DOI: 10.1002/acm2.12764
----------------------------------------------------------------------------------------------------------------------------------------------------------------------
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium,
provided the original work is properly cited.
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
KEY WORDS
FFF‐beam, lung SBRT, noncoplanar VMAT, RTOG‐0915, single‐dose
T A B L E 1 Characteristics of lung SBRT patients included in this study. Prescription dose was 30 Gy in one fraction.
Ipsilateral‐lung minus ITV
Patient no. Tumor location ITV (cc) PTV (cc) PTV diameter, d (cm) Mean HU Mean density (g/cc)
1 Midcentral, left lung 0.2 5.0 2.11 −761 0.24
2 Midcentral, left lung 0.3 5.1 2.12 −751 0.25
3 Upper lobe, left lung 0.7 6.4 2.28 −666 0.33
4 Lower lobe, right lung 0.7 8.2 2.48 −713 0.29
5 Upper lobe, left lung 10.1 41.1 4.22 −668 0.33
6 Midcentral, right lung 1.1 10.7 2.71 −858 0.14
7 Upper lobe, left lung 0.3 4.3 2.00 −716 0.28
8 Midcentral, right lung 13.6 37.5 4.09 −781 0.22
9 Upper lobe, left lung 0.8 5.2 2.13 −791 0.21
10 Upper lobe, left lung 1.4 11.0 2.73 −662 0.34
11 Upper lobe, left lung 2.1 14.8 3.00 −789 0.21
12 Midcentral, right lung 2.5 14.4 3.00 −823 0.18
13 Upper lobe, left lung 0.5 5.8 2.20 −661 0.34
ITV, Internal target volume; PTV, planning target volume; SBRT, stereotactic body radiation therapy.
SBRT/IGRT protocol by coregistering pretreatment cone‐beam CT sharp the dose falls off. The gradient distance (GD) is used to
with the planning CT scans. The patient setup, tumor matching evaluate dose sparing to normal lung volume.
online cone‐beam CT scan, and treatment delivery were monitored • Total number of monitor units, MU.
and verified by the treating physician and physicist. • Modulation factor, MF: ratio of total number of MU to the pre-
scription dose in cGy.
T A B L E 2 Plan quality evaluation for clinical 6X‐FFF and 6X‐FF (replanned) plans for all 13 VMAT lung SBRT patients.
Target volume Parameters 6X‐FF 6X‐FFF P‐value
PTV CI 1.09 ± 0.11 (0.98–1.39) 1.07 ± 0.08 (0.98–1.24) n. s.
GI 5.8 ± 1.3 (3.92–8.11) 5.5 ± 1.1 (3.81–7.23) P = 0.001
D2cm (%) 50.0 ± 5.5 (38.8–58.8) 48.0 ± 4.7 (37.8–55.1) P = 0.001
GD (cm) 1.1 ± 0.2 (0.79–1.52) 1.0 ± 0.2 (0.77–1.37) P = 0.001
HI 1.24 ± 0.1 (1.13–1.36) 1.20 ± 0.1 (1.10–1.25) P = 0.002
ITV Dmin (Gy) 32.8 ± 1.6 (28.7–34.5) 32.3 ± 1.3 (28.9–33.8) P = 0.04
Dmax (Gy) 37.0 ± 2.0 (33.9–40.9) 35.9 ± 1.5 (33.1–37.9) P = 0.002
Dmean (Gy) 35.2 ± 1.8 (32.9–38.8) 34.4 ± 1.2 (32.5–36.3) n. s.
ITV, internal target volume; n. s., not significant; PTV, Planning target volume; SBRT, stereotactic body radiation therapy.
The statistical significance at P < 0.05 are bold.
