time 20 seconds • HiRes Chest software provides a
mm images from inherently thinner
single-slice, high-resolution scan mode that
acquisition data sets;
minimizes dose for this critical application.
• For both helical and axial acquisitions.
• SmartPrep , standard on LightSpeed16,
High Resolution Chest • Image decomposition to: provides software for real-time monitoring
• Retrospective reconstruction of thin of contrast enhancement.
Coverage 200 mm images from data sets where thicker
Rotation 0.5 sec. • * Direct3D , makes 3D routine and instant
images were initially reconstructed;
Mode 16 x 0.625 mm by building the 3D model during axial image
• Facilitates more detailed image analysis reconstruction, reducing the need for user
Pitch 1.75:1 without need for re-scans; interaction.
mA 440
• Improves 3D visualization;
mAs 220 • * SmartScorePro , provides EKG-gated
• For both helical and axial acquisitions hardware and software on the Advantage
Speed 35 mm/s
Scan Time 5.7 seconds • Up to 2,881 multi-slice 1-sec scan rotations Workstation for coronary artery calcium
stored on disk. Facilitates retrospective scoring.
The net result is that in many cases, helical image decomposition during off hours.
scans on the Lightspeed16 are up to 12 times • * CardiacSnapShot, provides
• Eight fundamental scan modes to software and hardware to perform
faster than 4-slice CT systems. With the simplify the inherently complex nature of
Lightspeed16, users can routinely use a 0.5 EKG-gated reconstructions of the heart
multi-slice helical scanning. in SnapShot Segment mode.
second scan speed and 0.5625:1, 0.9375:1,
1.375:1, and 1.75:1 helical pitches. This • Standard set of clinically- proven - SnapShot Segment is a single sector
added performance, at the same image protocols derived from leading luminary protocol using information from one heart
quality, may allow you to: reduce contrast sites around the world. Up to 8,460 cycle to generate an image with temporal
media, perform better thin-slice CT protocols can be edited, modified and resolution of 250ms.
angiography exams, use thinner slices for stored on the system. - SnapShot Burst is a multi-sector
most exams, and perform longer helical • Remote Gantry Tilt from the Operator’s protocol using up to two sectors from two
exams without tube cooling delays. Console to increase exam speed, including different heart cycles to produce an image
built-in safety features to prevent with temporal resolution of 125ms.
The LightSpeed16 delivers this level of accidental contact of the gantry with the - SnapShot Burst Plus is a multi-sector
performance through a balanced design patient. protocol using up to four sectors of data

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 3 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

from four different heart cycles to • Completely protocol-driven scan control configuration, scan mode, scan rotation
produce images with temporal resolution with a dramatic reduction in number of speed, table speed, or image thickness
to 65ms. screens; changes. It takes the guesswork out of
• CTDI w, DLP (Dose Length Product), and • Highly flexible editing tools that allow easy setting scan technique when changing
Dose Efficiency display during scan tailoring of the exam to the patient; parameters (note: user must select initial
prescription provides patient dose Noise Index as well as maximum mA
• Large, 1024 color display; setting). Noise index enable 100% Image
informatbon to the operator.
• Leading edge, real-time image Quality reproducibity from one patient to
processing (MPR, MPVR, Volume another and from one user to another.
Interactive CT Technology Viewer*). • Protocol Pro™ protocol manager -
The LightSpeed16 CT Scanner supplies provides operator control of automated
In summary, primary benefits of Interactive
features (like AutoFilm, AutoStore, and
exceptional computer and image processing CT on the LightSpeed16 Scanner:
power that significantly enhances clinical AutoTransfer) on per patient basis.
• A natural scan control user interface
productivity, building upon the strength of the • Patient demographics and exam protocols
LightSpeed family - true interactive CT • Dramatic reduction in the number of can be preprogrammed in advance of
technology. screens; only 2 screens to set up first patient arrival through the Schedule Patient
scan and 1 screen for real-time feature.
Interactive CT embodies a variety of design monitoring while scanning;
• A preprogrammed selection of AutoView
choices all striving to enhance operator and • Easier and more flexible protocols
and Image Review Layouts allow simple
department productivity. A truly interactive CT • Much more flexible and intuitive customization of the image presentation to
system will: graphic prescription process with a match the anatomical area of interest -
• Provide a user interface beyond “intuitive” 1024 Localizer; without the complexity of free-form
to become purely “natural” - from the • View/Edit Wizard™ intuitively adjusts “windows”.
screens to the console hardware itself; dependent parameters automatically in • ImageWorks™ provides instant access
• Allow two users to review cases side - response to operator-initiated changes to advanced image processing features
by-side , with minimal interference; and highlights them for quick review; such as MIP, MPR, MPVR, Volume Viewer*,
also alerts the operator to incompatible
• Supply a truly multi-tasking environment Advanced Vessel Analysis*, CT Perfusion
dependencies requiring operator 2* and DentaScan*.
where even advanced image processing
intervention;
can take place quickly and simultaneously
with other processes underway; • DynaPlan Plus™ full screen display • No waiting for network transfer or
illustrates scan status pictorially, with image retrieval from archive media.
• Operate with a very high degree of real-time feedback.
automation, yet allow patient-specific • Background filming allows use of the full
changes to be easily made, with virtually • Large color screen screen for AutoView and image
no restrictions; • Extensive use of picture icons and color review/processing without interruption
cues enhance ease of use; when auto or batch filming. Special “one
• Be as self-teaching as possible, and have touch” controls provide on-screen viewing
on-line access to tutorials as needed. • Large on-screen controls and attractive
of camera progress during AutoFilm
color palette provide comfortable
without disrupting other image processes
One key element of this design is to combine viewing over extended periods.
in progress.
some of the best features from several • Enhanced multi-tasking allows operators
product families into one state-of-the art CT • ProView™ visualization algorithms
to review more than one exam
system. For example, the LightSpeed16 CT simultaneously, independently - even with
available to enhance anatomical structures
Scanner combines: without additional image reconstruction
AutoView and AutoFilm on/
time.
• SmartTools productivity software to • BrightBox™ dedicated controls for
automate every step of the examination, image next, prior, manual paging and • Operator console convenient to locate in
critical for ensuring the highest productivity trackball W/L helps make two person suite.
and throughput possible with the image review practical; • Computer, image processor and image
LightSpeed16 CT Scanner; reconstruction hardware completely
• Up to four 512x512 images from four
• Large screen interface for controlling different exams can be viewed on a integrated in base of console - no
scan acquisition easily, with virtually large 1024 color display. separate computer cabinet to site;
everything at a single glance; • Split table top allows unrestricted patient
• SmartmA™ User Interface - automatically
viewing while still supporting 2 large
• Excellent simultaneity and multi-tasking optimizes mA to maintain constant image
color monitors;
performance; noise when collimation/detector

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 4 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

