Jan Welinder, Christer Térnevik
Ulf Bostrém, Bengt Lindberg
Lars Atlegard, Bo Héjdefors
The Introduction of DECT
in Health Care Environments
sP
Swedish National Testing and Research Institute
Physics & Electrotechnics
SP REPORT 1997:182A
Abstract
The Introduction of DECT in Health Care Environment
This report gives some recommendations for the introduction and use of DECT (a cordless tele-
phone system) in health care environments. It is based on-several published practical investiga~
tions where medical equipment has-been tested together with DECT phones. A number of existing
recommendations for the use of transmitters have also been analysed,
‘The main conclusion is that DECT with very limited restrictions can be used in most areas,
Since the results may be unexpected due to the rather high field strength close to a DECT phone
some results regarding immunity to high frequency electromagnetic fields are also discussed.
Key words: DECT, EMC, medical equipment, immunity testing
‘Sveriges Provnings- och Swedish National Testing and
Forskningsinstitut Research Institute
SP Rapport 1997:18 SP Report 1997:18
ISBN 91-7848-676-9
ISSN 0284-5172
Boris 1997
Postal address:
Box 857, SE-501 15 BORAS
Sweden
Telephone + 46 33 16 50 00
Telefax + 46 33 1355.022B
CONTENTS
Abstract
Contents
1. Introduction
1.1 The DECT system
1.2 Similar telephony systems
2 Available investigations
2.1 Swedish report: Rapport om mobittelefont i sjukhusmiljs
2.2 Recommendations regarding the use of pocket telephones within
health cate institutions
2.3 A Study to Examine the interaction between Cordless PABX handsets
and Hospital Electronic Equipment
2.4 Electromagnetic compatibility of Medical Devices with Mobile
‘Communications
2.5 Other reports
2.6 Summary of the investigations
3 Published recommendations
3.1 Socialstyrelsen
3.2 VIFKA,
3.3 Telia Mobitel
3.4NHS
3.5 ANSI draft on-site test method
3.6 MDA,
4, Recommendations for introduction and use of DECT telephones
Appendix A. EM-field limits
A.1 Fields form handheld transmitters
A.2 Field strength immunity limits
Appendix B, Fields in buildnings
B.1 Reflection and scattering
B.2 Attenuation in various types of floors and walls
Appendix C. In Situ Testing
C1 Equipment
C.2 Method
Appendix D. Recommendations for the use of DECT systems in
Swedish health care
References
2A
2B
Bee
as
10
u
12
12
12
12
13
13
13
14
16
16
17
21
21
22
23
23
23
25
281. Introduction
‘The objective of this report is to establish.a set of recommendations that can be used
when DECT systems are introduced in the health care environment. It is a well known
fact that mobile transmitters can cause interference in electronic devices of various kinds
if this is not foreseen in the construction, However, the disturbance potential is not the
same for all (ypes of systems. DECT is a high frequency, low power system and can be
expected to generate less interference than many other systems like VHF-transceivers and
GSM mobile phones.
‘The results in this reports cannot directly be used for other environments without a
thorough analysis of technical and other differences. It is also strictly limited to possible
disturbances generated by DECT phones and does not deal with other transmitters or the
numerous other sources of electromagnetic disturbances (electrostatic discharges etc.)
that may interfere with the proper function of medical equipment. The type of equipment
covered is only medical equipment,
‘The recommendations are based mainly on reports already published, These are briefly
summarised but it is recommended for the reader to study the referenced documents more
carefully. The technical electromagnetic field limits are based on the reports and also on
experience from the EMC testing laboratory at SP which has handled a large number of
EMC-tests during recent years.
‘The group which has taken part in the work is:
Jan Welinder, SP
Christer TSrnevik, Ericsson Radio Systems
Ulf Bostriim, MTA, Clinical Engineering Dept,, University Hospital, Linképing
Bengt Lindberg, MTA, Sahlgrenska Universitetssjukhuset/Ostra
Lars Atlegard, ‘Telia Publicom,
Bo Hojdefors, Swedish National Board of Health and Welfare
‘The report was financed by Ericsson Business Systems AB.
