El Operating Manual
El Operating Manual
NOTE: This instrument has been tested and found to comply with the
limits for a Class A digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference when the instrument is operated in a commercial
environment. This instrument generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
reference manual, may cause harmful interference to radio
communications.
Operation of this instrument in a residential area is likely to cause
harmful interference in which case the user will be required to correct
the interference at his own expense.
01 ■ SE-1520 d
Introduction
e SE-1520 ■ 02
Notation
The following symbols are used in this manual and labels on this instrument
to call your attention to specific items.
03 ■ SE-1520 d
Caution Labels
3 Probe pin
Do not touch the probe pin with bare hands. When cleaning the probe pin,
wear protective gloves and use swabs to prevent exposure to pathogenic
microbes.
d SE-1520 ■ 04
Contents
Electrolyte measuring system, SE-1520 Operating Manual
Premise………………………………………………………………………………… 01
Introduction………………………………………………………………………… 02
Notation ……………………………………………………………………………… 03
Contents………………………………………………………………………………… 04
d SE-1520 ■ 06
07 ■ SE-1520 d
Chapter 1
Introduction
The SPOTCHEM EL SE-1520 is an electrolyte measurement
system using a exclusive E-Plate.
Chapter 1 contains functions and measurement principles of SE-
1520.
1-1 Outline
1-1-1 Features
1-1-2 Measurement Types
1-1-3 Measurement Principles
1-1-4 Specifications
1-4 Installation
1-4-1 Cautions
1-4-2 Setting Up
1-4-3 First Operation after Installation
1-4-4 Precautions in Instrument Relocation
d SE-1520 ■ 1-01
1-1 Outline Chapter 1 Introduction
1-1-1 Features
1-02 ■ SE-1520
1-1 SP-4430の概要 Chapter 1 Introduction
Er Ej Es
The identical structure of ion selective electrode is used for both reference electrode and
measurement electrode. When reference solution is measured with the reference electrode,
electric potential of reference electrode Er is obtained. Also, when the desired ion in
measurement solution is measured with the measurement electrode, electric potential of
measurement Es is obtained.
By setting up a solution junction part between the reference and measurement solutions,
measurement electric potential E is obtained between the reference electrode and
measurement electrode. There is a relation of Nernst equation between generated electric
potential and ion activity (ion concentration). Ion concentration is obtained by measuring
potential difference E.
E = Es − Er + Ej
2.303RT
E = (
log (as) − log (ar))+ Ej
ZF
2.303RT
E = log (as) + Eo
ZF
[note]
Es:Electric potential by ion activity in sample
Er :Electric potential by ion activity in reference solution constant)
Ej :Electric potential of solution junction (constant)
2.303RT
Eo:Ej − log (ar) (constant)
ZF
as :Ion activity in sample
ar :Ion activity in reference solution constant)
SE-1520 ■ 1-03
Chapter 1 Introduction
1-1-4 Specifications
1-04 ■ SE-1520 e
1-2 Shipping Carton Chapter 1 Introduction
Open the package and confirm that all items are included.
1・Analyzer(SE-1520)
Analyzer
No. Item Description Amount
1 Analyzer SE-1520(SPOTCHEM EL) 1unit
2 AC adapter 1unit
3 Power cord 1unit
4 Accessory box 1box
5 Operating manual This booklet 1copy
6 Pipetting tip 100pcs. 1box
Accessory box
No. Item Description Amount
1 Check plate 1sheet 1unit
2 Cotton buds 5pcs. 1set
3 Plate tray 1unit
4 Thermal printer paper 1roll
b SE-1520 ■ 1-05
1-3 Parts Description and Function Chapter 1 Introduction
3・Built-in printer
1・Display
1 Display Displays information, such as operating state of the unit and error messages.
2 Operator panel Start or Stop measurement, and enter various kinds of values.
3 Built-in printer Thermal line printer for printing out measurement results and parameter
settings.
7 Magnetic-card reader Insert magnetic card, if a new item is added or lot card is loaded
1-06 ■ SE-1520 b
1-2
1-3 梱包品の確認
各部の名称とはたらき Chapter 1 Introduction
Item Functions
ー/. (hyphen/period) Select items, move the cursor, switch the pages on
the display, and enter minus sign and decimal point.
SE-1520 ■ 1-07
1-4 装置の設置 Chapter 1 Introduction
1・Power switch
2・Terminal
3・RS-232C
4・CONT.
