Professional Documents
Culture Documents
EDMA: 12.02.01.02.00
Immunological and turbidimetric test for determination of human IgE in serum and plasma
PACKAGE INSERT
eIFU Indicator
https://www.biomed.de/en/ige-product-documentation/
To get access to this secured website-area, please enter the following password:
LSYNVabj7
!
BIOMED Labordiagnostik GmbH • Bruckmannring 32 • 85764 Oberschleissheim • Germany • +49 89 3157000 • www.biomed.de Version 2018-05
Principle of Method and Reaction Process: Test Components and Active Contents:
The reagent is based on an immunological reaction with R1-solution:
anti-human-IgE antibodies (latex based) that bind the Glycine-buffer, Sodium Chloride, bovine serum albumin, EDTA,
serum/plasma IgE. The agglutination is proportional to the preservative.
IgE concentration in the sample. The resulting light
scattering is measured by turbidimetry at 570 nm. R2-solution:
BIOMED Labordiagnostik GmbH • Bruckmannring 32 • 85764 Oberschleissheim • Germany • +49 89 3157000 • www.biomed.de Version 2018-05
2
Storage and Shelf Life: Normal Values – Reference Values in Serum:
The reagents have a shelf life as indicated in the expiry For adults, hypersensitivity is unlikely at 0-25 IU/mL,
date when unopened and kept between 2 - 8 °C protected possible at 26-100 IU/mL, likely at >115 IU/mL.
from light. When opened, cooled and protected from light The reference values are dependent on age as well as on
in the device: 7 weeks shelf life. other factors. Each laboratory should check and if
necessary determine the reference area validity for their
Automated Analyzer Application: own patient groups.
Suitable for chemistry analyzers Beckman Coulter AU
400 / 480 / 640 / 680 / 2700 / 5400 / 5800 / DxC700.
Sensitivity: 12,11 IU/mL (analytical detection limit)
Please refer to the instrument manual from Beckman Corresponds to the lowest measurable activity, which can
Coulter, especially concerning: be differentiated from zero. The calculation is made of
three standard deviations (n=30) of the lowest standard
- Installing procedures and requirements (zero standard).
- Operating mode
- Function Antigen Surplus (High-Dose-Hook-Effect):
IgE-Concentrations higher than 60000 lU/mL.
- Performance, working instructions
- Calibration procedures For the avoidance of antigen excess, serums with unclear
clinical characteristics should first be tested by means of
- Operational precautions, risks and limitations
protein electrophoresis.
- Information about service and maintenance
Performance Data:
Sample Material:
Type of precision conc. IU/mL sample type %CV n d
Use serum or plasma (heparin, EDTA, sodium citrate) taken
with standard single-use sampling tubes. Samples Repeatibility (SD) 67 Serum L1 2,23 6 5
containing precipitates must be centrifuged before 409 Serum L2 0,99 6 5
determination. Within-laboratory 67 Serum L1 1,28 6 5
It is recommended that blood sampling is carried out
409 Serum L2 1,31 6 5
according to NCCLS document M29-A4 and NCCLS
guideline NCCLS H3-A6. Total 67 Serum L1 2,41 6 5
at 20 - 25 °C: 7 days
at 4 - 8 °C: 7 days Method Comparison:
at -20 °C: up to 6 months
Methode n r y
Disposal Note:
Dilute samples 1+1 with physiological saline solution when
results are above linearity area. Multiply results by 2. Please dispose the external packing and the completely
emptied and rinsed bottles according to the national
Calculation Factor: disposal guidelines.
ng/mL x 0.42 = IU/mL lU/mL x 2.4 = ng/mL
BIOMED Labordiagnostik GmbH • Bruckmannring 32 • 85764 Oberschleissheim • Germany • +49 89 3157000 • www.biomed.de Version 2018-05
3
Device setup: Literature:
1. Thomas L, Labor und Diagnose,TH-Books Verlagsgesellschaft mbH,
Beckman Coulter AU 400 / 480 / 600 / 640 / 2012, 8, 1231-1240.
680 / 2700 / 5400 / 5800 / DxC 700: 2. Kjellman NIM, Johansson SGO, Roth A. Clinical Allergy 1976, 6, 51-59.
3. Ringel KP, Dati F, Buchholz E. IgE-Normalwerte bei Kindern.
Test Name: IgE Type: Serum Operation: Yes Laboratoriumsblätter 1982, 32, 26-34
4. Tiller, F-W, Stein, B. Das klinische Labor. 2005 ecomed MEDIZIN,
Sample: Volume 2.5 µL Dilution Volume 0 µL Verlagsgruppe Hüthig Jehle Rehm GmbH
Reagent 1: Volume 100 µL Dilution Volume 0 µL 5. Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes;
Preanalytical Variables. Brochure in: Samples: From the Patient to the
Reagent 2: Volume 53 µL Dilution Volume 0 µL µl Laboratory. Darmstadt: GIT-Verlag, 1996.
6. Passing H, Bablok W. A New Biometrical Procedure for Testing the
Wave length Pri. 570 Sec. - Equality of Measurements from Two Different Analytical Methods. J Clin
Method: END Chem Clin Biochem 1983, 21, 709-720.
7. Bablok W et al. A General Regression Procedure for Method
Reaction slope: + Transformation. J Clin Chem Clin Biochem 1988, 26, 783-790.
Measuring Point 1: First 12 Last 19 8. CLSI. Protection of Laboratory Workers From Occupationally Acquired
Infections: Approved Guideline-Fourth Edition. CLSI document M29-A4.
Measuring Point 2: First - Last - Wayne, PA: Clinical and Laboratory Standards Institute
9. Simundic* A et al. Compliance of blood sampling procedures with the
Linearity limit: - % No-Lag-Time: N CLSI H3-A6 guidelines: An observational study by the European
Min. OD: L -2.0000 Max. OD: H 2.5000 Federation of Clinical Chemistry and Laboratory Medicine (EFLM) working
group for the preanalytical phase (WG-PRE). Clin Chem Lab Med 2015;
Reagent OD limit: 53(9): 1321–1331, NCCLS H3-A6
First: L -1.0000 First: H 2.5000
Last: L -1.0000 Last: H 2.5000 Number of Tests:
Dynamic range: L 10 H 1000.0 AU 400 / 480 / 640 / 680 / 2700 / 5400 / 5800 /
Correlation factor: A 1.0 B 0.0 DxC700: 220-260 tests per kit
Caution
Manufacturer
Conformité Européenne
BIOMED Labordiagnostik GmbH • Bruckmannring 32 • 85764 Oberschleissheim • Germany • +49 89 3157000 • www.biomed.de Version 2018-05