Service Mode Babylog 8000

Table of Contents:

1 General description......................................................................................................... 3
1.1 Calling up the DS-Mode............................................................................................... 3
1.2 Left the Service-Mode.................................................................................................. 3
2 Structure of the service mode ......................................................................................... 4
2.1 Error list ....................................................................................................................... 4
2.2 Electronics ................................................................................................................... 4
2.3 Pneumatics .................................................................................................................. 4
3 Service Mode "Error list" ................................................................................................. 5
3.1 Error list ....................................................................................................................... 5
3.2 Reading out vector....................................................................................................... 5
4 Service Mode "User Interface" ........................................................................................ 6
4.1 Operating hours ........................................................................................................... 6
4.2 Output 02 sensor voltage with load test ....................................................................... 6
4.3 Output of storage battery voltage with load test........................................................... 7
4.4 Keyboard test............................................................................................................... 7
4.5 LED test ....................................................................................................................... 7
4.6 Display test .................................................................................................................. 7
4.7 Potentiometer test........................................................................................................ 7
4.8 Reading out jumper configuration ................................................................................ 8
4.8.1 Devices with software version 5.n ......................................................................... 8
4.8.2 Devices with software version 4.n ......................................................................... 9
4.8.3 Devices with software version 3.n and 2.n .......................................................... 10
5 Service Mode "PC Board" ............................................................................................. 11
5.1 Monitoring PCB.......................................................................................................... 11
5.1.1 02 sensor voltage................................................................................................. 11
5.1.2 Battery storage voltage ....................................................................................... 11
5.1.3 lnspiratory pressure............................................................................................. 12
5.1.4 Valve supply voltage = Vent +27V ...................................................................... 12

6173.3 Babylog 8000 Service Mode March 2000 1/30


5.1.5 Monitoring valve supply voltage .......................................................................... 12
5.1.6 Device temperature ............................................................................................. 13
5.2 Front Controller PCB ................................................................................................. 13
5.3 Flow PCB................................................................................................................... 14
5.3.1 Testing flow measurement .................................................................................. 14
5.4 I/O PCB...................................................................................................................... 16
5.4.1 02 voltage, 02 sensor voltage............................................................................... 16
5.4.2 Insp. and exp. pressure....................................................................................... 16
5.4.3 Control pressure sensors for air and oxygen....................................................... 17
5.4.4 Test channel X7 .................................................................................................. 17
5.4.5 Test channels for the A/D-converter measurement section................................. 17
5.5 Communication PCB / Testing RS232....................................................................... 18
6 Service Mode "Pneumatics".......................................................................................... 19
6.1 Service mode pneumatics “Measuring” ..................................................................... 19
6.1.1 Flow measurement values .................................................................................. 19
6.1.2 02 sensor voltages............................................................................................... 19
6.1.3 lnspiratory pressure............................................................................................. 20
6.1.4 Expiratory pressure ............................................................................................. 20
6.1.5 Control pressure AIR........................................................................................... 21
6.1.6 Control pressure O2............................................................................................. 21
6.1.7 P-bias (zero potential) of the airway pressure sensors ....................................... 21
6.1.8 Calibrating O2 and flow sensor ............................................................................ 22
6.2 Service mode pneumatics “Set. 1”............................................................................. 22
6.2.1 Testing the pressure regulators and input pressure sensors............................... 22
6.2.2 Testing the PEEP/PIP valve and current source ................................................. 22
6.2.3 Testing current source to PEEP/PIP valve .......................................................... 27
6.2.4 Testing airway pressure sensors......................................................................... 27
6.3 Service pneumatics "Set. 2"....................................................................................... 28
6.3.1 Valve bank (mixer) .............................................................................................. 28
6.3.2 Ejector valve Y1.3 and ejector............................................................................. 28
6.3.3 O2/Air switch over valve Y1.1, O2 calibration valve Y1.3 and O2 calibration valves
Y3.1 and Y3.2 in the inspiratory block........................................................................... 28
6.3.4 Safety valve Y1.2 ................................................................................................ 29
7 Special characteristics of micro processor 68000 ......................................................... 30

6173.3 Babylog 8000 Service Mode March 2000 2/30


1 General description
The Service-Mode (DS-Mode) enables the device error messages, which have been saved, to be
read out; with the frequency of errors, the read out of all important measurement values and the
targeted drive of D/A-converters and valves.
Operation takes place by means of a menu control, as is the case for the operating mode. The
individual functions are called up by softkeys. The references, warnings and alarms of the operating
mode are also active in the DS mode and can be ignored with the "Ok", "Confirm" or "Reset/Check"
key. Ignored messages are displayed again after 30 seconds if the reason for this still exists.

1.1 Calling up the DS-Mode


Do not connect any patient circuit system to the insp. and exp. nozzles during call up of diagnosis
mode.

Connect compressed O2 and Air supply (3 to 6 bar / 40 to 87 psi).

Enter service diagnosis mode:

Babylog with SW 5.0:


Press "Confirm" or "OK" and "Alarm silence" keys and switch on device simultaneously. Keep
both keys pressed until first service menu is displayed.

Babylog with SW 2.n, 3.n or 4.n:


Press "Reset/Check" and "Alarm silence" keys and switch on device simultaneously. Keep keys
pressed until first service menu is displayed.

Operation in the service mode is to be confirmed in each case within 2 minutes using the key
"Alarm silence", otherwise restart with error message "823".

The errors "002", "800", "801", "804" "805" or "823" can be ignored in the service mode.

Note: The units of the measurement values can be partially changed in some menus with the
"Cal. Config." (SW 5) or "Monitoring" (SW 2, 3 or 4) key between:

• hexadecimal values
• decimal values
• voltages or standard units (e.g. voltage in mV/V or pressure in mbar/bar/psi/cmH2)

1.2 Left the Service-Mode


The DS-Mode can be left by switching off the Babylog 8000.

