You are on page 1of 15

Ultrasound in Med. & Biol., Vol. 45, No. 12, pp.

32463260, 2019
Copyright © 2019 World Federation for Ultrasound in Medicine & Biology. All rights reserved.
Printed in the USA. All rights reserved.
0301-5629/$ - see front matter

https://doi.org/10.1016/j.ultrasmedbio.2019.08.018

 Original Contribution

ACOUSTIC DROPLET VAPORIZATION IN ACOUSTICALLY RESPONSIVE


SCAFFOLDS: EFFECTS OF FREQUENCY OF EXCITATION, VOLUME FRACTION
AND THRESHOLD DETERMINATION METHOD

TAGEDPMITRA ALIABOUZAR,* XIAOFANG LU,* OLIVER D. KRIPFGANS,*,y,z


J. BRIAN FOWLKES,*,y,z and MARIO L. FABIILLI*,y,zAGEDNTE
* Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA; y Applied Physics Program, University of
Michigan, Ann Arbor, Michigan, USA; and z Department of Biomedical Engineering, University of Michigan,
Ann Arbor, Michigan, USA
(Received 26 April 2019; revised 11 August 2019; in final from 23 August 2019)

Abstract—Ultrasound-induced vaporization of liquid perfluorocarbon (PFC) droplets into microbubbles, termed


acoustic droplet vaporization (ADV), has potential therapeutic and diagnostic applications. Recently, we demon-
strated how ADV—a threshold-based phenomenon—can modulate the release of biomolecules from composite
hydrogels, thereby stimulating regenerative processes, such as angiogenesis. These composite hydrogels, called
acoustically responsive scaffolds (ARSs), consist of monodispersed, micron size PFC emulsions embedded within a
fibrin matrix. This study investigated the effects of frequency of excitation (2.25, 5, 7.5 and 10 MHz) and volume frac-
tion (0.05%, 0.2% and 1% [v/v]) of monodispersed, double emulsions in the ARSs on the ADV threshold. We deter-
mined and compared the ADV thresholds via acoustic methods, including active detection, passive detection and
attenuation, as well as an echogenicity-based method using B-mode imaging. The ADV threshold determined via
these four techniques showed an increasing trend with frequency of excitation. Further analysis of the wave propaga-
tion showed that the amplitudes of high frequency harmonics were diminished in ARSs with high volume fractions of
emulsion. The ADV threshold inversely correlated with the volume fraction of emulsion at the lowest excitation fre-
quency. However, at higher frequencies, possibly due to the high acoustic reflectivity of the PFC emulsions, the ADV
threshold correlated directly with the volume fraction of the emulsion. Additionally, the ADV efficiency correlated
with the supra-threshold acoustic pressure. Overall, these results elucidate fundamental acoustic properties of the
ARSs, which can be used in future applications. (E-mail: mfabiill@umich.edu) © 2019 World Federation for
Ultrasound in Medicine & Biology. All rights reserved.
Key Words: Acoustic droplet vaporization, Ultrasound, Perfluorocarbon, Fibrin hydrogel, Regenerative
medicine.

INTRODUCTION which the phase transition becomes energetically favor-


able (Kawabata et al. 2006).
Over the past years, phase shift droplets of volatile per-
Perfluoropentane (C5 F12) has frequently been used
fluorocarbon (PFC) liquids have demonstrated better sta-
as the PFC phase in phase shift emulsions because of its
bility than conventional ultrasound contrast agents (i.e.,
bulk boiling point (i.e., 29˚C), which is close to body
microbubbles) and the ability to be formulated in the
temperature. However, recent techniques have enabled
sub-micron size range (Mullin et al. 2011; Sheeran et al.
production of stable, sub-micron size perfluorobutane
2013). These liquid PFC droplets can be vaporized in
(C4F10) and octafluoropropane (C3F8) droplets (Sheeran
situ into highly echogenic microbubbles by external
et al. 2012; de Gracia Lux et al. 2017; Li et al. 2018).
application of ultrasound pulses. The vaporization of the
Under physiologic conditions, these droplets are meta-
liquid PFC core using acoustic waves, termed acoustic
stable in the liquid state, despite being superheated.
droplet vaporization (ADV), occurs beyond a critical
Researchers have pointed to the suppression of homoge-
peak negative pressure (i.e., the ADV threshold) at
nous nucleation, rather than elevated boiling point
through Laplace pressure, as the likely mechanism for
Address correspondence to: Mario L. Fabiilli, Ph.D., University
of Michigan, 1301 Catherine Street, 3226 A Medical Sciences Building the metastability of superheated droplets (Mountford and
I, Ann Arbor, MI 48109-5667. E-mail: mfabiill@umich.edu Borden 2016). Alternatively, PFCs with higher bulk

3246
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3247

boiling points ultimately offer better thermal stability by (Giesecke and Hynynen 2003; Aliabouzar et al. 2019),
eliminating any potential spontaneous vaporization as active detection (Schad and Hynynen 2010; Reznik et al.
well as higher probability of repeated vaporization and 2011), high speed optical microscopy (Kripfgans et al.
re-condensation, which are desired for extended diagnos- 2004; Sheeran et al. 2013) and echogenicity-based tech-
tic and therapeutic applications (Asami and Kawabata niques (Kripfgans et al. 2000; Kawabata et al. 2006;
2012). In our recent paper, we investigated the use of Fabiilli et al. 2009; Williams et al. 2013)—have been
phase shift micro-emulsions with perfluorohexane (PFH) employed. Because of the differences in threshold deter-
(C6F14), perfluoroheptane (C7F16) and perfluorooctane mination techniques, droplet size distributions and
(C8F18) (Lu et al. 2019). experimental setups, it is difficult to directly compare
The ability to non-invasively and spatiotemporally ADV thresholds among the previous studies. Although
control ADV using focused ultrasound makes phase shift each technique has its own merit, a systematic study
droplets desirable for therapeutic applications (Kripfgans comparing different methods of ADV detection has not
et al. 2005; Fabiilli et al. 2010; Rapoport 2012; Moncion been completed before.
et al. 2016a). For example, phase-shift emulsions can be In this work, acoustic properties of fibrin gels and
formulated to contain a payload (e.g., drug), thereby ARSs, including attenuation and sound speed, were stud-
enabling localized payload delivery following ADV ied prior to ADV. The ADV thresholds in ARSs were
(Couture et al. 2012; Duncanson et al. 2014). We have determined as a function of frequency of excitation and
utilized ADV to control extravascular release of thera- volume fraction of monodispersed, double emulsions. We
peutic agents by incorporating phase shift emulsions determined and compared the ADV thresholds using
with a double emulsion structure of water-in-PFC-in- acoustic methods, including active and passive detection,
water (W1/PFC/W2) into implantable fibrin hydrogels. attenuation measurements and an echogenicity-based
These composite hydrogels, termed acoustically respon- method using B-mode imaging. Knowing how the acous-
sive scaffolds (ARSs), have been used to stimulate tic properties of ARSs, such as attenuation, change pre-
regenerative processes like angiogenesis via the ADV- and post-ADV can be a valuable tool to monitor and
controlled release of growth factors. For tissue regenera- quantify the release of a therapeutic payload in situ. Fur-
tion, the ability to spatiotemporally control the release of thermore, wave propagation and generation of harmonics
growth factors is desired given that regenerative pro- in ARSs was investigated. The efficiency of vaporization,
cesses are driven by the spatiotemporally controlled pre- which is relevant for drug delivery, was also measured in
sentation of growth factors. ARSs. Overall, these fundamental acoustic properties can
The impact of the following parameters on the be used to hone future therapeutic studies with ARSs as
ADV threshold within ARSs has been previously well as applications of ADV within viscoelastic media
reported: fibrin concentration, emulsion concentration, (e.g., tissue).
PFC core, droplet size and pulse repetition frequency
(Fabiilli et al. 2013; Moncion et al. 2016a). However,
the effect of the frequency of excitation used to generate
MATERIALS AND METHODS
ADV within ARSs is currently unknown. The ADV
threshold has been shown to increase (Sheeran et al. Double emulsion preparation and characterization
2013; Aliabouzar et al. 2018b) or decrease (Kripfgans Micron-sized, double emulsions with a W1/PFC/W2
et al. 2000; Schad and Hynynen 2010) with the fre- structure were prepared using a microfluidic-based tech-
quency of excitation within liquids. For homogenous nique following a previously described method (Moncion
nucleation in liquids, the length of the negative pressure et al. 2017). PFH (C6 F14, CAS# 355-42-0, Strem Chemi-
half-cycle, which is longer at lower frequencies, directly cals, Newburyport, MA, USA), with a bulk boiling point
correlates with the probability of nucleation (Church of 56˚C, was used as the PFC phase. The shell stabilizing
2002). Alternatively, the inverse correlation between the the primary emulsion (W1/PFC) was a fluorosurfactant
ADV threshold and frequency has been shown to result copolymer consisting of Krytox 157 FSH (CAS# 51798-
from non-linear propagation and super-harmonic focus- 33-5, DuPont, Wilmington, DE, USA) and polyoxyethy-
ing, which is only effective for larger droplets (Li et al. lene glycol (molecular weight: 1000 g/mol, CAS# 24991-
2014; Shpak et al. 2014). The use of higher frequency 53-5, Alfa Aesar, Ward Hill, MA, USA). The fluorosurfac-
ultrasound to generate ADV within ARSs would enable tant was dissolved at 2% (w/w) in PFH. The PFH solution
more tightly focused ADV as well as minimize any cavi- was combined with phosphate buffered saline (PBS, Life
tation-related damage to large molecular payloads or Technologies, Carlsbad, CA, USA) and then sonicated
adjacent cells. (Q55 with a CL-188 immersion probe, QSonica, LLC,
To measure the threshold of vaporization, different Newton, CT, USA) for 30 s while on ice. The resulting pri-
methods of detection—such as passive detection mary emulsion had a W1/PFC ratio of 1:2 (v/v).
3248 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

