Professional Documents
Culture Documents
r€(o8niz€d principtes
oD srlndardizarior e.tabtjshed in the D(is,on on prid.ipts tbr rhe
Devel0pfleflt 0f InlemalioDal Slardards, Guides and Recommendarions issueo -try tr'" woiio nuu" o.g"ntarion retrni.gi sai.re.j io r.-uo" 1r r r1 co--nr*.
.ll\aw
\ taaLEt
Desiqnation: D6667 - 14
5. Apparatus 6. Reagents
5.1 Funnce-An electric fumace held al a tempcmture 6.1 Puriy of Reagents-Reagent grade chemicals shall be
(1075 -F 25'C) suficient to pyrolyze the entire sample and used in tests. Unless otherwise indicated, it is intended thar all
oxidize sulfur to SO2. rcagents shall conlbrm to the specilications of the Committee
on Analytical Reagents of the American Chemical Society,6
5.2 Combustiort Tube-A quartz combustion tube con-
where such specifications are available. Other grades may be
structed to allow the direct injection of the sample into the
used, provided it is first ascertained that the reagent is oI
healed oxidation zone of the furnace. The combustion tube
suliciently high pu ty ro pennit its use withour lessening the
shall have side arms fbr the introduction of oxygen and canier
accumcy of the determination.
gas. The oxidation section shall be large enough (see Fig. l) to
ensure complete combustion of the sample (see IL3). Fig. I 6.2 Inert Gas-Argon or helium only, high purity grade
depicts a typical combustion tube. Other configurations are (that is, chromatography or zero grade),99.998 Vo min purity,
acceptable when precision is not degraded. nroisture 5 mg/kg max. (Warning or helium may be
-Argon
a compressed gas under high pressure (7.l)).
5.3 FIotN Co|t,ol The apparatus shall be equipped with
flow controllers capable of maintaining a constant supply of
" R?usent Che icdls, Anetua Ch?ntiLal Sd.ricr). Spe.ir.arir,,r. Aorerican
oxygen and carrier gas at the specified rates. Che'njcal Sociely. Washinglon. DC. i_or suggestions on the Esting of reagents nor
5.4 Drier Tube The apparatus shall be equippcd with a lisred by thc Anrerican Chenical Sociery. see Anular St.t .tatuts for lab.)rt1b^
Chenti&Ls. BDH Lrd.. Poole. Dorset. U.K.. and ihe Unite.l Stales phantocopeia
mechanism fbr the removal of water vapor lbrmed during dn.l Ndti.,tal font la^.U.S. Phamracopeial Convention. tnc. (USPC). Rockville.
sample combustion. This can be accomplished with a mem- MD.
450 mm
Copyight by ASTM lnt'l (all righls reserved); Fri Jan 2 7 05 : t 0:14 EST 201 7 .
Downloaded/prinred b)
valentyna K.avchuk (GiobuePetroleum Ltd.) pursuant lo License Agr€€ment. No fudh€r reproduciions authorized.
.dfr ooooz - ra
ARGON/OXYGB'I
CARRI€R LOOP
SEPTUM
FROM SAI.IPIE
II'ILET SYSIEM
FlG. 2 Example ot Orientation of Total Sulfur Analyzer and Gas or Liquid Sampling Vatve
Coplright by ASTM Infl (all rights reserved); Fri Jan 2? 05:10:34 EST20l7 3
Downloaded/pri ed by
Valentrna Kmvchuk (Globu.s-Petoleum Lld.) purslant !o License Agre€ment. No firther reproductions aurhorized.
{fl} ooeoz - ta
molecular sievcs. (Warning-Oxygen vigorously accelerates TABLE 2 Typical Operating Conditions
combustion and rnay be cornpressed gas under high pressure sample inlet system temperatlre 85 r 20'C
(7.1)) Sample injection system carrier gas 25-30 mumin
Furnace temperature 1075 t 25"C
6.4 C a I i b ru t i o n S t an d a
rds-Certifted cal ibration standards Furnace oxygen llow meter sefling 37t-450 mumirl
from commercial sources or calibmtion gases preparcd using Inlet oxygen flow meler setling 10 30 mumin
Inlet carrier flow meler setling 130 160 mumin
certilied permeation tube devices are required. Tabje I lists the Gas sample size 10 20 mL
sulfur source material and diluent matrices used during the LPG sample size 15 UL
interJaboratory study (Notes 2 and 3).