5・B.C.R.
6・Paper cover
1-08 ■ SE-1520
1-2
1-3 梱包品の確認
各部の名称とはたらき Chapter 1 Introduction
3・Eject button
4・Reflection board
4 Reflection board Informs the system of the condition of pipette operation in the event of sampling
SE-1520 ■ 1-09
1-4 装置の設置 Chapter 1 Introduction
1・Reference 2・Sample
solution button button
3・Ejection
button
9・Reference
solution
discharge
lever
10・Reference
solution 5・Nozzle cap
suction hook
lever
6・Reference
solution
cylinder 7・Reference
solution 4・Reflection
replacement mark board
11・Nozzle O-ring
8・ Nozzle cap
4 Reflection board Informs the system of the condition of pipette operation in the event of sampling
5 Nozzle cap hook Removed caps can be inserted and fixed here.
8 Nozzle cap Protection of reference solation nozzle tip and prevention of evaporation when not used.
9 Reference solution discharge lever Push down the lever when reference solution in the cylinder is discharged.
10 Reference solution suction lever Push down the lever when reference solution in the cylinder is drawn up.
1-10 ■ SE-1520
1-4 Installation Chapter 1 Introduction
1-4-1 Cautions
Allow a space of
10cm or more AC100-120V 50 / 60Hz
Before setting up the
analyzer, read the
following notes and
always take proper
AC220-240V
safety precautions.
● To avoid electric shock and/or fire, ● Connect the power plug directly to
use the attached power cord only an outlet, not via an extension cord
to connect with a power outlet. or power tap. The power supply for
the analyzer is 40 VA.
d SE-1520 ■ 1-11
1-3 各部の名称とはたらき Chapter 1 Introduction
Temperatures Humidity
10 ∼ 30℃ 20 ∼ 80%
1-12 ■ SE-1520 b
1-4 装置の設置 Chapter 1 Introduction
1-4-2 Setting Up
Required items Analyzer, AC adapter, power cord, exclusive cable (for communication external )
b SE-1520 ■ 1-13
1-3 各部の名称とはたらき Chapter 1 Introduction
1-3-3 装置背面の名称とはたらき
1-14 ■ SE-1520 d
1-4 装置の設置 Chapter 1 Introduction
b SE-1520 ■ 1-15
1-4 装置の設置 Chapter 1 Introduction
●Turn off the power and disconnect the power cable before
transporting the system. Failure to do so may cause system
malfunction.
1-16 ■ SE-1520 d
Chapter 2
Measurement
The SE-1520 is capable of sample measurement.
Chapter 2 explains the procedures and outline of each
measurement.
2-1 Outline
2-1-1 Operating Procedures
2-1-2 Exclusive E-Plate
2-1-3 Measurement
2-1-4 Calibration
2-2 Cautions
2-2-1 Cautions on Operation
2-2-2 Handling Samples
2-2-3 Handling Exclusive E-Plates
2-2-4 Handling Magnetic Card
2-2-5 Handling Twin Pipette and Twin Pipette II
2-2-6 Handling Reference Solution
2-3 Preparation
2-3-1 Preparation
2-3-2 Startup
2-3-3 Checks before Measurement
2-3-4 Preparation of Samples
2-3-5 Suction of Reference Solution and Samples
2-4 Measurement
2-4-1 Normal Measurement
2-5 Calibration
2-5-1 Overview
2-5-2 Magnetic Card Calibration
b SE-1520 ■ 2-01
2-1 Outline Chapter 2 Measurement
Press 【START】
Press 【STOP】.
or 【1】.
2-02 ■ SE-1520
1-2 梱包品の確認 Chapter 2 Measurement
2-1-3 Measurement
With this analyzer, normal measurement and check
measurement are available.
Normal measurement is used to measure the exclusive plate
and check measurement is the mode to measure the attached
check plate.
■ Normal Measurement
Press [1] key on the Main Menu to enter the Normal Measurement Mode
for sample measurement.
Measurement results are identified by the measurement number indicated
as “No. XXXX”, which is updated automatically and given sequent
number until the power is turned off.
■ Check Measurement
The attached check plate is measured to confirm that the measurement
function of the system is working normally. Run this measurement when a
false measurement result is obtained.