POKHREL ET AL. | 5
are plotted in Fig. 1. Traditional 6X‐FF beams systematically overpre- 0.6% vs 0.7%, 6X‐FFF vs 6X‐FF plan, respectively; all plan evaluation
dicted gradient indices (up to 12% higher) as a function of lower parameters were in favor of the 6X‐FFF beams. Major dosimetric
ipsilateral lung density suggesting that it could potentially deliver differences were observed in the values of intermediate dose‐spil-
higher intermediate dose‐spillage to the nontarget tissues compared lage and hot spots (tumor heterogeneity) with 6X‐FF plans compared
to 6X‐FFF beams. to 6X‐FFF plans. In addition, the absolute dose differences to the
Figure 2 shows an example case of radiosurgical dose distribu- OAR were up to 1.0 Gy with 6X‐FF plans, although, the OAR dose
tions in the axial view through the isocenter plane for a lung SBRT differences were shown to be clinically insignificant. However, the
patient (patient #6) with 6X‐FFF (right panel) and 6X‐FF (left panel). difference in the treatment delivery time was improved by a factor
This patient presented with the lowest average ipsilateral lung den- of 2.5 with 6X‐FF plans (17.65 min) vs 6X‐FFF (7.21 min) plans.
sity in the cohort. The PTV was 10.7 cc (2.71 cm diameter) and We calculated the dose profiles as a function of distance through
located in the midcentral right lung (island tumor). For example, in the isocenter position (from right to left and superior to inferior
this case, the CI, HI, D2cm, GI, GD, and V20 were 1.24 vs 1.39, 1.25 within D2cm) for patients with the lowest average ipsilateral lung
vs 1.36, 53.8% vs 58.8%, 7.23 vs 8.11, 1.37 cm vs 1.52 cm and density in this cohort. Figure 3 clearly shows that the dose distribu-
tions heavily depend on the lung density between the two plans. For
similar target coverage, the 6X‐FFF beams gave tighter dose distri-
butions (less intermediate dose‐spillage and more conformal dose
distributions), relatively small hot spots and fewer dose to the OAR
compared to traditional 6X‐FF beams. This is due to the low lung
density where the secondary electrons that have a relatively longer
range in lung parenchyma with 6X‐FF (average photon energy
1.75 MeV) beam compared to clinical 6X‐FFF beam (average photon
12
energy 1.28 MeV). In addition, the imbalance of the rates of sec-
ondary electron production in lung parenchyma and the tumor itself
caused dose buildup differences at the tumor center. These
appeared higher in lower lung densities (lung heterogeneities effect)
F I G . 1 . Scatter plot: Ratio between 6X‐FF and 6X‐FFF plans for with the 6X‐FF beams (see Fig. 3).
GI, D2cm, GD and HI as a function of average ipsilateral lung For the same planning objectives, optimization parameters, and
density for all 13 patients. Up to 12% higher values were observed
similar target coverage, very small, yet statistically significant differ-
for GI and GD with lower ipsilateral lung density with 6X‐FF plans.
In general, as the ipsilateral lung density decreases, the deviation in ences were observed for all normal lung parameters including V20Gy.
the gradient indices between the plans increases.
F I G . 2 . Comparison of dose distributions in the axial view for patient #6 (lowest average ipsilateral lung density of 0.14 gm/cc) with both
6X‐FFF (right panel) and 6X‐FF (left panel) beams. The red contour represents the ITV, orange represents the PTV (10.7 cc), green represents
the 95% isodose line that encompasses the PTV and the blue line represents the 50% isodose‐spillage (much tighter with 6X‐FFF beam). The
tumor was located in the middle of the right lung. The viewing plane intersection shows the isocenter location. The light blue ring was
generated to calculate D2cm around the target volume. ITV, Internal target volume; PTV, planning target volume.
6 | POKHREL ET AL.
F I G . 3 . Upper panel: Isodose distributions from 6X‐FF (left) and 6X‐FFF (right) plans in coronal views for patient #6 as shown in Figure 1.
