TRAINING MANUAL
CHROMATOGRAPH MAINTENANCE SCHOOL
FOR
LABORATORY AND PROCESS SYSTEMSTABLE OF CONTENTS
SECTION I
PRINCIPLES OF GAS CHROMATOGRAPHY Li
1.1 GAS CHROMATOGRAPHY Ll
1.2 TERMINOLOGY EI
1.2.1 TECHNIQUES El
1.2.2 APPARATUS 13
1.2.3 REAGENTS 13
1.2.4 OPERATING VARIABLES 4
1.2.5 DATA PRESENTATION 14
1.3 INSTRUMENT COMPONENTS 1s
1.3.1 CARRIER GAS 17
13.2 DETECTOR 17
13.3 VALVES 19
1.3.4 TEMPERATURE CONTROL 19
1.3.5 THE CHROMATOGRAPH COLUMN Eu
1.4 GENERAL COLUMN INFORMATION El
1.4.1 COLUMN PACKING 112
1.4.2 MULTIPLE COLUMNS AND VALVE CONFIGURATIONS 13
1.5 PERFORMANCE FACTORS 113
1.5.1 RESOLUTION 4
1.5.2 ANALYSIS TIME kis
1.6 APPLICATION ENGINEERING Lis
1.7 THE PROCESS CHROMATOGRAPH E16
1.7.1 THE ANALYZER 116
1.7.2 THE CONTROL UNIT Ly
1.7.3 OUTPUTS 17
1.7.4 SAMPLE SYSTEMS 17
1.8 PROBLEMS IN CHROMATOGRAPHY 19SECTION I
SAMPLE SYSTEM
2.1 INTRODUCTION
2.2 LOCATION AND MOUNTING
2.3 STREAM CONNECTIONS
2.4 COMPONENTS
2.4.1 FILTERS
2.4.2 OIL SEPARATORS
2.4.3 PRESSURE REGULATOR
2.4.4 VAPORIZER REGULATOR
2.4.5 LIQUID VALVE
2.4.6 STEAM TRACING
2.4.7 SCRUBBER
2.4.8 WATER COALESCER
2.4.9 CYCLONE SEPARATOR,
2.4.10 SAMPLE COOLER,
2.5 BYPASS FLOW
2.6 MULTIPLE STREAM SYSTEMS
SECTION Il
PROCESS ANALYZER
3.1 INTRODUCTION
3.1.1 CARRIER GAS FLOW
3.1.2 INSTRUMENT AIR FLOW
3.1.3 SAMPLE GAS FLOW AND BLOCKING VALVE.
3.2 HEATING SYSTEM
3.3 THE DETECTOR
3.4 INDICATING THERMOMETER
35 AL ES AND CONFIGURATIONS
3.5.1 SAMPLE VALVES
3.5.2 VALVE ACTUATION
Il
TH
m2
m2
2
m4
1-4
14
19
19
1-10
1-10
1-10
1-10
Tu
Tell
m1
M1
m1
4
ms
16
m6
m7
m7
m7
TL83.6 FLOW RATE ADJUSTMENTS
3.6.1 CARRIER FLOW
3.6.2 REFERENCE CARRIER
3.6.3 PRECUT CARRIER FLOW
3.7 ROUTINE MAINTENANCE
3.7.1 VALVE MAINTENANCE - 10-PORT
3.7.2 REPLACING DETECTOR FILAMENTS
3.7.3 COLUMN MAINTENANCE,
3.8 RECOMMENDED SPARE PARTS.
SECTION IV
LABORATORY ANALYZER
4.1 INTRODUCTION
4.2 THE INTEGRATOR
4.3 OPERATION
4.3.1 DESCRIPTION OF CONTROLS AND INDICATORS
4.3.2 PRINTED CIRCUIT BOARD PC-1
4.4 START-UP PROCEDURE,
4.5 VALVE MAINTENANCE - LIQUID
4.6 REPLACING HEATER
4.7 REPLACING BLOWER MOTOR
SECTION V
MC-503 CONTROL UNIT
5.1 INTRODUCTION
5.1.1 THE MC-503 HARDWARE
5.1.2 MC-503 SOFTWARE.
5.2 CONNECTING THE MC-503 ACM(s) TO PC
5.3 SOFTWARE INSTALLATION
5.3.1 EXECUTING THE MENU
5.3.2 A SAMPLE RUN WITH THE DEFAULT PARAMETERS
0-10
1-10
W-11
mil
W-11
m-12
1-16
17
1-18
IV-1
IV-10
IV-llSECTION VI
MC-503 SOFTWARE VEL
6.1 [A]: ANALYZER DEFINITION VI-2
62 [B]|: STREAM DEFINITION VI-2
6.3 [C]: COMPONENT DEFINITION Vi4
6.4 [D]: NAME VALVE DRIVERS VI-6
6.5 [E]: VALVE TIMES VI-6
6.6 [F]: ASSIGN SOLENOID DRIVERS V6
6.7 [G]: SET STREAM SEQUENCE VL7
6.8 [H]: SET LATCH BOARD MODULE, VET
6.8.1 RELAY/ALARM PARAMETERS Vi7
6.8.2 ANALOG OUTPUT PARAMETERS VES
6.9 [I]: RUN/STOP ANALYZER V8
6.10 (JJ: ENABLE/DISABLE STREAMS VEO
6.11 [K]: SET/CALIBRATE TEMPERATURE CONTROLLER VES
6.12 [L]: DISPLAY CURRENT ANALYSIS ‘I-10
6.13 [M): REPORT ALARM STATUS VEIL
6.14 [N]: DETAILED ANALYZER STATUS Vi
6.15 [O}: ACTIVATE DETECTOR PLOT Vi-12
6.16 [P]: PRINT DETECTOR PLOT FROM RECORDING Vi-13
6.17 (Q|: DISPLAY TREND VI-IS
6.18 IR]: DISPLAY STREAM HISTORY VL-16
6.19 [S]: STREAM ANALYSIS TREND VEIT
6.20 [T]: VIEW MODBUS REGISTERS VI-I7
SECTION VIL
THE ANALYZER CONTROL MODULE Va-1
7.1.1 YO DEVICES Vo-1
7.1.2 OPERATION OF INDIVIDUAL I/O DEVICES VI-3
7.2 CONSOLE-TO-ACM COMMANDS vu-4
7.2.1 ACM COMMANDS Vo-S7.2.2 DETAILED COMMAND DESCRIPTIONS Vo-6
7.3 SWITCH SETTINGS AND STATUS LEDs VO-1i
SECTION VII
TROUBLESHOOTING V1
8.1 MC-503 HARDWARE PROBLEMS VI-1
8.2 MC-503 SOFTWARE PROBLEMS Vm-2
8.3 MC-502 PROBLEMS ‘VI-3
8.4 PROCEDURE TO ZERO THE CHROMATOGRAPH BRIDGE vul4
8.5 RECENT UPDATES TO MC-503 SYSTEM. VuIL-S
85.1 INSTRUCTIONS TO USE NEW PASSWORD SYSTEM ‘VII-6
8.5.2 PULSED RELAY OUTPUTS VuL-7
8.5.3 AIC UTILITIES INFORMATION ‘VII-8
8.5.4 BACKING-UP THE MC-503 DATA ‘VI-10
8.5.5 INSTRUCTIONS TO COPY ALL ANALYZER SETTINGS VO-11v
n
12
13
14
15
16
7
18
19
20
MC-503D
INTRODUCTION
THEORY/ PHILOSOPHY
CALIBRATION
CALCULATIONS
GENERAL EDITING
[A]: ANALYZER DEFINITION
[B]: COMPONENT LIBRARY
(Cl: METHOD PARAMETERS
[DI-METHOD COMPONENTS
[E]-METHOD EVENTS
[F|‘STREAM DEFINITION
[G]:STREAM SEQUENCE
[H]-ALARM REGISTERS
(U;UTILITIES
[1]: RUN/STOP ANALYZER
[J]: STREAM ON/OFF METHOD
IK]: ANALOG OUTPUTS
[LI: CURRENT REPORTS
[M]: ANALYZER ALARMS
[IN]: ANALYZER STATUS
[0]: DETECTOR GRAPH
Ix-1
Xl
IX-1
IX5
X-8
1X10
TX-11
IX-13
IX-15
IX17
1X18
1X21
1-22
1X-24
1X25
1X27
1X28
1X30
1X32
1X-339.22
9.23
9.24
9.25
9.26
9.27
9.28
929
9.30
9.30.1
9.30.2:
9.30.3:
9.21.1 SCALES
IP}: PLOT
[Q]: GRAPHIC TEND
9.23.1 GRAPHIC TREND VIEW DIFINATION (SETUP)
[R]: HISTORY
[S]: LAST FIVE ANALYSIS
[7]: SERIAL DATA REGISTER
ANALYZER ALARM STATUS
ENVIRONMENT VARIABLE,
COMMUNICATIONS
APPENDIX
EXAMPLE: RATIO
EXAMPLE: SUMMATION
EXAMPLE: MODBUS,
GLOSSARY/ DEFINITION O F TERMS
1X40
1X33
1X35
1X36
1X37
1X39
IX-41
1X43
1X44
X45
1X47
X48,
X49
1X50
1X51SECTION I
PRINCIPLES OF GAS CHROMATOGRAPHY
11 GAS CHROMATOGRAPHY
Chromatography is a technique for separating the components in a mixture. This separation
is performed by distributing the components between two phases. One of these phases, called
the stationary or fixed phase, consists either of a solid adsorbent material (adsorption
chromatography) or of a film of liquid on an inert support (partition or absorption
chromatography). Whichever is used, it is packed into a tube of small diameter resulting in a
column. The mobile phase (carrier) consists of a gas which carries the gaseous sample
through the column. This carrier gas flows contimnously through the column at a constant
rate. At a predetermined time, a valve will introduce a fixed amount of sample into the
carrier gas stream. The cartier gas will push the sample through the column, where
separation of components results from the difference in the affinity between column material
and each of the sample components. Whether the nature of this affinity is adsorption,
solubility, chemical bonding, polarity, or molecular filtration, column material detains some
components longer than others. Consequently, all components flow through the column at
different rates and emerge in the inverse order of their affinity for the column material. At
the end of the column, a detector measures the amount of each component. Since each
component progresses through the column at a predictable rate, the elution time identifies