The primary emulsion was pumped at 0.5 mL/min Panametrics, Olympus NDT Corporation, Waltham, MA,
through an in-line filter (0.5 mm stainless steel frit, Cat# USA) were used to generate ADV within the ARSs. The
24993, Restek, Bellefonte, PA, USA) and into the inner specifications of the transducers, which operated as trans-
channel of a quartz microfluidic chip (Cat# 3200146, mitters at their central frequencies, were as follows:
Dolomite, Royston, United Kingdom) using a syringe 2.25 MHz (Model A305 S; 6 dB: 1.463.2 MHz),
pump (KDS-410, KD Scientific, Holliston, MA, USA). 5 MHz (Model V308; 6 dB: 3.066.52 MHz), 7.5 MHz
Simultaneously, 50 mg/mL Pluronic F68 (CAS# 9003- (Model A321 S; 6 dB: 5.569.92 MHz) and 10 MHz
11-6, Sigma-Aldrich, St. Louis, MO, USA) in PBS was (Model A315 S; 6 dB: 6.7513.03 MHz).
pumped at 2.5 mL/min through an in-line filter and into All transducers were calibrated in the free field at
the outer channels of the chip using a second syringe the focus using an in-house built fiber optic hydrophone
pump (78-0388, KD Scientific). The chip was mounted with a fiber diameter of 100 mm. Each ADV transducer
on an inverted microscope (DMIL, Leica Microsystems, was driven by a pulsed waveform (pulse duration, 5.4
Buffalo Grove, IL, USA), which enabled visualization of ms; pulse repetition frequency, 100 Hz) generated by a
emulsion production. Emulsions were stored at 4˚C until dual-channel function generator (33500 B, Agilent Tech-
use. The size distribution, coefficient of variation and nologies, Santa Clara, CA, USA) and amplified by a
concentration of the double emulsion (n = 5), analyzed gated radiofrequency (RF) amplifier (GA-2500 A Ritec
using a Coulter Counter (Multisizer 4, Beckman Coulter, Inc., Warwick, RI, USA). The amplified signals were
Brea, CA, USA) with a 50 mm aperture tube, were 13.2 viewed in real-time on an oscilloscope (HDO4034, Tele-
§ 0.8 mm, 2.6% § 0.2% and 3.44 £ 108 § 0.57 £ 108 dyne LeCroy, Chestnut Ridge, NY, USA).
droplets per mL, respectively.
Acoustic characterization of ARSs and fibrin-only gels
ARS preparation prior to ADV
ARSs were prepared by first dissolving bovine fibrin- Acoustic properties of ARSs containing 1% (v/v)
ogen (Sigma-Aldrich) in PBS at 20 mg/mL clottable pro- double emulsion, prior to ADV studies, as well as fibrin-
tein while under gentle vortex mixing for 30 s. Tubes only gels were determined using a substitution technique
containing the fibrinogen solution and PBS were degassed (Selfridge 1985; Zell et al. 2007; Parker et al. 2010). We
in a vacuum chamber (Isotemp vacuum oven, Model 282 measured ultrasound attenuation coefficient and sound
A, Fisher Scientific, Dubuque, IA, USA) at approximately speed in the above mentioned gels in the water tank
6 kPa for 60 min to minimize the amount of dissolved gas, with focused, single-element transducers centered at
which could act as cavitation nuclei. ARSs were made by 2.25 MHz, 5 MHz and 10 MHz at room temperature
combining the prepared fibrinogen, PBS, varying volume (21BC). Transducers were driven with a pulser-receiver
fraction of double emulsions (0.05%, 0.2% and 1% [v/v]) (5077 PR, Olympus, Center Valley, PA, USA) operating
and bovine thrombin (Thrombin-JMI, King Pharmaceuti- in pulse echo mode at low pressures to avoid undesirable
cals, Bristol, TN, USA). The final concentration of fibrino- non-linear propagation and ADV. Attenuation and sound
gen and thrombin in the ARSs were 10 mg/mL and speed were calculated as described previously (Fabiilli
2 U/mL, respectively. The ARS mixture was aliquoted et al. 2016; Aliabouzar et al. 2018c).
into custom-made cylindrical sample chambers (outer
diameter, 27 mm; inner diameter, 20 mm; length, 12 mm)
Determination of ADV thresholds in ARSs
made of polyvinyl chloride pipe. The open ends of the
cylindrical chamber were sealed by Tegaderm membrane Active detection technique. For active vaporization
(3 M Health Care, St. Paul, MN, USA), as shown in detection, a spherically focused immersion transducer with
Figure 1a, which permitted polymerization of the ARS a center frequency of 2.25 MHz (Model A305 S; 6 dB:
within the chamber. During polymerization, the ARSs 1.463.2 MHz), the same diameter and geometric focus as
were stored at 4˚C for 3 h before use. Fibrin-only gels (i.e., the transmitters, was used as a detector. The detecting
without double emulsions) were made in a similar manner transducer was connected to the pulser-receiver operating
as a control. Fibrin-only gels and ARSs were used on the in pulse-echo mode. The ADV and detecting transducers
same day of preparation. were placed at twice their focal lengths apart and confo-
cally aligned. Each ARS and fibrin-only gel was positioned
Ultrasound exposure setup and parameters between the two transducers, such that the axial focus was
All ADV threshold measurement experiments were positioned approximately in the middle of the ARSs and
conducted in an acrylic tank (32 cm £ 13 cm £ 13 cm) fibrin-only gels (approximately 32 mm from the surface of
filled with circulating, deionized (DI) water at 37˚C. Four the transmitter) and allowed to thermally equilibrate in the
single-element, spherically focused, immersion trans- 37˚C water bath for 15 min before the ADV experiments.
ducers (diameter = 19.05 mm, focal length = 38.1 mm, A schematic representation of the active detection method
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3249

Fig. 1. Schematic representation of three setups used to determine the ADV threshold in ARSs: (a) active detection, (b)
passive detection and (c) B-mode imaging methods. For ADV threshold determination via attenuation measurements,
the setup in panel (a) was used, and the signal from the back wall of the chamber was time gated and analyzed. The trans-
mitting and detecting transducers were confocally aligned, and the ARSs, having a thickness of 12 mm, were positioned
32 mm from the surface of the transmitter such that the axial focus was positioned approximately in the middle of the
gel. ADV = acoustic droplet vaporization; ARS = acoustically responsive scaffold; PVC = polyvinyl chloride.

