You are on page 1of 5

Refrigerator

ZL-2

TIANJIN GUOMING MEDICINAL EQUIPMENT CO.,LTD.


Introduction

 Brief introduction

  ZL-2 consists of main box, compressor, condenser, cooling bath (evaporator)


and fan etc.
  The Zl-2 refrigerator and the HW-2 constant temperature bath constitute a
constant low temperature bath, which is high efficiency. It cools water from 30
℃ to 10℃ in 45 minutes.
  ZL-2 main box consists of bottom, foot, box cover, front panel, and rear panel
etc.

Power on switch Rear panel

AC input socket Name plate

Water outlet Water inlet

  Front panel includes:


● Product model
● Manufacturing
  Rear panel includes:
● Power on switch
● AC input socket
● Water inlet
● Water outlet
● Name plate

 Function description

● Power on switch

  Power on switch is the main power switch. In ON position, unit is in working


status, inversely, unit stops working.
  Before connecting to AC power, please ensure it’s in OFF position!

● AC input socket

ZL-2 Refrigerator Manual                                       1


  AC input socket is connected to the power supply with power cord offered in
this unit.
  Please make sure AC power supply is connected properly!

● Water inlet

  The cooling fluids flow in the ZL-2 through this water inlet.

● Water outlet

  The cooling fluids flow out the ZL-2 through this water outlet.

 Specification                           

  Continuous work:                      >24hr


  Power:                 220V/60Hz/600W
  Dimension:                480×380×280MM3
  Weight:                           27KG

 Environment requirement                      

  Usage temperature                  +18℃~+28℃


  Store temperature                  -10℃~+50℃
  Humidity                        20%~80%

 Power supply                           

  Range of input power supply      220V±10%VAC 60Hz


  Max input power                        600W

Installation

 Unpacking and Checking              

  Open the package, take out manual, instrument, and accessories. And check items
according packing list. Any mismatch/damage happened, please contact marketing
department immediately.

 Unit settle

  It must be placed on stable work station/table and put damping pad between table
and unit.
  Minimum space required around test is 0.5m, to ensure heat escaping. Clean
Instrument periodically to prevent it from abnormal cooling.

ZL-2 Refrigerator Manual                                       2


 Pipe line connect                        

  Connect the ZL-2 and the HW-2 with attached plastic pipe. The inlet and the
outlet are on the rear panel. The inlet of the ZL-2 is connected to the outlet of
the HW-2, while the outlet of the ZL- 2 is connected to the inlet of the HW-2.
  The pipe suit with maximum temperature thimble is the shortest practical
length.

  Soak the pipe in hot water to soft it before connecting the pipe.

 Connect to the power supply                    

  Link the socket & AC output socket of the HW-2 with attached power cord.

 Cooling fluids

  Suggest using distillation water over +8℃.


  Suggest using mixture of tap water and laboratory grade ethylene glycol below
+8℃.
  Do not use automobile anti-freeze.

 Cooling bath

  The blocked cooling system consists of cooling bath inside ZL-2 and constant
temperature bath. It gets up preset temperature in 30 minutes when distillation
water is used.

Operation                              

Not open when room temperature below +5℃ .


Not open without cooling fluids or no circulation.
Not open when fluids temperature below +5℃ or over40℃.

ZL-2 Refrigerator Manual                                       3


 Power on

  Turn the power switch to I position, the power indicator lights, the bath runs.
After the fluids without bubble flow smoothly, Turn the power switch of refrigerator
to I position, the power indicator lights, the refrigerator runs.

 Power off                             

  First turn the power’s switch of the refrigerator to O position, the power
indicator off, turn it off. Turn the power switch of the bath to O position, the
power indicator off, the bath off.
  When the refrigerator shut off, allow about five minutes before restarting the
unit. Otherwise it will be damaged.

Maintenance                             

  Must clean box and bath periodically, do not use steel brush, it result in damage
to the equipment.
  Clean it at least on a month.

ZL-2 Refrigerator Manual                                       4

You might also like