Nilm 2-Other sulfur sources and diluent matcrials may be used if TABLE 3 Typical Sulfur Calibration Ranges and Standard
precision and accuracy are not degraded.
Concentrations
Noir 3
Calibration slandards are typically re mixed and re certified
on a regular basis depending upon frequency of use and age. These Curve I Curve lll
calibration slandards may have a uscful life of about 6 to 12 months. Suliur (mg/kg) (mg/kg)
Sulfur Sulfur (mg/kg)
Blank Blank Btank
6.5 Quality Control (QC) Samples, preferably are porrions 5 10 50
of one or more gas or LP gas matcrials that are stable and 10 50 100
reprcsentative of the samples of interest. r00 200
7. Hazards
7.1 High temperature, flammable hydrocarbons, and gases type represenhtive of the samples to be analyzed (Note 4).
under high pressures occur in the test method. Use materials Tablc 3 is representative of typical ranges, but narower ranges
that are rated fbr containing these pressurized hydrocarbons in than those indicated may be used if desired. However, the
all sample containe$ and sample transfer apparatus. Exercise method precision using narrower mnges than those indicated
extra care when using flammable materials near the oxidative has not been determined. Ensure the standards used fbr
turnace. calibration bracket the concentrations of the samples being
analyzed.
8. Sampling
8.1 Obtain a sample in accordance with Practices F307, N(nri 4 The nunber of sland ds us€d per curve may vaty.
D1265, D3700, D5287, or GPA 2174 or cPA 2166. Analyze 10.2 With the sample valve in the load position, connect Lhe
samples as soon as possible after taking from bulk supplies to pressurized sample container to the sample valve of the sanrple
prevent loss of sulfur or contamination due to cxposure or inlet system.
contact with sample containers. 10.3 Obtain a quantitative measurement of the injected
8.2 If tbe sample is not used imntediately, then thoroughly material by filling the sample loop of the sample valve system
mix it in its container prior to taking a test specinen. The use for the matrix being analyzed (see Table 2) (Notes 5 and 6).
of segregated or specially treated sample containers can help NorL 5-lnjecdon of a constant or similar sample size lbr all malerials
reduce sample cross-contamination and improve sample sta- analyzed in a selected operaling range plomotes consistent combusdon
bility. condition. and mr) simplify resuh calculdtion!.
Nonr 6
An automatic sample transfer and injection device may be
9. Preparation of Apparatus used.
9. I Assemble and check the apparatus for leaks according to 10.3.1 Flush the sample loop with suficient calibrant to
manufacturer's insructions. assurc that the material to be injected is representative.
10.3.2 For LPG samples, if bubbles are present in the
9.2 Typical appamtus adjustments and conditions are listed
viewable portion of the liquid column, flush the sample Ioop to
in Table 2.
inlroduce a new liquid full sample porrion.
9.3 Adjust instrument sensitivity and baseline stabiliry and
10.4 Slart the analyzer and inject the calibration material
perform instrument-blanking procedures tbllowing manufac-
according to the manufacturer's instructions.
turer's guidelines.
Coplright by ASTM Int'l (allrights reserved); Fri Jan 27 05:10:34 EST 20t7
Downloaded./printed by
Valenlyna Kmvchuk (Globus-Petroleum Lid.) pursuanl to License Agreement. No limher reproductions authorized.