is shown in Figure 1a. The ADV threshold was determined 10 MHz at 37˚C. A second transducer with a center fre-
by exposing each ARS to pulses of increasing acoustic quency of 2.25 MHz was used as an active detector in the
pressure. Following exposure at each pressure level, the pulse-echo mode. Effective attenuation in ARSs was mea-
pulse-echo signal from the detecting transducer was sured by time-gating the echo from the back wall of the
acquired directly from the oscilloscope (sampling rate of Tegaderm window and monitoring the change in its ampli-
100 MHz) and stored for offline processing. tude as the peak negative pressure was incrementally
increased. The signal amplitude at each peak negative pres-
Passive detection technique. For the passive detec- sure was compared with the signal received by the active
tion method, ADV was generated in the ARSs using the detector from the ARS for a sham exposure (i.e., 0 MPa,
aforementioned 2.25 MHz and 10 MHz transducers. A cap- ADV transducer off). Thus, no correction for reflection
sule hydrophone (HGL-1000, frequency range 120 MHz, from the Tegaderm membrane was required.
Onda, Sunnyvale, CA, USA) was placed 5 cm away from
the focus of the transducer and used as a broadband B-mode imaging. In an acrylic tank (30 cm £
receiver to detect backscattered signals from the ARSs dur- 60 cm £ 30 cm) filled with 37˚C DI water, echogenic-
ing ultrasound exposure as shown in Figure 1b. The RF sig- ity-based studies were conducted at two excitation fre-
nals, collected with the hydrophone and digitized by the quencies of 2.25 MHz and 10 MHz for ARSs containing
oscilloscope, were stored for further analysis. 0.05% (v/v) emulsion. In this method, a diagnostic ultra-
sound imaging system (ZS3, Mindray North America,
Attenuation measurements. For the ADV threshold Mountain View, CA, USA) with a 4 MHz linear array
measurements, the droplet to bubble transition was deter- (Mindray, C4-1) was used to record B-mode images for
mined by monitoring the effective attenuation of ultrasound detection of bright echoes from bubbles created by
propagating through the ARSs containing varying volume ADV. The mechanical index (MI) of the array was set to
fractions of emulsions. For this study, we employed the the lowest value of 0.31, which did not generate any
active detection setup illustrated in Figure 1a. The ARSs apparent ADV within the ARSs. Based on a MI of 0.31
were exposed to ultrasound at 2.25 MHz, 5 MHz or at 4 MHz, the maximum peak rarefactional pressure in
3250 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

the imaging plane was 0.6 MPa. B-mode images were ADV, the effective attenuation of the medium and mean
analyzed using ImageJ software (http://imagej.nih.gov). pixel intensity of the ARSs were plotted, respectively, as a
Detection for ADV bubbles was evaluated in terms of function of increasing peak negative pressure. As in other
pixel intensity in a given region of interest (ROI). ROIs, studies, we found that the peak-to-peak RF voltage, funda-
consistent in size for a given frequency, were selected at mental response, effective attenuation and the mean pixel
the ADV location in each B-mode image to measure the intensity in the ARSs remained relatively constant at
mean pixel intensity (Lo et al. 2006). acoustic pressures below the ADV threshold (i.e., sub-
threshold values). At pressures above the ADV threshold,
Vaporization efficiency measurements the metrics all increased from the sub-threshold values and
The efficiency of ADV, which was the ratio of the remained at elevated levels. Data were then fit with a
number of droplets remaining after ADV to the number piece-wise linear function with two pieces using GraphPad
of droplets before ADV, was assessed by modifying pre- Prism software (GraphPad Software, Inc., La Jolla, CA,
viously described methods (Fabiilli et al. 2010; Moncion USA). The ADV threshold was defined as the peak nega-
et al. 2016b; Moncion et al. 2017). ARSs containing tive pressure corresponding to the intersection of the two
0.05% and 1% (v/v) emulsion were exposed to acoustic linear fits (Radhakrishnan et al. 2013).
pressures at and above the ADV threshold at an excita-
tion frequency of 2.25 MHz. The focal region (i.e., loca- Statistical analysis
tion of ADV) in each ARS was removed using a 6 mm All statistical analyses were performed using Graph-
biopsy punch (Miltex, Inc. York, PA, USA) and then Pad Prism software. All data are presented as the mean
digested in 0.25% trypsinethylenediaminetetraacetic and corresponding standard deviations. ADV thresholds
acid (Life Technologies) for 3 h in the fridge. As a con- are listed as peak negative pressure. For each frequency
trol, this procedure was repeated for ARSs that were not and concentration, measurements were repeated twice per
exposed to ultrasound. The number of droplets in the gel in three independently produced ARSs (n = 6) at each
digested samples was counted using a Coulter counter excitation amplitude unless otherwise stated. The 95%
and compared with the control to find the percentage of confidence intervals of the calculated parameters are listed
vaporized emulsions at each supra-threshold pressure. in the format of [SL, SH], where SL is the lower bound
value and SH is the upper bound value. Significant differ-
Criteria for determining ADV thresholds ences between groups were determined using a one-way
Upon vaporization, a PFC droplet undergoes a volu- analysis of variance followed by Tukey’s multiple com-
metric expansion of about 125-fold, as shown by Kripf- parisons test, with a significance level of 0.05.
gans et al. (2005), Rapoport et al. (2009) and Sheeran
et al. (2011), resulting in a sudden increase in scattering.
RESULTS AND DISCUSSION
For the active and passive detection methods, the peak-to-
peak RF voltage and the fundamental frequency compo- Acoustic characterization of fibrin-only gels and ARSs
nent of the scattered response from the ARSs were plotted prior to ADV measurements
as a function of peak negative pressure. Similarly, for The frequency-dependent attenuation and speed of
attenuation measurements and B-mode imaging following sound in fibrin-only gels and ARSs with 1% (v/v)

Fig. 2. (a) Frequency-dependent attenuation and (b) speed of sound in fibrin gels and ARSs with 1% emulsions (n = 6).
Speed of sound did not significantly change in either scaffold as a function of frequency. Speed of sound was signifi-
cantly (p < 0.05) higher in fibrin than ARS with 1% emulsion. ARS = acoustically responsive scaffold.
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3251

emulsion are shown in Figure 2. The attenuation data frequency, where n is approximately between 1 and 2
were compiled from all three transducers (i.e., 2.25, 5 for soft tissues over a broad frequency range from
and 10 MHz), thus covering the range of 112 MHz. 150 MHz (Madsen et al. 1998; Chaffaı et al. 2000;
Linear fits of the attenuation data yielded slopes of 0.08 Rabell-Montiel et al. 2018).
§ 0.005 dB cm1 MHz1 (R2 = 0.91, [0.07, 0.09]) and
0.14 § 0.006 dB cm1 MHz1 (R2 = 0.94, [0.13, 0.15])
ADV threshold determination in ARSs
for fibrin gels and ARSs, respectively. The measured
attenuation of the Tegaderm membrane, which consti- Active detection method. Figure 3 shows the peak-
tuted the window of the sample chamber, with respect to to-peak voltage and the fundamental component of the
free field water was 0.12 § 0.07 dB cm1 MHz1, scattered response, acquired using the active technique,
which is close to the value of 0.2 dB/MHz reported by from ARSs containing varying volume fractions of dou-
Deng et al. (1996). ble emulsion as well as fibrin-only gels (i.e., control) at
Note that the frequency-dependent attenuation val- excitation frequencies of 2.25 MHz and 10 MHz. Scat-
ues are with respect to a water-filled sample chamber to tered responses from ARSs remained constant before
account for attenuation by the Tegaderm windows. The ADV (Fig. 3). However, once the bubbles were formed
speed of sound in water at room temperature (21˚C) was as a result of ADV, the scattered signal increased sharply
taken to be 1483 m s1 (Greenspan and Tschiegg 1959; because of the higher impedance mismatch and com-
Duck 2013). The attenuation coefficients of fibrin gel pressibility of gas-filled bubbles compared with liquid
and ARSs show a linear dependence on frequency, while droplets. Using piece-wise linear fitting (as described
the speed of sound has negligible frequency-dependence above), the ADV thresholds were 2.55 § 0.11 MPa, 3.15
within the range studied here. Attenuation coefficient § 0.13 MPa and 3.67 § 0.04 MPa for ARSs with 1%,
has been reported to be proportional to the nth power of 0.2% and 0.05% (v/v) emulsions, respectively, at