Lower panel: Corresponding radial dose profiles drawn through the isocenter. Bottom left shows the dose profiles from left to right and the
bottom right shows the dose profiles from superior to inferior directions (within D2 cm). The green lines on each graph show that both plans
were normalized to get 95% of the target to receive 100% of the prescription dose. For identical target coverage, the ITV and the out‐of‐field
doses (intermediate dose‐spillage and dose to the OAR) were higher with 6X‐FF beams. ITV, Internal target volume; OAR, organs‐at‐risk.
T A B L E 3 Normal lung dose statistics between clinical 6X‐FFF and 6X‐FF VMAT plans for all 13 lung SBRT patients. Mean ± standard
deviation (range) and P‐values were presented.
OAR Parameters 6X‐FF 6X‐FFF P‐value
Lungs minus ITV V20Gy (%) 0.6 ± 0.4 (0.1–1.6) 0.6 ± 0.4 (0.1–1.5) P = 0.004
V10Gy (%) 2.7 ± 1.8 (0.6–6.3) 2.6 ± 1.7 (0.6–6.0) P = 0.002
V5Gy (%) 6.8 ± 3.7 (2.0–15.0) 6.4 ± 3.5(1.7–14.5) P = 0.015
MLD (Gy) 1.2 ± 0.6 (0.6–2.4) 1.2 ± 0.5 (0.5–2.3) P = 0.002
Maximal dose to 1000 cc of lung (Gy) 0.9 ± 1.0 (0.1–3.1) 0.8 ± 0.8 (0.1–2.5) n. s.
ITV, internal target volume; n. s., not significant; OAR, organs‐at‐risk; SBRT, stereotactic body radiation therapy.
The statistical significance at P < 0.05 is bold.
These values were uniformly lower with clinical 6X‐FFF beams com- (Dmax and D10cc) were evaluated per SBRT protocol guidelines. It
pared to 6X‐FF beams (see P‐values in Table 3). was observed that the volumetric dose differences to the heart,
A comparison of other OAR dosimetric parameters for 6X‐FFF esophagus, and ribs were statistically significant (see P‐values in
and 6X‐FF plans for all 13 lung SBRT patients is presented in Table 4. Table 4) between the two plans. Overall, the doses with 6X‐FF plans
Critical organs such as spinal cord (Dmax, and D0.35cc), heart (Dmax were higher by 1–15% for most of the critical organs, suggesting
and D15cc), esophagus (Dmax and D5cc), ribs (Dmax and D1cc), and skin that the average values of absolute dose differences could be higher
POKHREL ET AL. | 7
T A B L E 4 Average dose statistics (mean ± SD) for 6X‐FFF and 6X‐FF VMAT plans for all 13 lung SBRT patients.
Dose to OAR Parameters 6X‐FF 6X‐FFF P‐value
Spinal cord (Gy) Dmax 3.5 ± 2.6 (0.2–8.6) 3.4 ± 2.5 (0.1–7.8) n. s.
D0.35cc 3.2 ± 2.3 (0.1–7.6) 3.1 ± 2.3 (0.1– 6.9) n. s.
Heart/pericardium (Gy) Dmax 7.9 ± 4.8 (0.2–18.0) 7.7 ± 4.6 (0.1–17.6) n. s.
D15cc 3.8 ± 2.4 (0.1–9.3) 3.7 ± 2.3 (0.1–9.0) P = 0.02
Esophagus (Gy) Dmax 4.0 ± 2.0 (0.2–7.0) 3.7 ± 1.9 (0.2–6.7) P = 0.01
D3cc 2.3 ± 1.5 (0.1–4.7) 2.1 ± 1.4 (0.1–4.5) P = 0.04
Ribs (Gy) Dmax 22.2 ± 6.8 (12.6–31.7) 21.6 ± 6.8 (11.4–31.5) P = 0.001
D1cc 16.9 ± 4.4 (10.0–24.3) 16.5 ± 4.3 (9.4–24.0) P = 0.01
Skin (Gy) Dmax 8.9 ± 3.0 (4.5–15.2) 9.0 ± 2.6 (5.5–14.5) n. s.
D10cc 4.7 ± 1.7 (2.8–7.2) 4.7 ± 1.6 (2.9–7.9) n. s.
