CERTIFICATE OF CALIBRATION “Romie Akedist Nasional
LABORATORIUM KALIBRAS!
UgeI2hIN
Issued By _ PT. Caltest Engineers Certificate Number
Date of Issue 21-Jan-14 435314013257
LABORATORIUM KALIBRASI
PT CALTEST ENGINEERS
Calt@st Those industrial Eat Block 2-V
JI. Engku Putri, Batam Center Batam Island Indonesia
‘Tel : +62 778 471704, +62 778 471705 — a
Fax +62 778 471706
info@ecaltestengineers.com
ENGINEERS
Customer
PT IKM INDONESIA
Puri Industrial Park 2000
J. Engku Putri Blok. C No 10 A, Batam Centre
Batam Indonesia
Instrument
Instrument: Pressure Gauge
Manufacturer: NLB corp
Model Number: -
Serial Number: NIL
Tag Number: 2
Range: 0 - 30000 psi
Calibration Date: 16-Jan-14
Ree Due Date: 16-Jan-15
Environmental Conditions
‘Temperature: 232°C Mains Voltage: 220V + 10%
Relative Humidity: 55 = 10% Mains Frequency: 60Hz
‘Comments
The Instrument Calibration Centre of Caltest Engineers is accredited to ISO/IEC 17025 under the
KAN Accreditted Laboratories ( KAN Cert No. LK-121-IDN)
“Calibration of this instrument has been accomplished using standards maintained by the Instrument
Calibration Centre of Caltest Engineers.
Calibration Method
-IK-P-01
“Guide to the Expression of Uncertainty in Measurement 2008
Traceability Information
“The Standard used to calibrate is traceable to IS unit through the Calibration Lab as below:
Instrument Description SN Traceability
Pressure Indicator 3282438 LK-121-IDN
Pressure Indicator 1331298, LA-2007-0370-C
ceasorocs.a5 rovesY
CERTIFICATE OF CALIBRATION Konite Ares Nason]
LABORATORIUM KALIBRAS]
Lk-21DN
Issued By _ PT. Caltest Engineers Certificate Number
Date of Issue 21-Jan-14 435314013257,
Page 2 of 2
Instrument
Instrument: Pressure Gauge
Manufacturer: NLB corp
Model Number:
Serial Number: NIL-
Tag Number: 2
Range: 0 - 30000 psi
Calibration Date: 16-Jan-14
Ree Due Date: 16-Jan-15
Calibration Result
Reading of Reading of Standard Correction
Ce ‘Ascending Descending ‘Ascending Descending
psi psi psi_ _psi psi
0 0 0 0 0
5000 4962 4902 38 9B
10000 9926 9826 “74 “174
15000 14916 14891 84 -109
20000, 19935 19825 65 175
25000 24920, 24826 -80 174
30000, 29875, 29875, 125 7125
Remarks:
The estimated measurement uncertainty associated inthis calibration is within = 180 psi at 95% confidence level,
with coverage factor 2.
~ Accredited calibration up to 16000 psi
- Traceability available for range 16000 psi to 30000 psi