SE-1520 ■ 2-03
2-1 Measurementの Chapter 2 Measurement
2-1-4 Calibration
2-04 ■ SE-1520
2-2 Cautions Chapter 2 Measurement
Temperatures between
The optimal 20℃ and 25℃
environment Humidity between
30% and 60%
● Before turning on the power, be ● The temperature controlling function ● If you find any abnormal
sure to read 1-4-1 “Cautions” works to obtain accurate measurement operation, odor or smoke, turn off
again and operate the results under temperatures between the power immediately and
analyzer in a proper 10℃ and 30℃ . However, it is unplug the analyzer. There may
environment. recommended to perform measurement be danger of damage to the
in the optimal environment at analyzer, injury or fire.
temperatures between 20℃ and 25℃
with humidity between 30% and 60% for
more accurate measurement results.
● In case of malfunction, contact ● Do NOT place bottles containing ● Do NOT move the analyzer
your distributor for repair. Do samples on the analyzer. They during measurement. Vibration to
NOT attempt to repair or remodel may tip over and leak into the the analyzer during measurement
the analyzer yourself. It may inside of the analyzer resulting in may cause malfunction and
result in damaging the analyzer damage. incorrect measurement results.
or injury.
b SE-1520 ■ 2-05
2-1 Measurementの Chapter 2 Measurement
Back to room
temperature
● Whole blood, serum, plasma also ● Prepare fresh blood (within 1 hour ● Stir sample without foaming then
urine are used as samples with after collection) and fresh urine. distribute it before starting the
this analyzer. Be aware that they Allow refrigerated blood samples measurement.
may be contaminated with back to the room
pathogenic microbes that may temperature before measurement.
cause infectious disease. Use
protective gloves or take other
preventive measures with
extreme caution when handling
these samples. Incorrect or
imprecise procedures may result
in exposure of the operator and
other persons nearby to
pathogenic microbes
Back to room
temperature
● Use exclusive E-Plate. Read the ● Check before use. ● Prepare exclusive E-Plate before
notes attached to each E-Plate Do NOT use exclusive E-Plate measurement.
and use it before the expiry date. that are cracked or leaked even Necessary number of exclusive
prior to the expiry date. E-Plate should be taken out of
the refrigerator to allow them to
reach room temperature before
measurement.
2-06 ■ SE-1520
2-1 Measurementの Chapter 2 Measurement
● Do NOT place magnetic cards ● Do NOT scratch the magnetic ● Do NOT use the magnetic card
near magnetic objects (magnets , surface (stripe). attached to the exclusive E-
TV sets. etc) otherwise. The information stored on the Plate for other devices or
The information stored on the card may become unreadable machines except for SE-1520.
card may become unreadable with the magnetic card reader. The card may get stuck.
with the magnetic card reader.
2-1 Measurementの
2-2-5 Handling Twin Pipette and Twin Pipette II
Do NOT use the twin pipette (for SE-1510) with reflection board on the both sides of it.
● Use either the exclusive twin ● The nozzle tip of twin pipette II
pipette or the twin pipette II. for the reference solution is
Do NOT use pipetting tips other narrow and sharp: it can be
than the exclusive one. damaged easily and also may
By using a soft cloth or swab, wipe cause injury by scratching or Do NOT place things in the cylinder
off the dust or dirt from the Tip sticking. window of the pipette or push the
fitting part (e.g. O-ring) of twin Be careful in handling and do not cylinder window with a ball-point
pipette or twin pipette II, if handle it roughly.
pen or other objects. Do NOT drop
necessary.
the pipette. Cylinder breakage
may cause injury.
b SE-1520 ■ 2-07
2-1 Measurementの Chapter 2 Measurement
2-08 ■ SE-1520
2-1 Measurementの
2-2-6 Handling Reference Solution
1. Handling Reference Solution
Back to room
temperature
● Store reference solution in a ● Remove the solution from the ● Always close the lid after use.
refrigerator (2-8℃). refrigerator and allow it to reach If the lid is left open, the reference
Use it within three months after room temperature before solution may become concentrated
opening. measurement. and lead to the incorrect
measurement results.