T A B L E 5 Comparison of average values of treatment delivery parameters (and range) between clinical 6X‐FFF and re‐optimized 6X‐FF plans
for all 13 lung SBRT patients.
Parameters 6X‐FF 6X‐FFF P‐value
Total monitor units (MU) 9034 ± 2159 (6245–15131) 9040 ± 2045 (6435–14684) n. s.
Modulation factor (MF) 3.0 ± 0.72 (2.1–5.04) 3.0 ± 0.68 (2.15–4.89) n. s.
Beam‐on time, BOT (min) 15.1 ± 3.6 (10.41–25.22) 6.5 ± 1.5 (4.6–10.5) P < 0.001
with 6X‐FF plans (of the order of 1.0 Gy) compared to clinical 6X‐
FFF plans. However, we predict the difference will not be clinically
significant.
F I G . 5 . Upper panel: Dose distributions from 6X‐FF (left) and 6X‐FFF (right) in coronal views for another example patient (patient #10) who
had the highest average ipsilateral lung density of 0.34 gm/cc. The PTV measured 11.0 cc and was located in the left mid‐lung adjacent to the
chest wall. Ribs, normal lung contour, and D2cm ring are shown. Lower panel: Corresponding radial dose profiles drawn through the isocenter
location. Bottom left shows the dose profiles from left to right and the bottom right shows the dose profiles from superior to inferior
directions (within D2cm). The green lines on each graph show that both plans were normalized such that 95% of target coverage received
100% of the prescription dose. For similar target coverage the ITV and out‐of‐fields dose (intermediate dose‐spillage and dose to the OAR)
were comparable with both plans. ITV, Internal target volume; OAR, organs‐at‐risk; PTV, planning target volume.
and V5Gy)32–34, and higher dose to skin.35 For instance, Pettersson breathing patterns along with the MLC modulation, gantry rotation,
29
and his colleagues reviewed 68 NSCLC patients treated via SBRT and dose‐rate changes are less likely to average out with a relatively
with 45 Gy in three fractions. Thirty‐three patients had complete shorter beam‐on time. However, it has been shown in the previous
clinical follow‐up and radiographic follow‐up exceeding 15 months. study (with 2400 MU/min) that the interplay effect causes clinically
Among these, 13 ribs fractures were found in seven patients. In their insignificant dose blurring (<3.0%) when using two or more arcs.36Our
study, the logistic dose–response curve showed that the risk of radi- noncoplanar VMAT lung SBRT plans (with 1400 MU/min) uses 3–6
ation‐induced rib fractures following lung SBRT treatments were partial arcs, therefore we do not expect clinically significant dose blur-
related to the dose to 2 cc of the rib. For the median follow‐up of ring due to MLC interplay. The change in respiratory patterns between
29 months, they showed that the 2 cc of rib receiving (3 × 9.1 Gy) the CT simulation and the time of treatment has also been previously
and (3 × 16.6 Gy) had 5% and 50% chances of rib fracture, respec- studied. 37–39Although, it has been reported that there were only small
tively. In the current study, utilizing 6X‐FFF beams for VMAT plan- changes (within ± 3 mm) due to intrafractional and interfractional
ning, all OAR metrics including rib, lung, and skin were lower motion in lung SBRT treatments, the mean patient setup time from
compared to traditional 6X‐FF beams. Furthermore, all OAR dose tumor localization to the end of treatment CBCT scan was about
metrics were well below the RTOG requirement, suggesting a low 40 min.39 It was suggested that an isotropic 5‐mm PTV margin around
probability of acute or late toxicities. the ITV (similar to our PTV margin) was sufficient to address the
Dose discrepancies due to interplay effects can be concerning for potential motion errors. In this study, our average beam on time was
single‐dose SBRT lung treatments with FFF‐VMAT plans. Changes in 6.5 min for a single dose of lung SBRT treatment. This potentially
POKHREL ET AL. | 9
decreased the variation of intrafraction motion error due to coughing 4. Fakiris A, McGarry R, Yiannoutosis C, et al. Stereotactic body radio-
or pain making a geographic miss less likely and improving the patient therapy of early‐stage non‐small‐cell lung carcinoma: four‐year
results of prospective phase II study. Int J Radiat Oncol Biol Phys.
stability as well as the clinic workflow.
