Professional Documents
Culture Documents
Ref: 122001
Prefix: AN
Service Manual
SM Rev. 2.0
Address Headquarter
ACUTRONIC Medical Systems AG
Fabrik im Schiffli
8816 Hirzel / Switzerland
Table of Contents
Sales and Service Information .................................................................... 2
Table of Contents........................................................................................ 3
1 Preface ....................................................................................... 5
1.1 Document Structure .......................................................................................................5
1.2 List of abbreviations........................................................................................................6
1.3 Service manual revision history.....................................................................................6
2 Hardware information................................................................ 7
2.1 Serial number description..............................................................................................7
2.2 Hardware Revision History .............................................................................................7
8 Servicing ...................................................................................59
8.1 Cleaning and disinfection ........................................................................................... 59
8.2 Transportation .............................................................................................................. 59
8.3 List of the test equipment and special tools............................................................. 59
8.4 Alarms off....................................................................................................................... 60
8.5 Service-Password off.................................................................................................... 60
8.6 Spare parts list .............................................................................................................. 60
8.7 Handling of the used components ............................................................................ 60
1 Preface
This service manual is intended for use by ACUTRONIC Medical Systems AG trained and authorized service
personnel.
ACUTRONIC Medical Systems AG does not condone or approve of service activity on its products by other
than ACUTRONIC Medical Systems AG trained or authorized personnel and ACUTRONIC Medical Systems AG
is not responsible for any unauthorized repairs or modifications, or any repairs or modifications made by
unauthorized procedures.
Use of the incorrect part, or failure to exercise due care in the installation, removal, servicing, checkout or
calibration of parts and equipment or the use of unauthorized accessories, may result in damage to the
equipment which may in turn result in damage in property and injury (including death) to persons. The
purchaser and installer of these parts shall bear full responsibility and liability for the above.
All maintenance performed within the applicable warranty period must be authorized in advance by an
ACUTRONIC Medical Systems AG service representative in order to retain the warranty status of the subject
unit.
Please carefully read the Service manual before using or servicing the ventilator.
The table of contents will assist you in quickly locating the respective topic.
2 Hardware information
2.1 Serial number description
Example 1: AN10AA-0001 (Before 1st of January 2016)
Example 2: AN-03002 (From 1st of January 2016)
From 1st of January 2016 the serial number have been changed to AN, for the device and xxxxx is a sequential
number: AN-xxxxx
AN11xx-xxxx 1.1
• External power connector with integrated filter
≥ AN11BB-0219 1.1
• Changing from “FAS” to “MIM” valves
≥ AN11CC-0497 1.1
• Colour coded internal tubing
≥ AN11FC-0597 1.1
• Adaptions in fan holder
≥ AN11KC-0727 1.1
• New Mainboard 2.0
≥ AN11ED-1119 1.1
• SD-card inserted
≥ AN11FD-1177 1.1
• New blender (new pressure input regulators Kuhnke)
AN12xx-xxxx 1.2
• New Mainboard 2.1
≥ AN12KD-1365 1.2
• New air filter (blue) build in in upper and lower cover
≥ AN-01657 -
• Changed the rotary knob to plastic
≥ AN-01804 -
• New front foil without edge around the display
≥ AN-01948 -
• New manifold with adapted fittings for the membrane holder
≥ AN-02135 -
• Blue Inspiratory port
≥ AN-03000 -
• New housing type (Hakama)
≥ AN03014 -
• New front foil
fabian +nCPAP evolution | SM-Rev. 2.0 7 / 99
Ref: 122010.02EN / Date: 29.09.2018
3. Important preliminary remarks
3.2 Warranty
We warranty that this product is free from defects in material and workmanship and, when properly used,
will perform in accordance with applicable specifications. If within two years after the original shipment it is
found not to meet this standard, it will be repaired at Acutronic Medical Systems AG, or replaced at no charge
when returned to an Acutronic Medical Systems AG service facility. Any modification to the equipment,
rework or repair by other than Acutronic Medical Systems AG approved personnel will render this warranty
null and void. Acutronic Medical Systems AG shall not be liable for any indirect, special or consequential
damages, even if notice has been given of the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING, BUT NOT
LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
The policy of ACUTRONIC Medical Systems AG is to maintain product repair capability for a period of five
years after the official sales stop.