each component qualitatively. This process is demonstrated in Figure I-1.
12 TERMINOLOGY
Following are definitions of the terms frequently used in this manual:
12.1 TECHNIQUES
Gas-Liquid Chromatography - A description for all chromatographic methods where the
mobile phase is gas and the stationary phase is a liquid distributed on a solid support.Figure J-1 Components separation in a column.
RECORDER
outputGas-Solid Chromatography - All chromatographic methods where the mobile phase is a gas
and the stationary phase is an active solid such as alumina, activated carbon, silica gel, or
molecular sieves.
Gas Chromatography - All chromatographic methods where the mobile phase is gas. It is
the collective name for gas-liquid and gas-solid chromatography.
122 APPARATUS
Sample Valve - Device for introducing a fixed volume of liquid or gaseous sample into the
column.
Switch Valve - A valve which routes the gases in the chromatograph through alternate
channels according to the needs of the analysis. ‘The sample valve often performs this
function.
Column - A tube packed with stationary phase which separates components of a sample.
Detector - A part of the analyzer which detects the separated components. The detector
continuously measures the change in composition of the carrier gas as it emerges from the
column. A change in composition is indicative of the amount of sample component mixed
with the carrier gas.
123 REAGENTS
Carrier Gas - The mobile phase of a chromatograph system. It is a gas that is used to move
the sample through the column.
Liquid Phase - A non-volatile liquid coated on a solid support which constitutes the packing
for a gas-liquid (absorption) column. The sample components are dissolved in this liquid
phase as the sample moves through the columm and the separation of the components depends
I-3on their vapor pressure (volatility) in liquid phase.
Solid Support - An inert porous solid such as diatomaceous earth. The column efficiency is
influenced by particle size and surface area of the solid support. Active solids may be
modified to serve as solid support. It is also possible to use the inner wall of a capillary tube
for the same purpose.
‘Active Solid - A porous solid which is capable of separating sample components without any
liquid coating. In this case, the separation mechanism is adsorption rather than absorption.
1.2.4 OPERATING VARIABLES
Column Pressure - Column pressure affects the carrier and sample gas flow rates. It
depends upon the column selected and the carrier gas pressure. Column pressure is controlled
by the Carrier Gas Pressure Regulator and indicated by a pressure gauge mounted on the
regulator.
Carrier Gas Flow Rate - The volume flow rate of cartier gas at column outlet measured at
ambient temperature. It is controlled by the metering valves mounted on the carrier regulator.
Column Temperature - Column temperature is of prime importance to the sample analysis
and depends on the oven temperature.
125 DATA PRESENTATION
Chromatogram - A time plot of the detector response to sample components as they elute
from the column. This plot can be obtained from @ computer.
Printer - It provides the print-out of normalized and non-normalized analysis, and
chromatogram of a sample.
Component Peak - A response from the detector at the time of elution of the component
I-4from the column,
Peak Height - The distance between the baseline and the top of a peak. The peak height
commonly used as a quantitative measure of the component concentration.
Resolution - The degree of separation between two adjacent peaks.
Elution Time - Elapsed time between the injection of a sample and the detection of the
maximum peak height for each sample component by the detector. Since, for a given
column, every component has a characteristic elution time, it is used as a means of
component identification.
Baseline - The portion of a chromatogram between two peaks when no sample component is
being eluted. The detector output is normally adjusted to make the baseline coincide with the
electrical zero of the recording/printer unit.
Integrated Area - The area under a peak. This area is proportional to the quantity of the
‘component and is usually measured with an integrating device.
13 INSTRUMENT COMPONENTS
‘The industrial applications of the principles discussed in this section are the foundation of the
Process gas chromatograph. Basically, a process chromatograph needs a sample inject valve
to repeatedly inject fixed quantities of sample into the system, a regulated supply of carrier
gas to move the sample, a column to separate the sample into its individual components, a
detector to detect the components, temperature control to regulate the heater, and air for the
heater and valves actuation. The AIC Process Chromatograph incorporates these requirements
and is designed to provide automatic and continuous monitoring of process streams. Its
design incorporates the use of three modules: an analyzer, the MC-503 ACM, and 2 IBM
compatible PC. The essential elements of an AIC instrument are shown schematically in
Figure 1.2.“squouungsuy OY J0y WUT UWLOISKS om, Z-] ANB L
=
{| |
$0 (ATER) |
fainanon NOW! IG Fn |
sap9 ens oqo roe A,
“| doo =f in
sing yu py —L_ [a | wean) oe a I}
, (i ni] ny
i (myes
suas exdwog : |
01 OL al tL _
1 INA va
Seve | has —P-— |
Laas — oar <4 Sa 1
way +
at mu oa]
au a oy
a
of
I-613.1 CARRIER GAS
Since the carrier gas is the mobile phase, its selection is very important to achieve optimum
results from Gas Chromatograph. The carrier gas must be inert. It should not react with
column materials or the components. Since most systems utilize a thermal conductivity
detector, the carrier gas should have thermal conductivity properties different from those of
the gases to be analyzed. The majority of organic compounds have a thermal conductivity of
3 to 7 thus hydrogen or helium with thermal conductivity of 50 and 40 respectively, should
give the best results. However, because of safety factors, helium is generally preferred. It is
a colorless, inert gas, and will not bum. Nitrogen, argon, and carbon dioxide are less
frequently used as carrier gas because they decrease the sensitivity of the thermal conductivity
detector. For hydrogen flame ionization detector, any gas is acceptable as a carrier gas.
