Professional Documents
Culture Documents
RIGAKU CORPORATION
ME0B-13 20200515
CONTENTS
CONTENTS
2. ANALYSIS .................................................................................................. 14
2.1 USE OF SUITABLE ANALYSIS TYPE ............................................................................... 14
2.2 EZ ANALYSIS .................................................................................................................... 15
a. Configuration of Screen for EZ Analysis .......................................................................... 16
b. Procedure for Sample Setting in EZ Analysis .................................................................. 19
c. Detailed Description of Each Area ................................................................................... 22
2.3 ANALYSIS ......................................................................................................................... 64
a. Before Analysis ................................................................................................................ 64
b. EZ Scan ........................................................................................................................... 86
c. SQX Analysis (Option) ..................................................................................................... 90
d. Quantitative Analysis ....................................................................................................... 93
e. Analysis Control ............................................................................................................... 95
f. Control ID ........................................................................................................................ 103
g. Sample ID Table............................................................................................................. 106
h. Cycle Repeat Analysis ................................................................................................... 108
i. Fixed Time/Period Analysis .............................................................................................. 111
j. LIMS Analysis Command Interface (Option) ................................................................... 112
GLOSSARY.................................................................................................... 796
Tool Bar
Work Area
Flow Bar
Area Description
Work Area Each program is displayed and operated.
Menu Bar Area A menu used in a running program is displayed.
Tool buttons of a running program and the ZSX Guidance main
Tool Bar Area
menu are displayed.
The positioning of a screen started by a qualitative or quantitative
Flow Bar
application program is displayed.
Expression Meaning
Mark Meaning
Means a precaution.
Means a note.
Program Menu
The ZSX Guidance menu consists of the tool bar shown below.
The normal menu is shown above, while the small menu is shown below.
In the case of the small menu, when the mouse cursor is placed on a button, the Help
function displays its menu item name.
In a description for an operating procedure, the operation of a button on the program menu
is expressed as follows.
“Click Utility on the menu.”
The user who has logged on to “ZSX Guidance” is displayed on the tool bar.
When this user name is clicked, a menu to log off from “ZSX Guidance” is displayed.
Flow Bar
EZ Analysis
Analysis operations can be done easily on an analysis screen on which display screens
necessary for daily routine analyses have been put together. For details, see “2.2 EZ
ANALYSIS”.
Analysis
This menu controls analyses. All measurements including the standard sample
measurement are possible using this menu. For details, see “2.3 ANALYSIS”.
Data Processing
This menu carries out data processing for qualitative analysis results and quantitative
analysis results. This menu has submenus of Qual Result (3.3) (including the functions
of Qualitative Data Handling (3.3.1) and SQX Calculation (3.3.2)), Result Display (4.4)
and Quantitative Simulation (4.5). See a section written in parentheses after each
submenu item.
Qualitative Application
This menu creates qualitative analysis applications. Applications for the qualitative
analysis and SQX analysis are created from this menu. This menu starts the flow bar.
For details, see “3.2 CREATION OF QUALITATIVE APPLICATION”.
Quantitative Application
This menu creates quantitative analysis applications. Applications for the empirical
calibration method and FP quantification method are created from this menu. This
menu starts the flow bar. For details, see “4.2 CREATION OF QUANTITATIVE
APPLICATION”.
Utility
This menu controls information for analyses such as the compound registration and
sensitivity libraries. This menu has submenus of Atomic Symbols, Compound Table,
Sample ID Table, Universal Standard Samples, Sensitivity Library (optional), Material
Standard Tables (optional), Program Operation and System Management. See section
3.4 for Sensitivity Library and “6.1 UTILITY” for the other submenu items.
Maintenance
This menu controls information concerning the condition of the system such as the
spectrometer check and maintenance record. This menu has submenus of Check List,
Self Diagnostics, PC Maintenance, Error Log, Record for Spectrometer Status Check,
Record for Self Diagnostics, Record for Check Analysis, Record for Drift Correction,
Record for PHA Adjustment and Test Measurement. For details, see “6.2
MAINTENANCE”.
Startup / Shutdown
This menu starts up and shuts down the system. This menu has submenus of System
Startup, Tube/Atmosphere Change, System Shutdown and Status Display.
Options that the system does not have are not displayed on the screen.
When the ZSX Guidance is terminated, the measuring control programs and ZMonitor are
also terminated automatically.
The measuring control program unit consists of more than one program. The names of
programs in operation can be displayed using the following procedure:
The X-ray fluorescence analysis is basically a surface analysis. So make sure that
sample surfaces are free from contamination.
Metal Sample
Powder Sample
Crushing
Samples are crushed to make sample grains small and uniform. A vibrating mill is
generally employed.
Pressing
A fixed quantity of powder sample is pressed intact or after binder or the like for sample
forming is added, in order to form a pelletlike sample. Before pressing, sample is
generally filled into a ring or cup as a support to retain the sample.
In the case of a powder sample, especially in the case of ore and material for the ceramic
industry, X-ray intensities may not correlate with concentrations because of grain size
effects or mineral effects. In such a case, the sample is vitrified to make it uniform before
an analysis. This sample is called a glass fusion bead. Li2B4O7 (or other materials) is
generally used as flux for vitrification.
Casting
Metal powder such as chips is re-fused to make a block of metal sample. A centrifugal
casting machine is generally employed. After re-fusing, samples are treated like solid
metal samples.
Liquid sample and powder sample that is not pressed are usually measured in the
helium atmosphere. Sample is put in a dedicated cell and set in a sample holder.
When PP is used as the window material of a liquid sample holder for the instrument of
the high output (4 kW) type, the window material may be damaged. Avoid taking a long
time (10 minutes or more) for a measurement.
For the details of the handling of the sample holders, see the instruction manuals of the
liquid and powder sample holders.
Film Sample
Sample Holder
When the instrument is the ZSX Primus, sample size that can be housed in a sample
holder is up to 51 mm dia. 40 mm. Set a sample to a sample holder with the analysis
side facing down. If the sample moves during a measurement, fix it using a ring. In the
case of a film sample, put a hollow cup on it.
When the instrument is the Primus III+, sample size that can be housed in a sample
holder is up to 51 mm dia. 30 mm. Set a sample to a sample holder with the analysis
side facing up.
When the instrument is the ZSX Primus IV, sample size that can be housed in a sample
holder is up to 52 mm dia. 30 mm. Set a sample to a sample holder with the analysis
side facing up.
Sample size to be set to sample holder is 52mmDia. x 30mmH maximum in case of ZSX
Primus IVi. Set the sample surface to be analyzed downward. If it seems to move in the
holder during measurement, fix it with a ring. Put a hollow cup over the sample in case
of film shape sample
A metal sample, pressed powder sample or glass bead sample is set in the sample holder
directly. Liquid sample is poured into a dedicated cell and then loaded into the sample
holder.
When a liquid sample cell is used for the ZSX Primus/Primus IVi, cut a film so that the film
or retainer ring does not protrude from the sample holder. Otherwise, the sample holder
may be stuck in the instrument.
Select a sample mask suitable for the sample diameter. Though a sample mask diameter
is generally the same as a measurement diameter, they can be different when the system
has a diaphragm. A diaphragm diameter should be the same as or smaller than the
sample mask diameter.
2. ANALYSIS
2.1 USE OF SUITABLE ANALYSIS TYPE
The ZSX Guidance has the following two types of analyses. Use a suitable one for your
purpose.
EZ Analysis
Analysis operations can be done easily on an analysis screen on which display screens
necessary for daily routine analyses have been put together.
Analysis
It is possible to make analyses with complicated sequences such as the repeated
analyses.
Since EZ Analysis has been designed for daily routine analyses, the following functions
have been omitted. When any of these functions is necessary, use the “Analysis” program.
2.2 EZ ANALYSIS
In EZ Analysis, operational procedures are simple and easy to understand thanks to
graphic display, and it is possible to carry out easily analyses, analysis reservations and
checking of analyzed results.
EZ Analysis consists of a reservation screen and a progress screen, and the screen can be
switched by clicking the tab for each screen. Each screen is divided into areas according
to functions.
The sample ID table area is displayed only when a sample ID table to be displayed has been
registered using the [Option] setting of the EZ Scan program.
The sizes of the areas on the screen can be changed freely by dragging each boundary line.
To return the sizes of display areas on the screen to the default sizes, follow the
procedure described below:
When there is no reserved or analyzed sample ID on the startup of EZ Analysis, the system
makes the sample position A-1 selected.
When a sample position is selected, the sample type and application selected previously
are automatically copied.
Reservations for the successive analyses of samples using the same application can be
made only by selecting sample positions continually.
3. To change the sample position, drag and drop the image of a sample position in the
ASC display area.
5. When EZ Scan has been selected for the analysis type, select a sample type,
measuring range, diameter and measuring time. If necessary, set detailed
information using the Advanced... button.
6. In the case of an analysis type other than EZ Scan, select an application that has
been created beforehand.
To carry out the sample film setting for EZ Scan, use the sample ID setting screen.
8. To change the analysis order, right-click a sample ID and select [Priority analysis].
That sample ID is set at the top of reserved ones.
9. When samples have been set, click Start to start analyses.
10.When the analyses have been completed, check analyzed results in the analysis
result display area.
By selecting a measured sample ID in the ASC display area on the Reserve tab
screen, analyzed results can be checked.
11.To check or reanalyze qualitative analysis data, check an SQX calculation or make a
calculation again, right-click a sample ID on the reservation and progress display
area or analyzed result screen or the position of an analyzed sample in the ASC
display area, and select [SQX Calculation] or [Qual Data Handling].
Sample positions for sample IDs are displayed graphically on a screen resembling the
ASC.
Sample positions can be set here.
Progress such as during an analysis or during transportation can be understood at a
glance.
The image of the position of a selected sample ID is enlarged.
A sample type icon or a sample name can be selected and displayed in the position of a
preset sample ID.
The ASC display area displays the positions of reserved sample IDs.
Before analyses, put sample holders on the ASC of the spectrometer in accordance with
reserved sample IDs.
The number of samples displayed in the ASC display area varies depending on the number
of trays that the spectrometer has. Up to 12 samples can be loaded on one tray.
Sample Type
Sample Information
A calculation Program
An analysis
error has occurred operation
setting is
during an reservation
incorrect.
analysis. position.
To display sample type icons in the ASC display area, make the following setting:
To display sample names in the ASC display area, make the following setting:
When setting has been made such that sample names are displayed on samples, a sample
name can be edited in the ASC display area by double-clicking a selected sample position.
Set a sample position by specifying a number for the X-direction and an alphabetic
character for the Y-direction, as shown below:
In the case of the ASC48, the range allowed for the sample position varies depending on
the number of trays (up to four). If the number of trays is one, the range is from A-1 to B-6
and up to 12 samples can be set.
In the case of the ACS96 (option), up to 96 samples can be analyzed successively.
The ASC54/60 applies only to ZSX Primus IVi.
1. In the ASC display area, drag the image of a sample ID the position of which is to be
changed.
2. Drop it in a destination sample position.
Copying
When a sample position where a sample ID has not been set is selected in the ASC
display area in EZ Analysis, the sample ID selected previously is automatically copied.
1. Select a sample ID image to be copied or an ID line in the progress display area.
2. Select a sample position where a sample ID has not been set in the ASC display area.
Reservations for the successive analyses of samples using the same application can be
made only by selecting sample positions continually.
A sample ID for which an analysis has been complete as well as an ID in the preset table
can be selected for the sample ID copy function. Thus, an analyzed ID on the progress
display tab can also be copied.
1. Select a sample ID image to be copied or a sample ID on the progress display tab.
2. Right-click the mouse and select [Copy].
3. Right-click a sample position to which to paste in the ASC display area.
4. Select [Paste].
1. Select a sample ID image to be cut or a sample ID on the progress display tab.
2. Right-click the mouse and select [Cut].
A sample ID that has been cut can be pasted in the same sample position or pasted in any
position by right-clicking a sample position for pasting in the ASC display area and
selecting [Paste].
All sample IDs, analyzed sample IDs or specified sample ID can be cleared. Source
sample IDs remain in the preset table after the copying operation, while they do not
remain in the table after the cutting operation.
1. Drag the image of a sample ID to be cleared in the ASC display area.
2. Drop it in an area outside the ASC panel screen.
In EZ Analysis, the analysis order of an arbitrary sample ID can be set at the top or an
arbitrary sequence number.
Reserve
1. Right-click the image of a sample ID the analysis order of which is to be changed or a
sample ID on the progress display tab.
2. Select [Priority analysis].
Progress
1. To change the analysis order, click the up- and down-arrow buttons on the right of
the sample ID.
<Progress screen>
1. Click the sample ID to change the order.
2. Change the order by pressing right side Up/Down button of sample ID.
3. As the Status changes to Reserved, click Analyzet to go to <Reserved>.
To change the order of the sample placed in the inlet sample chamber during
measurement;
1. Select the priority sample by clicking its sample ID (or figure).
2. Select [Priority Analysis] by right click.
It takes some time to change the sample, as the atmosphere of the inlet sample chamber
needs to be restored to atmospheric pressure.
When the set contents of any sample ID are incorrect, the warning mark appears on the
image of the sample and an analysis is impossible unless a correction is made.
When the mouse cursor is put in the sample position, the contents of a problem are
displayed. Make a correction according to the displayed contents.
When a problem arises and an analyzed result cannot be calculated after an analysis is
completed, such as when library sensitivity has not been registered, the error mark
appears.
When the mouse cursor is put in the sample position, the contents of a calculation error
are displayed. Make a correction according to the displayed contents, and make an
analysis again.
The status of a set ID line changes as shown below until it is deleted from the preset
table after an analysis is complete. This status is shown by the description and color of
the status item of a sample ID and the color of a sample position in the ASC display area.
After Analysis
Checking of analysis results is possible on the Reserve tab screen or using Result
Display.
1. Right-click the image or analyzed result of a sample ID for which an analysis is
completed or a measured sample ID on the Progress tab screen.
2. Select [Qual Data Handling].
1. Right-click the image or analyzed result of a sample ID for which an analysis is
completed or a measured sample ID on the Progress tab screen.
2. Select [SQX Calculation].
This area is used to make setting for a sample ID selected in the ASC display area or
sample ID setting area.
Select an analysis type and an application that has been created beforehand.
In the case of EZ Scan, detailed setting can also be made here.
Like the progress display area, this area can also be used to edit memoranda such as the
sample names and operator names of sample IDs and output file names.
It is possible to choose whether or not to set a file name for EZ Scan and the qualitative
analysis using [Option] of the EZ Scan program.
Sample ID
At the time of an analysis, the Progress tab screen is used to check identification
information such as an analysis type, application name and sample name and settings
such as the result output.
Operation in the sample ID setting area is impossible when no sample ID has been
selected and for a sample ID for which an analysis is completed.
The setting procedure for EZ Scan is described below. When more than one sample needs
to be analyzed, repeat the procedure. The file name of EZ Scan becomes the same as the
sample name by default. When the same sample name as an already saved file name is
set, a subdivision number of “_1” to “_999” is automatically added to the end of the file
name.
When a sample name is not entered, registration is conducted using a name that consists
of a year (four digits of the Christian era), month (two digits), day (two digits) and time (two
digits of an hour and two digits of a minute).
Due to the software specifications, the measurement range may be indicated as B-Cm or
F-Cm
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of sample
film name is made X-ray absorption correction by the film itself and the impurity correction
in the film at a same time, therefore, the impurities setting for the film which used to be done
per measurement diameter are not needed. Refer to the instruction manual “PROCEDURE
FOR ADDITION OF SAMPLE FILM INFORMATION” regarding details of sample film. (This
is the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
As a ZSX Primus IV/IVi‘s particular function, the setting up of fixed angle measurement can
be made on the screen of EZ Scan setting. It is possible to make measurement accuracies
better for objective elements. The best suited analysis line is selected automatically during
the measurement. (This is the function of ZSX Primus IV/IVi only, and applicable to the
equipment shipped newly.)
1. Select <Fixed angle elem.> and then click … on the right side.
2. As the fixed angle measurement screen (periodic table) comes out, click the objective
elements.
3. Set a common measuring time for the objective elements. The default values is 2 seconds,
i.e. peak measuring time: 2sec., background measuring time 2sec. x 2 points
(Background subtraction is curried out by 2 points method). Click OK after setting.
Sample Type
Sample types that can be selected for EZ Scan are described below:
Measuring Range
The measuring range <B-Cm> can be selected when the system has a crystal with which
all elements from B to Cm can be measured.
Diameter
Selection can be made from the following measuring diameters by the diaphragm
changer.
ZSX Primus, ZSX Primus IV/IVi
Selection of measuring diameter
Auto,0.5,1,10,20,30,35mm
10,20,30,35mm
When <Auto> is selected for the measuring diameter, the sample mask diameter is
automatically checked using X-ray measurement and measurement is carried out with the
optimum measuring diameter. However, when the liquid has been selected for the sample
type and the spectrometer has the helium flush mechanism and sealing option, the
diameter is fixed to 30mm because Helium/Vacuum is set for the atmosphere.
Measuring Time
Time Description
With this setting, analysis precision is lower but a
Short
result can be obtained more quickly.
(This is the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped
newly.)
When Advanced... is clicked on the EZ Analysis screen, the setting screen for each
sample type appears and information can be set according to a sample.
It is possible to set the sample film correction on the sample ID setting screen.
The sample film correction is used to correct X-ray absorption by a sample film. When the
impurity correction has been set for a selected sample film, the default impurity correction is
automatically set.
If the analysis line is affected by higher order line interference, the selection of <Add MC's
to eliminate high order lines> will automatically use the optimum optical condition data with
the least interference for SQX calculation. (This is the function of ZSX Primus IV/IVi only,
and applicable to the equipment shipped newly.) Details please refer to the operation
manual “SQX Analysis Considering Influence of Hight-Order lines”.
The set contents are the same for powder, polymer and a liquid.
When binder is set, analyzed values are calculated in the quantification calculation with the
effect of the binder taken into consideration. Set components in a binder compound
beforehand using the “Compound Table” program.
When the system has the scattered ray FP method option, <Estimate> for a balance can be
selected, and Rh-K Compton scattering rays and Thomson scattering rays are measured.
Set 20mm for the measuring diameter. (For ZSX Primus IV/ IVi, 20mm and 30mm can be
selected depending on the specifications.)
It is possible to set the sample film correction on the sample ID setting screen.
1. Select a component type from <Oxide>. In fusion bead samples, each component is
contained as an oxide.
2. Specify a balance. When a balance is to be specified, select a compound or an element
for the balance. The optional <Loss on ignition> can also be specified for the balance.
3. When an optional matching library is to be used, select <Search> or <Specify>. When
<Specify> is selected, select matching library data registered beforehand.
4. To make the impurity correction, select <Use>, impurity data registered beforehand
and a correction method from <Concentration> and <Intensity>.
5. As for the sample size, set sample weight, diameter, area, etc. when a sample
thickness affects analyzed values.
6. Select the type of flux used to prepare a fusion bead, and enter sample weight and
flux weight.
7. When oxidizer has been used, select its type and enter oxidizer weight.
The setting items of the flux, sample weight, flux weight and oxidizer weight are added to
the sample ID setting screen. Set those sample information.
When the optional loss on ignition is selected for a balance, analyzed values are calculated
with the loss on ignition regarded as a balance. This feature is useful for a sample
containing a loss on ignition.
When fluorescent X-rays with short wavelengths, such as the Zr-K line, are measured, the
thickness of a fusion bead affects analyzed values. In such a case, set a sample size.
Sample - Filter
This setting is used to analyze liquid with the instillation filter paper method using filter
paper such as the UltraCarry or MircoCarry.
When Rigaku’s filter such as the UltraCarry, UltraCarryLight or MicroCarry is selected for
the filter type, a material, weight and effective diameter are automatically displayed. When
using the UltraCarry or UltraCarryLight, set 30mm for the measuring diameter. When using
the MicroCarry, set 20mm for the measuring diameter.
It is possible to set the sample film correction on the sample ID setting screen.
This setting is used to analyze a thin film sample, the thickness and composition of a
thin film on a coated plate, paper with the main component regarded as the balance and
a thin sample.
When a thin sample such as paper, which consists mainly of organic matter, is to be
analyzed, specify cellulose or the like for a balance and set sample weight, diameter, etc.
as the sample size.
It is possible to set the sample film correction on the sample ID setting screen.
Regarding substrate setting, fixed values for the base substrate and the number of layers
under the top layer for analysis can be set and the screening analysis of the top layer
becomes possible. (This is the function of ZSX Primus IV/IVi only, and applicable to the
equipment shipped newly.)
2. Select the objective template to be deleted, and then click Delete .
When a sample name is not entered, registration is conducted using a name that consists
of a year (four digits of the Christian era), month (two digits), day (two digits) and time (two
digits of an hour and two digits of a minute).
The check analysis is carried out to ensure that prepared calibration curves are in
normal operation. Ordinarily the same sample is analyzed for assessment. By executing
the check analysis for the same sample several times a day, X-R control charts can be
drawn to check the relationship between the average values and ranges of analyzed
values. The setting procedure for check samples is described in <4.2 h. Setting of
Analysis Control Information>.
If necessary, an output format that is a combination of a file format and a transfer method
can be designated on the “Result output” list. When an output format that has a transfer
method other than RS232C or TCP/IP is selected, enter a file name.
When the number of standard samples and the number of analyzed components are
small, calibration curves can be prepared each time before analyzing unknown samples.
However, in ordinary quantitative X-ray fluorescence analysis, only the calibration of
X-ray intensities is necessary before analyzing unknown samples. This calibration of
X-ray intensities is called the drift correction. By calibrating X-ray intensities using
several samples, the same calibration curves can be used for a long period of time.
Analysis time can be shortened significantly. In the case where the drift correction is
insufficient, calibration curves must be re-created. For the setting procedure for the drift
correction, see “4.2 f. Measurement of Standard Samples”.
If necessary, an output format that is a combination of a file format and a transfer method
can be designated on the “Result output” list. When an output format that has a transfer
method other than RS232C or TCP/IP is selected, enter a file name.
For the easy selection of an application, a search function that narrows down objects
using the form and application name is available.
・Search Box
A search can be made by entering a character string. When “*” is set for the
character string, all character strings are used for that part.
If any of the positions of registered sample IDs overlaps with any of sample positions to
be used for analyses in Program Operation when EZ Analysis is started, the screen
shown below will appear:
If the screen is displayed, delete or move the sample ID(s) for EZ Analysis so that any
sample position does not overlap.
The position of any registered sample ID must not overlap regardless of whether its status
is reserved or analyzed.
Add New ID
Edit Sample ID
1. Select a sample image to be edited in the ASC display area on the Reserve tab screen,
or select an ID to be edited in the sample ID setting area.
2. Change the set contents according to editing contents in the sample ID setting area.
A sample name and an operator name can be set for the sample information. The sample
name is used to identify each sample and the operator name to identify an operator.
Those are treated as memoranda. They can be used for sorting or searching in Result
Display described later.
When setting has been made such that sample names are displayed on samples, a sample
name can be edited in the ASC display area by double-clicking a selected sample position.
For the display of a sample identification name such as a sample name and an operator
name, refer to the sample ID structure in “6.1 h. System Management”.
A folder in which applications for the SQX analysis, quantitative analysis, etc. are stored
must be selected. Ordinarily all applications are stored in the common folder. If you
create a folder other than the common folder (AnlCond), the folder in which a specific
application is stored must be selected before that application is selected.
For a user for whom a data storage folder has been specified beforehand using the User
Administration program, the specified folder or common folder is used for storage.
A sample name automatically becomes the default file name of a qualitative analysis
result. When the same sample names are set successively, three-digit successive
numbers are automatically added to file names. If the first file name has a number at its
end, the number with the digit(s) is automatically incremented. Therefore, if you set
one-digit number, the 10th file name will return to 1.
A folder to save a result file can be specified using <Folder> in the sample ID setting
area. A subfolder can also be created and specified by pressing ... .
This sample name is specified for processing in the qualitative data handling, SQX
calculation, etc.
When a sample name is not entered, registration is conducted using a name that consists
of a year (four digits of the Christian era), month (two digits), day (two digits) and time (two
digits of an hour and two digits of a minute).
Using the “Data Transfer Setting” program on the “System management” menu, setting
can be made for output to a text file or an HTML file. When <Specify file name> has been
selected for “File name type”, enter a file name for the result output when setting a
sample ID.
This operation is possible in the case of an application for which setting has been made
for manual input data in the SQX analysis, quantitative analysis, etc.
In the case of an application for which setting has been made for manual input data, add
the following operation in the ordinary ID setting procedure:
Although the results of analyses specified in the sample ID setting area on the Reserve
tab screen are generally saved in the following folders, quantitative analysis results and
qualitative analysis results can be saved in subfolders that have been created:
Other Types
(Check Analysis, Drift C:\Rigaku\SxIdv\UserData\AnlRslt Unrestricted
Correction)
Specify an output folder in the sample ID setting area on the Reserve tab screen.
As an example, let us assume that the logon user is “rigaku” and the data storage folder
“rigaku” has been created beforehand using User Administration.
It is also possible to create a subfolder directly under the Common folder. To create a new
subfolder, select the above-mentioned Common folder and then click Create folder....
As for folder designation, specify a folder under the above-mentioned common folder
(AnlRslt) or user folder (AnlRsltU).
By creating a data management folder beforehand for each logon user using User
Administration, data management can be carried out easily.
To display buttons only for analysis types to be used, follow the procedure described
below:
3. Click OK .
It is possible to register five typical applications used frequently. Applications that can
be set for Favorite are ones registered in the common folder (AnlCond) including
subfolders.
The procedure to set applications is as follows:
3. Click OK .
In the case of the calling using the menu item [Load from Sample ID Table], the
following two methods can be used to specify sample positions:
In the case of the calling using a sample ID button in the upper part of the ASC panel,
sample positions registered in a sample ID table are used intact.
A sample ID table to be called from EZ Analysis must meet all of the following conditions:
・Only analysis jobs that can be used in EZ Analysis are contained.
・Only sample IDs using the center measurement for the mapping mode are contained.
・Only analysis conditions using the same atmosphere as the current atmosphere of the
instrument are contained.
When sample positions registered in a sample ID table are used, the following conditions
must also be met:
・Only IDs using sample positions that have not been used by EZ Analysis are contained.
(Example: When the sample position A-1 has been reserved or analyzed, it is
impossible to call a sample ID table using the sample position A-1.)
・The same sample positions are not used duplicately in a sample ID table.
To select sample IDs on the EZ Scan screen, follow the procedure described below:
1. Click a sample ID button in the upper part of the ASC panel. The contents of
sample IDs in the sample ID table are automatically set.
In the case of the calling using a sample ID button in the upper part of the ASC panel,
sample positions registered in a sample ID table are used intact.
The statuses of sample IDs set in the reservation table and the spectrum being
measured are displayed.
The operational procedure for the preset table is described below. The maximum number
of IDs that can be set in the preset table is the maximum number of samples that can be
put on the ASC. A sample ID remains even when an analysis is completed.
Analysis types that can be set for an ID are the following five types:
In EZ Analysis, only one sample ID can be set at a time in the same sample position
regardless of whether it has been reserved or analyzed.
Blank Line
Spectral Display
This screen displays a qualitative analysis chart during the qualitative analysis. The
scale for X-ray intensities can be changed.
Each measuring range is indicated sequentially in turn with tab form in the spectral display.
(This is the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped
newly.)
The analyzed results and X-ray intensities of a selected analyzed sample ID are
displayed.
In the quantitative analysis etc., the display of analyzed results and X-ray intensities
can be switched by clicking the specified position of a measured sample on the ASC to
select the sample ID and clicking Analyzed result or Intensity .
To make operation easier, EZ Analysis does not have the following analyzed result
functions.
If any of theses functions is necessary, use “4.4 Result Display”.
Operation in the sample ID setting area is impossible when no sample ID has been
selected and for a sample ID for which an analysis is completed.
An analyzed result for which the analyzed sample ID has been deleted on the EZ Analysis
screen can be displayed by selecting [Data Processing] [Result Display] on the menu. It
is also displayed on the analyzed result screen of the “Analysis” program.
2.3 ANALYSIS
a. Before Analysis
“Analysis” on the main menu performs qualitative analyses and quantitative analyses
and makes adjustments for those analyses. The three analysis types shown in the table
below are available for routine analysis. Special types of measurement such as the PHA
adjustment and quantitative analysis setup are called analysis control. The routine
analysis is intended for an operator and the analysis control is intended for a manager of
the system or analyses. It is recommended that a PHA adjustment be performed before
routine analyses.
For the qualitative (SQX) analysis and quantitative analysis, applications must be
created beforehand. See chapters 3 and 4 for the procedures to create them.
The EZ scan performs the qualitative (SQX) analysis and creates measuring conditions
directly by dialogs. This type is recommended for beginners in the X-ray fluorescence
analysis.
Screens for analysis and the setting procedure for each analysis are described below.
Make a startup check and start up the system to make it ready for analyses.
The Analysis program has the main screen <Sample ID Setting> and sub screens
<Analyzed Result>, <Qual Analysis Chart>, <ASC Status> and <Operation Status>.
The <Analyzed Result>, <Qual Chart> and <ASC Status> screens overlap each other in
the initial status, but they are displayed in the foreground when the button in the upper
right part of the <Sample ID Setting> screen is clicked. When the main screen <Sample
ID Setting> is closed, all the sub screens are closed.
Sample ID Setting
(Main Screen)
Sample ID Setting
This screen appears in the foreground when the corresponding button in the upper right
part of the Sample ID Setting screen is clicked. All results of the qualitative and
quantitative analysis are displayed on this screen. When an analysis or measurement is
complete, the newest result is automatically displayed. Old data can be displayed by
clicking an ID line in the upper part of this screen. The upper part is called the ID
selection area and the lower part called the result display area. The results of the
analysis for the ID in the light blue cursor line selected in the ID line selection area are
displayed.
In the quantitative analysis, analyzed results and X-ray intensities can be switched by
clicking Analyzed Result or Intensity in the lower part of the screen.
When Display Latest Data is clicked, the newest analyzed results are displayed. The
analyzed result display is updated each time the analysis of one sample is complete.
The “Analyzed Result” screen has the same functions as those of “Result Display”. For the
details of operations, see <4.4 RESULT DISPLAY>.
This screen appears in the foreground when the corresponding button in the upper right
part of the Sample ID Setting screen is clicked. This screen displays a qualitative
analysis chart during the qualitative analysis. When the qualitative analysis starts, this
screen is automatically displayed in the foreground. The scale for X-ray intensities can
be changed.
Each measuring range is indicated sequentially in turn with tab form in the spectral display.
(This is the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped
newly.)
This screen appears in the foreground when the corresponding button in the upper right
part of the Sample ID Setting screen is clicked. Sample positions designated in the
preset table are displayed. They do not necessarily correspond to sample holders really
placed on the ASC.
This screen appears when [Window] [5:Operation Status] is selected on the menu.
The analysis or measurement status for an ID line is displayed. This screen consists of
three screens - the progress of the analysis for the ID line, spectrometer status and
mapping measurement positions. They can be switched using the button in the lower
right part of the screen. Examples of the spectrometer status and the progress of the
analysis are shown below.
Display of Screens
When the Analysis program is started, screens are arranged automatically. If any screen
size or position has been changed, the screens cannot be displayed as before using the
<Tile Windows> or <Cascade Windows> functions. In that case, they can be displayed as
shown below by selecting [Window] [Initial Position] on the menu.
The operating procedure for the preset table on the Sample ID Setting screen is
described below. Up to 999 ID lines can be set in the preset table. An ID line for which an
analysis or measurement has been complete moves to the ID list on the Analyzed Result
screen. That is, an ID line in the preset table is deleted each time an analysis is complete.
Analysis or measurement types that can be set in an ID line are of the following four
types:
ID Type Description
EZ Scan Setting for a qualitative analysis
Setting for a qualitative analysis, quantitative analysis
Sample ID
and analysis control
Control ID Setting for a control ID such as atmosphere change
Cycle Repeat Analysis Setting for cycle repeat analysis
In the above figure, line numbers 1 through 7 are called the ID lines. Line numbers 8, 5
through 7 are called the blank line and selected lines, respectively.
Blank Line
The blank line to set a new ID line is always displayed in the preset list regardless of the
number of preset IDs. To add a new ID line, this blank line is selected. This blank line
cannot be copied or deleted even when it is included in selected IDs.
Cursor Line
This is a display line to select a specific ID in the preset table. The cursor line is moved
with the up and down arrow keys or by clicking the mouse.
It is used to designate a position to insert an ID, to identify an already preset ID line to
be changed and in similar cases.
Selected Lines
Take special care for selected ID lines because they can be easily deleted by pressing the
Delete key.
The operating procedure for ID lines such as the EZ scan, sample ID, control ID and
cycle repeat analysis is described below. Basic operations are creation, insertion, edition,
copying and deletion. More than one line can be selected. For the detailed setting
procedure for each ID, see each paragraph.
Add New ID
Insert New ID
Edit ID Line
Copy
The copy function can reproduce the previous ID line(s) or an ID line or lines in any
position. Each procedure is described below.
To copy one preset ID line is called “Copy of ID above”. To copy two or more preset lines is
called “Batch Copy of ID above”. The processing can be carried out regardless of whether
the line next to the copied line is a blank line or another ID line. In “Copy of ID above”
and “Batch Copy of ID above”, sample positions are updated successively.
An ID line for which an analysis has been complete as well as a line in the preset table
can be selected for this copy function. Thus, an analyzed ID line from the result display
can also be copied. The procedure is the same. A copied ID line can be pasted to a blank
line or another ID line.
Clear ID Line(s)
All ID lines or specified ID lines can be cleared. Clearing of specified ID lines means
cutting, but the most recent cleared ID lines can be pasted. Source ID lines remain in the
preset table after the copying operation, while they do not remain in the table after the
cutting operation. In the case of clearing of all IDs, the pasting operation is impossible.
The procedure to select two or more ID lines for copying or cutting of ID lines is
described below. Three types of selections can be used: selection of all ID lines, selection
of successive ID lines and selection of ID lines in arbitrary positions.
or
For the ID line currently in analysis, the progress of the analysis is generally displayed
or the progress of a cycle repeat analysis is displayed. Each of these is described below.
The status of a set ID line changes as shown below until it is deleted from the preset
table after an analysis is complete. This status is shown by the color of the SEQ#. The
default colors are shown below.
Background Character
SEQ# Description
Color Color
001 Preset White Blue
002 Waiting for analysis Green Blue
003 During preparation (waiting for loading) Yellow Red
004 During analysis Red Yellow
005 Analyzed (waiting for unloading) Violet White
The cycle repeat analysis is carried out in one ID line. Therefore, in the case of a
10-times analysis, that ID line remains until measurement is repeated 10 times. When
that ID line is double-clicked, the “Progress of Cycle Repeat Analysis” screen appears.
Care must be taken to note that the ID line of an analyzed result is moved to the area of
analyzed results. The “Progress of Cycle Repeat Analysis” screen is closed automatically
when the cycle repeat analysis is complete.
As described before, items for the EZ scan, sample ID, control ID and cycle repeat
analysis can be set in ID lines. Common operations concerning IDs except the control ID
are described below.
Measuring
Common Sample Sample Manual
Point File Name
Operation Position Information Input Data
Coordinates
SQX
Necessary Necessary Possible Necessary Possible
Analysis
Sample ID
Quantitative
Necessary Necessary Possible Possible Possible
Analysis
Analysis
Necessary Necessary Impossible Possible Possible
Control
Cycle Repeat
Necessary Necessary Possible Possible Possible
Analysis
The setting of measuring point coordinates can be selected only in the case of a system
equipped with the mapping mechanism.
Set a sample position by specifying a number for the X direction and an alphabetic
character for the Y direction, as shown below:
A sample position can be set from the image display of the table using Select…. In the case
of the ASC48, the range allowed for the sample position varies depending on the number of
trays. If the number of trays is one, the range is from A-1 to B-6 and up to 12 samples can
be set.
A sample name and an operator name can be set for the sample information. The sample
name is used to identify each sample and the operator name to identify an operator.
Those are treated as memoranda. They can be used for sorting or searching in Result
Display described later.
To enter a sample name, double-click the cell to activate the vertical cursor. An operator
name can be entered in the same way, but it can also be selected from a list that has been
registered beforehand.
A folder in which applications for the SQX analysis, quantitative analysis, etc. are stored
must be selected. Ordinarily all applications are stored in the common folder. If you
create a folder other than the common folder, the folder in which a specific application is
stored must be selected before that application is selected. Select an objective folder from
<Folder> below the application selection list of the analysis information.
If you want to permanently record an analyzed result, you must specify a file name. If
you do not set a file name, an analyzed result will be lost after it is completed. Care
should be taken because old data will be overwritten if the same file name is specified.
When the same IDs are set successively, three-digit successive numbers are
automatically added to file names. If the first file name has a number at its end, the
number of the digit(s) is automatically incremented. Therefore, if you set one-digit
number, the 10th file name will return to 1.
A folder to save a result file can be specified using <Folder> below the file name input
field. A subfolder can also be created and specified by pressing Browse... .
Using the “Data Transfer Setting” program on the “System management” menu, setting
can be made for output to a text file or an HTML file. When <Specify file name> has been
selected for “File name type”, enter a file name for the result output when setting a
sample ID.
A mapping analysis is carried out according to coordinates that are preset in a mapping
table. Include the following operation in the ID setup.
When carrying out a mapping analysis, be sure to use a sample holder for the mapping.
To display a mapping analysis result using the “Mapping” program, a mapping file must
be created. Although a mapping file can be created using the “Result Display” program
after an analysis, it can be created automatically by selecting <Create mapping file>
when setting a sample ID.
A subfolder can be created and specified by pressing Browse… on the “Mapping file name”
screen.
This operation is possible in the case of an application for which setting has been made
for manual input data in the SQX analysis, quantitative analysis, etc. In the case of the
mapping analysis, manual input data must be set for each measuring point. If you want
to set the same value for all points, click Copy All points .
In the case of an application for which setting has been made for manual input data, a
value must be entered for each measuring point. Include the following operation in the
ID setup.
Copy All points copies the value of the coordinates where the cursor is placed.
Although analyzed results that appear on the Analyzed Result screen after analyses or are
displayed by the Result Display program are generally saved in the following folders,
quantitative analysis results and qualitative analysis results can be saved in subfolders
that have been created:
Other Types
(Check Analysis, Drift C:\Rigaku\SxIdv\UserData\AnlRslt Unrestricted
Correction)
As an example, let us assume that the logon user is “rigaku” and the data storage folder
“rigaku” has been created beforehand using User Administration.
It is also possible to create a subfolder directly under the rigaku folder. To create a new
subfolder, select the above-mentioned rigaku folder and then click Create folder....
As for folder designation, specify a folder under the above-mentioned common folder
(AnlRslt) or user folder (AnlRsltU).
By creating a data management folder beforehand for each logon user using User
Administration, data management can be carried out easily.
When a sample ID is clicked on the Sample ID Setting screen, the folder to save the data of
that sample ID is displayed in the upper part of the screen (only a subfolder).
The folder for saving that has currently been set is used for a sample ID that has been set
newly or created using <Copy of ID above>. However, the common folder (or its subfolder)
or the logon folder (or its subfolder) is used to store sample IDs that have been read from a
sample ID table.
Of the above-mentioned folders to save analyzed results that appear on the Analyzed
Result screen or are displayed by the Result Display program, specify a folder for display
using these programs. The storage location is not restricted to the above-mentioned
common folder AnlRslt and its subfolders. It is also possible to display data in folders in the
user folder AnlRsltU.
1.Click the title bar or a sample ID in the Analyzed Result window to move the focus
to the Analyzed Result window.
2.Select [View]→[Select Folder].
Or
1. Click Browse... on the Analyzed Result screen and select a folder.
PHA adjustment results, drift correction results and check analysis results are
unconditionally saved in the AnlRslt folder. If the folder for data displayed in the Analyzed
Result window is not the AnlRslt folder when carrying out these measurements, results will
not be displayed automatically on analyses. Therefore, it is generally advisable to specify
the AnlRslt folder.
b. EZ Scan
Setting items will be described below. For the sample position and sample information,
see the paragraph <2.3 a. Before Analysis>.
For items other than the measuring time and file name, see the paragraph “2.2 EZ
Analysis - Make Setting for EZ Scan”.
Measuring Time
Time Description
“Shortest” is the function of ZSX Primus IV/IVi only, and applicable to the equipment
shipped newly.
File Name
Set a file name to store qualitative spectral data. A default file name is displayed by
searching the last sample IDs already created including analyzed sample IDs as shown
below.
The setting procedure for the EZ scan will be described below. When more than one
sample needs to be analyzed, repeat the procedure. Care must be taken because old data
will be overwritten if you set the same file name as one that has already been used.
A component type etc. for each sample type are default settings for new registration in
the advanced setting in EZ Analysis, as shown below:
Component
Sample Type Balance Others
Type
Metal & Alloy Metal None None
Powder Oxide None None
Polymer Metal CH2 None
Fusion Bead Oxide None Flux: Li2B4O7, dilution ratio: 10
Filter Metal ― MicroCarry, sampling: 50L
Film & Coating Metal None None
Liquid Metal H2O None
To change the measuring condition or output information, see “Edit EZ Scan Condition”
in “3.2 a. Application File”.
When detailed information is different from the above default contents, change the sample
information using the SQX calculation program, and recalculate analyzed values. For the
detailed operational procedure, see “3.3.2 SQX CALCULATION”.
A qualitative analysis that carries out quantification to some extent is called the
semi-quantitative analysis.
The operation flowchart of the SQX analysis is shown below. The sample transportation,
operation control, etc. during an analysis are carried out automatically.
Unlike the EZ scan, the SQX analysis has various functions such as setting for manual
input data and the optional mapping analysis that are applicable to specific purposes.
The standard measurement will be described first and then details will be explained.
For descriptions of setting items, see the paragraph <2.3 a. Before Analysis>.
It is necessary to enter the sample-size condition when the analysis is carried out by
using SQX Scattering FP method. Measure the sample size in advance before analyzing
and enter it.
The base size shows the actual size of the analysis surface and has three kinds of the
setting way as follows.
Mass per unit area is calculated with input data of mass and base size. Height is used to
correct intensity calculation against depth and the value of height requires less
precision than mass and base size.
d. Quantitative Analysis
The analysis for process control or quality control is called quantitative analysis. To
execute quantitative analysis, applications must be prepared beforehand. Key words
such as “non-destructive”, “fast”, “high precision” and “low cost” are known as features of
the X-ray fluorescence analysis. However, it is essential that careful consideration be
given to all aspects of quantitative analysis for the method to be used correctly. For the
creation of quantitative applications, see chapter 4.
The recommended procedure of the routine control analysis is shown below. When an
operating procedure for the control analysis has been established, follow it. For the
operating procedure for the check analysis, drift correction, etc., refer to the following
paragraphs for the analysis control.
PHA adjustment
Drift correction
When the results of the PHA adjustment and check analysis are satisfactory, you can
carry out the quantitative analysis on unknown samples. The setting procedure for it is
described below. For descriptions of setting items, see the paragraph <2.3 a. Before
Analysis>.
e. Analysis Control
The analysis control means the PHA adjustment, check analysis, drift correction and so
on in the quantitative analysis. Analysis control items other than the PHA adjustment
should be executed if necessary and are classified as follows:
<Library sample> is only effective when the SQX software option is attached.
Update Selection
“Update Selection” is designated for <Drift corr. sample>, <Bias corr. sample>,
Standard sample> and <Scattering drift library> of Library sample. In these types of
measurement, in parallel with a series of measurement, calculation control such as
preparation for calculation, data storage and calculation instructions must be carried
out to obtain various coefficients. By setting “Update Selection” properly, this calculation
control can operate correctly.
If <Last sample> is not measured, coefficients will not be updated for Drift Correction
Sample, Bias Correction Sample, Standard Sample or Scattering drift library.
Calibration curve coefficients for the empirical calibration method and sensitivity
constants for the FP method are ordinarily calculated using the Regression Calculation
program in the quantitative application flow. However, they can also be automatically
calculated and saved using Update Selection for standard samples after measuring
them.