Fig. 3. Peak-to-peak voltage and fundamental components of the scattered response for ARSs containing varying vol-
ume fractions (0.05%, 0.2% and 1% [v/v]) of double emulsions at the excitation frequencies of (a, b) 2.25 MHz and (c,
d) 10 MHz, recorded using active method. A linear piece-wise function was fit to the data points. The ADV threshold
was defined as the pressure where the first two lines of the piece-wise function intersected (indicated by arrows in
Fig. 3a). ADV = acoustic droplet vaporization; ARS = acoustically responsive scaffold.
3252 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

2.25 MHz. A similar trend in the threshold was observed sub-threshold pressures, the fundamental response from
for the fundamental component of the scattered response ARSs with 1% (v/v) and 0.2 % (v/v) double emulsions
for ARSs at this excitation frequency (Fig. 3b). Note that was significantly greater than fibrin alone, as shown in
the scattered fundamental response from PFH double Figure 4a at 2.25 MHz. The fundamental response of an
emulsions seemed to saturate at higher excitation pres- ARS with 1% (v/v) double emulsion was 7.1 and 16.6-
sures at 2.25 MHz. Similar saturation behavior can also fold higher compared with ARSs with 0.2% (v/v) and
be seen in ADV studies of PFC liquids (Reznik et al. 0.05% (v/v) emulsions, respectively. This was because
2011; Moncion et al. 2017) as well as numerical predic- of a lower speed of sound in PFC liquids, which resulted
tions of scattered responses from microbubbles (Sarkar in larger acoustic impedance mismatch between PFH
et al. 2005). At an excitation frequency of 10 MHz, the emulsions (approximately 0.8 MRayl) and water
ADV thresholds of ARSs with 0.2% and 0.05% (v/v) (approximately 1.48 MRayl). The echogenic properties
emulsions were 5.16 § 0.07 MPa and 4.76 § 0.09 MPa, of PFC droplets have been investigated in previous stud-
respectively (Fig. 3c and 3d). For ARSs with 1% (v/v) ies (Couture et al. 2006; Rapoport et al. 2009). Note that
emulsions, no ADV was detected at an excitation fre- Couture et al. (2006) reported a linear correlation
quency of 10 MHz for the maximum peak negative pres- between the scattered response and volume fraction of
sure of 6 MPa studied here. (0.03%0.5%) sub-micron PFH particles. Such linear
A comparison of the slope (a) of each line above behavior also demonstrates that the measurements are
the ADV threshold in Figure 3a shows a three-fold not affected by multiple scattering.
increase for ARSs with lower emulsion volume fractions Similarly, at 10 MHz, the slope of the sub-threshold
of 0.2% (a = 0.11, R2 = 0.96, [0.08, 0.13]) and 0.05% region was not significantly different than zero for all
(a = 0.1, R2 = 0.98, [0.08, 0.11]) compared with 1% ARSs. The fundamental response of all ARSs with vary-
emulsion (a = 0.03, R2 = 0.95, [0.02, 0.04]). The steeper ing volume fraction of emulsions was higher than fibrin-
slopes in ARSs with lower volume fraction of emulsions only gels (Fig. 4b). At 10 MHz, the average fundamental
could be an indication of a higher vaporization effi- response of ARSs with 1% (v/v) emulsion was 7.8- and
ciency. It has been shown in previous studies that higher 17.5-fold higher than ARSs with 0.2% (v/v) and 0.05%
volume fractions of vaporized droplets within the focus (v/v) emulsion, respectively.
could shield the acoustic energy from propagating fur- A summary of the average ADV thresholds in ARSs
ther, therefore hindering additional ADV (Xin et al. of varying emulsion volume fraction at different excita-
2017). Similarly, in Figure 3c, slopes of 0.29 (R2 = 0.99, tion frequencies, determined using the active method, is
[0.28, 0.32]) and 0.21 (R2 = 0.96, [0.16, 0.25]) were mea- displayed in Figure 5. Overall, the ADV threshold corre-
sured for ARSs with 0.2% (v/v) and 0.05% (v/v) emul- lated with the excitation frequency, with the dynamic
sions, respectively. range of the threshold at each frequency correlating with
Additionally, we studied the effect of emulsion vol- the emulsion concentration. At 2.25 MHz, the ADV
ume fraction on the fundamental response of the ARSs threshold correlated inversely with emulsion concentra-
for sub-threshold pressures (i.e., 02.5 MPa) as shown tion because of the enhanced probability of nucleation
in Figure 4. In this region, the slope of the fundamental (Reznik et al. 2011; Lu et al. 2019). However, at higher
response was not significantly different than zero. For frequencies (i.e., 7.5 MHz and 10 MHz), an inverse

Fig. 4. At sub-threshold peak negative pressures (i.e., 02.5 MPa), the fundamental response of ARSs at 2.25 MHz (a)
and 10 MHz (b) correlated with emulsion volume fractions. Statistically significant differences (p < 0.05) are denoted as
follows: a: versus 1%; b: versus 0.2%; g: versus 0.05%. ARS = acoustically responsive scaffold.
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3253

Fig. 5. Average ADV thresholds in ARSs with varying volume fractions (0.05%, 0.2% and 1% [v/v]) of double emulsion
as a function of frequency of excitation, determined using the (a) active and (b) passive detection methods.
ADV = acoustic droplet vaporization; ARS = acoustically responsive scaffold.

correlation was observed between the ADV threshold resolution. Note that in the context of transient cavita-
and emulsion concentration in the ARS. At each fre- tion, Roy et al. (1990) showed that thresholds measured
quency, there were significant differences in the ADV by the active method were lower than or equal to those
thresholds of ARSs with varying volume fraction of detected by passive methods.
emulsions, with the exception of ARSs with 0.2% (v/v)
and 0.05% (v/v) emulsions at 5 MHz. In addition, when Attenuation measurements. Droplet-to-bubble tran-
comparing the same concentration across different fre- sition was also studied by measuring the effective attenu-
quencies, no significant differences in the threshold were ation through the ARSs at sub- and supra-threshold
observed between ARSs with 0.05% (v/v) emulsion at pressures at 2.25 MHz. The results are shown in Figure 6.
2.25 MHz, 5 MHz and 7.5 MHz, as well as ARSs with The effective attenuation of the ARSs increased sharply
0.2% (v/v) emulsion at 5 MHz and 7.5 MHz. above the ADV threshold because of the presence of
vaporized droplets in the acoustic path and was higher at
Passive detection method. With the passive detec- an excitation frequency of 2.25 MHz (Fig. 6a) compared
tion technique, the profiles of peak-peak voltage and fun- with 5 MHz (Fig. 6b) and 10 MHz (Fig. 6c). The gener-
damental responses displayed similar trends as in ated bubble cloud correlated inversely with frequency,
Figure 3 (data not shown). Figure 5b shows the ADV which was consistent with the size of the focal volume.
threshold of ARSs, with varying emulsion volume frac- The focal volume, calculated from the transducer specifi-
tion, at excitation frequencies of 2.25 MHz and 10 MHz cations, at 2.25 MHz was approximately 30 mm3 and 0.3
obtained using the passive detection method. There was mm3 at 10 MHz (Olympus 2006).
no significant difference in thresholds for ARSs with Similar to Figure 5, at each excitation frequency,
0.2% (v/v) and 0.05% (v/v) emulsions at 10 MHz. Pas- there was a significant difference in ADV thresholds
sive thresholds displayed a similar trend with respect to between ARSs with varying volume fraction of emul-
frequency and volume fraction, as was observed in sions, except for ARSs with 0.2% (v/v) and 0.05% (v/v)
Figure 5a, using active detection. Active and passive emulsion at 5 MHz. There were significant differences
techniques have been previously used as classic methods among ADV thresholds when comparing the same vol-
for detection of inertial cavitation in vitro (Roy et al. ume fraction of emulsions across frequencies, with the
1990; Holland et al. 1996). While active methods have exception of ARSs with 0.05% (v/v) at 2.25 MHz and
higher sensitivity to detect weak or single bubble oscilla- 5 MHz.
tions, emissions detected by passive techniques are more The measured attenuation was due to absorption
associated with resonance-sized bubbles or clouds of and scattering of bubbles generated by ADV. As men-
bubbles (Holland et al. 1992; Ding et al. 2015). The tioned earlier, upon vaporization, the PFC phase under-
major drawback of active detection is production of goes a volumetric expansion, which is approximately
acoustic bursts that might interfere with the transmitting 125-fold in water (Rapoport et al. 2009; Sheeran et al.
signal. However, unlike passive detection, active meth- 2011). The presence of a viscoelastic medium such as
ods can be used to determine where and when vaporiza- gels will reduce this expansion factor (Qamar et al.
tion is taking place, allowing both spatial and temporal 2010; Sheeran et al. 2011). Since the produced bubbles
3254 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