132 DETECTOR
Two types of detectors are normally used in process work: a thermal conductivity detector
and a hydrogen flame ionization detector. The thermal conductivity detector may use either
themnistors or hot wire filaments as the sensing element. Detectors with the hot wire filament
are arranged in a wheatstone bridge configuration with each of the four filaments located in
separate chambers within a metal block. Two of the detectors are continuously exposed to
the carrier gas only (reference side), while the other pair (active) are exposed to column
effluent comprised of carrier gas plus separated components (Figure F-3). A highly regulated
voltage, supplied by the detector power supply, is placed across the bridge which causes a
current to flow through the filaments. The heat produced by this current is carried away by
the cartier gas flowing across the filaments. As long as the detectors detect carrier gas only,
the bridge is balanced and no electrical output is produced. The light gases, such 2s helium,
Temove heat more efficiently than heavy gases. When sample mixed with cartier gas elutes
from the column and passes across the active detectors, the temperature of the filaments
increases. Since he Slaments have « positive temperature coefficient of resistence (hei
resistance increases with increasing temperature) and the voltage across the bridge is constant,
the current through the filaments should change, which causes an unbalanced bridge.-[ Power’
SUPPLY
RECORDER
[
OF FILAMENTS
NATCHED SET
REFERENCE
ER WwW
caRRi
couse: eFriuent 7
REFERENCE CARRIER OUT
Figure J-3 The thermal conductivity detectors with hot wire filaments.The unbalanced bridge produces a signal which is proportional to the amount of sample
moving across the detector. Sensitivity of the detector can be varied by changing the voltage
impressed across the bridge. Generally thermal conductivity detector is used for high parts-
per-million (ppm) concentration range. The second detector is hydrogen flame ionization
type, in which the sample components are ionized by combustion which produces a current.
Column effluent is mixed with hydrogen and bumed in a Teflon chamber. Combustible
materials containing carbon atoms produce ions which are collected by impressing a voltage
between the bumer tip and the collector (Figure I-4). The small current set up by flow of ions
is measured by an electrometer amplifier and sent to the ACM.
The flame ionization detector is sensitive to the parts per billion (ppb) concentration range of
sample. Since this detector measures ion current, it is only sensitive to components which
produce ions on combustion. This selectivity sometimes can be used to advantage such as in
the measurement of acetylene in oxygen or air with a full scale of only 10 ppm acetylene and
with almost complete insensitivity to the oxygen or air.
13.3 VALVES
‘The sample valves used in AIC instruments are multi-port rotary type operated by a
pneumatic air cylinder. The valve parts which are wet with sample are constructed of 400
series stainless steel and glass filled Teflon, making them resistant to corrosive materials
present in the sample.
The multi-port configuration of the valve serves the dual function of sample introduction and
column switching. Only complex applications will require a second valve.
13.4 TEMPERATURE CONTROL
A column is the heart of a chromatograph’s ability to do the job for which it is designed.
‘Consequently, factors which affect the functioning of a column are of primary interest. One
such factor is temperature. Resolution, sensitivity, and elution time are functions associated
with column which are affected by temperature. For example, if the column temperature is
too low, the sample will move through the column slowly resulting in a prolonged analysis
1-9Stack
Collector
Cathode anc Ignitor
tector Air In
Figure 1-4 Flame ionization detectortime and very low sensitivity since the recorded peaks will spread out in width. If the
column temperature is too high, the components will elute quickly and the recorded peaks
will crowd together, making resolution difficult. It is also essential that the column
temperature be such that the components of the sample remain in the vapor state when
passing through the column.
‘The high column temperature reduces the life of a partition column because of evaporation of
the liquid phase. Evaporation will cause the column to deteriorate and produce a decrease in
the component retention times.
The stability of the temperature is also very important during the column operation. Because
of unattended operation of a process gas chromatograph for long periods of time, it is vital
that there exist a method of temperature control. In the AIC Gas Chromatograph, various
column configurations are installed in a hot air bath. The air bath consists of an insulated
oven, a heater above which a stteam of air is constantly flowing, and an electronic control
system which automatically compensates for any temperature shifis. Such system eliminates
the necessity of placing different components of the analyzer at equal distance from the
heater. The temperature in the air bath analyzer is maintained at +.1° centigrade.
13.5 THE CHROMATOGRAPH COLUMN,
‘The column is the single most important factor affecting the ability of a chromatograph to
handle a specific application. It generally consists of an 1/8, 3/16 or 1/4 inch outside
diameter tube packed with a resolving material which may be either an adsorbing agent or a
Partitioning liquid on an inert support. The column is usually fabricated from stainless steel,
though Teflon due to its inert chemical characteristics, is enjoying increased use. Small plugs
of glass wool in either ends of the column retain the packing. Length of the column may
vary from a few inches to 30 ft. depending on the application.
14 GENERAL COLUMN INFORMATION
As previously mentioned, there are two types of column packing materials: active solids and
inert solids covered with a liquid. The common adsorption-type columns are activatedalumina, silica gel, activated carbon, and molecular sieves. They are nommally used for
separation of fixed gases and light hydrocarbons. Adsorption columns are susceptible to
contamination. The commonly used adsorbents can be contaminated by water, oxygenated
compounds, and in some cases, carbon dioxide. These contaminants reduce resolution and
sensitivity by reducing the amount of surface area available for sorption of sample
components. The contamination can be reversed by purging the column with dry carrier gas.
This procedure is a reactivation process and allows the column to be reused.
Partition columns are considerably more versatile than adsorbent types and have a much
wider range of application. They are also less susceptible to contamination. Some examples
of partition columns are:
* Dow Coming Silicon Oil on Firebrick
* Carbowax on Firebrick
° — Oranite on Teflon
14.1 COLUMN PACKING
While a column supplied with an AIC process chromatograph can be obtained on short notice
from AIC, occasionally a customer desires to produce his own columns. For an adsorbing
type column, all that is required is to pack the column tube with the correct mesh of the
proper material and then bake the column out for about 24 hours before installing it in the
analyzer. With partition columns, it is sometimes necessary to prepare the liquid phase and
apply it to the solid support before the column may be packed. The procedure for applying
the liquid phase is as follows:
+ The important factor in preparing a partition column is the ratio of liquid to
solid support. This value is generally known so all that is required is to weigh
the comect amount of liquid and solid support.
+ Add sufficient amount of solvent (acetone, methanol, ether, etc.) to the liquid
phase to produce 2 mixture which will completely wet the solid support.+ Pour the solid support into the liquid mixture and stir thoroughly.
+ Evaporate the solvent from the mixture by warming and stirring. Do not use
high temperature as this might cause the liquid phase to boil off. Once the
material is completely dry, pack it in a suitable tubing. While the tube is being
filled, it should be vibrated or tapped to ensure that the material settles
correctly. ‘The appropriate fittings are then attached and the column is installed
in the instrument. It should be purged with carrier gas at operating temperature
until the recorder baseline is stable.
142 MULTIPLE COLUMNS AND VALVE CONFIGURATIONS
It is often impossible to find a single column which will separate all of the desired
components. The solution most often adapted is to use two or more columns and,
occasionally, an additional valve. For many analyses, two columns are used when analysis
with a single column is too slow or has other disadvantages. For example, if the sample from
the process stream contains components which are harmful to the main column, it is advisable
to use a second column which will separate the harmful components from the primary
components before they reach the main column. The harmful components then can be
backflushed to vent.
There are literally hundreds of column packing materials available today, with additional ones
coming into use every year. The resulting possibilities in valve and column configuration are
staggering in their number. Yet, out of these complexity of columns, the best configuration
for the analysis desired must be determined. AIC personnel, with years of experience, are
qualified to deal with the problems of column selection. Consequently, AIC offers a wide
variety of column systems as standard modifications.
15 PERFORMANCE FACTORS
Once a column for a patticular application is chosen, its actual performance with respect to
resolution, analysis time, and sensitivity must be matched to the customer's requirements by
I-13suitable variation of the operating parameters. Predicting the actual performance frequently
presents difficulties because the desired performance factors are often mutually incompatible.
