Pratap Kollu1, Seok Soo Yoon 2*, Kim Gun Woo1,2, C. S. Angani1, and Cheol Gi Kim1*
1
Department of Materials Science and Engineering, Chungnam National University, Daejeon 305-764, Korea
2
Department of Physics, Andong National University, Kyoungsangbuk-do 760-749, Korea
(Received 6 September 2010, Received in final form 28 September 2010, Accepted 29 September 2010)
In our present work, we developed a GMI (giant magnetoimpedance) sensor system to detect magnetic fields in
the milli gauss range based on the asymmetric magnetoimpedance (AGMI) effect in Co-based amorphous
ribbon with self bias field produced by field-annealing in open air. The system comprises magnetoimpedance
sensor probe, signal conditioning circuits, A/D converter, USB controller, notebook computer, and program for
measurement and display. Sensor probe was constructed by wire-bonding the cobalt based amorphous ribbon
with dimensions 10 mm × 1 mm × 20 µm on a printed circuit board. Negative feedback was used to remove the
hysteresis and temperature dependence and to increase the linearity of the system. Sensitivity of the milli gauss
meter was 0.3 V/Oe and the magnetic field resolution and environmental noise level were less than 0.01 Oe and
2 mOe, respectively, in an unshielded room.
Keywords : milli guass meter, amorphous soft magnetic ribbon, magnetic sensor, magnetoimpedance, negative feed-
back
Fig. 6. GMI profile of the sensor with negative feedback. Fig. 7. Resolution check by decreasing the magnetic field by
steps of 0.01 Oe.
The Vo of the gauss meter system follows the relation [3] A complete system for detecting the magnetic fields in
milligauss range was successfully developed using self
A biased GMI. It was successfully developed and its per-
V o = ---------------- H . (3)
1 + AS formance was checked. The sensitivity was improved to
Since AS = 103.8 in the present gauss meter system, we ~0.3 Oe/V with negative feedback system. Electrical line
can get the approximation: noise was removed by using a 12V battery and the 100
kHz component was eliminated by the end-filter. The
1 resolution of the gauss meter was 1 µT.
V o ≈ --- H, if AS >> 1. (4)
S
The slope of 0.293 V/Oe in Fig. 6 shows good agree- Acknowledgements
ment with the 1/S = 0.289. This means that the Vo with
This research was financially supported by a grant to
negative feedback linearly depends on H and the sensi-
MEMS Research Center for National Defense funded by
tivity depends only on the coil parameters, such as N, l,
Defense Acquisition Program Administration.
and Rf . Since these coil parameters have negligible temper-
ature dependence in normal operation range of gauss meter,
we can avoid temperature calibration problem. References
Other important performance characteristics of gauss [1] T. Morikawa, Y. Nishibe, H. Yamadera, Y. Nanomura, M.
meter system are resolution and noise level. In order to Takeuchi, and Y. Taga, IEEE Trans. Magn. 32, 5 (1996).
test these performances, change in the output voltage Vo is [2] P. Kollu, J. Lin, K. W. Kim, D. Y. Kim, S. S. Yoon, and
monitored by altering the magnetic field in uniform steps C. G. Kim, J. Appl. Phy. A 90, 533 (2008).
by using HP 6632A programmable power supply. Fig. 7 [3] S. S. Yoon, P. Kollu, K. W. Kim, Y. Cha, D. Y. Kim, and
shows the change of Vo for the 0.01 Oe step which is the C. G. Kim, IEEE Trans. on Magn. 45, 2727 (2009).
minimum controllable step in our laboratory. The change [4] P. Delooze P. Delooze, L. V. Panina, D. J. Mapps, K.
of 0.01 Oe (1 μT) in the magnetic field is clearly detect- Ueno, and H. Sano, IEEE Trans. Magn. 39, 3307 (2003).
− 198 − Prototype Milli Gauss Meter Using Giant Magnetoimpedance Effect in Self Biased Amorphous Ribbon − Pratap Kollu et al.
[5] C. G. Kim, K. J. Jang, H. C. Kim, and S. S. Yoon, J. machinery 25, 1 (2010). http://d.wanfangdata.com.cn/
Appl. Phys. 85, 5447 (1999). Periodical_bjjxgyxyxb201001013.aspx
[6] K. Mohri, T. Uchiyama, and L. V. Panina, Sens Actuators [10] S. Yabukami, H. Mawatari, N. Horikoshi, Y. Murayama,
A 59, 1 (1997). T. Ozawa, and K. Ishiyama, J. Magn. Magn. Mater. 290,
[7] T. Kanno, K. Mohri, T. Yagi, T. Uchiyama, and L. P. 1318 (2005).
Shen, IEEE Trans. Magn. 22, 3358 (1997). [11] http://www.aichi-mi.com/
[8] H. Hauser, R. Steindl, C. Hausleitner, A. Pohl, and J. [12] C. G. Kim, K. J. Jang, D. Y. Kim, and S. S. Yoon, Appl.
Nicolics, IEEE Trans on Inst. and Meas. 49, 648 (2000). Phys. Lett. 75, 2114 (1999).
[9] G. P. Jian and T. G. Qing, Journal of Beijing Institute of [13] www.smekorea.com