You are on page 1of 2

Lot 59

Antibacterial and antiviral filter for mechanical ventilation (MN 136).

Specifications YES/NO
The circuits, the ventilator and the environment must be protected against
contamination generated by the patient.
The patient should be protected against nosocomial infections.
The filter must meet the following characteristics:
Transparent, made of polypropylene, made without latex.
With filter medium made of synthetic fiber mixture.
With carbon dioxide sampling port, provided with valve, protective cover fixed
with support band, and luer-lock connector.
With 15mm / 22mm conical connectors according to ISO 5356-1 standard.
Antibacterial and antiviral efficiency ≥ 99.9999%. YES
Proven efficacy against HIV, HCV, M. TUBERCULOSIS, ms2 Virus,
Pseudomonas spp, Serratia marcescens, prions.
It must be able to be used for a volume of air flow between 90ml and 1500ml.
Dead space of maximum 25 ml.
Weight up to 25 g.
Resistance to filtration: <100 Pa, at a flow of 30 L / min.
It must be possible to use it for 24 hours, unclogable during this period.
Individual packaging.
Sterile packaged product.
Packaging and labeling according to EU-MDD (EU Medical Device Directive)
(product name, description, code, validity, batch, sterilization method,
dimensions, manufacturer's address, etc.). CE mark.

Lot 141

Antibacterial filter with humidification (MNb137 with paper)

Specifications YES/NO
Disposable filter for adults.
Inside to present electrostatic filtration medium and hydrophobic humidification
and heating medium made of folded paper.
For the connection between the circuit and the catheter mount, at the end close to
the patient.
Connectors 22M / 15F and 22F / 15M, according to ISO 5356-1, with the end
close to the patient clearly inscribed.
Transparent polypropylene case, which allows visual confirmation of efficiency..
Does not contain latex
Assembled without adhesives.
Duration of use up to 48 hours.
Tidal volume between 200-1500ml. 150 - 1500
Maximum weight 30 g. YES
Maximum dead space 55 ml. YES
Maximum resistance 105Pa at 30lpm and 160Pa at 60lpm.
Humidification at 500ml - minimum 30 mg / l.
Minimum filtration area of 27 cm2.
The water loss at a volume of 500ml should be a maximum of 12 mg / Liter.
The filter should be non-pyrogenic, with a maximum value of 0.25 Eu / ml.
To be in accordance with the ISO 10993-1 standard regarding biocompatibility.
Bacteria and virus filtration efficiency ≥ 99.9998%. YES
Sodium chloride filtration efficiency ≥ 99.5%.
Port for carbon dioxide sampling, provided with valve, protective cover fixed
with support band, and luer-lock connector.
Individual packaging.
Sterile packaged product.
Packaging and labeling according to EU-MDD (EU Medical Device Directive)
(product name, description, code, validity, batch, sterilization method,
dimensions, manufacturer's address, etc.). CE mark.

Lot 399

Disposable mechanical ventilation circuit (MN 178-01 for circuit and MN 135-02 for
humidification chamber)

Specifications YES/NO
Made out of:
1 piece neonatal patient ventilation circuit set
heated inhale limb to reduce condensation YES
heated, robust exhale limb, made out of material with removal of condensation
properties by evaporating water before condensation
without water traps YES
heating resistance inserted in the spiral
for flow rates> 4L / min
circuit length: minimum 1.5m YES
compatible with Fisher & Paykel MR850 humidifier
1 piece humidification chamber
Compatible with Fisher & Paykel MR850 humidifier
Includes self-filling system YES
Double floating mechanism to ensure a constant level of water inside the room
Cylindrical shape with slightly rounded edges at the top to minimize
condensation
Compressible volume: min 280ml
Flow resistance at 60l / min: 0.5 cmH2O (0.5 mbar)
22 mm ports
Water supply hose included, with a minimum length of 50 cm

Lot 400
NEOPUFF circuit (MN 108)

Specifications YES/NO
Disposable
Compatible with Fisher & Paykel Neopuff system.
Circuit with T-piece
Mask of different sizes (35mm, 42mm, 50mm, 60mm, 72mm)

You might also like