‘The measurements usually made on the basis of peak area, often require a different degree of
resolution than deemed necessary in non-industrial applications. Minor components of low
peak area must be resolved from the tail of 2 major component, if they are to be measured
quantitatively. Stability and rapid analysis times required for industrial applications dictate
that tailing effects must be kept to a minimum. This may exclude the use of certain
desirable partition liquids with excessive vapor pressures which are rapidly lost from the
column. Also, the inert support should be chosen to eliminate surface effects that introduce
tailing or non-linearity. Some general rules which might assist in determining the best
column for a given application are as follows:
15.1 RESOLUTION
‘Within limits, resolution is increased by increasing the length of a column, reducing the oven
temperature and, to a lesser extent, reducing the carrier flow rate. These factors increase the
retention time and generally reduce the sensitivity. Further, the point of diminishing returns
is reached rapidly because diffusion effects assume greater significance. If resolution between
two components is poor, it might be improved by decreasing sample size. However, if
sensitivity is a problem, small samples may not provide sufficient peak height for trace
components. Generally, when resolution between two components is poor on a column, it is
better to look for another column that is more suited to the application,
152 ANALYSIS TIME
Generally, the conditions of an application should be such that the analysis time is as short as,
possible. This procedure results in narrower peaks, greater sensitivity, and more sample
cycles per unit time. Diffusion effects is also minimized which results in better
reproducibility of peak height. Analysis time decreases with higher temperatures and, to a
lesser extent, with greater carrier flow rates. Unfortunately, both of these factors may have
damaging effects on resolution and column life. If column size (length and volume) is kept
I-14to a minimum, then analysis time can be shortened.
16 APPLICATION ENGINEERING
Since each process chromatograph is designed to fit a specific application, it is necessary to
coordinate or blend the various elements which are involved in order to achieve the desired
results. This process is called application engineering. Generally, when a customer orders an
instrument from AIC, he will request that we custom build the columns, select a valve
configuration, and determine the operating conditions necessary for each specific task. When
the chromatograph is delivered, all that is required to place it into operation will be the
installation procedure.
Occasionally, a customer will purchase only the “hardware” of a chromatograph system.
‘Under these conditions, the column selection and determination of operating parameters, such
as peak sequence and time cycle, will be customer's responsibility. In other words, customer
will perform the application engineering.
While it is impossible to provide a complete discussion of variables affecting the operation of
a chromatograph, we will nevertheless consider some of the basic principles of the operating
‘process.
‘Once the different elements of an instrument have been interconnected and are operating
properly, the temperature of the analyzer must be adjusted to some arbitrary value. Other
parameters, such as flow, are adjusted according to information given in the operating
‘manual, and the instrument is ready for a manual run. A sample is injected into the column
and a chromatogram is produced on the recorder. Depending on the results of this
chromatogram, adjustments are made to the various parameters to improve performance.
For instance, if the peaks of the chromatogram are too much apart then the temperature of the
analyzer can be raised and/or the cartier gas flow rate can be increased. If the resolution is
poor, then temperature and flow rate should be decreased.
Another area of possible adjustment is sensitivity. If peak height is inadequate, then the
‘bridge voltage can be increased. The normal operating voltage range of the bridge is 8 to 12,
volts, but higher voltage is permissible. However, very high bridge voltage will increase
noise and decrease the filament lifespan from overheating. An altemative is to increase the
I-15sample size by changing the sample loop on the gas sampling valve or the sample disk in the
liquid sampling valve.
When a multi-column and/or a multi-valve configuration is used, it is necessary to determine,
by experimentation, the best time to energize and de-energize the valve(s). For example, a
single and precut configuration is designed to precut or remove certain components of the
sample and prevent them from entering the main column by de-energizing the valve when the
components of interest have eluted from the precut column. If the valve is not de-energized
quickly enough, components which might damage the main column are permitted to elute.
Early precutting, on the other hand, might prevent a peak of interest from eluting.
Several manual chromatographic analyses are performed to determine the complete cycle
time, precutting or column switching time, and automatic zeroing point. Once these
parameters have been fixed, an automatic sequence may be programmed, regulating the
operation of the entire chromatographic system.
1.7 THE PROCESS CHROMATOGRAPH
A process chromatograph must be mgged, resistant to corrosion, and suitable for use around
explosive gases, because of the environment in which it is often situated. Also, because it is
often operated for long period of time unattended, it should be completely automatic.
‘The purpose of a gas chromatograph is to analyze a process stream. The unit should be
located close to the sample source for increased efficiency. But, most process plants are
actually operated from a control room which may be some distance from the process stream.
In order to fulfill these conflicting requirements, the process chromatograph consists of three
basic units: the analyzer, control unit, and data acquisition and presentation unit.
17.1 THE ANALYZER
Just as the column is the heart of an analyzer, the analyzer can be considered the heart of gas
chromatograph. The system components which are primarily responsible chromatographic
r-16analysis: sample valve, columns, and detector, are located in the analyzer. In addition, flow
and temperature controllers, which affect these primary components, are situated at the
analyzer.
Electrical components are installed in explosion proof enclosures for operations in hazardous
area. Also, the analyzer is designed to accommodate additional valves and columns, thereby
increasing the number of applications for which it can be used.
1.7.2 THE CONTROL UNIT
The unit is solid state and microprocessor controlled with a crystal time base and no relay
contacts. It is capable of the simultaneous operation of up to ten analyzers, with up to 10
streams per analyzer and up to 20 components per stream. The programmed functions
incorporated with the MC-503 are slope sensitivity, component detection, peak area
determination, internal elution time determination of each component, automatic component
elution time adjustment, continuous base-line zero tracking, and auto zero to compensate for
Jong term drift.
Input to the MC-503 can be from thermal conductivity, flame ionization, or flame photometric
detector. There are 4 status alarm inputs from each analyzer, and an RS-232 input from
another computer or CRT terminal. Power requirements are 115V 10A, 50/60 Hz, 150 watts,
plus analyzer load.
173 OUTPUTS
‘The MC-503 interfaces with an IBM compatible PC through a standard RS-232 port or
‘current loop serial port A hard copy of all analysis may be obtained through a parallel
Ftinter. Also available are up to thirty two (32) 4-20 ma outputs for trend recording or closed
loop control.
= )) SAMPLE SYSTEMS
In order for the accurate and reliable analysis with a chromatograph, it is necessary that the
I-17sample delivered to the analyzer be representative of the process stream being monitored.
‘This requirement places the sample system on a level of importance equal to that of the
analyzer. Some of the operations which a sample system may be called on to perform are:
+ Take a representative sample from vessel to process line.
+ Transport it to the instrument without excessive delay.
+ Remove particulate matter.
+ Filter mists and aerosols.
+ Vaporize liquid samples.
+ Prevent vapor samples from condensing.
+ Measure and control flow and pressure through the analyzer.
+ Remove unwanted components.
+ Accommodate introduction of calibration standards.
+ Accommodate analysis of more than one stream with one analyzer.
The AIC sample conditioning system is designed to assist analyzer in
performing the specific job for which the chromatograph system, was designed.
Basically, each sample system regulates pressure, filters and controls the flow
of streams to the analyzer, and introduces a calibration standard either manually
or by means of a valve controlled by the programmer. Generally, AIC sample
system is mounted on a rack with the analyzer directly above it.
I-1818 PROBLEMS IN CHROMATOGRAPHY
‘As might be expected with an instrument of enormous application possibilities, problems are
sometimes encountered. Generally, these difficulties are results of inadequate information.
For example, occasionally, when a process analyzer is connected to a plant stream, the stream
is found to contain components which were not known to be present. These unexpected
components in the sample may cause problems by eluting from the column at the same time
as a peak of interest or by interfering with the operation of the sample system. To correct
this problem, it may be necessary to substitute a different column, or to adjust the flow,
pressure and temperature parameters. If a plant sample is available at the time of original
application engineering, it is possible to minimize such problems.
‘A gas chromatograph must be calibrated with a sample of known composition. The improper
standard sample will result in incorrect chromatographic analysis.