6173.3 Babylog 8000 Service Mode March 2000 3/30


2 Structure of the service mode
The service mode includes the Error list, Electronics service mode and Pneumatics service mode.

2.1 Error list


→ Error List → Decrease error no.
→ Increase error no.
→ Clear all errors
→ Vector

2.2 Electronics
→ Electronics → User Interface → 1 Operational data
→ 2 Keyboard and display
→ 3 Potentiometer test
→ 4 Option switches

→ PC Boards → 6 PCB Monitor


→ 7 PCB Flow
→ 8 PCB Front controller
→ 9 PCB I/O
→ 10 PCB Communication

2.3 Pneumatics
→ Pneumatics → Measuring → V-Cal. (flow sensor calibration)
→ 02 -Cal. (O2 sensor calibration)
→ Pressure P-bias (zero potential)

→ Set. 1 → Decrease pressure


→ Increase pressure
→ Potentiometer Fi02
→ Potentiometer Insp. Flow

→ Set. 2 → Valve selection left


→ Valve selection right
→ Selected valve on/off

6173.3 Babylog 8000 Service Mode March 2000 4/30


3 Service Mode "Error list"
With this function the device errors saved can be read out with the frequency of errors and the error
memory can be cleared. The status of the processor 68000 at the last processor error is saved with
its address in the vector, An error analysis, under some circumstances can be carried out using this
data following a system failure. An important and helpful component in troubleshooting is the error
list of the unit. There are two possible ways to read out error messages:

3.1 Error list


Babylog is in the service mode, press softkeys "Error list".

Use softkeys " ⇓ " and " ⇑ " to read out the error list.

See "Error List" of Babylog 8000 for detailed descriptions of the errors.

Use softkey "Clear" to erase the error list.

Press softkey "Return ".

3.2 Reading out vector


Babylog is in the service mode, press softkeys "Error list" and then "Vector".

Any deviations of the normal operation of the processor 68000 due to errors or error conditions e.g.
bus errors, RESET, address error and interrupts result in the processor carrying out an exception
condition and loads a vector table with a hexadecimal number. This vector number not only informs
the processor that an error or exception has arisen but, in addition to this, also tells it how to react.
This vector is stored with other processor data and can be read out and cleared in the
Babylog 8000.

Note: The vectors with processor data are principally provided to assist in the development
department and electronic workshop.

Use softkey "Clear" to erase the vector list.

Press softkey "Return " twice.

6173.3 Babylog 8000 Service Mode March 2000 5/30


4 Service Mode "User Interface"

4.1 Operating hours


Babylog is in the service mode, press softkeys "Electronics", "User Interface" and then
"1 Operational data".

Reading out the operating hours counter with the possibility of setting. Before renewing the Time-
Keeper RAM on the PCB CPU (every 4 years) this value is to be read out and after renewal to be re-
entered in the Babylog 8000.

Press softkey "Return " three times.

4.2 Output 02 sensor voltage with load test


Babylog is in the service mode, press softkeys "Electronics", "User Interface" and then
"1 Operational data".

Two voltages are read out:

• U1 Voltage from the first half of the 02 sensor, read out from master through the PCB I/O and
PCB 02 amplifier.

• U2 Voltage from the second half of the 02 sensor, read out from the PCB Monitor through the
PCB 02 amplifier.

Desired values at 21 Vol. % 02: U1 = U2 = 9.2 mV up to 26.0 mV


Desired values at 100 Vol. % 02: U1 = U2 = 43.6 mV up to 123.6 mV
Difference in voltage permitted: Smaller (U1 + U2) / 8 + 1 mV

Note: The 02 concentration at the sensor is undefined and matches the 0 2 setting before the device
was last switched off. Specification of set values through the pneumatics menu.

Load test by pressing softkey "O2Test":


On the PCB 02 amplifier, the master channel of the 02 sensor is pulled down as a result, the voltage
U1 is noticeable reduced (recognition two channel Dräger 02 sensor). If the voltage U2 is also
reduced, the sensor is not a Dräger sensor (not permissible).

Press softkey "Return " three times.

6173.3 Babylog 8000 Service Mode March 2000 6/30


4.3 Output of storage battery voltage with load test
Babylog is in the service mode, press softkeys "Electronics", "User Interface" and then
"1 Operational data".

Note: The storage battery remains loaded for 60 seconds after the key "BaTest" (battery test) is
pressed. Desired value for charged battery storage unloaded: U ≥ 9.2 V. When loaded the voltage
only sinks by 0.1 V to 0.2 V for a charged storage battery.

Press softkey "Return " three times.

4.4 Keyboard test


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

4.5 LED test


Babylog is in the service mode, press softkeys "Electronics", "User Interface", "2 Operational
data", and then "Keyboard and display" and then "LEDs".

In the LED display test all LEDs apart from the alarm LEDs are driven one after another and then
simultaneously.

Press softkey "Return " four times.

4.6 Display test


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

4.7 Potentiometer test


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

6173.3 Babylog 8000 Service Mode March 2000 7/30


4.8 Reading out jumper configuration
Note: Check for proper configuration in line with installed software version, for software version 2.n,
3.n and 4.n also in line with present scaling of PEEP/CPAP potentiometer.

Info about different versions of the Babylog 8000:

84 10 280 US version For USA and Canada


84 11 111 Canadian version For USA and Canada
84 10 564 Japanese Version For Japan
84 10 270 French version For France
84 09 200 / 84 10 300 International versions All other regions

Note: All devices in the USA, Canada and Japan require the option jumper X8/0 (limitation to 150
bpm) independent from the installed software version!

- = option jumper not fitted


 = option jumper fitted

All jumpers are located on the Front Controller PCB (devices with LCD) or Front PCB (devices with
ELD).

4.8.1 Devices with software version 5.n


Babylog is in the service mode, press softkeys "Electronics", "User interface" and then "4 Option
switches".

