You are on page 1of 4

Basic Research—Technology

A Comparison of the Dentin Cutting Efficiency


of 4 Pointed Ultrasonic Tips
Matthew P. Godfrey, DDS, James C. Kulild, DDS, MS, and Mary P. Walker, DDS, PhD

Abstract
Introduction: Pointed ultrasonic tips can be used for
several applications including troughing around
intracanal obstructions. With the variety of pointed
C urrently, ultrasonic instruments are used in many different phases of endodontic
treatment such as access refinement, locating calcified root canal systems, removal
of attached pulp stones, removal of intracanal obstructions, increasing the action of
ultrasonic tips available, the purpose of this in vitro irrigating solutions, placement of mineral trioxide aggregate, and surgical endodontics
investigation was to compare the dentin cutting (1). Ultrasound is sound energy above the range of human hearing, which is
efficiency of 4 commonly used pointed ultrasonic tips. 20 kHz (2). Ultrasound energy used for endodontic instrumentation is based on the
Methods: The ultrasonic tips (n = 5 tips/group) piezoelectric principle in which a crystal is used that changes dimensions when an
included in the study were the following: CPR-3D electric charge is applied (1). This crystal deformation is converted to mechanical
(Obtura Spartan, Algonquin, IL), BL 6A (B&L Biotech, oscillation without producing heat (2). This lack of heat production represents one
Bala Cynwyd, PA), PUENDO2 (Dentsply Tulsa Dental of several advantages of piezoelectric ultrasonics in endodontics (2).
Specialties, Tulsa, OK), and WH1 (eie2, San Diego, The removal of intracanal obstructions is 1 phase of endodontic treatment vastly
CA). The tips were attached to a testing apparatus improved by ultrasonics. Pointed ultrasonic tips can be used for this purpose to trough
that produced linear movement and a 15-g axial force around obstructions to maximize the benefits of subsequently applied vibratory or
during instrumentation of a human dentin specimen. extraction forces (3). In order to conserve tooth structure, a small pointed tip is
For all tips, instrumentation was completed at the able to not only remove any cement but also a thin shelf of dentin around the perimeter
same power setting, which fell within the power setting of the obstruction (3). Previously, trephine drills were used for this purpose, which led
range recommended by each manufacturer. Dentin to complications including the destruction of root structure and the perforation of
specimens were weighed at baseline and after 6 root trunks (3). With ultrasonics, troughing around root canal obstructions can be
minutes of instrumentation to measure dentin loss to performed in a predictable and controlled manner (3).
the nearest 0.01 mg. A qualitative analysis of the shape Various investigations have evaluated the cutting efficiency of ultrasonic tips. Some
and surface topography of new and used ultrasonic tips studies tested the cutting efficiency of one ultrasonic tip over time (4), whereas others
was performed via scanning electron microscopy. examined the efficiency of various rounded ultrasonic tips over time (5, 6). In
Results: A 1-factor analysis of variance and Tukey a comparison of the cutting efficiency of 4 ultrasonic tips, Dentkos and Berzins (6)
post hoc analysis of dentin removal revealed a statisti- included 1 pointed tip. Lin et al (7) compared the cutting efficiency of various pointed
cally significant difference between the 4 ultrasonic ultrasonic tips, but dental stone was used as the cutting medium. With the variety of
tips (P < .05). The CPR-3D removed more dentin than pointed ultrasonic tips available on the market, it would be valuable to evaluate the
the other 3 tips, whereas the BL 6A removed more cutting efficiency of various tips to allow the clinician to choose a tip most appropriate
dentin than the WH1. Scanning electron microscopic for a particular purpose. The purpose of this in vitro investigation was to compare the
analysis revealed the CPR-3D as having the least change efficiency of 4 commonly used pointed ultrasonic tips to remove dentin.
to tip shape and topography as compared with the other
tips. Conclusions: Within the limits of this study,
CPR-3D showed the greatest dentin removal, which Materials and Methods
may be linked to the stability of the CPR-3D tip shape Previously extracted third molars were collected from the Oral Surgery Clinic at
and topography. (J Endod 2013;39:897–900) the University of Missouri-Kansas City School of Dentistry, Kansas City, MO, according
to an institutional review board–approved protocol. In this protocol, no patient
Key Words identifiers were associated with the teeth, and all patients signed waivers to permit
Ultrasonic efficiency, ultrasonic instrumentation, ultra- the use of their teeth for research purposes. Teeth were mounted in dental stone,
sonic tips and dentin blocks were prepared by grinding the occlusal surfaces flat with a model
trimmer (Ray Foster Dental Equipment, Huntington Beach, CA). Desiccation of the
specimens was accomplished by storing the dentin blocks in the same environment
in which all tests were conducted. Specimens were weighed daily to the nearest
From the University of Missouri-Kansas City, School of
Dentistry, Kansas City, Missouri. 0.01 mg on a Mettler-Toledo AG 245 Lab Balance (Mettler-Toledo Inc, Columbus,
Address requests for reprints to Dr Matthew P. Godfrey, OH). Complete desiccation was confirmed by a lack of change in weight for
650 East 25th Street, Kansas City, MO 64108. E-mail address: 3 consecutive days as in a previous study (4).
godfmatt@yahoo.com Dentin specimens were mounted to the slow-speed alternating current motor of
0099-2399/$ - see front matter
Copyright ª 2013 American Association of Endodontists.
the testing apparatus (Obtura Spartan, Algonquin, IL), which provided a 6-mm linear
http://dx.doi.org/10.1016/j.joen.2013.04.016 movement at an approximate rate of 1.5 mm/s. A Satelec ultrasonic handpiece (Acteon
Group, Mount Laurel, NJ) was mounted on an adjustable inclinable slide, which
provided a constant and continuous axial force on the dentin specimen (8). An axial