The two years warranty can only be guaranteed when the yearly maintenance with the preventive
maintenance kit was done. All maintenance performed within the applicable warranty period must be
authorized in advance by an ACUTRONIC Medical Systems AG service representative in order to retain the
warranty status of the subject unit.
Description of:
Service: measures to maintain desired condition
Repair: to restore desired condition
Maintenance: inspection, service and if applicable, repair
Always clean and disinfect equipment or equipment components prior to any maintenance – including when
returning the equipment for repair.
Every 12 months
Perform the following work:
• Maintenance and safety inspections according to the following documents:
o TS-AA-17e, Calibration and Test instruction, See Software Release Package, website
o TS-FB-27e, Test protocol for service & maintenance, See Software Release Package, website
• Preventive Maintenance kit for the device, see “Spare Part Price List” on the website.
Every 4 years
Perform the following work:
• Maintenance and safety inspections according to the following documents:
o TS-AA-17e, Calibration and Test instruction, See Software Release Package, website
o TS-FB-27e, Test protocol for service & maintenance, See Software Release Package, website
• Preventive Maintenance kit for the device, see “Spare Part Price List” on the website.
Maintenance and the safety check may only be carried out by specialists trained by
Acutronic Medical System AG , who have suitable measuring equipment and test
facilities.
Original packaging should be kept for the possible return of the unit. Only original
packaging ensures optimal protection of the equipment during transport. If it is
necessary to return the equipment during the warranty period, the manufacturer is
not liable for damages caused by incorrect packaging.
Original packaging should be kept for the possible return of the unit. Only original packaging ensures optimal
protection of the equipment during transport. If it is necessary to return the equipment during the warranty
period, the manufacturer is not liable for damages caused by incorrect packaging.
3.5 Contact
Phone: +41 44 729 70 80
+41 44 729 70 86
E-mail: GMB-CH-AMS-Support@vyaire.com
4 System overview
Component name
2 Connection for nCPAP system based on flow generators (single limb systems)
8 Pressure sensor
Warning: Do not connect Ethernet, Nurse call, USB, R232, Flow Sensor ports to
anything other than specified devices.
PIN
2/1 Normal open (NO)
2/3 Normal Closed (NC)
Plug:
Manufacturer: Binder
Series 680, 4 pins
PIN
3/4 12 - 24 V DC
1/2 GND
3. Mainboard
4.3.2 NetDCU
The NetDCU PCB can be found on the Mainboard, connected via pin strips. The NetDCU contains an
exchangeable SD memory card to store the trends, log files and special font characters. Log and trend files
can be exported via USB. The NetDCU uses embedded WinCE operation system, therefore the used USB
device has to be WinCE compatible (for example: USB1.1). For more information about the downloading of
the log files, trends and device information please see Chapter 4.6.
The Device ID of the fabian and licenses are stored in the inner memory of the NetDCU, therefore at replacing
the NetDCU module, the fabian will get a new device ID (and a new service-password) and the licenses has to
be reinstalled as well (See Licensing submenu and License handling, Chapter 5.8.).
5. NetDCU
For detailed description on the topic please read the Battery operation Chapter in the IfU.
In this document the power module expression means the sum of the following components:
- Power PCB
- Powerpack NiMH/ Inner battery
- Power supply “XP”
- Power inlet module
The most important components of the Power PCB are shown on the picture below.
6. Power PCB
Numbers in
Comp. name on PCB Function
picture 6.
1 LED 1 “Power OFF” LED (See Chapter 4.2)
2 5V OK (See Chapter
LED 2
4.2)
3 12 V OK (See Chapter
LED 3
4.2)
4 Power board in normal condition
LED 4
(See Chapter 4.2)
5 LED 5 State of charge (See Chapter 4.2)
6 F3 (SMD, 3 AT, 125 V” MELF 6125” Fuse3: Safety OFF function (See Chapter 4.2)
6.1mm x 2.7mm)
7 F2 (SMD, 3 AT, 125 V” MELF 6125” Fuse2: 24V connection (Power off alarm)
6.1mm x 2.7mm) (See Chapter 4.2)
8 AKKU NiMH Powerpack (Internal battery) connector
9 ICP2 programmer connector
10 External power supply connector
11 24V wires
12 Mainboard connection
e.g: MIM AKKU_EVOL V.2.1 04.06.2012 Hardware version (HW-Rev)
Exchange of the Powerpack (Internal battery) is part of the 4 year preventive maintenance.