This function can be used for an application for which data has been examined using the
Regression Calculation program and calibration curve coefficients and matrix correction
coefficients for the empirical calibration method or sensitivity constants for the FP
method have been registered. This function is effective when X-ray intensities have
changed greatly after replacing the X-ray tube of the spectrometer.
In this automatic update, only calibration curve coefficients for the empirical calibration
method or sensitivity constants for the FP method are calculated. Matrix correction
coefficients for the empirical calibration method or theoretical intensities and overlap
correction coefficients for the FP method are not calculated. (Values that have been
registered are used intact.)
The standard samples from <First sample> to <Last sample> are used for calculation.
Sample selection used in the Regression Calculation program is also reflected.
The PHA adjustment should be executed before routine unknown analyses. In the case
of 24-hour continual operation, PHA adjustment should generally be carried out once a
day. After the P-10 gas cylinder or PC window is exchanged, the energy axis should be
checked and readjusted, if necessary. The axis is ordinarily stable and does not shift. The
PHA adjustment should be carried out about 30 minutes after aging and any adjustment
of the P-10 gas flow rate. Both of the two detectors are adjusted using the built-in
samples of the instrument. The PHA adjustment can be started from the Analysis menu
or the Spectrometer Startup menu. For the procedure from the Spectrometer Startup
menu, see <1.2 b System Startup> described before.
Since the system has two detectors, the PC and SC, set IDs for both of them.
In the case of an instrument for which Step 6 is unnecessary, the PHA adjustment sample
has been housed in the instrument, and the built-in sample is automatically measured to
make the PHA adjustment.
The check analysis is carried out to ensure that prepared calibration curves are in
normal operation. Ordinarily the same sample is measured for assessment. By executing
the check analysis for the same sample several times a day, X-R control charts can be
drawn to check the relationship between the average values and ranges of analyzed
values. The setting procedure for check samples is described in <4.2 h. Setting of
Analysis Control Information>.
If necessary, an output format that is a combination of a file format and a transfer method
can be designated on the “Result output” list. When an output format that has a transfer
method other than RS232C or TCP/IP is selected, enter a file name.
When the number of standard samples and the number of analyzed components are
small, calibration curves can be prepared each time before analyzing unknown samples.
However, in ordinary quantitative X-ray fluorescence analysis, only the calibration of
X-ray intensities is necessary before analyzing unknown samples. This calibration of
X-ray intensities is called the drift correction. By calibrating X-ray intensities using
several samples, the same calibration curves can be used for a long period of time.
Analysis time can be shortened significantly. In the case where the drift correction is
insufficient, calibration curves must be re-created. For the setting procedure for the drift
correction, see <4.2 f. Measurement of Standard Samples>.
If necessary, an output format that is a combination of a file format and a transfer method
can be designated on the “Result output” list. When an output format that has a transfer
method other than RS232C or TCP/IP is selected, enter a file name.
Analyzed results of unknown samples using prepared calibration curves may have
constant biases from expected values. This is because, for example, the preparation
method for the standard samples on the market that are used to prepare the calibration
curves is different from that for the unknown samples. The operation to correct this
phenomenon is called the bias correction. For the setting procedure for the bias
correction, see <4.2 h. Setting of Analysis Control Information>.
If necessary, an output format that is a combination of a file format and a transfer method
can be designated on the “Result output” list. When an output format that has a transfer
method other than RS232C or TCP/IP is selected, enter a file name.
The standard sample measurement must always be carried out to prepare calibration
curves or FP sensitivity curves. Because quantitative analysis in the X-ray fluorescence
analysis employs the standard sample comparison method, standard X-ray intensities
are necessary. Samples used to create calibration curves are called standard samples
and the procedure to measure them is described below. For the setting procedure for
standard samples, see <4.2 c. Setting of Standard Samples>.
If necessary, an output format that is a combination of a file format and a transfer method
can be designated on the “Result output” list. When an output format that has a transfer
method other than RS232C or TCP/IP is selected, enter a file name.
Reference X-ray intensities are necessary for the X-ray fluorescence analysis. Standard
samples are generally used to create calibration curves, but in some cases standard
samples that have concentrations similar to those of unknown samples are not available.
In those cases, the FP quantification method can be used to perform quantitative
analyses without standard samples. Spectrometer sensitivity is referenced in that FP
quantification method. A collection of spectrometer sensitivity values is called a
sensitivity library. The procedure to measure those library samples is described below.
For the sensitivity library, see <3.4 SENSITIVITY LIBRARY(OPTION)>.
f. Control ID
Spectrometer
Description
Control
Used to stop temporarily the execution of this ID line to
Temporary stop
change a sample holder or to wait for the arrival of a sample
Atmosphere Switches the analysis chamber atmosphere to the one
change suitable for the sample.
Raises the tube voltage and current slowly to the specified
Tube aging
values.
X-rays and the power of the spectrometer can be turned off
Automatic power
automatically after analyses by specifying in Program
off (Option)
Operation
This function can be used, for example, when you have set ID lines to analyze unknown
samples but want to stop a series of analyses because it takes a long time to prepare
samples. This can also be used when you want to stop measurement while a sample is
exchanged in the case of one-sample loader. To cancel a temporary stop or to terminate
measurement, follow the message.
This function is used to change the analysis chamber atmosphere according to the
application to be analyzed or measured.
When the spectrometer does not have the helium flush mechanism, <Helium/Vacuum>
cannot be selected. When the spectrometer has the helium flush mechanism, <Air> cannot
be selected.
It is possible to use this function only when the program operation option is attached.
The contents of a method to select must be set beforehand using Program Operation.
g. Sample ID Table
As described before, an area to place IDs in order of analysis is called the preset table.
When you must set the same sample IDs each time to this preset table, you can save
time and trouble by creating a sample ID table. For example, when the same contents
must be set every time as in the case of the PHA adjustment, a sample ID table in which
IDs have been set beforehand can be loaded to the preset table. This sample ID table is
also used for the cycle repeat analysis described later. In addition, sample IDs set in the
Analysis program can be registered in a sample ID table. For further details, see <6.1 c.
Sample ID Table>.
After insertion, existing ID lines move and follow the inserted sample ID table.
If the number of blank lines in the preset table is smaller than the number of IDs in the
specified ID table, that table cannot be loaded.
The EZ scan ID and cycle repeat analysis ID cannot be saved in a sample ID table. And
“Secure vacuum” and “Power off” in the control IDs cannot be saved in a sample ID table.
The cycle repeat analysis can be used to check the performance of the system. The
dynamic mode and the static mode are available in the cycle repeat analysis. Before
executing a cycle repeat analysis, ID lines must be set beforehand in a sample ID table.
A sample ID table must be set even when one sample is to be used for a cycle repeat
analysis. Only one ID line is set, but an analyzed result is displayed in one ID line for
each analysis. Up to a maximum of 999 repeats can be set.
The output of calculated results can be selected from among the raw intensity, X-ray
intensity and concentration. The “interval” means a delay time from each completion of
measurement to the next measurement. The maximum interval is 99 hours and 59
minutes.
Output Description
Automatization of correction (Auto drift correction) can be made using the sample ID
table for automatic drift correction which is set the check sample setting (lower limit,
upper limit, having been set, and drift correction. Labor-saving of analysis preparation
work is realized with this function.
5. Makes judgement automatically if it is in the range and repeat 4. and 5. above to see
if it is out again.
Auto drift correction is made up to two times. If it is the case that the check analysis after
second drift correction is still out of range, make calibration curves again.
1. Select [Edit] -> [Add new ID] -> [5:Auto drift correction…].
2. Auto Drift Correction screen is indicated and the sample ID table can be selected.
Select the objective sample ID table from the pull down list, then click OK .
3. Auto drift correction sequence is set in the sample ID setting screen. Set the
designated sample on the sample tray.
4. Click Analyze to start analysis. Judgment of tolerance and drift correction are
executed automatically.
Make the dedicated sample ID table for auto drift correction beforehand.
To start specified analyses at a specified time, the fixed time analysis and the fixed
period analysis can be used. For example, with the fixed time analysis, a PHA
adjustment can be made at a specified time once a day. With the fixed period analysis, a
check analysis can be made at constant time intervals.
It is possible to use this function only when the program operation option is attached.
The fixed time analysis and fixed period analysis have priority over all analyses except
the one now in progress. So sample IDs for the SQX analysis, quantitative analysis, etc.
that have been set on the Analysis screen are carried out after the fixed time/period
analysis.
The fixed time/period analysis is not valid unless the Analysis program on the main
menu has been started. For the setting procedure, see <6.1 g. Program Operation>.
This option provides functions to interface with the LIMS, reads an analysis request file
created by the LIMS and registers a sample ID based on the content of the file.
The ZSX Guidance reads an analysis request file in “C:\EXTLIST”. Set the LIMS
beforehand such that an analysis request file is created in “C:\EXTLIST”.
Create beforehand a quant application that will be used for an analysis. When it has
a component that requires manual input data, make setting such that input is
possible.
After reading an analysis request file, make an analysis in the ordinary procedure
using the registered sample ID.
The analysis request file can have any file name and the extension of the file is fixed
with TXT.
The analysis request file may consist of the sample analysis request of more than one
sample. Sample analysis request for one sample is written in one line.
Line change codes are CRLF(0DH 0AH).
Analysis request 1(CRLF)
Analysis request 2(CRLF)
-
-
Analysis request n(CRLF)
Sample request for each sample is written in one line and the analysis request consists
of three or more fields depending on number of manual inputs and the fields are
separated by a TAB((HT,09H).
Field 1 HT Field 2 HT --- HT Field m
Format of Field
Field Manual input data The value of the N-th manual input data is set.
3+N N
The intensity measurement cannot be used and only the quantitative analysis is
made.
The measurement is only for center measurement and not for mapping measurement.
The manual input data are set in the sample ID file following the order of quant
application. If number of manual input set in analysis request is larger than the
number of manual input registered in the quantitative analysis application, the rest
of manual input data are disregarded. If the number of manual input data set in
analysis instruction is smaller than the number of manual inputs registered in the
quantitative analysis application, the software display an error message and any
analysis request cannot be registered to the sample ID file for the file.
A null field cannot be set. When a sample name need not be set (there is no input memo in
the sample ID structure), set a dummy character string such as “dummy” or “abc”.
1. Select [File] [External Analysis Request...] on the “Sample ID Setting” screen.
2. Select an analysis request file.
3. Click OK .
Sample IDs are registered on the “Sample ID Setting” screen according to the selected
analysis request file.
When sample IDs have been registered normally, the analysis request file is deleted.
If at least one of the analysis commands in the analysis request file is erroneous, no sample
ID will be registered and the analysis request file will not be deleted.
1. Select [File] [Delete External Analysis Request File...] on the “Sample ID Setting”
screen.
2. Select an analysis request file to be deleted.
3. Click OK .
4. Click Yes on the dialog to confirm the deletion.
3. QUALITATIVE ANALYSIS
3.1 QUALITATIVE ANALYSIS
Start
Measurement
End
It is very useful to carry out a qualitative analysis to know what components are present
in the sample. The semiquantitative analysis program augments the qualitative
analysis, providing approximate quantities of those components. In this system, this
semiquantitative analysis program is called the SQX. To perform qualitative analysis
and semiquantitative analysis, you must create a qualitative application.
Since the application procedure for qualitative analysis is the same as an SQX
application without the setting for the SQX calculation, an example of the creation of an
SQX application is described below. The operational flow to create the application is
shown below:
Start
Use template?
No
Yes
Selection of template Creation of original application
End
When Qual Application is clicked on the menu, steps are displayed on the flow bar in the
right part of the screen to show the creation procedure.
Each step displayed on the flow bar is described below:
An example of the creation of an SQX application using the template for metals and
alloys is described below. Default conditions will be accepted. When operations for an
application file are complete, it is registered. Check the set contents and change them if
necessary.
The folder specified by the user with User Administration in “6.1 h. System Management”
has been selected as the default.
SQX Scattering FP method (Option) has the capability of estimating the extent of influence
by non-measurable elements to measurement element using scattered X-rays.
Consequently, it makes possible to obtain the accurate analysis on elements to be
interested without measuring ultra light elements.
When SQX Scattering FP method is employed, don’t specify the elements ranging from B
to O as measurement elements because the component estimation of these ultra light
elements is done by using scattered X-rays.
12.Click Next .
13.When a sample film is used, make setting for it and then click Next .
When SQX Scattering FP method is used, only 20 mm can be set for the measuring
diameter under the condition that the sealing (option) is not in use. (Selection of
measurement size, 20mm or 30mm, can be made depending on ship date of ZSX Primus
IV/IVi.)
17.Click Next .
18.Click Next .
a. Application File
There are two kinds of templates: a group to carry out the SQX calculation and a group
for qualitative analysis alone. A template for qualitative analysis is simply an
application without the SQX calculation. Templates for the SQX calculation are
classified into seven types: the metal & alloy, powder, glass fusion bead, polymer, filter,
film & coating and liquid. For those templates, only the SQX calculation conditions are
different. All other measuring conditions are the same. There is no need to prepare
many applications for SQX analysis because a recalculation can be made after
measurement by changing calculation conditions.
When creating an application without using a template, select <Create your own
application>.
When a part of an existing application has been changed, the old contents will be
overwritten if that application is saved under the same name. For comparison with past
data, it is recommended that an application with which measurement was made at least
once be kept without changing its contents.
Register as Template
Delete Application
EZ Scan applications are edited from the following nine applications when setting is
made for EZ Scan:
Although operational procedures for qualitative applications are the same as those for
ordinary applications, alterations become invalid to the EZ Scan settings at the time of
an analysis such as the measuring diameter, atmosphere, metal/oxide, compound table
and print option and to the contents that are set using EZ Scan in System Parameters.
Outline of template condition for EZ Scan (For ZSX Primus IV/IVi, in case of standard four
crystal composition)
Application Spectrum Filter Slit Crystal Detector Scanning angle
name (deg)
**_Solid_E Heavy Al25 S2 LiF(200) SC 5.000-90.000
**_Liquid_E Heavy(1) Ni400 S2 LiF(200) SC 13.000-19.000
Light(Ca-K) Al25 S4 LiF(200) PC 110.000-147.000
Light(Cl-Al) Be30(opt) S4 PETH/PET PC 62.000-147.000
Light(Mg-F) Be30(opt) S4 RX26 PC 35.000-77.500
**_Solid_F Heavy Al25 S2 LiF(200) SC 5.000-90.000
**_Liquid_F Heavy(1) Ni400 S2 LiF(200) SC 13.000-19.000
Light(Ca-K) Al25 S4 LiF(200) PC 110.000-147.000
Light(Cl-P) Be30(opt) S4 GeH/Ge PC 90.000-144.000
Light(Si-Al) Be30(opt) S4 PETH PC 105.500-147.000
Light(Mg-F) Be30(opt) S4 RX26 PC 35.000-77.230
**_Solid_N Heavy Al25 S2 LiF(200) SC 5.000-90.000
**_Liquid_N Heavy(1) Ni400 S2 LiF(200) SC 13.000-19.000
**_Solid_S Ca-KA Al25 S4 LiF(200) PC 110.000-116.000
**_Liquid_S K -KA Al25 S4 LiF(200) PC 133.000-140.000
Cl-KA Be30(opt) S2 GeH/Ge PC 90.000-96.000
S -KA Be30(opt) S4 GeH/Ge PC 107.000-114.000
P -KA Be30(opt) S4 GeH/Ge PC 138.000-144.000
Si-KA Be30(opt) S4 PETH/PET PC 106.010-112.010
Al-KA Be30(opt) S4 PETH/PET PC 138.990-147.990
Mg-KA Be30(opt) S4 RX26 PC 35.000-41.000
Na-KA Be30(opt) S4 RX26 PC 43.000-49.000
F -KA Be30(opt) S4 RX26 PC 70.000-77.000
The “Light” condition which straddles over numbers of light elements is the exclusive
measuring condition for ZSX Primus IV/IVi. (This is the function for ZSX Primus IV/IVi only,
and applicable to the equipment shipped newly.)
Above-mentioned applications for EZ Scan are used for both “EZ Analysis” and “Analysis”
program. For contents to be set at the time of an analysis, such as the sample type, see
“2.2 EZ ANALYSIS” and “2.3 b. EZ Scan”.
When EZ Scan based on the film & coating model is performed, long wavelength X-rays of
heavy element (L line, M line) become detectable. (This is the function for ZSX Primus
IV/IVi only, and applicable to the equipment shipped newly.)
When creating an application, an application folder can be created on the screen to enter
an application name. Although the <Common> folder is ordinarily used, another folder
can be specified. An application folder is created on the “Application Name Input” screen,
which appears when creating a new application.
An application folder is created also when a new folder is specified for an analysis condition
file in User Administration in <6.1 h. System Management>. The procedure described here
is used to create an application folder independent of users.
For a single component in the light element range, setting can be made by clicking an
element symbol.
For a component in the heavy element range in the case of the full range qualitative
analysis, setting can be made by clicking Ti-U .
Elements in the heavy element range (Ti to U) can be measured with one crystal and that
measurement is called the full range qualitative analysis. It is displayed as “Heavy” or “Hv”.
1. Select a component to be deleted from the ones displayed on the left side of the
screen.
2. Right-click the mouse to display a pop-up menu.
3. Click Delete .
To delete all qualitative analysis ranges, click Delete all. In that case, the fixed angle
analysis conditions are also deleted.
This function is effective only when the SQX software option is attached.
When there is a component of special importance, measure it by adding a fixed angle
analysis. One can obtain better statistical data on X-ray intensities using fixed angle
measurements. Although a long measuring time can also be set in the qualitative
measurement, it is virtually impossible because the total measuring time becomes very
long. The fixed angle analysis measures three positions, a peak position and two
background positions. Because a longer measuring time for an individual element can be
set compared to the qualitative analysis, X-ray intensities with smaller statistical errors
are obtained.
1. Click Fixed angle elem. in the “Fixed angle analysis” frame on the “Select Element
Range” screen.
2. Select an element name for a fixed angle analysis on the periodic table.
3. When you selected the element name for the full range qualitative analysis (Heavy),
select an individual line name.
4. Click Next .
The fixed angle measurement cannot be set for a component for which the qualitative
analysis is not made. Setting cannot be made for Ti to U unless the full range qualitative
analysis (Heavy) is carried out.
1. Click Fixed angle elem. in the “Fixed angle analysis” frame on the “Select Element
Range” screen.
2. Select a component to be deleted from the elements displayed in light blue on the
periodic table.
This function is effective only when the SQX software option is attached.
The overlap line detection is a function to make a quantitative analysis only from a
2-theta scan measurement result without making a fixed angle measurement, assuming
that a sample surely contains an element regardless of an automatic identification
result. The overlap line detection should be set when an element of interest may not be
automatically identified because of the overlap of the characteristic X-rays of another
component.
1. Click Overlapped elem. in the “Fixed angle analysis” frame on the “Select Element
Range” screen.
2. On the periodic table, select the element name of a component for which to carry out
the overlap line detection.
3. When you selected the element name for the full range qualitative analysis (Heavy),
select an individual line name.
4. Click Next .
The overlap line detection can also be added at the time of the SQX calculation. See
“Change Measured Element Line” in “3.3.2 a. Detailed Description of SQX Calculation”.
1. Click Overlapped elem. in the “Fixed angle analysis” frame on the “Select Element
Range” screen.
2. Select a component to be deleted from the element names displayed in orange on the
periodic table.
To delete all fixed angle measurement and overlap line detection conditions, click Delete
all .
1. When a measurement is to be made using “Sample film”, set a sample film.
Absorption by the sample film is corrected at the time of a quantification calculation.
n case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “PROCEDURE FOR
ADDITION OF SAMPLE FILM INFORMATION ” regarding details of sample film. (This is
the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
2. When Show option is clicked, detailed set contents are displayed in the lower part of
the screen.
The option information setting is used to set items that are not ordinarily used and to display
memoranda and comments on analysis results.
1. When binder is to be used, set information on it. Select <Used> and a binder
component and set sample weight and binder weight.
2. When a sample film is to be used, set information on it. This setting is used to correct
the absorption of X-rays by the sample film.
3. When a sample has been pulverized, set information on a container. This setting is
used to display a comment on an analyzed result.
4. When a sample thickness affects analyzed values, make setting for “Sample weight
thickness”. When <Input for each sample> is selected, a sample size is set for each
sample at the time of an analysis.
When binder is set, analyzed values are calculated with the absorption of X-rays by the
binder taken into consideration.
Make setting for “Sample weight thickness” when fluorescent X-rays with short
wavelengths are to be measured and X-ray intensities are to affect sample thicknesses.
When a weight thickness varies for each sample, select <Input for each sample>.
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “PROCEDURE FOR
ADDITION OF SAMPLE FILM INFORMATION ” regarding details of sample film. (This is
the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
1. Enter the kind and weight of flux and sample weight in “Sample and flux”. Analyzed
values are calculated using an obtained dilution ratio.
2. When a sample has been pulverized, set information on a container. This setting is
used to display a comment on an analyzed result.
3. Set information on oxidizer, which is added when producing a fusion bead, and make
setting for “Non-wetting”. The information on the oxidizer is used to calculate
analyzed values. “Non-wetting” is used to display a comment on an analyzed result.
4. Set a bead diameter when you want to estimate the effect of a thickness on analyzed
results.
1. Select a filter type. When “Other” is selected, enter a filter material, filter weight and
effective diameter.
2. Select a filter material.
3. Enter the weight of the instillation part of a filter.
4. Enter a sampling quantity for instillation.
5. Select an output unit in liquid for final analyzed results from <mass%> and <ppm>.
6. When a sample film is to be used, select <Used> and a sample film.
When Rigaku’s filter such as the UltraCarry, UltraCarryLight or MicroCarry is selected for
the filter type, a material, weight and effective diameter are automatically displayed. When
using the UltraCarry or UltraCarryLight, set 30mm for the measuring diameter. When using
the MicroCarry, set 20mm for the measuring diameter.
C6H10O5 (cellulose) is set for the filter material of the UltraCarry, UltraCarryLight and
MicroCarry. If a material is not displayed when setting the <Other> filter, enter and register
a compound name using “Special compound” in the “Compound Table” program.
The filter weight means the weight of the part of a filter where instilled liquid diffuses. In the
case of light elements, because of the ununiformity of diffusion, analyzed results can be
improved by analyzing liquid with known concentrations using the SQX analysis and
obtaining weight that leads to accurate analyzed values.
Since detection sensitivity for fluorescent X-rays is not uniform on a filter, a value slightly
different from the actual diameter of the instillation part of a filter must be used as the
effective diameter to obtain accurate analyzed values. When using the <Other> filter,
analyze liquid with known concentrations using the SQX analysis, obtain an effective
diameter that leads to accurate analyzed values, and set that diameter.
Since filters such as the UltraCarry, UltraCarryLight and MicroCarry consist of C, O and H,
C and O cannot be analyzed.
When setting <Other> for the filter type, obtain beforehand suitable weight and sampling
diameter using various standard liquids, and set them.
When a gel sample such as grease is to be analyzed using the SQX analysis (FP method)
after sample processing using the filter paper method, use the Film & Coating for the
sample model.
SampleType - Liquid
1. When diluent is to be used, set information on it. Select <Used> and a diluent
component and set sample weight and diluent weight.
2. When a sample film is to be used, set information on it. This setting is used to correct
the absorption of X-rays by the sample film.
3. When a sample thickness affects analyzed values, make setting for “Sample weight
thickness”. When <Input for each sample> is selected, a sample size is set for each
sample at the time of an analysis.
When diluent is set, analyzed values are calculated with the absorption of X-rays by the
diluent taken into consideration.
Make setting for “Sample weight thickness” when fluorescent X-rays with short
wavelengths are to be measured and X-ray intensities are to affect sample thicknesses.
When a weight thickness varies for each sample, select <Input for each sample>.
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “Sample Film
Information Registration” regarding details of sample film. (This is the function of ZSX
Primus IV/IVi only, and applicable to the equipment shipped newly.)
1. When a measurement is made using a sample film, set information on it. The
correction of absorption by a sample film is made at the time of the quantitative
calculation.
Measuring Condition
Select a measuring diameter according to the sample mask diameter of a sample holder.
When the sample spin is set, a sample rotates during a measurement. In the case of a
sample that consists of a single crystal or has segregation, select <No> for the sample spin
because a sample rotates during a 2-theta scan and therefore an abnormal peak may
appear. Liquid samples should not be rotated.
When the sealing (option) is in the “IN“ position because the atmosphere is Helium/Vacuum
or when the attenuator is used, only 30mm can be set for the measuring diameter.
When SQX scattering FP method is used and the partition is not used, only 30mm and
20mm of measurement diameter is selectable. (This function is only for ZSX Primus IV/IVi
and applicable only for the equipment shipped newly.)
In the case of the helium condition, because the atmosphere absorbs X-rays, components
that have atomic numbers smaller than 9-fluorine cannot be measured.
In the <Helium/Vacuum> mode, the sample chamber is set in the helium atmosphere and
the spectroscopic chamber in a vacuum.
For the qualitative application, it is possible to set individually the selection of APC ON/OFF,
evacuation speed Slow/Fast, vacuum degree High/Intermediate/Low of inlet sample
chamber. (This function is only for ZSX Primus IV/IVi and applicable only for the equipment
shipped newly.)
Sample Information
This information cannot be set for an application of the sample type “Qualitative”.
Component Type
The metal, oxide or compound table can be used for the conversion method.
The compound table can be created using the utility menu. For details, see “6.1 b.
Compound Table”.
Metal
As in the case of a metal sample, a quantification calculation is made in the SQX
calculation assuming that a simple substance (for example, Fe or Si) exists in a sample.
Oxide
A quantification calculation is made in the SQX calculation assuming that a standard
oxide such as Fe2O3 or SiO2 exists in a sample. Select this conversion for a pressed
sample or a glass fusion bead sample made of oxide powder.
Compound Table
A quantification calculation is made in the SQX calculation assuming that each element
exists in a sample in a form defined in the compound table prepared by a user.
For example, in case of sample which is mixture of components without oxygen like metal
or sulfide and oxide like SiO2, element or compound can be set in each element of
chemical compound list beforehand.
Balance
This item is not displayed when the sample type is “Filter”, “Film & Coating” or
“Qualitative”.
This item is generally set when a component with the maximum concentration is to be
defined as the balance component for the SQX calculation.
An element or a compound can be set as the balance component. However, in the case of
a compound, it must be registered beforehand as a compound.
When a check mark is placed for “Estimate non-measuring component” on “Select
Element Range”, <Estimate> is set for the balance.
The balance component is treated in the concentration calculation as the residual, which is
100% minus the total of the other components.
Film Setting
This item can be set only when the sample type is “Film & Coating”.
Set this detailed information when analyzing a coating layer on a thin film or a
substrate.
Regarding substrate setting, fixed values can be set for base substrate and number of
layers under the top layer for analysis, and the analysis of the top layer becomes possible.
(This is the function of ZSX Primus IV/IVi only)
The SQX result of multi layers under the analyzed layer (top layer) including substrate can
be imported. By substrate settings with fixed values for the base substrate and the layers
below the top layer for analysis, the analysis of the top layer makes possible. (This is the
function of ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
Fixed/Input Components
When there is a fixed value or a component value to be entered manually at the time of
an analysis, set information in the “Fixed/Input Components” dialog that appears when
the Setting... button is clicked.
When <Fixed> is set for “Type”, set information to enter an analyzed value manually.
Calculation Setting
When a matching library is to be used for the SQX calculation, make the following
setting:
Correction Setting
Setting contents for the sample film in the sample preparation information dialog are
displayed for the sample film absorption correction, and a correction is made for the
absorption of fluorescent X-rays by the sample film.
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “PROCEDURE FOR
ADDITION OF SAMPLE FILM INFORMATION” regarding details of sample film. (This is
the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
Measurement is generally made in order of wavelength (from the shortest to the longest)
to carry out an automatic identification analysis and to shorten the analysis time. The
measuring order from the shortest wavelength has been set as the default. When you
want to know the qualitative analysis result of a certain component first, you can change
the measuring order in a selected measuring range.
Because a fixed angle analysis is made immediately after the qualitative analysis of that
component, the measuring order of fixed angle analyses components cannot be changed.
To reset measurement in order of wavelength (from the shortest to the longest), click
Wavelength.
Tube wattage: For all measuring conditions keeping tube voltages (kV) related library
conditions unchanged, tube current (mA) is changed based on the designated wattage
(1.5kW/2.4kW/3kW).
Protection filter: Primary X-ray filters equipped are arranged for the measurement line
automatically to prioritize the protection of X-ray tube window. In case of
specifications without Be30 filter, no filter is set to lighter element range
than Cl-Ka.
Time : Continuous scan speeds are changed at the time. The change can be made
easily while checking total measuring time.
(This is the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped
newly.)
Measuring Condition
The measuring condition of a spectral line to be selected must have been created
beforehand using a library.
The library measuring conditions are displayed for the items that cannot be changed such
as tube voltage. When <Qualitative> is selected for application type, these condition can
be changed.
For example, when you want to make an SQX calculation using a special spectral line,
you can select a library. For example, when it is preferable from the viewpoint of
analysis precision that the K line be changed to the L line in the case of samples thinner
than the bulk thickness, make the selection here. However, the measuring condition of a
spectral line must have been created beforehand using a library.
To carry out measurement when X-ray intensities exceed the upper limit of counting
linearity, use the attenuator or reduce the current. The pre-measurement function,
which will be described later, can also be used to set the attenuator or change the
current automatically for measurement. The upper limit of X-ray intensities for the
qualitative analysis is in general 1000 kcps for both the SC and the PC.
The step scan or the continuous scan can be set for the scan method. In general, the step
scan is used for the qualitative analysis. A scan speed that is not available in the step
scan can be set in the continuous scan, which is effective especially for a high-speed
scan.
A scan range can be widened. Care must be taken when narrowing the range because
the SQX calculation may be affected.
A range of 5 to 118 degrees can be measured for the SC and a range of 13 to 148 degrees
for the PC.
In the application by SQX Scatter FP method (Option), the 2-theta angle of Start and End in
“2 theta scanning condition” of <Scatter> cannot be changed. (This is limited to the
specifications of which scattering sensitivity library contains LiF(220) in it.)
In general, when changing the measuring time, the time for each step is changed.
Because the time for each step is limited, the step size itself may be changed. In that
case, special care must be taken because it may affect the SQX calculation. In the case of
the continuous scan, the speed (degrees per minute) is changed.
Intensity
Adjustment Description
Method
Pre-measurement is not made, so measures against counting
No
loss are not taken.
Pre-measurement is made at the peak angle set in the 2-theta
Pre-measurement
scan condition. Whether counting loss occurs is checked at the
at peak
peak position.
Pre-measurement is made in the angle range set with the start
and end angles in the 2-theta scan condition. If there is a peak
Pre-scan
that causes counting loss in that range, the system lowers the
sensitivity and makes measurement again.
The system makes main measurement without making
pre-measurement, automatically distinguishes the range where
High speed counting loss occurred directly from the main measurement
(ZSX Primus IV) result. When that range is detected, sensibility is automatically
lowered and measurement is made again. Use this method when
the throughput in the qualitative analysis is needed.
To adjust the sensitivity automatically, the attenuator is used or the tube current is reduced,
depending on the system configuration.
Because the fixed angle analysis uses the same optical system as the qualitative
analysis for measurement, fixed angle measurement conditions are the same as
qualitative analysis conditions. In addition, since the fixed angle analysis is made
immediately after the qualitative analysis, background positions are automatically set.
Therefore, only the peak position and measuring time can be set as measuring
conditions exclusive for the fixed angle analysis.
1. Select a fixed angle component that has a FA mark as its spectral line name.
2. Click Property .
3. Select an angle for measurement for “Peak angle”.
4. Set measuring times in the peak and background angle positions for “Time”.
5. Click OK .
A peak angle in the qualitative analysis is obtained automatically. Therefore, if spectral lines
overlap, the peak angle shifts. Taking that into consideration, in the fixed angle analysis
X-rays are counted in the designated peak position.
Background intensities are measured in each position on both sides of an objective peak.
To enhance measurement accuracy, the designated peak and BG angles of fixed angle
measuring condition can be changed. Release the check box of <auto>, then input peak
angle and BG angle. (This function is applied depending on ZSX software version.)
f. Output Condition
Items to be printed just after measurement with the SQX or qualitative application are
selected. Selection can be made for the chart output, peak list, SQX calculation results
(option) and material judgment (option). All items can be printed for an SQX analysis
application. For a qualitative analysis application, selection can be made only for the
chart output and peak list.
Print
Description
Information
A qualitative analysis chart measured using the designated
Chart Output
measuring conditions is printed.
Peak List A list of detected peaks is printed.
Semi quantitative analysis values normalized so that the
SQX Result total of the concentrations of detected components will be
100% are printed.
Based on an SQX result, a material judgment result
Material
obtained from the condition set in the material standard
Judgment
table is printed.
SQX Application
Qualitative Application
In the qualitative data processing, setting is made for the element/oxide conversion, the
identification elements, the smoothing and the peak detection sensitivity. Those are
described below:
Element/Oxide Conversion
1. Place a check mark for the “Element/Oxide conv.” check box, and click Setting .
2. Click an element to be converted on the periodic table. The setting is switched to an
element or to an oxide each time an element is clicked.
3. Click OK .
For example, trace heavy elements in rock generally exist as oxides in a sample. To output
them as the concentrations of elements, specify the conversion of these trace heavy
elements into elements.
Identification Elements
Smoothing
This is the function of smoothing processing for a qualitative spectrum. As the number of
points used for smoothing becomes greater, the chart becomes smoother. However, if the
number of points is too large, the spectrum will be deformed. Because smoothing can
also be carried out after measurement, it is recommended that the default value be used.
There must be an odd number of points used for smoothing.
The peak detection sensitivity is the value to determine the sharpness of peak detection
in the automatic peak search. To detect a peak, the threshold value is used to judge
whether a peak on the spectrum is a real peak or a fluctuation of X-ray intensities. The
peak detection sensitivity coefficient is the indicator to change that threshold and its
standard value is 1.0. As the value becomes larger than 1.0 (the sensitivity becomes
higher), smaller peaks can be detected. As the value becomes smaller (the sensitivity
becomes lower), only peaks with higher peak/background ratios are detected. The range
allowed for the peak detection sensitivity is 0.3 to 2.0.
1. Set the number of smoothing points using [Calculation] [Smoothing] in the
Qualitative Data Handling program.
2. Set a peak detection sensitivity coefficient using [Calculation] [Peak Search] in
the Qualitative Data Handling program.
Elements that were not detected in the fixed angle measurement can be added to a
group of output elements.
The Qual Result program reprocesses qualitative analysis and SQX analysis results and
consists of the qualitative data handling process and SQX calculation.
SQX Calculation:
In the SQX calculation, though a calculated result is automatically displayed after
measurement, a recalculation can be made after detailed setting. For example, a
recalculation can be made after adding a new peak using the Qualitative Data Handling
program. And a calculation condition can be changed. For example, the metal conversion
can be changed to the oxide conversion. The standard unit of concentrations used here is
mass%, but this can also be changed.
When an analyzed peak overlaps with another interfering line, corrected intensities are
calculated using overlap correction coefficients that have been registered beforehand
and then analyzed values are calculated.
In the SQX calculation, a collection of spectrometer sensitivity values called a common
library is used for a quantification calculation. Though the library registered on
shipping is generally used, a calculation can also be made using a matching library. For
further details, see “3.4 SENSITIVITY LIBRARY”.
When the logon user (or the group to whom he/she belongs) does not have “Change
authority of Analysis Result”, it is impossible to save a changed result. When another user’s
analyzed result folder without “Permit changing by other users” is displayed, it is impossible
to save a changed result. In those cases, [Read Only] is displayed in the title bar.
A measured result file consists of charts obtained with the measuring conditions of two
or more components. In the example shown below, a file consists of a full range
qualitative analysis result of heavy elements of Ti to U and individual qualitative
analysis results of light elements such as Ca and K.
An SQX calculation is made after qualitative data handling is complete for all measured
results.
Heavy Ca K
2 Angle
In the measuring range of heavy elements, one crystal covers an element range of Ti to
Cm(U). That is possible because wavelengths of components are adjacent and the
resolution of the crystal is high.
In the range of light elements, unlike the range of heavy elements described above,
because wavelengths of components are dispersed and the resolution of the crystal is low,
the entire wavelength range cannot be measured at a time. The individual qualitative
analysis is therefore used.
when EZ Scan based on the film & coating model is performed, long wavelength
X-rays of heavy element (L line, M line) become detectable. (This is the function for
ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
Select first a measured result file for an identification analysis. In the ZSX Guidance,
two or more measuring condition graphs can be displayed simultaneously. Two or more
files can be opened simultaneously and overlaid as reference data. In that case, only one
file is an object of processing and the other files are not changed. The ordinary procedure
is to terminate the data handling of one file, proceed to the next file and then finally
compare the data items by overlaying them.
Start
Identification analysis
The procedure for peak identification in the qualitative data handling is shown below.
According to this procedure, all peaks on the spectral chart are identified. In general, the
results of heavy elements are identified first. Heavy elements generate X-rays with long
wavelengths from a sample as well as spectral lines ordinarily used for analyses.
Therefore, in an analysis of light elements, it is effective to make an identification
analysis using the analyzed results of the heavy elements.
Start
End
Peaks from the X-ray tube include Thomson scattering rays such as the K, L and M lines,
Compton scattering rays and their higher-order rays.
Pair peaks mean peaks in pairs such as the K and K1 lines and the L and L1 lines.
A file of qualitative analysis results is opened to display a chart of each measured result.
On the displayed chart, data processing items such as the identification analysis and the
net intensity update can be executed.
A measurement result file saved in the selected folder can also be selected on the
dropdown list on the left of Browse... .
When the program is started, the “Target File Selection” screen appears.
a. Marker Setting
A combination of an element name and an X-ray line name is called a marker. A function
to attach a marker to an identified peak is called a marker setting. Basically, a marker
setting is carried out automatically after a qualitative analysis is complete. But in some
cases editing operations are necessary to identify an unknown peak, to delete an
unnecessary peak, and so on. Using this marker setting function, you can check and
modify automatically identified peaks.
To check an attached marker or to check an element identified with the 2-theta search,
follow the procedure described below. For example, when your system has the Rh tube,
scattered rays of the characteristic X-rays of the target material always enter the
detector. In that case, designate Rh as the element name and check its marker position.
X-ray lines that cannot be detected with the optical system used in the measuring condition
are eliminated from objects of searching.
When “All Lines” is selected for “Level”, minor line names are also displayed as well as
major line names.
“Relative int.” represents the relative intensities of lines in the same series. For example,
KA 150 in the K line corresponds to LA 100 in the L line.
The scale can be changed with the marker editing dialog displayed.
When there is a peak without a marker on the chart, you can attach a marker using the
peak angle. For example, let’s assume that there is a peak without a marker near a
2-theta angle of 57.5 degrees with the measuring condition “Heavy”. You can search for a
marker from that 2-theta angle and identify the peak.
When “All Lines” is selected for “Level”, minor lines are displayed as well as major lines.
When “Higher orders” is checked, the higher-order lines of the designated line are also
retrieved.
When <Identified elements> is selected, only the lines of the elements identified in this file
are displayed. This function can be used to retrieve the higher-order lines of heavy
elements and lines with long wavelengths in the light element range.
The scale can be changed with the marker editing dialog displayed.
Delete Marker
Markers that were attached erroneously and unnecessary markers can be deleted.
b. Data Processing
Procedures for data processing such as the background fitting and the peak
deconvolution are described below. Although X-ray intensity calculation has been made
automatically after a qualitative analysis, operations for an SQX calculation are carried
out here. The purpose of these procedures is basically to update net intensities.
Background Fitting
This method calculates a background for the entire 2-theta angle range measured with
the measuring condition, and uses the calculated background to subtract the intensity of
the continuous X-rays.
Calculated background intensities can be subtracted from the original spectrum. See the
“Subtract” menu in “Calculation”.
Background intensities of a designated range are subtracted. The point designation and
the range designation can be used. A straight line is generally used for fitting. But when
the objective peak overlaps with another peak or when the background is oblique,
another function is used. For the selection of background processing, see “4.3
QUANTITATIVE ANALYSIS TECHNIQUES”.
Point Designation
Range Designation
After entire range background fitting, that background is subtracted. This function is
valid only just after a background calculation.
Peak Deconvolution
When an objective peak overlaps with a peak of another element, it may be necessary to
deconvolution the two peaks. The function deconvolution and the standard profile
deconvolution can be used for peak deconvolution. The function deconvolution separates
a waveform regarding it as a single shape. That is, a calculation is made regarding a
peak width and the shape of a peak foot as constant. On the other hand, the standard
profile deconvolution separates a peak using a spectrum (standard profile) without
overlap that has been extracted beforehand. The standard profile deconvolution can
therefore carry out separation more accurately taking each peak shape into
consideration.
The 2-theta adjustment function is used when peaks of light elements shift due to the
sample.
To execute the standard profile peak deconvolution, standard profiles must have been
determined beforehand. Using pure substances such as metals and reagents,
measurement without counting loss or overlap must be made. It is useful for calculations
that will be made later to set a wider extraction range.
The peak list function is used to check net intensities and background intensities of a
processed component. A peak displayed on that list can also be deleted. The following
items are displayed:
When you want to display fixed angle measurement data (FA) and overlap line detection
data (OV), click Fixed angle and Overlapped lines .
It is possible to copy displayed data to the clipboard. Press the Ctrl+C key to copy data to
the clipboard, and press the Ctrl+V key to paste data to the spreadsheet software Excel (R)
or the like.
Background angles are not output at the time of the printing of a peak list. If necessary,
output them via the clipboard.
FA or OV information is not printed at the time of the printing of a peak list. If necessary,
output them via the clipboard.
c. Calculation
Various calculations such as smoothing and peak search are described below. Because
the peak search and the automatic identification are carried out automatically after a
qualitative analysis, they generally need not be executed.
Smoothing
Smoothing is applied to a measured spectrum when the spectrum is not smooth and
shows significant statistical errors. Take care because a spectrum can be deformed if the
number of smoothing points is too large.
Peak Search
This function deletes markers attached automatically to peaks and runs a new peak
search. Since this operation is executed automatically after a qualitative analysis, use
this manual procedure to search for peaks again after changing detection sensitivity.
As the sensitivity value becomes smaller, only peaks with higher peak/background ratios
are detected.
This function clears all spectral line names. Peak positions alone are checked again. Using
the automatic identification, spectral line names can be reattached.
Automatic Identification
This function performs an identification analysis automatically after a peak search with
different sensitivity. When there are special components that require special attention,
they can be selected or eliminated here from objects of identification.
The peak width of fluorescent X-rays is determined by an element and line type. By
knowing beforehand the width of a certain peak, overlap of peaks can be checked. Using
that peak width, angular resolution is assessed.
In general, the abscissa axis of a chart is displayed in the 2-theta angle. In that case, for
example, when one wants to compare the spectra of the same element using different
crystals, comparison by overlapping the spectra is impossible because the abscissas are
expressed in the 2-theta angle. In such a case, the abscissas can be displayed in the
wavelength () or the energy (E).
Once the axis is changed from the 2-theta angle to the wavelength, it cannot be returned to
the 2-theta. It is recommended that <Add> be selected.
Normalization can be used when at least one reference file has been designated. This
function normalizes intensities using the spectrum of the reference file. The point
designation and the area designation can be selected.
Point Designation
Area Designation
The menu of [Normalize…] can be selected when only one chart window is displayed.
Reset Normalization
This function calculates intensity differences using data in a reference file. For example,
data of filter paper before dripping is set as a blank sample in a reference file and that
data is subtracted from data of filter paper after dripping.
d. Scale Change
To edit markers and process the data of the displayed spectrum, the procedure described
below can be used to change the displayed area to make the process easier:
Expand Display
This function is used, for example, to expand part of a full range qualitative analysis
chart and to check background.
To change the scale to the one designated in the measuring condition, click Default.
Display Legend
In the case of the overlapped drawing of two or more files, the legend is displayed by
default.
The full scale of a chart is adjusted according to the peak top of an objective element. A
peak higher than that of the objective element is displayed off scale. Since that peak top
cannot be checked, marker setting is impossible. In such a case, the vertical full scale
can be set according to the highest intensity in the displayed range.