JOE — Volume 39, Number 7, July 2013 Dentin Cutting Efficiency of 4 Ultrasonic Tips 897
Basic Research—Technology
force of 15 g, which was established by a previous study (4), was re- TABLE 1. Dentin Removal (mg) after 6 Minutes of Instrumentation
checked and recalibrated between each group by using a digital mini
Tip (n = 5) Mean dentin removal (SD)*
scale (model 1479V; Tanita Corporation, Arlington Heights, IL) that
was fastened to the testing apparatus via an attachment (Fig. 1). CPR-3D 5.13 (1.61)a
BL 6A 3.29 (1.05)b
Pointed ultrasonic tips of similar size and shape were chosen to PUENDO2 2.06 (0.38)b,c
be included in this investigation. The 4 types of pointed ultrasonic tips WH1 1.40 (0.47)c
included in the investigation were the CPR-3D (Obtura Spartan,
Superscript letters indicate tip subsets.
Algonquin, IL), the BL 6A (B&L Biotech, Bala Cynwyd, PA), the
*There was a significant difference (P # .05) in dentin removal between tips.
PUENDO2 (Dentsply Tulsa Dental Specialties, Tulsa, OK), and the
WH1 (eie2, San Diego, CA). For all ultrasonic tips, the Satelec P5
Newtron XS US unit (Acteon Group) was used at a power setting of
Results
The mean and standard deviation dentin removal values are
30%, which fell within the power setting range recommended by
presented in Table 1. There was a significant difference in dentin
each manufacturer. The ultrasonic tips were firmly attached to the
removal as a function of tip type (P < .05). The CPR-3D removed
handpiece, with each tip being placed at a 90 angle to the dentin
more dentin than the other 3 types, whereas the BL 6A removed
specimen. Adjustments to the dentin specimen allowed the ultrasonic
more dentin than the WH1. Representative scanning electron
tips to remain in dentin during instrumentation and not extend into the
microscopic (SEM) images at 100 of a new and used tip of each
peripheral enamel (8).
type are presented in Figure 2. The CPR-3D showed little change after
Instrumentation was completed for a total of 6 minutes for each
6 minutes of use (Fig. 2B), whereas the others showed substantial
tip. After each 20-second interval, a 1-second blast of compressed air
changes to the tip shape and topography. For example, with the BL
was blown onto the specimen to remove debris. The dentin specimen
6A, there was pitting after use (Fig. 2D), whereas the PUENDO2 showed
was weighed before and after instrumentation to the nearest 0.01 mg
what appears to be the removal of its zirconium nitride coating (Fig. 2F).
on a Mettler-Toledo AG 245 Lab Balance (Mettler-Toledo Inc) (8).
With the WH1, there was a shearing of the tip, which resulted in a less
This same process was performed for 5 ultrasonic tips from each of
pointed instrument (Fig. 2H).
the 4 respective tip groups. A 1-factor analysis of variance and the Tukey
post hoc analysis (a = 0.05) were used to determine whether there
were statistically significant differences in dentin removal between the Discussion
4 ultrasonic tip groups. Ultrasonic tip cutting efficiency is an important factor to consider
To potentially explain possible differences in dentin removal during endodontic treatment. The results of this in vitro investigation
efficiency that might be related to changes in the cutting surface, indicated that the CPR-3D removed significantly more dentin across
a qualitative analysis of the effects of usage on tip shape and surface time than the BL 6A, the PUENDO2, and the WH1. These 4 ultrasonic
topography was performed using scanning electron microscopy tips were chosen because they all had pointed tips of similar size.
(FEG ESEM XL 30; FEI, Hillsboro, OR). Images were captured at Five tips of each respective type were used in this investigation, and
25, 100, and 300 magnification of both a new and used none of them fractured.
ultrasonic tip from each respective group, and all images were Considerable effort was made to remove confounding variables in the
reviewed by 3 evaluators. present investigation. The same power setting was used for all the tips,
which fell within each respective manufacturer’s guidelines. Besides using
human dentin, a clinically relevant substrate and the same axial force (15 g)
as established from a previous study (4) were included in the protocol.
Although the tips were of a similar size, the tip coatings varied.
The CPR-3D has a diamond coating; the PUENDO2 has a zirconium
nitride coating, whereas the WH1 and BL 6A have no surface coating.
The results of the current study are consistent with another investiga-
tion that reported superior cutting efficiency of diamond-coated tips
(7). In that investigation, diamond-coated tips required significantly
less time than stainless steel tips or zirconium nitride tips to penetrate
a 3-mm block of dental stone (7). In the current study, the pointed
diamond-coated tip displayed superior cutting efficiency of dentin.
Besides the tip surface coating effect on dentin removal efficiency,
another factor to consider is tip size. In an effort to choose pointed
tips from various manufacturers, the exact size of the tips used in
the study varied between brands. The WH1, which happened to be
the smallest, removed significantly less dentin than the CPR-3D and
the BL 6A, which were larger. The CPR-3D, the largest ultrasonic tip
used, removed significantly more dentin than the other 3 tips. In
addition to the diamond coating, the larger size of the CPR-3D may
have contributed to its superior efficiency at removing dentin.
However, a previous investigation of various-sized diamond-coated
tips reported that the largest tip produced the least efficiency at 2
different power settings (6). They reported that the surface area or
Figure 1. A testing apparatus showing the ultrasonic tip in contact with the flat size of diamonds might not be the most important factor for cutting
dentin surface and the apparatus inclined to provide a constant 15-g axial efficiency (6). Furthermore, they reported that the sharpest tip
force. The reciprocating jig holds the dentin specimen. produced the second most dentin removal (6). This is contrary to