Power inlet module and Power Supply “XP” join to the same point of the back panel as ground potential
(Rear panel schematics above refers to it, as terminal studs for potential equalization).
8. Power inlet module and Power supply “XP” connection with potential equalizator
4.3.4 Manifold
In case of the failure of the Manifold please exchange the complete Manifold block, and do not disassemble
it. Except if the damage is occurred on the proximal connector, the expiration valve membrane, the ribbon
cable or the tubes. The exchange of the expiration valve membrane is part of the yearly preventive
maintenance.
9. Main components of the Manifold (Left: removed Manifod, Right: fabian without housing, Manifold is
still in its original position)
Detailed description about the front side of the Manifold please see above (Chapter 4.1).
4.3.5 O2 sensor
The O2 sensor monitors the O2 level in the air between the proportional gas blender and the inspiratory tube
of the manifold, and located in the rear panel of the fabian. It calibrates itself automatically after every restart
and in case of continuous long-term operation daily (~40-45 sec/24h).
The O2 sensor built up from the O2 cell and the calibration adapter. The exchange of the O2 cell is part of the
yearly maintenance (check warranty period). The replacement of the cell might be needed more frequently
(for example: high pressure occurred on the cell).
11. Main components of the O2 sensor (upper) and O2 sensor tubing (lower)
12. Main components of the O2 sensor (upper) and O2 sensor tubing (lower)
The above mentioned three blocks (input block, proportional block and flow sensor block) are located under
the Blender PCB, as a complete unit.
The Calibration valves and the Input pressure regulators are easily accessible on the sides of the blocks and
exchangeable.
The Blender contains two pieces 1D (one in the air side and one in the Oxygen side) multiuse flow sensors.
These are accessible after the remove of the Blender PCB. In case of defect or failure of the Blender flow
sensor never try to clean it, but exchange the sensor or the complete Blender. Do not touch the pins of the
flow sensor.
14. Position of the flow sensors in the Proportional gas blender (Blender PCB is dismounted)
Numbers in pictures
12, 13 and 14, or Component Comment
name on the
Blender PCB
1 Gas inlet O2
3 Pressure regulators
4 Calibration valves
5 Pressure sensors
6 Proportional valves
15. Tube connection between the Manifold, Proportional gas blender and O2 sensor
16. Inner tubing of the Proportional gas blender (The color code does not match with the above
mentioned)
(Mixed Patient gas connector to O2 Sensor [number 8 in the table above] is not signed on the picture)
There are several drivers on the market available, which supports our acuLink protocol. For the drivers itself
the manufacturer are responsible and have documentation on the supported versions.
Via the GUI in the Service Mode – Network menu you can check the type of the PDMS protocol and interface,
and the network settings.
5 Failure analysis
5.1 Alarms messages on device
Below listed the most relevant alarms with main priority class I.
In case of this type of alarms corresponding messages are blinking on the display, highlighted red and beep
sounds. (If system failure occurs, a yellow triangle also blinking on the display.)
This table is intended to help you determine and resolve the cause of an alarm message. For more
information, please refer the IfU.
The log file contains the message text of the occurred fault.
An alarm with higher sub priority can hide a lower sub priority alarm. Alarms are categorized into 9 sub
priority levels; 1 is the highest, 9 is the lowest sub priority level.
Sub
Alarm text Alarm Type Possible cause Solution
priority
-Fan is not moving 1
- Fan rotation
Cooling fan defect System failure Clean/replace cooling fan
value is out of
tolerance
- Controller PIC is 1
out of order
- One of the
connections - NetDCU check
COM interface System failure between NetDCU
and Controller PIC - Mainboard check
is failed
- The multiplexer is
failed
1
Internal - NetDCU check
DIO interface System failure
communication
- Mainboard check
- Shortcut on 1
Power PCB
- NetDCU check
I2C interface System failure - Port expander
failure - Mainboard check
- Shortcut on I2C
1
Internal - NetDCU check
parallel interface/PIF System failure
communication
- Mainboard check
- Monitor PIC is out 1
of order
- One of the
- NetDCU check
SPI interface System failure connections
between NetDCU - Mainboard check
and Monitor PIC is
failed.
checksum 2
System failure - Reinstall controller PIC SW
conductor/controller
Sub
Alarm text Alarm Type Possible cause Solution
priority
PIC - Communication check
2
checksum monitor - Reinstall monitor PIC SW
System failure
PIC
- Communication check
Low physical 2
memory - please System failure HW malfunction NetDCU check
reboot!