1. Click .
When changing the chart display area, it can be scrolled horizontally and continuously
using the procedure described below instead of using the common tool bar:
1. Click the icon on the tool bar, and the cursor will change to a right and left
arrow.
2. Drag the mouse right or left and hold it. The area scrolls continuously in the
direction of dragging.
In processing a full range qualitative analysis result, the measured range is so wide that
marker setting and data processing may be difficult. In such a case, the scale change
operation is useful. After an ordinary scale change, a selected part is displayed, and you
cannot see the original wide range of the measured data. For example, to edit the
markers, it may be difficult to check pair peaks. In that case, a wide-range screen and a
smaller region processing screen can be displayed simultaneously. That wide-range
screen is called the panorama. When you set a scale range to be changed on that
panorama window, the scale of the processing window is changed. The scale of the
panorama window is not changed.
When the full scale is adjusted to a high-intensity peak, you cannot see small peaks.
When the full scale is adjusted to a small peak, high peaks are displayed off scale. In
such a case, by displaying an expanded view of a small peak in the same window, both
can be displayed in optimum full scales. That operation is called the zoom. The
procedure shown below would print the window with the expanded view displayed:
3. Select the specific section of the chart to enlarge by dragging a box over it to the scale
line.
The selected area will be enlarged and displayed in the chart.
Because the display cannot be saved in a file, print a hard copy of the screen or use the
“Active Chart” printing. (See the last part in “h. Printing of Qualitative Chart”.)
When a scale change or the like is executed, the zoom setting is canceled.
Set the degree of the filling of an area below a line that indicates a spectrum when a
spectral graph is displayed.
When <0> is selected, the area is filled with a selected color. When <100> is selected,
only the line of a spectrum is displayed. When an absorption spectrum is displayed or in
a similar case, changing the transparency makes comparison easier.
When a spectrum is printed on the printer, only the line of the spectrum is output
(transparency: 100%) regardless of this style setting.
This function changes styles of spectrum display such as colors and line widths.
Functions to compare charts by overlapping two or more data items and to add or delete
measuring conditions are described below. Because qualitative measurement data is
generally used for the SQX calculation described later, it is recommended that measured
results necessary for the calculation that have been processed be saved.
Overlapped drawing of two or more files and overlapped drawing in one file can be used.
Up to 12 measuring conditions can be overlapped. For both drawing methods,
overlapped drawing is impossible when conditions such as the number of steps and the
units of the abscissa are not identical.
This function is used, for example, to overlap measured results of a normal sample and
of an abnormal sample. Up to 11 overlapping files can be selected. The types of
applications used can be different. However, results of reference files with measuring
conditions different from that of the target file can not be overlaid.
For example, the crystals LiF and Ge can be used for the Ca-K line. Because the
2-theta angle is different for each crystal, the abscissa axis is converted beforehand to
wavelengths. By overlapping those data items, comparison can be made to determine
which profile has higher intensities or better resolution, for example.
This function is used, for example, to delete components of B to O from a file in which a
qualitative analysis was made for B to U and to change that file to a qualitative file of F
to U. Care must be taken because deleted components cannot be restored.
For example, let’s assume that you made a full range qualitative analysis and want to
measure a component again with a slightly widened measuring range. With a
conventional system, all components must be measured again with a loss of time. In
such a case, to shorten the measuring time, only a changed measuring condition can be
measured again and the result can be added to the original file. Because a measuring
condition in a reference file is copied, it is not deleted.
An added component is placed at the bottom of the measuring condition list of the target
file.
When <Use added spectrum for SQX analysis> is selected, the identification result of a
spectrum added to the target file can be used for an SQX analysis.
Three methods can be used to print qualitative charts: printing of all qualitative charts,
printing of selected components and printing of charts that are selected as objects of
processing (described as “active” hereafter). In all cases, one file is processed. Printing of
all qualitative charts can be executed even when files have not been selected with the
Qualitative Data Handling program. Common setting items and individual setting
items for each printing method are shown below:
Those items are set beforehand using the print setting function, but some of them can be
changed on printing.
Set Header
Character strings set as the header are printed in the upper part of a printing sheet. The
header is divided into the left, center and right areas and an arbitrary character string
can be set for each of them. In the initial setting the common header has been set for
each chart, but individual headers can also be set.
To print a chart in the All Spectra mode, the scales of all qualitative charts can be set.
For the individual scale setting, each spectrum is displayed in its full scale reflecting
each result measured with each measuring condition. In the case of the common scale
setting, the full scale is set according to the highest peak in the measuring conditions.
The default setting is the individual scale. Setting can also be made for the printout of
file names and measurement dates and times. Because those cannot be set on printing,
they must be set here beforehand. In the cases of Selected Spectra and Active Chart,
those printouts cannot be set.
In the All Spectra mode, all measured results are printed in one sheet.
In the Selected Spectra mode, up to 9 graphs in the case of Portrait or up to 10
graphs in the case of Landscape can be printed.
In the Active Chart mode, one graph is printed in one sheet.
In the All Spectra mode, Landscape and Portrait can be set for the printing direction.
Each printing image is shown below. Basically, all charts are printed on one sheet.
Partial Qual.
Chart
Full Range
Chart
Portrait Landscape
In the Landscape printout, the full range qualitative analysis chart is located in the
lower part and partial qualitative analysis charts are located in the upper part. In the
Portrait printout, the full range qualitative analysis chart is located in the lower part
and partial qualitative analysis charts are located in the upper and middle parts. The
number of those rows cannot be set, but the number of divisions in each row can be
preset. When the number of measuring conditions is larger than that preset number of
divisions, a second page is used.
Printing
Setting for printing in the All Spectra mode can be made when the Qualitative Data
Handling program is started. Setting for printing in the Selected Spectra and Active
Chart modes can be made when a qualitative analysis file is opened or when a chart is
displayed. Though the printout format is basically set using [Print Setting...], it can be
changed on printing. The printer, bit map file or a metafile can be used as the output
device for printing.
Printer
The printer is ordinarily used as the output device for printing. When your system is
equipped with a color printer, a qualitative analysis chart displayed on the screen can be
printed.
A displayed image is filed and can be pasted using word processor software etc. When
the image size is changed, however, the image may be deformed. Typically the file size is
large.
Metafile
This is an image file in which a method to create an image is recorded and can be pasted
using word processor software etc. that can read metafiles. Even when the image size is
changed, the image is not deformed. The file size is smaller than that of a bit map file.
In the case of printing in the All Spectra mode, the output device, paper direction,
header setting, etc., can be changed. Only in this mode can printing be executed without
displaying a chart using the Qualitative Data Handling program. In addition, the
components that need not be printed can be selected and the method to display the scale
can be changed only through the All Spectra mode.
In the All Spectra mode, one sheet is basically used to display all measuring conditions.
Components that should not be printed on the same page can be printed on another page,
or settings can be made so as not to print them at all.
Only in the case of the All Spectra mode, the scale mode can be changed.
This <Selected Spectra> mode is valid when a target file and a reference file are selected.
Overlapped charts are output in the same image as the one on the screen. The number of
graphs that can be printed simultaneously varies depending on the paper direction. The
output order can be designated. In the case of “Vertical”, graphs are laid out in the
vertical order. In the case of “Horizontal”, graphs are laid out in the horizontal order.
When a chart/charts is/are displayed, the chart(s) in the active condition can be printed.
This function can be used, for example, to print a chart/charts after background fitting
or zooming. Needless to say, overlapped charts can be output intact.
Qualitative analysis results that can be transferred are qualitative element information,
an SQX analysis result, a peak list and 2-theta - X-ray intensities. The transmission
format and the spreadsheet format can be used as the text format. The disk saving and
the external transmission can be used as the transfer method. The text formats and
transfer methods that can be selected vary depending on the data type.
Objects of Transmission
Data Type Contents
Qualitative element Detected elements and the analyzed values or
information X-ray intensities
SQX analysis result Components and analyzed results
Peak list Element lines and net intensities
2-theta - X-ray 2-theta scan angle and X-ray intensity for each
intensities spectral line
In qualitative analysis information, analyzed values are output when SQX analysis result is
obtained. When SQX analysis result is not obtained in such a case that only qualitative
analysis has been done, X-ray intensities are output.
More than one qualitative analysis result file can be selected for qualitative element
information and an SQX analysis result. For a peak list and 2-theta - X-ray intensities,
select one file and then select spectral lines to be transferred.
The procedure to save an SQX analysis result into the disk is used as an example for
explanation.
When <Peak list> or <2-theta - intensities> is selected in the “Data type” frame, the
“Spectrum Selection” screen appears after clicking OK.
When <2-theta - intensities> is selected in the “Data type” frame, “Text format” can be
selected from <Transmission format> and <Spreadsheet format>. And the format can be
selected from <CSV> and <TAB>.
To transfer <Peak list> or <2-theta - intensities>, select spectral lines after the dialog on
the “Qual Data Transmission” screen.
To return to the dialog on the “Qual Data Transmission” screen, click Cancel.
A measurement result file saved in the selected folder can also be selected on the
dropdown list on the left of Browse... .
The unit ppm can be used for a component with a specified concentration or less in SQX
calculation results by setting system parameters in <6.1.h System Management>.
The contents of display on the screen and printing output can be set.
2. In the “Result Output Selection” dialog, place check marks for items to be displayed
or printed.
Up to four items can be printed.
3. Click OK .
Component names, analyzed values and units in final results are unconditionally
displayed and printed.
Item Contents
Detection limit after taking account of the
Det. limit
background subtraction.
Analysis line type used for an analyzed value
El. line
calculation.
Intensity Net X-ray intensity.
The item “w/o normal” means an analyzed value
before normalization to 100%. The item “liquid
w/o normal/liquid conc.
conc.” means an analyzed value converted to that of
original liquid in the case of a filter sample.
Depth from a sample surface at which an analysis
line is measured. When a sample is thinner than
Analyzing depth (mm) this thickness, the sample thickness affects an
analysis value. In this case, set a sample thickness
using the sample size setting function.
Intensity ratio of analysis line to the intensity of a
sample that has an obtained constitution and an
infinite thickness. This item is valid only when a
Intensity ratio sample thickness is set. When this value is 1.0, an
intensity is the same as that of a sample with an
infinite thickness and an analyzed value is not
affected by a thickness.
Component name for a quantification calculation
Component (Org) before conversion when the element/oxide
conversion is specified.
Quantification calculation result before conversion
Result (Org)
when the element/oxide conversion is specified.
When a new marker was added or net intensities were updated using the Qualitative
Data Handling program, a recalculation can be made. The SQX calculation is made
using a result processed by Qualitative Data Handling.
In the SQX calculation, the metal (elemental) conversion, the oxide conversion or the
compound table is set for the component type. In general, the metal conversion is set for
metal samples, residual components in polymer, etc., and the oxide conversion is set for
powders, glass, etc. The compound table is used to make a quantification calculation
using a combination of compounds set by a user.
Two methods can be used for a calculation using a combination of metals and oxides.
One is to specify a compound table that has been created beforehand. The other is to
change component names on the “Component Condition” screen.
In the case of rock samples, for example, an SQX calculation may be made using the
oxide conversion for main components and the metal conversion for minor components.
In such a case, component names can be changed on the “Component Condition” screen.
When a compound table is to be used, select it for the component type. Specify each
component on the “SQX Calculation” screen according to the following procedure:
For example, when analyzing polyethylene, the compound CH2, the main component, is
not measured and is treated as the balance. In a calculation, components other than
measured ones are treated as the residual (balance component). In the SQX calculation,
that balance component can also be set. When the sample model is a thin film, the
balance component must be set. Only one balance component can be set.
If binder is used when pressing a powder sample, binder must be set because it affects
analyzed results.
This setting is possible when the sample type is the powder or polymer.
To analyze a fusion bead and obtain analyzed values, a flux type and a dilution ratio
must be set because a sample is diluted with flux.
4. When oxidizer is used for fusion, select <Used> and the compound name of the
oxidizer, and enter added oxidizer weight
5. Click OK .
When the sample type is the powder, polymer or fusion bead and a sample thickness
affects analyzed values in the case of some analysis line, a quantification calculation
with the weight thickness of a sample taken into consideration becomes possible by
setting a sample size.
Elements that were not detected in the fixed angle measurement can be added to a
group of output elements.
1. Select <Estimate> for “Balance”. The option <Estimate> is displayed only for the
result file of measurements using a qualitative application created with a check
mark placed for “Estimate non-measuring component”.
2. Select a scattering ray library name to be used for the SQX scattering ray calculation
from the pull-down list. (In case of the specifications of scattering library with
LiF(200), this is done automatically.)
3. Click Setting... for “Sample size” and enter sample size information.
4. Click OK .
It is possible to operate <Estimate non-measuring component> only for the result files of
qualitative analysis measurement, of which application is to be prepared by SQX Scattering
FP method.
In case of the specifications using scattering library with LiF(200), the selection of “Estimate
non-measuring component” is possible for the analysis of measurement diameter 30mm
and 20mm qualitative analysis result file. (This is applicable to the sample model of powder,
polymer and liquid only.)
Filters, films, coating samples, etc. are treated as thin films. A thin film sample coated
with a metal film or the like is treated as a film sample with a substrate. Solution on
dripping paper is treated as a filter sample.
The procedure for the SQX calculation for a film sample without a substrate is used as
an example for explanation.
1. Place a check mark for “Substrate” in the “Advanced Setting” dialog, and push the
Setting... button.
When Rigaku’s filter such as the UltraCarry, UltraCarryLight or MicroCarry is selected for
the filter type, a material, weight and sampling diameter are automatically displayed.
The SQX calculation uses all components detected in a qualitative analysis. If, for
example, Cl detected in measurement of a metal sample can be judged evidently to be a
contaminant, it can be excluded from the calculation.
Since more than one component cannot be selected, select one by one.
Let’s assume that Co in the heavy element region is to be measured with the full range
qualitative analysis. When the main component is Fe, the ordinary measured element
line Co-K overlaps with Fe-K1. In that case, Co-K1, which does not overlap, is
generally used as the measured element line for a calculation. Thus, when one
component has two or more measured element lines, selection can be made from them
for a calculation.
5. Click OK .
Add Component
Even a component the peak of which is not detected can be added when overlap line
detection data is displayed.
It is possible to display the spectrum of each component directly on the SQX calculation
screen, enabling the checking of spectra used for the SQX calculation.
Because it is impossible to display the spectra of two or more components, select one
component at a time.
A chart in the Heavy range displays the range of +/-5 deg. to the angle of an analysis line.
This function is used to set information about a component with a known concentration
as a fixed value. Two or more fixed values can be set.
A component name other than an element name must have been set beforehand in the
compound table.
Unlike the common library, a library registered using standard samples for each kind of
unknown is called the matching library. When a matching library that has
concentrations similar to those of unknown samples is available, a more accurate SQX
calculation can be made. However, a matching library can be used only for the bulk
model and cannot be used for the thin film model.
The impurity correction, sample film correction and helium atmosphere correction can
be used. The sample film correction is automatically made when the sample film setting
described below has been carried out:
<Helium atmosphere correction> can be selected when the atmosphere specified in the
application is Helium.
For example, if filter paper itself contains a certain component, that component must be
subtracted when calculating concentrations of unknown samples deposited onto the
filter. In that case, the impurity correction can be carried out by executing a qualitative
analysis of filter paper itself and registering X-ray intensities. X-ray intensities or
analyzed values can then be subtracted.
Since a sample film absorbs X-rays generated from a sample, a correction must be made.
When the sample type is not “fusion bead”, a sample film can be set on the “Advanced
Setting” screen.
When the helium atmosphere is selected, the atmosphere absorbs X-rays. That
absorption has a greater effect on lighter elements than heavier ones. By using this
correction, library samples need not be measured in the helium atmosphere. This setting
is possible only when the helium atmosphere has been selected.
Model Unit
Bulk and Thin Film (Except Filter) Unit preset for analysis value
Thin Film (Filter) Unit preset for liquid concentration
b. Impurity Data
The procedure to register, change and delete impurity data used for the SQX calculation
is described below:
The registered contents of impurity data can be changed. A component that is not used
for calculations can be deleted.
c. Printing
The procedure to print an SQX calculation result is described below. The procedure to
change the output order of components and their output number of digits is also
described.
Print Result
The mode to print in order of an atomic number (from the smallest to largest) and the
mode to print in order of an analyzed value (from the largest to smallest) can be selected.
The number of digits of a printed calculation result can be set. For example, when an
analyzed result is 12.3456, it is output as shown below:
When the reliability of analyzed result is doubtful, setting can be made so that the value
will not be output. The output is called the trace output. For a component with a
concentration lower than detection limit in bulk samples, “Trace” is output instead of the
analyzed value.
It is possible to judge the material of the sample in the SQX Calculation program. To
judge material, material judgment standards registered using <Material Judgement
Standards> are retrieved. Although material can be judged on making an SQX analysis,
it can be judged again here.
In the SQX calculation, an overlap correction is made for interfering lines using
theoretical intensity overlap correction coefficients that have been stored beforehand in
this system. The correction is made when the measured element line is a K line in the
K series or an L line in the L series. Only primary rays are taken into consideration as
interfering lines.
The overlap correction is not made for measured element lines after peak deconvolution
and for ones for which intensities have been entered using the SQX Analysis program.
When a measured element line is identified together with another interfering line, an
overlap correction is made by calculating net intensities at an angle calculated
theoretically.
The automatic overlap correction is used not only in the “SQX calculation” in the data
processing but also in the data processing in the “Analysis” program.
It is possible to export a CSV file of SQX analysis result that can be read using spread
sheet program.
Operate as follows when calculated result is displayed in the main window of SQX
calculation.
Data such as component, result, unit, detection limit, El. Line and intensity are output
as displayed.
In the case of the matching library, sensitivity values are registered using standard
samples for each kind of unknown material. In an SQX analysis calculation, a library
that has components similar to those of unknown samples is retrieved and more
accurate concentrations can be obtained.
It is also possible to register a matching library directly from the SQX calculation.
1. Click Browse... of the target file and select an SQX analysis result file to be
registered in a matching library.
2. Click the <SQX Calculation> tab to display an SQX result.
3. Select [File] [Save Matching Library...].
4. Enter and select the following items on the Save Matching Library screen:
5. Click OK . The matching library and sensibility values are automatically registered.
When entering standard values, use the concentrations of component names in an SQX
calculation result.
1. Click the qualitative analysis result of one point in SQX mapping analysis results.
2. Click Calc Condition... and change calculation conditions if necessary.
3. Select [File] [Lump calc]. The number of mapping measuring points and the point
number of the data being calculated are displayed on the screen during a calculation.
The recalculation result is automatically reflected in the mapping result and Result Display.
The function called the library drift correction is available to use libraries for a long term.
Libraries depend on the spectrometer conditions, which change over long periods of time.
To correct libraries using a small number of samples without remeasuring all library
samples is called the standardization of libraries.
Common Library
The common library has been registered using standard substances such as pure metals
and reagents without using standard samples for each kind of samples. By measuring
library samples for representative elements and saving sensitivity values beforehand,
the analyzed value of an element that is not registered in the library can be calculated
by estimating its sensitivity from those of elements that have atomic numbers before
and after that of the analyzed element.
The sensitivity library that Rigaku created by measuring many standard substances is
called the master common library. That master common library reproduced with many
library samples is called the common library.
This library can be used for both the bulk sample model and the thin film sample model in
qualitative or quantitative analysis.
In the case of the matching library, sensitivity values are registered using standard
samples for each kind of unknown material. In an SQX analysis calculation, a library
that has components similar to those of unknown samples is retrieved and more
accurate concentrations can be obtained.
A matching library that a user registers is called the user matching library.
This library can be used only for qualitative analysis and cannot be used for quantitative
analysis.
This library can be used only for bulk samples and cannot be used for thin films.
Like the drift correction in the quantitative analysis, calibration of X-ray intensities to
use created sensitivity libraries for a long term is called the standardization of libraries.
That operation is called the drift correction and its purpose is to re-establish libraries
using a small number of samples. This system has samples for the library
standardization, so it is recommended that setting be made for drift correction.
The scattering sensitivities used in SQX Scattering FP method (option) mean the
equipment sensitivities of Compton scattering X-rays (Rh-KC) and Thomson scattering
X-rays (Rh-KT), which originate from the target material (Rhodium) in X-ray tube.
The scattering sensitivity libraries are assorted on each sample type.
In the SQX analysis, quantification values are calculated using the common library and
matching libraries. In this case, the following three correction functions can be used
separately or in combination with each other:
Impurity Correction
Among other uses, this function is used to eliminate components contained in filter
paper for drip samples and impurity lines from the X-ray tube.
To analyze liquid samples, a film is used to protect the sample. This function is used to
correct absorption of X-rays by the film.
Atmosphere Correction
For samples like liquids that cannot be measured in the vacuum condition, the sample
chamber atmosphere is changed to the helium condition or the like. This function is used
to correct absorption of X-rays by the helium atmosphere.
The operation flow to register each type of library is shown below. The common library
uses standard substances and the matching library uses standard samples. Other than
that difference, the same procedure is used for both types of libraries.
Library drift corrections to calibrate X-ray intensities are registered using other
procedures. In the detailed descriptions below, the procedure for each purpose is
explained.
Start
Common or matching
library? No
Yes
Create measuring condition
End End
b. Measuring Condition
Because all the measuring conditions for Scattering sensitivity library (option) are fixed, the
user cannot edit them.
Library
Usually, the common sensitivity library named <Standard> has been set beforehand. In
general, because a library is spectrometer sensitivity, one measured element line has
one sensitivity value. For example, when the same component has different measured
element lines such as Cd-K and Cd-L, a library can have two or more components. To
create a new library of the same measured element line as that registered already in the
common library, an original library can be created.
A combination of those libraries can also be used.
The procedure to create a new original library other than the common library is
described below:
Measuring Conditions
The measuring conditions are classified into common conditions, qualitative conditions
and quantitative conditions.
Conditions Description
Common The excitation condition, optical condition and PHA condition are
Condition set.
Qualitative
The scan method, angle range, etc. are set.
Condition
Quantitative
The background subtraction, measuring time, etc. are set.
Condition
For the setting procedure for each condition, see “3.2 e. Setting of Measuring Conditions”,
“4.2 d. Setting of Analysis Information” and “4.3 a. Measuring Condition”.
For the detailed setting contents, see “3.2 e. Setting of Measuring Conditions”.
For the detailed setting contents, see “4.2 d. Setting of Analysis Information”.
c. Library Samples
Library samples are classified into ones for the common or matching library and drift
correction samples for intensity correction. The former require standard values to
calculate spectrometer sensitivity values. The latter do not require standard values
because they are used only for intensity correction.
Samples to recalibrate intensities are called library drift correction samples. If standard
samples are stable for a long term, they can also be used for library drift correction
samples.
This is the library sample to be used for obtaining the equipment sensitivities of
Compton scattered X-rays (Rh-KC) and Thomson scattered X-rays (Rh-KT) that
originates from the element of X-ray tube material when SQX Scattering FP method
(option) is employed. The scattered X-ray intensities are collected with the measurement
of library samples after setting up components and their concentrations of library
samples.
This is a drift correction sample used for scattering sensitivity library. It is necessary to
prepare drift correction samples for each library. Either the method to correct by one
point or the method to correct by two points can be chosen as the drift correction
method on each library but the use of the method is generally selected.
The procedure to create a new library sample is described below. A library sample of an
aluminum plate using the measured element line Al-K will be used as an example.
By clicking Add in the Measuring Condition frame on the main screen, the common library
with a different measuring condition can be registered additionally for the same component.
Select the common library with a different measuring condition for a library in the line of an
additional measurement.
“Sealing” is displayed only when the spectrometer has the helium flush mechanism and the
sealing (option). Usually select <OUT> for “Sealing”.
Though Al alone is set for measurement in the above example, two or more measured
element lines can be set such as K-K and Br-K in the case of KBr of common library
samples.
The total of the standard values of components must be within a range of 99.0 to 101.0.
It is possible to add a different common library condition using Add in the Measuring
Condition frame.
A measuring condition can be changed if necessary. But part of the excitation and optical
conditions and the PHA condition cannot be changed in the library measurement. The
conditions concerning intensity shifts and the angle range can be set here.
In the case of the ZSX Primus, ZSX Primus II or ZSX Primus IV/IVi, the attenuator can be
selected only when the measurement diameter is 30 mm and the sealing is “OUT”.
In the case of the ZSX Primus III+, the attenuator can be selected only when the
measurement diameter is 30 or 20 mm.
The procedure to set a matching library sample is the same as that for a common library
sample except that bulk samples alone can be used. Two or more measured element lines
can be set.
5. For the setting procedures after this, see the paragraph for a common library
sample.
It is also possible to register a matching library sample from the SQX calculation of the
[Qual Result] program.
It is impossible to save a library sample registered from the SQX calculation after it is
updated. The sample size, sample film correction, etc. are also stored and can be
displayed.
The library drift correction sample has no model such as bulk or thin film and does not
need standard values. Other than those differences, the setting procedure is the same as
that for a common library sample.
5. For the setting procedures after this, see the paragraph for a common library
sample.
This setup functions only when SQX Scattering FP method (option) is attached.
In the case of a scattering library sample, only scattered rays are set for a measurement
line.
When preparing the scattering library, it is possible to use the standard sample that only
consists of non-measuring elements such as carbon and hydrogen.
It is also possible to set compounds such as oxides.
Compton and Thomson, which are scattered rays for the scattered ray correction, are
displayed additionally in fields for components.
15. Select the line of scattered rays to be edited, click Edit... in the Measuring
Condition frame, and edit a measuring condition.
16. Click OK .
17. Select Common Conditions by clicking.
In SQX Scattering FP method, it is possible to use only 20mm in diameter only when
Sealing is OUT and to select 30 mm in diameter only in case that Sealing is IN under He
atmosphere.
5. Refer to the setting procedure on Scattering Library Sample for the following
settings:
A library sample can be deleted. Care must be taken because a deleted sample cannot be
restored.
Saved information about library samples can be printed. The content of a sample or a
list of sample names can be printed. In the case of the content of a sample, measuring
conditions can also be printed.
2. Select a library type for measurement on the “Library Sample Meas.” screen.
3. Select sample names.
4. Set a sample position.
5. Set sample information if necessary.
6. Click OK .
When there are many library samples to be measured, you can easily select necessary
samples by clicking All and then clicking unwanted samples to deselect them.
To delete an ID or add one in the preset library samples on the ID list, reselect library
samples.
The sample information that is set on the screen where library samples are selected is
common to all IDs. To identify IDs more distinctly, change sample information
individually. Sample positions can also be changed individually.
It is possible to add, insert, cut, copy and paste sample ID of library sample as done in
sample ID edit in Analysis program.
e. Sensitivity Calibration
Select Library
The SFP method that takes the photoelectron enhancement effect into consideration can
be used for boron (B) to oxygen (O) in bulk samples. The SFP sensitivity list displays
only those elements. Carry out the following procedure after selecting library:
Change Scale
1. Click an element (rectangular mark) with the mouse on the graph.
2. Select <Update>, <Do not update> or <Delete> in the “Sensitivity” frame.
3. Click OK .
After making selection for Sensitivity Update, carry out the following procedure:
Save is displayed beside List under the title bar of the “Library Update” screen. A
sensitivity library can also be updated using this Save.
Print Graph
<Print Library List> and <Graph Print> can be operated regardless of the library that has
been displayed.
4.Select a calculation method from < Alpha method > or < Alpha-beta method >.
5.Select the drift correction sample at High side. If <Alpha-beta method> method is
used, it is necessary to select the drift correction sample at Low side too.
6.Click Alpha-beta reset when resetting the coefficients currently stored.
7.Click OK .
8.Select [File] [Save] in due order when the setting is completed.
Reference intensities are automatically stored when the first measurement of drift
correction samples is carried out.
The setup of drift corrections for scattering library is completed at the initial operation
only. The coefficients of drift correction are automatically updated at every
measurement of scattering drift correction samples.
1. Click Scattering Library Update on the flow bar of the sensitivity library.
The window of Scattering Library Update is displayed.
Print Graph
1. Click Print .
2. Select the output from <Printer>, <Bitmap>, <Metafile> in ”Output” .
3. When either <Bitmap> or <Metafile> is selected, create <File output folder> and
<File name>.
4. Click OK .
Display Result
Close Graph
3.5.1 OUTLINE
In the SQX analysis, which is a standardless analysis method, lower limits of detection
can be reduced by using the fixed angle measurement for the analyses of Cd, Pb, Hg, Cr
and Br, which are object elements of the WEEE, RoHS and ELV Directives. Reports on
these hazardous elements can also be drawn up. This section describes the method and
contents of the SQX analysis to analyze elements including hazardous ones using the
hazardous element analysis option.
Qualitative applications for the SQX analysis of the sample types of metal, powder and
polymer have been registered beforehand as templates. By using these templates, the
analyses of Cd, Pb, Hg, Br and Cr can be made with high precision.
To analyze other samples, prepare qualitative applications referring to the template of
metal, powder or polymer.
“Metal_RoHS” and “Powder_RoHS” should be used for bulk samples, whose X-ray
intensities are not affected by thicknesses. “Polymer_RoHS” should be used for thin film
samples, whose X-ray intensities are affected by thicknesses. Therefore, when making an
analysis using “Polymer_RoHS”, enter beforehand the coating weight (mg/cm2) of an
analysis sample using Input Data... in the “Sample ID Setting” window on the Analysis
screen. When analyzing a polymer sample, use a disklike or pressed sample.
The procedure to prepare a qualitative application is described below, using the analysis of Cu
alloy with an analysis diameter of 20mm as an example:
6. In the right part of the “Select Element Range” screen, fixed angle analysis conditions
have been set beforehand to analyze trace elements with sufficient precision. If any
element (for example, Br or Hg) need not be analyzed, delete a setting for Br-Kα or
Hg-Lα on the left part of the screen. For deletion, right-click a spectral line to be
deleted and select [Delete]. (The measurement time can be shortened.)
7. Click Next .
8. Set information for <Sample Preparation> and click Next .
9. On the “Parameters” screen, first select a measuring diameter.
10. Then set <No> for the sample spin.
12. When the measuring diameter is 10mm, make necessary changes such as
lengthening the scan measuring time.
13. Click Next . Change settings on “Output information” if necessary.
14. Select <Output detection limit for undetected element (fixed angle)>.
When a check mark is placed for <Output detection limit for undetected element (fixed
angle)>, elements that were not detected in the fixed angle measurement are also
displayed on an analysis result and their detection limits are output. When you want to draw
up a hazardous element analysis report, be sure to place a check mark.
subfolder, click Browse... , select an appropriate folder and then select the data file.
Like Br shown above, when a check mark has been placed for <Output detection limit for
undetected element (fixed angle)> on “Calculation Condition”, a detection limit is also displayed
for an element with an analysis result of zero.
When a liquid sample has been analyzed using a filter, a detection limit for the concentration of
the original liquid sample is displayed, as shown below:
By placing a check mark for <Output detection limit for undetected element (fixed angle)> after
an analysis is completed, detection limits can also be calculated for undetected elements in the
fixed angle measurement.
4. Place a check mark for <Output detection limit for undetected element (fixed angle)>.
5. Click <Calculate from qualitative analysis result.>.
6. Click OK .
7. Click Calculate.
Since the RoHS Directive defines values in concentration, select <mass%> for the unit of
each component.
b. DRAWING OF REPORT
5. Excel automatically starts and a report is displayed. For saving in a file, execute
[Save as].
(1) (2)
(3) (4)
“2. Analysis Results” and “3. Sample Spectrum” are created automatically.
To change the full scale in “3. Sample Spectrum”, select an appropriate file by selecting
Data Processing Qual Result in ZSX Guidance, make the spectrum of an relevant
element displayed on the “Spectrum” tab, change the full scale and carry out the procedure
of Output Hazard. Elem. again.
6. A new work sheet is added and a report in the Excel format is drawn up
automatically. After the report is drawn up, save it with another name in a folder
such as “My Documents” by selecting [Save As] on the [File] menu.
Because the default template files in the following folder are used as master files, do not
overwrite or delete any of them:
7. To print the report, make a setting for [Print Area], check a printing image using [Print Preview]
of Excel and select [Print].
3.5.4.CUSTOMIZATION OF TEMPLATE
a. LAYOUT OF REPORT
A report with a free layout can be drawn up by changing the layout of a report template.
<Example of Report>
Analyzed by:
1. Sample Information
Sample name
Analysis group
Sample weight
Material
2. Analysis Results
Method X-Ray Fluorescence Analysis
Preparation
Atomic symbol Cd Cr Pb Hg Br
Judgment
Analyzed value
LLD
Unit
3. Sample Spectrum
<<Notes>>
b. DESCRIPTION OF NAMES
A report template can be easily customized by changing the positions of names. Details are
explained below:
Constitution of Names.
Names Contents
ANL_DATE Analysis date
SAMPLE_IMG Sample image (when available)
ANL_VAL_01 ~
Analysis value
ANL_VAL_10
JUDGE_01 ~JUDGE_10 Judgement
LLD_VAL_01 ~
Lower limit of detection
LLD_VAL_10
Manually input memo in sample ID
SMID_MEMO_01 ~
(Example: 01: Sample name, 02:
SMID_MEMO_09
Analysis company name)
SPECTRUM_IMG_01 ~
Spectral chart (image)
SPECTRUM_IMG_10
UNIT_01 ~UNIT_10 Unit
A spectral chart only of the individual qualitative analysis is output. A full range
qualitative analysis chart for Ti through U is not output.
The sheet name of a report template must be “Report”. If not, a report will not be
drawn up.
Although any value can be entered for the company standard, set a value equal to or
smaller than an RoHS standard value. On a report, an analysis value smaller than a
company standard value is treated as “passed”, one larger than an RoHS standard
value as “failed” and one between an RoHS standard value and a company standard
value as “warning”.
d. ANALYSIS RESULT
Expression Condition
N.D.-: No
Analysis A data item does not exist or an analysis value is zero.
detection
Value
[Numeric value] An analysis value exceeds zero.
Lower N.D.: No detection A lower limit of detection is zero.
Limit of
[Numeric value] A lower limit of detection exceeds zero.
Detection
An analysis value is below a company standard value.
OK In the case of no detection, a lower limit of detection is
below a company standard value.
An analysis value is equal to or greater than a company
standard value and below an RoHS standard value.
Warning In the case of no detection, a lower limit of detection is
Judgement
equal to or greater than a company standard value and
below an RoHS standard value.
An analysis value is equal to or greater than an RoHS
standard value.
NG
In the case of no detection, a lower limit of detection is
equal to or greater than an RoHS standard value.
NG
Analysis Value
RoHS standard and company standard values can be used as the upper and lower limits of
thresholds for judgements in screening.
4. QUANTITATIVE ANALYSIS
4.1 QUANTITATIVE ANALYSIS
Quantitative analysis is generally used for control analyses such as process control and
quality control. To perform control analyses, it is important to select the appropriate
method. Various analysis methods are described below for your information.
The empirical calibration method and the Fundamental Parameters (FP) method are
the two primary quantitative analysis methods.
The criterion to select the empirical calibration method and the FP quantification
method is shown below. The FP quantification method cannot be used with an
application template.
In the FP quantification method, the empirical calibration method can be used to obtain
the quantification values of some components. In this case, after determining analyzed
values using the empirical calibration method, calculation proceeds for the other
components using the FP method. Reciprocally, in the empirical calibration method, the
FP method cannot be used for any component.
In Chapter 4, the following marks are attached to descriptions for items concerning the
empirical calibration method and the FP method:
Mark Description
The relationship between the concentrations (or thicknesses) and corresponding X-ray
intensities of a component is called the calibration curve. The method that analyzes
unknown sample values based on such a calibration curve is called the empirical
calibration method.
The empirical calibration method (standard sample comparison method) uses the
relationship between measured fluorescent X-ray intensities and concentrations
obtained from standard samples similar to unknown samples and the fluorescent X-ray
intensities of unknown samples to perform quantitative analyses.
When effects of coexisting elements (absorption and enhancement of X-rays and overlap
of peaks) cannot be neglected in the empirical calibration method described above, a
calibration curve that can correct for those effects with calculations is used for
quantitative analyses. This is called the correction quantification method.
This is another method commonly used to correct effects of coexisting elements. Samples
that have known concentrations and contain an internal standard element added in a
constant proportion are used as standard samples. Calibration curves created from the
relationship between the X-ray intensity ratios (of the analyzed element to the internal
standard element) and concentrations are used for quantitative analyses. This method is
called the internal standard method.
Additive Method
When standard samples are not available, a certain quantity of an analyzed element or
material that contains it (that has chemical composition similar to that of measured
material) is added to the unknown sample in several concentrations. A quantification
factor is calculated for the change of the concentration and the X-ray intensity. That
method is called the additive method.
Standard samples that have concentrations similar to those of unknown samples must be
selected to reduce effects of coexisting elements and avoid error factors characteristic of
samples.
b. FP Quantification Method
The following two methods can be used to obtain spectrometer sensitivity values:
1. Use standard samples that have composition similar to that of unknown samples
2. Use standard substances such as metals and reagents (sensitivity libraries)
Unlike the empirical calibration method, all components constituting the sample
are used for quantification calculation. Any unmeasured components must be
defined as the fixed value, manual input value or balance component, so that the
total of the concentrations is 100%.
In the thin film FP quantification method, information about the structure of layers
is necessary.
Samples necessary for the quantitative analysis using the empirical calibration method
or the FP quantification method are described below. Those samples are classified into
the standard samples, drift correction samples and check samples.
Standard Samples
Samples to correct long-term drifts of X-ray intensities used for calibration curves or
sensitivity calibration curves are called the drift correction samples. There are no
limitations of the concentrations and shapes of those samples. They must be stable over
long periods of time. For example, metal drift correction samples can be used for the drift
correction of a calibration curve for powder samples. If the standard samples described
above are stable, some of them can be used as drift correction samples.
Check Samples
To ensure that existing calibration curves or sensitivity calibration curves can be used in
normal operation, a check sample is run. Unlike the drift correction sample, the check
sample must be similar to the unknown samples and have known concentrations. Like
the drift correction samples, however, if the standard samples are stable, some of them
can be used for check samples.
The operational flow to create a quantitative application is shown below. Unlike the
procedure to create a qualitative application, many steps are necessary to create a
quantitative application. Specifically, in the flowchart shown below, the measuring
condition setup and the regression calculation may have many operational parts
themselves and may become complicated depending on the purpose of analysis. Their
operational procedures are therefore described separately in detail. In general, a
quantitative application must be created through thoughtful consideration and cannot
be set in one process. Therefore, one cannot typically follow this flow diagram in one
direction. It is necessary to repeat steps such as changing preset values, and by a trial
and error method of reconsidering and checking a good quantitative application is
developed. You need not set all items. It is possible to set necessary items alone.
Start
g. Regression Calculation
End
When the menu item “Quant Application” is selected, <Application File> is started
automatically.
An example using the Steel template in the Metal & Alloy category is described below:
When a template is used, the empirical calibration curve method must be used.
Other Templates
Templates have been registered only for generally established analysis methods.
Applications for which standard samples are available on the market are set
preferentially as templates.
To create an application without using a template, select <Create your own application>.
When registering an application by entering its name, a folder for registration can be
created.
The Anlcond folder in the “Select Folder” window is the <Common> folder.
The folder specified by the user with User Administration in <6.1 h System Management>
has been selected as the default.
To display an application that has already been created or modify a part of it, first select
it.
When the mouse pointer is placed at the application name, the memo of the application is
displayed.
When a part of an existing application has been changed, the old contents will be
overwritten if that application is saved under the same name. For comparison with data
of the past, it is recommended that an application that has been used at least once be
kept without changing its contents.
1. Select <Rename and save the application> on the “Application File” screen.
2. Click Next > .
3. Select <Application> for “File type”.
4. Enter a new application name that is not identical to any existing file.
5. Select a folder to save the application.
6. When drift correction coefficient have been registered for the selected application,
the “Reset drift correction coefficients and reference intensities” check box is
displayed. Place a check mark when reference intensities for the drift correction for
the new application are to be remeasured.
7. Click Finished .
Register as Template
1. Select <Rename and save the application> on the “Application File” screen.
2. Click Next > .
3. Select <Template> for “File type”.
4. Enter a new template name that is not identical to any existing template.
5. Click Finished .
1. Select <Change analysis type/sample model> on the “Application File” screen.
2. Click Next > .
3. Enter a new application name that is not identical to any existing file name.
4. If necessary, select <Analysis type change> or <Sample model change of FP method>.
5. When the sample model is to be changed, select an application type.
6. Select a folder.
7. Click Finished .
To change the empirical method to the FP method, the analysis unit of a component must
be “mass%”, “at%”, “ppm” or mol%”.
When flux is set, change to the thin film is impossible. When the number of components is
100 or more, change to the thin film is also impossible.
Change from the thin film to the bulk is possible only in the case of a single layer. In this
case, a substrate is not counted as the layer. When the measurement line of a substrate is
used as a measurement line for a film thickness, change is impossible.
Delete Application
The application information includes setting items for the component selection and the
sample preparation.
When an application has been created using a template, a standard set of components
has been selected. On the “Component Selection” screen, unnecessary components can
be deleted from the list and additional ones can be added. Select a component name used
for the quantitative calculation from the periodic table and include the choice of element,
oxide or special compound. Select a unit and an analysis type. The analysis type is
<Measure[EMP]>, <Balance>, <Input> or <Fixed>.
1. Select <Element> for “Type of component to add” and click an objective element
symbol in the periodic table.
This setting performs quantification calculations with the addition of a standard oxide.
1. Select <Oxide> for “Type of component to add” and click an objective element symbol
in the periodic table.
When a special compound is set, element symbols are not distinguished by color. But an
element that is set as a measured component in two or more places such as Si and Si3N4
is displayed in light blue.
When the sample type is the fusion bead and <All compounds> is selected for the special
compound, <Ig> is displayed as the component name. <Ig> is used for the component of
ignition loss in fusion bead samples.
Since an ignition loss component in a fusion bead can be automatically added in the
sample preparation information setting, it need not be set here.
Set here only components in a sample. Flux in a fusion bead and binder for analyzing a
powder sample should be set in the sample preparation information setting.
The component <Ig> of ignition loss in fusion bead samples is selectable only when the
fusion bead correction option is attached.
The units displayed at the completion of the quantification calculation can be set. When
setting the units for concentration, mass proportions cannot be mixed with molecular
proportions or atomic proportions.
Filters, films and coated samples are treated as thin film samples. Unlike bulk samples,
setting for a multi-layer (the same for a single layer) is needed. The order of layers
cannot be obtained using the X-ray fluorescence analysis. The structure of layers must
be set beforehand. In the case of coated samples, their bases can be set as substrates.
Set Substrate
Add Layer
1. Select a layer on the “Component Selection for Film” screen and right-click it.
2. Select <Insert layer above> or <Insert layer below> on the menu.
3. Set information on the layer in the procedure described in the paragraph “Add
Layer”.
To make an accurate analysis, sample information suitable for a sample must be set. For
example, to make an accurate analysis for a powder sample or a fusion bead, accurate
information is necessary such as a compounding ratio (binder weight/sample weight)
when mixing powder with binder or a dilution ratio (flux weight/sample weight) when
fusing sample using flux to make a fusion bead sample. Using this software, setting
contents for such sample information can be reflected in analysis conditions and the
theoretical matrix correction coefficients calculation. When an application is created
using a template, a sample preparation method has been preset. However, this
information is an example and should be treated as merely information for guidance.
Although setting items that are treated as memoranda need not be set, it is
recommended that they be set to record a sample preparation method.
Setting items for sample preparation information vary depending on a sample type.
Sample types are shown below:
When you want to proceed to the next step without setting this item, click Cancel and select
an item on the flow bar directly.
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “PROCEDURE FOR
ADDITION OF SAMPLE FILM INFORMATION ” regarding details of sample film.
1. When binder is used to improve the formability of a powder sample, click <Used>
under “Binder/Grinding aids” and select a binder compound. Enter sample weight
and binder weight.
2. To set an option, click Show option . An additional option display area appears in the
lower part of the screen.
3. Set information on pulverizing, pressing dies and sample weight thickness. These
items are used for the analysis diagnosis but are not reflected in analysis conditions.
