OOOmA XMSh <> will give the same
action.
2, The LCD will display the current calibration value. The left value is HGB calibration
value and the right HCT calibration value.
3. ‘The HGB calibration value which is on the cursor can be changed. Input anew setting
value with numeric keys and press to set the new calibration value into the
instrument. If change of the HGB calibration value is not necessary, press
without pressing numeric keys. Then the cursor will move to the HCT calibration value.
‘Set the HCT calibration value in the same way.
4. After setting of HCT calibration value by pressing 119.8%
HCT : 104.0% ~ 105.0%
NOTE
HGB AND HCT FACTORS CAN BE CHANGED WITH STEPS OF MAXIMUM 5%, IF
A HIGHER CALIBRATION IS REQUIRED, THIS CAN BE DONE WITH SOME
STEPS.
Example :
HGB : 100.0% TO 108.0%
First enter the value 104.0% and then the second time 108.0% (2-times
4%)
A calibration of >5% at once will be ignored
5. Ifil's necessary to calibrate the other parameters: WBC, RBC and PLT. Press first
<9> <.> <0> and then press