• All versions 84 10 280 (see rating plate under “Typ”) with software version 5.0:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - - -  - 
• All versions 84 11 111 and 84 10 564 (see rating plate under “Typ”) with software version 5.0:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - -   - 
• All versions 84 09 200, 84 10 300 and 84 10 270 (see rating plate under “Typ”) with
software version 5.0:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - -   - -

6173.3 Babylog 8000 Service Mode March 2000 8/30


Press softkey "Return " three times.

4.8.2 Devices with software version 4.n


Babylog is in the service mode, press softkeys "Electronics", "User interface" and then "4 Option
switches".

Note: This software version is not approved in the USA!

• All versions 84 10 280, 84 11 111 and 84 10 564 (see rating plate under “Typ”) with
software version 4.n:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - - X X - 
Option switches PCB Front X8:

Bit 0 = fitted
Bit X = Option switches 2 and 3, jumper fitted or not fitted depending on scaling of
PEEP/CPAP potentiometer:

PEEP/CPAP = 0, 5, 10, 15 ⇒ 2 not fitted, 3 not fitted


PEEP/CPAP = 0, 3, 6, 9, 12, 15 ⇒ 2 fitted, 3 not fitted
PEEP/CPAP = 0 to 25 ⇒ 2 and 3 fitted

• All versions 84 09 200, 84 10 300 and 84 10 270 (see rating plate under “Typ”) with
software version 4.n:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - - X X - -

Option switches PCB Front X8:

Bit 0 = not fitted


Bit X = Option switches 2 and 3, jumper fitted or not fitted depending on scaling of
PEEP/CPAP potentiometer:

PEEP/CPAP = 0, 5, 10, 15 ⇒ 2 not fitted, 3 not fitted


PEEP/CPAP = 0, 3, 6, 9, 12, 15 ⇒ 2 fitted, 3 not fitted
PEEP/CPAP = 0 to 25 ⇒ 2 and 3 fitted

Press softkey "Return " three times.

6173.3 Babylog 8000 Service Mode March 2000 9/30


4.8.3 Devices with software version 3.n and 2.n
Babylog is in the service mode, press softkeys "Electronics", "User interface" and then "4 Option
switches".

• All versions 84 10 280, 84 11 111 and 84 10 564 (see rating plate under “Typ”) with
software version 3.n or 2.n:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - - - X - 
Option switches PCB Front X8:

Bit 0 = fitted
Bit X = Option switch 2, jumper fitted or not fitted depending on scaling of
PEEP/CPAP potentiometer:

PEEP/CPAP = 0, 5, 10, 15 ⇒ 2 not fitted


PEEP/CPAP = 0, 3, 6, 9, 12, 15 ⇒ 2 fitted

• All versions 84 09 200, 84 10 300 and 84 10 270 (see rating plate under “Typ”) with
software version 3.n or 2.n:

Location 7 6 5 4 3 2 1 0 Bit DSMode


PCB Front X7 - -
PCB Front X8 - - - - - X - -

Option switches PCB Front X8:

Bit 0 = not fitted


Bit X = Option switch 2, jumper fitted or not fitted depending on scaling of
PEEP/CPAP potentiometer:

PEEP/CPAP = 0, 5, 10, 15 ⇒ 2 not fitted


PEEP/CPAP = 0, 3, 6, 9, 12, 15 ⇒ 2 fitted

Press softkey "Return " three times.

6173.3 Babylog 8000 Service Mode March 2000 10/30


5 Service Mode "PC Board"
The respective measurement data from the most important PC boards are issued. In addition to this,
the possibility exists of presetting the desired values for the D/A-converter.

5.1 Monitoring PCB


Reading out of all important values from the Monitoring PCB.

The output is divided into two screens:

• Potentiometer values: For a detailed test instruction refer to "Preventive Maintenance" (Test
Certificate).

• Other values: Values of all other input channels.

5.1.1 02 sensor voltage


Babylog is in the service mode, press softkeys "Electronics", "PC Boards", "6 PCB Monitor" and
then "Other values".

Voltage from the second half of the 02 sensor, read from the Monitoring PCB through the 02 amplifier
PCB. The display value is converted to the sensor voltage.

Desired values at 21 vol. % 02: U = 9.2 mV up to 26.0 mV


Desired values at 100 vol. % 02: U = 43.6 mV up to 123.6 mV

Note:
The 02 concentration at the sensor is undefined and is equivalent to the 02 setting before the device
was last switched off. Presetting of defined values via the pneumatics menu.

Press softkey "Return " four times.

5.1.2 Battery storage voltage


Babylog is in the service mode, press softkeys "Electronics", "PC Boards", "6 PCB Monitor" and
then "Other values".

Desired value for charged storage battery ≥ 9.2 V

Press softkey "Return " four times.

6173.3 Babylog 8000 Service Mode March 2000 11/30


5.1.3 lnspiratory pressure
Babylog is in the service mode, press softkeys "Electronics", "PC Boards", "6 PCB Monitor" and
then "Other values".

Standard value from the inspiratory pressure sensor.

For a detailed test instruction of the airway pressure measurement refer to "Preventive
Maintenance" (Test Certificate).

Press softkey "Return " four times.

5.1.4 Valve supply voltage = Vent +27V


Babylog is in the service mode, press softkeys "Electronics", "PC Boards", "6 PCB Monitor" and
then "Other values".

For Monitoring PCB with Watch-dog PCB: desired value = 27 V


For Monitoring PCB without Watch-dog PCB (from 1991/1992): Not relevant.

Press softkey "Return " four times.

5.1.5 Monitoring valve supply voltage


Babylog is in the service mode, press softkeys "Electronics", "PC Boards", "6 PCB Monitor" and
then "Other values".

If the voltage exists after the safety relays K1 and K2 on the Pneumatics drive PCB the relay
contacts are closed.