• Front and back work surfaces can be and physician flexibility to tailor technique Helical Multi-slice Modes:
set during installation within a range of to specific needs, optimizing patient dose,
vertical heights that help accommodate a and providing the power needed to perform
Simplified scan prescriptions and easy-to-
variety of siting requirements - a broad spectrum of axial and helical
use default protocols make the LightSpeed16
especially surrounding the height of the examinations.
fast and efficient in patient set up.
console relative to the window into the
scanning suite.
System Specifications Multi-slice acquisitions and short intergroup
• Work surface accommodates a large delays significantly reduce potential mis-
variety of working conditions and individual Scan Modes: registration between scans by increasing the
operator preferences. number of scans possible in a patient breath
The LightSpeed16 CT system can perform hold. Contrast agents may be better utilized
• Wide work surface has room for
virtually any clinical application due to its wide as well due to significantly faster scan
papers, patient charts, logbooks or other
variety of scan modes. With the LightSpeed16 acquisition.
peripheral devices;
, body CT studies are easier to perform and
• Wide area also helps make two person more productive than ever before. Helical protocols are almost identical to
operation more comfortable and “classical” axial scan protocols. At the
practical; beginning of a study, the operator selects the
Helical:
• All console components (monitors, type of exam with the anatomical
keyboard, mouse and BrightBox) are • Continuous 360° scanning with table programmer, and indicates the desired scan
free-standing and can be easily moved incrementation and no interscan delay. range - either manually, or from a Scout.
on the front work surface; • Scans can be acquired in a wide variety of
• All surfaces made of high-impact resin speeds. After completing the prescribed exam, the
that hides marks and are easy to clean. system remains ready to continue with
additional helical scans or a set of axial
• Direct network connection means a Axial:
scans.
multi-suite Ethernet card is not required for • Up to 16 contiguous axial slices acquired
a gateway out of the suite - saving costs simultaneously with each 360° rotation, The operator may reconstruct helical scans
and simplifying installation. with the time between scans set by the prospectively with up to 90% overlap, and
• Learning Solutions provides an on-line user-selected interscan delay (ISD) or retrospectively, at any arbitrary table location
Operator manual detailing system operation intergroup delay (IGD). in 0.1 mm increments.
via a multi-media CD-ROM player integrated • Scans may be easily clustered in groups to
into the front of the operator console. allow multiple scans in a single breath hold. Helical Multi-slice Modes
Learning Solutions can also be accessed
on a stand alone PC providing flexibility and • Minimum scan-to-scan cycle time of only
1.5 sec with table moves of ≤ 10 mm (any The complex nature of helical multi-slice
productivity for on demand learning of scanning has been simplified by grouping
system operation scan time).
all critical acquisition parameters within 8
basic scan modes, all optimized for image
Scout™:
Advanced Tube & Generator quality and speed. For eight-slice
Technology • Single radiographic plane for scan acquisition: 0.625: 1; 0.875: 1; 1.35: 1 and
localization and graphical prescription of 1.675: 1. For 16-slice acquisition: 0.5625:
prospective reconstruction; 1; 0.9375: 1; 1.375: 1 and 1.75: 1. These
• The Performix X-ray tube, with 6.3 MHU of • Extended range matches helical scannable
clinically derived multi-slice scan modes
storage and 53.2 kW operation, provides offer a wide range of selections that
range.
increased helical performance with greater carefully balance acquisition speed,
patient throughput and virtually no tube image thickness, artifact level and
cooling. Advanced technology in the Helical Scans retrospective image reconstruction
Performix tube includes a metal-ceramic flexibility.
frame for long life, a high-speed bearing for Slip ring technology has advanced axial
scanning by enabling scans with zero This simplified user interface guides the
sub-second scanning, a high-efficiency
interscan delay and simultaneous table user in the choice of scan parameters.
motor to accelerate the large anode and
movement. The user selects a pitch mode, a desired
efficient cooling for high throughput and
image slice thickness and table travel per
superior helical performance.
rotation. The user interface also displays
• Wide range of technique factors (10 to 440 the resulting choice of retrospective
mA, in 5 mA increments) gives the operator

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 5 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

image thicknesses available for each Helical Scan Parameters


choice of acquisition parameters.

The 16 slice helical acquisition modes 8-SLICE HELICAL MODES Scan Speeds: Full 360° rotational scans
provide table speeds from 5.625 in 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 seconds.
mm/rotation up to 35 mm per rotation, Table Speed (mm/rotation)
Slice
Scan Technique:
enabling scan speeds that are up to 12 Pitch
Thick-
times faster than 4-slice helical scanners. 0.625:1 0.875:1 1.35:1 1.675 — kVp: 80, 100, 120, 140
ness
(mm) :1
— mA: 10 to 440 in 5 mA increments
Prospective Multiple-
Thickness Reconstruction: 1.25 6.25 8.75 13.5 16.75 — Power: 0.8 to 53.2 kW

2.5 6.25 8.75 13.5 16.75 — Focal Spot Selection:


For all helical scan modes, the operator can
choose to reconstruct images prospectively 12.5 17.5 27.0 33.5
Small spot for up to 24 kW
in any of the defined nominal slice 3.75 6.25 8.75 13.5 16.75
thicknesses. 12.5 17.5 27.0 33.5 Large spot for greater than 24 kW

In addition to the initial reconstructed slice 5 6.25 8.75 13.5 16.75 Single Acquisition: 120 second scan
thickness, the operator has the option to 12.5 17.5 27.0 33.5 maximum.
prospectively specify additional images to be
7.5 6.25 8.75 13.5 16.75 Helical Tilt: helical acquisition is possible
reconstructed from a single raw data set.
These images can be reconstructed at any of 12.5 17.5 27.0 33.5 with the gantry tilted to a maximum of 30
the defined nominal slice thicknesses 10 6.25 8.75 13.5 16.75 degree, in half degree increments.
available for a given table speed and scan
12.5 17.5 27.0 33.5 Multiple Acquisition Maximum Scan Time:
mode.
Multiple scans may be acquired in one
This effectively facilitates later, more detailed series to produce up to 1,500 contiguous
image analysis without additional patient helical images. Up to 2,881 seconds
scans and subsequent dose and image helical coverage is possible in multiple
registration concerns. 16-SLICE HELICAL MODES series.

Table Speed (mm/rotation) Minimum Inter-Group Delay (IGD): 5


seconds between adjacent helical scans
Slice Pitch
Thick-
ness 0.5625 0.9375 1.375 1. 75:1 Scan Fields of View :
(mm) :1 :1 :1
— 25 cm for adult head

0.625 5.625 9.375 13.75 17.5 — 25, 50 cm for body

1.25 5.625 9.375 13.75 17.5 — 25 cm for pediatric head


11.25 18.75 27.5 35
2.5 5.625 9.375 13.75 17.5 Helical Scan Enhancements
11.25 18.75 27.5 35
Full simultaneity allows complete image
3.75 5.625 9.375 13.75 17.5
display, processing and analysis, as well
11.25 18.75 27.5 35 as image archival and filming, concurrent
5 5.625 9.375 13.75 17.5 with scanning and reconstruction -- even
11.25 18.75 27.5 35 when acquiring helical images in a multi-
slice mode.
7.5 11.25 18.75 27.5 35
10 11.25 18.75 27.5 35 Confirm Rx to X-Rays on: < 15 sec. for
any state of tube and gantry; < 10 sec.
with the gantry rotating