- Safety relay 2
Check optically the 2 safety
failure
relays on the Mainboard ->
Safety relay defect System failure - Soldering
Exchange safety relay or
material across the
Mainboard
safety relay
Either one of the 2
watchdogs are
failed or one of the - Power board check
Voltage monitoring System failure
measured voltage - Mainboard check
values are out of
tolerance.
No power or too 2
less power from
battery.
-Fuse might be Exchange battery, check
Battery defect Electrical power
blown fuse.
- Battery failure
- Connector to the
battery is failed
Exchange exhalation 2
Exhalation Exhalation
System alarms membrane holder or
calibration calibration failed
membrane.
Proportional gas 2
check Blender System alarms blender or control Check blender
failure
O2 sensor calibration Error occurred 4
Sensor alarms Repeat calibration.
failed during calibration.
Oxygen sensor defect Sensor alarms O2 sensor failure Replace O2 sensor. 4
The numbering of watchdog LEDs depends on the hardware version, it is D700-703 or D600-603.
The log file contains the error code and the message text of the occurred fault.
Error
Message Possible Cause
Code
- Contact problem
Hot wire Air AIR: U101, L100, R117, 118, T101, R113
8
(See picture 18. below.)
- Blender Flow sensor
- Contact problem
Hot wire O2 O2: U101, L101, R119, 120, T102, R114
16
(See picture 18. below.)
- Blender Flow sensor
19. Part of Blender PCB (Left side is the O2 side, Right side is the Air side)
20. Menu
Press “Tools” button to save the Trends, Log files and the Device info.
The “save” buttons are displayed once the USB key is recognised. Then push the appropriate button(s) and
follow the instructions.
Push the “Yes” button and do not disconnect the USB stick while saving the data.
24. Log files and Trends in folders, Device info in a text document
26. Logs contain the data from the SD Card and FFDISK (Both are on the NetDCU)
Download the corresponding software package and use the documents from 06 Actual Test Instruction
folder.
28. Use the documents from the 06 Actual Test Instruction folder
29. Menu
The Service-password is given by the last 5 digits of Device-ID. (For example on the picture above the Service-
password is E8A90). The Device-ID is stored in the memory of the NetDCU (Chapter 3.2.2.).
In service mode you can check the System information, by pressing the “System” button.
Information which cannot be found in Info menu (without service mode) but in Service Mode – System menu
are the following (see pictures 29 and 32.):
- Interfaces
- State Conductor PIC
- PhysMemory
- FFDISK
- Hardware Configuration
In order to check the information of the Power module navigate back to the Service Mode main menu and
choose the Accuboard submenu.
It is possible to check the actual state and the driver version of the communication protocols in Service
mode -> Interfaces submenu.
If Patient Data Management System (PDMS) is available, it is useful to monitor to following network data.
Network submenu accessible via the Service mode menu.
Flow correction default setting is 0.960 in both NEO and PED ranges. Modify it only at the instruction of
Acutronic Medical Systems AG.
Always take care of the correct setting of the Altitude.
Typical values of P-insp (P-mix), Oxy1, Oxy2, P-AIR and P-OXY see in table below:
The State Monitor PIC column shows if the events are occurred/ conditions are met.
0 – Inactive
1 – Active
Perform a “Burn In” after replacing the calibration valves (on the Manifold) for 15minutes. (You can hear the
sound of the valves clacking.) After this period the fit of the sealing rings is okay to close properly the
calibration path.
41. Service Mide- Display (Brightness setting is the same as in the user Menu)
During the touch screen calibration process carry out the next steps by following the instructions.
After a successful calibration you have to push once on the touchscreen and then the calibration is stored. If
you don’t do it within the 30 seconds, the calibration is not stored and you have to redo it again.
The “reset complete” applies the default settings and sets the parameters to their default values.
By “Enable factory settings” it makes “factory defaults” option accessible for the end-user in the general
Ventilation submenu.
Available licenses can be monitored in licensing submenu. The licenses are related to Device-ID, therefore if
the Device-ID is exchanged (for example: in case of the exchange of the NetDCU), all the previous licenses are
lost (contact Acutronic Medical Systems AG).
The Licensing is managed by Acutronic Medical Systems AG at factory side. Should it be necessary to re-enter
the keys, please contact Acutronic Medical Systems AG.