4. In the case of an ignited sample analysis, select <Converts before Ignited> and set
the component name of an ignition loss. A quantitative analysis value can be
converted into a concentration before ignition during an unknown sample analysis.
Ignition loss weight is automatically set in analysis parameters as an input
component.
Items below << Hide option on the screen are displayed only when Show option is clicked.
In the case of an ignited sample analysis, set analyzed values after ignition for the standard
values of standard samples. Register beforehand the component name of ignition loss
weight using the “Compound Table” program.
When the component names of binder and ignition loss are not displayed, register them
beforehand as binder and ignition loss using the “Compound Table” program.
Binder information is used in the theoretical matrix correction coefficient calculation in the
FP and empirical methods.
When you want to proceed to the next step without setting this item, click Cancel and select
an item on the flow bar directly.
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “PROCEDURE FOR
ADDITION OF SAMPLE FILM INFORMATION ” regarding details of sample film.
Checking <set to each sample>, it becomes possible to change each kind of film for
standard sample setting and when analyzed by FP method. (This is the function of ZSX
Primus IV/IVi only, and applicable to the equipment shipped newly.)
1. Select a flux type under “Sample and flux”, and set flux weight and sample weight
for criteria.
2. When a component that becomes an ignition loss, such as water of crystallization or
a carbonic acid radical, exists in a sample and a correction is to be made for it, select
<Corrects for LOI in sample>.
3. When the ignition loss correction is to be made, select <Balance> or <Manual input>
for the handling of an ignition loss. If <Balance> is selected, the effect of an ignition
loss can be corrected even when the ignition loss is unknown. To use the ignition loss
value of a standard sample or an unknown sample, select <Manual input>. It is
automatically set in analysis parameters as an input component. The ignition loss
component is handled as <Ig>.
4. To use oxidizer, select <Used> and set an oxidizer type. For “Weight”, enter oxidizer
weight for a criterion. An addition ratio (oxidizer residual compound weight/sample
weight) is automatically set.
5. To prepare a fusion bead using a dilution ratio that is not constant, select <Corrects
for flux ratio>. Select <Each weight> or <Flux ratio> for “Input in analysis”.
6. When <Corrects for flux ratio> is selected, <Corrects for flux evaporation> can be
selected and setting becomes possible to make a correction for flux that evaporates
during fusion. In this case, fusion bead weight and sample weight are used, and they
are defined as input items in analysis parameters. And <Corrects for oxidizer
weight> can be selected.
The ignition loss correction and flux evaporation correction can be set only when the
system has the fusion bead correction option.
When any oxidizer is not displayed, enter and register a compound name as oxidizer using
“Special Compound” in the “Compound Table” program, and set a residual oxide at the
same time.
When <Corrects for flux evaporation> is not selected, the dilution ratio means (Flux
Weight/Sample Weight). When a correction is made for evaporation, the dilution ratio
means (Bead Weight/Sample Weight - 1).
When <Corrects for flux evaporation> is selected, only <Balance> can be selected for an
ignition loss.
7. To set an option, click Show option . An additional option display area appears in the
lower part of the screen.
8. When <Each weight> was selected for an input in an analysis for the dilution ratio
correction on the previous screen, set a component name when entering weight.
When <Corrects for flux evaporation> is not selected, enter the component names of
sample weight and flux weight. When <Corrects for flux evaporation> is selected,
enter the component names of sample weight and bead weight. When <Corrects for
oxidizer weight> is selected, enter the component name of oxidizer.
9. In the case of an ignited sample analysis, select <Converts before Ignited> and set
the component name of an ignition loss. A quantitative analysis value can be
converted into a concentration before ignition during an unknown sample analysis.
Ignition loss weight is automatically set in analysis parameters as an input
component.
Settings for pulverization, bead preparation condition and fusion bead diameter are used
for the analysis diagnosis but are not reflected in analysis conditions.
When a component name to enter weight is not displayed, enter a component name and
select a corresponding component type using “External Parameter” in “Compound Table”.
When the FP method is used, the concentration of solution can be calculated from a
sampling quantity, effective diameter and filter weight.
When the FP method is used, calculate the concentrations of elements in a filter using filter
weight and set them as the concentrations of elements in standard samples.
When Show option on the screen is clicked, setting for the type of a hollow cup to support
a filter appears. This setting is used for the analysis diagnosis but is not reflected in
analysis conditions.
1. Click <Used> under “Binder/Grinding aids” and select a binder compound. Enter
sample weight and binder weight.
2. To set an option, click Show option . An additional option display area appears in the
lower part of the screen.
3. Set information on pulverizing, pressing dies and sample weight thickness. These
items are used for the analysis diagnosis but are not reflected in analysis conditions.
Items below << Hide option on the screen are displayed only when Show option is clicked.
When the component names of binder and ignition loss are not displayed, register them
beforehand as binder and ignition loss using the “Compound Table” program.
Binder information is used in the theoretical matrix correction coefficient calculation in the
FP and empirical methods.
When you want to proceed to the next step without setting this item, click Cancel and select
an item on the flow bar directly.
1. When diluent is to be used, set information on it. Select <Used> and a diluent
component and set sample weight and diluent weight.
2. To set an option, click Show option . An additional option display area appears in the
lower part of the screen.
3. Set sample weight thickness. This item is used for the analysis diagnosis but are not
reflected in analysis conditions.
Items below << Hide option on the screen are displayed only when Show option is clicked.
When the component names of diluent are not displayed, register them beforehand as
diluent using the “Compound Table” program.
When you want to proceed to the next step without setting this item, click Cancel and select
an item on the flow bar directly.
In case of ZSX Primus IV / IVi, the sample film which has the suffix (N) at the end of film
name is made X-ray absorption correction by the film itself and the impurity correction in the
film at a same time, therefore, the impurity setting for the film which used to be done per
measurement diameter is not needed. Refer to the instruction manual “PROCEDURE FOR
ADDITION OF SAMPLE FILM INFORMATION” regarding details of sample film.
Checking <set to each sample>, it becomes possible to change each kind of film for
standard sample setting and when analyzed by FP method. (This is the function of ZSX
Primus IV/IVi only, and applicable to the equipment shipped newly.)
By using FP method, it makes possible to correct the measurement error caused by the
effect of sample thickness change for the analysis line having high energy in liquid
samples.
Especially for the case of tube below optics it is effective for the correction of the error due
to the difference of sample amount put in the sample holder or the error caused by the
height difference of liquid due to the difference of specific weight.
The correction of geometry effect which changes ratio of detectable area and
non-detectable area becomes possible for ZSX Primus IVi only. Sample cell CH1095 are
recommended to use for liquid sample.
In both the empirical calibration method and the FP quantification method, standard
samples must be set. In the case of the FP quantification method, the total of the
concentrations of each standard sample must be within a range of 99 to 101%. However,
it is possible to change to within 97~103% in case of bulk FP method. In the case of thin
film samples, setting must also be made for the film thickness or weight. The number of
digits can be set and the setting for a certified value or an uncertified value can be made.
In the case of bulk samples, standard values can be copied from the Universal Standard
Samples, a database of standard values, or a sensitivity library can be used without
setting standard samples in the FP quantification method.
The standard values of bulk samples are set here. Analyzed values obtained using
another analysis method are set as standard values for the X-ray fluorescence analysis.
Set one sample in each line. When a standard value for a component is not available, set
a blank in the appropriate cell. That component will not be measured. If you want only
to measure X-ray intensity for a check, enter a zero.
1. Double-click the “Name” cell and enter a standard sample name.
2. If desired, double-click the “Note” cell and enter a comment.
3. Double-click the cell of each component, and enter a standard value.
When the balance component is not set in the component selection of bulk FP method,
the selection of total standard sample concentration is indicated at the upper right of
standard sample setting screen.
When <99 – 101%> is selected, theoretical intensities are calculated by normalized total of
concentration of all standard samples to be 100%. But in case of selection of <97~103%>,
theoretical intensities are calculated without normalization, therefore, it is effective to the
case that the total concentration of standard samples are off from 100% significantly.
In the case of film or filter samples, make sure to enter film thicknesses or weight values
for each layer.
Standard values can be treated as uncertified values. That setting does not affect a
result of the regression calculation. The setting merely indicates that those standard
values are uncertified values. Those standard values are displayed in parentheses like
(1.23) to distinguish them from certified values.
1. Right-click the mouse in the cell of the standard value of a component that you want
to set as an uncertified value.
2. Select <Uncertified>.
To restore the previous setting, follow the same procedure and select <Certified>.
An arbitrary component can be set as a balance component. Only one balance component
can be set in each sample or layer.
1. Right-click the mouse in the cell of a component to be set as a balance component.
2. Select <Calculation Balance>.
When you have a standard sample registered in the Universal Standard Samples,
standard values can be copied from that database. Information about certified values
and uncertified values are also copied.
Standard sample information such as manually input standard values can be saved to
the Universal Standard Samples. Saved standard sample information can be used for
other applications. The information is saved in the “House standards” folder of the
Universal Standard Samples.
When the “Standard Samples” screen appears, X-ray intensities are not displayed. To
check X-ray intensities after measuring standard samples, follow the procedure
described below:
For the equipment with DMCA (ZSX Primus IV/IVi), X-ray intensities which are already
measured standard sample cannot be input directory. If necessary, select the column of
X-ray intensity and make right click, and enter the values after deleting the content by
clicking <Delete> on the popup screen. Note that the extraction of X-ray intensity by PHA
modification to create a calibration curve cannot be performed because DMCA data is
lost due to manual input of X-ray intensity.
The standard values, X-ray intensities and background intensities of standard samples
can be copied to Excel files or the like.
1. Select a cell area to be copied on the “Standard Samples” screen, and right-click the
mouse.
2. Select [Copy] on the menu.
3. Paste data using the software of the destination of copying, such as Excel.
The standard values, X-ray intensities and background intensities of standard samples
can be copied from Excel files or the like.
1. Copy data using the software of the source of copying, such as Excel.
2. Select a start cell for pasting on the “Standard Samples” screen, and right-click the
mouse.
3. Select [Paste] on the menu.
The standard values, X-ray intensities and background intensities of standard samples
can be exported to CSV files, which can be read by spreadsheet software.
CSV files with the predetermined format can be read to set the standard values, X-ray
intensities and background intensities of standard samples.
The format of CSV files that can be imported is the same as that of exported CSV files.
Export a CSV file and check the file format.
Thin film standard sample data can also be imported and exported.
When standard samples are set, typical values are automatically calculated. Typical
values are important when considering measured element lines in the case of a thin film
etc. The average values of the set standard samples are calculated as the typical values
of components. To change them, follow the procedure described below:
When Set Average is clicked, the averages of the standard values of standard samples are
recalculated.
When the sample model is the bulk and the units of all analysis components have been set
to mass%, ppm or ppb, the total of typical values is displayed. In the case of an FP method
application, set typical values such that the total falls within the range of 90.0 to 110.0%. In
the case of an empirical method application with which the theoretical matrix correction
coefficient calculation or analysis depth estimation is to be carried out, also set typical
values such that the total falls within that range.
1. On the “Analysis Information” tab, the type of each component can be changed, and
an extra component can be added or deleted.
2. On the “Measuring Condition” tab, set the measuring condition etc. of each
measured component.
3. On the “Calculate Parameter” tab, set the formula and coefficients of a process
parameter, a calculated component, etc.
In the case of a measured component, the FP method or the empirical calibration method
can be selected. The empirical calibration method can be used for one or several
components if applicable in the FP method application, but the FP method cannot be
used in the empirical calibration method application.
A component measured for correction for which the quantification calculation is not
made is also set here.
The non-measurement components, including any balance component, fixed value,
manual input, flux component, etc., can also be set.
For example, in a quantitative application to obtain analyzed results of Cr and Ni, you
can set a process parameter when you also want to report the ratio of Cr to Ni. One can
set a process parameter regardless of analyzed components. Like a process parameter,
information such as a known component or an additive can be used for a calculation and
a component for which external data is used can also be used for a calculation. The
following four types of calculated parameters are available:
In general, when there is no overlap of peaks, the K or L line is used for the measured
element line. So those measured element lines have been automatically selected. The
selected element can also be changed only for a measured element line to analyze a film
thickness in thin film samples.
Change Units
The procedure to change the reported units of a bulk sample or a thin film sample is
described below:
Units set here are used for standard samples. Units set in limit range conditions are output
as analyzed results.
Density
This setting is made only for a film thickness component in a thin film sample. To
calculate a film thickness, a film density must be set. On initial setting, the default
density is calculated from the concentration ratios of the components. When the
accurate density of the film is known, set it as a fixed value.
The balance component is not analyzed to calculate its quantification value. Its value is
calculated as the residual of the other components.
Known Components
The fixed value, input value or calculated value can be selected as the known
component.
When the concentration of a component is known for all measured samples, that
concentration can be set as the fixed value. Fixed values can be set for two or more
components.
This setting is used when the value of a known component varies for each sample. Two
or more manual input components can be set.
This setting is used when a component constituting a sample is treated like a process
parameter. Two or more calculated components can be set.
Additive / Flux
The known components described above are constituents of a sample. However, flux for a
fusion bead, oxidizer, binder for a powder sample, etc. are not contained in an original
sample and this program only displays them. Set those components using the program
for the sample preparation information.
Advanced Settings
In the advanced settings, one can set the number of standard sample preparations, enter
a memorandum and select printed items for an analysis.
In case that the application is set by the bulk FP method, the selection of <Normalize
application result to 100%> is displayed in the initial state. When this is changed to
deselect, analysis values before normalization to 100% become application analysis
values. This is effective for the analysis of sample of which total concentration is off from
100% significantly.
Extra components are classified into three types: a measuring condition for correction, a
process parameter and an external parameter. Although they are not analyzed directly,
they can be used for correction of analyzed components and for calculation.
The process parameter is used to calculate and report analyzed results such as the ratio
of two components. The calculation formula of a process parameter can be set arbitrarily
but complicated functions cannot be used.
The external parameter utilizes external (non-X-ray) data for calculations. For example,
in the case of a bead sample, sample weight and flux weight are set with external
parameters. Those external parameters are used to calculate a dilution ratio with the
formula for a calculated flux component. The external parameter name is registered
beforehand as a compound.
From the saved sample preparation information and analysis condition, problems etc.
concerning analysis are checked and displayed. Evaluation is carried out based on the
typical composition that was set on the “Composition Information” screen. When the
sample type is Film & Coating, the measured element line for the thin film analysis can
be evaluated.
Powder
Contamination from the crushing container or die is checked. When the sample film
is used in the loose powder method or the like or when the atmosphere is not a
vacuum, the fluorescent X-rays of light elements may be absorbed and the analysis
may be difficult. The absorption of X-rays by the sample film or the atmosphere is
checked. When the sample size and weight have been set, whether the sample is
thick enough to generate the measured element line is examined. For example,
when measuring the Ce-K line in a sample that consists mainly of light elements,
X-ray intensities may vary depending on the sample quantity because the Ce-K
line is not absorbed so much by the sample. This phenomenon is checked.
Polymer
In the analysis of chlorine in polyvinyl chloride, chlorine may volatilize affected by
X-rays from the X-ray tube. This phenomenon is checked.
Filter
Whether elements contained in the hollow cup are objects of analysis is examined.
Volatile elements such as chlorine are checked.
Liquid
Depending on the sample film used and the atmosphere (helium or air), the
fluorescent X-rays of light elements may be absorbed and the analysis may be
difficult. The absorption of X-rays by the sample film used and the atmosphere is
checked.
・Evaluation
Using film thicknesses and composition set with Composition Information,
analysis precision with a combination of set measurement lines is estimated,
and whether analysis is possible or not is judged. If analysis is impossible,
problems will be displayed.
When an interfering line (overlap line) that overlaps with an analysis line
exists, that interfering line is also displayed. Interfering lines are as follows:
When the influence of overlap exists, consider adopting the overlap correction.
When a sample thickness reaches the bulk region, “Int. ratio to infinite for analyzing
depth” becomes 1.0.
When a sample thickness is thicker than “Analyzing depth”, the sample thickness has
also reached the bulk region. “Analyzing depth” means a sample thickness at which an
X-ray intensity realizes the preset “Int. ratio to infinite for analyzing depth”.
This function can be used when setting has been made such that the total of typical
concentrations falls within the range of 90.0 to 110.0%.
To make an analysis after changing a measurement line, setting for a measurement line in
a measuring condition must be changed.
For an FP method application using the thin film model, a calculation can be made for
how much X-ray intensities change relatively when a layer thickness is changed.
If the gradient of a graph is small in the standard sample range or thickness range
assumed for unknown samples, analysis precision will worsen.
When an application that has the sample type of <Film & Coating> has been selected,
measured element lines can be retrieved.
To estimate the analysis precision of thin films, values of Composition Information are used.
The average of the values of set standard samples is set as the typical value of each
element by default. Therefore, when any standard sample has been changed, click Set
Average on the Composition Information screen for updating.
When composition information differs, optimum measured element lines may change. In
this case, reset the thickness and composition of an actual analysis sample as the
composition information, and carry out Evaluation again.
When different layers contain the same element, the simultaneous analysis of the thickness
and composition of each layer is generally difficult. By measuring the different
measurement lines of the same element, analysis may become possible. We recommend
that the composition and thickness of each layer be fixed or input manually for each
sample.
Intensity Calculated
Description
Theoretically
Intensity of X-rays generated in each layer. “Total
Attenuated
intensity” is the total of these.
X-ray intensity calculated assuming a single-layer
Not attenuated
thin film that consists of that layer alone
Degree of the absorption of fluorescent X-rays
Attenuation ratio
generated in each layer by upper layers
X-ray intensity ratio assuming that the thickness of
each layer is infinite.
Ratio to infinite
When this value is 1.0, the thickness cannot be
measured.
Detailed setting is made for each measurement of a sample component set in the
analysis information.
This setting is the same for the entire application that is now being created. The
measuring diameter of unknown samples and that of standard samples must be
identical. Therefore, the measuring diameter is generally set according to that of
unknown samples.
When the sealing (option) is in the “IN“ position because the atmosphere is Helium/Vacuum
or when the attenuator is used, only 30mm can be set for the measuring diameter.
<Air> cannot be selected for “Atmosphere” when the spectrometer has the helium flush
mechanism (option).
This can be selected only when the measuring diameter is 30(28)mm. When the
atmosphere is <Helium/Vacuum>, the sealing is unconditionally set to “IN”. The sealing
can be selected also when the atmosphere is <Vacuum>.
Order of Measurement
In this spectrometer, considering the arrangement of the crystals, the measuring order
is set so that measurement will begin with components at low 2-theta angles and
proceed to higher angles. This function is called the optimization of the measuring order.
Regardless of the setting order of the components, measurement is made in the
optimized order to shorten the measuring time. To make measurement in the order of
the components in the setting list, follow the procedure described below:
In general, when the measuring time of each component is made longer, measurement
precision becomes better. However, it would take a long time to change the measuring
time of each component each time. This system therefore has a function to lengthen the
measuring time collectively by multiplying it by a coefficient. An example of doubling
the measuring time is described below:
Setting of APC
The APC (Automatic Pressure Control) is used to keep the vacuum level constant when
making measurements in the vacuum condition. This is especially effective for X-rays of
light elements with long wavelengths. The APC can be set for each application.
When a hygroscopic sample is measured, it may take a long time to heighten the vacuum
degree of the pre-evacuation chamber or a vacuum error may occur. In such a case, carry
out a measurement with the APC turned off.
The target vacuum degree for pre-evacuation can be selected from the three levels of the
high, middle and low. When the high vacuum is set, variations in the vacuum degree in
the measurement chamber accompanying sample loading can be decreased. On the
other hand, when the low vacuum is set, a time required for sample loading can be
shortened. Set the target vacuum degree according to a sample to be measured and
necessary analysis precision.
Following items are functions to change the measuring condition. (This is the function of
ZSX Primus IV/IVi only, and applicable to the equipment shipped newly.)
・Tube wattage: Tube voltage (kV) and tube current (mA) of all measuring conditions
are changed to set values.
・Protection filter: Primary X-ray filters equipped with the system are automatically
arranged for measurement line to protect X-ray tube on a priority
basis. Filter OUT is set for the area lighter than Cl-Ka in case of
the specifications without Be30 filter.
・Time: All measuring time set previously are changed to 2 second thoroughly. It can be
changed easily while checking the total measuring time. Minimum setting
time is 2 second.
In the Analysis and EZ Analysis, measurement diameter for each sample can be
changed as below when sample ID is set in the application.
Measured intensities to different measurement diameter are made with conversion factors
stored in advance. To improve the analysis accuracy it is recommended to make the
application having same measurement diameter as analysis sample.
When the attenuator is set in the private measuring conditioning type, setting of
measurement diameter for every analysis sample cannot be made. Adequate attenuation
of X-ray intensities using tube current and primary X-ray filter are needed in advance.
Type Description
Private Used only in one application
Library Used with library measuring conditions
Common Used in two or more applications
Since the private conditions and common conditions can be changed arbitrarily, there
are basically only two types of setting, the private (common) conditions and the library
conditions.
In the case of the ZSX Primus, ZSX Primus II or ZSX Primus IV /IVi, <1/10> can be
selected for the attenuator only when the measurement diameter is 30 mm and the sealing
is “OUT”.
In the case of the ZSX Primus III+, the attenuator can be set only when the measurement
diameter is 30 or 20 mm.
For the optimum PHA condition, see “4.3 QUANTITATIVE ANALYSIS TECHNIQUES”.
When you use a template, the recommended PHA conditions have been set. Change any
necessary items.
For the optimum peak and background conditions, see “4.3 QUANTITATIVE ANALYSIS
TECHNIQUES”. When you use a template, the recommended peak and background
conditions have been set. Change any necessary items.
Background subtraction coefficients can be set only in the case of the 1 or 2 points method.
Special Setting
The following four measuring methods can be used for the quantitative analysis:
Measuring Method Description
A peak and backgrounds are measured with a measuring
Standard
time specified for each of them.
The optimum measuring times for a peak and backgrounds
Fixed with which the specified relative precision (%) is obtained is
precision determined using pre-measurement and then main
Time measurement is carried out.
optimization A peak is measured with the specified time. To measure
Optimize backgrounds, the optimum measuring times are determined
BG time using pre-measurement and then main measurement is
carried out.
A specified angle range is measured using a step scan to
Integration obtain integral intensities. When background measurement
has been specified, backgrounds are subtracted.
In “Fixed precision measurement”, to avoid too long measuring times, the measuring time
is limited to the total of the measuring times that have been set for a peak and backgrounds.
If the total measuring time obtained from the relative precision exceeds this maximum
measuring time, the maximum measuring time will be divided using the ratio of each
measuring time.
A calculation parameter can be copied from the application currently in dialog or from
another application.
1. Select a parameter to be set and right-click the mouse. The pop-up menu of [Copy]
appears.
2. Click [Copy].
3. Select a source application from <Current application> and <All applications>.
4. When <All applications> was selected, select an application.
5. Select a calculation parameter to be copied.
6. Click OK .
If any component set in the calculation parameter to be copied is not included in the
application in dialog, an error message will be displayed and the parameter cannot be
copied.
Special standard samples can be set only when the application is created with the FP
quantification method. In the FP method, unlike the empirical calibration method,
standard samples that have composition different from that of unknown samples can be
used. For example, when the composition of the film of unknown samples is different
from that of standard samples in the analysis of thin film samples, these special
standard samples can be set. The setting procedure is the same as that of ordinary
standard samples.
The example of setting described below uses an alloy of Ti, Cr and Cu as standard
samples when components of metal unknown samples are Fe, Cr and Ni. Cr is used as
an example of a measured component. An application with the FP method is created
without using a template.
Since the standard samples are for the FP quantification method, the total of the
concentrations must be within a range of 99 to 101%.
Set Special Standards of Thin Film Samples
For thin film samples, film thicknesses or weight values must be set in addition to the
equivalent setting items for bulk samples described before.
Delete File
Setting of the case that the measurement diameter of samples set with special
standard samples are different from the measurement diameter of application;
Special standard sample can be used as standard sample for different size sample from
analysis sample. Intensity measured with different measurement diameter than
application is translated into for the diameter in the application and registered. (This is
the function of ZSX Primus IV/IVi only, and applicable to the equipment shipped
newly.)
1. Click New sample , and after input sample name, click OK .
2. Click Input components and select the components/elements of the sample to be set
using the periodic table.
3. Click Property for each element/component and input standard values. The Total
concentration must be 100mass%.
4. Confirm the measurement line for the sensitivity registration element, and then click
OK .
5. Repeat the steps above for all components to complete the setting.
6. Select adequate measurement diameter from the pull down list to set measurement
diameter.
Setting for the case that the measurement diameter of special standard sample is
the same for the application but the counting loss takes place on the measured
intensity of the measurement line of the special standard sample;
If the unknown sample is a thin film and the bulk sample is used as a special
standard for creating the sensitivity curve, counting loss often occurs when
measuring the special standard.
In this case, it is possible to measure under different measuring conditions than for
the application to reduce the intensity of measurement line of a special standard
sample so that measurements are made in the suitable counting rate. For example,
setting lower tube current than the application, or applying attenuator (available only
the case of 30mm measurement diameter) for special standard can be made. The
measured intensity is converted considering the application measuring condition and
the sensitivity registration is made using that intensity. (This is the function of ZSX
Primus IV/IVi only, and applicable to the equipment shipped newly.)
1. Click New sample , and after input sample name, click OK .
2. Click Input components and select the components/elements of the sample to be set
using the periodic table.
3. Select same measurement diameter as application from the pull down list to set
measurement diameter of special standard sample setting screen.
4. Click Property for each element/component and input standard value. Total of
concentration must be 100mass%.
5. Uncheck the <Use measuring condition of application> for the measurement line of
element of which sensitivity registration is made, and then set tube current which
does not make counting loss or set attenuator (available only for 30mm
measurement diameter). Both settings at a same time are also available.
6. Follow above steps for all components repeatedly to complete the setting.
The optimum measuring conditions must be determined using the 2-theta scan and PHA
scan. Since the optimum conditions vary depending on an application, an expression
such as “THIS condition is optimum, set it” is impossible. You must understand samples
well and search for the optimum conditions by trial and error. Therefore, you may have
to go through this measuring condition setup procedure only once, or may have to repeat
it several times creating calibration curves and reconsidering the measuring conditions.
The measuring condition setup procedure is described below:
The excitation conditions, optical conditions and PHA conditions are the components of
the measuring conditions. To create a calibration curve or FP sensitivity curve, it is
fundamentally important to measure X-ray intensities of a measured component only.
That is, you must check that there are no overlapping peaks using the 2-theta scan and
that there is no overlap of energy using the PHA scan. From the viewpoint of angular
resolution or energy resolution, it may be difficult to create measuring conditions free
from all those overlaps. In that case, carry out correction calculations such as the
overlap correction in the regression calculation (Calibration program) described later.
The conditions monitored by the measuring condition setup are classified into the
excitation conditions, optical conditions and PHA conditions.
When checking each condition, it is recommended that the check be made using the
samples that have the highest and lowest concentrations in the concentration or film
thickness range of the standard samples of an individual component.
Excitation Conditions
Determine an excitation condition so that the highest intensity can be measured without
counting loss. For each measured component, carry out a 2-theta scan using the sample
with the highest concentration or film thickness in the range of the concentrations or
film thicknesses of the standard samples. If an unknown sample with a concentration
out of the range of the standard samples may be measured, create an excitation
condition taking this sample’s intensity into consideration.
In the case of the ZSX Primus or ZSX Primus III+, the upper limit of intensities that do
not cause the counting loss in the quantitative analysis is approximately 1000 kcps for
the SC and 2000 kcps for the PC. In the case of the ZSX Primus IV/IVi, the upper limit of
intensities that do not cause the counting loss is approximately 1800 kcps for the SC and
3000 kcps for the PC. Therefore, when X-ray intensities approach those values, take
sufficient care. When measuring intensities exceeding 1000 kcps using the SC or PC,
precision can be improved by adopting the integral measurement for the PHA setting or
setting a wide window.
Background can be lowered by using the primary filter. The primary X-rays from the
X-ray tube give energy necessary for excitation. When the primary filter is used, the
excitation efficiency is lower. However, even if the excitation efficiency lowers a little,
lower background may lead to a better calibration curve. For example, the characteristic
X-rays of the X-ray tube target material irradiate a sample and high efficiency and high
intensities are obtained for measured components having atomic numbers near that of
the target material, but these characteristic X-rays are scattered by the sample and
result in high background. Although the excitation efficiency lowers a little, by using the
primary filter and measuring a spectrum with lower background and a better
peak/background ratio, a better calibration curve can be created.
Optical Conditions
Improper setting of the optical conditions may cause poor accuracy of a calibration curve
or sensitivity curve. Check the following items with the 2-theta scan measurement:
For ordinary bulk samples, use an ordinary measured element line. For thin film
samples, a measured element line with a suitable escape depth must be selected.
Basically in the empirical calibration method, standard samples that have composition
similar to that of unknown samples must be used. In such a case, since the compositions
(referred to as “matrix” hereafter) are nearly identical, backgrounds have generally
similar shapes. In that case, it is not necessary to measure background positions. By
determining only a peak angle and a measuring time in that position, a calibration curve
with sufficient reliability can be created. That measuring method is called the fixed
angle measurement or fixed angle analysis.
Unlike the case described above, when standard samples of a different kind are added to
standard samples of the same kind, the background shape may vary according to the
change of the matrix. In that case, background positions as well as peak positions must
be measured. In qualitative analysis, background is estimated from a qualitative
analysis chart. In the fixed angle measurement, a background shape must be
reproduced with a small number of measured positions. Especially in the FP
quantification method, be sure to set the background measurement. For functions to
estimate the background shape, see “4.3 QUANTITATIVE ANALYSIS TECHNIQUES”.
Descriptions in paragraphs 2) and 3) are for a case where a measured element line is a
single peak. In fact, the peak of another component may overlap with the measured
element line. In such a case, further caution is suggested. It is especially important to
know constituting components, so it is recommended that all components be checked
beforehand. It is important to check the peaks of all components using the qualitative
analysis. The selection of monitored samples is also important. When peak overlap
cannot be eliminated with the optical condition, it may be eliminated by executing the
overlap correction.
Determine the measured angle positions and measuring times of a peak and a
background (referred to as “BG” hereafter) by executing a 2-theta scan. According to the
flowchart shown below, determine the optimum measuring condition concerning the
2-theta angle for each component. Since the PHA condition should have been adjusted so
that the peak position will be at 200, carry out a measuring condition setup procedure
with the initial lower limit set to 100 and the upper limit set to 300.
Start
Same BG shape?
No
Yes
BG fitting
Determine BG position
End
When using standard samples to set up a measuring condition, samples that must be
measured can be retrieved from the standard sample table and added collectively.
When a check mark is placed for “Add heavy scan to identify overlap”, a 2-theta scan
condition for heavy elements is added. By referring to the automatic identification result
of this 2-theta scan measurement result, fluorescent X-rays overlapping with each
measurement line, such as the overlap of the L-line of a heavy element with an element
spectral line, can be identified more accurately. (There are two types of setting screen
depends on software version as below.)
Set Conditions
Usually, prepare a sample with high X-ray intensities and one with low X-ray intensities.
Make setting for each sample as described below:
The upper area is used to select data, while the lower area is used to display results.
The sizes of the display areas can be changed using the division line.
The peak angle in the scan condition must be placed between the start angle and the end
angle.
If the atmosphere at the startup of Measuring Condition Setup is different from the one set
in the application, data display and editing will be possible but measurement will be
impossible.
If there is a problem with the measurement data, such as too high X-ray intensities, a
warning mark will appear in the data display area. By placing the mouse cursor on this mark,
the contents of the warning can be checked.
When a data item for which a measurement is now being made is selected, it is displayed.
A data item for which a measurement has been completed can be displayed, and various
processes for it are possible during a measurement.
1. Select a 2-theta scan data item for which measurement has been made.
2. Select [View] [Select Reference Data…].
3. Select another data item that has the same measured element line.
4. Click OK .
5. Check the overlapping of peaks.
Two data items for the same element and with different peak heights are displayed so that
you can see backgrounds before and after the peaks.
Specify Intervals
Specify Points
Elements that may overlap with the selected measurement line are retrieved from
typical concentrations that have been set using “Composition Information” and
displayed for overlap correction setting.
When the overlap correction has been set, measure standard samples and then calculate
overlap correction coefficients using the “Calibration” program.
The overlap rate is the ratio of an interfering line intensity to a measurement line intensity
and can be used for a criterion. A larger value means the greater degree of overlap.
Overlap correction coefficients can be calculated only when the peak deconvolution was
carried out and net intensities were updated.
If a sample name entered when adding a condition differs from the sample name of a
correcting line, that correcting line cannot be selected. Be sure to register a sample using
the same name.
Once the peak and backgrounds have been determined, a precision can be estimated.
Two methods are available for estimation. One is to calculate a precision after
determining a measuring time, and the other is to calculate a measuring time after
determining a precision. When the relative precision is used in precision calculation, 1.0
is used for the relative concentration in the calculation. When the absolute precision is
used, an arbitrary value can be entered as the concentration. This function is effective
when analyzing samples for which a high precision or a shorter measuring time is
needed.
Calculate Precision
Heighten Precision
The method to calculate a measuring time with which a precision is 0.001 mass% for a
concentration of 2.0 mass% in the ordinary analysis is described below as an example.
“Maximum measuring time” is set to 100 seconds.
Even if conditions have been determined through calculations, they will not be used for
analyses unless you save them in the files. Each time a measuring condition for an
analysis element has been determined, save it in the file using “Save Measuring
Condition”.
An item for which a new condition is left blank has not been changed from the old condition.
PHA Condition
Overlap of spectral lines on a qualitative chart can be corrected using the optical
condition described above and with overlap correction, if necessary. However, there is
one more item to be taken into consideration. It is the energy overlap of X-rays that
enter the detector, and can be checked with a PHA scan. X-rays of all wavelengths that
satisfy the Bragg’s formula enter the detector and are counted. If X-rays other than the
measured element line overlap, the accuracy of the calibration curve may suffer. A
condition to avoid that is called the PHA condition. A method to count objective X-rays
alone by setting a lower limit to eliminate noise and setting an upper limit to eliminate
higher-order X-rays is called the differential measurement. The range between the lower
limit and the upper limit is called the window. When there is no energy overlap issue,
the integral measurement method can be used. In the integral measurement, a lower
limit alone is set without setting an upper limit.
2) Escape Peak
Argon gas is used for the detector F-PC. When X-rays with energy higher than the Ar-K
absorption edge enter the detector, they excites Ar-K fluorescent X-rays and X-ray
signal pulses corresponding to energy lower by that of Ar-K are generated. When a
differential curve is drawn using the PHA scan, a peak appears at a pulse height lower
than an ordinary peak pulse height by this energy loss. This is called an escape peak. In
the case of the SC, which uses NaI, an escape peak corresponding to I-K appears.
Fluorescent X-rays generated from a sample are separated by the crystal and enter the
detector to be counted. Then part of X-rays with high energy may excite the crystal and
generated secondary X-rays may enter the detector to be counted. Unlike an escape peak,
pulses generated from that impurity line must be excluded from the window. For
example, when measuring Na-K, the M line of Tl that constitutes the crystal is
generated and enters the detector. In that case, by setting the lower limit to about 150,
the impurity line from the crystal can be eliminated.
Carry out a PHA scan, check energy overlap and determine measuring conditions. When
many parts of the PHA condition must be changed, you may have to execute the 2-theta
scan described in the previous paragraph again.
Start
Overlapping drawing
Integral measurement?
No
Yes
Set upper limit No upper limit
End
1. Select a PHA scan data item for which measurement has been made.
2. Select [View] [Select Reference Data…].
3. Select another data item that has the same measured element line.
4. Click OK .
To select a PHA scan data item, first select a sample with higher intensities.
In general, escape peaks are conspicuous with components with high intensities and
noises are conspicuous with components with low intensities. Determine upper and lower
limits taking account of noises, escape peaks, etc.
1. Select data to be printed from measurement lines that were measured.
2. Select [File] [Print].
3. Select <Printer>, <Bitmap> or <Metafile> for “Output”.
4. When an item other than <Printer> was selected for “Output”, enter a file name and
make setting for <File output folder>.
5. Click OK .
1. Select data to be output from measurement lines that were measured.
2. Select [File] [Output Text File].
3. Select <CSV> or <TAB> for “Spreadsheet”.
4. Click Browse... and specify a folder for file output.
5. Enter a file name.
6. Click OK .
2. Select standard sample names to be measured on the “Run Standards” screen.
3. Set a sample position.
4. Set sample information if necessary.
5. Make setting for “Repeat times” if necessary.
6 Click OK .
When there are many standard samples to be measured, you can easily select the desired
samples by clicking All and then clicking unnecessary samples to deselect them.
On the “Run Standards” screen, you need not select <Intermediate>,<First sample> or
<Last sample> for the update selection information, which is necessary on the “Sample ID
Setting” screen.
To delete an ID or add one in the preset standard samples on the ID list, reselect
standard samples.
The sample information and number of repetitions that are set on the screen where
standard samples are selected are common to all IDs. To identify IDs more distinctly,
change the sample information individually. Sample positions and numbers of
repetitions can also be changed individually.
It is possible to add, insert, cut, copy and paste a standard sample ID like the sample ID
setting in the Analysis program.
In case that the measuring component is added after standard sample measurement,
or the measuring conditions of particular measuring component are changed, it is
possible to measure the analysis line of the added/changed measuring component
without having to remeasure all standards. Setting procedure is same as ordinary
quantitative application.
Measured X-ray intensities change (drift) with time. To correct for such a change, one
implements the drift correction method. The drift correction is performed to update
prepared calibration curves for a long period. The same operation is also necessary in
the FP quantification method. Since it typically takes a very long time to prepare
calibration curves etc. before each analysis, X-ray intensities are calibrated with several
drift correction samples. The check analysis is used to check that calibration curves are
in normal operation. If an abnormality occurred in the check analysis, the drift
correction should be run.
The drift correction is classified into two steps. Initially, the drift correction samples are
measured when calibration curves are created, and the reference X-ray intensities are
registered along with the intensities of the standard samples and the calibration curves.
Subsequently, if a check analysis shows a drift in intensity, the drift correction samples
are measured before analyzing unknown samples. The drift corrections are ratios of the
reference intensities of drift correction samples measured when calibration curves were
created to the intensities of the drift correction samples measured before analyzing
unknown samples.
The preparation for drift correction to register reference intensities is explained here.
The correction method is selected from the alpha method and the alpha-beta method
that are described below:
Alpha Method
This is also called the one-point method. Only one sample that has about half the
highest X-ray intensity in the concentration range of a calibration curve or
sensitivity curve is set.
Alpha-Beta Method
This is also called the two-point method. Two samples that have a high intensity
and a low intensity in the concentration range are set. This method is used when a
concentration range is wide and when X-ray intensities including background
intensities are used.
Since the drift correction corrects X-ray intensities, information about concentrations is
unnecessary with these samples. There is no limitation of the types of samples, except
that they must be stable for a long term and have X-ray intensities appropriate for the
alpha method or alpha-beta method in the concentration range. For example, the drift
correction of pressed powder samples can be carried out using metal samples. If
standard samples used to create calibration curves satisfy the requirements for drift
correction samples, they can be used also for drift correction.
The flowchart of the procedure is shown below. As described before, when using
standard samples for drift correction, they need not be measured again. When standard
samples are not used for drift correction, X-ray intensities must be measured after the
following setting:
Start
Setting for drift correction samples can be made on the standard sample table.
Automatic setting of drift correction samples from standard samples is also possible. The
alpha method or alpha-beta method is selected and drift correction sample names are set
here.
The standard values and, if already measured, X-ray intensities are displayed for each
standard sample name. Optimum drift correction samples are selected from this table.
Select a sample for each component and click High or Low , and that standard sample
will be set as a drift correction sample. Drift correction samples can also be set with Auto
selection .
3. Select a sample appropriate for the component and click High .
4. If necessary, select an appropriate sample and click Low .
5. Select High or High and Low for each component.
6. Check the selected drift correction samples and click OK .
To select drift correction samples from standard samples, the following three selection
criteria are used:
Criterion Description
Alpha
The alpha method alone is set.
method
Method The alpha-beta method is also set if possible.
Alpha-beta
The alpha method is selected if the condition is
method
improper.
Setting is made only for components for which drift
Unsaved
correction samples have not been selected.
Component Setting is made also for components for which drift
Saved correction samples have been selected and samples are
reselected.
X-ray intensities are compared and samples are
selected.
Measured
When measured intensities exist, the initial selection
condition is set.
Data
Standard values are compared and samples are
selected.
Standard
When measured intensities do not exist, the initial
selection condition is set.
6. If necessary, select a sample not selected automatically and click High or Low .
7. Click OK .
When standard samples are not suitable for drift correction samples, other samples can
be registered as drift correction samples. In that case, after setting drift correction
samples, they must be measured to register reference intensities.
In the advanced settings, selectable items for the drift correction are the fixed time
multiplier, measuring condition, short term limits, maximum effective period and
sample preparation times.
The optimum time to analyze unknown samples is set for the measuring time in an
application. Since the drift correction is designed to recertify calibration curves using a
small number of samples, a longer measuring time may be used. But if the measuring
time in the application is changed, it must be returned to the previous value before
analyzing unknown samples. Therefore, by setting the fixed time multiplier, the
measuring time for the drift correction can be adjusted without changing measuring
conditions for unknown samples.
The range allowed for the fixed time multiplier is from 0.1 to 99.9.
The following two setting methods can be used to set a measuring condition for the drift
correction:
Measuring
Description
Condition Used
Current The measuring condition of the current application is used
Application intact.
Other Common When another measuring condition can be used in common,
Measuring that condition is used.
Condition A condition for a different measured element line can be used.
The common measuring condition means a measuring condition for which <Common> is
specified as the measuring condition type. See the paragraph “Advanced Setting of
Measuring Condition” in “4.2 d. Setting of Analysis Information”.
1. Select <Use meas. condition of current application> or <Use other common
measuring condition> for “Measuring condition to use”.
2. When <Use meas. condition of current application> was selected, click OK to
terminate the “Advanced Setting of Drift Correction” screen.
3. When <Use other common measuring condition> was selected, click “Select”.
After a drift correction is carried out, correction coefficients can be checked by comparing
them to the previous correction coefficients.
1. Select <Alpha ratio change> or <Beta change> for “Short term limits”.
2. Enter a preset value.
3. In order not to update coefficients values in out of limits, select <Do not update
drift coefficients when a check error occurs>.
4. Click OK on the “Advanced Setting of Drift Correction” screen.
The effective period from the previous drift correction can be set. When the preset period
has passed and unknown sample analysis is executed without executing drift correction,
a warning is given. If you do not want to use this function, select <Unlimited>.
In the case of metal samples which have their surface is ground before each
measurement, when using the averages of values measured several times for drift
correction, the number of sample preparation times is set. By using averages, the
dispersion of data due to sample preparation can be smoothed. The range allowed is up
to 10 times. Take care because this setting is different from the number of repetitions of
measurement.
1. Select the number of preparations for “Sample prep. times to update”.
2. Click OK on the “Advanced Setting of Drift Correction” screen.
When setting sample IDs in Analysis program for a component set with more than one
sample preparation times, set sample IDs by the number of sample preparation times for
the drift correction samples and set <Last sample> at the last sample ID of drift correction
sample. Note that drift correction coefficient update is not carried out when the number of
measured sample IDs of the drift correction sample is smaller than sample preparation
times.
On the drift correction sample list, you can see which components in which applications
use a registered drift correction sample for their drift corrections. If necessary, you can
enter a memorandum.
When samples other than standard samples are to be used as drift correction samples,
measure them using the “Run Drift Corr. Samples” program, and register reference
intensities.
2. Select sample names to be measured on the “Run Drift Corr. Samples” screen.
3. Set a sample position.
4. Set sample information if necessary.
5. Make setting for “Repeat times” if necessary.
6. Click OK .
When there are many samples to be measured, you can easily select the desired samples
by clicking All and then clicking unnecessary samples to deselect them.