898 Godfrey et al. JOE — Volume 39, Number 7, July 2013


Basic Research—Technology

Figure 2. SEM images of ultrasonic tips: (A) a new CPR-3D, (B) a used CPR-3D, (C) a new BL 6A, (D) used BL 6A, (E) a new PUENDO2, (F) a used PUENDO2, (G)
a new WH1, and (H) a used WH1.

the current investigation in which the sharpest and smallest tip (WH1) WH1 tip (Fig. 2H). Overall, it appeared that the CPR-3D underwent the
was among the least efficient tips studied. least amount of change to tip shape and topography, and this could be
In order to examine any change in ultrasonic tip shape and a possible explanation as to why the CPR-3D removed more dentin across
topography, a SEM qualitative analysis was performed on randomly time. This knowledge may aid in the manufacturing process because the
chosen used (Fig. 2B, D, F, and H) and unused tips (Fig. 2A, C, E, and ultrasonic tips with the greatest stability of shape and topography may be
G) of each respective type. A previous investigation that included SEM the most effective cutting ultrasonic tips.
imaging of new and used diamond-coated tips reported that wear was
observed as diamond loss from the cutting surface (6). In this investiga- Conclusion
tion, all 4 ultrasonic tips showed signs of wear in either surface topog- Within the limits of this investigation, the CPR-3D showed the
raphy or overall tip shape. However, the CPR-3D showed the least greatest dentin removal, which may be linked to the stability of the
amount of surface and tip change, whereas with the other 3 tips, there CPR-3D tip shape and topography.
was notably more change. For example, there was only a minimal loss
of diamond crystals and some rounding of remaining crystal edges
(Fig. 2B) with the CPR-3D. In contrast, the PUENDO2 exhibited a consid- Acknowledgments
erable loss of the zirconium nitride coating at the tip (Fig. 2F), whereas The authors acknowledge the donation of ultrasonic tips by
the BL 6A showed surface pitting (Fig. 2D) and there was a shearing of the the respective companies mentioned as well as the use of testing

JOE — Volume 39, Number 7, July 2013 Dentin Cutting Efficiency of 4 Ultrasonic Tips 899
Basic Research—Technology
equipment from Obtura Spartan and the donation of a Satelec P5 4. Lawlor K, Yelton C, Kulild J, et al. Dentin removal efficacy over time of the Buc-1
Newtron XS US unit from Acteon Group. ultrasonic tip. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2010;109:
e107–9.
The authors deny any conflicts of interest related to this study. 5. Yelton C, Lawlor K, Kulild JC, et al. Comparison of the efficiency of four different ultra-
sonic tips to remove dentin over time. J Endod 2010;36:529–31.
References 6. Dentkos TR, Berzins DW. Evaluation of cutting efficiency of orthograde ultrasonic tips
1. Plotino G, Pameijer CH, Grande NM, et al. Ultrasonics in endodontics: a review of the by using a nonstatic model. J Endod 2008;34:863–5.
literature. J Endod 2007;33:81–95. 7. Lin YH, Mickel AK, Jones JJ, et al. Evaluation of cutting efficiency of ultra-
2. Stock CJ. Current status of the use of ultrasound in endodontics. Int Dent J 1991;41: sonic tips used in orthograde endodontic treatment. J Endod 2006;32:
175–82. 359–61.
3. Iqbal MK. Nonsurgical ultrasonic endodontic instruments. Dent Clin North Am 2004; 8. Cloutier DL, Kulild JC, Walker MP. A novel methodology to evaluate the use of an ultra-
48:19–34. sonic tip. J Endod 2011;37:1264–7.

900 Godfrey et al. JOE — Volume 39, Number 7, July 2013

You might also like