Fig. 6. Effective acoustic attenuation of ARSs containing varying volume fractions of double emulsions as a function of
peak negative pressure at excitation frequencies of (a) 2.25 MHz, (b) 5 MHz and (c) 10 MHz. The attenuation was mea-
sured at 2.25 MHz using a pulse echo technique. ARS = acoustically responsive scaffold.

due to ADV are expected to be around 4050 mm in shown in Figure 7 (af) at 2.25 MHz and (gl) at
diameter, which accounts that each double emulsion 10 MHz at increasing peak negative pressures. Specular
droplet contains 67% (v/v) PFH, attenuation due to scat- reflection from the two windows of the sample chamber
tering will be more prominent since the product of the can be seen in these figures. Based on mean pixel inten-
wave number and scatterer size approaches one. sity measurements, the ADV threshold in ARSs with
At the frequencies studied here, we did not observe 0.05% (v/v) emulsions was 3.58 § 0.26 MPa (n = 5) at
any change in the speed of sound in the ARSs as ADV 2.25 MHz and 4.65 § 0.23 MPa at 10 MHz (n = 5).
bubbles were generated. Note that speed of sound in a The change in the shape of the bubble cloud from
mixure of bubbles will not depend on frequency at exci- ellipsoidal (Fig. 7d) to a tadpole (Fig. 7f) due to acoustic
tation frequencies much higher or lower than the reso- shadowing was consistent with the previously published
nance frequency of the bubbles at low volume fractions theoretical predictions (Xin et al. 2017). Variabilty in
of bubbles (Commander and Prosperetti 1989; Povey cloud shape and size, in a backward propagating direc-
1997; Wilson and Roy 2008). Bubbles with a diameter tion toward the transducer, with increasing amplitude
of 4050 mm will have a resonance frequency in the and pulse number, was also observed by Lo et al. (2006).
hundred-kilohertz range and therefore will be non-reso-
nant at frequencies studied here. Comparison of ADV threshold detection methods
Figure 8 is a comparison of the ADV thresholds
B-mode imaging. Acoustic methods used here to measured using four different techniques of active, pas-
study ADV are based on scattering emission and are 1-D sive and attenuation along with B-mode imaging. For
detection techniques, with the inability to spatially map the different techniques, thresholds were not statistically
the distribution of vaporization. However, echogenicity different, except for ARSs with 0.05% (v/v) and 0.2%
based techniques, such as conventional B-mode imaging, (v/v) emulsion at 10 MHz. With the attenuation method,
can effectively reflect the spatial distribution of ADV. B- ADV thresholds in 0.05% and 0.2% (v/v) ARSs were
mode images from ARSs with 0.05% (v/v) emulsion are 5.14 MPa (5.2% higher) and 5.59 MPa (8% higher)
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3255

Fig. 7. B-mode images of ARSs with 0.05% (v/v) double emulsion exposed to ultrasound at increasing peak negative
pressures of (a) 1.1 MPa, (b) 3.2 MPa, (c) 4.2 MPa, (d) 4.8 MPa, (e) 5.8 MPa and (f) 6.1 MPa at 2.25 MHz frequency.
Exposures were repeated at 10 MHz at peak negative pressures of (g) 0.9 MPa, (h) 2.7 MPa, (i) 4.1 MPa, (j) 5 MPa, (k)
5.8 MPa and (l) 6.1 MPa. Direction of the sound wave is shown by the arrow. Scale bar shows 5 mm. ARS = acoustically
responsive scaffold.

Fig. 8. Comparison of ADV thresholds determined using different techniques including active, passive, attenuation and
echogenicity methods at (a) 2.25 MHz and (b) 10 MHz. Statistically significant differences (p < 0.05) are denoted as
follows: a: versus 0.05% (v/v) attenuation and b: versus 0.2% (v/v) attenuation. ADV = acoustic droplet vaporization.

compared with thresholds measured using other techni- frequency of 2.25 MHz. At higher applied pressures, the
ques. This difference was possibly because of the small supra-threshold focal volume increased, along with the
size of the generated bubble cloud having a negligible ADV efficiency (Fig. 9a). This has also been previously
effect on attenuation of the propagating signal. observed (Radhakrishnan et al. 2016). ADV efficiency
increased linearly with the applied peak negative pres-
ADV efficiency in ARSs sure for ARSs with 0.05% (v/v) emulsion while it satu-
In Figure 9, we determined the vaporization effi- rated, possibly due to acoustic shadowing, at higher
ciency at 0 MPa (i.e., sub-threshold), at the ADV thresh- pressures for the ARSs with 1% (v/v) emulsion. This is
old and at supra-threshold pressures for ARSs containing also consistent with the calculated slopes of lines above
0.05% and 1% (v/v) double emulsions at an excitation the ADV threshold shown earlier in Figure 3. Recently,
3256 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

Fig. 9. (a) Efficiency of vaporization in ARSs containing 1% (n = 5) and 0.05% (v/v) (n = 6) double emulsions as a func-
tion of peak negative pressure at the excitation frequency of 2.25 MHz. (b) Corresponding attenuation in the ARSs as
the ADV efficiency increased at an excitation frequency of 2.25 MHz. The effective attenuation of the gel was measured
at 2.25 MHz. ADV = acoustic droplet vaporization; ARS = acoustically responsive scaffold.