②
③
● Exchange of reference solution ③ Draw the reference solution ● If it is left open, it will evaporate
① Soak the tip of the nozzle in the again. or become contaminated and
reference solution container, and After exchanging the reference correct measurement results will
push up the reference solution solutions, lock the reference not be obtained.
suction lever. solution suction lever at the ● After drawing, close the
② After drawing of the reference lowest position and the container lid firmly. If it is left
solution, push the reference reference solution discharge open, it will allow evaporation or
solution suction lever down to the lever at the highest position. contamination of the solution,
bottom and discharge all of it. Due to quantity of use, pull the which may lead to incorrect
reference solution suction lever measurement results.
up. When the lever at the top, Put the nozzle cap back tightly.
can measure about fifty times.
b SE-1520 ■ 2-09
2-3 Preparation Chapter 2 Measurement
2-3-1 Preparation
Protective glove ○ ×
Twin pipette
or twin pipette Ⅱ ○ ×
Exclusive E-Plate ○ ×
Magnetic card × ○
2-10 ■ SE-1520
2-2 測定時の注意 Chapter 2 Measurement
2-3-2 Startup
SE-1520 ■ 2-11
2-3 測定の準備 Chapter 2 Measurement
2-12 ■ SE-1520 b
2-2 測定時の注意 Chapter 2 Measurement
2-3-2 装置の始動
b SE-1520 ■ 2-13
2-2 測定時の注意 Chapter 2 Measurement
1. Prepare samples
● Prepare samples to measure.
Allow refrigerated samples to
reach room temperature before
use.
2. Stir samples
● Stir samples by inverting or
other mixing methods. without
making foam.
2-14 ■ SE-1520 b
2-3 測定の準備 Chapter 2 Measurement
①Make sure the pipette tip is firmly attached to the twin pipette.
②Exchange the pipette tip in each measurement. Do NOT use it again.
After suction of the reference solution is completed, close the lid of the reference
solution container tightly. If it is left open, solution may evaporate or
contaminated and may lead to incorrect measurement.
b SE-1520 ■ 2-15
2-3 測定の準備 Chapter 2 Measurement
①Make sure that the pipetting tip is firmly attached to the twin pipette.
②Exchange the pipetting tip for each measurement. Do NOT use it again.
2-16 ■ SE-1520 b
2-3 測定の準備 Chapter 2 Measurement
d SE-1520 ■ 2-17
2-4 Measurement Chapter 2 Measurement
1. Prepare samples
● Prepare samples (see 2-3-4
“Preparation of Samples”).
2-18 ■ SE-1520 b
2-2 測定時の注意 Chapter 2 Measurement
2-3-5 検体の分注操作
SE-1520 ■ 2-19
2-4 Measurement Chapter 2 Measurement
If not enter measurement No. and ID, to“7. Set the E-Plate on the table”.
Normally measurement No. is allocated automatically. ID is not entered in the such way.
If the lot number is not same, a correct data can not be obtained.
In this case, calibrate by a magnetic card.
2-20 ■ SE-1520 b
Chapter 2 Measurement
9. Start pipetting
● Make sure that a message “Apply
sample” is displayed.
REF. SAMPLE.
d SE-1520 ■ 2-21
2-4 Measurement Chapter 2 Measurement
2-4-1 通常測定(連携測定)
10. Measure
● When the twin pipette is pulled
out, the plate transfer part
automatically moves to the
measurement part and electric
potential is measured. About one
minute later, the measurement
result is printed. During this
operation, the upper right corner
of the display shows the
remaining time to the end of the
measurement.
2-22 ■ SE-1520 b
2-5 Calibration Chapter 2 Measurement
2-5-1 Overview
IMPORTANT If the measurement is performed when the magnetic card calibration is not
performed correctly, errors occur and measurement may not start.
In this case, be sure to perform “ Magnetic Card Calibration.”
Pr i n t i ng…
● Press [2].
Lot information of all items
recorded in the system is
printed.
When the printing is completed,
you return to the Calibration
Menu.
SE-1520 ■ 2-23
2-4 Measurement Chapter 2 Measurement
3. End calibration
● When calibration is not
1 . Me a s u r e 2 . S u bme n u
performed consecutively, press
[STOP] key to return to the
3. Ca l . ( 1/1 )
MAIN MENU
2-24 ■ SE-1520 b
2-6 Measurement Result Chapter 2 Measurement
(1)Normal printing
1
2
6
SE-1520 ■ 2-25
2-6 測定結果のみかた Chapter 2 Measurement
2-6-1 通常測定結果の印字
1
2
3
Basically the survey mode printing is performed in the same way as normal
printing. However, the following is printed in addition to the measurement
value of each channel.
