SARRP Related Messages on MP1

Disclaimer: These are only recommendations.


If you have any questions please contact Xstrahl Service.

Description Action
If first occurance, restart X-rays and proceed. If it
E-12 Over mA
continues, contact Xstrahl Service

Detector Panel Shield Verify that the detector panel shield is in the correct
E-16 Position incorrect / High position for the focal spot selected. If it is corrected
cooler water temp and error does not clear, contact Xstrahl Service

MP1 Powered off not in MP1 was left on when SARRP was powered off. Self-
E-64
mode 0, CPU Reset clearing error when X-rays are energized.

Keyswitch not in HT during Turn Key to the "HT" position and retry the exposure.
E-66
exposure If it does not clear, contact Xstrahl Service

Multiple E-12 errors, the system has locked out for


E-78 kV Breakdown Lockout
safety. Contact Xstrahl Service

Exposure time is set to "0". Adjust exposure time to


E-86 Exposure parameters error
greater than zero as required.