we demonstrated that the ADV efficiency for ARSs with et al. 2000; Kawabata et al. 2006; Schad and Hynynen
PFH emulsion was less dependent on ultrasound expo- 2010), which is opposite to the current results. The
sure parameters, such as lateral raster spacing and num- potential mechanisms proposed for this inverse correla-
ber of axial planes, compared with ARSs with higher tion have included droplet deformation (Kripfgans et al.
boiling point PFCs. In general, a high ADV efficiency is 2004), inertial cavitation (Fabiilli et al. 2009), non-linear
desired in applications where ADV is used to trigger propagation and super-harmonic focusing due to the low
release of a therapeutic agent from an ARS (Lu et al. speed of sound in PFCs, as well as the presence of higher
2019). harmonics at the focus (Li et al. 2014; Shpak et al. 2014;
As seen in Figure 9b, the effective acoustic attenua- Miles et al. 2016). However, all these studies were done
tion correlated with the ADV efficiency. Monitoring the in water, which is a non-attenuating, non-linear medium
droplet-to-bubble transition by measuring the change in (Khelladi 2012). The Goldberg number, defined as the
the acoustic properties of the medium, such as attenua- ratio of non-linear distortion over attenuation, is much
tion, could enable new methods of monitoring the dose higher for water than tissue (Szabo et al. 1999). For
of a therapeutic payload released by ADV within an example, the Goldberg ratio for water at 5 MPa and
ARS in situ. Previously, acoustic mapping has been used 5 MHz is 266 and for fibrin gel is 12 (i.e., similar to soft
to co-localize cavitation activity with drug delivery tissues such as breast and liver). This indicates that non-
(Choi et al. 2014). Further studies are required to investi- linear distortion is easier to achieve in water than tissue.
gate the growth mechanism of ADV bubbles in the vis- To further investigate this, we studied wave propa-
coelastic fibrin gels over the physiologically relevant gation and generation of harmonics in ARSs with 1%
time points and their potential impact on the payload (v/v) emulsion, fibrin gels and free-field water using a
release. Parameters such as PFC boiling point, volume fiber optic hydrophone as a passive receiver. Waveforms
fraction of the emulsions, dissolved gas concentration, in the time domain, recorded by the fiber optic hydro-
ambient pressure, temperature and fibrin concentration phone at each excitation amplitude and frequency, were
will impact the physical stability of an ARS post-ADV. converted to the frequency domain using the fast Fourier
We have previously shown that the volumetric expan- transform (FFT) to obtain the harmonics. Figures 10a
sion of ARSs following ADV correlated inversely with and 10b show the fundamental and third harmonic
the bulk boiling point of the PFC liquid, with a 233% responses as ultrasound at 2.25 MHz propagated through
volumetric increase for ARSs containing perfluoropen- the three different media. Analogous data at 10 MHz is
tane emulsions (Lu et al. 2019). For ARSs implanted displayed in Figures 10c and 10d. Attenuation dimin-
in vivo, the percent of fibrin remaining in ARSs contain- ished the generation of harmonics by decreasing their
ing PFH emulsions was 39% § 9.1% at day 14 post- amplitudes, with a greater impact at 10 MHz versus
ADV (Moncion et al., 2016b). 2.25 MHz. Because of the finite active element size of
the fiber optic hydrophone (100 mm) and associated spa-
Free field and in situ wave propagation tial averaging errors (wavelength of 50 mm at 30 MHz),
Earlier studies of ADV have shown a decreasing harmonics above the third were not analyzed. However,
trend in threshold with increasing frequency (Kripfgans higher harmonics, which are more effective for the
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3257

Fig. 10. Effect of the propagating media on generation of harmonics at (a, b) 2.25 MHz and (c, d) 10 MHz. Waveforms
were recorded in the free field and in the presence of a fibrin gel and an ARS containing 1% (v/v) double emulsion using
a fiber optic hydrophone. Recorded waveforms were Fourier transformed using the FFT to obtain the harmonics. For
ARSs containing 1% (v/v) double emulsions, measurements were done up to 4 MPa at 2.25 MHz and 5.5 MPa at
10 MHz to avoid ADV at room temperature. ADV = acoustic droplet vaporization; ARS = acoustically responsive scaf-
fold; FFT = fast Fourier transform.

focusing effect (Li et al. 2014; Shpak et al. 2014), are frequencies. Although several authors have shown negli-
expected to be even more attenuated. In addition to the gible dependence of ADV threshold on the number of
effect of the viscoelastic fibrin matrix on ultrasound cycles and microsecond pulse lengths (Lo et al. 2007;
propagation, the effect of dissolved gas concentration Reznik et al. 2011; Aliabouzar et al. 2018a), the nucle-
should be considered as well. All the presented experi- ated vapor bubble may be disrupted when exposed to
ments were conducted in DI water (i.e., non-degassed more cycles at high pressures because of inertial cavita-
medium), in which small changes in pressure could tion or re-condensation (Reznik et al. 2013).
impact the dissolved gas content. This could affect sound The effect of droplet concentration in the ARS on
propagation, with a greater impact at higher frequencies wave propagation was investigated in a similar manner.
(Povey 1997). It is also noteworthy to mention that the Figures 11a and 11b compare the frequency spectrum of
patterns of homogenous nucleation in pure PFC liquid an acoustic wave with a peak negative pressure of 3.5
droplets, studied by previous authors using high speed MPa as it propagated through water and in an ARS con-
cameras (Kripfgans et al. 2004; Shpak et al. 2013), may taining 1% (v/v) double emulsion at 2.25 MHz and
not necessarily be the same for double emulsions, where 10 MHz, respectively.
heterogeneous nucleation may occur. Further studies As shown in Figure 11a, the frequency spectrum of the
using high speed cameras are required to further eluci- signal from an ARS containing 1% (v/v) emulsion was sim-
date the vaporization dynamics of double emulsions. ilar to that of water at 2.25 MHz at this high-pressure ampli-
The other factor that could impact the trend of ADV tude. However, the higher order harmonics were highly
threshold with frequency is the constant pulse length of damped in the presence of an ARS with 1% (v/v) emulsion
5.4 ms that was used throughout the ADV studies, which at 10 MHz compared with the free field (see Fig. 11b).
yields an increasing number of cycles for higher Thus, the double emulsion in the ARS can act as a low pass
3258 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

Fig. 11. Frequency spectrum of the pressure waveform, normalized to the amplitude of the fundamental component, in
free field water and in the presence of an ARS containing 1% (v/v) double emulsion at a peak negative pressure of 3.5
MPa at (a) 2.25 MHz and (b) 10 MHz. ARS = acoustically responsive scaffold.

filter at higher frequencies of excitation, decreasing the dependence of ADV on droplet concentrations over the
amplitudes of higher harmonics. Although droplets are range 0.002%0.1% (v/v). The inverse correlation
weak, linear scatterers of ultrasound, compared with gas- between ADV threshold and volume fraction of emul-
filled microbubbles, the reflection due to the acoustic sions, as a result of increased number of vaporization
impedance mismatch can be strong (Marsh et al. 2002; Cou- events at the focus, was only observed here at the lowest
ture et al. 2006; Rapoport et al. 2009). Reflection due to the frequency (2.25 MHz), where the focal volume was
impedance mismatch of micron-size PFH emulsions was almost 100 times larger than that at 10 MHz. At higher
shown earlier in Figure 3. Therefore, as the volume fraction frequencies, because of the combined effects of focal
of PFH emulsions in the ARSs increased, the scattering volume and acoustic impedance mismatch, lower vol-
cross section and attenuation due to scattering also ume fractions have lower ADV thresholds.
increased.
Note that in the long wavelength limit (i.e., Ray-
leigh scattering), scattering intensity is proportional to
CONCLUSION
the fourth power of frequency and sixth power of particle
size (Povey 1997), which could explain the trend In this work, we systematically investigated the
observed with the ADV threshold and volume fraction at effect of frequency of excitation on the ADV thresholds
higher frequencies. Couture et al. (2006) showed a linear in ARSs containing varying volume fractions of mono-
increase in the overall scattering from sub-micron PFH dispersed, double emulsions. We studied the ADV phe-
droplets with concentration as well as frequency. Fur- nomenon through acoustic means, including active and
thermore, several studies reported significant enhance- passive detection techniques, attenuation measurements
ment (up to 32 dB) in surface reflectivity of substrates and enhanced echogenicity of B-mode images. The
when targeted with sub-micron PFH emulsions at high ADV thresholds determined via these four techniques
frequencies, while less enhancement was observed when were consistent with respect to the frequency of excita-
the same particles were in suspension (Lanza et al. 1998; tion and emulsion volume fraction. Results showed that
Marsh et al. 2002). ADV threshold increased with the frequency of excita-
Previous studies have reported a decreasing trend of tion. We showed that the propagating media affected the
ADV threshold with concentration due to the enhanced generation of non-linearly distorted waves, diminishing
probability of nucleation at the focus (Reznik et al. the amplitude of harmonics at higher frequencies and
2011; Arena et al. 2015). In these aforementioned stud- higher volume fractions of double emulsions and there-
ies, a single excitation frequency was used to phase-tran- fore potentially inhibiting super-harmonic focusing. The
sition polydispersed droplets at concentrations up to ADV threshold decreased with increasing volume frac-
0.3% (v/v). Because of the polydispersed droplet popula- tion of double emulsions in the focal volume at the low-
tion, an increase in concentration also increased the like- est excitation frequency studied here. However, at
lihood that larger droplets traversed the region of higher frequencies, as a result of high acoustic reflectiv-
excitation and therefore reduced the ADV threshold. At ity of PFH emulsions, the ADV threshold correlated
10 MHz, Williams et al. (2013) observed no significant directly with the volume fraction of emulsions. Future
ADV in acoustically responsive scaffolds  M. ALIABOUZAR et al. 3259