1・ Printing of measurement values (concentration value, various marks) is
the same as in normal printing.
2・ Concentration values (before correction by coefficient
of correlation correction) are printed.
3・ Temperature conversion coefficient: Measurement value = Kt*
Measurement value befor conversion.
4・ Unit conversion coefficient: Measurement value =Ku* measurement
value befor conversion.
5・ Correlation correction coefficient: Measurement value= Acor×
Measurement value before correction + Bcor
2-26 ■ SE-1520
Chapter 3
Sub Menu
The SE-1520 has a main menu and five sub menus.
Sub menus are classified by setting items.
Chapter 3 explains how to perform Sub Menu settings.
3-1 Overview
3-1-1 Contents of Each Menu
b SE-1520 ■ 3-01
3-1 Overview Chapter 3 Sub Menu
1.Measure 2.Submenu
3.Cal (1/1)
START 7 P
QRS 8 TUV 9 W
XYZ
-
MENU 0 ?# . ENTER
START 7 P
QRS 8 TUV 9 W
XYZ MENU 0 ?#
-
. ENTER
START 7 P
QRS 8 TUV 9 W
XYZ 1.Print 2.Input
FEED 1 2 ABC 3 DEF
3.Initialize (1/1)
-
STOP 4 GHI 5 JKL 6 MNO
MENU 0 ?# . ENTER
-
MENU 0 ?# . ENTER
START 7 P
QRS 8 TUV 9 W
XYZ
MENU 0 ?#
-
. ENTER
FEED 1 2 ABC 3 DEF
-
MENU 0 ?# . ENTER
4.Mode 5.Clock START 7 P
QRS 8 TUV 9 W
XYZ 1.Survey
(2/2) (1/1)
STOP 4 GHI 5 JKL 6 MNO
-
MENU 0 ?# . ENTER
START 7 P
QRS 8 TUV 9 W
XYZ Date 〈00ー02ー15〉YMD
Time 〈11:45:00〉
STOP 4 GHI 5 JKL 6 MNO
-
MENU 0 ?# . ENTER
3-02 ■ SE-1520
2-2 測定時の注意 Chapter 3 Sub Menu
2-2-1 操作上の注意
Parameter Menu
Item Description Reference Item
Maintenance Menu
Item Description Reference Item
2. T a b l e
Clean the table. 3-4-2
Clean the table
Mode Menu
Item Description Reference Item
SE-1520 ■ 3-03
3-2 Measurement Results Menu Chapter 3 Sub Menu
If [STOP] key is pressed Measurement results (up to 50 samples) stored in memory can
during operation, the
previous screen is restored. be printed out. The printing type can be chosen from the
following.
▲
3-04 ■ SE-1520
2-2 測定時の注意 Chapter 3 Sub Menu
2-2-1 操作上の注意
to avoid inconsistency.
● Press [ENTER] key.
The 2-digit number of the I D< >
"year" section of the date The Enter ID screen is
indicates the last 2 digits of
year and is interpreted as displayed.
follows:
00 - 89→2000 - 2089
90∼99→1990 - 1999
● Enter an ID search pattern for
I D <99ー12ー25ーA B C D>
the sample(s) to print by using
numbers, alphabets and signs
up to 13 characters in all.
▲
4. End Printing
● Press [STOP] key to return to
1 . Me a s u r e 2 . S u bme n u
the MAIN MENU.
3. Ca l . ( 1/1 )
SE-1520 ■ 3-05
3-3 測定結果メニュー Chapter 3 Sub Menu
If [STOP] key is pressed Measurement results (up to 50 samples) stored in memory can
during operation, the
previous screen is be transmitted.
restored. The transmitting type can be chosen from the following.
▲
3-06 ■ SE-1520
2-2 測定時の注意 Chapter 3 Sub Menu
2-2-1 操作上の注意
4. End transmission
● Press [STOP] key to return to
1 . Me a s u r e 2 . S u bme n u
the MAIN MENU.
3. Ca l . ( 1/1 )
SE-1520 ■ 3-07
3-3 測定結果メニュー Chapter 3 Sub Menu
1 . R e s u l t s 2 . P a r am
● Select [START] or [STOP] key.
When deleting→
3 . M a i n t e n a n c e ( 1/2 )
If [START] key is pressed, the
measurement results are
deleted and the MEASUREMENT
RESULTS MENU is restored.