On the “Run Drift Corr. Samples” screen, you need not select <Intermediate>,<First
sample> or <Last sample> for the update selection information, which is necessary on the
“Sample ID Setting” screen.
When the first sample is measured, measurement results that have not been used for the
drift correction coefficient calculation are reset.
To delete an ID or add one in the preset samples on the ID list, reselect drift correction
samples.
The sample information and number of repetitions that are set on the screen where drift
correction samples are selected are common to all IDs. To identify IDs more distinctly,
change the sample information individually. Sample positions and numbers of
repetitions can also be changed individually.
It is possible to add, insert, cut, copy and paste a drift correction sample ID like the
sample ID setting in the Analysis program.
The Calibration program creates calibration curves or sensitivity curves in the empirical
calibration method or the FP quantification method. After measuring standard samples,
calculations are made for each component to prepare for quantitative analysis.
Calculations are made by setting various conditions such as regression formulae,
background subtraction methods and matrix correction methods. Better calibration
curves or sensitivity curves are created through trial and error. In some cases, you must
return to recheck the measuring conditions. For further details, see “When Accuracy of
Calibration Curve Is Bad” in “4.3 b. Empirical Calibration Method”.
This software features the capability to carry out two or more regression calculations
and display their results simultaneously. You can consider and then save the results. For
example, as shown below, a result with a linear formula (right) and one with a quadratic
formula (left) can be compared and the better one can be saved:
In the Calibration program, a calculation is made for each component according to the
following procedure:
Start
End
The mark on the left of a component name indicates that the component has been
registered after a regression calculation or sensitivity calibration. When calculation has
been executed at least once for a component and the result is saved, this mark is
displayed.
Formula Type
The ratio method, linear formula, quadratic formula, cubic formula and logarithmic
formula can be used for the formula in the empirical calibration method. The ratio
method, linear formula and quadratic formula can be used for the formula in the FP
quantification method. The general form of those formulae is shown below. The constant
terms of each formula are shown in the following table. When the number of samples is
small, some formulae cannot be selected.
<Use calculated> in the options is used to calculate matrix correction coefficients and is
not selected here.
W A I 3 B I 2 C I D
When the calibration constant is registered in the empirical calibration method, <Used
calculated> can be selected in the formula. This can be applied to calculating correction
coefficients with the calibration constants registered as fixed.
<Log[curved]> is used to measure the film thicknesses of coatings etc. Although they can
also be measured using the FP method for thin films, <Log[curved]> can be used to
measure film thicknesses that cannot be measured using the FP method, such as films
containing organic matter.
The matrix correction, overlap correction and internal standard correction are available
for correction calculation. The matrix correction and internal standard correction can be
set only in the case of the empirical calibration method and cannot be used in the FP
quantification method. The overlap correction can be used for both the empirical
calibration method and the FP quantification method.
Matrix Correction
The matrix correction recognizes the condition that concentrations or film thicknesses
do not correlate with X-ray intensities because of coexisting components in a sample.
Matrix correction method can be selected from <Correct concentration> and <Correct
intensity>.
<Correct concentration> method normally used is expressed as a general formula shown
below including overlap correction and used for calculation.
RijFj
Wi ( AIi 3 BIi 2 CIi D)(1 Ki AijFj QijFjFk )
1 Wi / 100
BijFj DijFjFk Ci
The correcting values of Fj and Fk in the correction terms of Aij, Qij, Rij, Dij and Bij can
be used for either concentration correction or intensity correction. When expressed in an
uppercase character, concentration correction is used. When expressed in a lowercase
character, intensity correction is used.
Two methods are used for the matrix correction model: the dj correction method using a
binary-sample reference calibration curve and the j method using a straight-line
reference calibration curve obtained from samples with 0% concentration and with 100%
concentration. The JIS model is one of the dj method, and examples of the j method are
general correction methods of the Lachance Trail model and the de Jongh model and
special correction methods of the Rasberry Heinrich model and the Claisse Quintin
model.
The JIS, Lachance Traill and de Jongh models are widely used, but the other models are
not used so widely because they are in the stage of empirical formulae. This system
employs the general formula applicable to all correction formulae.
In general, the Aij term is employed for absorption/enhancement correction and the Bij
or bij term for overlap correction. The other correction terms are called special correction
terms.
formula shown below including overlap correction and used for calculation.
RijFj
Wi CIi(1 Ki AijFj QijFjFk )
1 Wi / 100
BijFj DijFjFk D
The meaning of each correction term is same to <Correct concentration> method. The
basic calibration formula when <Correct intensity> method is selected is Wi = CIi+D.
The calibration graph is plotted with the corrected X-ray intensities. Moreover, as the
special correction for fusion bead sample, <Correct intensity and BG> method can be
selected. This method is effective when the optional glass bead correction is included,
and D term of calibration intercept of <Correct intensity> method is made to D (1+RF)
and D term is corrected with dilution rate RF. When oxidizer correction method is
carried out, the correction with residual oxidant/sample weight ratio is also added.
Method Description
Enter published values Published values such as ones in JIS G 1256 for iron
manually. and steel analysis are entered manually.
Obtain with Coefficients are calculated with regression calculation
calculation. using many standard samples.
Theoretical correction coefficients are calculated with
Obtain theoretically.
the FP method. This method is generally used.
There are three matrix correction methods available namely <Correct concentration>, <
Correct intensity > and <Correct intensity and BG>, and the default setting is <Correct
concentration>.
<Correct intensity> is able to be selected only when the calibration formula is set to
<ratio> or <linear>. <Correct intensity and BG> is available for fusion bead sample.
The procedure to enter published values manually is described below. In this method,
values are entered manually for each necessary component and regression calculation is
carried out.
When the number of standard samples is small, do not use this method because accurate
calculation is impossible.
Here it is assumed that theoretical matrix correction coefficients have been calculated
and registered. For the procedure to calculate correction coefficients, see “Calculate
Theoretical Matrix Correction Coefficients” in “4.2 g.4 Calculation”.
Though concentrations are ordinarily used for term A in the general formula, X-ray
intensities can also be used.
To use any of the special correction methods, the Rasberry Heinrich model and the
Claisse Quintin model, follow the procedure described below. Set term R for the former
and term Q for the latter.
Overlap Correction
When the overlap of spectral lines exists, the overlap correction can be made. Overlap
correction coefficients to be used for this overlap correction can be calculated using the
regression calculation. The overlap correction can be applied to both the empirical
calibration method and the FP quantification method. Term B or b in the general
formula for the matrix correction is the overlap correction term.
In an application with the FP method, the FP method is selected initially for the
calculation type. In that case, an overlap correction coefficient can be calculated using
measured or theoretical intensities (option) or a fixed value can be entered.
The procedure to enter a fixed value for an overlap correction coefficient is described
below:
If no possible overlap line is found, the “Possible Overlap Lines” screen will not appear and
a message will be displayed.
Also in the case of the FP method, a correcting component can be set from possible
overlap lines.
If no possible overlap line is found, the “Possible Overlap Lines” screen will not appear and
a message will be displayed.
The overlap correction by <Theoretical intensity> is usable only when Quant. theoretical
overlap correction option is attached.
The matrix correction adjusts the calibration for effects of coexisting elements, while the
internal standard correction can correct errors resulting from thicknesses of samples. In
general, a sample that contains an element that has a peak near that of the measured
element line is added to all analyzed samples, and the ratios of the X-ray intensities of
the analyzed component to those of the added component are used for analyses. That
added component is called the internal standard sample.
As an example of such an application, one can use background intensities near the
measured peak without adding standard material to the samples. That method is called
the scattered X-ray internal standard correction. The internal standard correction
cannot be used for the FP quantification method.
In the case of a component for which background intensities are measured, <Background>
is displayed as an option for the internal standard correction. The above screen shows a
case where background intensities are measured for Si.
The automatic background subtraction and the scattered X-ray internal standard
correction are described below. Those functions are valid only in the case of the FP
quantification method and cannot be used for the empirical calibration method.
These functions are usable only when Quant Scattering FP method option is attached.
This function measures background intensities when measuring standard samples and
estimates background intensities when measuring unknown samples. The setting
procedure is described below. For further details, see “4.3 QUANTITATIVE ANALYSIS
TECHNIQUES”.
The options in the “Calculation method of BG calibration” frame can be selected when the
net intensity or gross intensity of the designated line has been saved.
When selecting samples to be used for calculation, this software distinguishes the
analyzed values of each sample. That is, the certified values are displayed in blue and
the uncertified values in yellow, so that you can distinguish them at a glance.
The procedure to select samples for calculation is described below:
Clicking Deselect All makes all samples non-selected state. When a few standard samples
are selected while a lot of standard samples are there, this function is effective.
This operation is valid after a regression calculation is executed and the result is saved.
The sample selection condition on saving is recorded, which helps when comparing it to
other sample selections.
1. Select a component for which a calculation was made using other sample selection.
2. Click Sample Selection .
3. Click Saved Selection .
4. Click OK .
When a regression calculation is carried out for a component that has a wide range of
concentration, samples with high concentrations may come near the calibration curve
and ones with lower concentrations may go off the curve. In such a case, the weighting
calculation is useful. The designation by measuring precision or effective digits of
analyzed values can be used. In the weighting calculation, the designation by measuring
precision to calculate using measured X-ray intensities is generally employed, but the
designation by the number of digits after the decimal point can also be set.
The setting to shift a calibration curve to samples with low concentrations using the
weighting calculation is described above. The fixed point calculation is a function to
make a calibration curve to be created pass through a designated sample. This fixed
point calculation can be used simultaneously with the weighting calculation.
g.4 Calculation
In most cases, one calibration curve can be used for the whole range of concentrations.
When the matrix of samples changes at a certain value or when a concentration range is
very wide, it may be impossible to cover the whole range with one calibration curve. In
such cases, quantitative analyses are possible by dividing the calibration curve and
using an individual calibration curve for each range. Up to three regions can be set and
individual divisions are displayed as Low (L when abbreviated), Middle (M when
abbreviated) and High (H when abbreviated).
The calibration curve division and calibration range setting is not available in the FP
method.
Printing of Result
Various formats are used to print regression calculation results depending on whether
the calibration curve division is used, whether the number of matrix correction
coefficients is 10 or less, whether the overlap correction is used for FP sensitivity
calibration and whether matrix correction coefficients alone are to be printed.
Graphs can be output not only to the printer but also to bit map files or metafiles.
Print Result
1. Click Print on the “Empirical Calibration Graph” or “FP Calibration Graph” screen.
2. Select <Printer>, <Bitmap> or <Metafile> for “Output”.
3. When the “Output” setting is not <Printer>, make setting for <File name> and <File
output folder>.
4. Click OK .
Selection for <File output folder> can be made easily by clicking Browse… .
Change Scale
You can designate an area using the common tool buttons and change the scale. You can
also click the “Scale Change” button on the “Empirical Calibration Graph” screen and
enter a value manually to change the scale.
To restore the initial scale designated in the measuring condition, click Default.
When there is a balance component, select the de Jongh model, JIS model or “Auto”.
When the base component is designated as the uncorrected component, set the
component with the highest concentration as the base component. To calculate the
correction coefficients of the base component, set the component with the second highest
concentration as the second base component.
When the main component is not included in the application condition like iron in steel, click
Addition to add the component as the <Balance> component and make a calculation.
When flux for a fusion bead, oxidizer and binder for a powder sample are all added to a
sample, <Flux> is displayed for the component type.
In the data table processing, measured X-ray intensities and analyzed results are
displayed in a table. Its contents can be checked and printed, and standard values can be
corrected.
Measured X-ray intensities and analyzed results are displayed in a table. Peak
intensities and background intensities can be displayed.
It is possible to copy a displayed data table to the clipboard. Press the Ctrl+C key to copy
data to the clipboard, and paste data to the spreadsheet software Excel or the like.
The contents of the data table are output to a text file that can be handled by a
spreadsheet program or the like.
This function is used to reconsider background subtraction after obtaining results using
quantification calculation.
When the “Subtract BG” setting is changed from <Yes> to <No> and a calculated result is
registered without background subtraction, delete background measurement from the
measuring condition of that component using “Parameters”.
This function is used to consider the optimum measured element line after obtaining
results using quantification calculation. To execute this function, two or more measured
element lines must have been registered beforehand as intensity correction element
lines of extra components. X-ray intensities must also have been measured.
The ordinary overlap correction uses measured X-ray intensities, while this function
uses theoretical X-ray intensities for overlap correction. This function can be used for an
application with the FP method.
Theoretical intensities for the overlap correction have been calculated for lines for
correction with * marks.
This feature functions only when Quant. theoretical overlap correction option is attached.
This function is valid only in the case of thin film samples. To analyze thin film samples,
an appropriate initial value must be set. In the initial condition, the average of the
standard samples used for calculation has been set. To change this value, use this
function. When the component type is “Fixed”, the value cannot be changed.
The calibration curve coefficients of each component can be displayed, and entered
manually if necessary. Setting can also be made for the calibration curve division.
The matrix correction coefficients of each component can be displayed, and entered
manually if necessary. Matrix correction coefficients (term A and term K) and overlap
correction coefficients (term B and term C) that are ordinarily used can be displayed as a
table. When <Display each component> is selected, special setting is possible as in the
case of the “Calibration” program. For the setting procedure for the manual input of
constants etc., see “4.2 g.2 Setting for Correction Calculation”.
The “Matrix Correction” tab is displayed when there is an empirical component.
Display and Manual Input of Overlap Correction and Internal Standard Correction
The overlap correction coefficients of each component can be displayed, and entered
manually if necessary. For the setting procedure for the manual input of constants etc.,
see “4.2 g.2 Setting for Correction Calculation”.
3. If necessary, enter a coefficient in the “a(quadratic)” box in the “BG sensitivity
coefficients” frame.
4. If necessary, enter a coefficient in the “b(linear)” box in the “BG sensitivity
coefficients” frame.
5. If necessary, enter a coefficient in the “c(constant)” box in the “BG sensitivity
coefficients” frame.
6. Click OK .
This screen is to check information obtained with regression calculation. Therefore, when
<Auto BG subtraction> is selected, any of the background sensitivity coefficients must not
be blank.
This function used by DMCA data makes re-extraction of X-ray intensity possible from
the DMCA data which saved by standard sample measurement. It is possible to change
the setting width of the PHA using the measurement data of the measured standard
samples without re-measuring the standard samples and reduce the burden of PHA
setting before standard sample measurement..
PHA:150-300 PHA:150-250
accuracy: 0.0060 mass% accuracy: 0.0024 mass%
3. Click Recalc . on the calibration graph, then input any value of LL and UL into the
<PHA> column in the recalculation screen and click OK .
4. The calibration line with newly input PHA setting width is indicated. Click Save and
make PHA setting conditions and intensities of standard samples in the standard
sample set renew.
The re-extraction of X-ray intensities by the newly changed PHA setting are done for the
standard intensities of drift correction using private conditions automatically.
If PHA condition width change is carried out after unknown sample measurement was done,
the re-extraction of X-ray intensities of unknown samples are not done. The PHA condition
width must be decided before unknown sample measurement.
The analysis control information is comprised of three parts: the limit ranges, check
samples and bias correction samples. The detailed setting procedures for these items are
described below, primarily for bulk samples. The only significant difference in the case of
thin film samples is that layer numbers are added and similar settings must be made for
each layer.
To control quantification values of analyzed samples, upper limits and lower limits are
set. The analyzed results on unknown samples are evaluated to determine if they are
within those standard ranges. When quantitative analysis is carried out, the limit range
check is made unconditionally except when <No limit> is set for the upper and lower
limits. If an analyzed value is out of the preset upper and lower limit range, its result
will be displayed. One can choose to set an upper and lower limit or only an upper limit
or only a lower limit for any quantification value. Two or more limit range conditions can
be created if desired.
The limit range name that is automatically issued first is an application name. That first limit
range cannot be deleted. Any editing such as deletion of a component is also impossible for
the first limit range.
When creating a new limit range condition, setting can be made so that a balance
component and a fixed value component that need not be set will not be displayed in the
table.
Delete Component
In the case of a limit range condition that has the same name as the application, a
component cannot be deleted.
Add Component
<No limit> can also be set by entering a zero in a cell for an upper or lower limit.
Change Unit
The unit selected here is used for an analyzed result. If it is different from the one set in the
application, the result will be converted using the unit selected here and output. For a
conversion factor, see the factor described in “6.1 f. Unit Setting”. However, a conversion
between mol% or at% and mass% is made with mol weight taken into consideration.
Enter Note
This function is used, for example, to fix the total of the analyzed values to a value
different from 99 mass%.
Set the number of sample preparation times when making a calculation for bias
correction using the averages of repeated measurement. This procedure is designed to
smooth the dispersion resulting from sample processing such as grinding.
When setting a sample ID on the Analysis screen for a component for which 2 or a larger
number was specified as “Bias correction sample prep. time”, set the corresponding bias
correction sample ID in the specified number of lines and set <Last sample> for the last line.
If the number of measured samples is less than the specified number, bias correction
coefficients will not be updated.
Check analysis is the method used to determine whether prepared calibration curves are
still acceptable. A sample with known concentrations is analyzed and the analyzed
values are calculated using the prepared calibration curves. The X-R control analysis, in
which the same check sample is measured three to five times a day (specified times) and
judged results of averages X and range R is recorded, is generally performed. The check
sample condition is created based on limit range conditions.
Besides, it is used for judgment of tolerance at the check analysis in case of execution of
auto drift correction. (For ZSX Primus IV/IVi only)
Delete Component
Add Component
Enter Note
For example, in the case of iron and steel samples, when the heat history in the
preparation of standard samples used to create the calibration curves is different from
that of unknown samples, a constant analysis error may appear. To correct this error of
analyzed values based on this calibration curve is called the bias correction. Usually,
samples with known analyzed values are measured after a calibration curve is prepared
and then the bias correction is set if necessary.
One can select either an intercept correction or a linear function for the bias correction to
utilize the calibration curves. The following formula is used for the bias correction.
Either correction method can be selected for each component.
W ' W A W B
Item Description
W’ Quantification value after bias correction
W Application result
A, B Correction coefficients
For example, let’s assume that true values are 1.1 times as large as calculated
quantification values for all samples. The method to correct that bias between true
values and quantification values is called the A method. In that case, set the
correction coefficient A to 0.1.
For example, let’s assume that true values are calculated quantification values
plus 0.5% for all samples. The method to correct that bias between true values and
quantification values is called the B method. In that case, set the correction
coefficient B to 0.5. This procedure is called the intercept correction method.
Correction by Function
This method is a combination of the A method and the B method, and is called the
AB method.
A bias correction condition is set for each limit range condition. Two bias correction
samples must be set for the AB method, and one sample for either the A or the B method.
6. Select <A method>, <B method> or <AB method> for “Method”.
7. Enter (a) sample name(s).
8. Enter the true value(s) of that component for “Standard Value”.
9. Enter (a) tolerance(s).
10.Click OK .
11.Repeat steps 4 to 10 for other components for which to carry out the bias correction.
The coefficient(s) is/are calculated automatically and need not be entered. Or only (a)
coefficient(s) can be entered to carry out the bias correction without setting a sample name.
In “Print Quant Application”, the following items can be printed. For items with
“Select…” buttons, detailed items can be selected.
To select all standard samples in each frame, click Select All under each frame.
Operational procedures for the following screens that appear when Select... is pressed
are the same. Refer to “Example of Operations”.
Example of Operations
1. Select <Sample> for “Drift correction” on the “Print Quant Application” screen.
2. Click Select... on the right of <Sample>.
3. Select drift correction samples to be printed.
4. Click OK .
a. Measuring Conditions
The characteristic X-ray spectral line used for measurement in quantitative analysis
such as the empirical calibration method or FP quantification method is called the
measured element line. The selection of this measured element line may affect analysis
precision and accuracy significantly. The measured element line to be selected varies
depending on the analyzed component, sample model, etc. The two distinct sample
models are bulk samples and thin film samples.
When the sample thickness increases from zero, the generated X-ray intensity changes
linearly. The region where this linear correlation holds true is called the thin film region.
That relationship can be used for the film thickness analysis for thin films etc. However,
the intensity does not increase indefinitely as the thickness increases. When the
thickness exceeds a certain threshold value, the X-ray intensity does not change as a
function of thickness. That region is called the bulk region, and the region between the
thin film region and the bulk region is called the medium thickness region. It is known
that X-rays with short wavelengths have high energy and, therefore, high penetration
through materials. Since X-rays with short wavelengths can therefore exit from the
sample surface even when they are generated in deep regions, those X-rays may be used
for concentration analyses for bulk samples. On the other hand, because X-rays with
long wavelengths have low penetration through material, those X-rays are used
especially for concentration analyses for thin film samples.
Purpose Description
Excitation Condition
Excitation
Description
Condition
Determines the kind of primary X-rays that irradiate a
Kind of Target
sample.
Tube Voltage Voltage applied to the target
Tube Current Current that flows through the filament
Primary Filter Eliminates unnecessary X-rays from the tube.
Tube Voltage
The voltage applied to the target increases overall intensity and primarily affects the
generation of the continuous X-rays. As the applied voltage is increased, the continuous
X-rays become more powerful. On the other hand, for a specific element to be detected, it
generates X-rays when it is excited with energy higher than a certain level. Even if an
element to be excited receives energy higher than the necessary level, the excitation
effect does not change. The relationship between optimum excitation voltages and
excited elements is shown below:
However, if the tube voltage and current are changed for each element in the
quantitative analysis, an extra measuring time will be necessary to change the voltage
and current and X-ray intensities will be unstable. Therefore, the same tube voltage and
current are generally used.
From the viewpoint of excitation efficiency, you must also consider the penetration depth
of the excitation X-rays. In general, as the applied voltage is increased, the excitation
X-rays penetrate deeper into a sample. And as the applied voltage is increased,
background intensities become higher.
Tube Current
This is the current that flows through the X-ray tube (electrons from the heated filament
to the anode). It significantly affects the intensities of X-rays generated from a sample.
The tube current is directly proportional to X-ray intensities to a first approximation. As
the current is increased, X-ray intensities become higher. The optimum excitation
condition should be selected from the combination of the tube voltage and current.
However, the maximum load of the tube must also be taken into consideration. The
maximum load is expressed by the following formula:
The tube voltage and current must be set so that the voltage multiplied by the current
will not exceed the maximum load.
Primary Filter
This is located between the X-ray tube and the sample and used to reduce any
interference of continuous X-rays or characteristic X-rays generated from the X-ray tube
to analyzed X-rays. This is also used to lower background. When the primary filter is
used, X-ray intensities are reduced. However, since background intensity is also lowered,
peak/background ratios become larger and lower limits of detection are improved. The
kinds and purposes of the primary filters are shown below:
Filter Description
Absorbs the Rh K-line.
Ni400
Effective for analyses using the K-lines of Ru, Rh, Pd, Ag, Cd, etc.
Ni40 Effective for trace analyses using the Pb L-line and As K-line.
Al125 Effective for trace analyses of Ti, Cr and Fe.
Absorbs the Rh L-line.
Al25
Effective for analyses of Cd-L etc. in thin samples such as films.
Optical Condition
The optical condition is set for the spectrum analysis of generated X-rays. The optical
condition includes the slit, analysis crystal and detector. The attenuator is used if
necessary to obtain suitable X-ray intensities. Details are described below:
Item Description
Slit Determines the resolution of a detected peak.
Crystal Separates X-rays generated from a sample.
Detector Counts separated X-rays.
Attenuator Reduces the intensities of generated X-rays.
Selection of Slit
The slit creates a parallel beam of X-rays generated from a sample. This is also an
important item to determine the resolution of a peak. Though the crystal used must also
be taken into consideration, the angular resolution is generally important for heavy
elements because there are many adjacent spectra. On the other hand, for light elements,
the sensitivity is important because there are not so many adjacent spectra.
Slit Description
S2 Standard slit for the detector SC. High resolution is obtained.
S4 Standard slit for the detector PC.
S8 Resolution lowers, but high X-ray intensities are obtained.
(ZSX PrimusIV)
Resolution higher than that of the S2 is obtained.
S1
Analysis Crystal
The analysis crystal separates X-rays that have a wavelength that satisfies the Bragg’s
formula and directs them to the detector. In general, selecting a crystal that has high
reflection intensities and high resolution contributes to analysis precision and accuracy.
Since very light elements generate X-rays with long wavelengths, a synthetic
multi-layer device (RX series) is used as the crystal for light element analysis.
Crystal Description
LiF(200) Best in resolution and reflection intensities
Ge Does not reflect secondary X-rays.
Has demerits of the large coefficient of thermal
PET
expansion and fragility.
TAP Used to separate Mg, Na, F and O.
RX25 – RX85 Synthetic multi-layer series.
Detector
The detector counts separated X-rays from the various crystals. In the ZSX, the
scintillation counter (SC) and the proportional counter (PC) are employed to cover a
wide range of measurement conditions. These detection methods are listed below:
Detector Description
Detector for heavy elements to be used with the ordinary
SC receiving slit.
Generally used for wavelengths of 0.01 to 0.3nm.
Detector for light elements to be used with the ordinary
PC receiving slit.
Generally used for wavelengths of 0.3 to 20nm.
In the case of the ZSX Primus or ZSX Primus III+, a counting range in which the
counting loss does not occur is up to 1000 kcps for the SC and up to 2000 kcps for the PC.
In the case of the ZSX Primus IV, it is up to 1800 kcps for the SC and up to 3000 kcps for
the PC. However, to carry out measurement exceeding 1000 kcps with the PC, set the
integral measurement in the PHA condition or set a wide window.
Attenuator
Sealing
In the analyses of liquid samples, the measuring chamber is filled with helium so as to
reduce the absorption of X-rays by the atmosphere. In the case of the helium
atmosphere, a seal with a thin film is placed between the spectroscopic chamber and the
sample chamber to separate the two chambers in this spectrometer.
By using the seal, it is possible to set only the sample chamber in the helium
atmosphere keeping the spectroscopic chamber in a vacuum. And this method can
reduce the consumption of helium gas.
PHA Condition
Fluorescent X-rays generated from a sample are separated by the crystal and counted by
the detector. When one X-ray photon enters the detector, one signal (pulse) is generated.
The number of pulses in a unit time is counted and defined as the X-ray intensity (kcps).
X-rays are separated according to the Bragg’s formula (n=2dsin), but X-rays with the
wavelength ', which is 1/n times the wavelength ('=n), also enter the detector. The
relationship between the wavelength (nm) and the energy E (keV) is as follows:
1.24
E
So, when the energy of the wavelength and that of the wavelength ’ are E and E’
respectively, the relationship E’=nE exists and the X-rays signals can be separated by
energy. Pulses with different heights therefore enter the detector successively, and they
are treated as time dependent data. A graph that has pulse height as the abscissa and
the number of pulses as the ordinate is called a PHA scan (see the figure below):
Number of Pulses (kcps)
Pulses counted by the PHA are shown below. Since these pulses may all be generated
simultaneously, the PHA condition must be set to count pulses only of an objective
component.
Signal pulses of For example, pulses generated when the W-M line of the
fluorescent X-rays from crystal RX26 (RX25) used to measure the Na-K line enter
crystal the detector.
Integral Measurement
Noises
No upper limit
Pulse Height
The fluorescent X-ray spectrum is expressed by a chart shown below. The abscissa
represents the 2-theta angle, which is the rotation angle of the detector. The standard
comparison method, in which the intensities of an unknown sample are compared to a
calibration curve prepared from a correlation between standard samples concentrations
and their corresponding X-ray intensities and the concentrations of the unknown
samples are calculated, is called quantitative analysis. In quantitative analysis, X-ray
intensities, which become the ordinate axis of a calibration curve, are measured with the
fixed-time counting method. The fixed-time counting method means collecting X-ray
intensities during a specified time while fixing the crystal and detector at a certain angle.
This angle is typically determined by a 2-theta scan prior to quantitative analysis. In
general, the characteristic X-ray intensity of a measured component is counted at its
peak position with fixed-time counting and a calibration curve is created.
100
90
80
Count Rate [ - ]
70
60
50
40
30
20
10
0
10 20 30 40 50 60 70 80 90
2-theta [deg.]
When background intensities are sufficiently low compared to peak intensities (the
peak/background ratio is large), a good calibration curve may be obtained by measuring
only a peak position. However, if background intensities are higher than variations due
to differences in concentrations (the peak/background ratio is small), a calibration curve
with poor analysis precision will be created. Generally in such a case, background
positions are measured in addition to the peak to create a calibration curve using net
intensities after subtracting background intensities. Thus, the peak and background
condition obtains net intensities to prepare a calibration curve. The methods to subtract
background intensities are shown below:
Background
Description
Subtraction
One or two background positions are designated.
1/2-Point Method
Only the linear method can be selected.
Up to 16 background positions are designated.
Multi-Point
Selection can be made from the linear, quadratic, cubic, hyperbolic
Method
and Lorentz functions.
To calculate net intensities and create a calibration curve, background intensities are
measured. The criterion to select a background subtraction method is shown below.
When deciding whether to measure backgrounds, sufficient consideration is necessary
because the measuring time is extended as the number of measuring positions becomes
larger.
1-Point Method
As shown on the left side figure below, when the background intensities on both sides of
a peak are identical, the net intensity can be calculated by measuring only one
background position. In the case of the 1-point method, setting the background position
on the high angle side is recommended for the angular resolution.
Net Net
Intensity Intensity
Background Background
Intensity Intensity
= =
2 2
2-Point Method
As shown on the right side figure above, when the background intensities on both sides
of a peak are not identical but the background intensity can be fit with a straight line,
the net intensity can be calculated by measuring two background positions.
The background subtraction coefficient is calculated with the 2-point method described
above. But, in real measurement, only one background position is measured and
background intensities are calculated. That background measuring position must be a
background position used for the calculation.
As shown on the left side figure below, in the case of the quadratic or cubic formula for
the multi-point method, background is expressed using a function. A quadratic or cubic
curve is used instead of a straight line. Three or more background positions must be
measured in the case of the quadratic formula, and four or more background positions
must be measured in the case of the cubic formula.
Net Net
Intensity Intensity
Background Background
Intensity Intensity
2 2
So far background calculations for gentle background shapes such as continuous X-rays
have been described. As shown on the right side figure above, estimating the background
of a small peak on the shoulder of a giant peak by employing the Lorentz function or
hyperbolic function to approximate the large peak better is useful. Four or more
background positions must be measured in the case of the hyperbolic function, and five
or more background positions must be measured in the case of the Lorentz function.
2 2
If you cannot determine which measured element line is optimal for a calibration curve
or a sensitivity calibration curve, set several element lines and perform a measurement.
In the regression calculation described later, a measured element line can be changed
when a component has two or more element lines. In that case, you can set two or more
element lines as <correction lines>.
For example, to analyze tin (Sn) in the analysis of polymer samples, you can set the
Sn-L line as correction lines in addition to the Sn-K line, which is generally used as
the measured element line. In the regression calculation, if the accuracy of the
calibration curve using the Sn-K line deteriorates because of thickness, you can change
the measured element line to the Sn-L line and carry out a regression calculation.
×
×
×
×
Concentration/
Film Thickness
Accuracy is bad
a.Right standard
Enter right values
values? No
Yes
Yes
c.Right measuring
Set right conditions
conditions? No
Yes
d.Peak overlap
Overlap correction
effects? Yes
No
e.Matrix correction
Make sure that the correct standard values have been entered.
Do peaks overlap? Is there any overlap of energy? Make checks again using the
2-theta scan and the PHA scan respectively. In such a case, to solve the problem
sooner, it is important to check a sample that was not monitored in the measuring
condition setup described before. Carry out a 2-theta scan and a PHA scan
specifically for a sample off the calibration curve. If it seems to be impossible from
a 2-theta scan result to eliminate overlap of spectral lines, it is recommended that
the overlap correction described later be made.
×
×
If the accuracy of the calibration curve does not improve, carry out matrix
corrections.
If the accuracy of a calibration curve does not improve even after measurement is
optimized using the measuring condition and overlap correction, coexisting
elements affect the measured component. Basically all samples are affected more
or less by coexisting elements. Although in most cases using standard samples
with the same composition as unknown samples can solve the problem, matrix
corrections are used to correct any residual effects. The following three methods
are available, but in general the method to calculate theoretical correction
coefficients using the FP method is employed:
Matrix Correction
Description
Method
For example, in the case of iron and steel samples,
Use published
matrix correction coefficients published by JIS are
correction coefficients
available.
Coefficients are calculated using many standard
Calculate using the
samples that have concentrations that do not
regression calculation
correlate with each other.
Calculate using the FP Theoretical matrix correction coefficients are
method calculated using the FP method.
Matrix Correction
X-ray intensities of all components in a sample are affected by the other components to a
greater or lesser degree. In general, fluorescent X-rays generated from a certain
component are absorbed by all the components in the sample, including itself. In some
cases, secondary enhancement occurs by fluorescent X-rays generated within the sample.
The phenomenon that X-ray intensities do not correlate with a calibration curve due to
these issues is called the matrix effect. Correcting this phenomenon caused by the
matrix effect using a regression calculation is called the matrix correction.
Matrix correction can be selected from [Correct concentration] and [Correct intensity].
[Correct concentration] normally used is expressed as the linear equation below
including overlapping correction and used for calculation.
RijFj
Wi ( AIi 3 BIi 2 CIi D)(1 Ki AijFj QijFjFk )
1 Wi / 100
BijFj DijFjFk Ci
The correcting values of Fj and Fk in the correction terms of Aij, Qij, Rij, Dij and Bij can
be used for either concentration correction or intensity correction. When expressed in an
uppercase character, concentration corrections are used. When expressed in a lowercase
character, intensity corrections are used.
In general, the Aij term is employed for absorption/enhancement correction and the Bij
or bij term for overlap correction.
[Correct intensity] is different at the intercept D of calibration line compared with
[Correct concentration], but as it is based on the same point of view, it is expressed as
below and used for calculation.
RijFj
Wi CIi(1 Ki AijFj QijFjFk )
1 Wi / 100
BijFj DijFjFk D
To carry out matrix corrections in the empirical calibration method, the following three
methods can be used: a) use published correction coefficients, b) calculate
experimentally using the regression calculation and c) calculate theoretically using the
FP method. These methods are described below in detail:
Matrix correction coefficients for iron and steel are published. They are described
in JIS G 1256 and are available for Fe base and Ni base systems. Those correction
coefficients can be set as fixed values in the regression calculation. They basically
correct absorption/enhancement terms.
1) Variations in concentration must not correlate with each other among the
standard samples.
For example, let’s consider samples that have Fe as their main component and
contain Al and Si. Let’s assume that concentrations of components correlate with
each other. For example, when Al concentrations increase, the Si concentrations
increase, and when Al concentrations decrease, Si concentrations decrease. Such
samples cannot be used to calculate matrix correction coefficients. When
calculating correction coefficients, concentrations of components must not
correlate with each other.
Overlap Correction
The peak of a measured component may overlap with the peak of another component. In
general, a measured element line that does not overlap with another peak is selected.
However, if a spectral line with low relative intensity is selected, the accuracy of a
calibration curve may deteriorate. In such a case, overlap correction is made. The degree
of effect varies depending on the concentration of an overlapping element, so that effect
is corrected. Up to five components can be used as correcting components. For the
overlap correction, the above-mentioned matrix correction method or the X-ray intensity
method described below can be used. The method using X-ray intensities can be used for
both empirical method and FP method.
I ia I ib B j I j
Item Description
Iia X-ray intensity after correction
Iib X-ray intensity before correction
Ij X-ray intensity of correcting element j
Bj Overlap correction coefficient
This correction is used for the application in which a constant amount of material is
added to an unknown sample for correction. A component that generates X-rays with a
wavelength near that of a measured component is added. Generated fluorescent X-rays
are affected by other components in a sample and their intensities change. Since the
X-ray wavelength of the added component is near that of the measured component, the
degree of effect from other components is regarded as almost identical. This method can
therefore correct the effects of the matrix.
Ii
Ia
Ij
Item Description
Ia X-ray intensity after correction
Ii X-ray intensity of corrected component
Ij X-ray intensity of correcting component
In the background ratio method, ratios of net X-ray intensities to background X-ray
intensities are used for the ordinate axis of the calibration curve. Generated fluorescent
X-rays are affected by other components in a sample and their intensities change. Since
the background X-ray wavelength can be regarded as similar to the peak X-ray
wavelength, the degree of effect from other components is regarded as almost identical.
This method can therefore correct the effects of the matrix. This method is especially
effective in analyzing minor heavy elements in samples that have light elements as their
main components.
I n et
Ic
I BG
Item Description
Ic X-ray intensity after correction
Inet Net X-ray intensity before standardization
IBG Background X-ray intensity
When the concentration range of a component is wide and a single regression calculation
cannot cover the entire range, a calibration curve can be divided. This method is called
the calibration curve division, and a calibration curve can be divided into one, two or
three regions. The calibration curve formula can be different in each region. For example,
a linear formula can be used in the low concentration region and a quadratic formula can
be used in the other wide concentration regions. The three divisions are called Low,
Middle and High divisions respectively and the Low division is used for a calibration
curve when there is no division.
c. FP Quantification Method
The applicable type of the FP quantification method judged from prepared standard
samples is described using the flowchart shown below. The accuracy of <I.Calculate like
empirical calibration method> is the best with regard to indeterminate factors.
Start
In the empirical calibration method, standard samples that have the same
concentrations as unknown samples are necessary. But, in the FP quantification method,
even when such standard samples cannot be prepared, quantitative analyses are
possible. There is no requirement for standard samples when calculating theoretical
intensities. This is the case not only with bulk samples but also with thin film samples.
Seriously consider the following in the quantitative analysis of thin film samples using
the FP method:
In the case of a single-layer thin film, setting the constituent components is sufficient. In
the case of a multi-layer thin film, the order of layers must also be set. Since theoretical
X-ray intensities can be calculated only when the order of layers is known, it must
always be set. Conversely, the order of layers cannot be obtained using the FP method.
In the thin film analysis, information about a thickness is calculated as weight. The
density is therefore necessary to analyze a thickness. A density may be calculated from
components constituting the layer from the quantitative analysis. However, to generate
a more accurate analysis, when the density of a layer is known, set it as a fixed value.
When executing the weight conversion, a density is unnecessary.
When the Ni+P layer is so thin that Al-K can penetrate it, this setting can be
used. That is, the degree of absorption of Al-K is used for a calculation.
In the above example, all components can be measured. But when a sample model is
complicated and usable X-rays are limited, the method like Item 3) can be used. Thus, by
considering a sample model and extracting analyzable X-rays, the thin film FP
quantification method can be applied to a complicated system.
When Standard Samples for Thin Film Samples Are Not Available
In the thin film FP quantification method, when standard samples that have the same
layer structure as thin film samples are not available, set standard samples as shown
below. Let’s consider the sample of an Al plate with Ni and P coated on it. Let’s assume
that only bulk samples of P and Ni are available as standard samples.
Al No substrate No substrate
The layer structure is set for each standard substance like the sample model. For
example, in the case of the P sample, Al is designated for the substrate but its thickness
is set to zero, and Ni is designated for the composition but its concentration is set to zero.
In the case of the Ni sample, by using the same setting, it can be regarded as the same
structure as that of unknown samples and can be processed as a standard sample.
The types of usable scattered X-rays and their symbols are as follows:
These are the Thomson scattering X-rays of the characteristic X-rays of the target
element. Since the Thomson scattering X-rays are scattered without losing energy,
they appear in the diffraction position of the target element.
“T” is attached after ordinary spectral names like Rh-KT and Rh-K1T in the
case of the Rh target, for example.
These are the Compton scattering X-rays of the characteristic X-rays of the target
element. Since the Compton scattering X-rays are scattered with some loss of
energy, they appear at an angle higher than the diffraction position of the target
element by about one degree.
“C” is attached after ordinary spectral names like Rh-KC and Rh-K1C in the
case of the Rh target, for example.
These are the scattered X-rays of the continuous X-rays and are the so-called
background.
Like the theoretical intensity calculation for fluorescent X-rays, theoretical intensities
are calculated for the scattered X-rays from the composition of a sample and physical
constants. Using the atomic scattering factor for the Thomson scattering X-rays and
using the Compton scattering cross section for the Compton scattering X-rays,
theoretical intensities are calculated taking absorption by a sample into consideration.
Theoretical background intensities are calculated as the sum of Thomson scattering
X-ray intensities and Compton scattering X-ray intensities.
For example, the Thomson scattering X-rays of primary X-rays are set as the measured
element line, and the thickness is analyzed by comparing calculated theoretical X-ray
intensities with measured X-ray intensities. In that case, standard samples can be set by
assigning that measured element line to the component for thickness measurement. The
setting procedure is the same as the ordinary one except that scattered X-rays are set as
the measured element line.
Let’s assume that iron in iron ore powder or copper in copper concentrate powder is to be
analyzed using the empirical calibration method. You can use the method to carry out
matrix correction with a calibration curve whose ordinate axis is the ratio of Fe-K or
Cu-K X-ray intensities to scattered X-ray intensities of the characteristic X-rays of the
X-ray tube target. Since the calibration curve is created using intensity ratios to
scattered X-rays, this method is called the scattered X-ray ratio method. That is used to
reduce the mineral effect or grain size effect characteristic of powders.
For example, in the case of the Rh target, Rh Compton scattering X-rays are set as the
measured element line for correction. When selecting the measured element line of Fe,
set Rh-KC as the internal standard. Thus, ratios to Compton scattering X-rays are
used for the ordinate axis. To calculate the theoretical matrix correction coefficients,
select the scattered X-ray ratio method in a calculation condition for Fe. After that,
make the appropriate setting corrections in the same way as the ordinary calculation of
matrix correction coefficients.
Geometry effect
In case of the measurement of liquid sample put in the sample cell, the detectable area of
fluorescence X-ray excited in the sample varies according to the height of the liquid sample.
This area is defined by the position of optics of incident side (X-ray tube side) and receiving
side (analyzing crystal・detector side) (Geometry effect).
low low
incident emission
incident
Application example of correction of geometry effect (For ZSX Primus IVi only)
Following is the FP sensitivity curve of Cd-Ka in case of use of 5g and 10g oil multi
element samples. In case of conventional model, sample weight is taken into account but it
is obvious that FP sensitivity curve divides into two lines caused by the difference of sample
weight, 5g or 10g, which makes difference of sample height. This is caused by that the
difference of sample height gives the difference of detection area (geometry effect). On the
other side, in case of geometry effect correction model, same sensitivity curve for both 5g
and 10g are obtained. This fact proves successful correction of error due to difference of
sample height.
The Result Display program manages all sample IDs, which include the history of
measurement such as quantitative analysis results. As for data, this program collects
mainly quantitative analysis and qualitative analysis results. Since qualitative analysis
results have different data processing methods, they are treated separately with the
Qual Result menu, but recalculation results are reflected in this Result Display data.
This program also handles, on the maintenance record menu, data to monitor the
stability of the quantitative analysis such as check analysis results and drift correction
analysis results, and spectrometer control data such as PHA adjustment records.
Functions of this program are as follows:
1) All quantitative analysis and qualitative analysis results are recorded and can be
displayed freely by selecting sample IDs.
2) By copying a measured sample ID to the “Sample ID Setting” screen, the same
measurement can be made again.
3) Quantitative analysis results can be collected and a plotting chart can be drawn.
4) Result data can be saved by designating a file format.
5) Collected data can be transferred to an external device.
6) Collected data can be printed.
When the Analysis program has been started, this program becomes the “Analyzed
Result” screen.
The upper part is called the data selection area and the lower part the result display area.
The size of the display area can be changed using the dividing line.
a. Result Display
Since a sample ID entered in the Analysis program is recorded intact, its analyzed result
can be displayed by designating that sample ID. Though the main purpose is to display
quantitative analysis results, qualitative analysis results can also be displayed.
However, since this function cannot process qualitative analysis results, by processing
data using the Qual Result menu, its results can be reflected. In some cases,
remeasurement becomes necessary after checking a measured result. In that case, that
sample ID can be copied and used in the Analysis program. Display functions and
retrieval functions will mainly be described below:
By selecting a sample ID line left in the data selection area, an analyzed result can be
displayed. Since there is no blank line like the sample ID preset table in the Analysis
program, a result is always displayed.
Switch Display
In the default condition of the quantitative analysis result display, quantification values
such as concentrations and film thicknesses are displayed. X-ray intensities can also be
checked here.