2009;75:677–6822.
In summary, each 6X‐FFF and 6X‐FF plan was rigorously evaluated 5. Timmerman R, McGarry R, Yiannoutsos C, et al. Excessive toxicity
using the dosimetric parameters listed in the Tables 2–5. All parame- when treating central tumors in a phase II study of stereotactic body
ters were deemed acceptable per SBRT protocol suggesting that 6X‐ radiation therapy for medically inoperable early‐stage lung cancer. J
Clin Oncol. 2006;24:4833–39.
FFF plans are dosimetrically superior to 6X‐FF plans. Furthermore, 6X‐
6. Onishi H, Shirato H, Nagata Y, et al. Stereotactic body radiotherapy
FFF plans would also deliver much faster lung SBRT treatments which (SBRT) for operable stage I non‐small‐cell lung cancer: can SBRT be
would potentially improve patient compliance and clinic efficiency. comparable to surgery? Int J Radiat Oncol Biol Phys. 2011;81:1352–
While evaluating the target coverage and OAR doses as a function of 1358.
7. Crabtree T, Denlinger C, Meyers B, et al. Stereotactic body radiation
average ipsilateral lung density, it was observed that the island tumors
therapy versus surgical resection for stage I non‐small cell lung can-
with surrounding ipsilateral low lung density showed much higher vari- cer. J Thor Cardiovasc Surg. 2010;140:377–386.
ation between the modalities (6X‐FFF vs 6X‐FF plans) compared to 8. A randomized phase II study comparing 2 stereotactic body radiation
lesions located near the chest wall with a higher ipsilateral lung den- therapy (SBRT) schedules for medically inoperable patients with
stage I peripheral non‐small cell lung cancer; RTOG 0915. 2014 (1‐
sity; suggesting that FFF‐beam improves dose coverage at tumor–lung
67).
interface. Dose‐limiting toxicity after hypofractionated dose‐escalated
9. Videtic G, Stephans K, Woodly N, et al. 30 Gy or 34 Gy? Comparing
radiotherapy in NSCLC patients is still an issue in lung SBRT treat- 2 single‐fraction SBRT dose schedules for stage I medically inopera-
ment.40 Utilizing 6X‐FFF beams for VMAT SBRT lung planning may ble non‐small cell lung cancer. Int J Radiat Oncol Biol Phys.
potentially reduce dose to OAR, help enhance dose to tumor periph- 2014;90:203–208.
10. Videtic G, Paulus R, Singh A, et al. Long‐term follow‐up on NRG
eries, and deliver much faster treatments. However, while optimizing
Oncology RTOG 0915 (NCCTG N0927): a randomized phase 2 study
VMAT SBRT lung plans, planners are advised to pay special attention comparing 2 stereotactic body radiation therapy schedules for medi-
to the ipsilateral lung density and the tumor location on a per‐patient cally inoperable patients with stage I peripheral non‐small cell lung
basis as a function of beam modality. cancer. Int J Radiat Oncol Biol Phys. 2019;103:1077–1084.
11. Vassiliev O, Kry S, Chang J, et al. Stereotactic radiotherapy for lung
cancer using a flattening filter free Clinac. J Appl Clin Med Phys.
2009;10:14–21.
5 | CONCLUSION 12. Vassiliev O, Titt U, Pönisch F, et al. Monte Carlo study of photon
fields from a flattening filter‐free clinical accelerator. Med Phys.
2006;33:820–827.