After registration you can download the appropriate version of the Demo License Generator from the
Acutronic Medical Systems AG website.
The License, which generated with Demo License Generator, is valid for 3 months.
It is possible to set up the license keys via the update USB stick also. The license key (available from
Acutronic Medical Systems AG) must be located in the SETUP\ LICENSE directory on the USB update stick.
Start fabian in manual update mode:
Set the install type to manual in SETUP\ AcuInstall.ini file (Leave all of the other settings unexchanged).
#### Installtype########################################################################
# define what and how to install (see below)
#
# MANUAL: enables the manual installation, does not read the [CONFIGURATION]
# AUTO: installs the update using the current configuration of NetDCU,
# does not read the [CONFIGURATION]
# COLDBOOT: for production or a complete first installation
#########################################################################################
INSTALLTYPE=MANUAL
During the manual installing process on the fabian press “Licensing” button, then “install license from USB”
(Modul Licensing or PDMS Licensing) button and follow the instructions.
If you switch off the service-password during the service or repair process, please do not forget to switch it on
after the maintenance.
It is possible to switch off the system alarm (this is the same function as in Chapter 7.4. mentioned) with
usage of the System Alarms slidebar.
If you exchanged the main-PCB, you have to set the running hours of the device here in manually.
When you put in a “Next service date” the user will have the following information in the info-bar visible:
This has no impact to the functionality of the device. It shows only, that the next maintenance has to be
done.
7 Software update
For Software update instructions please read the Update description. After registration you can download
the appropriate version of the needed documents from the ACUTRONIC Medical Systems AG website.
After software update fill the Software update form out and send it back to ACUTRONIC Medical Systems AG.
8 Servicing
8.1 Cleaning and disinfection
Clean and disinfect the house and the screen of the fabian with a damp cloth at regular intervals. To clean
the system we recommend mild soap-suds. While cleaning and disinfecting the fabian disconnect the unit
from mains supply. Hold the device horizontally and make sure that no fluids penetrate in the device. For
disinfection refer to your infection control or housekeeping protocol that references how to disinfect
equipment of this nature. If you disinfect the device by spraying, make sure not spray in the system openings.
Cleaning and disinfection procedures according to country specific regulations. The device may not be
sterilized under any circumstances.
Cleaning of the Flow Sensor and the Membrane holder with membrane according to the IfU Sterilisation /
Cleaning / Disinfection (Chapter 15.). Preparation of the test sets, maintenance kit and spare parts according
to the manufacturer information.
8.2 Transportation
Only original packaging ensures optimal protection of the equipment during transport. If it is necessary to
return the equipment during the warranty period, the manufacturer is not liable for damages caused by
incorrect packaging.
If you switch off the service-password during the service or repair process, please do not forget to switch it on
after the maintenance.
Treat the electrical and electronic waste in accordance with the local recycling and waste management
regulations.
54. Flowchart of the disassembling process (Accessable components are in the ovals)
To replace the cell of the O2 sensor, the removal of the housing is not necessary. Screw the 2 knurled head
screws out from the O2 sensor module (rear panel of the fabian) and take the sensor out.
Slightly push backwards the pawl on the connector and disconnect it from the cell. Screw out the cell from
the calibration adapter.
After the exchange of the cell reconnect the cable and fasten the sensor module with screws.
Take care on the direction of the membrane, the TOP sign have to be on the upper side.
After the replacement of the Expiratory membrane, screw the Expiratory tube port back.
To replace the Proximal connector, the removal of the housing is not needed. In case of its damage, screw
the connector out and replace it.
To replace the User knob, the removal of the housing is not necessary. In case of its damage, pull the knob
out and replace it.
61. Placement of the User knob on the Plastic front panel (Contact with the Encoder)
Remove the 4 screws from the front part of the module, loosen the remaining 2 on the back part and bend
the Power. module backwards (>90°).
The following components are accessible after the opening of the upper side of the fabian:
- Power module (Power PCB, Power supply “XP”, Powerpack/ Internal battery)
- Mainboard
- NetDCU
- Connector PCB
- Power inlet module
After the reparation, fasten the Power module and the housing with screws in its original position.
To access the Powerpack please read first the disassembling instructions above in Chapter 8.2.
Disconnect the AKKU cable, which provide contact between the Power PCB and the Powerpack. After the
removal of the 3 fastening screws the metal plate can be slided away and the Powerpack/ Betterypack is
already accessible. During the exchange of Powerpack/ Betterypack handle the anti-skidding foil with care.