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 6 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Anatomical Programmer: a ten-region Reconstruction Matrix: 512 Prospective Multiple Reconstruction


anatomical selector allows quick and (PMR): Up to 3 sets of reconstructions
easy access to 90 user-programmable Display Matrix: 1024. can be pre-programmed as part of the
protocols per region. Separate selector scan protocol prior to acquisition. The
Display FOV: Freely variable center/off-
for adult and pediatric exams. There are operator can select different start/end
center, prospective/retrospective target
four selection tabs to select: GE, User, location, slice thickness, interval,
selection.
Service and Most Recent Patient. reconstruction algorithms and display
Copy/Paste is supported for easy CT Number Scale: -1024 to 3071 HU fields of view for each reconstruction.
modification and copying of protocols. This frees the operator from sitting at the
Helical Reconstruction Times: console and directly contributes to
Ten user-defined regions. Each region increased productivity.
has one default protocol displayed with  Reconstruction time as fast 6
the anatomical selector for very fast images per second Prospective Recon: Operator may
access to most commonly used protocols initiate full recons at any table location in
 Typical 0.17 sec. image-to-image increments of 1/10 the image thickness;
Protocols include preset scan time, kVp, recon in normal 16 slice recon mode. image thickness remains constant.
mA, scan mode, slice thickness and
spacing, table speed, scan FOV, display  Maximum image-to-image cycle Retrospective Recon: Operator may
FOV and center, recon algorithm and time is ± 10% for prospective and initiate full recons at any table location in
special image acquisition and processing retrospective image-to-image 0.1 mm increments; image thickness
options display. This applies for 512 matrix; remains constant.
any display FOV; in AutoView (all
Any scan parameters may be edited for layouts); with concurrent filming and Retrospective Image
each scan or all scans - either before or image archival for all scan modes. Decomposition: The operator has the
during an exam. The number of scans option to retrospectively decompose the
may also be easily changed.  Iterative bone processing original raw data set and reconstruct
increases time by 250 milliseconds. additional images at any of the defined
AutoScan™: Fully automates longitudinal Iterative bone processing, which is nominal image thickness available for a
table movement and start of each scan. always enabled for adult head given table speed and scan mode.
scanning, reduces image artifacts in
AutoVoice™: 3 preset (English) and 17 head scans stemming from X-ray
user-defined messages automatically beam hardening effects.
deliver patient breathing instructions;
especially useful for multiple helical Minimum DFOV: 9.6 cm
scanning.
Minimum Pixel Size: 0.19 mm
Trauma Patient: Allows patient scans
and image display/analysis without Queued Recon: Requests will be
entering patient data before scanning. processed continuously and
simultaneously with other processes on
Biopsy: Simplified prescription for single the system including scanning.
or multiple scans around an arbitrary Prospective recon will be prioritized over
table position aids biopsy studies. retrospective recon.

Advanced Artifact Reduction (AAR) Priority Recon Queuing: One touch


Filter significantly reduces streaking selection marks most recent rotation for
artifacts when highly absorbent objects next available recon. Available during or
are in the field of view – ie: large after scanning.
shoulders, screws.
Images annotated to indicate continuous
scan acquisition with table
Helical Image Reconstruction incrementation:

Reconstruction Algorithms : Soft Tissue, HE (helical) + Pitch


Standard, Detail, Bone, Bone Plus, Lung
and Edge Table speed

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 7 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

and coronal images of the 8” Catphan,


Contiguous Helical Coverage MTF module with bead source #CTP445.
Helical Scan Protocols
Multiple helical scans can be performed in
All protocols assume 120 kVp scans succession with only 5-second delays
under typical clinical conditions. between helical scans, providing up to Detail Algorithm
2,881 contiguous rotations (with up to
Single Helical Scans: 1,500 images in one series). MTF (z-axis) is 12.4 lp/cm @ 10%, 15
lp/cm @ 0% which is a 0.333 mm limiting
(The following chart assumes 120- resolution
Scan Maximum second helical with 5-sec IGD)
Time mA In Plane Helical Scan Image Quality
10 sec. 440 Total Scan Total Elapsed Max
Time (sec.) Time (sec.) mA For details of scan technique please refer
20 390-440
to the technical reference manual.
30 330-380
150 155 100-200
40 290-350 200 205 80-150 High-Contrast Spatial Resolution
50 260-320 450 465 50-70
60 230-310 On GE Performance phantom:
70 210-300
80 180-280
Helical Scan Image Quality Standard Algorithm

90 170-270 • 0.584 mm limiting resolution


Volumetric Helical Scan Image
100 160-260
Quality • 4.0 lp/cm @ 50% MTF
110 150-250
16
120 140-230 With LightSpeed being a sub-millimeter • 6.5 lp/cm @ 10% MTF
isotropic CT scanner, it is now possible to
Multiple Helical Scans specify coronal and sagittal image quality: • 8.5 lp/cm @ 0% MTF
(contiguous helical coverage with 5- 1. Visual Measurement: Hi-Res Algorithm (Edge).
second IGD):
Reformatted resolution is demonstrated • 0.324 mm limiting resolution
Scan No. Max on the Nuclear Associates AAPM High
Time IGD Scans mA Contrast Resolution Insert #76-413. • 8.5 lp/cm @ 50% MTF
10 sec. 5 sec. 3 340-420
4 300-400 0.5mm • 13.0 lp/cm @ 10% MTF
5 270-390
• 15.4 lp/cm @ 0% MTF
6 240-360
20 sec. 5 sec. 2 300-360 Line pair values decrease with larger
3 240-330 focal spot (by 5% with Standard and by
4 180-310 7% with Hi-Res); limiting resolution is
30 sec. 5 sec. 2 230-310 unaffected.
3 180-280
Low-Contrast Detectability
4 140-240
40 sec. 5 sec. 2 190-290
3 140-230 On 8 inch (20 cm) CATPHAN phantom:
4 110-190 5 mm @ .3% at 13.3 mGy
0.4mm
50 sec. 5 sec. 2 170-270 3 mm @ .3% at 37.2 mGy
3 120-200 An Effective 0.5 mm voxel size is clearly
4 90-160 Noise:
seen in coronal and sagittal views.
60 sec. 5 sec. 2 140-240 On either an AAPM water phantom or GE
3 100-170 2. Statistic Measurement:
Quality Assurance phantom:
Volumetric MTF is computed from the X,Y
0.32% +/- 0.03% at 28.5 mGy (2.85 Rad)
and Z axis of a 0.2mm bead using sagittal

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 8 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

