5.

After removing the tip cone, wipe the Important Notes


piston, the O-ring and the tip cone with
TROUBLE SHOOTING
ethanol and a lint-free cloth. 1. Procedure should take place at 20 (± 0
Trouble Possible Reason Correction
Note : Models up to 10µl have a fixed O-ring location inside the tip 0.50) C. constant temperature.
cone, so do not remove or maintain it.
2. The weighing beaker, distilled water, Droplets left Unsuitable tip, non-uniform Use new tip
6. Before replacing tip cone, it is inside the tip wetting of the plastic
pipettor & tips must be at the same
recommended to grease the piston
temperature. Tip holder (cone) scratched Change the
sligthtly using the silicone grease
or damaged tip cone
provided. 3. Use an analytical balance with 0.01 Research micro pipette with smart grip
mgs readability. Aspirate & discard
7. After reassembling, use the pipettor the organic solvent
(without liquid) several times to make 4. Pre-rinse the tip 3 to 5 times before Organic Solvent as liquid. several times before
sure that the grease is spread evenly. pipetting. actual pipetting by
the same tip.
8. Check the pipettor calibration. 5. Divide the weight of the water by its
density (at 200C, 0.9982) to get the Leakage or Tip incorrectly attached Attach firmly.
Note : Never disassemble the upper part of the pipette. To avoid
Pipetted volume
losing or damaging fragile parts, reassemble the pipette volume. too small Unsuitable tip. Use new tip.
immediately.

Foreign particles between Clean the


Recalibration tip and tip cone. tip cone.

Insufficient amount of grease Clean & grease


CALIBRATION 1 Place the service tool into grooves at the on piston and O-ring. O-ring and piston.
base of the push button as shown below.
Turn it clockwise to increase & O-ring not correctly Change the
Each pipette has been checked & calibrated positioned or damaged. O-ring.
anticlockwise to decrease the volume.
at factory with procedure conforming to EN-
Incorrect operation. Follow instructions
ISO 8655 standards. It is recommended to 2 Repeat the 'Checking calibration' carefully.
check the calibration at least once a year, for procedure.
regularly used pipette. Calibration altered. Recalibrate according
to instructions.
Inaccuracies
Unsuitable for the particular Use correct pipetting
Checking calibration liquid pipetting technique. technique.

Instrument damaged. Send for repair.


1 Fit new tip onto the tip cone.
Push button Piston contaminated. Clean & grease
2 Pipette distilled water into pre-weighed jammed or O-ring
beaker at least five times & record the moves erratically Penetration of solvent vapours. and piston
weight each time.
Tip ejector Tip cone Remove ejector
3 Compare the results with the permitted jammed or contaminated collar and clean tip
moves erratically. from outside. cone’s outer surface
variation chart given below. The with ethanol.
calibration of the pipette must be set even
if only one of the results falls outside the Volume setting is not Click stop mechanism Send for repair.
properly click stopped. damaged.
permitted range.
Push button does not Use of excessive force Send for repair.
turn for volume setting. beyond the range of pipette.
Volume Variation Volume Variation
permitted permitted
1 µl ± 0.15 µl 50 µl ± 0.80 µl
2 µl ± 0.20 µl 100 µl ± 1.50 µl STORAGE
5 µl ± 0.30 µl 200 µl ± 2.00 µl
When not in use, it is recommended that your
10 µl ± 0.30 µl 500 µl ± 5.00 µl pipettor is stored in a vertical position. Fine Care Biosystems
20 µl ± 0.40 µl 1000 µl ± 10.00 µl
Leaving the pipette on its side can cause
For user in a normal lab environment, the limits may be
liquids to leak into the body of the pipette and
doubled. cause corrosion.
PIPETTE DESCRIPTION PIPETTING TECHNIQUE PIPETTING RECOMMENDATIONS MAINTENANCE

These pipettes are available as fixed or A. Forward Pipetting Aspirate liquid into the pipette only To maintain the best results from your
variable, general purpose micropipettes for pipettor, each unit should be checked every
sampling and dispensing accurate when a tip is attached to its tipcone.
day for cleanliness. Particular attention
amounts of liquid in micro volumes.
should be paid to the tip cone(s).
These pipettes operate on the air Allow liquids, tips, and pipettes to
This pipettor has been designed for easy in-
displacement principle and use disposable equilibrate to the ambient temperature.
tips of any standard make. house service. However, we also provide
complete repair and calibration service.
All pipettes are equipped with a built-in tip Pre-rinsing of tip 5 times with the liquid Please return your pipettor to your local
ejector. distributor for repair or calibration. Before
to be dispensed is recommended. This is
They cover the volume range of 0.1 µl to important especially when dispensing returning, please make sure that it is free
5000 µl. 1. Press the operating button to the first stop. liquids which have a viscosity and from any contamination.

