These pipettes are available as fixed or A. Forward Pipetting Aspirate liquid into the pipette only To maintain the best results from your
variable, general purpose micropipettes for pipettor, each unit should be checked every
sampling and dispensing accurate when a tip is attached to its tipcone.
day for cleanliness. Particular attention
amounts of liquid in micro volumes.
should be paid to the tip cone(s).
These pipettes operate on the air Allow liquids, tips, and pipettes to
This pipettor has been designed for easy in-
displacement principle and use disposable equilibrate to the ambient temperature.
tips of any standard make. house service. However, we also provide
complete repair and calibration service.
All pipettes are equipped with a built-in tip Pre-rinsing of tip 5 times with the liquid Please return your pipettor to your local
ejector. distributor for repair or calibration. Before
to be dispensed is recommended. This is
They cover the volume range of 0.1 µl to important especially when dispensing returning, please make sure that it is free
5000 µl. 1. Press the operating button to the first stop. liquids which have a viscosity and from any contamination.
Variable volume pipettes 2. Dip the tip attached to the pipette into the solution to a density different from water. Check the performance of your pipettor
depth of about 1cm, and slowly release the operating regularly e.g every 3 months and after every
Volume Range Increment button. Wait for a while, then withdraw it from the
in-house service or maintenance.
0.1-2.5 µl 0.01 µl liquid touching it against the edge of the reservoir to Pre-rinse the tip several times before use
remove excess liquid adhering to the outer surface of
0.5-10 µl 0.1 µl the tip.
when pipetting liquids at temperature
5-50 µl 0.5 µl different from ambient.
3. Dispense the liquid into the receiving vessel by gently
50-200 µl 1.0 µl pressing the operating button to the first stop. After a Cleaning Your Pipettor
200-1000 µl 5.0 µl second, press the operating button to the second stop.
This will empty the tip completely. Remove the pipette While pipetting, the pipettor should be
10-100 µl 0.5 µl from the vessel sliding it up the wall of the vessel.
To clean your pipettor, use ethanol and a soft
vertically straight and tip should be
100-1000 µl 5.0 µl cloth or lint-free tissue. It is recommended to
4. Release the operating button to the ready position. dipped only a few millimeters into the clean the tip cone regularly.
2-20 µl 0.1 µl
liquid.
20-200 µl 1.0 µl
500-5000 µl 50.0 µl B. Reverse Pipetting
Always control the push button In-House Maintenance
Fix volume pipettes movements with the thumb for
Volume consistency. 1. Hold down the
2.5 µl 200 µl tip ejector.
5 µl 250 µl
Wipe the tip only if there is liquid on the 2. Place the tooth
10 µl 500 µl
outside of the tip, being careful to avoid of the opening
20 µl 1000 µl
touching the tip’s orifice. tool between the
25 µl 2000 µl
tip ejector and
50 µl 2500 µl the tip ejector
100 µl 5000 µl Don't keep pipette in your hand while collar to release
Note: Special volume pipettes other than specified above 1. Press the operating button to the second stop. not working, to avoid transferring of the locking
may be supplied on demand. body heat. mechanism.
1 0 0 0
2. Dip the tip attached to the pipette into the solution to a
depth of about 1cm, and slowly release the operating
Multi(8) Channel pipettes button. This action will fill the tip with a volume that 3. C a r e f u l l y
Volume Range Increment is larger than the set volume. Wait 1-2 seconds and Use the correct pipette tip designed for release the tip
withdraw the tip from the liquid, touching it against ejector and
0.5-10 µl 0.1 µl the edge of the reservoir to remove excess liquid.
use with the particular pipette.
5-50 µl 0.5 µl remove the
3. Dispense the liquid into the receiving vessel by ejector collar.
10-100 µl 0.5 µl pressing the operating button gently and steadily to Select the correct pipetting technique
30-300 µl 1.0 µl the first stop. This volume is equal to the set volume. 4. P l a c e t h e
Hold button in this position. Some liquid will remain (e.g. Reverse, forward etc.) depending
in the tip, which should not be dispensed. on the nature of the liquid. wrench end of
Multi(12) Channel pipettes 4. The liquid remaining in the tip can be dispensed back
the opening
Volume Range Increment into the original solution by pressing the button to the tool over the tip
0.5-10 µl 0.1 µl second stop or disposed together with the tip. Using excessive force to turn the push cone and turn it
5-50 µl 0.5 µl 5. Release the operating button to the ready position. button outside the range specified for it anti-clockwise.
10-100 µl 0.5 µl Note : Reverse pipetting technique is recommended for viscous may jam the mechanism and damage
30-300 µl 1.0 µl
solutions, solutions having tendency to foam or for dispensing the pipettor.
very small volumes.