For our 30 Gy single‐dose lung SBRT treatments, 6X‐FFF plans showed
13. Tambe N, Fryer A, Marsden J. Determination of clinically appropriate
dosimetrically superior isodose distributions, lower OAR doses, and flattening filter free (FFF) energy using treatment plans and delivery
much faster treatment deliveries compared to traditional 6X‐FF plans. measurements. Biomed Phys Eng Express. 2016;2:065016.
Additionally, the isodose distributions were significantly affected by the 14. Fu W, Dai J, Hu Y, et al. Delivery time comparison for intensity‐-
modulated radiation therapy with/without flattening filter: a planning
ipsilateral lung density and tumor location as a function of beam modal-
study. Phys Med Biol. 2004;49:1535–47.
ity. The dose enhancement at the tumor periphery was achieved by pre- 15. Xiao Y, Kry S, Popple R, et al. Flattening filter‐free accelerators: a
scribing dose at the tumor margin (rather than prescribing dose at the report from the AAPM therapy emerging technology assessment
tumor center) in addition to using 6X‐FFF beams. Given that FFF‐beam work group. J Appl Clin Med Phys. 2015;16:12.
16. Lu J, Lin Z, Lin P, et al. Optimizing the flattening filter free beam
are already widely available in clinics, we hope this study will help in
selection in RapidArc®‐based stereotactic body radiotherapy for
implementation of 6X‐FFF beams for fast, effective, and safe treatment
Stage I lung cancer. Br J Radiol. 2015;88:20140827.
of NSCLC patients treated with SBRT. Clinical follow‐up results of these 17. Navarria P, Ascolese AM, Mancosu P, et al. Volumetric modulated
single‐dose SBRT lung patients are underway. arc therapy with flattening filter free (FFF) beams for stereotactic
body radiation therapy (SBRT) in patients with medically inoper‐able
early stage non‐small cell lung cancer (NSCLC). Radiother Oncol.
CONFLICT OF INTEREST 2013;107:414–418.
18. Ong C, Verbakel W, Cuijpers J, et al. Stereotactic radiotherapy for
None. peripheral lung tumors: a comparison of volumetric modulated arc
therapy with 3 other delivery techniques. Radiother Oncol.
2010;97:437–442.
REFERENCES
19. Nagai A, Shibamoto Y, Yoshida M, et al. “Safety and efficacy of
intensity‐modulated stereotactic body radiotherapy using helical
1. Benedict S, Yenice K, Followill D, et al. Stereotactic body radiation
tomotherapy for lung cancer and lung metastasis. Biomed Res Int.
therapy: the report of AAPM Task Group 101. Med Phys.
2014;473173:2014.
2010;37:4078–4100.
20. Chan M, Kwong D, Law G, et al. Dosimetric evaluation of four‐di-
2. Zimmermann F, Geinitz H, Schill S, et al. Stereotactic hypofraction-
mensional dose distributions of CyberKnife and volumetric‐modu-
ated radiation therapy for stage I non‐small cell lung cancer. Lung
lated arc radiotherapy in stereotactic body lung radiotherapy. J Appl
Cancer. 2005;48:107–114.
Clin Med Phys. 2013;14:4229.
3. McGarry R, Papiez L, Williams M, et al. Stereotactic body radiother-
21. Holt A, Vliet C, Mans A, et al. Volumetric‐modulated arc therapy for
apy of early‐stage non‐small‐cell lung carcinoma: phase I study. Int. J.
stereotactic body radiotherapy of lung tumors: comparison with
Radiation Oncology Biol Phys. 2005;63:1010–1015.
10 | POKHREL ET AL.
intensity‐modulated radiotherapy techniques. Int J Radiat Oncol Biol fracture following lung SBRT. Int J Radiat Oncol Biol Phys. 2008;72:
Phys. 2011;81:1560–1567. S36.
