The Fersenius kabi COM.TEC cell separator System will perform all types of apheresis
procedures. One of its main procedures is to collect single donor platelets. Because of the
unique abilities of this device, the procedures are individualized by the control of several
algorithms. Prior to the beginning of the run, a platelet yield is predicted using the entered
platelet count. Based upon this information, changes can be made in the operating parameters
to optimize the collection. Leuko-poor thrombocytes are produced which can be stored for 5
days. The collect concentration monitor, an on-line device, provides information on the use of
thrombocyte concentration and yield, without the use of probes.
I. PRINCIPLE:
The system draws whole blood from a donor or patient, adds anticoagulant, separates the
blood components, collects or removes specific components and returns uncollected
components to the donor or patient. In therapeutic plasma exchange and red blood cell
exchange procedures, appropriate replacement fluid is continuously returned.
II. Specifications:
Five (5) self-loading peristaltic pumps with one of them for plasma recirculation inside
the separation chamber in order to stabilize the interface and regulate the hematocrit
on it, in productive procedures.
With online help to advice the operator about the steps to follow.
With dedicated printer (included) to register all the events occurred during the
procedure. With comprehensive printout for reporting.
Double needle Patient/Donor vein access, with single needle capabilities in selected
procedures.
Exchangeable printed layout to help the operator on the mounting of the different kits
Automatic saline perfusion system in case that procedure holds, or type “Keep Vein
Open” system.
Hemolysis/Red cell sensor for Plasma line monitoring and Platelet’s sensor on Cell line.
Wireless connection capabilities to be linked with dedicated data management
software.
Eleven (11) treatment and productivity programs present, with the following
characteristics:
a) Platelet:
1. Double needle PLT collection with Five (5) days storage bags.
2. Single needle PLT collection with Five (5) days storage bags.
b) Stem Cells:
6. Granulocyte collection.
7. Bone Marrow stem cell processing with automatic transfer between the donation and
reinfusion bag (BMSC).
All Stem Cell procedures with CD34+ cell estimation and procedure time
forecast.
c) Therapy:
10. Therapeutic Plasma exchange with isovolemic and automatic fluid balancing and Buffy
Coat reinfusion.
Disposable kit
Gloves
ACDA
Normal Saline
Fistula Needle
Physio Ball Hand exerciser
Surgical tape
Emergency kit
V. Operating Procedure:
Check power supply.
Clean all sensor and valves with gauze piece.
Switch on power button by press button 1.
Select program C5L for Platelet pheresis.
Install C5L
Open pump lids.
Hang up blood lines.
Close red inlet clamp just beyond Y piece to the pre sampling clamp.
Close white needle clamp.
Hang up bags and containers.
Close clamp between concentrate bags.
Close clamp at pc sample bag.
Insert four pump segments hit<Turn pumps>.
Close pump
Insert drip chamber in air detector.
Insert plasma line in valve 4.
Insert plasma line Hemolysis detector.
Insert diversion line in valve 5.
Insert return line in valve 1.
Insert inlet line to cell pump in cell detector.
Insert ACD drip chamber in ACD detector.
Insert ACD pump segment.
Insert red saline line in valve 2.
Insert blue line in valve 3.
Connect the pressure lines.
Insert separation chamber.
Lock housing and rotor.
Insert centrifuge lines in tubing guide (hook at rotor).
Close centrifuge door.
Insert air protector in holder.
Open bags and containers.
Set saline and ACD level.
Blue return line clamp opened.
Prepare priming.
Start priming.
Feed donor Data gender, Hight, weight, Hematocrit and platelet count.
Select program and platelet yield.
Prepare donor.
Connect inlet and return pressure lines.
Press continue to start draw.
When the required volume obtained disconnect inlet pressure line and platelet
collection bag and rinse back.
After the rinse back complete disconnect the return line and apply bandage.
Remove the tubing set and discard.