Another factor which greatly affects instrument accuracy is the physical state of a sample. A
chromatograph column accepts only vapor samples. However, since the chromatograph is
required to monitor both gas and liquid process streams, it is important to vaporize the liquid
sample so that it is acceptable to the column. The partial vaporization of samples will result
in different sample volume, from one sample to another, hence large inaccuracies in analysis
of sample will result. The similar difficulties can occur with a vapor stream. If condensation
of vapor samples occurs, the sample will not be in vapor phase and measurement errors will
again result. To facilitate vaporization and prevent condensation, it may be necessary to heat
the sample lines. Steam tracing normally provides the heat source. However, the incorrect
tracing may develop cold spots, resulting in condensation of sample.SECTION IL
SAMPLE SYSTEM
2.1 INTRODUCTION
It is generally recognized that a comectly designed and built sampling system is as important
as analyzer for proper analysis of the sample. This section will provide the necessary
information for operation and maintenance of the Sample System.
Each sample system is designed for the specific application, so the customer is assured of
receiving the correct system for his needs. However, certain basic components are used in
every sample system. A discussion of the various components utilized in the AIC sample
system is included in this section.
‘The function of the sample system is to prepare the sample for analysis. This preparation
includes controlling and measuring flow, regulating and measuring pressure, filtering the
streams to remove particulate matter, vaporizing liquid samples, and maintaining the
‘temperature of vaporized samples. The unit also handles selection of the stream to be
analyzed and introduction of a standard sample upon instructions from the controller.
‘The components of the sample system are housed in a heavy duty steel cabinet (3 standard
sizes available for AIC sample system) with a latched and gasketed door. If required, an air
bath heated system can be incorporated. The unit may be rack or wall mounted, and is
suitable for installation in almost any area of a plant. This characteristic is extremely
advantageous when one considers that the unit should be located as near the sample point as
possible,
2.2 LOCATION AND MOUNTING
Several factors should be considered before positioning the sample system at a location. It
should be positioned as close as possible to the analyzer to minimize the response time. At
II-1the same time, it should be located near sample points so that a constant supply of fresh
sample can be assured. In both cases, placement should be in either an instrument room or
house set up for that purpose, or in a semi-protected area.
Once the location for the sample system has been determined, it will be necessary to mount
the unit. Generally, when the sample system is shipped, it has already been mounted, by
AIC, in a house or on a rack with the analyzer. If this is not the case, then the cabinet
should be mounted vertically against a panel or wall and secured with four 7/16 bolts.
23 STREAM CONNECTIONS
‘The sample system should be located near sample points and a rapid flowing stream from
sample probe (Figure II-1) should be piped to the immediate vicinity with a side stream
directly to the conditioner. Sample lines are generally 1/4” SS, and there is a sample inlet
and bypass for each stream. Since the unit is capable of analyzing eight streams per analyzer,
there can be as many as sixteen stream connections. When a calibration standard is utilized,
there will be one 1/8” connection to be made. All of the aforementioned comnections are
made at the side of the sample box. A vent line is located at the bottom of the unit which
should be connected to a suitable vent and should not be returned to the process stream.
24 COMPONENTS
As previously stated, each AIC sample conditioner is custom built for its particular
application. A functional description of these components follows.
2.4.1 FILTERS
A suitable filter is the Swagelok bypass unit provided with most sample systems furnished by
AIC. This filter has two flow paths. The main path is tangential to the filter elements. This
flow helps to sweep away particulates accumulated on filter elements and minimizes the
tendency of plugging of filter element. A side stream is taken through the filter element.
qI-2aqoud
aidueg 1
03 wy
p
“74
“Tamdyy
ye Rie
‘BAN CROWS SNA CD)
HON
ous
ment]
vag
ORD al a
To mana!
||
Ho
-od_fe
L-
zhy—e}
Pan2.4.2 OIL SEPARATORS
oe eS
Often the sample might contain entrained mist, such as sulfuric acid, absorption oil,
compressor lubricant, or heavy ends. Since these materials are difficult to vaporize, they
might collect and cause line plugging or damage to the chromatograph column. They should
‘de removed using a suitable oil separator or other knockout device.
2.4.3 PRESSURE REGULATOR
‘The chromatograph accepts a vapor or liquid sample, depending on the of type of sample
valve used, at a pressure between 1 and 300 + 1 psig. Vapor samples should be provided to
the analyzer at a pressure of 1 to 10 psig regulated to + 0.5 psig. Liquid samples up to 2
maximum of 800-1000 psig can be analyzed. A special all-stainless pressure regulator, AIC
Part No. 12125, or equivalent, is recommended for vapor sample regulation, This regulator
can also be used for liquid samples when a liquid sample valve is provided. Brass or
aluminum regulators can be used depending on the nature of the sample. A filter should
always be used up-stream of the regulator to protect the seat. Most regulators have built in
filters ahead of the seat.
24.4 VAPORIZER REGULATOR
The AIC Vaporizer Regulator is a combination of vaporizer and pressure regulator designed
to vaporize liquid samples with boiling points up to 300°F and to regulate the pressure of
vaporized samples precisely to within + 0.1 psig. The unit can be mounted at the sample
point, but is generally installed in the sample conditioning box. Heat for vaporization of
liquid sample is provided by steam (maximum steam pressure 300 psig) or an electrical
heating clement. The unit is weatherproof and a spring loaded design allows mounting in any
Position, since gravity has no effect on its operation. Maximum up-stream pressure range is
2000 psig, and an automatic vent eliminates the possibility of downstream pressure exceeding
the preset value. The downstream pressure can be set to any value between 8-30 psig by
tuming the adjusting screw at the top of the regulator.
rI-4Specifications
+ Stability: For input pressure between 50 and 700 psi, the regulator will
maintain the output pressure within 0.5 psi with an upstream pressure change
of 50 psi, or within 0.1 psi with upstream pressure variation of 10 psi. For
input pressure between 20 and 50 psi, the regulator will maintain output
pressures within 0.1 psi for upstream pressure variations of 5 psi.
+ Flow Capacity: Maximum flow capacity of regulator, while maintaining the
stability as stated above, is approximately 2200 co/min. A variation of flow
rate from 10-2000 ce/min. will not change the outlet pressure more than 1 psi.
+ Pressure Ranges: Maximum inlet pressure 700 psig, outlet pressure 8-30 psig.
: ‘Vent Flow Rate: Varies with downstream pressure but should be less than
300 ccfmin.
+ Maximum Steam Pressure: 300 psig.
+ Size: Space clearance required for mounting, 4 1/2 inches width x 7 3/8 inches
height.
+ Weight: Approximately 10 pounds.
. Construction Materials: Components wetted by the sample are made of AISI
‘Types 302, 420, and 440 stainless steel, Monel and Teflon.
Installation
Figure IL-2, shows the mounting dimensions and sizes of connectors required for installation
of vaporizer regulator. It is recommended that a valve be installed in the sample line between
TI-5ADJUSTING ALLEN CAP SCREW
aq| o
738" Ft VENT OUT 1/8" NPT
VAPOR QUT 1/8" “NPT
vate,
STEAM WOR CONDUTT FoR USE __ TTI ai foes OE Nee eden
WITH ELECTRICAL HEATER hpi LUN HEATER 15 USED 1/4" NPT.
v2" NPT
Lup SAMPLE We" wer HTT
NOTES -
MATERIAL WEIGHT
AIST Types 302, 420 and 440 Stainless Steel, Monel, and Teflon TO Lbs.
Max. Steam Pressure 300 PSIG
Sample Inlet Pressure 700 PSI
Sample Outlet Pressure 8-30 PSI
Flow Capacity 0-2,000 cc/min.
Figure IL? Installation drawing for AIC vaporizer regulator.the regulator and the sampling point to shut off the sample for periodic cleaning of the porous
metallic filter in the vaporizer regulator.
‘The regulator is rugged and corrosion-resistant and designed to withstand the environmental
conditions of outdoor locations.