The intensity type of displayed data when <Intensity> is selected or data for transfer
can be selected.
1. Select two or more sample IDs in the data selection area.
2. Select [View] [Multi data Draw].
Sort by Item
This operation selects an objective analyzed result from a list of data. For example, when
sorting by the analysis type is selected, measured data can be sorted for each analysis
type regardless of analysis dates and times or sequence numbers. Data items are
arranged by default in the order of the analysis date and time.
1. Click title items such as Type , Application and Sample name in the data selection
area.
Displayed analyzed results or X-ray intensities can be copied to an Excel file or the like
thorough the clipboard.
The copying and pasting operations are possible even when two or more sample IDs are
selected and two or more data items are displayed.
Data Retrieval
When the number of measured samples becomes large and the analyzed result data
quantity increases, retrieval using the data sorting function alone may be difficult. In
addition, to draw a plotting chart described later, only samples of the same application
must be selected. In such cases, specific desired data alone can be retrieved and
extracted to the selection area. Retrieval items can be selected from the analysis date
and time, type, application and input memo. That procedure is described below:
There is a filter function with the dropdown button in the column header.
1. In the data selection area, click the dropdown button of an item for which to carry
out data extraction. A list of options is displayed according to the item. A list of
analysis types is shown below:
In a column for which a filter has been set, the filter icon is displayed instead of the
dropdown button.
Recalculation Procedure
Data to be displayed by the Result Display program can be specified using a folder for
saving on analyses and can be saved in another folder using the Result Display program.
A folder for display using the Result Display program is specified with the following
procedure:
The folder for display is always displayed in the upper part of the Analyzed Result
screen.
Folders to save quantitative analysis results on analyses are the folders AnlRslt and
AnlRsltU and their subfolders.
The specified folder for display is saved. If it is different from the one for saving on analyses,
results are not displayed automatically on analyses, In this case, change the folder and
then reselect the one for general use.
Collection processes such as statistics calculations and plotting charts are performed on
stored quantitative analysis data. Basically, in the quantitative analysis, management
of analyzed results is necessary. Collected statistics calculation results can be printed
and transmitted. The procedure to draw plotting charts is described below. However,
plotting charts for X-ray intensities cannot be drawn.
For the operational procedure for the X-R control analysis to obtain statistics of results
of check analyses, see “6.2 g. Record for Check Analysis Result” in the Maintenance
menu.
Plotting Chart
A chart to check the trend of analyzed results for each component is called the plotting
chart. Statistics calculation results such as averages, maximum values, minimum
values, ranges, standard deviations and coefficients of variation are displayed. A
histogram can also be drawn.
The operation flow for the collection of analyzed results is shown below:
Start
Draw histogram?
No
Yes
Plotting chart + histogram
End
The ordinary procedure to draw a plotting chart is described below. The following
procedure displays a plotting chart and statistics. To set other necessary processes,
make setting changes by referring to procedures described later before clicking OK .
1. Retrieve objective measured results with the application name or the like.
2. Select measured results for processing in the data selection area.
3. Select [View] [Plotting].
4. Select a component for processing.
5. Click OK .
After the above operation, a time-series plotting chart is displayed on the left and
statistics are displayed on the right, as shown below:
On the initial drawing, the vertical scale is automatically determined from analyzed
results. That vertical scale can be set arbitrarily. An absolute setting or a relative setting
can be used. In the absolute setting, enter an integer or a real number. In the relative
setting, enter a percentage.
Horizontal Scale
To set the horizontal scale, a time format or a data points format can be used. In the time
format, start and end dates are designated and a time-series plotting chart is drawn. In
that case, data intervals are time intervals and therefore may not be consistent. On the
other hand, in the data points format, a plotting chart with a constant interval is drawn.
Compare those formats and draw a plotting chart suitable for your purpose.
To set a date, the standard date setting function of Windows can be used.
To check the variation condition of results drawn on a plotting chart, reference lines can
be displayed. For the reference lines, one, two or three times the standard deviation ()
or the limit range values set in the application can be selected.
Display Histogram
A step-function graph that has the horizontal axis of a quantified physical parameter
characterizing an event and the vertical axis of the number of its observations is called
the histogram. This program displays a histogram rotated clockwise by 90 degrees to the
right of a plotting chart. The number of divisions for the horizontal axis is set to 7 to 13
and the right and left parts from the center are divided into three to six divisions
respectively, in order to indicate the degree of shifts from the average.
When a histogram is not displayed, statistics are displayed to the right of the plotting
chart. When a histogram is displayed, statistics are not displayed. If desired, a detailed
histogram and a statistics graph can be drawn.
2. Double-click the histogram on the right side of the display area.
A detailed histogram is drawn from a plotting chart. Therefore, when the plotting chart is
closed the detailed histogram is also closed.
Updated results of drift correction can be printed using Record for Drift Correction Result.
PHA adjustment results can be printed using Record for PHA Adjustment Result.
The result of a library sample cannot be printed in the “Result Display” program. Print the
data using the “Library Sample” or “Qual. Data Handling” program.
Print Setting
6. Enter a format name for Single Result Print for “Format”.
7. Enter a title in the “Title” frame.
8. Click Setting in the “Header” frame.
9. Select an item to be output as header information under “Type” for each item.
10.Enter a heading under “Heading” for each item.
11.Select <Left>, <Centered> or <Right> under “Align Text”.
12.Click OK .
13.When <Analyzed result> was selected for “Print Format”, select <Final result> or
<Application result> in the “Data” frame.
14.Select data to be output in the “Data” frame.
15.Select <Unit> if necessary.
16.When <Analyzed result> was selected for “Print Format”, select <Color output for
out of range> if necessary.
17.If necessary, select <Reporting date> and enter its format.
18.Select <Logo output> if necessary.
When <Input memo> was set for the type of item, its contents can be entered in the printing
dialog. It can be used to print an operator name or the like.
An example is shown below in which the paper direction is <Portrait> and all header items
are aligned to the left under a title:
Entered Title
An example is shown below in which the paper direction is <Landscape> and header items
are aligned to the left, center and right under a title:
Entered Title
12.Click OK .
13.Select data to be output in the “Data” frame.
14.Select <Unit> if necessary.
15.Select <Color output for out of range> if necessary.
16.Select a sample name width if necessary.
17.Select <Output statistics> in the “Statistics” frame.
18.Select output items in the “Statistics” frame.
19 Select <Reporting date> and enter its format.
20.Select <Logo output> if necessary.
21.Click OK .
For the sample name width, specify the number of printing units required for the sample
name. When the sample name width is 1, approximately 10 characters can be used. When
the sample name is long, increase the sample name width.
When two or more sample IDs to be output have been selected, <Multi-result> is selected
on the “Print” screen.
Quantitative analysis results can be transferred by designating their sample IDs. As for
the text format, the transmission format, spreadsheet format, HTML format and output
hazardous elements format (option) are available. As for the transfer method, the disk
saving, e-mail and external transmission are available. The text format and transfer
method that can be used vary depending on the data type.
Objects of Transfer
Analysis Type Item
Intensity
Quantitative Analysis
Analyzed result
Intensity Measurement Intensity
Intensity
Check Analysis
Analyzed result
Drift Correction Sample Intensity
Standard Sample Intensity
Intensity
Bias Correction Sample
Analyzed result
Intensity
Qualitative Analysis
Analyzed result
In the case of analysis types that have analyzed results, selection can be made from the
transfer of analyzed results only and that of both analyzed results and intensities.
To output intensities, select the type of intensities from <Net intensity before drift
correction>, <Net intensity after drift correction> and <Internal std and overlap correction>.
When the output hazardous elements format (option) is selected, data to draw up a report
on hazardous elements can be output.
In the case of mapping measurement data, more than one sample ID can be selected in the
spreadsheet format. In other cases, only one sample ID can be selected.
The procedure to transfer quantitative analysis results using the CSV format in the
spreadsheet format is used as an example for explanation.
To use e-mail for transfer, its destination must be set beforehand using [Utility] [System
Management] [Data Transfer Setting].
To use the external transmission, its transmission condition must be set beforehand using
[Utility] [System Management] [System Parameters].
Only when <Transmission format> was selected for “Text format”, <RS232C
Transmission> or <TCP/IP Transmission> can be selected for “Trans. method”.
When <Output hazard. elem.> (option) is selected for “Text format” and data transfer is
carried out, XX.csv and XX_data.csv (XX: specified file name) are created in the specified
folder.
A quantitative analysis result can be saved in a specified folder by selecting a sample ID.
By specifying a folder for data display with Result Display, saved data can be displayed.
By selecting a sample ID, the “Qual Data Handling”, “SQX Calculation” or “Mapping”
program can be started directly according to a data type.
Even when a qualitative analysis result or a mapping file is specified, a data processing
program cannot be started if that file does not exist in the specified folder.
When a mapping file is specified in the qualitative analysis, “Qual Data Handling” and “SQX
Calculation” cannot be selected.
Since already measured X-ray intensities are used, there is no item concerning X-ray
intensity measurement. Since check samples are unnecessary, there is no item
concerning setting for them. Consideration on a quantitative analysis is possible without
measurement. For example, consideration is possible on the effective depth of the X-rays
of a certain component.
The operational procedure is basically almost the same as that for the quantitative
application. For details, see “4.2 CREATION OF QUANTITATIVE APPLICATION”.
Only items concerning the simulation are described below.
The operational flow for the Quantitative Simulation is shown below:
Start
a. Creation of Application
e. Regression Calculation
End
For <Rename and save the application> and <Delete an application>, the operational
procedures are the same as in the case of the quantitative application. See “4.2 a. Creation
of Application File”.
In the case of <Quant application>, selection items on the right do not appear. After
pressing Next>, the “Select an existing quant application” screen appears.
When a template is used, only the empirical calibration method can be selected.
To create an application with the FP method, select <Create your own application>.
For the operational procedure for “Application Information” (“Component Selection” and
“Sample Preparation”), see “4.2 b. Setting of Application Information”.
Most operations are the same as those for standard sample setting for the quantitative
application. As for copying from other applications, a function to copy an application
dedicated for the simulation can be used.
The operational procedure for analysis conditions (analysis parameters and special
standard samples) is also almost the same as that for the quantitative application.
Differences in the case of the simulation are described below. As in the case of the
quantitative application, special standard samples are set only in the case of the FP
quantification method.
Differences in options for the component type on the “Property of Sample Component”
screen for each component are shown below:
The item of “External parameter” does not appear on the “Property for an Extra
Component” screen. The other operational procedures are the same as those for the
quantitative application.
The measuring diameter and atmosphere need not be set. The Advanced button does not
appear. The angle and time are not displayed. The incident angle is added as a setting
item.
The following items are displayed in the case of the quantitative simulation:
Item Description
Element line(El. line) The same as the quantitative application
A target element and an X-ray tube type are
Target
selected.
Tube voltage(kV) The same as the quantitative application
Take off angle A takeoff angle is entered.
A filter element is selected.
Filter
A filter thickness is entered.
For the target, selection can be made from the following three types. The initial setting is
the target of the X-ray tube used.
For the operational procedure for the calculated parameter, see “Add Process Parameter”
and “Setting of Calculated Parameter” in “4.2 d. Setting of Analysis Information”.
Special standard samples can be set only when the application has been created with the
FP quantification method. In the FP method, unlike the empirical calibration method,
standard samples that have different composition from that of unknown samples can be
used.
In the case of the Quantitative Simulation, the item of the background intensity does not
appear in the “Measurement” frame on the “Property for Component” screen. The other
items are the same as those for the quantitative application.
d. Regression Calculation
In the regression calculation for the Quantitative Simulation, all items concerning
background in a quantitative application do not appear. Other differences are also
described below:
The BG Subtract/Internal Standard button changes to the Internal Standard button on the
“Correction for FP Method” screen that appears when the Correction button is clicked,
and <Scattering internal standard correction> alone is set.
Sample Selection
Differences are shown below concerning <Weighting> on the “Sample Selection” screen
that appears when the Sample Selection button is clicked:
Others
In the case of the Quantitative Simulation, the “Auto BG subtraction” tab cannot be
used.
When unknown samples have been measured using a quantitative application created
beforehand, one can recalculate the results for the sample after changing contents of the
application. It does not mean that measuring conditions are really changed requiring
measurement of the standard samples again. For example, in an application for which
background measurement was carried out, a recalculation can be made by creating a
calibration curve using gross intensities without subtracting background intensities.
Those changed contents are not reflected in the original application and are valid only in
this Quantitative Simulation program.
When <Quant application> is selected on the “Application File” screen, a result file can
be read. Based on quantitative analysis results using the quantitative analysis and
intensity measurement, recalculation and unknown sample quantification can be
carried out.
Two or more result files can be selected. Select necessary ones referring to sample names,
measuring coordinates and analyzed dates/times.
When sample data has already been set, selected result data is added after that. However,
if the number of samples exceeds 100, data will not be added any more. When result data
for which a sample name has not been set is read, “no name” is set as its sample name.
<Cut >, <Copy>, <Paste > and <Insert> can also be performed by right-clicking at an
objective line number or sample name.
When <Measure (EMP)> or <Measure (FP)> has been set for a component type, it can be
changed to <Fixed>.
X-ray intensities etc. for “Unknown Calculation” can be copied to an Excel file or the like.
X-ray intensities etc. for “Unknown Calculation” can be copied from an Excel file or the
like.
1. Copy data using the software of the source of copying, such as Excel.
2. Select and right-click a cell from which to start pasting on the “Unknown
Calculation” screen.
3. Select [Paste] on the menu.
In “Print Quant Application”, the following items can be printed. For items with
“Select…” buttons, detailed items can be selected.
To select all standard samples in each frame, click Select All under each frame.
Setting Items
Qualitative Spectrum
Component Input
Measurement and Composition
Analysis of Standard Samples
Sample Information
Input
Sample Form &
Standard Values
Standard Sample
Measurement ・Automatic Theoretical
Matrix Correction
Calibration Curve
Coefficient Calculation
Creation:
・Automatic Overlap
Matrix Correction
Overlap Correction Correction Setup
Theoretical matrix correction coefficients are used for the matrix correction. The
concentration correction is employed for the overlap correction.
Shown below are sample types, quantification element ranges and component forms to
be used for the automatic quantitative application setup:
Fundamental
Quantification Balance
Sample Type Component
Element Range Component Setting
Form
Metal & Alloy Mg - U Metal Auto/Manual
Powder F-U Oxide Auto/Manual
Na - U
Fusion Bead Oxide Auto/Manual
(Excluding Tc - Cd)
Ceramics &
F-U Oxide Auto/Manual
Glass
Auto (Carbon)
Polymer F-U Metal
/Manual
Since settings are automatically made using the qualitative analysis data of standard
samples, this function cannot automatically create optimum analysis conditions. In any
of the following cases, create an application using the ordinary quantitative application
setting procedure:
In the case of a trace element that is difficult to detect using the ordinary
qualitative analysis.
When a sample has a small diameter and detection is difficult even for
components that are not trace elements.
When a sample is formless and smaller than the measuring diameter.
When the ununiformity effect of a sample is large.
Although the internal standard method and addition method are known as correction
methods for the quantitative analysis, it is impossible to use any of those special
corrections.
Optical Conditions
Measuring conditions are set with the standard common library conditions using the
standard crystals (LiF(200), PET(H), Ge(H) and RX26 (RX25)) and the slits S2 and S4.
Therefore, it is impossible to use other crystals, PHA conditions, slits or the output
conditions of the X-ray tube.
Start
End
Only <Metal & Alloy>, <Powder>, <Fusion Bead>, <Ceramics & Glass> and <Polymer>
can be used as the sample type.
The application information setup consists of setting items of the component selection
and sample preparation information.
Only the empirical method can be used as the analysis method for a measurement
component.
When a special compound exists in a powder sample or the like, carry out the compound
registration beforehand.
Shown below is a screen on which Fe has been set as the balance component:
The automatic setup function may judge that components other than quantification
components affect the quantification components and automatically set them additionally.
The FP method is defined for added components (manual addition is impossible).
Shown below is the “Component Selection” screen after only Cr is set as a quantification
component and the automatic setup is carried out. Mn, Fe, Ni and Mo have
automatically been added as FP method components. It is impossible to delete
automatically added components.
Automatically Added
Components
For the automatic setup of analysis conditions, it is necessary to set sample information
correctly. As shown below, the setting contents of the sample preparation information
vary depending on a sample type. Sample types are shown below:
1. When a sample film is used for an analysis sample, set information on the sample
film. Information on the sample film setting is used at the time of calculations in the
standard sample SQX analysis.
1. When binder is used, click <Used> for “Binder/Grinding aids” and select the
compound of the binder. Enter the weight values of a sample and binder.
2. When a sample film is used, set information on it. Information on the settings of
“Binder/Grinding aids” and “Sample film” is used at the time of calculations in the
standard sample SQX analysis.
1. Select a flux type in the “Sample and flux” frame, and set flux weight and sample
weight to be used as standards.
2. When it is necessary to set other items such as the loss on ignition and oxidizer, set
them. Information on the above setting items is used at the time of calculations in
the standard sample SQX analysis.
When the effect of the loss on ignition exists, select <Corrects for LOI in sample> in the
“Loss on ignition (LOI)” frame. It is possible to correct the effect of the loss on ignition.
In organic matter such as a polymer, the intensities of characteristic X-rays with high
energy may be affected by a sample thickness. By setting a sample thickness, it becomes
possible to obtain the SQX analysis values of standard samples after the effect of a
thickness is corrected.
Use the same sample preparation process for both standard samples and unknown
samples. It is important that they have the same sample thickness.
Set analysis values obtained using another analysis method as standard values for the
X-ray fluorescence analysis. Set the data of one sample in one line. When a standard
value is not available, leave its cell blank.
1. Double-click the “Name” cell and enter a standard sample name.
2. If necessary, enter a comment in the “Note” cell.
3. Double-click the cell of each component and enter its standard value.
4. Click Next .
The automatic setup function may judge that components other than quantification
components affect the quantification components and automatically set the other
components and standard intensities additionally. The values of the standard sample SQX
are used for the standard values of added components.
Select <Set the analytical conditions in automatic> on the “Create a New Application”
screen, and Run Standards SQX will be displayed additionally on the flow bar.
The SQX analysis of the standard samples set using “Standard Samples” is made.
2. On the “Run Standards SQX” screen, select the names of standard samples to be
measured.
3. Set the position of the first standard sample.
4. Click OK .
When all the samples have been measured, the “Composition Information” screen
automatically appears.
Since the qualitative analysis of all elements is made, the measuring time for one sample is
12 to 15 minutes or so. As the number of quantification components becomes larger, the
measuring time becomes longer.
Based on the standard values and SQX analysis result values of standard samples,
components and their typical values are automatically set. These typical values are used
at the time of the theoretical matrix correction calculation.
Example of Composition Information When Analysis Sample Is Stainless Steel and Only
Cr Has Been Set As Quantification Component
Mn, Fe, Nb and Mo are automatically added as components to be measured
because they are regarded as components that affect the quantification component
Cr.
Components to Be
Measured (Automatically
Added)
1. Click the <Analysis Information> tab and check the automatically set element lines.
Suitable element lines are automatically set based on qualitative analysis results and SQX
analysis results of standard samples.
2. Click the <Measuring Condition> tab. Optical conditions have automatically been
set for each element line. Check them.
Components to Be
Measured (Automatically
Added)
Peak and background angles of element lines of the library are also automatically set.
3. Select an element line, click Property and check the detailed contents of the settings
of each element line.
Items displayed using dimmed characters, such as the excitation conditions, optical
conditions and PHA conditions, have automatically been set and cannot be changed. The
peak and background angles have automatically been set as well and cannot be changed.
Only the peak and background measuring times can be changed. Set them while
considering required precision.
“Optimize MC’s” is a program to set various measuring conditions based on the 2-theta
scan data and PHA data of each element line.
In “Run Standards SQX”, PHA data for each element line candidate is obtained besides
qualitative analysis spectra, and a spectrum and PHA data extracted for each element
line to determine measuring conditions are automatically stored. The data of two or
more line types may be stored depending on an element. Stored line types are K (Ka) line
and L (La, Lb1) line or L (La, Lb1) line and M (Ma) line included in the measuring range
of the 2-theta scan.
If necessary, it is possible to check peak and background measurement angles,
interfering lines and PHA setting conditions.
The 2-theta scan data and PHA data of samples with the maximum and minimum
concentrations have been stored for each quantification component.
Overwritten
Figure
Enlarged
Figure
It is impossible to change a peak or background angle. When you want to change an angle,
turn off the automatic setup mode. For the procedure to turn off the automatic setup mode,
refer to <4.6.2 CANCELLATION OF AUTOMATIC QUANTITATIVE APPLICATION
SETUP>.
When you want to change a PHA condition, turn off the automatic setup mode. For the
procedure to turn off the automatic setup mode, refer to <4.6.2 CANCELLATION OF
AUTOMATIC QUANTITATIVE APPLICATION SETUP>.
Check the calculated tentative calibration curve. Since the linear calibration curve
model is used for the theoretical matrix correction coefficient calculation, only a linear
expression can be used for a regression formula.
1. Click Result on “Empirical Calibration Result”. The result of a calibration curve
calculation is displayed. It is possible to check calibration curve coefficients, accuracy,
correlation factor and matrix overlap corrections.
When at least one coexisting element affects a quantification component greatly, the matrix
correction is made for the quantification component using set measurement components.
(The base component or analysis component may not be used for the correction depending
on a matrix correction model.)
When it is judged that overlap affects the element line of a quantification component, the
overlap correction using concentrations is automatically set.
It is possible to check tentative intensities used for a tentative calibration curve. “Run
Standards SQX” data is used for these tentative intensities.
1. Click the Recalc. button on the regression calculation result graph.
2. The “Recalculation of Empirical Calibration” screen is displayed. The “#” marks in
the X-ray intensity fields mean tentative intensities.
Described below is the procedure to change an element line selected by the automatic
setup. Element lines that can be changed are only those of optical conditions displayed
in <g. Measuring Condition Setup (Optimize MC’s)>. When an element line is changed
and there are two or more optical conditions for it, the optimum optical condition is
automatically selected. After the automatic selection, the intensities of the element line
are reextracted from the qualitative analysis data of standard samples, and various
corrections are automatically registered.
The procedure to change the element line of Pb from La to Lb1 is described below as
an example:
3. Change the line type from La to Lb1 at <Element line>, and click OK . X-ray
intensities and various corrections are automatically set for the specified line type
from the qualitative analysis data of standard samples, and a tentative calibration
curve can be checked.
As for the operational procedure after this, carry out the measurement of standard
samples, regression calculation and setup of analysis control information in that order
using set measuring conditions. For these items, refer to <4.2 f. Measurement of
Standard Samples> in <CHAPTER 4 QUANTITATIVE ANALYSIS> and sections after
that.
Select a quantitative application file for which you want to cancel the automatic
setup.
1. Click Quant Application , select <Modify an existing application> and click Next > .
2. The “Modify an Existing Application” screen appears. Select a relevant application
and click Finish .
3. Click Application File on the flow bar, select <Change analysis type /sample model>
and click Next > .
4. Make sure that <Change the analytical condition setting method> has been selected,
enter a new application name for “Application name”, specify a folder for saving and
click Finish .
If two or more measuring positions are set at the same coordinates, it will measure at the
same one by that number of positions.
Measuring coordinates on a sample are so defined that the angle of the markers of a
sample holder for mapping is 180 deg., as shown below:
Optical marker
.
. X
.
for mapping and optical marker
(For ZSX Primus IVi (For ZSX Primus
only) IVi only)
The optical marker is used to determine the direction of a holder from an image obtained
with the camera in the sample observation position in the spectrometer. The X-ray
marker is used to determine the direction of a holder in the measuring position by
measuring the neighborhood of the marker.
The center of a sample is the center of coordinates.
In the case of the r-theta coordinates, the start point of theta is the opposite side of the
markers and theta increases in the counterclockwise direction.
By designating measuring positions with the r-theta method, A mapping table is created.
The method to designate measuring points with crossing points and the method to add
measuring points on the graphic screen can be used. In both methods, when a measuring
point is designated on the graphic screen, its coordinates are displayed in the list in
order of designation.
Change Coordinates
After designating measuring positions using the r-theta coordinates or the X-Y
coordinates, the coordinates can be transformed to the other one. In the measuring point
coordinates list, the displayed measuring point coordinates are converted.
When a measuring point is designated on the graphic screen, the coordinates of the
measuring point are displayed in the list. The value in a cell of r(mm), theta(deg), X(mm)
or Y(mm) can be transformed. The bottom line of the measuring point list is blank. By
entering a value in that line, the measuring point can be added.
1. Double-click an objective “r(mm)” or “X(mm)” cell in the measuring point list.
2. Enter an r or X value.
3. Double-click an objective “theta(deg)” or “Y(mm)” cell in the measuring point list.
4. Enter a theta or Y value.
When one of coordinate value is entered in the bottom line of the measuring point list, the
value of the previous line is copied for the other of coordinate value.
When the Enter key alone is pressed in the bottom line of the measuring point list with both
cells left blank, the input cursor moves to the top cell.
All measuring points or designated measuring points can be deleted. In the case of the
deletion of designated measuring points, the measuring point designated last can be
pasted. In other words, it means temporary cutting. In the case of the deletion of all
measuring points, pasting is impossible.
All measuring points are deleted from the coordinate list and the graphic screen.
Or you can designate all measuring points and then delete them.
1. Select [Edit] [Select All].
2. Select [Edit] [Cut].
Measuring points can be deleted from the measuring point coordinates list.
Or you can designate a range of coordinates lines and then delete it.
1. Select the first coordinates line of measuring points to be deleted in the measuring
point list.
2. Select the last coordinates line of measuring points to be deleted with the Shift key
pressed down.
3. Select [Edit] [Cut].
When you select one measuring point, you need not press the Ctrl or Shift key.
To deselect measuring points, select an arbitrary coordinates line with the Ctrl and Shift key
released.
Two or more measuring points or one measuring point can be inserted. To insert one
measuring point, the previous coordinates are inserted preliminarily in the insertion
position, so change those values. To insert two or more points, measuring points must be
kept in a memory using the deletion or copy operation. The copy operation is as follows.
To select two or more points, select [Edit] [Select All], make arbitrary selection using the
Ctrl key, or select a range using the Shift key. See the paragraph “Deletion of Measuring
Point”.
1. Select the coordinate line for insertion in the measuring point list.
2. Select [Edit] [Insert].
3. Double-click an “r(mm)” or “X(mm)” cell in the line for insertion.
4. Enter the r or X value.
5. Double-click a “theta(deg)” or “Y(mm)” cell in the line for insertion.
6. Enter the theta or Y value.
This operation is possible when two or more deleted or copied measuring points have
been kept in a memory.
1. Select the coordinate line for insertion in the measuring point list.
2. Select [Edit] [Paste].
The diaphragm determines the effective area for a measurement. You can see the
effective measurement area by a circle with the diameter of the diaphragm drawn
around the measuring point.
This program guides you using the flow bar in a series of operations from loading a
sample to unloading it. Carry out an analysis according to the procedure shown below.
Sample Loading
No
Magnified picture?
Yes
Take magnified picture
Select application
Analysis
Yes
Repeat analysis?
No
Sample Unloading
Be sure to set the measuring diameter in the analysis parameters. Select <No> for the
sample spin.
a. Sample Loading
In the micro mapping analysis, a sample holder dedicated to mapping analysis is used.
Prepare a sample that can be set in that holder. When a sample is very small, take
measures such as embedding it in resin.
When a holder for mapping is loaded, it is moved to the photographing position and an
image of a sample is obtained automatically using the camera. The sample holder is then
moved to the measuring position and, after detecting the direction of the sample holder
using its X-ray marker, a specified position is measured.
A sample table can be displayed by clicking Select on the “Sample Load” window. Thus,
you can set a sample position visually using the sample table display.
In the “Loading sample” window, the image by the camera is displayed from loading a
sample till placing it in the photographing position of the camera.
When a sample has been loaded and images have been taken, the whole image is
displayed. The whole image is displayed on the left, and a position or area is specified on
the right. A magnified image can be taken to observe some area in detail. In that case,
the whole image is displayed on the left and the magnified image on the right.
Whole Image
Magnified Image
When a position or area is specified on a magnified image, the whole image is not saved.
When the whole image is necessary, make a screen copy and save it in a bit map file or the
like.
In the case of a permeable sample, its image may be dark. Make an adjustment by
inserting aluminum foil or the like beneath the sample.
Magnified image?
No
Yes
Display magnified image
Select the point designation, line designation or plane designation with a button on the
whole image display or magnified image display. Then specify a measuring position or area
using the mouse.
A magnified image is obtained by expanding a part of the whole image. Therefore, the
definition of the image decreases a little.
To change the setting, click before clicking OK and then change the setting, or click
All Clear and then make setting again.
Applicable only to the case of full screen mapping used with full screen.
When is clicked, it toggles to display the grid on the right side of the image.
Change Color
It may be difficult to recognize each part of a sample image depending on the light and
shade of the sample. In this case, it may become easier to recognize when the brightness
of the sample image is adjusted.
The brightness of a sample image is set with the sample image displayed on the
Mapping Position Setting screen. The procedure after the Mapping Position Setting
screen is displayed is described below:
2. Enter a number for the item Brightness. A larger number (up to 255) makes an
image brighter.
3. Click OK .
4. Click Sample Load in the flow bar again, check the position of the loaded sample and
click OK . A sample image is obtained again with the set brightness and redisplayed
on the Mapping Position Setting screen.
To carry out measurement for the same measuring position or area using another
application, click Setting Sample ID on the flow bar again without unloading the sample
after an analysis using the first application is complete.
In the case of the line analysis and the plane analysis, set a value to partition the
specified area. For example, when the number of partitions is 5 in the line analysis, the
number of measuring points is 6. Likewise, when the numbers of partitions for X and Y
are 3 and 4 respectively in the plane analysis, the number of data is 20 (=(3+1)(4+1)).
In the case of the point analysis, only one sample ID line is displayed. In the case of the
line analysis and the plane analysis, up to 10 sample ID lines are displayed according to
the specified number of areas. In that case, the number of partitions and an application
cannot be set individually for each sample ID line.
In the case of the point analysis, up to 999 points can be specified. In the case of the line
analysis and the plane analysis, up to 10 areas (999 points for each area) can be specified.
The procedure to set a sample ID for the qualitative analysis is described below as an
example. For the sample ID, see <2.3 a. Before Analysis> in the instruction manual. See
also procedures in <2.3 c. SQX Analysis> and <2.3 d. Quantitative Analysis> for your
information.
In the window of sample ID setting, sample unloading after completion of measurement can
be preset.
Unload Sample
1. Click Mapping Position Setting or Sample ID Setting on the flow bar.
Start
No 2-dimensional display
Bird’s-eye view?
Yes
No
Bar graph (3-D) display Bird’s-eye view (3-D) display
When point or line designation is done in Micro mapping, only bar graph can be displayed.
A mapping file can be created automatically by specifying the creation of a mapping file
in the sample ID setting of the Analysis or Micro Mapping program. It can also be
created from an analyzed result using the Result Display program.
1. Click Browse... for the result output on the Sample ID Setting screen.
2. Click a root folder in the “Select Folder” window.
3. Click Create folder… .
4. Enter a file name to create.
5. Click OK .
6. Click OK .
A mapping file is in a subfolder of the common folder (AnlRslt) or the user folder (AnlRsltU).
The following two files are created for each mapping data item:
Mapping Data File MPDT_(Specified File Name).ho
Image File MPDT_(Specified File Name).bmp
When an image file does not exist, only a mapping data file is created.
a. Display Method
The 2-dimensional display and the 3-dimensional display are available in the mapping
data display. In the case of the 3-dimesnional display, the bird’s-eye view and the bar
graph are available. In the case of the 2-dimensional display, analyzed results or X-ray
intensities are displayed. In the case of the bird’s-eye view, the Z-axis is added to display
ups and downs. Even areas that were not actually measured are displayed in a graph
through data processing. On the other hand, in the case of the bar graph display, it
displays the analyzed results or X-ray intensities only for area that was measured.
In the 3-dimensional display, the bird’s-eye view and the bar graph are available.
3-Dimentional Mapping
Bar Graph
Displayed data can be switched to analyzed results such as concentrations and film
thickness or to X-ray intensities.
1. Click the unselected item from <Analyzed result> and <Intensity> in the “Displaying
data” frame.
2-Dimentional Mapping
3-Dimentional Mapping
Switch Scale
The scale for displayed data can be switched to the linear display or to the logarithmic
display.
1. Click the unselected item of <Linear> and <Log> in the “Scale” frame.
Switch Component
When two or more components were measured, the displayed component can be
changed.
1. Select a component name at <Data name> in the “Basic setting” frame.
When mapping data has more than one measured component, up to nine components
can be displayed simultaneously. But this function can be used only in the case of the
2-dimensional display.
When more than one data are displayed, displaying methods can be switched between
using common scale for all data and using different scale for each data.
When a check mark is displayed at the left in the menu of [2-Dimensional Common
Scale], common scale is used. Common scale is useful to compare the amount among
elements. Different scaling among elements is useful to see the distribution of each
element.
Scale change can be done by changing the values in <Minimum> and <Maximum>.
When different scaling among elements is selected, the values are effective to the
element selected in <Data name> of <Basic setting>.
When a sample was measured in Micro Mapping program and sample image is
displayed in mapping window, measured area can be displayed on the sample image.
Displaying measured area for each measuring position designation method is shown
below.
When a check mark is displayed at the left in the menu of [Measured Area], the
measured area is displayed.
When measuring positions are specified by point designation, the number of order in
measurements can be displayed on the sample image.
To display or erase the measuring numbers , operate as follows.
When a check mark is displayed at the left in the menu of [Measuring Number], the
measuring numbers are displayed.
100%
Maximum 200%
Maximum 50%
100%
Steps :4
Minimum : 0%
Maximum : 50%
Steps :4
Minimum : 0%
Maximum : 100%
Shades of a single color or strata of two or more colors can be used as display colors.
Even when more than one data item is displayed simultaneously, the same colors are
used for all graphs. In the case of the 3-dimensional display, the single color display
alone can be used.
1. Select a color from the seven colors at <Color:> in the “Basic setting” frame.
For example, when the scale of displayed data is divided into 10 parts, an arbitrary
display color can be set for each step by adjusting sliders for red, blue and green. This
setting is always used when the 2-dimensional display is selected, not for each mapping
data item.
The color scale under the 2-dimensional mapping image is displayed when only one data
is selected as shown above.
When more than one image are displayed, sectional views of all the images can be
displayed by dragging across the sample image.
To create an image file, select a menu of [Transfer Image File] or specify a image by
using the mouse.
Specify an image when the image is displayed on the screen by the next operation.
In the field of <file name>, a default file name is displayed by the name combining mapping
file name, type of image and component name (line name) to be able to distinguished.
1. Move the mouse pointer on a position of an image to create an image file.
2. Click the right button of the mouse.
3. Click the [Transfer Image File] of the pop-up menu.
4. Set the dialog of [Transfer Image File] and click OK .
In the dialog of [Transfer Image File], the image specified by the mouse is selected when
the dialog appeared. The operation in the dialog is the same as that in [Create an image file
from menu] described above.
1. Select [Mapping Display] [Measuring Points] to display measuring points on the
sample image.
2. Click one measuring point on the sample image or in the measurement data area in
the lower part of the screen.
The selected measuring point is displayed in yellow on the sample image or in the
selection color in the measurement data area. When the mapping data has been
displayed using a bar graph, the frame of that measuring point is highlighted.
Up to 12 measuring points can be selected by holding down the Ctrl key while selecting.
The same operation is possible also from the measurement data of the analyzed result
grid.
It is possible to export various items of mapping data, such as images, graphs, analysis
values and X-ray intensities, to individual sheets in an Excel file. It is also possible to
export analysis values and X-ray intensities to a CSV file.
No sheet is created when there is no data item to be exported. For example, for the result of
a measurement using a mapping table from the Analysis screen, no image sheet is created
because there is no image.
Measuring coordinates are always output in the X-Y coordinates as in the case of
displaying an image. When an expanded image was used in the micro mapping, the center
of the image area becomes the origin.
When a mapping file is deleted, both a selected mapping data file and its image file are
deleted.
f. Print
Programs controlling data necessary for analysis and measurement using this system
are collected in the Utility menu. This menu consists of Atomic Symbols, Compound
Table, Sample ID Tables, Mapping Tables (optional), Unit Setting, Universal Standard
Samples, Sensitivity Library (optional), Material Standard Tables (optional), Program
Operation (optional and partly standard) and System Management. Descriptions are
given in “3. QUALITATIVE ANALYSIS” for Sensitivity Library. All other items on the
Utility menu are described in this chapter.
The operational procedure for the sensitivity library is explained in “3.4 SENSITIVITY
LIBRARY”. In this chapter, “e. Material Standard Tables” follows “d. Universal Standard
Samples”.
a. Atomic Symbols
This program provides various kinds of element information for the X-ray fluorescence
analysis. You can see general information such as atomic numbers, element names,
atomic weights and densities, and individual analysis information such as main
measured element lines, their wavelengths, absorption edge energy values, crystals
used and 2-theta angles.
You can see the elemental information and optical data for the lines normally used as
measured element lines for the wavelength dispersive X-ray fluorescence spectrometer.
For the elements from calcium (Ca) through uranium (U), one of the two lines such as
the K line and the L line can be selected as the measured element line and information of
each of them is displayed.
Item in Element
Description
Information
KA and LA represent K and L respectively.
Line name
For example, KA of aluminum represents Al-K.
Wavelength Displays the wavelength of the measured element line.
Edge energy When a tube voltage higher than this value is applied to
(Absorption edge energy) the target, fluorescent X-rays can be generated.
Single crystal to analyze that measured element line.
Crystal
The RX series represents the artificial multi-layers.
Displays the detector angle for each crystal.
2-theta This corresponds to the abscissa of a qualitative
analysis chart.
b. Compound Table
Presettable Items
Term Description
Any name can be set as a compound name. Uppercase
Compound
characters and lowercase characters are distinguished. In the
Name
case of a mixed compound, a MIX mark is attached.
This item must be set to register a compound for measurement
Analyte
for the quantitative calculation or the like.
This item must be set in the case of a component for which to
analyze a thickness using the quantitative film FP method.
Density
When a calculation is to be made as weight, this item need not
be set.
The type is used for the sample preparation information and
the setting for the SQX calculation. Compounds for which
“Flux” is set are listed when setting flux.
Types that can be selected are as follows:
Normal
Flux
Type Binder
Oxidizer: Presettable only for special compound
Non-wetting
Diluent
LOI (Flux)
Compounds other than “Normal” and “Binder” are used in the
fusion bead method.
Standard oxides, flux, etc. that have been registered beforehand cannot be changed
because they are already used for library samples etc.
Set “LOI (Flux)” as a loss on ignition component when the loss on ignition (LOI) correction
is to be made in the evaporation correction for flux in an analysis using the fusion bead
method. Set the same constituting elements as those in flux to be used.
For example, the procedure to set SiO2 under a compound name “TEST” at compound
number 33 is described below. Si is set for the analyte and 2.650 for the density, but the
compound is not registered here.
1. Click [Special compound] from the list in the left part of the screen.
2. Double-click the compound name cell at number 33.
3. Enter “TEST”.
4. Set “Si” in the “Analyte” cell.
5. Set a density in the “Density” cell.
6. Select <Normal> in the “Type” cell.
7. Set “Si” in the “Symbol” cell of “Element1”.
8. Enter the number of atoms “1” in the “Number” cell of “Element1”.
9. Set “O” in the “Symbol” cell of “Element2”.
10.Enter the number of atoms “2” in the “Number” cell of “Element2”.
To register the compound, select [File] [Save] after the procedure described above.
A registered compound can be checked. In addition to the preset number of atoms of each
element, density, etc., the calculated concentration ratio and molecular weight are
displayed.
You can click any cell in a compound line. Because double-clicking in a cell is used to
change the preset contents, use single-clicking to select a compound line.
Insert Element
Delete Element
1. Click [Special compound] on the list in the left part of the screen.
2. Click an arbitrary cell in the compound number 34 line.
3. Select [Edit] [Mixed Compound...].
4. Enter “TEST2” for “Compound name”.
5. Enter “34” for “Special Comp. No. ”.
6. Set “Si” for “Analyte”.
7. Select <Normal> for “Type”.
8. Enter “SiO2” for “Compound 1”.
9. Enter “1” for “Weight ratio1”.
10.Enter “Al2O3” for “Compound 2”.
11.Enter “2” for “Weight ratio 2”.
12.Click OK .
A mixed compound is used when mixed flux of Li2B4O7 and LiBO2 is used or in a similar
case.
Each mixed compound will have a “MIX” mark attached to the side of its compound name.
1. Select and right-click a compound registered as oxidizer for a special compound.
Execute “Residual Compound” on the menu that appears.
2. Set a residual compound and a residual ratio. A calculated value can also be used for
the residual ratio.
For the residual compound, set a compound that is left in a fusion bead when oxidizer is
specified for the sample type of a fusion bead in a quantitative application. For the residual
ratio, set the ratio of the weight of a residual compound in a fusion bead to original oxidizer
weight. Use a residual compound for the quantification calculation component, and use the
ratio of residual compound weight to a sample weight for the addition ratio.
If no residual compound is left after a fusion bead is created, select [None] for “Residual
compound”.
1. Select and right-click [Compound list] on the list in the left part of the “Compound
Table” screen. Execute “New Table” on the menu that appears.
2. Set a component type, list name and memo. The specified component type becomes
the default setting.
1. Select and right-click a compound list to be deleted on the list in the left part of the
“Compound Table” screen. Execute “Delete” on the menu that appears.
1. Select and right-click a compound list to be copied on the list in the left part of the
“Compound Table” screen. Execute “Copy” on the menu that appears.
2. Select and right-click the place of a compound list on the list in the left part of the
“Compound Table” screen. Execute “Paste” on the menu that appears.
3. Set a list name and a memo.
1. Select [External parameter] on the list in the left part of the “Compound Table”
screen. Set a parameter name and a type on the table.
Term Description
Normal parameter
Sample weight
Flux weight
Bead weight
Oxidizer weight
Type Loss on ignition
The items from the sample weight to the oxidizer weight are
used for a correction by setting each weight value at the time of
an analysis and calculating a dilution ratio.
The loss on ignition is used to convert analyzed values to ones
before ignition using the concentration of an ignition loss
entered at the time of an analysis when analyzing a powder
sample after ignition.
Printing
To print the contents of the compound table, compound name list printing, compound
structure printing including calculated concentration ratios etc. and parameter name
list printing for external parameters can be used. The compound structure printing has
two modes: the printing of the whole table and the printing of selected compounds.
1. Select [Standard oxide], [Special compound] or [External parameter] from the list in
the left part of the screen.
2. Select [File] [Print].
3. Select <Compound name list>.
4. Click OK .
1. Select [Standard oxide] or [Special compound] from the list in the left part of the
screen.
2. Select [File] [Print].
3. Select <Compound structure>.
4. Click OK .
5. Select <Compound Selection>.
6. Select compound names.
7. Click OK .
1. Select [External parameter] from the list in the left part of the screen.
2. Select [File] [Print].
3. Select <Parameter name list>.
4. Click OK .
1. Select [Standard oxide] or [Special compound] from the list in the left part of the
screen.
2. Click the compound name cell of a compound to print detailed information.
3. Right-click the mouse.
4. Select <Information Print>.
c. Sample ID Table
To measure samples with the Analysis program, you must set ID lines. When you want
to measure two or more samples, you must set an ID line for each sample. For routine
analysis that will analyze a series of samples many times a week, ID lines can be
registered in a table and they can be loaded when necessary for an analysis. This process
creates sample ID tables.
By loading a sample ID table for an analysis, ID setting becomes easy. For example, it is
recommended that routine analyses such as PHA adjustment and check analysis be
preset in tables.
Sample ID tables must be registered for cycle repeat analysis and for program operation.