CTDI: • Nominal Thickness: 0.625, 2.5, 5, 7.5, 10 • Minimum scan-to-scan cycle of 1.5
mm seconds possible for 0.5 sec. scan speed
On CTDI Head and Body Dose Reference with minimum ISD's.
Phantoms: 4i Mode:
• Produces 4 images per rotation Scan Fields of View:
CTDIvol expressed in mGy/100 mAs for
IEC pitch 1: (normalized to a pitch of 1). • Nominal Thickness: 2.5, 3.75, 5 mm • 25 cm for adult head
• 25, 50 cm for body
Head 20.9 mGy/100 mAs 8i Mode:
• 25 cm for pediatric head
Body 10.6 mGy/100 mAs • Produces 8 images per rotation
• Nominal Thickness: 1.25, 2.5 mm Scan with no table incrementation,
Axial Scans contiguous image location, or skipped
16i Mode: image location are possible. Overlapped
Multi-slice acquisitions and short interscan axial scans are not possible.
• Produces 16 images per rotation
delays significantly reduce potential missed
registration between scans by increasing the • Nominal Thickness: 0.625, 1.25 mm
Axial Image Reconstruction
number of scans possible in a patient breath
hold. Contrast agents may be better utilized Axial Scan Parameters Reconstruction Algorithms: Soft Tissue,
as well due to significantly faster scans.
Standard, Detail, Bone, Bone Plus, Lung and
Scan Time: Edge.
Axial Multi-slice Prescription • 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 2.0, 3.0 and 4.0
Reconstruction Matrix: 512
second full scans (360° acquisition)
Simplified scan prescriptions and easy-to- Display Matrix: 1024.
use default protocols make the Lightspeed16 Scan Technique:
CT Scanner fast and efficient in patient set- • kVp: 80, 100, 120, 140 Display FOV: Freely variable center/off-
up. Axial protocols are nearly identical to center, prospective/retrospective target
• mA: 10 to 440, in 5 mA increments
helical scan protocols. selection.
• Power: 0.8 to 53.2 kW
• Focal Spot Selection: CT Number Scale: -1024 to 3071 HU
Axial Multi-slice Modes
• Small spot for up to 24 kW Axial Image Reconstruction:
The LightSpeed16 acquires axial scans in • Large spot for greater than 24 kW
• Reconstruction time as fast as 6 images
sets of 8 or 16 contiguous images in one
Scan Plane Geometry: per second.
360° rotation.
• +/- 30° gantry tilt, in 0.5° increments • Typical 0.17 sec. image-to-image recon in
For each rotation of the gantry, the 16 slice recon mode.
LightSpeed16 collects 16 rows of scan • Longitudinal positioning in 0.01 mm per slice
data. There are five reconstruction increment. Gantry display in 0.5 mm • Maximum image-to-image cycle time is ±
modes available for creating images from increments. 10% for prospective and retrospective
the multi-slice scan data (1i, 2i, 4i, 8i, and image-to-image display. This applies for
Interscan Delay (ISD): 512 matrix; any display FOV; in
16i). By using 1i, 2i, 4i, and 8i
reconstruction modes, scan data can be • Minimum ISD with table movements of AutoView (all layouts); with concurrent
combined prior to image reconstruction to 0 - 10 mm: 1.0 sec. filming and image archival for all scan
modes.
create slices with reduced partial-volume • Minimum ISD with table movements of more
artifacts. This is particularly useful for than 10 mm and up to 20 mm: 1.3 sec • Iterative bone processing increases time by
posterior-fossa imaging. 250 milliseconds.
• User-selectable.
Prospective Multiple Reconstruction
Inter Group Delay (IGD):
1i Mode: (PMR): Up to 3 sets of reconstructions
• Minimum IGD is the same as minimum ISD; can be pre-programmed as part of the
• Produces 1 image per rotation also user-selectable. scan protocol prior to acquisition. The
• Nominal Thickness: 1.25, 5, 10 mm operator can select different
Scan-to-Scan Cycle:
reconstruction algorithms and display
2i Mode: fields of view for each reconstruction.
• Produces 2 images per rotation This frees the operator from sitting at the

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 9 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

console and directly contributes to Standard Scans: On GE Performance phantom:


increased productivity.
Standard Algorithm
The operator has the option to
Scan Acquisition • 0.584 mm limiting resolution
reconstruct the original raw data set at
Time ISD m A Scans Time
any of the defined nominal slice
1 sec. 1 440 18-45 0:35-1:27 • 4.0 lp/cm @ 50% MTF
thicknesses.
sec.
• 6.5 lp/cm @ 10% MTF
Reconstructions can be prescribed down 1 1 400 24-55 0:47-1:49
to 1/16 the original acquisition image • 8.5 lp/cm @ 0% MTF
1 1 360 32-68 1:03-2:15
thickness for images acquired in the 1i
scan mode, down to 1/8 the original 1 1 340 37-76 1:13-2:31 Hi-Res Algorithm
thickness for 2i mode, and down to 1/4 1 1 320 43-86 1:25-2:49
the original thickness for 4i mode. • 0.324 mm limiting resolution
1 1 300 50-97 1:37-3:11
Similarly, additional reconstruction 1 1 280 58-110 1:45-3:37 • 8.5 lp/cm @ 50% MTF
supports partial-volume artifact reduction 1 1 260 66-122 1:55-3:59
by reconstructing images with 4, 8, or 16 • 13.0 lp/cm @ 10% MTF
1 1 240 74-135 2:07-4:21
times the acquisition image thickness.
1 1 220 84-152 2:21-4:49 • 15.4 lp/cm @ 0% MTF
These reconstruction features effectively 1 1 200 94-168 2:37-5:21 Line pair values decrease with larger
facilitate later, more detailed image
2 1 200 37-77 1:13-2:33 focal spot (by 5% with Standard and by
analysis without additional patient scans
2 1 180 42-86 1:23-2:51 7% with Hi-Res); limiting resolution is
and subsequent dose and image
unaffected.
registration concerns.

• Cluster Scans (All cluster protocols Low-Contrast Detectability


The following table illustrates the
retrospective reconstruction image assume 9-second clusters of five slices, 1-
On 8 inch (20 cm) CATPHAN phantom:
thicknesses available for each acquisition second scans with 1-second interscan
thickness and mode: delays and 7 seconds between clusters). 5mm @ .3% at 13.3 mGy
3mm @ .3% at 37.2 mGy
Prospective Perscription # Scans Acquisition
Slice Recon mA (Clusters) Time Noise:
ScanMode Thickness Slice Thicknesses 440 15-45 (3-11) 0:41-2:49
16i - 0.625mm On either an AAPM water phantom or GE
400 25-55 (5-13) 1:13-3:21 Quality Assurance phantom:
8i - 1.25mm
16 row 0.625 340 35-75 (7-18) 1:45-4:41
4i - 2.5mm
2i - 5.0mm 0.32% +/- 0.03% at 29.3 mGy (2.93 Rad)
320 45-85 (9-20) 2:17-5:13
16i - 1.25mm
300 50-95 (10-22) 2:33-5:45 CTDI:
8i - 2.5mm
16row 1.25
4i - 5.0mm 280 60-110 (12-25) 3:05-6:33
On CTDI Head and Body Dose Reference
2i - 10mm 260 75-125 (13-28) 3:21-7:21 Phantoms:
8i - 1.25mm
4i - 2.5mm 240 85-145 (14-31) 3:37-8:09
8 row 1.25 CTDI expressed in mGy/100 mAs:
2i - 5.0mm 220 100-170 (16-34) 4:09-8:57
1i - 10mm 200 115-185 (18-38) 4:41-10:01 Head 16.8 mGy/100 mAs center
8i - 2.5mm
8row 2.5 4i - 5.0mm 180 135-250 (22-43) 5:45-11:21 17.0 mGy/100 mAs surface
2i - 10mm
Axial Scan Image Quality Body 5.50 mGy/100 mAs center
Axial Scan Protocols 10.0 mGy/100 mAs surface
For details of scan technique parameters,
CTDI100 expressed in mGy/100 mAs:
All protocols assume 120 kVp scans please refer to the technical reference
under typical clinical conditions. manual. Head 18.7 mGy/100 mAs center
17.9 mGy/100 mAs surface
High Contrast Spatial Resolution:

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 10 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Body 5.4 mGy/100 mAs center • Image monitor for image display, ImageWorks:
11.2 mGy/100 mAs surface analysis, processing, and management
• Each monitor provides a 1280 x 1024 ImageWorks is a desktop environment
CTDIvol expressed in mGy/100 mAs: designed to take advantage of the
high resolution, flicker-free display
Lightspeed16 CT Scanner System computer
Head 18.4 mGy/100 mAs • Optional 18” LCD Monitors.
and image processor.
Body 9.3 mGy/100 mAs • Scan control keyboard assembly with
intercom speaker, microphone and volume Standard features include archive, network
Scout Scans controls and manual film control, as well as some
• Three button mouse with mouse pad advanced image processing such as multi-
ScoutView™ scans provide excellent planar reformatting (MPR), multi-projection
• BrightBox (trackball assembly) volume rendering (MPVR), and MR image
detail for anatomical localization in
conjunction with scan prescription. • Two wide work surfaces display. It also has optional add-on packages
for Volume Viewer*, CT Perfusion2* and
Scan locations may be prescribed at the All these devices are free-standing and can DentaScan*.
operator console either graphically (via be easily moved to accommodate a large
mouse), or explicitly (keyboard entry) variety of working conditions and individual The ImageWorks desktop also provides a
from a Scout scan. operator preferences. gateway for DICOM 3.0 image transactions,
either through a local area network, or via
Prescription of scans with multiple gantry Split table top allows unrestricted patient DICOM-formatted MOD media.
angles are also available on a single viewing while still supporting 2 monitors.
Scout. Each work surface can be adjusted at Learning Solutions:
installation to help accommodate a variety of
siting requirements. The LightSpeed16 provides an on-screen,
Scout Scan Parameters on-line operator manual via a multi-media CD-

Aperture: 1.25 mm effective aperture


Desktop Overview ROM player integrated into the front of the
operator’s console. Learning Solutions is also
The user interface utilizes the paradigm of viewable on a stand-alone PC providing
Table speed: 100 mm/sec.
managed work environments for a more flexibility and productivity for on-demand
Scan Technique: intuitive clinical workflow. learning of system operation.
• kVp: 80, 100, 120, 140
Virtually all clinical operations are managed
• mA: 10 to 440 in 5 mA increments through three “virtual desktops” or
• Power: 0.8 to 53.2 kW applications managers: Exam Rx,
Exam Rx
ImageWorks and Learning Solutions.
Orientation: AP, RLAT, PA, LLAT (preset); Operators can effortlessly move back and
or any angle from 0° - 359° (manually forth between these environments simply by
selected). clicking on an icon. Xtream TM technology
enhances multi-tasking architecture and Patient Scheduling
Axial scan prescription lines indicate
maintains simultaneously all processes so no
scan location to nearest 1 mm table
work is lost or disrupted as desktops are Patient demographics and exam protocols
position.
switched. can be pre-programmed in advance of patient
Scouts longer than 1,000 mm are auto arrival by selecting Schedule Patient from the
minified to fit the display. Exam Rx: scan/recon monitor. This productivity
enhancement allows entry of all or some of a
The Exam Rx desktop environment provides patient’s demographic data, as well as pre-
User Interface the clinical tools necessary for comfortable, selection of the exam protocol.
efficient control of patient studies.
The Lightspeed16 Operator Console utilizes a This feature is available any time a patient
computer workstation with the following user These tools include patient scheduling and exam is not currently underway.
interface features: data entry, exam protocol selection, protocol
• Two 20-inch monitors viewing and editing, scan data acquisition, This feature uses the same interface as New
image reconstruction, image display and Patient selection for simplified, consistent
• Scan/recon monitor for scan and recon
routine analysis, AutoFilm or manual filming, programming.
control with no image display
AutoStore and AutoTransfer.

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 11 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Patient information can be easily recalled to display the 1024 Localizer each time a new Scan Data Acquisition:
set up an immediate exam via List/Select series is requested.
Scheduled Patient on the scan/recon monitor. • Default protocols also include preset scan • Full-screen DynaPlan Plus illustrates scan
Pre-programmed patient exams can also be time, kVp, mA, slice thickness, scan mode, status graphically, with real-time feedback
recalled from the New Patient screen table speed, image interval, gantry tilt, scan while the exam is underway. Scans,
automatically by entering the patient ID field-of-view, display field-of-view and programmed delays (prep, breathing, inter-
number. center, recon types, and breath timing group), and even AutoVoice
parameters. announcements are clearly shown before
and during scanning.
Patient Data Entry: • Any scan parameter can be edited for
each scan or all scans either before or • AutoScan: Fully automates longitudinal
Patient data can be entered as part of New during an exam. Scans can be easily table movement and the start of each scan
Patient set-up, or can be recalled from the list added or removed from the prescription. • AutoVoice: Preset (English) and user-
of pre-scheduled patients. recorded messages automatically deliver
• Scan/recon control uses only 2 screens to
set up first scan - New Patient and Protocol patient breathing instructions, especially
Trauma Patient ID allows patient scans and useful for multiple or multi-pass helical
image display/analysis without entering View/Edit.
scans
patient data before scanning.
• Full Simultaneity allows scan and recon to
Protocol View/Edit: work concurrently with image display,
Exam Protocol Selection: • A single, full screen View/Edit table allows processing and analysis (including
fast and easy examination and modification computationally intensive features such as
One of the main contributions of the of exam parameters before scanning MPR, MPVR and 3D*/MIP) while still running
LightSpeed16 CT Scanner System to begins image archival, filming and networking
department productivity is its simplified exam processes.
• Exam parameters can be changed for just
set-up.
one scan, or for all scans in a series
• Exam parameter set-up has been greatly Dose Computation & Display
• When used in conjunction with the 1024
simplified through the exclusive use of
Localizer, changes made in the View/Edit
protocols CTDIvol (CTDI volume), DLP (Dose Length
table that affect the number of scans,
• Protocols can be easily selected in one of image interval, starting/ending locations, tilt,
Product), and Dose Efficiency computation
three convenient ways: and display during scan prescription provides
or display FOV are automatically shown on
patient dose information to the operator.
• A large, graphical Anatomical the 1024 Localizer
Programmer located on the New Patient • Any changes made directly on the 1024 CTDI vol is a dose index defined by IEC 601-2-
screen Localizer display using the mouse and the 44. This index is computed automatically by
• A default list of the “top 10” most on-screen cursor controls are also the Lightspeed16 CT System and reported on
commonly used protocols located near reflected automatically in the View/Edit the Exam Rx screen. CTDIvol is a single
the anatomical programmer table number consisting of 2/3 of the CTDI100
• A numerical entry peripheral dose plus 1/3 of the CTDI100 central
• View/Edit Wizard intuitively adjusts
dose that is divided by the helical or axial
• Two Anatomical Programmers - one for dependent parameters automatically in
pitch factor.
adults and one for pediatrics - provide response to operator-initiated changes and
quick and easy access to 8,460 user- highlights them for quick review. It also CTDI100 is a dose index based upon CTDI
programmable protocols (total). Each alerts the operator to incompatible dose measurements over a 100 mm volume,
programmer has ten anatomical regions dependencies requiring operator as defined in IEC 601-2-44.
with 90 protocols for each region intervention.
Dose Length Product (DLP) is given in
• Default protocols have been expanded • Tab card groupings for Timing, Recon and mGy*cm and is computed and displayed for
through Protocol Pro - a “behind the Filming help organize the large number of each group prior to the scan. Additionally, an
scenes” protocol manager - that allows parameters available within each protocol. accumulated DLP is displayed for the entire
preselection of automated features like • As many as 8,460 protocols can be stored exam, as the exam prescription progresses.
AutoVoice, AutoFilm, AutoStore and on the Operator Console. The final exam accumulated DLP provides a
AutoTransfer on a per-exam basis. convenient measure for maintaining patient or
• Protocol Pro also provides preselection of procedure dose management statistics.
two different window/level settings per
Dose Efficiency is automatically
image for AutoFilm and can automatically
computed and displayed on the Exam Rx