If the downstream sample line is to be steam-traced, the steam should be brought first to the
regulator, then into the tracing line.
If the user desires, a downstream pressure gauge, flow meter, and flow control valve can be
installed.
Operation
After the installation of regulator, turn on the steam supply and allow a few minutes for
warm-up. Unscrew the adjusting screw on the top of the regulator all the way and open the
sample valve to admit sample. Adjust the regulator to give the desired output pressure as
indicated by the downstream pressure gauge. Set the sample flow through the analyzer to the
desired value with the downstream flow control valve, and re-adjust the regulator output
pressure if required. The regulator will now operate without further attention.
Maintenance
‘The vaporizer regulator should require litle servicing except for occasional cleaning of the
porous metallic filter. To access the filter, remove the bottom cap and the screw in the center
of the filter. If the unit is to be dis-assembled (Figure II-3), it should be done only by a
qualified instrement technician. Great care must be taken to avoid buckling the metal
diaphragm during disassembly. The unit may be retumed to AIC for reconditioning when
required.
I-7Ayquiasse s0yvjnBo1 sozpodes E11 82n3}yVaporizer Regulator Recommended Spare Parts List
DESCRIPTION PART NO. ory
Filter 2010 1
Gasket 2018 2
Gasket 2017 1
Gasket 2019 i
Gasket 2020 i
Spring 2016 1
Ring, Retainer 13005-2 1
Diaphragm 2007 1
Pin Ball Release 2008 1
Ball, Precision 2009 1
Spring, Retention 2006 1
2.45 LIQUID VALVE
‘When samples are introduced through a liquid valve to the analyzer, no vaporization facilities
are required, but the sample must be completely liquid.
24.6 STEAM TRACING
It is necessary to maintain some sample lines at higher temperatures to prevent condensation.
One effective method of accomplishing this is to employ steam tracing or jacketing. If steam
tracing is employed, the steam lines should be close-coupled to the sample lines and
components to assure good heat transfer, and the two lines lagged. When multiple stream
manifolds are supplied, all the solenoid valves and other components should be steam traced
and lagged. Care should be taken to leave no cold spots. If vaporizing regulators are not
utilized, the sample cabinet should be heated.
qI-92.4.7 SCRUBBER
It is sometimes required to remove certain materials from the sample prior to analyzer, such
as large amounts of water or acid. An effective method of removing these materials is to
pass the sample through a stainless steel, or Monel scrubber. The scrubber is packed (usually
by the customer) with 2 material suitable for the specified application. A scrubber can be
made from a length of two or three inch pipe with end caps and sample IN and OUT fittings
at both ends. The end caps are removed to replace the packing.
24.8 WATER COALESCER
The water coalescer removes free water or water vapor from Hiquid or gaseous samples which
contain 10% or less free water and no solid particles. When water coalescer is used with a
sample which is heavier than water, it should be installed in an inverted position. As the
water and gas mixture flows through the polymer membrane, the water separates from the gas
and flows down the outside of the membrane and vent out with the bypass flow.
24.9 CYCLONE SEPARATOR
‘The cyclone separator is used to remove entrained water in a gas stream. The cyclone
separator separates water from gas by centrifugal force, the water collects on the sides of the
separator and drains at the bottom of the separator. Inlet and outlet comnections are 1/4"
NPT.
2.4.10 SAMPLE COOLER
The sample cooler is a counterflow heat exchanger for cooling of high temperature samples.
It is used to drop temperature of liquid samples to prevent the liquid sample from affecting
analyzer temperature and becoming two-phase. It can also be used to separate moisture in a
hot gaseous sample.
II-1025 BYPASS FLOW
Sample lines should be kept as short as possible to maintain rapid instrument response. For
liquid samples, a speed loop should be connected to a tee and the inlet port of the vaporizer
regulator (if used) with a bypass flow at this point. For either gas or liquid samples,
additional bypass flow is desirable at either the stream selection manifold or the analyzer unit.
The bypass flow can be vented or recycled to a low pressure point in the process. Bypass
flow should be chosen so that the transit time of the sample is minimal.
2.6 MULTIPLE STREAM SYSTEMS
‘The sampling techniques described previously apply to each individual stream in multiple
stream installations. In addition, there are special considerations relating to operation of the
sample system for multiple stream instruments. The system uses the backflush system, where
the sample, from the stream being analyzed, flows not only through the valve to the analyzer,
but also circulates through the other solenoid valves, to purge the lines. If a solenoid fails to
seal completely after a period of use, any leakage will be purged out the vent line. The
system includes for each stream a tee, for sample bypass, a stainless steel 3-way solenoid
valve, and a needle valve in the vent port of the solenoid valve. The needle valve can be
used to adjust the bypass flow. For proper conditioning of sample, these components can be
mounted in an insulated oven for electrical or steam heating.
I-11SECTION DL
PROCESS ANALYZER
3.1 INTRODUCTION
‘An analyzer is generally considered to be the heart of a process chromatograph system. As
stated in Section I, it is the analyzer’s components: the sample valve, columns, and detector,
which are primarily responsible for performing the chromatographic analysis. The sample
valve repeatedly injects a fixed quantity of sample into the column. A carrier gas carries the
sample through the column, where it is separated into its individual components. These
‘components then elute in the inverse order of their affinity to the column packing and flow
across the detector where they are measured.
‘While the sequence of component elution is dictated by the chemical and physical properties
of the column, the time of elution is determined by carrier flow rate and column temperature.
Since elution time identifies each component qualitatively, and since all automatic
programming functions are based on this information, the importance of precisely controlled
carrier flow and oven temperature is obvious.
Basically, an analyzer (Figure IN-1 and Figure I1J-2) consists of an air bath unit with precise
temperature control, a detector, and columns and valves for a specific application. On AIC
Model 600 analyzer, the regulators and pressure gauges are mounted outside where they are
readily available for monitoring and adjustment. The detector, valve and columns are located
inside the analyzer where they benefit from a constant temperature environment. All
automatic functions regulating the operation of the analyzer are controlled from the PC.
3.L1 CARRIER GAS FLOW
In theory, any gas which is not a component of the sample may be used as a carrier.
However, only very few gases make efficient carriers. Helium, because of its inertness and
high thermal conductivity, is the most frequently used. Carrier gas flow rate must be
III-1Carrier Regulator Carrier In Hester Air Press
Gauge
Carrier Pressure
Gauge
Heater Air Reguiat
Thermometer
Column Out
Figure IIL1 Exterior of AIC model 600 Analyzer.oaueg wanquseduey,adjustable over a wide range in order to control the elution rate. At the same time, it is
important that the flow should be constant throughout sample analysis.
Carrier gas is supplied to the analyzer from the cylinder at a constant pressure, generally 100
psig. It first enters the analyzer at the carrier regulator (Figure I-1). The carrier gas
regulator is provided with three outputs. Two of these, marked secondary, are adjustable
from 0 to 100 psig, while the third supplies carrier gas at the primary unregulated pressure.
From one of the secondary ports, the gas goes through the sample valve and to the main
column. A 0-100 psig pressure gauge is also attached to this port and it measures carrier gas
pressure at the column. Gas from the other secondary port passes through a metering valve
(mounted on the regulator) and goes to the reference side of the detector. The unregulated
port is in service only when a precut column configuration is utilized. In this case, the carrier
‘gas passes through a metering valve and is used to backflush the precut column. The port is
plugged when the precut operation is not used.
3.1.2 INSTRUMENT AIR FLOW
An air supply at 40-50 psig to the analyzer is required. The air is used for heating and valve
actuation for sample introduction and column switching (Figure Il-1). It enters the analyzer
through 1/4” tubing and is connected to a tee on the four-way solenoid used to actuate the
sample valve. From the tee, the air is piped to the inlet of a single stage regulator. A
Pressure switch is connected to one port of the regulator.