The measurement sequence carrying out auto drift correction is also set up as sample ID
table for auto drift correction. (Dedicated for ZSX Primus IV/IVi)
The operational procedure is basically the same as that for the preset table in the
Analysis program. EZ Scan and cycle repeat analysis cannot be selected when creating
or adding an ID line. For example, the procedure to create a sample ID table for a PHA
adjustment for the SC and PC is described below:
Now the sample ID table for a PHA adjustment for the SC and PC has been created.
Subsequent PHA adjustment can be performed easily by loading that sample ID table.
There is not setting for sample position since PHA adjustment sample is built in the
equipment. In regular analysis such as quantitative analysis, there is setting for sample
position.
Unlike the preset table in the Analysis program, some restrictions are imposed on a
sample ID table. One concerns EZ Scan and another the secure operation in the control
IDs. EZ Scan need not be registered in a sample ID table and therefore is excluded. Since
a sample ID table may be loaded by the program operation, the secure and power on/off
functions cannot be set in control IDs.
Dedicated sample ID table for auto drift correction is made. The order of the sample ID
setting is check samples and Drift Correction samples.
6. In case of single position of drift correction sample in the sample ID table setting
screen, select <Last sample> in the update selection. In case of two, select < First
sample> and <Last sample>, and in case of three or more, set up <First sample>
and <Intermediate>, and <Last sample> for the last sample.
7. Select [File] and [Save] in turn.
It is necessary to set up preliminary check sample name for check sample setting of aimed
quantitative application, tolerance of low limit and high limit of measuring component, and
drift correction samples which is set the reference intensities.
The structure of the Universal Standard Samples is shown below. It has supplier folders
such as NIST (National Institute of Standards and Technology) and JSS (Japanese Steel
Standard), and each of them has sample type folders. The information on each standard
sample is registered in the appropriate sample type folder. A user’s standard sample is
registered in the “House standards” folder.
NIST
JSS
More…
Existing Data
Standard sample information such as NIST and JSS has been preregistered You can see
this information. You can also change its contents, but take care because the original
data cannot be restored if it is changed.
Your own samples can be treated as universal standard samples. By registering your
own standard sample, the same sample can be used easily for two or more applications.
Although “House standards” has been prepared as a supplier, you can create a new
supplier. Folders for various sample types will be created within those folders.
To add a new standard sample, follow the flowchart shown below. As described in
“Concept of Universal Standard Samples”, a sample type folder is created in a supplier
folder. Since all the standard sample data is registered in that sample type folder, a new
folder must be created if there is not an appropriate folder.
Start
End
When creating a new supplier folder, a new sample type folder is also created.
1. Select [File] [New] [Create Supplier...].
2. Enter a new folder name for “Supplier”.
3. If desired, enter a vendor name for “Vendor”.
4. Enter contact information for “Contact”.
5. Click OK .
6. Enter an arbitrary sample type name for “Sample type”.
7. If desired, enter a memorandum for “Information”.
8. Set all components constituting the standard sample in the “Components” frame.
9. Set the units of the preset components in the “Unit” cells in the “Components” frame.
10.Click OK .
To create a new sample type folder in an existing supplier folder, follow the procedure
described below:
1. Open an existing supplier folder.
2. Select [File] [New] [Create Sample Type...].
3. Enter an arbitrary sample type name for “Sample type”.
4. If desired, enter a memorandum for “Information”.
5. Set all components constituting the standard sample in the “Components” frame.
6. Set the units of the preset components in the “Unit” cells in the “Components” frame.
7. Click OK .
A standard sample is set in an existing or new folder. All standard values set in the
following procedure will be treated as certified values. To set them as uncertified values,
see “Change Standard Values to Uncertified Values”.
To “open data” hereafter means select [File] [Open] and select a supplier and a sample
type to load standard sample data to be changed.
A standard value that was set as a certified value can be changed to an uncertified value.
Only one balance component can be set for one sample. To cancel the balance setting and
set a standard value, first delete the balance component.
If there is a duplicate standard sample name, the sample type cannot be saved. Be sure to
change its name. However, the same name can be used for a different application.
Delete Folder
When a sample type folder is deleted, all standard samples in that folder are also deleted.
When a supplier folder is deleted, all sample type folders and standard samples in that
folder are also deleted. Take care because deleted folders cannot be restored.
To delete all sample type folders, select the check box <All>.
A selected supplier folder is deleted. But the <House standards> folder cannot be
deleted.
To print all standard samples belonging to a selected sample type, click All.
When analyzed values are calculated using the SQX analysis, material standards such
as ones for stainless steel and an aluminum alloy can be displayed (option). The
database of material standards used for that judgment is registered using the Material
Standard Tables program. Though a material standard is generally set with upper limits
and lower limits, setting with no upper limit or no lower limit is also possible. The sum of
the analyzed values of two elements can also be set as an upper or lower limit.
Although general material standards have already been set, a user’s standards can also
be set.
Existing Data
Material standards such as ASTM and JIS have been preregistered. You can see that
information. You can also change its contents, but take care because the original data
cannot be restored if it is changed.
By registering your own material standard, material judgments can be made after an
SQX calculation. A new standard folder is created and folders for various categories will
be created in that folder.
To add a new material standard, follow the flowchart shown below. A category folder is
created in a standard folder. Since each material standard data is registered in that
category folder, a new folder must be created if there is not an appropriate folder.
Start
End
A new standard folder is created and a new category folder is also created.
When there is not a lower limit, leave the lower limit cell blank.
When there is not an upper limit, leave the upper limit cell blank.
Both upper and lower limit cells can be left blank. When a value is entered in one of them,
“No limit” is automatically set to the other cell.
Registered upper and lower limit values can be set as the balance component. The
balance setting is made for one component.
Registered upper and lower limits can be deleted. The deletion operation is done for each
component.
If there is a duplicate material standard name, the standard cannot be saved. Be sure to
change its name.
For example, in the case of the nickel alloy, the total of the analyzed values of Ni and Co
may be set as its lower limit. In that case, enter the total value in the cells of both
elements and set the standard using the “Summation” function.
1. Enter total values in the upper and lower limit cells of both components to
use total values as their standard.
2. Select [View] [Category Information...].
3. Click Summation .
4. Select <Apply summation>.
5. Select components to be added.
6. Click OK .
7. Click OK again.
There is no mark to indicate the summation setting on the material standard table after the
summation setting. When judging analyzed results, the total value of the components set
for the summation is used.
Deletion of Folder
When a category folder is deleted, all material standards in that folder are also deleted.
When a standard folder is deleted, all category folders and material standards in that
folder are also deleted. Take care because deleted folders cannot be restored.
f. Unit Setting
A unit to be used for the quantitative analysis can be added. A unit added here can be
used for both an empirical method application and a limit range condition but can be
used only for a limit range condition in the case of an FP method application. When a
quantitative application and a unit are changed in a limit range condition, an analyzed
value is output after a conversion using a factor that is set in the unit.
Add Unit
Units that have been registered beforehand are displayed in gray in their display fields and
cannot be changed.
Units that can be used for calculations in the FP method and theoretical matrix correction
coefficient calculation are mass%, ppm, at% and mol% for concentrations, Angstrom, nm
and um for thicknesses and g/m2, mg/m2, mg/cm2 and ug/cm2 for weight thicknesses.
When another unit including an added unit is used, use any of the above units in a
quantitative application also in the empirical method, and set another unit in a limit range
condition.
This program performs the fixed time or periodic analysis and controls the automatic
startup and shutdown of the system. The following functions are available in Program
Operation:
Three functions except Tube Aging in Program Operation are optional. Only the Tube
Aging (standard) is displayed when the option is not attached.
The setting items for Auto Power Off are displayed as selection items in <Power off
method> on the “Control ID Setting” screen of the Analysis program and in <Method> for
the automatic power off on the “Spectrometer Shutdown” screen.
Fixed time analysis and periodic analysis may include <Auto start Spectrometer Startup>
and <Auto start analysist>. (This is the function for ZSX Primus IV/IVi only, and applicable
for the equipment shipped newly.)
In Time Preset Analysis, setting is made to automatically start analyses using a sample
ID table at a specified time or at specified intervals. That setting is valid when the EZ
Analysis or Analysis program on the main menu is activated. An analysis or
measurement need not be running at the time. However, when the fixed time or the end
of the specified period is approaching, the time preset analyses are performed
preferentially before other analyses that have been set. A fixed time or a period is set
and a sample ID table is selected here.
Setting of every day/every week/every month can be made as schedule function. (This is
the function for ZSX Primus IV/IVi only, and applicable for the equipment shipped
newly.)
Setting of sample ID table for auto drift correction (For ZSX Primus IV/IVi only)
Start time/schedule of auto drift correction is set using the sample ID table preliminary
made for the auto drift correction. (This is the function for ZSX Primus IV/IVi only, and
applicable for the equipment shipped newly.)
The measurement of the time interval starts when the EZ Analysis or Analysis program is
started on the main menu.
When turning on the X-rays after they have been turned off for a long time, the X-ray
tube voltage and current must be raised slowly to the specified load step by step. Raising
the tube voltage and current slowly is called aging. If the period when the X-rays have
been turned off is longer, a longer time should be taken to raise the tube voltage and
current. Only the target values for aging are set here and actual operation is started in
the Analysis program or on a system startup.
Some combinations of tube voltage and current values cannot be selected because of
restrictions on the X-ray tube output or filament current.
When the tube voltage is selected, only those tube current values that can be combined
with it are available for selection.
Two methods can be used for Auto Power Off. One is the method to keep the X-rays and
the system turned on and the other is the method to turn off the X-rays. When the
X-rays are to be turned off, selection can be made from the method to keep the system
turned on and the method to turn it off. The methods of turning off set here become the
selection items in the Analysis program or on a system shutdown.
The setting items for Auto Power Off are displayed as selection items in <Power off
method> on the “Control ID Setting” screen of the Analysis program and in <Method> for
the automatic power off on the “Spectrometer Shutdown” screen.
Setting is made to keep the system and the X-rays turned on with the X-ray output at
the lowest setting of 20kV-2mA. Because two or more control methods can be created,
they are given names to distinguish them and then saved.
The software can be set to turn off the X-rays and the system.
When the system is not used, the load of the X-rays can be automatically reduced to save
electric power. Following setting is available.
h. System Management
Items to be set in System Management are conditions that should be determined before
operating the ZSX and information to add or change according to the operating condition
of the ZSX. These include the classification of users, the function assignment to users
and the management method for analyzed results. This program has the following five
functions:
User Administration
System Parameters
Sample ID Structure
Data Transfer Setting
Record Voice
User Administration
User Administration sets users who operate the instrument and places them in groups
in order to assign functions and set locations to which files are saved.
The following are registered at the time of shipment as the initial conditions. Only users
who belong to the Administrators group can utilize these functions.
The users of this software are unrelated to those of Windows. Register users using this
software in addition to those of Windows.
User Administration sets access rights concerning folders, files and functions of the data
processing system.
The administrator(s) of a group basically impose(s) limitations for each group on
functions of the data processing system. A user can belong to only one group.
Access rights are set for an individual folder for each group, and it is possible to give
authority for updating files of analysis conditions and analyzed results to other groups.
Add Group
It is called the addition of a group to add a group of access right settings concerning
folders, files and functions of the data processing system.
6. For a user of the group to be set, choose whether each item is displayed or not on the
main menu. Only the chosen menu items are displayed. On an example of the main
menu screen shown below, only four menus items have been chosen:
7. Select analysis types that can be set in the Analysis program.
The selection of analysis types is applied to both <EZ Analysis> and <Analysis> programs.
8. Set data change authority for the six items including analysis conditions.
When the instrument does not have the mapping option, the check box [Change authority
of Mapping Tables] is not displayed.
When the instrument does not have the Program Operation option, the check box [Change
authority of Program Operation] is not displayed.
When the usage of individual folders for each user is set for analyzed results and analysis
conditions, subfolders that have the set user name are automatically created directly under
\AnlRsltU for analyzed results and \AnlCondU for analysis conditions.
When individual folders for each user are not used for analyzed results or analysis
conditions, the common folder is used.
14.Click OK .
Add User
It is called the addition of a user to add a user to a group that has access right settings
concerning folders, files and functions of the data processing system. It is possible to set
two or more users in one group.
Because even yourself cannot read the registered password, be sure to write it down.
Analysis result files are used for quantitative analysis results, qualitative analysis results
and mapping files.
The set subfolders described above become default folders. It is also possible to create a
subfolder for analysis conditions or analyzed results when an application file is created or
when a sample ID is created at the time of an analysis, respectively.
Delete User
Delete Group
When at least one user who belongs to a group exists, it is impossible to delete that group.
System Parameters
The system parameters are conditions and values to control data concerning the
analysis system. The parameters are divided into data processing, printing,
measurement setting, transmission condition, EZ Scan, initialization and others. When
the system has an ASC, the tab for an automatic sample loader is added.
<No Limit> is set in the initial condition. To change setting of some period to <No Limit>,
right-click an objective period and select <No Limit>.
It is possible to output to printer as a PDF file using Microsoft Print to PDF to the printer.
Click Setting… on the Printer and set as follow.
1. Select <Use the following printer>, and select <Microsoft Print to PDF> from the
pull-down list.
2. Select <Print to file>.
3. Click Browse… for saving place of file to make folder.
4. Input “PDF” on the extension screen.
5. <Open the file after output> is normally not necessary to be selected.
Use [Microsoft Print to PDF] without fail. No guarantee of other pdf printers working.
It is possible to designate any tray to be able to set only helium atmosphere application
as a dedicated tray for helium atmosphere.Designated sample tray and sample position
are indicated with blue.
Evacuation ASC position setting in the event of a two-stage sample holders stack
error (For ZSX Primus IVi only)
Select the standard to be used for transmission from RS232C and TCP/IP.
The procedure to change the number of retry actions and waiting time for the ENQ
response will be used as an example.
Set EZ Scan
Sample types of metal and oxide can be selected in EZ scan. In the analysis of liquid, a
balance component should be specified. Accordingly, recalculate specifying the balance in
SQX calculation program after EZ scan analysis. Or create a qual application for the liquid
and analyze using the qual application.
When SQX software is not attached, there is no setting of <SQX result> and <Material
judgement> in “Print” frame and <Sample film absorption> in “Helium atmosphere
Correction”.
When the radio button <Initialize the spectrometer only if necessary> is selected, the
spectrometer will be initialized only if it has not been initialized when the ZSX Guidance
icon on the Desktop is double-clicked to start the ZSX Guidance program. Usually, <Do
not initialize the spectrometer> is not used.
Set Others
Uncheck the check box of <Show file name input field of the qual. Analysis result> if it is
not necessary to indicate it. Sample name becomes file name.
Sample ID Structure
The sample ID consists of the following items. The outline of each item is shown below:
Multiple Content
Option Meaning of Item
Settings Setting
Analysis type (quantitative analysis,
Transmit,
Type intensity measurement, check analysis, Impossible
text
drift correction, … , PHA adjustment)
Number to indicate a position on the turret
Position Impossible Transmit
or ASC
Application Analysis condition name Impossible Transmit
Mapping Mapping measurement type Impossible None
Times Number of repetitions Impossible None
Update selection (first, last,
Flag Impossible Text
intermediate sample)
Displays the contents of the memo in the
Memo Impossible None
application etc.
Indicates whether data has been input or
Input data Impossible Text
not. The value is not displayed.
File name Qualitative analysis file name Impossible None
Memo Memo entered on ID setting.
(See on Possible Transmit
the next page.)
Incremental setting is possible.
Selections Options are registered on a table and
(See on Possible Transmit
the next page.)
selected on ID setting.
When setting is made on the sample ID structure screen, the caution message shown
below appears. Convert measurement result data according to the message. For the
conversion, see “6.4 CONVERSION OF SAMPLE ID”.
The item at the left end of the sample ID is item number 1. Items after that follow to the right
in numerical order.
The maximum number of characters to be output as the sample name is 40. The
transmission sample ID is used only in the transmission format and there is no limitation on
its number of characters.
When <Output this item in sample name> is selected for <Memo> or <Selections>, set the
number of characters by actually printing data because the printing space is limited.
Delete ID Item
The option <Space> has the same function as <Cut> on the right-click menu.
To make or alter the settings for data transfer, you must have the administrative rights
for it. Preset contents are administered for each group. Obtain network information
beforehand from an administrator of the network.
Disk saving, FTP transfer and e-mail are available as the transfer method. Up to 20
destinations can be registered for one group.
Add Destination
Settings to send e-mail directly using the ZSX Guidance are made from the SMTP Info.
button on the <Destination> tab. See “6.5 SETTING FOR E-MAIL” for details.
When <Dialog> is selected, a dialog to determine whether to transfer data or not appears in
the sample ID setting on an analysis.
An output file name is created from a file name entered in setting an ID.
All output settings for e-mail on the panels are validated. To stop e-mail transfer,
deselect this.
The voice output setting for the error information and analysis progress is validated.
Record Voice
This function administers sound files used when an instrument error occurs during an
analysis and in similar cases. Sound output is determined according to a type of a
phenomenon (an analysis error, analysis progress, instrument error, etc.).
WAV files and files that multimedia supports can be handled.
When an existing file name is designated, the following procedure is added on the “Edit
Voice” screen:
Analyzed results can be transferred using the Result Display program after analyses.
If <Send e-mail> is not selected on the “Destination” tab on the “Data Transfer Setting”
screen, data will not be transmitted even when a destination is designated.
Errors such as hardware errors of the instrument and progress of analysis can be
transferred.
Setting for transfer is made using “Data Transfer Setting” in “System Management”.
Two or more destinations can be designated.
If <Send e-mail> is not selected on the “Destination” tab on the “Data Transfer Setting”
screen, data will not be transmitted even when a destination is designated.
To restart the system, select <Initialize spectrometer> and <Age X-ray tube> with the
“Spectrometer Startup” program.
Operation of <Turn off the X-ray>, <Set vacuum to Secure condition> and <Turn off
the power of spectrometer> in Auto Power Off
5 minutes later (when <Turn off the power of spectrometer> has been selected)
The X-ray generator and the heat exchanger are turned off.
The spectrometer is turned off.
The personal computer and the printer are not turned off.
To restart the system, terminate the ZSX Guidance, turn on the spectrometer and restart
the ZSX Guidance.
Energy-Saving Operation
The energy-saving operation sets the specified standby condition when no measurement
has been made during the wait time to switch to the energy-saving mode. The wait time
starts when an analysis is completed or suspended. This function is valid when <Use
Energy saving operation> has been selected on the “Program Operation” screen.
When time is up
The message “ZSX is shifting to standby mode” appears.
The system starts to switch to the standby mode.
When an analysis is started using the Analysis program in the standby mode, recovery is
made automatically.
The upper four items in the above table should be set according to the requirements of
the external communication device. For the lowest one, selection for the RS232C or
TCP/IP transmission should be made from <No>, <Yes> and <Dialog> for each analysis
type in the quantitative analysis.
<Drift corr. update> or <Bias corr. update> can be selected in the “Data selection” frame. By
using these options, update results such as coefficients can be transmitted.
When you have changed any setting for the transmission or the sample ID structure, you
must restart the ZSX Guidance to activate the changed contents.
When selection for the RS232C or TCP/IP transmission is made from <No>, <Yes> and
<Dialog> for each analysis type, options on the “Sample ID Setting” screen and
transmission processing just after an analysis are as follows:
Since qualitative analysis results must be transmitted after checking and editing them,
a function to transmit them just after an analysis is not available.
6.2 MAINTENANCE
Programs to manage the mechanical condition of this system are collected on the
Maintenance menu. This section describes Spectrometer Test, Maintenance Records
and Test Measurement. The Maintenance menu consists of the following:
Maintenance
Sub-Item Description
Item
Displays a list of items that an operator
Check List
should check daily.
Carries out a status check, tests of
Self-Diagnostics driving units, vacuum test,
reproducibility test, etc.
Carries out assistant driving to replace
Spectrometer
PC Maintenance the PC window and executes PC wire
Test
cleaning.
The analysis chamber is returned to the
air mode, and the lift of the sample
Sample Chamber
chamber is lowered, as automatic
Maintenance
preparation to pull out the sample
chamber.
Record of spectrometer errors and
Error Log
communication errors
Record of spectrometer status
information obtained using the
Spectrometer Check
spectrometer check in the
self-diagnostics
Record of results of various checks and
Self-Diagnostics
tests obtained using the self-diagnostics
Record of X-R values of check analysis
Maintenance Check Analysis
results
Records
Record of alpha and beta values of drift
Drift Correction
correction results
Record of variations of PHA adjustment
PHA Adjustment values and resolutions of PHA
adjustment results
Record of calibration curve constants
Calibration Curve that were changed using the regression
Constants calculation and automatic calibration
update function
Test Executes the fixed angle measurement,
Measurement 2-theta scan and PHA scan.
a. Check List
Check List displays everyday check items for the spectrometer so that the spectrometer
supervisor can inspect the contents of work on the screen. Check List includes periodical
check items such as ones shown below. For each of those items, you can see the contents
of work and carry out an inspection.
Especially when maintaining the spectrometer abroad, by using this function, a user can
see an inspection procedure at a glance and carry out the displayed procedure easily.
Inspection work is very important to maintain the reliability of the spectrometer for a
long period.
The contents of inspections by this function are recorded and displayed as marks on a
graph in Record for PHA Adjustment Result, Record for Drift Correction Result, and
Record for Check Analysis Result, etc.
b. Self-Diagnostics
Self-Diagnostics includes three functions of “Status”, “Driving time” and “Vacuum test”.
Status
Driving time
Vacuum test
Status
The spectrometer status is checked, and the spectrometer I/O status is received and
displayed. Those results are recorded and can be redisplayed.
The integrated time when the X-rays have been turned on is displayed at “X-ray tube hour
integrator”.
The spectrometer status and driving part status can be obtained even if a hardware error
has occurred.
Driving Time
This function obtains a driving time by operating a mechanical driving unit individually
by the specified number of repetitions. Units to be tested are the sample
loading/unloading mechanism, the optical system such as the filter and slit, and other
mechanisms. Continuous operation is not carried out.
Vacuum Test
A sample holder alone is loaded, and the attained vacuum degree and the stability of the
vacuum degree are displayed.
c. PC Maintenance
This program has functions concerning the maintenance of the PC, which is the detector
for light elements. Center wire cleaning and the replacement of the PC window are
carried out here.
Replace PC Window
A very thin macromolecule film with thin metal evaporated on it is used for the PC
window. This window is a consumable that is damaged when it has been used for a long
period of time. You can make a preparation to replace this window through dialogs on
the screen.
When replacing the PC window, it is recommended that P-10 gas be kept flowing. By
keeping P-10 gas flowing, the detector inner wall can be kept out of contact with air.
Therefore, after the window is replaced, a stable condition can be restored sooner.
For the procedure for replacement, see “Procedure to Replace PC Window” in the
maintenance manual.
d. Error Log
Errors that occur in the spectrometer and in communications with the spectrometer are
automatically recorded. Errors are classified into spectrometer errors, X-ray generator
errors, spectrometer response errors, X-ray generator response errors, data processing
errors, communication errors and other errors. They are classified into three levels: an
error, a warning and a failure. A warning means an error with which an analysis can
continue and changes to a failure when the analysis cannot be continued. Using the
Error Log program, errors that have been recorded automatically in a time sequence can
be displayed and printed in detail.
Up to 200 errors can be recorded in the error log. If the number of errors exceeds 200, the
oldest data will be overwritten.
All recorded errors are deleted. Take care because deleted data cannot be restored.
The contents of an error are displayed in simplified form on the “Error Log” screen.
Detailed information of an error (the contents and factors of an error, maintenance
information, etc.) can be displayed and printed.
1. Select an error item to display detailed contents on the error log list.
2. Click Information .
3. Click Print if desired.
4. Click Close .
Of the diagnostics and test items in Self Diagnostics, results of <Spectrometer Status
Check> can be displayed using “Spectrometer Status Check Record”. The records of the
other items can be displayed using this “Self-Diagnostics Record”. The relationship
between the self-diagnostics items and display functions is shown below:
Spectrometer
— List of check items
“Status” status
tab Driving part
Driving part status —
status
Self-diagno
The data displayed on the screen can be copied to clip board by pressing the Ctrl+C key. It
is easy to paste it to spreadsheet like Excel®.
Reference Data
When a diagnosis has been carried out, reference data for comparison is necessary to
judge a diagnosis result. When our service technician has made an adjustment, he or she
leaves reference data with '*' attached to the diagnosis date and time. Judge a diagnosis
result by comparison to this reference data.
Performance makes it easy to compare the profile of each measurement result with the
shipping data.
1. Select aimed date and time.
2. Right-click, then click spectrum . The designated date and related shipping date file as
reference are selected.
3. Click Spetrum to display the spectrum.
Because the reference data set cannot be deleted, it does not appear in the “Diagnostic
date” list.
The check analysis is generally made several times a day and its results are recorded.
Monitoring the trends of those averages (X) and ranges (R) is called X-R control. Check
analysis results are recorded automatically and you can carry out the X-R control easily
using this program.
The data displayed on the screen can be copied to clip board by pressing the Ctrl+C key. It
is easy to paste it to spreadsheet like Excel®.
The specified number of check analyses are made each day, their statistics are calculated
and an X-R control chart is drawn. For a selected component, graphs of averages and
ranges are displayed in the upper and lower parts of the screen. Statistics are displayed
superimposed on the graph. Data of one day is displayed with one point. The cursor line
in the list interlocks with the cursor bar in the graph. An analyzed result out of its limit
range is displayed in red in the list.
By dragging the dividing line, the sizes of the graph areas can be changed. To display two
graphs evenly, double-click the dividing line.
Method Description
All dates All data items of the designated sample type are displayed.
The start and end dates are designated. The standard
Date from
calendar function of Windows is used.
For the last x Data items of the last months from the present date are
months displayed.
For the last x Data items of the last days from the present date are
days displayed.
For reference grid lines for the ordinate axis, the limit range values preset for the check
analysis or the standard deviation can be used. Upper and lower limits are set for
average values and an upper limit is set for ranges. In the case of the standard deviation,
+/-1 to +/-3 can be set.
For reference grid lines for the abscissa axis, grid lines can be drawn at intervals of the
specified number of days.
In the initial condition, <Auto scaling> has been set for the vertical scales of both the X
graph and the R graph. Other options cannot be selected when <limit range> is selected
for the reference grid lines. If you want to select <Absolute> or <Relative> for the X
(average) graph or <Absolute> for the R (range) graph, change the setting for the
reference lines described before.
Reference
X Value Graph R Value Graph
Grid Lines
Standard
Auto scaling only Auto scaling only
Values
Standard
Auto scaling/Absolute/Relative Auto scaling/Absolute
Deviation
None Auto scaling/Absolute/Relative Auto scaling/Absolute
In the case of <Absolute>, specify an absolute value in the unit used, for example, in
mass%. In the case of <Relative>, specify a value relative to the average in percentages.
Data Information
The specified number of check analyses are generally made each day. Up to five data
items can be used for the average (X) calculation. On the “Data information” page, the
number of data items to be used for the average (X) calculation and the handling of data
out of its limit range are set.
When the specified number of data items is less than the number of actual analyses, the
older data items are used. If the number of analyses in one day is larger than the
specified number, excessive analyses (data items exceeding “Number of data for
averaging”) will not be used for a calculation. Even if there is an out-of-range data item
in the specified number of data items and <Not included> has been set for “Out of range”,
averages will be calculated using the specified number of data items.
Components used for the X-R control can be printed out. The procedure to print a
component of a check sample that is not displayed now is described below as an example:
Whole data of designated items can be output as Excel file in a lump by clicking [File]
[Output to Excel file].
To continue to use a calibration curve or a sensitivity calibration curve for a long period
of time, the same intensities as those at the time of the creation of the calibration curve
or sensitivity calibration curve are required. Because X-ray intensities change with time,
the drift correction is used to correct their changes. Because drift correction coefficients
and are recorded automatically, the history of their corrections can be checked.
The data displayed on the screen can be copied to clip board by pressing the Ctrl+C key. It
is easy to paste it to spreadsheet like Excel®.
The alpha method and the alpha-beta method can be used for the drift correction. One
drift correction sample is used in the alpha method, and two drift correction samples are
used in the alpha-beta method. In the case of the alpha method, a graph of beta values is
not displayed. The cursor line in the list correlates with the cursor bar in the graph. An
analyzed result out of its limit range is displayed in red in the list.
By dragging the dividing line, the sizes of the graph areas can be changed. To display two
graphs evenly, double-click the dividing line.
Method Description
All dates All data items of the designated sample type are displayed.
The start and end dates are designated. The standard
Date from
calendar function of Windows is used.
For the last x Data items of the last months from the present date are
months displayed.
For the last x Data items of the last days from the present date are
days displayed.
For the abscissa axis, reference grid lines can be drawn at intervals of the specified
number of days.
The vertical scale can be changed to the automatic scale, absolute value or relative
value.
Maintenance markers are used to indicate when the maintenance of the SC or PC was
done or a detection circuit board was replaced. Colored arrows and characters for
identification are displayed on a list and a graph.
Measured element lines used for the drift correction record can be printed out. The
procedure to print a measured element line of a drift correction sample that is not
displayed now is described below as an example:
Whole data of designated items can be output as Excel file in a lump by clicking [File]
[Output to Excel file ]
In the PHA adjustment, which is one of the everyday control analyses, the results are
recorded automatically for each execution. In this program, variations in the energy
resolutions of the detectors can be checked. Furthermore, in the case of the PC, you can
check dates and times when maintenance operations such as the replacement of the PC
window and the wire cleaning were carried out.
The data displayed on the screen can be copied to clip board by pressing the Ctrl+C key. It
is easy to paste it to spreadsheet like Excel®..
From PHA adjustment results of each detector, you can check PHA adjustment values
and resolution values, which indicate the performance of a detector. This system has the
SC and PC as the detectors, and data of each of them is displayed separately.
1. Click PC or SC .
The cursor line in the list correlates with the cursor bar in the graph.
By dragging the dividing line, the sizes of the graph areas can be changed. To display two
graphs evenly, double-click the dividing line.
Method Description
All dates All data items of the designated sample type are displayed.
The start and end dates are designated. The standard
Date from
calendar function of Windows is used.
For the last x Data items of the last months from the present date are
months displayed.
For the last x Data items of the last days from the present date are
days displayed.
For the abscissa axis, reference grid lines can be drawn at intervals of the specified
number of days.
The vertical scale can be changed to the automatic scaling, absolute value or relative
value.
Maintenance markers are used to indicate when the maintenance of the SC or PC was
done and a detection circuit board was replaced. Colored arrows and characters for
identification are displayed on a list and a graph.
A PHA curve, which is a result of the PHA scan, can be displayed from an adjustment
result of each detector.
Printing
Whole data of designated items can be output as Excel file in a lump by clicking [File]
[Output to Excel file]
The record for calibration curve constants that were changed using the regression
calculation and automatic calibration update function can be displayed.
The data displayed on the screen can be copied to clip board by pressing the Ctrl+C key. It
is easy to paste it to spreadsheet like Excel®.
Detailed Result
Although up to three calibration curves can be stored for one component using the
calibration division function, in the record for calibration curve constants only the first curve
is displayed. By displaying a detailed result using the “Detail display” button, the constants
and graph of each calibration curve can be displayed and printed.
The record for calibration curve constants does not have a function to delete any saved
record.
k. Test Measurement
By creating measuring conditions freely, you can execute the fixed angle measurement,
2-theta scan and PHA scan.
Mode Description
Measurement is made at a fixed angle to check X-ray
Fixed Angle
intensities. This mode is used to adjust and check the
Measurement
spectrometer.
A qualitative analysis chart is generated using the 2-theta scan
measurement. This mode is used to determine whether there
2-theta Scan
are interfering spectral lines and to check peak and
background angles.
Energy distribution of X-rays that enter a detector is measured
at a fixed angle. This mode is used to check overlap of
PHA Scan
higher-order lines, escape peaks, fluorescent X-rays from a
crystal and the resolution of a detector.
Measured element lines, measuring conditions, etc. are set from the common menu.
The measuring mode can be switched using the fixed angle measurement,
Start
Measurement
Calculation processing
End?
No
Yes
End
In the fixed angle measurement, measurement is continued till you click Stop. In the
2-theta scan and PHA scan, measurement automatically stops after measuring a specified
range.
When a measured element and an element line are selected, the default measuring
conditions are set automatically.
Measured element lines, measuring conditions, etc. can be set also using a pop-up menu
that appears when the mouse is right-clicked.
Measuring Conditions
The measuring conditions are classified into “Common conditions”, “Sample” and
conditions for each measuring mode. Those are described in the table below. For details,
see “4.3 QUANTITATIVE ANALYSIS TECHNIQUES”.
Division Description
Classified into the excitation condition, optical condition and PHA
Common
condition. The PHA condition cannot be set in the PHA scan mode.
The sample position in the sample changer is specified. In the case of a
Sample system equipped with the mapping function, the measuring position can
also be designated.
Conditions In the fixed angle measurement, the measuring time and full scale are set.
for Each In the 2-theta scan, the scan mode, start and end angles, etc. are set.
Mode In the PHA scan, the number of steps, start and end positions, etc. are set.
The tube voltage can be set at intervals of 1kV. The tube current can be set at intervals of
1mA.
Some combinations of tube voltage and current values cannot be selected because of
restrictions by the X-ray tube maximum load or filament current.
Because the 2-theta angle set here is only used to determine the full scale, it need not be
set.
The sealing can be set to <IN> only when the diaphragm is set to 30(28)mm and the
attenuator to <1/1>. It may be impossible to change the sealing, depending on the current
atmosphere.
In the case of the ZSX Primus, ZSX Primus II or ZSX Primus IV/IVi, <1/10> can be selected
for the attenuator only when the diaphragm is 30 mm.
In the case of the ZSX Primus III+, <IN> can be selected for the attenuator only when the
diaphragm is 30 or 20 mm.
In the case of the PHA scan, the PHA condition cannot be set.
Specify a sample placed in the sample changer. Because the specified sample is loaded
automatically but is not unloaded, do an unloading operation if necessary.
In the case of a system equipped with the mapping function, an arbitrary measuring
position on a sample can be specified.
Condition setting for each of the fixed angle measurement, 2-theta scan and PHA scan is
described below:
For the fixed angle measurement, the goniometer angle and the measuring time at that
angle are necessary. Because the 2-theta angle is set in the optical condition on the
“Common condition” tab, the measuring time and full scale are set here.
In the 2-theta scan mode, setting is made for the scan method, number of steps, scan
range, measuring time, number of repetitions, speed and smoothing. When the scan
method is the step scan, the measuring time is set. When the scan method is the
continuous scan, the speed is set.
Set PHA Scan Condition (ZSX Primus, ZSX Primus II, ZSX Primus III+)
In the PHA scan mode, setting is made for the scan range, measuring time and number
of steps.
Display Format
Three functions of the data selection, scale change and graph setting are available as
auxiliary display functions for measured results.
The procedure described here is to change the scale after measurement. To change the
scale during measurement, use <Fixed angle measurement condition>.
Change Colors
Calculation Items
The angular resolution can be calculated in the 2-theta scan mode, and the energy
resolution in the PHA scan mode.
The angular resolution can be calculated from a qualitative analysis chart obtained from
a 2-theta scan. The angular resolution is assessed with a half width (FWHM) and a 10%
width.
The energy resolution can be calculated from a differential curve obtained from a PHA
scan. The energy resolution is assessed for the half width (FWHM). The total intensity
in the specified range divided by the total intensity in the PHA scanning range is called
the transmittance.
Printing
After the above procedure, a check mark is added before [Intensity Print] to indicate
that data is printed during measurement.
By repeating the same procedure, the check mark is displayed or erased to indicate that
the condition is switched.
It is possible to select <Bitmap> or <Metafile> for output. When the output is selected, set
File name and File output folder.
Quant. Data
Besides quantitative analysis results, result data items
displayed using the Result Display program such as check
Analyzed analysis results, drift correction results and PHA
Results adjustment results are saved.
Mapping Data
Mapping data prepared using a mapping analysis, a micro
mapping analysis or the Result Display program is saved.
Saving is possible only when the system has the optional
sample observation mechanism.
When User Setting Data, Analysis Conditions (Qual. Condition and Quant. Condition) are
saved at a time, the date and time of saving are automatically recorded on the check item
list.
Three startup procedures are available for the “Data Save Load” program.
It is possible to start the program from the data saving notification bar or menu bar
when ZSX Guidance is in operation or from the shortcut in the program list that the
Windows OS manages.
Data Saving
Data Loading
The user without access authority for the “Data Save Load” program is a user who belongs
to a group for which the User Administration program has not set access authority for the
“Data Save Load” program.
For the details of the access authority for a program, refer to <User Administration - Add
Group> in <6.1 h. System Management>.
Data saving (backup of analysis conditions etc.) must be carried out periodically. This
system has a function to indicate on the main screen of this software that data saving
has not been carried out for a specific period and carry out data saving from the main
screen.
When data saving notification is displayed, carry out data saving according to the
following procedure:
For the procedure to change the time that has passed from the last data saving, see “6.1.
h. System Management”.
a. Data Saving
Described below is the procedure to carry out data saving when ZSX Guidance is in
operation:
When <Specify a folder to save> was selected, be sure to designate a drive name or a
folder name.
To designate a new directory as a device for saving, click Make New Folder and enter a
new directory name.
The displayed “Data size (Reference)” indicates the real size of the selected data. Thanks
to compression, the size of saved data is smaller than this value.
When <analysis condition to save> is indicated, select <all> and execute Dsave.
When saved data is saved to disk media such as CD-R/DVD-R, the data has to be saved
to a hard disk temporary.
The procedure to save data has been described before in the paragraph “Designate
Device for Saving”.
For a procedure for saving to a CD-R/DVD, see the instruction manual supplied with the
writing software.
b. Data Loading
Described below is the procedure to start up the program from the shortcut in the
program list that the Windows OS manages:
1. Select [Data Save Load] under [Rigaku ZSX Guidance] from “All Programs”.
2. On the Logon screen, enter a user name and a password registered for ZSX
Guidance.
3. When an item for a language is displayed, select a language for display.
4. Click OK .
A user who belongs to a group for which the User Administration program has not set
access authority for the “Data Save Load” program cannot log on.
For the details of the access authority for a program, refer to <User Administration - Add
Group> in <6.1 h. System Management>.
The procedure to call the function called “All Programs” to display a list of programs varies
depending on the version of Windows. For the detailed procedure to call it, see <All
Programs> in “INSTRUCTION MANUAL OF WINDOWS FUNCTIONS”.
While this function of the data saving and loading program is in operation, the ZSX
Guidance cannot be started.
When <Specify a folder to load> is selected, be sure to specify a drive name or a folder
name.
Insert a CD-R/DVD-R in which data has been saved into the CD/DVD-R drive and
specify the CD/DVD-R drive as the device for loading in the procedure described below.
The procedure to load the saved data has been described before in the paragraph
“Designate Device for Loading”.
It is possible to load with the same steps as above in case of saving to USB.
Data files to be saved when each data item is selected are as follows:
QLAP_*.ho
Analysis QNAP_*.ho
Conditions STDS_*.ho
\Rigaku\SXIDV\UserData\anlcond ANLCOND.ZIP
USER_*.ho
\Rigaku\SXIDV\UserData\anlcondU ANLCONDU.ZIP
(Qual.) QLDT_*.ho
(Quant.) QLAS_*.ho
QLDS_*.ho
*.emf
QLDT_*.ho
SMDR_*.ho
Analyzed SMDR_*.bmp
Results
\Rigaku\SXIDV\UserData\AnlRslt QNDT_*.ho ANLRSLT.ZIP
(Qual. Data) \Rigaku\SXIDV\UserData\AnlRsltU SMDR_*.ho ANLRSLTU.ZIP
(Quant. Data) SMDR_*.bmp
(Mapping Data)
MPDT*.ho
MPDT*.bmp
Communication
Log (for \Rigaku\SXIDV\Log *.* LOG.ZIP
Maintenance)
Convert Sample ID
Do not start another program in the “Rigaku” folder. If another program is started, data may
be destroyed.
When the folder for saving measurement results has not been changed, select the
“\Rigaku\SXIDV\UserData” folder.
When the e-mail server does not require authentication, setting has been completed now.
Save the information that has been set according to “Save E-Mail Setting”.
When the e-mail server requires authentication, follow the steps described below:
9. When the outgoing mail server requires authentication when sending e-mail, place a
check mark for “Outgoing server (SMTP) requires authentication”.
10.Select <Following account and password are used at log on mail server (SMTP).> or
<Log on to incoming mail server (POP3) before sending mail.> according to the
setting of the e-mail server.
When <Following account and password are used at log on mail server (SMTP).>
is selected
When <Log on to incoming mail server (POP3) before sending mail.> is selected
11.Enter the name of the incoming mail server at “Incoming mail server (POP3)”.
12.Enter the port number of the incoming mail server at “Port number (POP3)”.
13.Enter an account name to log on to the incoming mail server at “Account name
(POP3)”.
14.Enter a password to log on to the incoming mail server at “Password (POP3)”.
When information has been set, save it according to the procedure described below:
Transmitted texts are classified into various types according to their purposes. The data
type determines the combination of transmitted texts and the transmission type.
Transmission Type
The transmission type is defined for a combination of texts or for similar information
items. The transmission formats are described later for each transmission type.
Transmission
Transmitted Data
Type
Quantitative analysis results, qualitative element information
1
and SQX analysis results
2 Drift correction coefficients, bias correction coefficients, etc.
3 Spectra, peak angles, X-ray intensities, etc.
Measuring conditions, 2-theta values of all measuring points and
4
X-ray intensities
The type is the code to classify data and is specified in the first transmitted text. The
receiver processes the succeeding texts according to this code.
Transmission
Type Data Result Type
Type
_1 1 Qualitative element information Qualitative
_2 1 Quantitative analysis results Quantitative
_3 1 SQX analysis results Qualitative
_4 3 Peak list Qualitative
_5 4 2-theta-intensity data Qualitative
_6 2 Drift correction results Coefficient update
_7 2 Bias correction results Coefficient update
_ : Space code
These results are output after each quantitative analysis or using the Result Display
program.
Output Data
Transmission
Analysis Job Analyzed X-Ray
Type
Results Intensities
Quantitative Analysis 1 Output Output
Intensity Measurement 1 Not output Output
Check Analysis 1 Output Output
Drift Correction Analysis 1 Not output Output
Bias Correction Analysis 1 Output Output
Standard Sample 1 Not output Output
In the case of an analysis job that has analyzed results, X-ray intensities can also be
output. By selecting <Output intensities for result> on the “Text Structure Setting”
screen on the “Transmission” tab in the System Parameter program in System
Management, both data items can be output. For X-ray intensities to be output, selection
can be made from net intensities before drift correction, net intensities after drift
correction and intensities after internal standard/overlap correction.
Transmis-
Analysis Job Output Data
sion Type
Drift Correction Sample 2 Drift correction update results
Bias Correction Sample 2 Bias correction update results
These results are output using the Qualitative Data Handling program.
Transmis-
Analysis Job Output Data
sion Type
1 Qualitative element information
1 SQX analysis results
Qualitative Analysis
3 Peak list
4 2-theta-intensity
b. Transmission Specification
Interface Specification
RS232C
Type RS232C-compliant serial interface
Communication
Data transmission from this system (PC) to external terminal
Method
Communication Selectable from 300, 600, 1200, 2400, 4800, 9600, 19200 and
Speed 38400 BPS
Start bit 1
Data bits Selectable from <7> and <8>
Data Constitution
Stop bits Selectable from <1> and <2>
Parity bits Selectable from <None>, <Even> and <Odd>
Flow Control Selectable from <RS/CS>, <Xon/Xoff> and <None>
Code Type Selectable from <JIS> and <ASCII>
Transmission
<Link control> can be selected or deselected.
Protocol
Error Detection Selectable from <None> and <BCC>
Ethernet
Communication
IEEE802.3
Method
Transmission
TCP/IP
Protocol
Connection 10BASE-T
After the power is turned on, the ER (DTR) signal of this system is always turned on. When
the receiver is a computer terminal or a similar device, connect the DR signal (pin No.6) of
the receiver and the ER signal (pin No.4) of this system
When the link control is not used or in the case of the RS/CS control, the RD (RXD) signal
of this system and the SD (TXD) signal of the external terminal need not be connected.