.ffi oosez - r+
10.5 Calibrate the insrrumcnr using one 01-the lbllowing GXd
lechniques. Sulfur. Ds/kr: (t)
|0.5.1 Multi-point Calibratiott:
where:
10.5.1.1 When the apparatus f-eatures an intcrnal self-
calibratio|r routine, analyze the calibration srandards and blank ./ = density ol standard mixture, g/ml,
three times using the procedures described in 10.2 - 10.4. s = density of sample, g/ml, and
10.5.1.2 Calibrate the analyzer according to the manufac- G = sulfur found in test specimen, mg/kg.
turer's instructions to yield sulfur concentration (see Section 12.2 For analyzers calibrated using a one point calibration,
l4). This curve is typically linear and system perfonnanca shall calculate the caiibration factor, (10.5.2.4).
be checked at ieast once per day, each day of use. (Nore 7).
A(
Nort 7 Othef calibrntioD cu e
K:
-- -. (2)
Iechniques may
whenbe uscd Mr x &.9
accuracy and precision are no( degnded. The frequency ofcalibration may
bc de(ermined by the use of quality conhol charts or olher quality
assurancc/qualily coDtrol techniqu€s.
| 0.5.2 One- po irtt Cal ib rat ion : (3)
10.5.2.1 Utilize a calibrarion standard (6.4) wirh a sulf'uf
contenl close to that ol'the samples to bc analyzed (125 70 whcre:
max.). A. = integrated detector rcsponse for calibration standard,
10.5.2.2 Follow the ilstrument n]anuf'acturer's instructions in counts, and
to establish an instrument zero (instrument biank) by conduat- Mc = mass of calibration standard injecred, in milligrams.
ing an analysis run without injection of the calibntion stan- either measured directly or calculated fiom measured
dard. volume injected and density.
10.5.2.3 Perlbrrn measurements of the calibration standard a M(:V\D( (4)
minimum of three times.
10.5.2.4 Calculate a calibration l'actor K, in couuts per whgre:
nanogram of sulfur (counts/ng S) as described in 12.2. Dc = density of calibration standard at measuremcnt
temperature, g/ml-,
ll, Procedure Vc = volume of calibration standard injected, !L,
11.1 Obtain a test specimen using the procedure described Sca = su;1'u..nn,ent of calibration standard, ml-/kg, and
in Section ti. Typically the sulfur concentration in thc test Scr = sulfur content oi calibration standard, mg/L.
specimen is less than the concantmtion of the highest standard 12.2.1 Calculate the averagc of the calibration factor (rl
and grealer than thc concentration of the lowest standard used and check thal thc standard devialion is within the tolerance
in the calibration. accepted. This calibration factor shali be esrablished evcry day.
ll.2 Measure the response for the test specimen usirg one I2.2.2 Calculate the sulfur contenr, Sj of the sample. in
of the procedures described in I0.2 - 10.4. mg/kg. using the following cquation:
ll.3 Inspcct the combustion tube and olhcr flow path A
(5)
components to verify complete oxidation of the test specimen. ' MxkrtZ
11.3.1 Reduce the ratc of injection or thc sample size, or
both, ol the specimen into the furnace whcn coke or sooting is
observed.
11.4 Cleanirtg and Re-calibration-Clean any coked or " yx(xF"
sooted parts according to the Inanufacturer's instructions. Afier whcre:
any cleaning oi adjustment, assemble and chcck the apparatus
for leaks. Repeat instrumcnt calibration prior to reanalysis
K - calibration lactol in counts per nanogram o1'sulf'ur, and
o1' M = mass 01'test specimen solution injected, in milligrams,
the test specimen. either measurcd directly or calculated from measured
I 1.5 To obtain one result, measure each test specimen three volume iniected and density.
times and calculate the average detector response.
14:V>\D (7)
ll.6 Density values needed for calculations are to be
measured using Test Methods D1070 or equivalent, at the
temperature at which the sample was tested (Note 8).
Nof! 8 When sample natdx composilions at€ known, other tech
niques may be used to derive samplc density, provided accuracy and
precision arc not dcgraded.
12. Calculation
12.1 For analyzers calibrated using an internal self-
calibration, calculate the sulfur content in the test specimen as
ibllows:
Coplrighr by ASTM Infl(all righrs reserv€d); Fri Jan 27 05:10:34 EST 201?