66. Components
Exchange the Powerpack (take care of the disposition of the cable on the Powerpack/ Betterypack), slide
back the metal plate and fix it with 3 screws. Connect the new Powerpack/ Betterypack with the Power PCB.
Under normal the battery has no wear and tear. The internal battery is part of the 4-year maintenance kit and
will be replaced in this time frame.
A fully charged battery must have an operating time of 2 hours (IPPV-Mode with standard paediatric settings
and connected to the gas supply).
To access the PowerPCB please read first the disassembling instructions above in Chapter 8.2.
After the repair or exchange of the board fasten it with screws and reconnect all of the above mentioned
cables. (Some version contains 6 plastic fixing cover on the metal legs between the PCB and the metal plate.
These can deform if the screws are fastened too tightly.)
To access the Power supply “XP” please read first the disassembling instructions above in Chapter 8.2.
Remove the 4 screws and the Power supply “XP” is already detachable.
After the repair or exchange of the board fasten it with screws and reconnect all of the above
mentioned cables.
To access the Connector board please read first the disassembling instructions above in Chapter 8.2.
With the help of Assembly key for binder connector (see spare parts list) the 3 screws, which fasten the
Connector board, are removable. (During the remove take care of the O2 sensor connector.)
After the repair or exchange of the board fasten it with screws and reconnect all of the above mentioned
cables. During the assembling do not forget to put the securecap of the VDC connector back.
9.2.1.5 NetDCU
To access the NetDCU please read first the disassembling instructions above in Chapter 8.2.
Remove the 3 fixing screws and slightly pull up the NetDCU (The NetDCU connects to the Mainboard with pin
straps). Please take care to keep the NetDCU parallel with the Mainboard during the removal, otherwise the
pins may be deformed. The connectors are SMD soldered and do easy brake off and this is not covered by
the warranty.
The position of the pin straps and place of the screws on the Mainboard see below.
71. Pins on the Mainboard and connectors on the lower side of the NetDCU
After the repair or exchange of the board fasten it with the pin connectors and the screws.
9.2.1.6 Mainboard
To access the Mainboard please read first the disassembling instructions above in Chapter 8.2.
(If there is not enough space to remove the Mainboard, first screw out the remaining 2 screws, which are
fasten the Power module.)
After the repair or exchange of the board fasten it with screws and reconnect all of the above mentioned
cables. During the replacement, suggested to fit first the 3 connectors (RS232, USB, Ethernet) into the holes
on the rear panel. Please take care to keep the O2 sensor cable in the notch, otherwise the cable may be
damaged. Please take care to keep the O2 sensor cable without tangle, otherwise next time the O2 cell may
not be exchangeable without the removal of the housing.
To access the Power inlet module please read first the disassembling instructions above in Chapter 8.2.
Remove the Mains cable lock and the 2 fixing screws (back side of the rear panel).
After the repair or exchange of the module fasten it with screws and reconnect all of the above mentioned
cables. During the assembling do not forget to put the Mains cable lock back.
To access the Loudspeaker and Fan module please read first the disassembling instructions above in
Chapter 8.2.
Disconnect the cables of the Fan and the Loudspeaker on the mainboard. (Both of them.)
Remove the 4 screws and pull out the Loudspeaker and Fan module.
The following components are accessible after the removal of the Loudspeaker and Fan module:
- Manifold
- Proportional Gas Blender
- Loudspeaker
- Fan
- O2 sensor (with calibration adapter)
After the reparation fasten the Loudspeaker and Fan module in original position with screws.
9.2.2.1 Manifold
To access the Manifold please read first the disassembling instructions above in Chapter 8.2.2.
In case of removal of the Manifold, the Complete Front Housing have to be also displaced. Therefore remove
the 2 Front Housing fixing screws, which are closer to the manifold and loose the other 2, which are farther
from the manifold. Bend the Complete Housing Front panel a little bit forwards (~20°). Do not open in too big
angle, because the Touch cable may come out from the Mainboard connector.
80. Lay the Manifold on the metal bulges and connect the appropriate tubes and cables
After repair or exchange of the Manifold slide it back under the front panel. To find the right position the
metal bulges help. Fasten it with 2 screws. Close back the Complete Front Panel and fix the 4 screws.