When not deleting→
If [STOP] key is pressed,
deletion is canceled and the
MEASUREMENT RESULTS
MENU is restored.
3. End Deletion
● Press [STOP] key to return to
1 . Me a s u r e 2 . S u bme n u
the MAIN MENU.
3. Ca l . ( 1/1 )
3-08 ■ SE-1520
2-2 測定時の注意 Chapter 3 Sub Menu
3-2-4 Wildcards
■ wildcards
Wildcards can be used for ID search. Wildcards are two special
characters, "?" and "*", indicating a single or any number of
characters.
SE-1520 ■ 3-09
3-3 Parameter Menu Chapter 3 Sub Menu
1. P r i n t 2. I npu t
● Press [2] key.
3. I n i t i a l i z e ( 1/1 )
The Parameter Menu is displayed.
3. End Printing
● Press [STOP] key to return to
the MAIN MENU.
1 . Me a s u r e 2 . S u bme n u
3. Ca l . ( 1/1 )
3-10 ■ SE-1520
2-2 測定時の注意 Chapter 3 Sub Menu
3-3-2 パラメータの入力
1
2
3
4
5
6
7
8
9
10
11
12
13
14
SE-1520 ■ 3-11
3-4 メンテナンス Chapter 3 Sub Menu
change the previous This function allows the results obtained by this device to match the results by
setting, press [ENTER] key another measurement method (reference method). Apply the regression equation Y
to go on to next "[ ]" or = aX + b, (X : the result measured by the SE-1520, Y : the result obtained by the
"< >".
reference method.) Enter coefficient values for a and b. For obtaining the
coefficients of correlation correction for a and b, contact your distributor.
■Normal value range
Data out of the normal value range are marked with ▲ or ▼, when printed.
■Correction Temperature
Measurement is always performed at 32℃ as the measurement
temperature in this analyzer. Converts the measurement results to the
ones measured at 25℃, 30℃, or 32℃ and outputs the converted results.
However only the enzyme items can be converted.
1 . R e s u l t s 2 . P a r am
● Press [2] key. 3 . M a i n t e n a n c e ( 1/2 )
The Parameter Menu is displayed.
1. P r i n t 2. I npu t
3. I n i t i a l i z e ( 1/1 )
● Press [2] key and the Enter
Password screen is displayed. P a s s w o r d <** >
Enter password "99". C a n c e l( S T O P )
"**" is displayed
3-12 ■ SE-1520
3-3 パラメータメニュー Chapter 3 Sub Menu
3-3-1 パラメータ設定内容の印字
●
decimal points by using the "- S e r um a〈 1 . 000〉
/." key. by using the numeric keys.
To enter a minus sign, press
S-02 b〈 0 . 000〉
Any number from 0 to 10000
[-/.] before pressing any
numeric keys. A decimal can be entered.
point can be entered. Press
[-/.] key after any numeric key S e r um a〈 1 . 000〉
is pressed. ● Press [ENTER] key and the
S-02 b〈 0 . 000〉
cursor moves to the entry
e.g.) To enter "-12 position for "B".
Press [-/.] [1] [2] Enter a coefficient factor "B" and
e.g.) To enter "3.4"
Press [3] [-/.] [4] press [ENTER].
e.g.) To enter ".5" Any number from -10000 to
Press [0] [-/.] [5]. If only [-/.]
[5] is pressed, -5 is entered. 10000 can be entered. Press
[ENTER] key and the next
▲
SE-1520 ■ 3-13
3-3 パラメータメニュー Chapter 3 Sub Menu
7. End setting
● If the setting is finished, press
1 . Me a s u r e 2 . S u bme n u
[STOP] key three times to
3. Ca l . ( 1/1 )
return to the MAIN MENU.
3-14 ■ SE-1520 d
3-3 パラメータメニュー Chapter 3 Sub Menu
2. Initialize parameters
▲
3. End Initialization
● When the initialization is
1 . Me a s u r e 2 . S u bme n u
finished, press [STOP] key to
3. Ca l . ( 1/1 )
return to the MAIN MENU.
▲
Present parameter
■Factory set parameter values
settings can be checked, if
it is necessary see 3-3-1 The following table shows the conditions of the factory
“Printing Parameters" for
details. setting.
Refer to this table to set each condition.