Display value Meaning


>3.2V Relays and supply voltage OK
1 V to 3.2 V Valve supply voltage defective (mains +27 V)
< l.0V Relays open-circuited

Test: The relays are open-circuited if the pressure is greater than -5 mbar produced at the
inspiratory pressure sensor.

Press softkey "Return " four times.

6173.3 Babylog 8000 Service Mode March 2000 12/30


5.1.6 Device temperature
Babylog is in the service mode, press softkeys "Electronics", "PC Boards", "6 PCB Monitor" and
then "Other values".

The temperature sensor is fixed on the PC board. For values lower -20 °C or higher +100 °C the
measurement fails.

Press softkey "Return " four times.

5.2 Front Controller PCB


This test function enables eight AD channels to be read out and desired values for the contrast
setting (view angle) of the LCD display by means of a D/A-converter. The eight channels are:
• 6 potentiometer voltages
• dynamic A/D-converter test channel, can be switched to two test voltages
• the Offset (GND - reference voltage).

Babylog is in the service mode, press softkeys "Electronics", "PC Boards" and then "8 PCB Front
controller".

Desired values potentiometer voltages: For a detailed test instruction refer to "Preventive
Maintenance" (Test Certificate).

Desired values test channel: The voltage can be switched from

U = 128 dec ± 10 dec, equivalent to 5.02 V (±0.4 V) to


U = 191 dec ± 10 dec, equivalent to 7.49 V (±0.4 V)

Desired value offset (GND reference channel):

U = 00 dec ±2 dec, equivalent to 0.00 V (±0.08 V)

The contrast setting (observation angle) can be set from

Umin = 200 dec, equivalent 7.84 V (down) up to


Umax = 255 dec, equivalent to 10.0 V (up)

Press softkey "Return " three times.

6173.3 Babylog 8000 Service Mode March 2000 13/30


5.3 Flow PCB
Both measuring wires in the flow sensor are evaluated by two independent measurement channels
for the Flow PCB. One measuring wire measure the amount of flow and the other measure the
recognition of the flow direction.

Four measurement values are read out:


• Sensor 1 Flow voltage of the direction recognition
• Sensor 2 Flow voltage of the measurement value
• Current limit Reference voltage, can be changed to test the current limit
• Ground Monitoring GND

The 12 bit D/A-converters provided for adjustment of the bridges can be driven together. This is a
manual adjustment and test of the measurement section.

The adjustment can also be carried out automatically as with the operating mode but then no
desired valued is issued for the A/D-converters.

Test measurement section sensor - sensor cable - Flow PCB.

Preliminary testing of the flow sensor:


Measurement of the resistance of the measuring wires at room temperature:
Testing value: R = 2.2 Ohm (±10%)

5.3.1 Testing flow measurement


Switch Babylog off.

Connect flow sensor and sensor cable so that they are ready for operation.

Enter service mode, press softkeys "Electronics", "PC Boards" and then "7 PCB Flow".

Set units of the measured values to decimal values by pressing the key "Cal. Config." (SW 5) or
"Monitoring" (SW 2, 3 or 4).

• Test value current limit "CurLim1" = 982 dec up to 1228 dec


Note: Voltage is generated from the A/D-converter reference voltage.

Possible error 1:

Reference voltage for A/D-converter out of tolerance.

Press softkey "Test" (the sensor may not be adjusted):

• Test value current limit "CurLim1" = 1525 dec up to 1638 dec

6173.3 Babylog 8000 Service Mode March 2000 14/30


Press softkey "Test" again.

Possible error 2:

Test circuit, A/D-conversion or reference voltage out of tolerance.

Set units of the measured values to volts pressing the key "Cal. Config." (SW 5) or
"Monitoring" (SW 2, 3 or 4).

The sensor voltages sensor 1 and sensor 2 must be able to be set to U = 4.37 ±0.1 V (equivalent
to 1793 dec) with softkeys " ⇓ " and " ⇑ " for adjusting the desired values.

Possible error 3:

The Voltage does not increase although the resistance value of the sensor is within the
tolerance range:

• High resistivity cable or measurement channel defective.

Test: Create a short circuit at the sensor socket at the cable between the contacts opposite
each other and at the socket on the device between contacts 1 and 2 (sensor 1) and 3 and 4
(sensor 2) (pin 1 upper right, counter direction anti-clockwise).
The voltage must be > 7 V each.

• Adjustment voltage not reached:

Resistance potentiometer wire in the flow sensor is outside the permissible tolerance
or the measurement channel is defective.

• Voltage on maximum (normal > 7 V):

Short circuit cable (remove plug) or measurement channel defective.

Press softkey "Return " three times.

6173.3 Babylog 8000 Service Mode March 2000 15/30


5.4 I/O PCB
The following measurement values are issued:

5.4.1 02 voltage, 02 sensor voltage


02 voltage of the first half of the 02 sensor is measured via the 02 amplifiers PCB. The output value
(units: decimal, hexadecimal or volts) is equivalent to the input voltage of the PCB. Whereas, an
output value in standardized units (e.g. mV) represents the voltage at the sensor.

Enter service mode, press softkeys "Electronics", "PC Boards" and then "9 PCB I/O".

Set units of the measured values to volts pressing the key "Cal. Config." (SW 5) or "Monitoring"
(SW 2, 3 or 4).

Desired values at 21 vol. % 02:

U = 9.2 mV up to 26.0 mV (standardized sensor voltage)


U = 0.6999 V up to 1.979 V (input voltage I/O PCB)

Desired values at 100 vol. % 02:

U = 43.6 mV up to 123.6 mV (standardized voltage)

Note: The 02 concentration at the sensor is undefined and is equivalent to the 0 2 setting before the
device was last switched off. Presetting of the set values occurs via the pneumatics menu.

Press softkey "Return " three times.