Do$nloaded/prinred b)
Valentyna Kmvcbuk (clobus-Petroleurn Lld.) pursuanl ro License Agreemenl. No 1irrlher reprcducrions authorized.
,dfr ooooz
D6667 - tc
14
APPENDIXES
(Nonmandatory Information)
X | . i Confirm the performance of the instrument or the test cause(s). The results of this investigation may, but not
procedure by analyzing quality control (QC) sample(s). necessarily, result in instrument re-calibration.
No r Xl.l In the absence of explicit requirements given in the tesl
X1.2 Prior to monitoring the measurement process, the user method, Xl.4 provides guidance on QC testing frequency.
of the test method needs to determine the average value and
control_limits of the QC sample. See Practice D6299 and Xl.4 The liequency of QC testing is dependent on rhe
MNL7.' criticality of thc quality being measured, the demonstrated
stability of the testing process, and customer requirements.
X L3 Record the QC results and analyze by control charts or
Generally, a QC sample should be analyzed each testing day
other statistically equivalent techniques to ascertain the statis-
with routine samples. The QC frequency should be increased if
tical control status of the total testing process. See practice
a large number of samples are routinely analyzed. However,
D6299 and MNL7.7 Investigate any out-of-control data for root
when it is demonstrated that the testing is under statistical
control, the QC testing frequency may be reduced. The QC
1
MNL 1. lt4ulual oi P,esentuti t ul Data CotLrd Chatl Anub.sis. 6th Ft' sample lesting precision should be periodically checked against
Sectior 3. ASTM International- the ASTM method precision to ensure data quality. See
?
Cop)'right by ASTM Inl'l (all rights reserved); Fri Jan 2 7 05 : t 0:34 EST 201
6
Downloadeaypri.led by
Vale.tyna Kmvchuk (clobus-Pelroleum Ltd.) pusuant to License Agreement. No funher reproductions authorized.
,(ffi ooooz-tc
Practice D6299 and MNL7.7 homogenous and stable under the anticipated stomge condi-
tl0ns.
X1.5 It is recommended that, if possible, the type of QC
sample that is rcguiarly tesled be representative of the material
Xl.6 See Practice D6299 and MNL77 for l'urrher guidanca
routinely analyzed. An ample supply of QC sample materiai
on QC and control charting techniques.
should be available for the intended period of use, and must be
X2.l Prior to sample injection, fully stabilize and ready the X2.8 For sample cylinders containing liquefied petroleum
apparatus fbr analysis according to the manufacturers' instruc- gases, use sufficient pressure (typically greater than 200 psig
[tons. container pressure) to allow the transfer of sample to the
sample inlet system withour the formation of bubbles in the
X2.2 Valves in the sample inlet system typically remain in transfer tubing and sight glass.
the load position except during sample analysis.
X2.9 Use standard gas I'low conditions for your analyzer.
X2.3 Once the sample valve has been filled and allowed to See Section 9of fte test method.
equilibrate, accomplish sample injection by executing a prompt
and full rotation to the sample valve to the iiject position. X2.10 The use of a filtering device prior to sample inrro-
duction is strongly recommended. This can greatly extend
X2.4 Leave the sampling valve in the inject posirion until valve service and prevent transfer line (tubing) blockage.
analysis has completed (instrument has returned to baseline
and integration has finished.) X2. I I The use of low or iron-free alloys and/or inert treated
materials can enhance the analysis for low level sulfur concen-
X2.5 The length o1' time required fbr detector response will tration.
depend upon the type of pyrotube utilized, canier gas, sample
size or sulfur concentration, or both- From 20 s to I min can be X2.12 The use of a strip-chart recorder or a software peak
typical (see Fig. X2.1). display can aid in the set up and normal operation of the sample
inlet system.