Correction 「25℃」「30℃」
temperature 「−−−−」
「−−−−」
(No correction)
SE-1520 ■ 3-15
3-4 Maintenance Menu Chapter 3 Sub Menu
3-16 ■ SE-1520
3-5 Mode Menu Chapter 3 Sub Menu
The Survey Mode is Switch to the Survey Mode. When measurement is done in the
canceled automatically if
the power is turned off. Survey Mode, raw data can be printed out besides ordinary
measurement results (see 2-6 "Measurement Result" for
details of printouts).
4 . Mo d e 5. C l oc k
● Press [MENU] key or [hyphen ( 2/2 )
(−)] key.
The SUB MENU 2/2 is displayed.
1. Su r v e y
● Press [4] key.
( 1/1 )
The Mode Menu is displayed.
SE-1520 ■ 3-17
3-6 Built-in Clock Adjustment Chapter 3 Sub Menu
3-5-2 コントロール測定モード
Set the date and time of the built-in clock. Once you set the
date and time, resetting is not necessary, however, some
adjustment may be required in a long period of use.
4 . Mo d e 5. C l oc k
● Press [MENU] key or [hyphen
( 2/2 )
(−)] key. The SUB MENU 2/2
is displayed.
P a s s w o r d <** >
● Press the [5] key.
C a n c e l( S T O P )
“Password Entering screen” is
displayed. Enter passward "99",
"**" is displayed.
Da t e <00ー02ー15>Y MD
● Press [ENTER] key.
T i me <1 1:4 5:00>
The Clock Adjustment screen
is displayed.
3. End Setting
● Press [STOP] key to return to
1 . Me a s u r e 2 . S u bme n u
the MAIN MENU.
3. Ca l . ( 1/1 )
3-18 ■ SE-1520
Chapter 4
Maintenance
After measurements are finished for the day, turn OFF
the power.
Chapter 4 explains the maintenance of the system.
4-1 Outline
4-1-1 Frequency of Maintenance
b SE-1520 ■ 4-01
4-1 Outline Chapter 4 Maintenance
Replacement of Thermal Printer Paper When a red line appears on the both sides 4-3-1
4-02 ■ SE-1520 b
4-2 Daily Maintenance Chapter 4 Maintenance
Separate used samples, tips and protective gloves from general waste and
dispose of them according to local regulations on biohazardous waste.
Separate used samples, tips and protective gloves from general waste and
dispose of them according to local regulations on biohazardous waste.
b SE-1520 ■ 4-03
4-3 Periodical Maintenance Chapter 4 Maintenance
4-04 ■ SE-1520
4-1 Maintenanceの Chapter 4 Maintenance
4-1-1 Maintenanceの目安
SE-1520 ■ 4-05
Chapter 4 Maintenance
Correctly insert the check plate so that the arrows on the check plate face
forward.
If an error occurs, clean the point of contact shown in the next page.
4-06 ■ SE-1520 b
Chapter 4 Maintenance
Separate used samples, tips and protective gloves from general waste and
dispose of them according to local regulations on biohazardous waste.
b SE-1520 ■ 4-07
4-1 Maintenanceの Chapter 4 Maintenance
4-08 ■ SE-1520 b
4-1 Maintenanceの Chapter 4 Maintenance
The tip where the pipetting tip is attached has 2 nozzle O-ring.
Worn-out nozzle O-ring leads to incorrect sealing or sampling.
Replace nozzle O-ring (large and small) about once a year.
b SE-1520 ■ 4-09
4-10 ■ SE-1520
Chapter 5
Troubleshooting
5-1 Error Messages
5-2 Trouble Messages
b SE-1520 ■ 5-01
5-1 Error Messages Chapter 5 Troubleshooting
4-1-1 お手入れの目安
If an error occurs, an alarm sounds and the error message is displayed. Press
[STOP] key to stop the alarm. The error is cancelled and the Main Menu is displayed.
To avoid the recurrence of errors, confirm that the Main Menu is displayed and take
the proper measures described below. Confirm the condition, turn off the power and
contact your distributor.
Wear protective gloves to prevent exposure to pathogenic microbes.
Separate used samples, tips and protective gloves from general waste and dispose of them
according to local regulations on biohazardous waste.
E01 Mispipetting
OK(ENTER)
● The plate is not set or set in the ● Wipe dirt off the table.
E05 opposite direction.
● The barcode was not read
● Make sure that foreign matter
is not adhered to the plate.
E05 Plate Miss Set. correctly. ● Set the plate correctly.