A delay time from one text to the following text can be set in a range of 0 msec to 30 sec.
(30000 msec). This time is called the text delay time. This means the minimum interval
between texts when they are transmitted consecutively without the transmission
protocol. Set this delay time according to the processing speed of the receiver.
Transmission Protocol
When the transmission protocol is used, texts and control codes are transmitted as
shown below:
Example without Text 4 Example with Text 4
Data Processing External Data Processing External
System Terminal System Terminal
ENQ ENQ
ACK ACK
Text 1 Text 1
ACK ACK
Text 2 Text 2
ACK ACK
Text 3-a Text 3-a
ACK ACK
Text 3-b Text 4-1
NAK ACK
Text 3-b Text 4-2
ACK ACK
Text 5 Text 4-n
ACK ACK
EOT Text 3-b
ACK
Text 4-1
ENQ ACK
ACK Text 4-2
Text 1 ACK
ACK Text 4-m
Text 2 ACK
:
ACK :
:
Text 3-a : :
: : Text 5 ACK
:
EOT
Line Matrix
The rule to exchange control codes and texts when the transmission protocol is used is
shown below using a line matrix for the data processing system:
Response A B C D
Status ACK NAK Invalid Response Time-Out
W To X after Terminated To W after sending Same as
Waiting for sending text after sending retry ENQ. C-W
response after EOT To A-Y on retry over.
sending
calling ENQ
X To X after To X after To Z after sending Same as
Waiting for sending next resending. demanding ENQ C-X
response after text. To A-Y on
sending text To Y after retry over.
sending last
text.
Y Terminated To Y after To Z after sending Same as
Waiting for after resending. demanding ENQ C-Y
response after sending EOT To A-Y on
sending last retry over.
text
Z Same as Same as B-W To Z after sending Same as
Waiting for A-W retry ENQ. C-Z
response after To A-Y on retry over.
sending
demanding
ENQ
S E
Contents of Text
For the text start code and text end code, preset values in the transmission condition
are used in the case of the RS232C or TCP/IP transmission, while fixed values are used
in the case of the file output.
Up to two bytes can be set for the text start code or text end code for the RS232C or TCP/IP
transmission.
Kinds of Texts
The order of transmission of texts varies depending on the transmission type. In the case
of transmission just after an analysis, the start text and end text can be omitted.
Start
End
To omit the start text and end text in the case of transmission just after an analysis, deselect
<Output start and end texts> in the transmission condition in System Parameters.
Texts 1, 2 and 5 have formats common to all transmission types. “_” (an underline) will
represent a space code hereafter.
Text 1
Text 1 indicates the start of transmission. In the case of transmission just after an
analysis, Text 1 and Text 5 can be omitted by making appropriate setting using the
“System Parameters” program.
1 2 3 4 5 6 7 8
S 1 _ T T NN E
Type
The data type is represented by two characters. “_” represents a space code.
The text start code is represented by one character hereafter for convenience.
Text 2
No. of components
Analysis Type
This item is edited when setting has been made in “Text to transmit” and “Select to
transmit” for the item <Type> on the “Sample ID Structure Setting” screen. Up to
16 characters can be transmitted.
This item is used for compatibility with old type analyzers.
Sample ID
This consists of the following four items and setting for transmission is made for
each item. A space code is added between items to be transmitted. When the
number of characters of “Text to transmit” for each item is larger than “Number of
characters”, the length of the text is limited to “Number of characters”. When
“Number of characters” is larger than the number of characters of “Text to
transmit”, the rest is padded with space codes.
Application
This item is edited when setting for transmission has been made in “Select to
transmit” for the item <Application> on the “Sample ID Structure Setting”
screen. Up to 16 characters can be transmitted and this number of characters
is used for <No. of Code Characters> in Text 1.
Sample Name
This is edited from each of <Memo> and <Selections> on the “Sample ID
Structure Setting” screen when setting to output it has been made. A space
code is added between the items. Up to 40 characters can be transmitted for
each item.
Measuring Coordinates
This item is edited when the system has the mapping measurement function
and setting for transmission has been made in “Select to transmit” for the
item <Mapping> on the “Sample ID Structure Setting” screen. The output
format is “-99.9,-99.9”.
Sample Position
This item is edited when setting for transmission has been made in “Select to
transmit” for the item <Position> on the “Sample ID Structure Setting” screen.
Up to four characters can be transmitted. When “Number of characters” is 3,
the sample position becomes, for example, “A-8” in the case of the X-Y type
automatic sample changer and “_12” (squeezed to the right) in the case of the
continual type automatic sample changer.
In the case of drift correction update results and bias correction update results, <Sample
ID> are padded with spaces.
Care should be taken because character positions in a text are different from values that
were preset using “Sample ID Structure Setting”.
Text 5
Text 5 indicates the end of transmission. In the case of transmission just after an
analysis, Text 1 and Text 5 can be omitted by making appropriate setting using the
“System Parameters” program.
1 2 3 4 5 6
S 5 _ 0 0 E
The fixed character string “00” is used for compatibility with old type analyzers.
e. Transmission Type 1
Described below are the formats and contents of texts for quantitative analysis results,
SQX analysis results and qualitative element information:
Start
Start text
Text 1
Text 2 Sample ID
N All components
completed?
Y
N All samples
completed?
Y
End
The transmission order in the case of the quantitative mapping measurement is shown
below.
Start
Text 2 Sample ID
N All components
completed?
Y
When both analyzed values and X-ray
N All data types intensities are transmitted in
completed? quantitative analysis results
Y
N All measuring
points completed?
Y
Text 2 Sample ID
N All components
completed?
Y
N When both analyzed values and X-ray
All data types intensities are transmitted in
completed? quantitative analysis results
Y
End
Text 3 in the transmission type 1 has two items with variable lengths and one item for
which the output size can be selected in a fixed range. For these items, representative
names of “Name of comp.”, “Unit” and “Result” are used on the program screen to set
their output sizes. The procedure to set the output size of each item is described below:
Unit (0 - 8 bytes)
This item follows <Component name>, <Line name>, <Element symbol and
Space> or <Element symbol, Line name and Space>. The default value is eight
bytes. The length of this item affects the text length.
Result (6 - 7 bytes)
This item is handled differently depending on the data type.
No. of
Data Item Name
Digits
Analyzed result or
Quantitative Analysis Result 6 or 7
X-ray intensity
SQX Analysis Result Analyzed result 6
Qualitative With SQX Analysis Result Analyzed result 7
Element
Information Without SQX Analysis Result X-ray intensity 7
Though selection can be made from 6 and 7 digits, the text length is not affected
because the length in the text is fixed to 7 digits. When the length is set to 6 bytes, the
first digit is padded with a space. This is a selectable function for compatibility with old
type analyzers.
Values are edited in the floating point format and squeezed to the right. However, only
quantitative analysis results conform to formats specified for each component under
<Digit> in the limit range condition.
Asterisks (**…*) are outputs for all digits for a large value that cannot be expressed.
1 2 3 4 11 12 13 19 20 21 22 23 28 29 30
S 3 _ C C C C C C C C U U U U U U U U L _ 9 9 9 9 9 9 9 _ E
Component name or Unit Analyzed result
line name (0 - 8 characters) or X-ray intensity
(0 - 8 characters)
Judgment result
Unit
A specified unit is edited in the case of an analyzed result, and “kcps” in the case of
an X-ray intensity.
The character positions shown above are examples and vary depending on the output
sizes. However, the preset numbers of characters are common to quantitative analysis
results, SQX analysis results and qualitative element information.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
S 3 _ C C C C C C C C U U U U U U U U 9 9 9 9 9 9 9 9 9 _ E
Component name Unit Comp. Analyzed result
(0 - 8 characters) (0 - 8 characters) No. (6 characters)
Component No.
The contents of a component are expressed as follows:
Analyzed result
The floating point format with six digits is used.
The character positions shown above are examples and vary depending on the output
sizes. However, the preset numbers of characters are common to quantitative analysis
results, SQX analysis results and qualitative element information.
The format varies depending on whether an SQX analysis result is included or not.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
S 3 _ E E - _ _ _ _ _ U U U U U U U U _ _ _ 9 9 9 9 9 9 9 E
Unit Analyzed result
(0 - 8 characters) (7 characters)
Element symbol
Element Symbol
The symbol of a detected element is output.
The format shown above is an example in the case where the output size of <Comp.
name> is eight bytes. It includes the following six bytes of “- _ _ _ _ _”.
Analyzed result
The floating point format with seven digits is used.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
S 3 _ E E - S S S _ _ k c p s _ _ _ _ _ _ _ 9 9 9 9 9 9 9 E
Line Unit X-ray intensity
name (0 - 8 characters) (7 characters)
Element symbol
X-ray intensity
The floating point format with seven digits is used.
The character positions shown above are examples and vary depending on the output
sizes. However, the preset numbers of characters are common to quantitative analysis
results, SQX analysis results and qualitative element information.
1 2 3 4 1 11 1 1 2 2 3 3 3 3 4 4
0 7 8 4 5 1 2 8 9 5 6
S 4 _ AAAAAAA HHHHHHH L L L L L L RRRRRRRS S S S S S SVVVVVVV E
L Standard Variation
Average Maximum Range deviation coefficient
Minimum
Variation
Fixed to 2 digits Fixed to 2 digits
Coefficient
The number of digits after a decimal point for an analyzed result conforms to the format
specified for each component under <Digit> in the limit range condition.
f. Transmission Type 2
Described below are the formats and contents of texts for drift correction coefficient
update results and bias correction coefficient update results:
The constitution of texts and the transmission order are shown below:
Start
Text 2 Sample ID
N All components
completed?
Y
End
52 53 54 69 70 85 86 95 96 105
9 9 9 9 .9 9 99 9 9 9 9 9 .99 9 99 E
“Low” “Low”
sample sample
ref. int. meas. int.
The character positions shown above are examples and vary depending on the number of
code characters. However, the text length is the same for the drift correction and bias
correction.
1 digit code
“S” and “L” have the same meaning for both coefficient and coefficient . Both short-term
and long-term judgment errors can occur simultaneously.
52 53 54 69 70 85 86 95 96 105
_ J HHHHHHHHHHHHHHHH LLLLLLLLLLLLLLLL 9 9 9 99 .9 99 9 9 9 9 99 .9 9 99
“High” “High”
“High” side sample name “Low” side sample name sample sample
std. value anal. value
9 9 9 99 .9 99 9 9 9 9 99 .9 9 99 E
“Low” “Low”
sample sample
std. value anal. value
The character positions shown above are examples and vary depending on the number of
code characters. However, the text length is the same for the drift correction and bias
correction.
1 digit code
“Over upper limit” means that a coefficient is larger than a standard value plus a
tolerance. “Below lower limit” means that a coefficient is smaller than a standard
value minus a tolerance.
“J” has the same meaning for both coefficient A and coefficient B.
g. Transmission Type 3
Described below are the formats and contents of texts for peak lists. Peak list data of
qualitative analysis results is transferred from the Qualitative Data Handling program.
The constitution of texts and the transmission order are shown below:
Start
Text 2 Sample ID
N All spectra
completed?
Y
End
1 2 3 4 9 10 18 19 27 28 44
S 3 _ S S - S S S PPPPPPPPP 9 9 9 9 9 9 9 9 9 EE-SSSSS-OOOOOOOO
Spectrum X-ray
name Peak angle intensity Line name 1
(6 char.)
45 61 62 78 79
EE-SSSSS-OOOOOOOO EE-SSSSS-OOOOOOOO E
Peak angle/Energy/Wavelength
The types of the abscissa axis and formats are shown below:
X-ray intensity
The length is fixed to nine digits and the floating point format with up to four
digits after a decimal point is used.
Symbol Item
EE Element symbol
SSSSS Line type
OOOOOOOO Order
Up to three line names can be output. When the number of line names is two or
less, vacant areas are padded with space codes. When the order is 1, the area after
“_” is padded with space codes.
Example : Rh-KB1_ _ -Compton_, Ni-KA_ _ _ _ _ _ _ _ _ _ _ _
h. Transmission Type 4
Described below are the formats and contents of texts for 2-theta-X-ray intensity data.
This data of qualitative analysis results is transferred from the Qualitative Data
Handling program.
The constitution of texts and the transmission order are shown below:
Start
Text 2 Sample ID
End
1 2 3 4 11 12 13 14 19 20 23 24 27 28 31 32 40 41 46 47 51
Measuring condition items for 2-theta-X-ray intensity data are output in the following
numbers of digits:
No. of
Item Name Remark
Digits
Spectrum Name 8 Symbol : EL-SSSSS
X-Ray Tube Target Element 2 Symbol : EE
2 digits for tube voltage and 3 digits for
kV-mA 6
tube current
Filter Name 4
Diaphragm Name 4 This means a measuring diameter.
Attenuator Name 4
Crystal Name 9
Slit Name 6
Detector Name 5
Preset Values of PHA 1 8
Used only in the case of PC+SC
Preset Values of PHA 2 8
measurement
Start The format varies depending on the
9
angle/energy/wavelength abscissa axis.
End The format varies depending on the
9
angle/energy/wavelength abscissa axis.
Step 7 “9.99999” format
Full Scale 9 “99999.999” format
No. of Data Items 5 “99999” format
“1”: 2-theta, “2”: Energy,“3”:
Type of Abscissa Axis 1
Wavelength
1 2 3 4 12 13 14 21 22
S 4 _ AAAAAAAAA _ 99999999 E
X-ray
2-theta intensity
2-theta angle/energy/wavelength
The types of the abscissa axis and formats are shown below:
X-ray intensity
The length is fixed to eight digits and the floating point format with up to four
digits after a decimal point is used.
GLOSSARY
Absorption Edge
1.24
V0
a b s
For example, to generate K series X-rays (such as K and K1) of Ti, energy higher than
the K absorption edge must be absorbed by a sample. In that case, voltage of about 5kV
must be applied.
Accuracy
This is the error between standard values and X-ray analysis values in the empirical
calibration method. This value is calculated as the root mean square of errors from a
calibration curve including errors caused by a sample. This is usually expressed as d.
(Ci Wi) 2
d
nm
Item Description
Ci Chemical analysis value
Wi X-ray analysis value
n Number of samples
m=2 for linear formula, m=3 for quadratic formula
m
The number of matrix correction coefficients is also added.
Reproducibility
Aging
To turn on the X-rays after they have been turned off for a long time, the X-ray tube
voltage and current must be elevated slowly to the specified load step by step. Raising
the tube voltage and current slowly is called aging. A longer period when the X-rays have
been turned off requires a slower speed to raise the tube voltage and current.
The automatic aging function automatically raises the tube voltage and current at an
appropriate speed according to the period during which the X-rays have been turned off.
Analysis Type
The analysis type is a term used to indicate the type of preset analysis using this system.
Examples of the analysis type are qualitative analysis and quantitative analysis, as well
as items for the control analysis such as the drift correction and PHA adjustment.
Analyzing Crystal
The single crystals used for X-ray spectroscopy are called analysis crystals. In a
wavelength dispersive X-ray fluorescence spectrometer, the crystal is located between
the sample and the detector, and the crystal and the detector are moved according to
Bragg’s formula to measure fluorescent X-rays of a specific wavelength. Oriented LiF, Ge,
PET, TAP, etc. are used as crystals.
Synthetic Multi-Layer
APC
This is an abbreviation for Automatic Pressure Control that is used to keep the vacuum
degree constant when the measuring atmosphere is set to the vacuum mode. Even in the
vacuum mode, the transmittance of X-rays varies depending on the vacuum level.
Therefore, the APC is generally used because a comparison with previous data is easier
if a constant vacuum degree is maintained.
In some cases, however, especially for samples with large surface areas such as activated
carbon and zeolites, measurement without the APC is more advantageous.
Apparent Concentration
Wi BijWj
Xi
1 AijWj
Xi Wi BijWj
Absorption term Aij is used for correction of X-ray intensity.
Application
The application is a set of measuring conditions and calculation conditions that have
been preset and reserved for the qualitative and quantitative analyses. To perform
qualitative analysis, a qualitative application must be prepared. To run quantitative
analysis, a quantitative application must be prepared. All information for analyses is
stored in this application.
Application Package
In the quantitative analysis, an application package contains all the following parts of
the analysis: measured components, measuring conditions, standard sample
information and correction information necessary to prepare calibration curves.
Standard samples to prepare calibration curves are also included. By measuring
standard samples after necessary sample preparation, regression calculations are made
automatically and calibration curves are calculated. Quantitative analysis of unknowns
can then be started.
Application Template
Application templates exist within the software that allows the user to quickly create a
qualitative or quantitative application. Measuring conditions and calculation conditions
optimum to analyses have been registered beforehand. Therefore, one can create either
qualitative or quantitative applications by changing only a few necessary parts.
Attenuator
Background Function
The following functions can be used for the background calculation. The linear,
quadratic or cubic formula is used to calculate continuous X-rays as background. To
calculate the background of a small peak on the foot of a large peak, the Lorentz or
hyperbolic function is generally used.
I i a1 a2 X i a3 X i a4 X i
2 3
Lorentz Function
a1
Ii a 4 a5 X i
X i a2 2 a3
Hyperbolic Function
a1
Ii a3 a 4 X i
X i a2
Item Description
Ii X-ray intensity
Xi 2-theta angle
a1 - a5 Coefficients
In this method, the intensity ratio of net intensity to background intensity is calculated
and used as the ordinate axis of a calibration curve. This is used, for example, to correct
effects in samples when analyzing trace heavy elements in samples that have light
elements as the main components. In the FP method, the background ratio method
cannot be set.
I n et
Ic
I BG
Item Description
Ic X-ray intensity after correction
Inet Net X-ray intensity
IBG Background X-ray intensity
Bias Correction
In the case of iron and steel samples, when the heat history in the preparation of
standard samples used to prepare a calibration curve is different from that of unknown
samples, a constant analysis error from the calibration curve may appear. The bias
correction fixes this error of analyzed values from the calibration curve. Usually,
samples with known analyzed values are measured after a calibration curve is prepared
and then the bias correction is set if necessary.
For the bias correction, one can correct only the intercept of a calibration curve or apply
a correction function to calculate correction values.
Binder
When a pellet sample prepared by pressing powder sample is breakable, powder and
binder (10% to 20% of sample weight) are mixed before pressing. Many kinds of binders
such as cellulose and boric acid are used.
Bragg’s Formula
Fluorescent X-rays generated from a sample reach the crystal or artificial multi-layer,
and reflect selectively when they satisfy the following formula. represents the crystal
angle and the detector angle 2 is used for the abscissa axis on a qualitative analysis
chart.
n 2d sin
Item Description
Wavelength of diffracted X-rays
d Lattice spacing of crystal
Crystal angle on diffraction
Order (1 represents the measured element line,
n
and 2 or larger represents the higher-order lines.)
Bulk Sample
A sample that can be regarded as infinitely thick for X-rays is called a bulk sample.
Thick samples such as metal disks and pressed powder samples are treated as bulk
samples. It is impossible to define generally the thickness that can be regarded as that of
bulk samples. Distinctly from bulk samples, finitely thick samples are called thin film
samples.
Wi AI 3 BI 2 CI D
Item Description
Wi Analyzed value
A, B, C, D Calibration curve coefficients
I X-ray intensity
Check Analysis
A check analysis is run to determine whether prepared calibration curves are in normal
operation. A sample with known concentrations is analyzed and analyzed values are
calculated. The analyzed values are judged against upper and lower limits registered in
the limit range conditions and judged results are output together with the analyzed
results.
The same check sample can be measured three to five times (specified times) a day and
results can be recorded for X-R control. Output components can be selected in the limit
range condition.
Compound
A compound is a component such as silicon dioxide (SiO2) that consists of two or more
bonded elements. Besides oxides, lithium tetraborate (Li2B4O7) and many other
multi-element materials are also called compounds. In programs using the FP method
such as the SQX analysis, calculation methods such as metal conversion and oxide
conversion are available, and the constituting elements and the ratio of the numbers of
atoms in a compound are important. Standard oxides that are routinely used have been
set in the compound table.
Diaphragm
This is a mechanism to limit a measurement area for an analysis. Care must be taken to
make sure that a sample mask with a diameter larger than that of the preset diaphragm
is used. Otherwise, the sample mask will also be measured. Use a diaphragm suitable
for the diameter of a sample mask.
Differential Curve
An energy distribution curve of X-rays that enter the detector after being separated by
wavelength using the crystal is called a differential curve. This is also called a pulse
height distribution curve. This curve is measured to check electronic noises, interference
of higher-order rays and the resolution of a detector.
PHA
Differential Measurement
Measurement using the upper and lower limits of the PHA is called the differential
measurement. The lower limit is set to eliminate electronic noises and the upper limit to
eliminate higher-order rays. A range between those upper and lower limits is called a
window.
Integral Measurement
Drift Correction
Since measured X-ray intensities change with time, calibration curves shift (drift) as
they are used for long periods of time. The drift correction adjusts the intensities to
compensate for this shift. The drift correction is performed to continue to use prepared
calibration curves for a long period of time. The same operation is also necessary in the
FP quantification method. The following two methods, the alpha method and the
alpha-beta method, can be used.
Alpha Method
This is also called the one-point method. Only one sample that has about half the
highest X-ray intensity in the calibration curve range is set.
Alpha-Beta Method
This is also called the two-point method. Two samples that have a high intensity and a
low intensity in the calibration curve range are set. This method is used when a
concentration range is wide and when X-ray intensities that include background
intensities are used.
Excitation Condition
Optical conditions to generate primary X-rays from the X-ray tube and irradiate a
sample with the primary X-rays are called the excitation condition. The target element,
tube voltage, tube current and primary filter are included in the excitation condition.
EZ Analysis
This is an analysis program in which functions necessary for daily routine analyses have
been put together.
It is possible to reserve and make analyses and check analysis results easily on a graphic
screen.
EZ Scan
EZ Scan is an analysis type to carry out the SQX analysis. This is treated as a simplified
mode in the SQX analysis. Though some restrictions are imposed on measurement and
control, advanced SQX analyses can be carried out without preparing applications for
analyses.
Fixed angle analysis is designed to measure a spectral line of a component by fixing the
crystal position and detector position at the peak position and a background position and
by counting in each position for a fixed time.
The function that performs a series of analyses automatically at a fixed time is called the
fixed time analysis. This can be applied to a periodical analysis such as the PHA
adjustment and drift correction. The Analysis program must have been started. Even
when a series of sample IDs is being analyzed in the Analysis program, IDs preset for
the fixed time analysis are analyzed preferentially after the analysis of the present ID is
complete.
Flow Bar
Flux
In the heavy element dilution and in the preparation of a glass fusion bead, material
mixed with sample in a constant proportion is called flux. For example, lithium
tetraborate (Li2B4O7) or sodium tetraborate (Na2B4O7) is generally used in the
preparation of a glass fusion bead.
FP Quantification Method
The quantification method using the fundamental parameter method is called the FP
quantification method. It is assumed that elements are distributed uniformly for X-rays
in a sample in the case of a bulk sample or in each layer in the case of a thin film sample
(homogeneous). On this assumption, fluorescent X-ray intensities can be expressed as a
function of physical constants (fundamental parameters) such as the concentrations of
components, mass absorption coefficients, fluorescence yields and spectrum of the X-ray
source. A correlation between theoretical fluorescent X-ray intensities calculated using
this FP method and measured X-ray intensities is expressed with a sensitivity
calibration curve. The FP quantification method is an analysis method to calculate
quantitative values using this sensitivity calibration curve.
Measurement to scan a wide range of the 2-theta angle with a combination of a crystal
and a detector is called the full range qualitative analysis. For example, “Heavy” is a full
range qualitative analysis to measure a range of 5 to 90 degrees with the LiF-SC optical
system. This full range qualitative analysis can check the existence of components from
Ti to U and the measured element lines of two or more components can be used for the
SQX calculation. On the other hand, the qualitative analysis method for the light
element region is called the partial qualitative analysis. This means measurement to
scan a partial 2-theta range with a combination of a crystal and a detector.
To separate overlapping peaks using a function, the following function is used for
approximation. Up to four peaks can be separated. The peak deconvolution function is a
combination of Gaussian and Lorentzian factors.
I i h r fg 1 r fl a i 2 b i c
ln 2i u 2
fg e W2
1
fl
(i u ) 2
1
W2
Item Description
fg Gauss function
fl Lorentz function
Fitting intensity (intensity of channel i)
h Peak intensity
Proportion of Gauss function to total
r
(0.0 -1.0)
u Data channel No. of peak
i Data channel No.
w Half width at half maximum of peak
a, b, c Coefficients
This software has optional correction functions for a dilution ratio and a loss on ignition
in a glass fusion bead analysis using the empirical method. These matrix correction
coefficients are calculated using the theoretical matrix correction coefficient calculation.
In the matrix correction method [Concentration correction] which is normally used,
following calibration equation is used.
When a loss on ignition component is set as a balance component (residual) in the loss on
ignition correction, a correction term for the loss on ignition component is not included,
and information on the concentration of the loss on ignition component is unnecessary.
The loss on ignition component of a glass fusion bead is named “Ig” when used in this
software. The dilution ratio is flux weight divided by sample weight, and the oxidizer
addition ratio is the residual weight of oxidizer divided by sample weight.
When the evaporation correction is made for flux, the value of (Bead Weight/Sample
Weight - 1) is used as the dilution ratio.
The X-ray fluorescence analysis is a method that analyzes sample surfaces 1 to several
micrometers deep. Therefore, in the case of powder samples, X-ray intensities may vary
depending on grain sizes and their non-uniformity. That phenomenon is called the grain
size effect. In such a case, the effect can be reduced using glass fusion beads or
pulverization.
Mineralogical Effect
Gross Intensity
Net Intensity
Higher-Order Line
All fluorescent X-rays that satisfy the Bragg’s formula at a certain angle enter the
detector. When the analyzed line has a certain wavelength , a line having the
wavelength ’ that satisfies =n’ (n>1) also enters the detector and is counted. The
X-rays with that wavelength ’ are called a higher-order line of the wavelength . The
wavelength with n=1 is called the primary line. For example, when Pb-L1 is analyzed
and a sample contains Sn, Sn-K1 appears as a secondary line.
Since the wavelength is inversely proportional to the energy, when equals 2’, E’
equals 2E in the term of energy. So a higher-order line can be separated using the
difference of detected energy.
PHA
Identification Analysis
One of the data processing features in the qualitative analysis is the automatic
identification analysis of detected peaks. A marker is attached to each peak detected in
the peak search. For example, Fe-K andSi-K are used as markers.
Impurity Correction
This is a correction method used in the SQX analysis. The impurity correction can be
used, for example, to correct for impurities in the flux of a glass fusion bead and impurity
lines from the X-ray tube. Selection can be made from concentration correction and
intensity correction. To correct impurity lines from the X-ray tube, it is generally
recommended that concentration correction be used because the intensities of impurity
lines vary depending on the matrix of a sample.
Integral Measurement
The integral measurement uses only the lower limit of the PHA with no upper limit.
This method eliminates electronic noise. However, care must be taken because all
higher-order lines are counted. To select the integral measurement, designate 0 (zero)
for the upper limit. This method can be used when main components with very high
intensities are measured and there is no effect of higher-order lines. In that case, stable
measurement is possible with almost no counting loss or variation in intensities caused
by changes in pulse heights.
Differential Measurement
The internal standard correction is a type of matrix correction where the intensity ratio
of an analysis spectral line to an internal standard spectral line is calculated. A certain
amount of a standard substance that can generate a peak near the peak of an analyzed
component is added. That standard substance is called the internal standard substance.
Since both peaks are affected by coexisting elements in similar ways, the matrix effects
can be corrected.
In some cases, the scattered rays of the characteristic X-rays of the X-ray tube or a
background intensity near the analyzed line are used as an internal standard spectral
line. This method is called the scattered rays internal standard.
The internal standard correction cannot be set in the case of the FP method.
Ii
Iic
Ij
Item Description
Iic X-ray intensity after correction
Ii X-ray intensity before correction
Ij X-ray intensity of internal standard
Loss on Ignition
This means a decrease in sample weight due to the evaporation of H2O and CO2 during
fusion when analyzing a powder sample containing water of crystallization or carbonate
using the glass fusion bead method. It is also called the LOI. This software has an
optional function to correct this loss on ignition in the quantitative analysis using the
glass fusion bead method. There is another method in which a powder sample is ignited
before an analysis using the powder method. On the other hand, Fe3O4 in a sample may
oxidize to Fe2O3 in the glass fusion bead method. This phenomenon is called a gain on
ignition (GOI).
Data entered from the keyboard on sample ID setting instead of measuring X-ray
intensities to calculate analyzed values are called the manual input data. Data is
entered for a component for which “Input” has been set for “Analysis” on the
“Component Selection” screen in the application information. Since this data is entered
on individual sample ID setting unlike fixed value setting, a value obtained with
another analysis method can be entered for a specific component in an unknown sample.
Matrix Correction
The calculation to correct the effects of coexisting elements in the empirical calibration
method is called the matrix correction.
Matrix correction method is selectable from [Concentration correction] and [Intensity
correction].
In case of [Concentration correction] which is normally used, the matrix correction
equation is expressed as general formula below and used for the calculation.
RijFj
Wi ( AIi 3 BIi 2 CIi D)(1 Ki AijFj QijFjFk )
1 Wi / 100
BijFj DijFjFk Ci
The correcting values of Fj and Fk in the correction terms of Aij, Qij, Rij, Dij and Bij can
be used for either concentration correction or intensity correction. When expressed in
uppercase characters, concentration corrections are used. When expressed in lowercase
characters, intensity corrections are used.
In general, the Aij term is employed for absorption/enhancement correction and the Bij
or bij term for overlap correction.
RijFj
Wi CIi(1 Ki AijFj QijFjFk )
1 Wi / 100
BijFj DijFjFk D
Measurement Atmosphere
Mineralogical Effect
The phenomenon that X-ray intensities vary depending on the differences of the
configuration of analyzed components and the kind of a mineral crystal is called the
mineral effect. In general, this is often seen in powder or block samples of rock and
mineral. This effect can be reduced by using glass fusion beads or by pulverizing
samples.
Net Intensity
→ Gross Intensity
Overlap Correction
This is the correction for the overlap of interfering rays with analysis rays. Correction
can be made for net intensities after the drift correction using up to three peaks. Since
this correction employs X-ray intensities, this can be used for both the empirical
calibration method and the FP method.
Oxidizer
This is an additive to oxidize a sample when a glass fusion bead is made from sulfide,
carbide or a metal sample. NaNO3, LiNO3, etc. are used as oxidizer.
→ Glass Fusion Bead
PC
SC
PC Wire Cleaning
The PC uses the P-10 gas as its detector gas. The P-10 gas consists of Ar gas and
methane gas mixed in a proportion of 9 to 1. When one X-ray quantum enters the
detector, Ar gas is ionized to emit electrons that ionize other Ar atoms in an avalanche
process. A large number of electrons reach the center wire and one pulse is counted. In
such a case, the methane gas, which is added to prevent subsequent discharges, absorbs
electrons, ionizes and sticks to the center wire. Since the methane gas that sticks to the
wire is not conductive, electrons cannot transfer appropriately, resulting in bad energy
resolution. PC wire cleaning is executed to restore the original condition.
Peak Deconvolution
Adjacent spectral lines in 2-theta scan data are separated into individual spectral lines.
Peak deconvolution has the following two methods.
Function Deconvolution
Overlapping peaks are separated by fitting those using functions. The waveform of each
peak is calculated as a combination function of the Gauss function and the Lorentz
function.
Overlapping peaks are separated using peak profiles that were measured and extracted
beforehand. More accurate separation can be carried out using the standard profile
deconvolution.
Peak Search
This is one of the qualitative data processing items. Peak search detects a peak from a
qualitative analysis chart. Using several points on the right and left of each measured
data point, the position where a parabola can replace those points is found. The program
takes statistical fluctuations of X-rays into consideration when setting requirements for
locating a peak.
Threshold 3 k 3 k I BG
Item Description
k Peak detection sensitivity constant (k=0.3-2.0)
IBG Background X-ray intensity
PHA
A unit that discriminates pulse heights according to the energy of X-rays that enter the
detector is called the PHA (Pulse Height Analyzer). The pulse heights of electronic
signals vary depending on incident energy. The PHA selects only signals in an objective
range using differences in pulse height. Pulse height is expressed by a scale with
arbitrary units and an adjustment is made so that the maximum of measured X-ray
pulses will be at 200. X-ray signals captured by the detector consist of the following.
Signal pulses of For example, pulses generated when the Tl-M line of the
fluorescent X-rays from crystal TAP used is excited and enters the detector when
crystal the Na-K line is measured.
PHA Scan
Measurement of the energy distribution of incident X-rays with a fixed detector angle is
called the PHA scan. The energy distribution is expressed with a pulse height
distribution curve. This scan is used to check electronic noises and the interference of
higher-order lines. It is also used to check the energy resolution of a detector.
PHA Adjustment
The PHA adjustment is carried out to set the pulse height of a measured element line to
200. The PHA adjustment can be made automatically using PHA adjustment samples.
Preparatory Evacuation
P-10 Gas
Gas for the detector that is always flowing through the PC is called the P-10 gas. In
general, P-10 gas, which is the mixed gas of Ar 90% and methane 10%, is employed.
Primary Filter
This is located between the X-ray tube and the sample and is used to reduce the
interference of continuous X-rays or characteristic X-rays generated from the X-ray tube
on the analysis X-rays. This is also used to lower background. When the primary filter is
used, X-ray intensities lower relatively. However, since lower background results in
smaller calibration curve coefficients, analyses can be made with higher precision.
Detection limits are also improved. The kinds and purposes of the main filters are shown
below.
Filter Description
Absorbs the Rh K-line.
Ni400
Effective for analyses using the K-lines of Ru, Rh, Pd, Ag, Cd, etc.
Ni40 Effective for trace analyses using the Pb L-line and As K-line.
Qualitative Analysis
The qualitative analysis to see whether a specific element exists in a sample is called the
partial qualitative analysis.
Quantitative Analysis
Scattered Rays
Quantitative Analysis
The analysis to determine the quantity of a known component is called the quantitative
analysis. The standard sample comparison method with the empirical calibration
method or the FP quantification method is used for analyses. In general, a qualitative
analysis is made first to know components that comprise a sample and then a
quantitative analysis is made.
Qualitative Analysis
Reproducibility
The standard deviation of analyzed values obtained by measuring the same sample
many times is called the repeatability. This value is used to know the so-called stability
of the spectrometer. The standard deviation of X-ray intensities is generally calculated
with the following formula.
( Ii I M ) 2
n 1
Item Description
Ii Each measured X-ray intensity (kcps)
IM Average X-ray intensity (kcps)
n Number of repetitions
The theoretical standard deviation of X-ray intensities is calculated with the following
formula.
IM
cal
t 1000
Resolution
The following two kinds of resolution are defined in the X-ray fluorescence analysis.
Energy Resolution
The energy resolution shows the performance of a detector and is expressed as the
percentage of the half-value width of a differential curve to the peak position. The
half-value width is an energy width where the intensity is half of the peak.
Angular Resolution
This term is used to describe the performance of an optical system and is the angle width
of a peak on a qualitative analysis chart. This is assessed with the half-value width or
the 10% width. The 10% width is an angular width where the intensity is 10% of the
peak.
The sample film correction adjusts for the absorption of X-rays by the film when making
a measurement using a polypropylene film (P.P. Film) or a polyester film (P.E. Film) for
analyses of a liquid sample or the like in the SQX analysis. Though a film designated by
Rigaku must be used, a library need not be registered beforehand using a sample with
the film.
Sample ID
Information that identifies each sample is called the sample ID. Identification
information such as the type of an analysis, an application name and a sample name,
and information about processing after an analysis such as result output are set on an
analysis. After an analysis is completed, a sample ID is used to retrieve an analyzed
result.
Sample ID Table
A series of specific analyses on certain samples can be set into a sample ID table. This
can be loaded as an individual analysis and designated for the program operation or
cycle repeat analysis. One can save time and trouble by registering a series of sample
IDs used frequently and routinely such as ones for the PHA adjustment and check
analysis.
Sample Model
Sample models must be set to calculate concentrations with an FP program such as the
SQX analysis. Those models are classified into the metal and the oxide.
Metal
Oxide
This choice is used for samples in which each element exists as a compound such as an
oxide. In the case of powder sample such as glass, concentrations are generally
calculated assuming that each component exists as an oxide. Standard oxides to be used
have been registered beforehand in the compound table.
Sample Spin
To rotate a sample during measurement is called the sample spin. When a sample
surface has directivity (in the case where a metal sample was ground using a belt or in a
similar case), measured intensities may vary depending on the sample direction if the
sample is not spun. By rotating a sample during measurement, directivity can be
canceled and analysis precision can be improved. In the case of liquid samples, however,
measurement is generally made without the sample spin.
SC
PC
Scattered Rays
Primary X-rays generated from the X-ray tube contain the characteristic rays of the
X-ray tube target element and continuous X-rays. Those X-rays are reflected and
scattered by a sample and become the scattered rays. There are two types of scattered
rays: Thomson scattered rays, which maintain the same energy, and Compton scattered
rays, which are observed with some loss of energy. Since these scattered rays contain
information about a sample, they can be used for analyses using the FP method. To
analyze the two types of scattered rays generated from the X-ray target used is the first
step of the identification analysis. Most of the background of fluorescent X-rays can
generally be regarded as scattered rays of continuous X-rays from the X-ray tube.
Sealing
In the analyses of liquid samples, the measuring chamber is filled with helium so as to
reduce the absorption of X-rays by the atmosphere. A seal with a thin film is placed
between the spectroscopic chamber and the sample chamber to separate the two
chambers in this spectrometer. By using the seal, it is possible to set only the sample
chamber in the helium atmosphere keeping the spectroscopic chamber in a vacuum. And
this method can reduce the consumption of helium gas.
Secure
Sensitivity Library
The relationship between X-ray intensities obtained by measuring library samples such
as pure metals and reagents and theoretical X-ray intensities calculated with the FP
method is called the spectrometer sensitivity. A collection of sensitivity values is called
the sensitivity library. A collection of sensitivity values obtained by measuring
representative elements using library samples without using standard samples for each
kind of samples is called the common library. It is also possible to analyze elements that
have not been registered in the library by estimating the sensitivity values of those
elements using the sensitivity values of elements before and after those elements in the
periodic table. Even when conditions for the diaphragm and the attenuator are different,
correction is made automatically. The sensitivity library can be used for both qualitative
and quantitative analyses.
Slit
The slit is a mechanism to make the fluorescent X-rays generated from a sample parallel.
By changing the type of slit, the angular resolution for spectral lines can be changed.
When spectral lines overlap, interference can be lessened by using a slit with high
resolution (Fine slit). For an artificial multi-layer such as RX61 to be used for the
measurement of boron (B) or carbon (C), when there is no adjacent spectral line, high
intensities are obtained by using a slit with high sensitivity.
Slit Description
Smoothing
This is the function of smoothing processing for a qualitative spectrum. As the number of
points used for smoothing becomes larger, the chart becomes smoother. But if the
number of points is too large, the spectrum will be deformed.
It BI 2 CI D
Item Description
It Theoretical X-ray intensity
I Measured X-ray intensity
B, C, D Spectrometer sensitivity coefficients
SQX Analysis
If interfering rays overlap with a peak for analysis in the SQX analysis, overlap
correction can be made using overlap correction coefficients stored beforehand in each
spectrometer. Analyzed values can be calculated using net intensities after the
correction. This correction has been registered mainly for the overlap of primary X-rays
and may not have been registered for higher-order rays etc.
When the samples of industrial wastes and organism materials are analyzed by
conventional FP method, the difficulty on XRF analysis takes place because the samples
mentioned before contain non-measurable components by XRF such as carbon and
hydrogen. In this application, SQX Scattering FP Method becomes the preferred
technique, which can calculate the concentration of such components by monitoring
scattered X-rays (Rh-KA Compton and Rh-KA Thomson scatterings).
Scattered Rays
Standard Profile
A standard profile is the spectral data of a single line such as the Ti-K or Zn-L line.
This is registered using a pure material etc., and is used for the standard profile peak
deconvolution. In the ZSX Guidance, this file is created in the standard profile
extraction in the Qualitative Data Handling program.
I i j Gji j a i 2 b i c
Item Description
Fitting intensity (intensity of channel i)
j Contribution coefficient of standard profile j
Gj Standard profile intensity (intensity of channel i+j)
i Data channel
j Channel direction correction value for standard profile j
a, b, c Coefficients
Synthetic Multi-Layer
Analyzing crystals that are artificially made of a heavy element that behaves as a
reflection layer and a light element that behaves as a spacer are called synthetic
multi-layers. These are used to analyze components in the longer wavelength region
where the Bragg’s formula cannot be satisfied with a single crystal. Synthetic
multi-layers are employed mainly for analyses of elements lighter than O and are
labeled as the RX series.
Analyzing Crystal
Matrix Correction
This term describes a sample that has a finite thickness for X-rays, such as metal-coated
electronic material or a coated sample. This sample does not have an enough film
thickness that all X-rays are coming from the film. The measurement intensities are
affected by its thickness. In analyses with the FP method, a film thickness and
concentrations can be analyzed simultaneously using this principle.
Bulk Sample
These are the voltage applied to the target of the X-ray tube and the current that flows
through the filament. The tube voltage and current are limited by the restrictions on the
filament current as well as the maximum load of the X-ray tube. If an improper selection
is made, an error message will be displayed. The X-ray intensity is roughly proportional
to the current. It is not linearly proportional to the voltage and depends on an X-ray tube
target, element and measured element line. In general, for heavier elements (shorter
wavelengths), the intensity becomes higher than a linearly proportional value as the
voltage becomes higher.
2-Theta Angle
A crystal is used for the spectral analysis of fluorescent X-rays. When the crystal angle is
theta, the detector angle is 2-theta because of the conformal reflection according to the
Bragg’s formula. Both the crystal and the detector move according to the wavelength to
be measured. The detector angle 2-theta is conventionally used for the abscissa axis of
the qualitative analysis spectral chart.
2-Theta Scan
The crystal and detector are rotated from low angle to high angle to measure an X-ray
intensity at each angle (wavelength or energy). This is used for the qualitative analysis
and the measurement of a peak shape. In general, a chart is drawn with the 2-theta
angle as the abscissa axis and the X-ray intensity as the ordinate axis. A chart can also
be drawn with its abscissa axis converted to the wavelength or energy.
Unit
The following units are used in this software. Units for thickness are used for thin film
samples.
Concentration Description
mass% Mass%
ppm 1/10000 of mass%
at% Ratio of numbers of atoms
mol% Ratio of numbers of molecules
Weight Description
g/m2 g/m2
mg/m2 mg/m2
mg/cm2 mg/cm2
g/cm2 g/cm2
The unit for density is fixed to g/cm3. The density of a layer can be selected from a value
fixed to the layer or one calculated from the density of each component. It is also possible
to add a unit.
The database that has information about standard samples on the market such as NIST
(National Institute of Standards and Technology) and JSS (Japanese Steel Standard) is
called the Universal Standard Samples. When you have a sample in this database, you
can set the information for that standard sample without entering standard values
manually. Furthermore, by adding your standard sample to the database, you can
handle it as a universal standard sample.
Weight Thickness
This means sample weight per unit area, for example, g/cm2. It corresponds to the mass
thickness in the case of a unit of mass. In the case of a coating sample, the coating
weight of a coating layer is also the weight thickness. It is also called an area density.
X-Ray Intensity
In the X-ray analysis, the X-ray intensity is expressed as the number of X-ray photons
that enter through the detector window in a unit time. In general, the number of counts
per second is expressed in units of kcps (kilo-counts/second). In the International System
of Units (SI), the X-ray intensity is expressed as the total electrical charge of ions
generated in air by electrons generated by interactions with 1kg of air through which
X-rays pass. Its unit is C/kg (Coulomb/kilogram).
X-R Control
Check analyses are carried out three to five times a day to control the precision of
analyzed values. Variations of check analysis results are monitored using an X-R control
chart. Check samples are registered with limit range conditions and the averages (X)
and ranges (R) (maximum values minus minimum values) of analyzed results are
assessed daily. Upper and lower limits in limit range conditions are used as control
limits for averages on an X-R control chart. R upper limits are used as control values for
ranges.