5.4.2 Insp. and exp. pressure


Enter service mode, press softkeys "Electronics", "PC Boards" and then "9 PCB I/O".

Set units of the measured values can be changed by pressing the key "Cal. Config." (SW 5) or
"Monitoring" (SW 2, 3 or 4).

Measure current values of the airway pressure sensors. The output value (units: decimal,
hexadecimal or volts) is equivalent to the input voltage of the PCB.

Whereas, an output value in standardized units (e.g. mbar, cmH20) represents the voltage at the
sensor.

Technical data respiratory pressure sensors:

Relative pressure sensors

6173.3 Babylog 8000 Service Mode March 2000 16/30


Sensitivity: 36.5 mV/mbar (±0.3 mV/mbar)

Offset: 1.740 V +0.12 V / −0.08V

For a detailed test instruction of the airway measurement refer to "Periodic Manufacturer's Service"
(Test Certificate).

Press softkey "Return " three times.

5.4.3 Control pressure sensors for air and oxygen


Enter service mode, press softkeys "Electronics", "PC Boards" and then "9 PCB I/O".

Set units of the measured values can be changed by pressing the key "Cal. Config." (SW 5) or
"Monitoring" (SW 2, 3 or 4).

Measuring current values of the control pressure sensors. The output value (units: decimal,
hexadecimal or volts) is equivalent to the input voltage of the PC board. Whereas, an output value in
standardized units (e.g. bar, psi) represents the voltage at the sensor.

Technical data control pressure sensors:

Absolute pressure sensor

Sensitivity: 1.5 V/bar (±0.008 V/bar)

Offset: 1.0 V (±0.1 V)

For a detailed test instruction refer to "Preventive Maintenance" (Test Certificate).

Press softkey "Return " three times.

5.4.4 Test channel X7


Enter service mode, press softkeys "Electronics", "PC Boards" and then "9 PCB I/O".

Set unit of the measured value can be changed by pressing the key "Cal. Config." (SW 5) or
"Monitoring" (SW 2, 3 or 4).

Test channel X7 is not used.

Press softkey "Return " three times.

5.4.5 Test channels for the A/D-converter measurement section.


Enter service mode, press softkeys "Electronics", "PC Boards" and then "9 PCB I/O".

6173.3 Babylog 8000 Service Mode March 2000 17/30


Set unit of the measured value can be changed by pressing the key "Cal. Config." (SW 5) or
"Monitoring" (SW 2, 3 or 4).

The input voltages can be switched with the “Toggle” key.

Desired values for D2 and D3:

U = 1642 dec up to 2252 dec = 4.50 V up to 5.50 V and


U = 2866 dec up to 3276 dec = 7.00 V up to 8.00 V

Press softkey "Return " three times.

5.5 Communication PCB / Testing RS232


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

6173.3 Babylog 8000 Service Mode March 2000 18/30


6 Service Mode "Pneumatics"
The pneumatics menu consists of the “Measuring” part, the “Set. 1” part, and the “Set. 2” part.

6.1 Service mode pneumatics “Measuring”


Reading out the current measurement values of the sensors in menu "Meas.". By means of the
menus “Set. 1” and “Set. 2” desired values can be specified.

6.1.1 Flow measurement values


Enter service mode, press softkeys "Pneumatics" and then "Meas.".

Flow in L/min with flow direction (positive for direction to patient, negative for direction to device).

Technical data for measuring:

Measurement range: up to 30 L/min

Accuracy: ±10 % of the measured value

Lower limit of the recognition direction:


300 mL/min with SW 2.00, 250 mL/min as of SW 3.n and higher

Tidal volume: Measurement range 0 to 999 mL

Press softkey "Return " twice.

6.1.2 02 sensor voltages


Test set-up:

Babylog without patient circuit system.

Connect O2 and Air pressure supply to Babylog 8000 (40 to 87 psi / 3 to 6 bar).

Babylog is in the service mode, press softkeys "Pneumatics" and then "Set. 1".

Settings in menu "Set. 1":

O2-conc. = 21
Insp. Flow = 5
Pinsp = 80

Press softkey "Return " and then "Meas.".

6173.3 Babylog 8000 Service Mode March 2000 19/30


Two sensor voltages are read out:

U1: Voltage of the first half of the 02 sensor, read out from the master by means of the I/O PCB and
02 amplifier PCB.
U2: Voltage of the second half of the 02 sensor, read out from the Monitoring PCB by means of the
02 amplifier PCB:

• desired values at 21 vol. % 02: U1 = U2 = 9.2 mV up to 26.0 mV


• Difference in voltage permitted: Lower (U1 + U2) / 8 + 1 mV

Press softkey "Return " and then "Set. 1".

Settings in menu "Set. 1":

O2-conc. = 100
Insp. Flow = 5
Pinsp = 80

Press softkey "Return " and then "Meas.".

Two sensor voltages are read out:

• desired values at 100 vol. % 02: U1 = U2 = 43.6 mV up to 123.6 mV


• Difference in voltage permitted: Lower (U1 + U2) / 8 + 1 mV

Press key "Return " twice

6.1.3 lnspiratory pressure


Enter service mode, press softkeys "Pneumatics" and then "Meas.".

Relative pressure from the inspiratory pressure sensor, the first value (master measurement) read
out from the I/O PCB and the second measurement value (monitor measurement) read out from the
Monitoring PCB.

Accuracy for master: Measured value ± 3% ±1 mbar.

For a detailed test instruction refer to "Preventive Maintenance" (Test Certificate).

Press softkey "Return " twice.

6.1.4 Expiratory pressure


Enter service mode, press softkeys "Pneumatics" and then "Meas.".

Relative pressure from the expiratory pressure sensor.

6173.3 Babylog 8000 Service Mode March 2000 20/30


Accuracy: Measured value ± 3% ±1 mbar

For a detailed test instruction refer to "Preventive Maintenance" (Test Certificate).