The air going to the regulator is used to heat the analyzer. The output of this regulator is
adjustable from 0 to 15 psig, and is normally set at 8 psig. From it, air travels to the heater
condulet, over the heater, and out into the analyzer. The safety pressure switch is an
electrical switch which is set to cut the voltage to the heater if air pressure should fail. It is
normally set for about 3 psig, but can be increased or decreased by turning the adjustment
serew on the top of the switch. No adjustment should be attempted until the set screw
securing the adjustment screw has been loosened.
TII-4The sample inject valve (and column switching valve, when used) is operated by air supplied
by the four-way solenoid. When this solenoid is de-energized, air is applied to the rear
chamber of the air cylinder, causing the shaft or piston arm of the cylinder to be extended. If
the solenoid is energized, air is shifted to the front of the cylinder, causing the piston arm to
retract. The sample inject valve assumes a new position for proper operation of the
chromatograph sequence.
3.13 SAMPLE GAS FLOW AND BLOCKING VALVE
Generally, by the time the sample reaches the analyzer, it has already been filtered, vaporized,
or condensed, reduced in pressure, etc., and no further preparation is necessary. Occasionally,
however, as in the case of a fairly uncomplicated sample, the sample handling equipment will
be attached to the side of the analyzer. Whichever is the case, the sample system drawing in
the modification section of this manual, gives a schematic presentation of the sample system.
Sample flow through the analyzer is usually on the order of 200 cofmin. for gas, and 100
cc/min. for liquids. It flows continuously through the sample valve, and leaves the analyzer
through the sample out port. Sample pressure of gaseous or vapor samples at the inlet of the
analyzer should be reduced to a value (about 5 psig) just sufficient to move the desired flow
through the analyzer. Liquid samples, on the other hand, must be under sufficient pressure to
‘prevent them from flashing, To achieve accuracy and reproducibility in the sample size, the
flow in the sample loop is blocked by the blocking solenoid just before sample introduction.
‘This action allows the pressure of the sample trapped in the sample loop to stabilize. When
the sample is a vapor, the blocking solenoid is located before the sample loop, and its
actuation allows the trapped sample to bleed to atmospheric pressure. Since this pressure is
essentially constant, the sample volume will be repetitious. In the case of a liquid sample, the
blocking solenoid is located after the sample loop. When the solenoid is energized, flow
through the sample loop is stopped, and the sample assumes the sample in pressure. Since
this pressure is held constant by regulators, a consistent sample volume is assured. This also
allows any gas bubbles to go back into solution at full sample pressure thus minimizing the
chances of a two phase sample.
rII-53.2 HEATING SYSTEM
‘An important feature of the AIC analyzer is the control of its temperature by a solid state
temperature controller. The system consists of an electronic controller, a sensor, and a heater.
‘The electronic controller portion is located in the MC-503 controller. Any adjustments to the
temperature of the analyzer are made at the console. The sensor is a solid state probe located
out of the stream of hot air from the heater assembly that distributes the air flow through the
oven. The heater is an air-flow-through type and is enclosed in a condulet mounted at the
bottom of the analyzer. Maximum power consumption of the heater is 800 watts and the
temperature control within the analyzer is approximately 20.1°C.
A high temperature limit thermoswitch is located in the detector condulet, and is designed to
remove power from the heater if the analyzer temperature exceeds 300°F. Any time this
switch is used, an investigation of the heater system should be undertaken to determine the
cause for high temperature.
33 THE DETECTOR
‘The standard detector used in the AIC Model 600 analyzer is in the form of a four element
bridge. Each of the filaments has a resistance value of approximately 30 ohms at 25°C.
Current is supplied to the bridge by a highly regulated power supply, located in the control
unit. This power supply is adjustable from 7 to 20 volts with the nomal range being between
8 and 12 volts. The filaments are mounted in a metal block, called the detector block, which
is in tum mounted in an explosion proof condulet known as, you guessed it, the detector
condulet.
‘These filaments are constructed from 2 combination of nickel and iron, and enjoy the
following advantages over other types of filaments:
+ They maintain good sensitivity at high temperature.
III-6= By operating hot, they discourage coating.
+ They are capable of surviving long periods of operation without a carrier gas.
In fact, some filaments of this type have remained operable after 24 hours
without carrier.
Despite their ruggedness, however, filaments eventually become mismatched, and it becomes
necessary to replace them.
3.4 INDICATING THERMOMETER
Each analyzer is equipped with a thermometer located on the door. It indicates the
temperature of the oven.
3.5 VALVES AND CONFIGURATIONS
‘The components which we have discussed so far in this section, detectors, heaters, etc., are
the same in almost every analyzer. It is the addition of the valves and columns which makes
a chromatograph uniquely able to perform a specific analysis.
‘There are literally hundreds of columns available for the separation of specific components,
and up to four of these columns normally can be installed in the Model 600 analyzer. The
direction of flow of carrier gas and sample through these columns is determined by the valves
and the way they are interconnected. AIC uses two types of valves; the standard 10-port
rotary valve, and the liquid valve. A total of two valves, in any combination, can be mounted
in the analyzer. The various ways in which the valves can be interconnected results in
ammerous column configuration possibilities.
3.5.1 SAMPLE VALVES
As previously indicated, AIC utilizes two types of valves, a ten port gas valve and a liquidvalve of 4, 8 or 10 ports. The gas valve is the more common, and a picture of the assembly
can be found in Figure III-3. ‘The lower section of this valve is constructed of 440 stainless
steel, and the upper rotary body is made of aluminum surfaced with fluorogold. No
lubrication between the mating surfaces is required. The valve cap contains an arrangement
which allows the pressure between the rotary and the lower body to be vatied by tightening
or loosening the cap. This construction pemmits the use of high carrier and sample pressures.
A general explanation of its operation follows. The upper surface of the lower body contains
10 equally spaced ports in a circle. The fluorogold surface of the upper rotary contains 5
‘grooves machined into it. Each groove is designed to cover two ports simultaneously. When
the valve is de-energized (the shaft on the air cylinder is extended) ports 1 and 2 are covered
by a single groove and hence are common to one another. The same is true of ports 3 and 4,
5 and 6,7 and 8, and 9 and 10. When the valve is energized, a different set of ports are
common. In this case 2 and 3, 4 and 5, 6 and 7, 8 and 9, and 1 and 10 are the connected
pairs.
‘The second type of valve is the liquid valve. AIC uses the Valco air-actuated 4 port. The
Valco valve is constructed of 303 SS and has pressure and temperatute ratings of 3000 psi
and 300°F. Multiple sample sizes (2 to 5 sl) are available by exchanging the shaft and seal
assembly. This valve is used primarily in conjunction with the AIC 10-port valve.
3.5.2 VALVE ACTUATION
Actuation of the liquid sample valve and the ten-port rotary valve is essentially identical in
Principal and mounting details. In both cases, actuation is accomplished through linkage
connected to a conventional air cylinder operated by air supplied from a four-way solenoid
valve. Pressure is applied alternately to opposite ends of the air cylinder (depending on
whether the four-way solenoid valve is energized or de-energized) with the other end vented
to atmosphere. When the solenoid valve is de-energized, pressure is applied to the rear of the
air cylinder and the cylinder piston arm or shaft is extended. When the coil is energized,
pressure is applied to the front chamber and the piston arm or shaft is retracted.
rII-83.6 FLOW RATE ADJUSTMENTS
As previously indicated, the flow of carrier gas through the system is one of the major
Parameters in proper operation of the analyzer. If accurate and repeatable results are to be
obtained, then this flow must be carefully controlled and regulated. Sensitivity, resolution,
and elution time are all critically affected by the cartier flow rate. Furthermore, each column
will have an optimum flow rate which will depend largely on the type of sample being run,
and the type and length of the column.
‘There are several methods of measuring the carrier flow rate, the most convenient being the
rotameter. Another, and a more accurate method, is to observe a moving soap bubble. With
this method a soap solution is placed at the bottom of a burette, and the carrier gas is
permitted to flow into the burette. The amount of time required for a soap bubble to move
‘over a measured volume will provide the necessary information for determining the flow rate
of the carrier gas. The procedures for adjusting carrier flow through the column, reference
flow, and precut carrier flow follow:
3.6.1 CARRIER FLOW
Carrier flow through the column should be set the first time before the analyzer is turned on.