22. Chetty IJ, et al. Report of the AAPM Task group No. 105: issues 32. Madani I, De Ruyck K, Goeminne H, et al. Predicting risk of radia-
associated with clinical implementation of Monte Carlo‐based pho- tion‐induced lung injury. J Thorac Oncol. 2007;2:864–874.
ton and electron external beam treatment planning. Med Phys. 33. Baker R, Han G, Sarangkasiri S, et al. Clinical and dosimetric predic-
2007;34:4818–4853. tors of radiation pneumonitis in a large series of patients treated
23. Chetty IJ, Devpura S, Liu D, et al. Correlation of dose computed with stereotactic body radiation therapy to the lung. Int J Radiat
using different algorithms with local control following stereotactic Oncol Biol Phys. 2013;85:190–195.
ablative radiotherapy (SABR)‐based treatment of non‐small‐cell lung 34. Guckenberger M, Baier K, Polat B, et al. Dose‐response relationship
cancer. Radiother Oncol. 2013;109:498–504. for radiation‐induced pneumonitis after pulmonary stereotactic body
24. Vassiliev O, Kry S, Wang H, et al. Radiotherapy of lung cancers: FFF radiotherapy. Radiother Oncol. 2010;97:65–70.
beams improve dose coverage at tumor periphery compromised by 35. Hoppe BS, Laser B, Kowalski A, et al. Acute skin toxicity following
electronic disequilibrium. Phys Med Biol. 2018;63:1–9. stereotactic body radiation therapy for stage I non‐small‐cell lung
25. Eclipse algorithms reference guide, Varian Medical System, 2018 cancer: Who’s at risk? Int J Radiat Oncol Biol Phys. 2008;72:1283–
26. Vassiliev O, Wareing T, McGhee J, et al. Validation of a new grid‐ 1286.
based Boltzmann equation solver for dose calculation in radiother- 36. Ong CL, Dahele M, Slotman B, et al. Dosimetric impact of the inter-
apy with photon beams. Phys Med Biol. 2010;55:581–598. play effect during stereotactic lung radiation therapy delivery using
27. Bush K, Gagne I, Zavgorodni S, et al. Dosimetric validation of flattening filter‐free beams and volumetric modulated arc therapy.
Acuros® XB with Monte Carlo methods for photon dose calcula- Int J Radiat Oncol Biol Phys. 2013;86:743–748.
tions. Med Phys. 2011;38:2208–2221. 37. Hugo G, Vargas C, Liang J, et al. Changes in the respiratory pattern
28. Kroon P, Hol S, Essers M, et al. Dosimetric accuracy and clinical during radiotherapy for cancer in the lung. Radiother Oncol.
quality of Acuros XB and AAA dose calculation algorithm for stereo- 2006;78:326–331.
tactic and conventional lung volumetric modulated arc therapy plans. 38. Bissonnette JP, Franks KN, Purdie TG, et al. Quantifying interfrac-
Radiat Oncol. 2013;8:149. tion and intrafraction tumor motion in lung stereotactic body radio-
29. Pettersson N, Nyman J, Johansson K‐A. Radiation‐induced rib frac- therapy using respiration‐correlated cone beam computed
tures after hypofractionated stereotactic body radiation therapy of tomography. Int J Radiat Oncol Biol Phys. 2009;75:688–695.
non‐small cell lung cancer: a dose‐ and volume‐response analysis. 39. Li W, Purdie TG, Taremi M, et al. Effect of immobilization and per-
Radiat Oncol. 2009;91:360–368. formance status on intrafraction motion for stereotactic lung radio-
30. Stephans KL, Djemil T, Tendulkar R, et al. Prediction of chest wall therapy: Analysis of 133 patients. Int J Radiat Oncol Biol Phys.
toxicity from lung stereotactic body radiotherapy (SBRT). Int J Radiat 2011;81:1568–1575.
Oncol Biol Phys. 2012;82:974–980. 40. Cannon D, Mehta M, Adkison J, et al. Dose‐limiting toxicity after
31. Dunlap NE, Biedermann GB, Yang W, et al. Chest wall volume hypofractionated dose‐escalated radiotherapy in non‐small‐cell lung
receiving more than 30 Gy predicts risk of severe pain and/or rib cancer. J Clin Oncol. 2013;31:4343–4348.