OK(ENTER)
E10 Power down
OK(ENTER)
E20 Card misread
OK(ENTER)
5-02 ■ SE-1520 b
4-2 毎日のお手入れ Chapter 5 Troubleshooting
4-2-1 試験片テーブルの清掃
E25 Wrong card
OK(ENTER)
● The plate barcode is not read ● Wipe dirt off the table.
E35 successfully.
● Abnormal barcode sensor
● Make sure that foreign matter is
not attached to the plate.
E35 BAR misread ● Table is dirty. ● Set the plate correctly.
OK(ENTER) ● Foreign matter is adhered to the plate.
● The hand-held barcode reader is not set ● Check the setting (e.g. baud rate) of
E40 correctly.(e.g.baud rate)
● The hand-held barcode reader is not
the hand-held barcode reader.
● Connect the hand held barcode
E40 BCR error connected correctly. reader correctly.
OK(ENTER) ● The hand-held barcode reader is broken.
SE-1520 ■ 5-03
Chapter 5 Troubleshooting
● Make sure the setting for the sample type and its
● Measurement result is lower than
W20 the lower limit of the normal
measurement range.
normal measurement range are correct.
● Make sure the Lot No. on the display is same as
● Make sure the setting for the sample type and its
● Measurement result is higher than
W21 the upper limit of the normal
measurement range.
normal measurement range are correct.
● Make sure the Lot No. on the display is same as
W21 Out of nomal U ***** the one on the laminated aluminum package of the
※Note E-Plate.
● If the same trouble occurs, contact your distributor.
4-1-1 お手入れの目安
When trouble occurs with the analyzer, measurement stops, an alarm sounds and the
trouble message will be displayed. Press [STOP] key to stop the alarm. The error is
cancelled and the Main Menu is displayed. According to the System Initializing
Confirmation Display, press [ENTER] key and the internal system of the analyzer is
initialized. When initialization is finished, the Main Menu is displayed.
To avoid recurrence of the trouble, confirm that the Main Menu is displayed, then take
the measures described below. Please confirm the condition, turn off the power and
contact your distributor.
IMPORTANT When trouble occurs during measurement, perform measurement again. It may influence
the result measured before the trouble occurred. When the measured result appears to
be abnormal, start the measurement again.
Separate used samples, tips and protective gloves from general waste and dispose of
them according to local regulations on biohazardous waste.
b SE-1520 ■ 5-05
5-2 トラブルメッセージ Chapter 5 Troubleshooting
4-2-1 試験片テーブルの清掃
Description and troubles Problems and causes Remedy
5-06 ■ SE-1520
Chapter 6
Appendix
6-1 Transmission Specifications
6-1-1 External Output Format
6-1-2 Block Structure
6-1-3 Format of Measurement Results
b SE-1520 ■ 6-01
6-1 Transmission Specifications Chapter 6 Appendix
Hand shake Control by CTS or RTS is possible. (By default setting, this
control is not performed.)
XON/XOFF control is not available.
Time gap Two second waiting time is inserted between each block
(from<ETX> to <STX>).
6-02 ■ SE-1520
6-2 毎日のお手入れ Chapter 6 Appendix
The block structure has a certain rule. One block consists of start, data and
end. This is illustrated as below in the following explanations.
● Start (S)
The start of each block is < STX >.
The start of block is described as S in the following illustration.
● Data
The data (text) of each block is the main body of transmission content,
and is described by ASCII character arrangement. < CR >, < LF >, < RS >
or < US > is sometimes involved in the data. Controlling characters other
than these are not included.
● End ( E )
The end of each block is < ETX >.
The end of block is described as E in the following illustration.
SE-1520 ■ 6-03
6-2 索引 Chapter 6 Appendix
The measurement results are the same as the "Normal format" in SE-1510 English mode. The program
is designed to receive the measurement results (normal format) in SE-1510 English mode receives the
measurement results of SE-1520 normally.
measurement
S result E
6-04 ■ SE-1520
6-2 After-sales Service Chapter 6 Appendix
6-1-1 外部出力仕様
■Warranty A warranty is included in the packaging box for this analyzer. The warranty is
necessary when the analyzer needs to be repaired. After filling in the required
information and confirming the described contents, keep the certificate in a safe
place.
■About
When the analyzer does not Please contact distributor.
repair
function well
SE-1520 ■ 6-05