Press softkey "Return " twice.

6.1.5 Control pressure AIR


Enter service mode, press softkeys "Pneumatics" and then "Meas.".

Absolute pressure measured behind the pressure reducer AIR at the valve block. The value
displayed in the DS Mode is the absolute pressure. The atmospheric pressure must be subtracted
from this absolute pressure. The result is the relative pressure.

Example:

Displayed value Pair is 2.76 bar, atmospheric pressure is 1.00 bar.

Relative pressure = displayed value – atmospheric pressure = 2,76bar –1.00 bar = 1.76bar

For a detailed test instruction refer to "Preventive Maintenance" (Test Certificate).

Press softkey "Return " twice.

6.1.6 Control pressure O2


Enter service mode, press softkeys "Pneumatics" and then "Meas.".

Absolute pressure measured behind the pressure reducer O2 at the valve block. The value displayed
in the service mode is the absolute pressure. The atmospheric pressure must be subtracted from
this absolute pressure. The result is the relative pressure.

Example:

Displayed value PO2 is 2.76 bar, atmospheric pressure is 1.00 bar.

Relative pressure = displayed value – atmospheric pressure = 2,76bar –1.00 bar = 1.76bar

For a detailed test instruction refer to "Preventive Maintenance" (Test Certificate).

Press softkey "Return " twice.

6.1.7 P-bias (zero potential) of the airway pressure sensors


Enter service mode, press softkeys "Pneumatics", "Meas." and then "P-bias".

6173.3 Babylog 8000 Service Mode March 2000 21/30


The respiratory system pressure sensors have a P-bias (offset) after being switched on, thus
normally they are in pressure-free state.

Testing value each U = 1.74 V +0.12 V / –0.08 V.

For a detailed test instruction refer to "Preventive Maintenance" (Test Certificate).

Press softkey "Return " three times.

6.1.8 Calibrating O2 and flow sensor


Enter service mode, press softkeys "Pneumatics" and then "Meas.".

Connect O2 and Air pressure supply to Babylog 8000 (40 to 87 psi / 3 to 6 bar).

By pressing "V-call" or "02-cal" keys the flow and 02 measurement section can be calibrated. The
calibration in the service mode is identical to the calibration in the operating mode. For the 0 2
measurement section there is a two-point adjustment first with air (21 vol. %) and then with 02 (100
vol. %).

Press softkey "Return " twice.

6.2 Service mode pneumatics “Set. 1”


Set the 02 concentration via potentiometer 02 vol. %, flow via potentiometer insp. flow and set the
pressure (drive PEEP/PIP valve) via soft keys.

The potentiometer "Pinsp" must be at its maximum position for high pressure values. Otherwise, if
the Monitoring PCB measures inadmissibly high pressure values, the valve supply voltage will be
switched off. The airway pressure = Pinsp - 0.7 (Pinsp - Pexp) measured by the airway pressure
sensors and the desired value for the PEEP/PIP valve are displayed. The following tests can be
carried out with the Set. 1 menu in connection with the measuring menu:

6.2.1 Testing the pressure regulators and input pressure sensors


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

6.2.2 Testing the PEEP/PIP valve and current source


Note: Testing output pressure of PEEP/PIP valve electrical triggered. The pressure is depending on
the value of the supply pressure.

Note: First check pressure regulators and input pressure sensors according to test step 6.2.1.

Test set-up:

6173.3 Babylog 8000 Service Mode March 2000 22/30


Only connect compressed Air supply (40 to 87 psi / 3 to 6 bar).

Remove expiratory valve and measure pressure at the right lip seal (expected max. pressure 100
mbar / cmH2O).

s e c r et i on s i g ht
g l a ss 2 M 0 75 8 2

Babylog is in the service mode, press softkeys "Pneumatics" and then "Set. 1".

Settings in menu "Set. 1":

O2-conc. = 21
Insp. Flow = 10
Pinsp = 80

Set unit of Pref to decimal values (dec.) by pressing the key "Cal. Config." (SW 5) or "Monitoring"
(SW 2, 3 or 4).

Important: When performing this test PEEP/PIP valve must be in normal installation position;
Babylog is therefore not to be tilted!

Seal off center lip seal with finger (outlet 5a to the injector, see arrow) when performing next
measurements and settings.

Set displayed value of Pref to 0 (= 0 mA) with softkeys " ⇓ " and " ⇑ ".

Press softkey "Return " and then "Meas.".

Read out value PAir in the menu "Meas." and calculate relative supply pressure to PEEP/PIP valve:

6173.3 Babylog 8000 Service Mode March 2000 23/30


Rel. supply pressure to PEEP/PIP valve = Displayed PAir − ambient pressure

The reference pressure gauge must indicate the following test value depending on the calculated
supply pressure to the PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve Value reference pressure gauge (Pref = 0)
in psi in bar in cmH2O in mbar
18.85 psi 1.30 bar - 13.3 ± 3.0 cmH2O - 13.0 ± 3.0 mbar
19.58 psi 1.35 bar - 14.1 ± 3.0 cmH2O - 13.8 ± 3.0 mbar
20.30 psi 1.40 bar - 14.8 ± 3.0 cmH2O - 14.5 ± 3.0 mbar
21.03 psi 1.45 bar - 15.6 ± 3.0 cmH2O - 15.3 ± 3.0 mbar
21.75 psi 1.50 bar - 16.3 ± 3.0 cmH2O - 16.0 ± 3.0 mbar
22.48 psi 1.55 bar - 17.1 ± 3.0 cmH2O - 16.8 ± 3.0 mbar
23.20 psi 1.60 bar - 17.9 ± 3.0 cmH2O - 17.5 ± 3.0 mbar
23.93 psi 1.65 bar - 18.7 ± 3.0 cmH2O - 18.3 ± 3.0 mbar
24.65 psi 1.70 bar - 19.4 ± 3.0 cmH2O - 19.0 ± 3.0 mbar

• Setting the pressure offset of the PEEP/PIP valve:


The pressure offset is set by turning the screw at the bottom of the valve. The control pressure
must be 1.7 bar measured with a reference pressure gauge directly at the PEEP/PIP valve. The
setting value is -19 mbar (±3 mbar), whereby this value is to be checked after knocking the valve
several times and after checking the reference pressure level. It is essential that the valve is
operated in the fixed position. After this offset has been adjusted, the screw is to be secured with
a safety lacquer. The current of the peep source of current must be checked previous to this
however.