X2.6 Sample inlet system can;er gas flows can be used to
manipulate sample combustion and detection characteristics. X2.13 Quality analytical results are best obtained when
However, excessive carrier flow rates (grcater than 30 ml-/min) calibration materials match the mafix of the samples analyzed.
can cause incomplete sample combustion (sooting). Injection of solvent (room temperature) liquids as calibrants or
samples is discouraged and can cause severe coking (sooting)
X2.7 Allow time fbr return of signal to baseline between of sample flow path components and yield poor analytical
injections. results.
X3.l Routinely inspecr rransf-er lines and fittings used to sample fansfer appamtus bcing employed.
connect the sample container to the sample inlet system for
leaks and stress weakening that can be caused by frequent X3.5 Use slow to moderate purge rates when filling the
bending, age, or use. sample loops during the sample-loading phase of the sampling
procedure.
X3.2 Inspect Quick-Connect fittings and other sample trans-
X3.6 After sample loop loading, allow pressure within rhe
Ier and venting apparatus routinely for leaks and proper
loop and vent system to equilibmte. Use ofa securs gas-bubble
opemtion.
monito ng device that vents to an arca free of potenlial
X3.3 Vent waste gases from the sample inlet system to an ignition sources can provide visual confirmation and aicl in
area away from the instrument operating environment and ensuring the injection of consistent sample sizes.
potential ignition sources. X3.7 After sampiing is complete, a contrclled venting or
X3.4 Pressures employed to maintain contents of sampling relief of pressure of the sample tmnsfer line is advised.
cylinders and to enable sample transfer are typically less than X3.8 After analysis, wirh rhe sample valves in the load
750 psig. Do not exceed the general overall pressure rating position, an inert gas connected to the sample inlet system can
(900 psig) of the sample inlct sysrem or the capaciry 01'rhe be utiliz"ed to purge sample residue.
Coplrighr by ASTM Int'l (all rights reserved); Fri Jan 27 05: I 0:34 EST 20 | 7
7
Downloaded/printed by
valentyna Kmvchuk (clobusPerroleum L1d.) pusuant lo License Agreement. No funh€r reproducrions authorized.
{ft ooooz-tc
.o
c
.9
an
SUMMARY OF CHANGES
Subcommittee D02 has identifed the location of selected chanses to this standard since the last issue
tD6667 - l0) that may impacr the use of this srandard. lApproved Oicr. 1,2014.)
(.1) Updated subsection 1.1. (4) Removed former footnote 6.
(2) Added footnote 3; renumbered subsequent footnotes. (5) Update Table 3 and Table 4.
(3) Updated Section 15, Precision and Bias, to reflect data from (6) Updated. X2.11 .
second ILS.
ASTM lntenational lakes no position respecting lhe validily of any patent rights assefted in connection with any item mentionect
in this stanckrd- Users of thb standard arc exptessly advised that ctetermination of the valdity ol any such patent ights, and the risk
of infringement ol such ghts, are entirely their own responsibitv
This standarcl is subEct to rcvision at any tinp by the rc$)onsible technical comn't6e aN nust be reviewed every live yeag anc!
if not rcvised, eilher reapprcved or withdnwn. Your comnents arc invited eithe, for revisjon of this standarcl ot lor adctitional danckrds
and should be adclressed to aSTM Intemational HeaclquadeE, Your comnents witt eceive careful consideration at a ngetino of the
responsible technical commiftee, whbh you nay aftencl lf you leel that you comnents have not rcceivec! a kit heeing yo; should
nake your views known to the ASTM Commiftee on Standards, al the adchess shown betow
Thb standard is copyngned by ASTM lnternational, 1(n Baff tkrbor Drive, PO Box C700, West Canshohocken, pA lg4a,-2gsg.
Uniled Stales. lndividual repints (single or multipte copies) of this standard may be obtained by contacting ASTM at the above
acntess ot at 610'832'9585 (phone), 610-832-9555 (lax), ot seNice@astm.ory @-nail); or through the AsrM website
(tttltw.asln.org). Pemission rights to photocopy the stanclatd may also be securcd from the Copytight Cteannce Centet, 222
Rosewood Dtive, Danverc, MA 01923, Tel: (97q 64&2600; hvr {ww.copyight.con/