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 12 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

screen. The dose efficiency is a image review/processing without • Screen Save


measure of how much of the Z-axis X- interruption during AutoFilm. • Gray Scale Enhancement
ray beam is used by the system, as
• ProView visualization algorithms are
defined in IEC 601-2-44.
Image Review Layouts: available to enhance anatomical structures
without additional reconstruction time:
AutoView Layouts: • Five flexible Image Review Layouts are
• Four Selections for enhancement of high
provided for those applications where
contrast objects where fine detail is
• Eight powerful AutoView layouts provide greater than 512 image display may be
required without aliasing (such as
exceptional flexibility in tailoring the 1,024 desired and AutoView is not required.
lungs)
image display to the user or the application • Image Review Layouts include:
at hand - without the complexity of free- • Three Selections for modifying
form “windows.” Note: uses short notation for screen options perceived levels of noise and low
• 1024 single image display contrast discrimination
• AutoView Layouts include:
• 768 single image display • Three ways are provided to adjust
• 1024 AutoView image
• Two 512 image display, horizontal window/ level of images in focus in order
• 768 AutoView image (matches the to meet a variety of clinical work
format
image size shown on the HiSpeed environments and user preferences:
Advantage 2.X Series OC monitor) • Two 512 image display, vertical format
• Six user-programmable keys on the
• 512 AutoView image + 512 Localizer • Four 512 image display
scan control keyboard (F6 - F11), plus
Scout with cut lines automatically • Each image display window can be further one key for returning to prior setting (F5)
showing the location of the AutoView subdivided into four more images,
• On-image through middle mouse button
image on the Scout increasing the total number of images that
• BrightBox trackball
• Two 512 AutoView images (same image can be displayed at once to 16.
but at different window/level settings) + • BrightBox image control is also available for
512 Localizer Scout with cut lines Image Review Layouts. Routine Measurements:
automatically showing the location of the
AutoView images on the Scout • Image measurement features provided
• 512 AutoView image + 512 AutoFilm
Image Access: within Exam Rx:
image • Box ROI
• Point and click interface along with a
• Last two 512 AutoView images pictorial directory (browser) allows for • Ellipse ROI
• Last four 512 AutoView images easy selection by exam, series or image • Trace ROI
• AutoLink which links the current series • Measure Distance
to a view port Routine Image Display: • Measure Angle
• Basic image review features such as • Grid On/Off
• Image display features provided within
window/level, magnification and flip/rotate • Hide Graphics
Exam Rx:
are available for AutoView images. • Erase
• Zoom/Roam
• Any window not used for AutoView is • Screen Save
• Explicit Magnify
available to independent, simultaneous • MIROI (Multiple Image ROI)
review of other exams. • Flip/Rotate
• Report Pixels
• Special BrightBox, a three-button trackball • ProView
device, provides independent control of • Display Normal
image next, prior, manual paging and • List/Select Display Preferences:
trackball window/level for any review • Ellipse ROI • Display settings available to tailor the
images in focus. This helps make two
• Measure Distance overall display (settings apply to all images
person operation practical.
• Grid On/Off in all exams):
• Regardless of the AutoView Layout used, • Annotation Levels
• Cross Reference
AutoFilm viewing is available anytime via an
• User Annotation • Inverse Video
on-image selection - without disrupting
other image processes in progress. • Exam/Series Page • Next/Prior Each View Port
Background filming allows full use of the • Hide Graphics • Next/Prior Series Binding
image display monitor for AutoView and
• Erase • Continuous Report Cursor

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 13 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Auto Image Management: Because some of the image analysis and • Pseudo Surface Shading Mode
display features of ImageWorks replicate • Predefined Cut Planes
those in Exam Rx, the next section describes
The Exam Rx work environment conveniently
only features that are incremental or • Volume Rendered Navigator views
provides for selection of AutoFilm, AutoStore
significantly different. • VR Preset save/recall
(to local or remote MOD), and AutoTransfer
(across a network). • 3D Rendered Lumen View
Image Analysis • Automatic Path Tracking
An AutoFilm Composer provides a simple
programming interface for automated filming • Path Bridging (in case of occlusions)
• Multi-Projection Volume
set-up. Reconstruction (MPVR): Quick and • SmartCursor™ for Easy Navigation

Batch Filming is accomplished through a


easy way to generate volumetric images • Synchronized Reformatted Views
for CT angiography without thresholding
single keystroke which automatically prints
data or removing unwanted anatomy. An • Cut visualization mode
an entire series at a time.
entire volume is used to generate images
in any plane, creating real-time frames of
Manual Image Filming: reference at the same time; *Advanced Vessel Analysis
• On-screen filming is available for any • Clinical utility is extended via two Advanced Vessel Analysis is the ultimate
analog or digital camera using a 3M-952 additional modes: tool to assess and quantify vascular
protocol. structures, including stenosis analysis,
- MIPS - enhances contrast and improves stent planning procedures, post stenting
• Images may be individually filmed manually
visualization of calcifications or vascular surgery follow -up.
via “drag and drop” to the on-screen Film
Composer. - Average - generates 2D radiographic • protocol driven tools to perform quick,
• Print Series permits automatic printing of an images; flexible and accurate quantitative analysis
entire series with one keystroke. of vascular anatomy
• Multi-planar Reformation (MPR): • provides maximum, minimum and mean
• Page filming permits creation of an entire
Provides real-time assessment of intraluminal diameter measurements
film with one keystroke.
anatomy in offaxis planes. Sagittal,
• Multiple image formatting allows filming of coronal, oblique and curved planar • provides cross-sectional areas of true
multiple images in a single film frame. reformations available; orthogonal sections of the aortoiliac
systems at selected anatomical points
• Film formats supported are 1:1, 2:1, 4:1,
6:1, 8:1, 9:1, 12:1, 15:1, 16:1, 20:1, 24:1
• Batch reformatting can also be • clinical benefits include: stenosis sizing,
defined and executed for later viewing if pre- and post- surgical assessment, stent
and 35-mm slide
desired; planning
Important note: The Lightspeed16 CT Scanner • Measurements in % stenosis or mm of
• Image Addition and Subtraction: Includes
comes standard with a DICOM Print Interface stenosis, and measurement of length and
image addition of more than two images
configurable for multiple DICOM Print dimension of stenosis.
at a time;
destinations. Connections with cameras that
do not support DICOM Print may require a
filming interface (purchased separately).
*Volume Viewer Image Display
To save further filming cost, the Operator
Console can directly print to a postscript Volume Viewer is an innovative and • Magnifying Glass allows quick 2X mag
printer such as the GE Color Printer available powerfull suite of productivity enhanvers window that can be moved over an image.
as an option. (Volume Rendering, Volume Analysis and • Image Scroll moves an image within its’
Navigator) that includes : own window.
ImageWorks • Dynamic Volume Review™ for Fast • Groupings allow application of
Screening window/level values,
ImageWorks software is designed to take • Curved Volume Of Interest magnification/minification, image scroll or
advantage of the Lightspeed16 CT Scanner’s flip and rotate to a user-defined image set.
• Protocol Management and Loading
computer and image processor. This desktop
• Save State stores user-selected image
environment includes image management and • Review Layout Presets
orientation and window/level with each
networking. • Multiple VR Objects Merge data set.