Press softkeys "Pneumatics" and then "Set. 1".

Set displayed value of Pref to 406 dec (= 63.2 mA) with softkeys " ⇓ " and " ⇑ ".

Press softkey "Return " and then "Meas.".

Read out value PAir in the menu "Meas." and calculate relative supply pressure to PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve = Displayed PAir − ambient pressure

The reference pressure gauge must indicate the following test value depending on the calculated
supply pressure to the PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve Value reference pressure gauge


(Pref = 406)
in psi in bar in cmH2O in mbar
18.85 psi 1.30 bar 3.1 ± 4.1 cmH2O 3.0 ± 4.0 mbar
20.30 psi 1.40 bar 1.5 ± 4.1 cmH2O 1.5 ± 4.0 mbar
21.75 psi 1.50 bar 0.0 ± 4.1 cmH2O 0.0 ± 4.0 mbar
23.20 psi 1.60 bar - 1.5 ± 4.1 cmH2O - 1.5 ± 4.0 mbar
24.65 psi 1.70 bar - 3.1 ± 4.1 cmH2O - 3.0 ± 4.0 mbar

6173.3 Babylog 8000 Service Mode March 2000 24/30


Press softkeys "Pneumatics" and then "Set. 1".

Set displayed value of Pref to 948 dec (= 147.4 mA) with softkeys " ⇓ " and " ⇑ ".

Press softkey "Return " and then "Meas.".

Read out value PAir in the menu "Meas." and calculate relative supply pressure to PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve = Displayed PAir − ambient pressure

The reference pressure gauge must indicate the following test value depending on the calculated
supply pressure to the PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve Value reference pressure gauge


(Pref = 948)
in psi in bar in cmH2O in mbar
18.85 psi 1.30 bar 23.5 ± 5.1 cmH2O 23.0 ± 5.0 mbar
20.30 psi 1.40 bar 21.9 ± 5.1 cmH2O 21.5 ± 5.0 mbar
21.75 psi 1.50 bar 20.4 ± 5.1 cmH2O 20.0 ± 5.0 mbar
23.20 psi 1.60 bar 18.9 ± 5.1 cmH2O 18.5 ± 5.0 mbar
24.65 psi 1.70 bar 17.1 ± 5.1 cmH2O 17.0 ± 5.0 mbar

Press softkeys "Pneumatics" and then "Set. 1".

Set displayed value of Pref to 1489 dec (= 231.6 mA) with softkeys " ⇓ " and " ⇑ ".

Press softkey "Return " and then "Meas.".

Read out value PAir in the menu "Meas." and calculate relative supply pressure to PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve = Displayed PAir − ambient pressure

The reference pressure gauge must indicate the following test value depending on the calculated
supply pressure to the PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve Value reference pressure gauge


(Pref = 1489)
in psi in bar in cmH2O in mbar
18.85 psi 1.30 bar 43.9 ± 5.1 cmH2O 43.0 ± 5.0 mbar
20.30 psi 1.40 bar 42.3 ± 5.1 cmH2O 41.5 ± 5.0 mbar
21.75 psi 1.50 bar 40.8 ± 5.1 cmH2O 40.0 ± 5.0 mbar
23.20 psi 1.60 bar 39.3 ± 5.1 cmH2O 38.5 ± 5.0 mbar
24.65 psi 1.70 bar 37.7 ± 5.1 cmH2O 37.0 ± 5.0 mbar

Press softkeys "Pneumatics" and then "Set. 1".

6173.3 Babylog 8000 Service Mode March 2000 25/30


Set displayed value of Pref to 2031 dec (= 315.8 mA) with softkeys " ⇓ " and " ⇑ ".

Press softkey "Return " and then "Meas.".

Read out value PAir in the menu "Meas." and calculate relative supply pressure to PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve = Displayed PAir − ambient pressure

The reference pressure gauge must indicate the following test value depending on the calculated
supply pressure to the PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve Value reference pressure gauge


(Pref = 2031)
in psi in bar in cmH2O in mbar
18.85 psi 1.30 bar 64.3 ± 5.1 cmH2O 63.0 ± 5.0 mbar
20.30 psi 1.40 bar 62.7 ± 5.1 cmH2O 61.5 ± 5.0 mbar
21.75 psi 1.50 bar 61.2 ± 5.1 cmH2O 60.0 ± 5.0 mbar
23.20 psi 1.60 bar 59.7 ± 5.1 cmH2O 58.5 ± 5.0 mbar
24.65 psi 1.70 bar 58.1 ± 5.1 cmH2O 57.0 ± 5.0 mbar

Press softkeys "Pneumatics" and then "Set. 1".

Set displayed value of Pref to 2572 dec (= 400 mA) with softkeys " ⇓ " and " ⇑ ".

Press softkey "Return " and then "Meas.".