1.1 The DECT system
DECT (Digital Enhanced Cordless Telecommunications) is « digital radio standard for
cordless telephony in home and working environments, DECT has been a standard in the
EU countries since 1993, and it has also been adopted in several countries outside of
Europe. Today DECT is used in more than 30 countries around the world.
‘The DECT standard is mainly used for multipte-user wireless communications in offices,
factories and other working environments. Such a business cordless system, which is also
called cordless PBX (Private Branch Exchange), is a system consisting of a number of
radio base stations which are connected to a central exchange. Users can make and receive
calls when in range of a base station (up to about 50 meters indoors). Like in a cellular
‘mobile telephony system (e.g. GSM), the radio connection ts handed over from one base
station to another when a user is moving within the area covered by the system. ‘The main
feature of DECT cordless systems is that very high densities of users are permitted,
However, unlike celular mobile telephone systems, DECT telephones generally cannot be.
used outside of the working environment.DECT is based on the TDMA (Time Division Multiple Access) technology, which means
that each of the radio channels (RF carriers) are divided into 24 time slots (12 duplex time
slots), One user is allocated two time slots, one for transmission and one for reception.
Consequently, each radio channel can be used for 12 simultaneous calls. The total time of
‘24 slots is 10 milliseconds, which means that a DECT telephone transmits with a time slot
repetition rate of 100 Hz.
DECT uses 10 radio channels in a 20 MHz frequency band between 1880 and 1900 MHz.
‘The telephones and the base stations have a peak ontput power of 250 mW, Since the
handsets only use 1 time slot out of 24 for transmission, the average oulput power is 10
mW.
In the table below, key system data for DEC are listed.
DECT system data
Peak power/mean power:
handset: 250 mW/ 10 mw
base station: 250 mW/ max 125 mW (depending on number of
channels used)
Frequency band 1.88 - 1.90 GHz
Multiple access method ‘TDMA (Time Division Multiple Access)
Duplexing method ‘TDD (Time Division Duplex)
Time slots per channel 24 (12 duplex slots)
Time slot duration 0.417 ms
‘Time slot repetition rate 100 Hz,
1.2 Similar telephony systems
A number of similar wireless telephony systems exist. The results in this report are not
valid for those systems, However, some conclusions may be drawn by the reader based
on the various systems technical specifications. For this reason the basic characteristics
of some commonly used systems are described below. The parameters in the technical
specifications have an influence which can be described as follows:
Power. Increasing power increases the probability of interference and interference
occurring at larger distance,
Frequency, A bigher frequency will normally decrease the risk of interference in an
adjacent equipment. This explains why electronic devices does not react to the presence.
of the modem telephony systems in a way like laboratory testing result seems to predict,
This is due to the fact that less power is received due the decreasing efficient antenna area
athigher frequencies (see appendix A), Experience shows that for a constant field
strength most devices including external cables and susceptible to electromagnetic fields
will show most of its reactions for frequencies below 400 MHz, As mentioned before
DECT work at 1.9 GHz. In rare cases the immunity will decrease at high frequencies due
to the better penetration of higher frequencies through slots and holes or due to the small
size of the device.
Modulation, Many devices like audio and video equipment are more susceptible for a
modulated signal (in this context AM-modulation or a time multiplexed carrier wave)
than for only the high frequency carrier wave, For other systems the degree of
interference may be depending on the peak or mean power instead, but as a general rulemore interaference can be expected with a TDMA system than with an analogue FM-
modulated systems when all other parameters are similar,
‘A general trend is that interference is less likely with modern systems that use higher
frequency and less power than older. E.g.a “walkie-talkie” has a considerably higher
interference potential than a GSM telephone.