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 14 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

• Window/Level values may be: • Batch film can typically enable you to Industry Standards
• Preset to provide six on-screen instant reduce filming images by 50%, thickening
the slab from 1.25 to 2.5mm with no • The LightSpeed16 CT Scanner System
window/level settings
information loss. complies with a wide variety of industry
• Set independently for up to 16 images standards to facilitate more rapid adoption
on the screen of features and performance improvements
• User-modified in discrete or variable Image Networking: as the computing and medical imaging
steps industry evolves.
• Adjusted real-time on-image by holding Exams can be selected and moved from
down the middle mouse button and MR Signa Advantage, CT HiSpeed
Advantage, Advantage Independent DICOM Conformance
moving the mouse
Console, Advantage Windows Standards:
• Cine mode provides paging in up to 4 view
Workstation to the Lightspeed16 CT
ports of up to 128 previously-stored CT or
Scanner’s Operator Console. This is • DICOM 3.0 Storage Service Class
MR images at full selected display frame • Service Class User (SCU) for image
performed via the AdvantageNet
rate. For more than 128 images, display send
(GenesisNet) point-to-point protocol. This
frame rate may be reduced.
protocol supports send, receive and • Service Class Provider (SCP) for image
• Cine mode also provides temporal, spatial pull/query (no broadcast). NOTE: receive
or manual playback loops. Because the Lightspeed16 creates
• DICOM 3.0 Query/Retrieve Service Class
• Text Page images in pure DICOM format, exams can
not be moved from the Lightspeed16 CT • DICOM 3.0 MOD Media Service Class on
Scanner System to MR Signa Advantage, 1.2- and 2.3-GB MOD media
Image Annotation CT HiSpeed Advantage or the Advantage • DICOM 3.0 Storage Commitment Class Push
Independent Console systems.
• Image annotation and cursor are • DICOM 3.0 Modality Worklist*
shadowed to permit ease in reading. Exams can also be selected and moved • DICOM 3.0 Modality Performed Procedure
between the Lightspeed16 CT Scanner Step*
Image Management System and any imaging system
• DICOM 3.0 Print
supporting the DICOM 3.0 protocol for
• Images may be stored and retrieved via network send, receive and pull/query. GE
Magnetic Optical Drive (MOD) media using systems that support this interoperability Filming Protocol:
DICOM 3.0 format. This allows interchange include CT HiSpeed CT/i, X/i, NX/i, QX/i,
with other imaging systems supporting • 3M-952 Standard
MR Signa LX, and Advantage Windows
DICOM 3.0 MOD media. Not all vendors 1.2/2.x/3.x/4.x.
implementation of DICOM 3.0 are identical, System Components
so please check with the manufacturer for Image transfer time using either
compatibility. AdvantageNet or DICOM 3.0 protocols is
approximately 0.1 second per 512 image Gantry:
• Off-line retrieval of all image files. Images
on a 100baseT Network.
may be viewed as soon as they are
Advanced slip ring design continuously
restored from MOD.
rotates generator, tube, detector and data
• Image retrieval time is approximately 30 Computer Based Training acquisition system around the patient.
images 512 per second. This innovative feature provides on-
• Aperture: 70 cm
screen, on-line support of Exam Rx and
ImageWorks functions via a multi-media • Tilt: ± 30°
VariViewerTM: CDROM player, integrated into a tower on • Tilt Speed: 1°/sec.
• VariViewer is an interactive axial review operator console for easy access. Use
• Focus to Detector: 95 cm
mode that can change the slice thickness of rich graphics and text to provide truly
reconstruction instantaneously. useful system help. • Focus to Isocenter: 54 cm

• The user selects the volume to be • Maximum SFOV: 50 cm


A well-indexed table of contents helps
specifically analyzed and chooses the slice speed access to information. The CD can • Rotational Speeds: 360° in 0.5, 0.6, 0.7 0.8,
thickness to be displayed as axial images. be viewed on the scanner or on a PC. 0.9, 1.0, 2.0, 3.0 and 4.0 sec.
The user can then save a number of recon • Remote Tilt from Operator’s Console.
images sets, viewing a large number of
slices for pure axial review and filming.

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 15 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

• Integrated breathing lights and countdown • 205 kg (450 lb) maximum allowed with • mA: 10 to 440 mA, 5 mA increments
timer. normal operation and +/- 1 mm positional
accuracy Maximum mA for each kVp selection:
• Integrated start scan button with
countdown to X-ray on. • Controls on gantry for elevation and cradle
incrementation. Foot pedals on both sides
Scan plane toward front of gantry for of table for fast elevation. Cradle position KVp Max mA
improved positioning access. controlled from OC for prescribed scans. 80 400
Biopsy and interventional studies have been • IV Pole integrated at the foot-end of the 100 420
facilitated through a more streamlined gantry table prevents IV lines from becoming 120 440
shroud, and bilateral table/gantry controls crossed and tangled, and ensures that the 140 380
and gantry display that maximize lines stay securely in place on the patient.
maneuverability while working next to the HiLight Matrix II Detector:
gantry. X-Ray Tube
Laser Alignment Lights: 21,888 individual elements composed by: 8
Performix ADV Metal-Ceramic Tube Unit. rows of 1.25 mm thickness and 16 rows of
• Define both internal and external scan Design optimized for exams requiring a 0.625 mm thickness, each containing 888
planes to ± 1 mm accuracy. large number of scans without tube active patient elements; 24 reference
• Operate over full range of gantry tilt; cooling. elements.
activated any time during exam (with tube • Heat Storage Capacity: 6.3 MHU 4 typical modes of data output:
stationary). • 8 x 1.25 mm (uses center 16 rows)
• Heat Dissipation:
• Coronal light remains perpendicular to axial • 8 x 2.5 mm (uses all 24 rows)
• Anode (max) 840 KHU/min
light as gantry tilts.
• Casing (cont) 300 KHU/min • 16 x 1.25 mm (uses all 24 rows)
Visual readout is easy to read from the table • Tube Unit: 6.9 kW continuous for 10 • 16 x 0.625 mm (uses center 16 rows )
side or from the operator console. minutes
70% geometric efficiency; 98% absorption
Gantry tilt controls are located on the side of • Dual Focal Spots: efficiency.
the gantry. • Small Focal Spot:
0.7 mm (W) x 0.6 mm (L) nominal
value
Data Acquisition System
Table (IEC 336/93)
12,288 available input channels.
• Single table, cantilever design with wide 0.9 mm (W) x 0.7 mm (L) traditional
height range methodology 1,640 Hz maximum sample rate.
• Vertical Range: 51.6 cm to 99.1 cm • Large Focal Spot: Effective analog to digital conversion range
• Vertical Scannable Range: 80.6 cm to 99.1 0.9 mm (W) x 0.9 mm (L) nominal greater than two million to one.
cm value
(IEC 336/93)
• Elevation Speeds: 5 mm/sec and 40 Scan/Control Unit
mm/sec 1.2 mm (W) x 1.2 mm (L)
traditional methodology
• Horizontal Range: 170 cm Located in base of Operator
• Horizontal Scannable Range: 170 cm • Maximum Power: 53.2 kW Console.
metal-free (axial) and 160 cm metal-free • Beam collimated to 55° fan angle.
(helical & Scout)
Average time to replace tube: < 10 hours
• Horizontal Speed: Up to 100 mm/sec Host Computer
• Table automatically recenters on scan
High Voltage Generation • Dual SMP 2.66.GHz Intel Xeon processors
plane with changes in vertical position
with 512KB L2 cache.
(after setting internal landmark with • High-frequency on-board generator.
alignment lights on) • Intel Hyper-threading technology.
Continuous operation during scan.
• Table Load Capacity: • 2GB DDR266 Dual Channel Memory with a
• 53.2 kW output power.
• 180 kg (400 lb) with +/- 0.25 mm throughput of 4.2GB/sec
• kVp: 80, 100, 120, 140
positional accuracy