Connect all of the above mentioned tubes and cables.
To access the Proportional Gas Blender please read first the disassembling instructions above in Chapter
8.2.2.
Screw out the 3 fixing screws and remove the Proportional Gas Blender.
If you detect failure related to one of the component of the Proportional gas blender, please contact service/
local dealer, and clarify if under the specific circumstances the Proportional gas blender is exchangeable
partially or in a whole block.
To access the Proportional valves please read first the disassembling instructions above in Chapter 8.2.2.2.
Disconnect the cable of the appropriate valve from the Blender PCB, by pressing slightly down the edge of
the connector. (Both of cables, if you have to replace both of the valves.)
80 / 99 fabian +nCPAP evolution | SM-Rev. 2.0
Ref: 122010.02EN / Date: 24.04.2019
9. Repair and various replacements/ Disassembly and assembly
Remove the 2 screws where the Proportional valve is fixed (on the side of the Blender) and pull out it from
the Blender.
84. 2 fixing screws of the Proportional valve on the side of the Blender
After the exchange of the valve(s), push it back into the blender, fasten it with screws and reconnect its
cable(s).
To access the Calibration valves please read first the disassembling instructions above in Chapter 8.2.2.2.
Disconnect the cable of the appropriate valve from the Blender PCB. (Both of the cables, if you have to
replace both of the valves.)
Perform a “Burn In” after replacing the Calibration valves for 15minutes (Please see Chapter 5.5).
Remove the 2 screws where the Calibration valve is fixed (on the side of the Blender) and carefully remove
the valve, with taking care all of its moving parts.
86. 2 fixing screws of the Calibration valve on the side of the Blender
After the exchange of the valve(s), put it back into the blender, fasten it with screws and reconnect its
cable(s).
To access the Input pressure regulators please read first the disassembling instructions above in Chapter
8.2.2.2.
Remove the 2 screws where the Input pressure regulator is fixed (on the side of the Blender)
82 / 99 fabian +nCPAP evolution | SM-Rev. 2.0
Ref: 122010.02EN / Date: 24.04.2019
9. Repair and various replacements/ Disassembly and assembly
87. 2 fixing screws of the Input pressure regulator on the side of the Blender
After the exchange of the Input pressure regulator, put it in right position and fasten with the screws.
To access the Blender PCB please read first the disassembling instructions above in Chapter 8.2.2.2.
Remove the 4 fixing screws and pull up the Blender PCB. The Inner flow sensor still connects to the PCB with
pins (on the bottom side of the board) therefore handle it with more care during the removal.
88. Proportional gas blender module with and without Blender PCB
After the repair or exchange of the Blender PCB, put it in right position and fasten with the screws. Connect
all of the above mentioned cables and tubes.
To access the Inner flow sensors please read first the disassembling instructions above in Chapter 8.2.2.2.4.
After the removal of the Blender PCB the Inner flow sensors are already accessible.
Screw out the 2 screws and remove the flow sensors with the help of a forceps.
Before inserting the new flow sensor read carefully the handling instructions of the manufacturer. Use
forceps and do not touch the pins of the sensor.
After the replacement fasten the inner flow sensors with the metal plate and screws.
9.2.2.3 Loudspeaker
To access the Loudspeaker please read first the disassembling instructions above in Chapter 8.2.2.
Separate the wires of the Loudspeaker from the wires of the Fan. Remove the 4 screws and take the
Loudspeaker out.
After repair or exchange of the Loudspeaker fasten it with the screws and lead the cables through the
wrapping hose.
Lead the cables through the appropriate notch under the Manifold to connect it to the Mainboard (Access
the mainboard: Please read Chapter 8.2.)
9.2.2.4 Fan
To access the Fan please read first the disassembling instructions above in Chapter 8.2.2.
Separate the wires of the Fan from the wires of the Loudspeaker. Turn the 4 rubber feet of the Fan into the
bigger holes and remove the Fan from the metal plate.
After repair or exchange of the Fan fasten it with its rubber feet and lead the cables through the wrapping
hose.
Lead the cables through the appropriate notch under the Manifold to connect it to the Mainboard (Access
the mainboard: Please read Chapter 8.2.)
9.2.2.5 O2 sensor
To access the O2 sensor please read first the disassembling instructions above in Chapter 8.2.2.
Remove the 2 knurled head screws from the outside part of the rear panel and pull out O2 sensor.
