2 RECALIBRATION TISCH | cet “ January 30, 2022 Environmental Lortglcate of Calbbeation Calibration Certification Information Cal. Date: January 30,2021 Rootsmeter S/N: 438320 Ta: 294 “*K Operator; Jim Tisch Pa: 754.4 mmHg Calibration Model #:_TE-S0254 Calibrator S/N: 3470 Vol.Init | Vol.Final | OVol. atime ro aH Run (m3) (m3) (m3) (rnin) | (mmHe)_| (in #20) 1 1 2 1 aia 33 2.00 2 3 4 1 1.0001 62 4.00 3 5 6 i 0.8989 81 5.00 4 7 8 1 08541 3.0 6.00 5 8 10 1 0.7088 128 8.00 Data Tabulation r Vstd Ostd Qa SH (Ta/Pay (m3) («-axis) (y-axis) Va (x-axis) 1.0021 | 0.7086 1aa2i | 0.9956 | _0.7040 0.8969 0.9982 | 0.9981 19819 | 0.9915 | 0.9916 1.2326 0.9957 | 1.1076 2.2669| 0.9892 | _1.1005 1.4099 0.9945 | 1.1643 2.3900| 0.9880| 1.1568 1.4365 0.9894 | 1.3958 2.8519 | 0.9830 | 1.3869 1.7738 | S m= 2.07344 m= 1.297812 TD. b= -0.04212 QA b= 0.0261 Q t= 0.99745 re 0.9975 Calculations: Vstd=[AVol((Pa-AP)/Pstdy(Tstd/Ta) Va=[AVol((Pa-AP)/Pa) ‘Qstd=[vsta/aTime (Qa=[Va/ATime For subsequent flow rate calculatio = [Pa ss) r waaprell Sart tappa) 0) ased= inl ( Jon GRRE) a= um((far( Taf) 8) ‘Standard Conditions std: | 298.15" Pst: | 760 mmHg BECDUSEATION US EPA recommends anual recalibration per 1998 40 Code of Federal Regulations Part 50 to 51, ‘AP: rootsmeter manometer reading (mmHg) ‘Appendix B to Part SO, Reference Method for the ‘Ta: actual absolute temperature (*k) Determination of Suspended Particulate Matter in a: actual barometric pressure (mm Ha) the Atmosphere, 9.2.17, page 30 by intercept im: slope J ‘AH; calibrator manometer reading (in H20) Tisch Environmental, ine wn 145 South Miami Avenue TOLL FREE: (877)263-7610 Village of Cleves, OH 45002 FAX: (513)467-$009