Read out value PAir in the menu "Meas." and calculate relative supply pressure to PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve = Displayed PAir − ambient pressure

The reference pressure gauge must indicate the following test value depending on the calculated
supply pressure to the PEEP/PIP valve:

Rel. supply pressure to PEEP/PIP valve Value reference pressure gauge


(Pref = 2572)
in psi in bar in cmH2O in mbar
18.85 psi 1.30 bar 84.5 ± 5.1 cmH2O 83.0 ± 5.0 mbar
20.30 psi 1.40 bar 83.1 ± 5.1 cmH2O 81.5 ± 5.0 mbar
21.75 psi 1.50 bar 81.6 ± 5.1 cmH2O 80.0 ± 5.0 mbar
23.20 psi 1.60 bar 80.1 ± 5.1 cmH2O 78.5 ± 5.0 mbar
24.65 psi 1.70 bar 78.5 ± 5.1 cmH2O 77.0 ± 5.0 mbar

Press softkey "Return " twice

6173.3 Babylog 8000 Service Mode March 2000 26/30


6.2.3 Testing current source to PEEP/PIP valve
The current can be measured on the I/O PCB at the test points X3/1 and X3/2 (DGND) as a voltage
drop.

Babylog is in the service mode, press softkeys "Pneumatics" and then "Set. 1".

Set unit of Pref to hex. values by pressing the key "Cal. Config." (SW 5) or "Monitoring"
(SW 3 or 4).

Set displayed value of Pref with softkeys " ⇓ " and " ⇑ ".

Info: Conversion ratio: 1 mV = 1 mA.

Desired value Pref Voltage via X3 PCB I/O


0 hex Smaller 5 mV
FFF hex 600 to 680 mV

Possible errors:
Relay K2 defective if no voltage drop is measured. The relay can be bridged by directly placing +5 V
from the testing plug X22/1 (right pin) on the Motherboard PCB (fold away front part) onto the
connection plug for the PEEP/PIP valve X8/1 (front pin). If this is also unsuccessful, there is a fault in
the peep source of current on the I/O PCB.

6.2.4 Testing airway pressure sensors


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

6173.3 Babylog 8000 Service Mode March 2000 27/30


6.3 Service pneumatics "Set. 2"

6.3.1 Valve bank (mixer)


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

6.3.2 Ejector valve Y1.3 and ejector


For a detailed test instruction refer to "Periodic Manufacturer's Service" (Test Certificate) as of
edition February 2000.

6.3.3 O2/Air switch over valve Y1.1, O2 calibration valve Y1.3 and O2
calibration valves Y3.1 and Y3.2 in the inspiratory block
Check by the 02 calibration and setting an inspiratory flow with the “Set. 2” menu with the respective
other type of gas. The test value is the sensor voltage of the 02 sensor in the measuring menu.

Connect O2 and Air pressure supply to Babylog 8000 (40 to 87 psi / 3 to 6 bar).

Babylog is in the service mode, press softkeys "Pneumatics" and then "Set. 2".

Switch on safety valve Y 1.2 (SaVal) and flow control valve 1 of the 02 valve block (nominal value 15
L/min).

Press softkey "Return ", "Meas." and then "O2-Cal".

The Babylog 8000 starts the calibration procedure with air.

Test value 02 sensor voltages master (1st value) and monitor (2nd value):
U1 and U2 = 9.2 mV up to 26.0 mV (make a note of displayed values)

Permissible voltage difference between U1 and U2 = Lower (U1 + U2) / 8 + 1 mV.

When the Babylog 8000 is switched over from air to O2 calibration gas (via Y1.1) the voltage at the
sensor increases:

Press softkey "Return " and then "Set. 2".

Switch off the flow control valve 1 of the 02 valve bank and switch on flow control valve 1 of the air
valve block (nominal value 15 L/min).

Test values for the 02 sensor voltage at 100 vol. % each:


U=U1 x 97 / 21 up to U = U1 x 100 / 21 and

6173.3 Babylog 8000 Service Mode March 2000 28/30


U=U2 x 97 / 21 up to U = U2 x 100 / 21

Note: When the calibration has been successfully completed then no fault exists.

Press softkey "Return " twice.

6.3.4 Safety valve Y1.2


Settings are carried out in menu "Set. 1" and "Set 2". The set value to be adjusted is the value
indicated on the reference pressure gauge and is set in menu "Set. 1" with softkeys " ⇑ "
and " ⇓ ".

Test set-up:

Connect O2 and Air pressure supply to Babylog 8000 (40 to 87 psi / 3 to 6 bar).

Connect reference pressure gauge (expected max. pressure 100 cmH2O) between inspiratory and
expiratory nozzle. Use short ventilation hoses or hoses 7x2.5, 200 mm between nozzles and Y
piece:

Babylog is in the service mode, press softkeys "Pneumatics" and then "Set. 1".

Settings in menu "Set. 1":

O2-conc. = 21
Insp. Flow = 5
Pinsp = 80

Adjust set value on reference pressure gauge to 80.0 cmH2O or 80 mbar


with softkeys " ⇑ " and " ⇓ ".

6173.3 Babylog 8000 Service Mode March 2000 29/30


Press softkey "Return " and then "Set. 2".

Switch off safety valve Y1.2 (SaVal).

Test value: The Pressure at the gauge must drop immediately.

Press softkey "Return " twice.

7 Special characteristics of micro processor 68000


The Microprocessor 68000 is an asynchronous processor, which - in contrast to a synchronous
processor (e.g. 6802), requires an acknowledgement signal DTACK from the peripherals after each
external read or write access. By means of this DTACK-signal a slow -speed peripheral may be
interfaced to a fast-speed processor. If e.g. PCBs generating a direct DTACK-signal for the CPU
68000 (I/O PCB, Monitoring PCB, Front adapter PCB + Flow PCB) are missing or the DTACK
generation is defective, the processor system cannot operate.

Overall failure
The system does not start after it is switched-on, thus the valves do not switch, continuous acoustic
alarm occurs and undefined display is present. The supply voltages are applied.

The main failure cause is at the:


• CPU 68000 PCB
• I/O PCB
• Monitoring PCB
• Flow PCB
• Front adapter PCB

With this minimum equipment - thus without pneumatics components and front part - the system has
to start.

6173.3 Babylog 8000 Service Mode March 2000 30/30