If application engineering was done by AIC, simply adjust the carrier regulator until the
Pressure is the same as the column pressure for the specific application. Tum the analyzer on
and allow it to reach operating temperature. Attach a rotameter or bubblemeter to the
“column out” port and measure the flow. If it differs from the value given, adjust the carrier
regulator until the values are the same. When using a rotameter, make certain it is the correct
type. For example, if helium is the carrier, be certain to use a helium meter. Since helium is
such a light gas, an air oxygen rotameter would indicate that the cartier flow was much lower
than it really is.
III-103.6.2 REFERENCE CARRIER
‘The reference carrier is that portion of the carrier which flows across the two reference
detectors. is rate is adjusted by the metering valve located on the carrier regulator. As in
the case of the main carrier flow, the reference carrier should be adjusted the first time while
the analyzer is off. Adjust the main carrier flow first, then, making certain that the precut
metering valve is shut off, attach a rotameter to the vent port on the analyzer and adjust the
reference carrier valve until the rotameter indicates a flow of about 15 cc/min. Anytime the
main carrier flow is adjusted, it will be necessary to readjust the reference flow.
3.63 PRECUT CARRIER FLOW
A precut carrier is needed only when a single and precut column configuration is used. In
this case, the precut carrier's function is to backflush the precut column and push the
‘unwanted components to vent. As a result, it is necessary to measure the precut cartier flow
rate at the same vent port where the reference flow rate is measured. To accomplish this,
simply set the reference cartier flow rate, and then adjust the precut metering valve attached
to the carrier regulator until the sum of the reference and precut flow rates is the desired
value. For example, if reference flow is 15 cofmin., and the desired precut flow is 30
ccc/min., then the flow at the vent port should be 45 cofmin. As a matter of fact, this is the
flow which will normally be measured at the vent port since 30 cc/min. is the correct precut
cartier flow rate.
3.7 ROUTINE MAINTENANCE
‘There are no requirements for periodic maintenance on the Model 600 Analyzer. Whenever
convenient, however, the operator should make the following general checks in order to
ascertain that everything is functioning properly.
* Check the pressure on the carrier cylinder. If it is at or below 200 psig, then
the cylinder should be changed.
TIt-11+ Occasionally, check fittings for tightness.
+ The 10-port vaive is mounted on an aluminum channel which is, in tum,
secured to the analyzer mounting pan. ‘The operating of the valve could loosen
the mounting and it should be checked occasionally for tightness.
+ About once 2 month, check the pressure switch by dropping the pressure below
the pre-set point. The heater will go off. Then retum the air pressure to 50
psig; the heater will come on.
‘When, by abnormal functioning of the instrument, the need for more specific service becomes
evident, one of the following procedures should be observed.
3.7.1 VALVE MAINTENANCE - 10-PORT
A picture of the valve assembly and a detailed drawing of the rotary valve is given in Figure
IL-3 and Figure III-4, respectively. The valve cap assembly can be removed for inspection of
the valve surfaces by turning it in a counter clockwise direction. After the cap has been lifted
off, remove the upper rotary by pulling it straight out and off the shaft. The surface of the
lower valve is 440 stainless steel and the upper rotary is surfaced with glass filled Teflon.
Should cleaning be required they can be washed with a suitable solvent not harmful to these
materials,
Buffing The Upper Rotary
After long use (several months), the surface of the upper rotary becomes somewhat rough. In
order to prevent leaks, it becomes necessary to refinish the fluorogold surface. The procedure
is as follows:
* Obtain a sheet of 800 grit sandpaper.
I1I-12200A en ae
fe Oia] set NOs NBL
ova
as “oaqea sed Aawjox wod woy, p11] aunty
on sua
ayes
on
wa paren
Tornoe shaman sate OLY
(1) aw @ 4960 G Cy — =»
TL NE yee lt
cooo-fLay the sandpaper, grit side up, on a hard, smooth surface.
Place the fluorogold surface on the paper and make about four figure eights.
Do not press down hard.
‘Tum the paper over so that the grit side is down, and buff the upper rotary
until it is smooth.
Removing The Lower Valve
‘Tum off the instrument air.
Disconnect the two fittings to the air cylinder.
Disconnect all fittings to the lower valve.
Undo the four screws securing the aluminum channel to the analyzer pan.
Remove the assembly from the analyzer.
Remove the four screws securing the valve to the channel and lift the valve off.
‘Make certain to note the positioning of the valve before removing it. Port 1
should be located at the top of the bracket (Figure III-3).
Unscrew the cap and lift off the upper rotary.
If the lower valve is to be cleaned, remove all the fittings.
Reassembly
Replace the fittings in the ports of the lower valve. Make certain each port has
a Teflon gasket in it.
III-14Reattach the lower valve to the channel. Port 1 should be at the top.
Place the upper rotary on the valve. Make certain that one of the marks in the
rotary line up with the scribe above Port 1 on the valve.
Screw on the valve cap assembly. Hand tighten or until the bottom of the cap
is about 1/32” off the valve body, plus 1/4 tum with wrench.
Replace the assembly in the analyzer and reattach all lines according to the
column configuration diagram for the specific application.
Replacing The Air Cylinder (Figure H11-3)
‘Tum off the instrument air going to the analyzer.
Disconnect the two fittings to the air cylinder.
Loosen the set screw in the clevis.
Unscrew the pivot mount bolt. The cylinder will be free of the channel and
attached only by its shaft.
‘Tum the cylinder in a counter clockwise direction until the shaft is unscrewed
from the clevis.
Remove the cylinder from the analyzer.
Pop off the snap ring and remove the pivot mount from the cylinder.
Secure a new cylinder and attach the pivot mount to it.
III-15Screw the shaft of the cylinder into the clevis until the end of the shaft is flush
with the surface of the clevis.
Screw in the pivot mount bolt. Make certain the cylinder is positioned as
shown in Figure III-3.
Remove the valve cap assembly, leaving the upper rotary on the shaft, with the
air shaft of the cylinder extended.
If the line on the upper rotary does not line up to the scribe on the valve body,
then the pivot bolt must be removed and the shaft screwed into or out of the
clevis as necessary.
Once the adjustment is correct, tighten the clevis set screw.
Replace the valve cap.
3.7.2 REPLACING DETECTOR FILAMENTS
Depending on the current and temperature to which they are subjected, detector filaments
eventually become mismatched. This condition is revealed through baseline drift and inability
of the zero controls to adjust the baseline. The four filaments must be replaced as a set (AIC
Part # 10331-1). To replace the filaments, proceed as follows:
Tum the power off at the analyzer.
Remove the detector condulet cover.
Undo the four 1/16 Swagelok fittings from the bulicheads in the condulet
Disconnect the two shielded cables from the terminal board. Note where each.
III-16connects so that they can be reconnected properly.
Loosen and remove the two retainer screws in the front of the detector block.
Remove the detector block assembly.
‘Remove the filament leads from the terminal strip to which they are connected.
Loosen and remove the retaining mut on one of the filaments. Replace the old
filament with a new one, making certain to take the Teflon sleeving off the
Jeads of the old filament and put them on the new filament. Tighten down the
retaining mut.
Repeat for the other three filaments.
With a meter, check the leads for shorts to the detector block.
Attach the filament leads to the correct positions on the terminal board.
Consult the analyzer schematic if there is any doubt.
Replace the detector block in the condulet. Connect the two cables and the
‘Swagelok fittings. Replace the detector condulet cover and place the unit back
in operation.
3.73 COLUMN MAINTENANCE
A contaminated or failing column normally manifests itself as a drifting baseline, a high
baseline, or an unstable baseline. Once a malfunctioning column has been detected, it can be
regenerated by flushing with carrier gas at the operating temperature, or, this failing, it can be
replaced. The procedure for replacement follows:
III-17