studies will employ the presented findings to optimize triggered perfluoropentane emulsion. Ultrasound Med Biol
regenerative studies using ARSs. 2010;36:1364–1375.
Fabiilli ML, Wilson CG, Padilla F, Martın-Saavedra FM, Fowlkes JB,
Acknowledgments—This work was supported by NIH grant R01 Franceschi RT. Acoustic droplethydrogel composites for spatial
HL139656 (MLF). and temporal control of growth factor delivery and scaffold stiff-
ness. Acta Biomater 2013;9:7399–7409.
Conflict of interest disclosure—The authors declare no competing Fabiilli ML, Phanse RA, Moncion A, Fowlkes JB, Franceschi RT. Use
interests. of hydroxyapatite doping to enhance responsiveness of heat-induc-
ible gene switches to focused ultrasound. Ultrasound Med Biol
REFERENCES 2016;42:824–830.
Giesecke T, Hynynen K. Ultrasound-mediated cavitation thresholds of
Aliabouzar M, Kumar K, Sarkar K. Effects of acoustic parameters and liquid perfluorocarbon droplets in vitro. Ultrasound Med Biol
bulk fluid properties on acoustic droplet vaporization threshold of per- 2003;29:1359–1365.
fluoropentane droplets. In: Proceedings of the 10th International Sym- Greenspan M, Tschiegg CE. Tables of the speed of sound in water. J
posium on Cavitation (CAV2018). New York: ASME Press; 2018a. Acoust Soc Am 1959;31:75–76.
Aliabouzar M, Kumar KN, Sarkar K. Acoustic vaporization threshold Holland CK, Roy R, Apfel R, Crum L. In vitro detection of cavitation
of lipid-coated perfluoropentane droplets. J Acoust Soc Am induced by a diagnostic ultrasound system. IEEE Trans Ultrason
2018b;143:2001–2012. Ferroelectr Freq Control 1992;39:95–101.
Aliabouzar M, Zhang GL, Sarkar K. Acoustic and mechanical charac- Holland CK, Deng CX, Apfel RE, Alderman JL, Fernandez LA, Taylor
terization of 3 D-printed scaffolds for tissue engineering applica- KJ. Direct evidence of cavitation in vivo from diagnostic ultra-
tions. Biomed Mater 2018c;13 055013. sound. Ultrasound Med Biol 1996;22:917–925.
Aliabouzar M, Kumar KN, Sarkar K. Effects of droplet size and per- Kawabata K-i, Yoshizawa A, Asami R, Azuma T, Yoshikawa H, Wata-
fluorocarbon boiling point on the frequency dependence of acoustic nabe H, Sasaki K, Hirata K, Umemura S-i. 1 F-3 site-specific con-
vaporization threshold. J Acoust Soc Am 2019;145:1105–1116. trast imaging with locally induced microbubbles from liquid
Arena CB, Novell A, Sheeran PS, Puett C, Moyer LC, Dayton PA. precursors. 2006 IEEE ultrasonics symposium. Vancouver, Can-
Dual frequency acoustic croplet vaporization detection for medical ada: IEEE; 2006. p. 517–520.
imaging. IEEE Trans Ultrason Ferroelectr Freq Control Khelladi H. Goldberg’s number influence on the validity domain of the
2015;62:1623–1633. quasi-linear approximation of finite amplitude acoustic waves.
Asami R, Kawabata K. Repeatable vaporization of optically vaporiz- Ultrasonic Waves. London: InTech; 2012.
able perfluorocarbon droplets for photoacoustic contrast enhanced Kripfgans OD, Fowlkes JB, Miller DL, Eldevik OP, Carson PL. Acous-
imaging. 2012 IEEE International Ultrasonics Symposium. Dres- tic droplet vaporization for therapeutic and diagnostic applications.
den, Germany: IEEE; 2012. p. 1200–1203. Ultrasound Med Biol 2000;26:1177–1189.
Chaffaı S, Padilla F, Berger G, Laugier P. In vitro measurement of the Kripfgans OD, Fabiilli ML, Carson PL, Fowlkes JB. On the acoustic
frequency-dependent attenuation in cancellous bone between 0.2 vaporization of micrometer-sized droplets. J Acoust Soc Am
and 2 MHz. J Acoust Soc Am 2000;108:1281–1289. 2004;116:272–281.
Choi JJ, Carlisle RC, Coviello C, Seymour L, Coussios C-C. Non-inva- Kripfgans OD, Orifici CM, Carson PL, Ives KA, Eldevik OP, Fowlkes
sive and real-time passive acoustic mapping of ultrasound-medi- JB. Acoustic droplet vaporization for temporal and spatial control
ated drug delivery. Phys Med Biol 2014;59:4861–4877. of tissue occlusion: A kidney study. IEEE Trans Ultrason Ferroe-
Church CC. Spontaneous homogeneous nucleation, inertial cavitation lectr Freq Control 2005;52:1101–1110.
and the safety of diagnostic ultrasound. Ultrasound Med Biol Lanza GM, Trousil RL, Wallace KD, Rose JH, Hall CS, Scott MJ,
2002;28:1349–1364. Miller JG, Eisenberg PR, Gaffney PJ, Wickline SA. In vitro charac-
Commander KW, Prosperetti A. Linear pressure waves in bubbly terization of a novel, tissue-targeted ultrasonic contrast system with
liquids: Comparison between theory and experiments. J Acoust acoustic microscopy. J Acoust Soc Am 1998;104:3665–3672.
Soc Am 1989;85:732–746. Li DS, Kripfgans OD, Fabiilli ML, Brian Fowlkes J, Bull JL. Initial
Couture O, Bevan PD, Cherin E, Cheung K, Burns PN, Foster FS. nucleation site formation due to acoustic droplet vaporization.
Investigating perfluorohexane particles with high-frequency ultra- Appl Phys Lett 2014;104 063703.
sound. Ultrasound Med Biol 2006;32:73–82. Li DS, Schneewind S, Bruce M, Khaing Z, O’Donnell M, Pozzo L.
Couture O, Urban A, Bretagne A, Martinez L, Tanter M, Tabeling P. In Spontaneous nucleation of stable perfluorocarbon emulsions for
vivo targeted delivery of large payloads with an ultrasound clinical ultrasound contrast agents. Nano Lett 2018;19:173–181.
scanner. Med Phys 2012;39:5229–5237. Lo AH, Kripfgans OD, Carson PL, Fowlkes JB. Spatial control of gas
de Gracia Lux C, Vezeridis A, Lux J, Armstrong A, Sirsi S, Hoyt K, bubbles and their effects on acoustic fields. Ultrasound Med Biol
Mattrey R. Novel method for the formation of monodisperse super- 2006;32:95–106.
heated perfluorocarbon nanodroplets as activatable ultrasound con- Lo AH, Kripfgans OD, Carson PL, Rothman ED, Fowlkes JB. Acoustic
trast agents. RSC Adv 2017;7:48561–48568. droplet vaporization threshold: Effects of pulse duration and con-
Deng CX, Xu Q, Apfel RE, Holland CK. In vitro measurements of trast agent. IEEE Trans Ultrason Ferroelectr Freq Control
inertial cavitation thresholds in human blood. Ultrasound Med Biol 2007;54:933–946.
1996;22:939–948. Lu X, Dong X, Natla S, Kripfgans OD, Fowlkes JB, Wang X, France-
Ding T, Yin H, Hu H, Bai C, Wan M. Cavitation mapping. Cavitation schi R, Putnam AJ, Fabiilli ML. Parametric study of acoustic drop-
in biomedicine. New York: Springer; 2015. p. 47–114. let vaporization thresholds and payload release from acoustically-
Duck FA. Physical properties of tissues: A comprehensive reference responsive scaffolds. Ultrasound Med Biol 2019;45:2471–2484.
book. York, United Kingdom: Academic Press; 2013. Madsen EL, Frank GR, Dong F. Liquid or solid ultrasonically tissue-
Duncanson WJ, Arriaga LR, Ung WL, Kopechek JA, Porter TM, Weitz mimicking materials with very low scatter. Ultrasound Med Biol
DA. Microfluidic fabrication of perfluorohexane-shelled double 1998;24:535–542.
emulsions for controlled loading and acoustic-triggered release of Marsh JN, Hall CS, Scott MJ, Fuhrhop RW, Gaffney PJ, Wickline SA,
hydrophilic agents. Langmuir 2014;30:13765–13770. Lanza GM. Improvements in the ultrasonic contrast of targeted per-
Fabiilli ML, Haworth KJ, Fakhri NH, Kripfgans OD, Carson PL, fluorocarbon nanoparticles using an acoustic transmission line model.
Fowlkes JB. The role of inertial cavitation in acoustic droplet IEEE Trans Ultrason Ferroelectr Freq Control 2002;49:29–38.
vaporization. IEEE Trans Ultrason Ferroelectr Freq Control Miles CJ, Doering CR, Kripfgans OD. Nucleation pressure threshold in
2009;56:1006–1017. acoustic droplet vaporization. J Appl Phys 2016;120 034903.
Fabiilli ML, Haworth KJ, Sebastian IE, Kripfgans OD, Carson PL, Moncion A, Arlotta KJ, Kripfgans OD, Fowlkes JB, Carson PL, Put-
Fowlkes JB. Delivery of chlorambucil using an acoustically- nam AJ, Franceschi RT, Fabiilli ML. Design and characterization
3260 Ultrasound in Medicine & Biology Volume 45, Number 12, 2019