Disconnect the O2 sensor-Mainboard cable, by pressing backwards the white pawl with a forceps on the
connector.
To the replacement of the O2 sensor adapter remove the 2 fixing screws and screw off the sensor cell.
96. Screws, which fasten the O2 calibration adapter to the rear metal plate
After the replacement of the O2 sensor connect the O2 sensor-Mainboard cable and fasten the sensor to the
rear panel with 4 screws. Attach all of the above mentioned tubes.
fabian +nCPAP evolution | SM-Rev. 2.0 89 / 99
Ref: 122010.02EN / Date: 29.09.2018
9. Repair and various replacements/ Disassembly and assembly
Remove the 4 fixing screws and take down the Complete Front Panel.
After repair or exchange of the Front Panel fasten it with screws and reconnect all of the above mentioned
cables. Take care of the placement of the cables. Do not forget to wipe the fingerprints and the
contaminations off from the inner side of transparent part of the panel with a dry, clean textile. (Assembling
suggestion: first clean the panel and lead the UDP and Touch wires through the notch and after that the
ribbon cable of the Keyboard. Fit the front panel on the Manifold and cover the fabian with it. Fasten the
Panel with screws. Connect the 3 cables to the Mainboard.)
97. Keyboard (wide green and transparent) and UDP, Touch (2 thinner) cables go through the holes on
the metal front panel
98. Cable of the touch screen connected in the right position (connecting surfaces of the cable look
downwards)
To access the LCD display please read first the disassembling instructions above in Chapter 8.2.3.
Disconnect the following cables from the Mainboard (To access the Mainboard remove the 4 screws close to
the front side, loosen the 2 close to the back side and bend the power module up).
100. Both sides remove the circled 2 screws and loosen the dashed circled screw, to access 2 of
the fixing screws of the LCD display
- LCD Display (Pull up slightly the black strap to disconnect the cable)
- Backlight
Remove the Loudspeaker and Fan module according to the instructions in Chapter 8.2.2.
Remove the remaining 2 screws.
After the reparation or during the replacement of the LCD display first lead the LCD display and the backlight
cables through the appropriate notches. Do not twist the LCD display ribbon cable.
103. LCD display (white) and Backlight (black and red) cables of the LCD through the holes on the
metal front panel
During the fitting of the LCD Display take care of the direction. (Thicker frame have to be closer to the
Manifold, thinner farther.)
Fasten the LCD Display with the 4 screws and connect the LCD display and backlight cables to the
Mainboard. Do not forget to wipe the fingerprints and the contaminations off from the LCD display with a dry,
clean textile. (And fix the Power module with 6 screws.)
9.2.3.2 Encoder
To access the Encoder please read first the disassembling instructions above in Chapter 8.2.3.
Screw out the fixing screw from the outer side of the Plastic front panel and remove the Encoder.
After the repair fix the encoder in its original position with the screw. Take care of the direction of its cable.
Push the User knob back.
109. Ribbon cable of the Encoder goes upwards to the direction of the Mainboard
10 Technical Specifications
See “Technical Data Sheet” on the ACUTRONIC Medical Systems AG website:
The minimum and maximum ambient temperatures in the clinical area will not affect the accuracy of the
flow, pressure or oxygen measurement.
Weight Approx. 10 kg
10.3 Ratings
Pressurised air supply 2.0-6.5 bar / max. 40 l/min. Connector: NIST
Medical Grade AIR
Oxygen supply 2.0-6.5 bar / max. 40 l/min. Connector: NIST
100-240 VAC, 0.8 – 0.2 A, 50/60Hz
Power supply external 12 - 24 VDC 4 A
(Charging only when more than 12.5V)
max. 70 W
Power consumption
Standby, connected to mains: 35 W (battery charging)
Internal Battery 13.2 V / 4500 mAh max. charging time: 5 hours
Fuse 3 AT, internal NiMH Battery
battery charge is continuously monitored
battery life approx. 3 hours in conventional mode (IPPV)
Sound pressure Max. 52 dB (A)
Device fuse T1AL
Safety class Type B ≙ EN 60601-1
Certifications of IEC 60601-1, 60601-1-2, 60601-1-8, 6060601-2-12
Classification to EC directive 93/42/EEC Class II b
IP protection IPX1,
Protected against falling water equivalent to 3-5 mm rainfall
per minute for a duration of 10 minutes. Unit is placed in its
normal operating position.