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 16 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Image Processor: • UltraSCSI interface • Protocols supported:


• Scan data disk drive: • DICOM 3.0 network send (one IP
• Nvidia Quadro4 980XGL AGP 8X graphics address at a time) and receive,
• High Performance Drive
with 128MB Memory pull/query, and storage commitment
• 36 GB, 3.5 inch form factor
• Graphics Processor Unit (GPU) Clock push;
• UltraSCSI interface
300Mhz • AdvantageNet (GenesisNet) point-to-
• Graphics Memory Clock 325Mhz point send, receive and pull/query (no
• Standard MOD drive: broadcast);
• Magnetic Optical Disk Drive • InSite point-to-point;
• Erasable, rewritable media • TCP/IP (for system administration)
Image Reconstruction Engine
• 2.3 GB, 3.5 inch form factor
(GRE)
• Assigned to DICOM 3.0 format image file. Compatible Options
• Stores 4,700 lossless JPEG compressed • A DICOM Print Interface is standard on the
• Custom-designed special purpose
512 image files per side system.
CT Image Generator
• Custom CT back projection • Off-line retrieval of image. Images may • English Keyboard
hardware accelerates 2D & 3D back be viewed as soon as they are restored
from MOD • French UIF & Keyboard
projection.
• Intel Hyper-threading Technology. • German UIF & Keyboard
• 32-bit floating point data format • DVD Ram: • Italian UIF & Keyboard
• 2GB DDR226 ECC Dual Channel • 4.7 GB per side, 5.25" half height form • Spanish UIF & Keyboard
Memory Standard (4.2 GB/sec). factor
• Portuguese UIF & Keyboard
• Transfer rate 2.7MB/sec
• ConnectPro HIS/RIS Interface with
Software Architecture: • Assigned to scan data file and protocol Performed Procedure Step (PPS) M4
file storage/retrieval. compatibility
• Software architecture based on industry
standards and client-server design •
• Color monitors:
• Volume Viewer
• 21 inch diagonal width
Peripherals: • DentaScan
• 1280 x 1024 dot resolution
• Cardiac imaging
Total of 254 GB system: • Non-interlaced, flicker-free presentation
• 76 kHz Horizontal deflection frequency • CT Perfusion 2
• Main system (host) disk drive:
• 72 Hz Vertical deflection frequency • Smart Step
• High Performance Drive
• 36 GB, 3.5 inch form factor • Sync on green • Advanced Vessel Analysis
• 15,000 RPM • LCD Color Monitor*: 18” diagonal width •
• Ultra320 SCSI interface • Scan control keyboard (English language) • Remote 21 inch Color Monitor
• Assigned to applications software and assembly with intercom speaker, • UPS
image files microphone and volume controls.

• 2 system disk drives (Image Disk) • Global modem to allow InSite connectivity
• LCD Monitor
• High Performance Drive • 3-Button Mouse
• 73 GB, 3.5 inch form factor each. • 3-Button Trackball
Siting Requirements
• 15,000 RPM
• Ultra320 SCSI interface Image Networking: Minimum Suite Size:
• Assigned to image files only
• Standard auto-configuring Overall Suite:
- 250,000 uncompressed 512 images 100BaseT/10BaseT Ethernet (UTP
• Scan data disk drive: connection) 15.5 feet (4.74 m) wide x 19.5 feet (5.96 m)
deep x 9 feet (2.75 m) high.
• High Performance Drive • Direct network connection; multi-suite
• 36 GB, 3.5 inch form factor ethernet card not required for gateway out
of suite

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 17 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Recommended Suite Size: general, suppresser filters are not


recommended.
Overall Suite: System Components Dimensions
Compatible UPS options are available if
19.5 feet (5.96 m) wide x 24.5 feet (7.49 m) desired. mm inches
deep x 9 feet (2.75 m) high.
Cooling Requirements: Gantry:
Exam Room: 13 feet (3.97 m) wide x 19.5
feet (5.96 m) deep. The cooling requirements do not include Width: 2,230 87.6
cooling for the room lighting, personnel or
Control Room: 11 feet (3.36 m) wide x 12.5 non-CT equipment present. Cooling Depth: 1,007 39.6
feet (3.82 m) deep. requirements are listed by subsystem to
allow planning for each room of the CT suite. Height: 1,887 74.1
Equipment Room: Not required. Power
Distribution Unit heat and noise output under Cooling requirements are given for minimum, Table:
all operating conditions with all available recommended and growth allowance
options is low enough to allow inclusion in scenarios. Width: 610 24
the patient scanning suite with the table and
• The minimum cooling figures assume
gantry. Length: 5,050 199
patient throughput of 3 patients per hour
Temperature and Humidity: and 75 scan rotations per patient. Height: 1,120 44
• The recommended cooling requirements
Exam and Control Rooms: 60°-75°F (15°- assume patient throughput limited by the Power Distribution Unit:
26°C) at 30%-60% relative humidity (non- tube cooling algorithm.
condensing). Width: 762 30
• The suite cooling can be sized for future
Equipment Room: If a separate equipment developments by using the growth Depth: 585 23
room is used to house the PDU, the allowable allowance figures. This cooling will
temperature range is 60°-84°F (15°-29°C) at accommodate more patients per hour Height: 1,270 50
30%-60% relative humidity (non-condensing). and/or potential future system
enhancements. Operator Console:
Temperature Rate of Change: 3°C/hour max.
Width: 1,238 48.7
Relative Humidity Rate of Change: 5%
Minimum Allowance Depth: 1,288 50.7
RH/hour max.
Subsystem Watts BTU/hr
Power Requirements: Gantry 5,440 18,600** Height: 1,257 49.5

The only facility input to the system is a 380 Table 200 700
to 480 V nominal, 3 phase Delta or Wye, PDU 1,000 3,400
50/60 Hz, 90 kVA service, 20 kVA average
Operator 2,165 7,400 Warranty
power; main disconnect to be located within
Console
5 feet (1.5 m) of the PDU. The facility must
also provide a protective disconnect device Optional Laser 800 2,730 The published Company warranty in effect
with low voltage, low energy local and multi- Camera on the date of shipment shall apply. The
point remote capability, in the line feeders to Company reserves the right to make
Recommended cooling values should not be changes.
the PDU.
used for calculating system input power
Complete, detailed specifications of all power requirements. All specifications are subject to change.
requirements are available upon request. For
** Recommended Allowance: 7,150/24,400
most installations, the Lightspeed16 CT Regulatory Compliance
Scanner System does not require any power Growth Allowance: 9,200/31,400
conditioning equipment to be used in This product is designed to comply with
conjunction with the PDU. Regulators are not applicable standards under the Radiation
recommended for use with this system. For Control for Health and Safety Act of 1968.
those sites with known large power line
transients, a suppresser filter for the system Laser alignment devices contained within this
computer and peripherals may be useful. In product are appropriately labeled according

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 18 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

to the requirements of the Center for Devices


and Radiological Health.

This product is a CE-compliant device which


satisfies regulations regarding Electro-
Magnetic Compatibility (EMC) and Electro-
Magnetic Interference (EMI), pursuant to IEC-
601.

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 19 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

* Option
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544
3384
Product Data Sheet – rev 3.1
Europe - July 2003
LightSpeed 16 GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Page 20 With Xtream Hong Kong - Fax: 852 2559 3588

CT Scanner System

Minimum Room Layout