Variable volume pipettes 2. Dip the tip attached to the pipette into the solution to a density different from water. Check the performance of your pipettor
depth of about 1cm, and slowly release the operating regularly e.g every 3 months and after every
Volume Range Increment button. Wait for a while, then withdraw it from the
in-house service or maintenance.
0.1-2.5 µl 0.01 µl liquid touching it against the edge of the reservoir to Pre-rinse the tip several times before use
remove excess liquid adhering to the outer surface of
0.5-10 µl 0.1 µl the tip.
when pipetting liquids at temperature
5-50 µl 0.5 µl different from ambient.
3. Dispense the liquid into the receiving vessel by gently
50-200 µl 1.0 µl pressing the operating button to the first stop. After a Cleaning Your Pipettor
200-1000 µl 5.0 µl second, press the operating button to the second stop.
This will empty the tip completely. Remove the pipette While pipetting, the pipettor should be
10-100 µl 0.5 µl from the vessel sliding it up the wall of the vessel.
To clean your pipettor, use ethanol and a soft
vertically straight and tip should be
100-1000 µl 5.0 µl cloth or lint-free tissue. It is recommended to
4. Release the operating button to the ready position. dipped only a few millimeters into the clean the tip cone regularly.
2-20 µl 0.1 µl
liquid.
20-200 µl 1.0 µl
500-5000 µl 50.0 µl B. Reverse Pipetting
Always control the push button In-House Maintenance
Fix volume pipettes movements with the thumb for
Volume consistency. 1. Hold down the
2.5 µl 200 µl tip ejector.
5 µl 250 µl
Wipe the tip only if there is liquid on the 2. Place the tooth
10 µl 500 µl
outside of the tip, being careful to avoid of the opening
20 µl 1000 µl
touching the tip’s orifice. tool between the
25 µl 2000 µl
tip ejector and
50 µl 2500 µl the tip ejector
100 µl 5000 µl Don't keep pipette in your hand while collar to release
Note: Special volume pipettes other than specified above 1. Press the operating button to the second stop. not working, to avoid transferring of the locking
may be supplied on demand. body heat. mechanism.

1 0 0 0
2. Dip the tip attached to the pipette into the solution to a
depth of about 1cm, and slowly release the operating
Multi(8) Channel pipettes button. This action will fill the tip with a volume that 3. C a r e f u l l y
Volume Range Increment is larger than the set volume. Wait 1-2 seconds and Use the correct pipette tip designed for release the tip
withdraw the tip from the liquid, touching it against ejector and
0.5-10 µl 0.1 µl the edge of the reservoir to remove excess liquid.
use with the particular pipette.
5-50 µl 0.5 µl remove the
3. Dispense the liquid into the receiving vessel by ejector collar.
10-100 µl 0.5 µl pressing the operating button gently and steadily to Select the correct pipetting technique
30-300 µl 1.0 µl the first stop. This volume is equal to the set volume. 4. P l a c e t h e
Hold button in this position. Some liquid will remain (e.g. Reverse, forward etc.) depending
in the tip, which should not be dispensed. on the nature of the liquid. wrench end of
Multi(12) Channel pipettes 4. The liquid remaining in the tip can be dispensed back
the opening
Volume Range Increment into the original solution by pressing the button to the tool over the tip
0.5-10 µl 0.1 µl second stop or disposed together with the tip. Using excessive force to turn the push cone and turn it
5-50 µl 0.5 µl 5. Release the operating button to the ready position. button outside the range specified for it anti-clockwise.
10-100 µl 0.5 µl Note : Reverse pipetting technique is recommended for viscous may jam the mechanism and damage
30-300 µl 1.0 µl
solutions, solutions having tendency to foam or for dispensing the pipettor.
very small volumes.