of fibrin-based acoustically responsive scaffolds for tissue engi- acoustic vaporization of submicron perfluorocarbon droplets. Ultra-
neering applications. Ultrasound Med Biol 2016a;42:257–271. sonics 2013;53:1368–1376.
Moncion A, Arlotta KJ, O’Neill EG, Lin M, Mohr LA, Franceschi RT, Roy RA, Madanshetty SI, Apfel RE. An acoustic backscattering
Kripfgans OD, Putnam AJ, Fabiilli ML. In vitro and in vivo assess- technique for the detection of transient cavitation produced by
ment of controlled release and degradation of acoustically respon- microsecond pulses of ultrasound. J Acoust Soc Am
sive scaffolds. Acta Biomater 2016b;46:221–233. 1990;87:2451–2458.
Moncion A, Lin M, O’Neill EG, Franceschi RT, Kripfgans OD, Put- Sarkar K, Shi WT, Chatterjee D, Forsberg F. Characterization of ultra-
nam AJ, Fabiilli ML. Controlled release of basic fibroblast growth sound contrast microbubbles using in vitro experiments and viscous
factor for angiogenesis using acoustically-responsive scaffolds. and viscoelastic interface models for encapsulation. J Acoust Soc
Biomater 2017;140:26–36. Am 2005;118:539–550.
Mountford PA, Borden MA. On the thermodynamics and kinetics of Schad KC, Hynynen K. In vitro characterization of perfluorocarbon
superheated fluorocarbon phase-change agents. Adv Colloid Inter- droplets for focused ultrasound therapy. Phys Med Biol
face Sci 2016;237:15–27. 2010;55:4933.
Mullin L, Gessner R, Kwan J, Kaya M, Borden MA, Dayton PA. Effect Selfridge AR. Approximate material properties in isotropic materials.
of anesthesia carrier gas on in vivo circulation times of ultrasound IEEE Trans Sonic Ultrason 1985;32:381–394.
microbubble contrast agents in rats. Contrast Media Mol Imaging Sheeran PS, Wong VP, Luois S, McFarland RJ, Ross WD, Feingold S,
2011;6:126–131. Matsunaga TO, Dayton PA. Decafluorobutane as a phase-change
Olympus N. Ultrasonic transducers technical notes [technical bro- contrast agent for low-energy extravascular ultrasonic imaging.
chure]. Waltham, MA: Olympus NDT; 2006. Ultrasound Med Biol 2011;37:1518–1530.
Parker N, Mather M, Morgan S, Povey M. Longitudinal acoustic prop- Sheeran PS, Luois SH, Mullin LB, Matsunaga TO, Dayton PA. Design
erties of poly (lactic acid) and poly (lactic-co-glycolic acid). of ultrasonically-activatable nanoparticles using low boiling point
Biomed Mater 2010;5 055004. perfluorocarbons. Biomaterials 2012;33:3262–3269.
Povey MJ. Ultrasonic techniques for fluids characterization. San Diego, Sheeran PS, Matsunaga TO, Dayton PA. Phase-transition thresholds
CA: Academic Press; 1997. and vaporization phenomena for ultrasound phase-change nanoe-
Qamar A, Wong ZZ, Fowlkes JB, Bull JL. Dynamics of acoustic drop- mulsions assessed via high-speed optical microscopy. Phys Med
let vaporization in gas embolotherapy. Appl Phys Lett 2010;96 Biol 2013;58:4513.
143702. Shpak O, Kokhuis TJ, Luan Y, Lohse D, de Jong N, Fowlkes B, Fabiilli
Rabell-Montiel A, Thomson AJ, Anderson TA, Pye SD, Moran CM. M, Versluis M. Ultrafast dynamics of the acoustic vaporization of
Acoustic properties of small animal soft tissue in the frequency phase-change microdroplets. J Acoust Soc Am 2013;134:1610–
range 1232 MHz. Ultrasound Med Biol 2018;44:702–713. 1621.
Radhakrishnan K, Bader KB, Haworth KJ, Kopechek JA, Raymond JL, Shpak O, Verweij M, Vos HJ, de Jong N, Lohse D, Versluis M. Acous-
Huang S-L, McPherson DD, Holland CK. Relationship between tic droplet vaporization is initiated by superharmonic focusing. Pro
cavitation and loss of echogenicity from ultrasound contrast agents. Natl Acad Sci 2014;111:1697–1702.
Phys Med Biol 2013;58:6541. Szabo TL, Clougherty F, Grossman C. Effects on nonlinearity on the
Radhakrishnan K, Holland CK, Haworth KJ. Scavenging dissolved estimation of in situ values of acoustic output parameters. J Ultra-
oxygen via acoustic droplet vaporization. Ultrason Sonochem sound Med 1999;18:33–41.
2016;31:394–403. Williams R, Wright C, Cherin E, Reznik N, Lee M, Gorelikov I, Foster
Rapoport N. Phaseshift, stimuliresponsive perfluorocarbon nano- FS, Matsuura N, Burns PN. Characterization of submicron phase-
droplets for drug delivery to cancer. Wiley Interdiscip Rev change perfluorocarbon droplets for extravascular ultrasound imag-
Nanomed Nanobiotechnol 2012;4:492–510. ing of cancer. Ultrasound Med Biol 2013;39:475–489.
Rapoport NY, Efros AL, Christensen DA, Kennedy AM, Nam K-H. Wilson PS, Roy RA. An audible demonstration of the speed of sound in
Microbubble generation in phase-shift nanoemulsions used as anti- bubbly liquids. Am J Phys 2008;76:975–981.
cancer drug carriers. Bubble Sci Eng Technol 2009;1:31–39. Xin Y, Zhang A, Xu LX, Fowlkes JB. Numerical study of bubble area
Reznik N, Williams R, Burns PN. Investigation of vaporized submi- evolution during acoustic droplet vaporization-enhanced HIFU
cron perfluorocarbon droplets as an ultrasound contrast agent. treatment. J Biomech Eng 2017;139 091004.
Ultrasound Med Biol 2011;37:1271–1279. Zell K, Sperl J, Vogel M, Niessner R, Haisch C. Acoustical properties
Reznik N, Shpak O, Gelderblom EC, Williams R, de Jong N, Versluis of selected tissue phantom materials for ultrasound imaging. Phys
M, Burns PN. The efficiency and stability of bubble formation by Med Biol 2007;52:N475.

You might also like