System: ct
‘Type/use: Analogue cordless PBX
Peak power/mean power
handset: 10mW/10mw
base station: 10 mW/10 mW
Frequency: 900 MHz
Multiple access method: FDMA (Frequency Division Multiple Access)
Duptexing method: FDD (Frequency Division Duplex)
‘Time slot repetition rate: :
System: cr
‘Typefuse: Digital cordless PBX
Peak power/mean power
handset: 10 mW/5 mW
base station: 20 mW /10mw
Frequency: 900 MHz
Multiple access method: FDMA
Duplexing method: ‘TDD (Time Division Duplex)
‘Time slot repetition rate: 500 He,
System: NMT450
‘Typefuse: Analogue mobile telephony
Peak power/mean power handset: 1,5 W/ 1,5W (sce note)
Frequency: 450 MHz
Multiple access method: FDMA
Duplexing method: FDD
‘Time slot repetition rate: -
System: NMT 900
Type/use: Analogue mobile telephony
Peak power/mean power handset: 1 W/ TW (see note)
Frequency: 900 ME
Multiple access method: FDMA
Duplexing method: FDD
‘Time slot repetition rate: rp
‘System: GSM
‘Typeluse: Digital mobile telephony
Peak power/mean power handset: 2. W/ 250 mW (see note)
Frequency: 900 MHz.
Multiple access method: TDMA
Duplexing method: FDD
Time slot repetition rate: 217 He,‘System:
‘Typesuse:
Peak power/mean power handset:
Frequency:
Multiple access method:
Duplexing method:
‘Time slot repetition rate:
DCS 1800 (GSM 1800)
Digital mobile telephony
1 W/ 125 mW
1800 MHz,
TDMA
FDD
27
Note: Other types of telephones with higher output power exist for these systems. The
figures relates to the handheld phones,2 Available investigations
Since it is widely recognised that radio transmitters in general may cause hazardous
electromagnetic interference in the health care environment several investigations haye
been carried out, This report is based on the reports described below, all including tests
with DECT.
Common properties for all the reports ate that it has for obvious reasons not been
possible to test all equipment in their normal environment. In many cases it has been
necessary to use some kind of simulator for the patient. Like in all similar testing it has
not been possible to test all modes of each equipment but in most cases the equipment is,
tested in a mode that as closely as possible simulates the normal functions. Some
functions of the equipment like alarms and the like may not have been active,
‘The tested equipment are of various kinds. The majority is medical devices but some are
‘used for e.g. chemical analysis. All are used in hospitals,
2.1 Swedish report: Rapport om mobiltelefoni i sjukhusmiljé
(Report on mobile telephones in hospital environment).
‘This project was financed by ‘Telia Mobitel, Sweden. Active participants were also the
‘Swedish National Testing and Research Institute and persons representing clinical
engineering departments from two hospitals (3).
Objective
‘The objective of the project was a broad investigation of possible interference with
transmitters in a hospital environment, Another objective was to investigate the
occurrence of ineidents in hospitals due to the use of transmitters. None of the reported
incidents include the use of DECT or any other system with similar or lower power.
‘Test method and result
The project included a part in which medical equipment was tested for immunity to
transmitters, A number of mobile transmitters, in total eleven different types, in the
frequency range 27 - 1900 MHz were selected. The output power varied from 10 mW to
20 W. The number of medical devices that were tested was 24. However, only a limited
number was tested together with DECT. The equipment was selected more'or less
randomly among the normally used equipment.
Real transmitters were tested in normal use in the vicinity of the medical equipment. ‘The
tests were performed in hospital environment but in a safe area using various means to
simulate real operation of the device, The equipment was in most cases placed on a
wooden table and the transmitter was switched on and moved from a distance of 5 m to
10 cm from the EUT (equipment under test). At fixed transmitter distances the EUT was
observed for any type of interference, This was repeated four times and the EUT was
rotated 90 degrees in between. Small hand held transmitters were also swept around the
EUT at 10 cm distance and placed on top of the EUT where that was possible,‘The test results were as below,
DECT: 6 devices were tested and none showed any susceptibility.
Other systems: 12 of 24 were susceptible to GSM or NMT and 20 of 24 for transmitters
in the power range 5-20 W.
2.2 Recommendations regarding the use of pocket telephones within health care
institutions
VIFKA Telecommunicatic is the Dutch trade body for the telecommunications industry,
‘The tests included were performed by TNO Prevention & Health, Leiden. More complete
reports arc published by TNO in dutch [1], (21.
Objective
‘The objective was to find recommendations for use of pocket phones in health care
enyironment. This was one of the reports where this clearly was stated as the objective.
‘The reason for starting the project was a ban of the use of pocket phones advised by the
‘Dutch National Health Sector Inspectorate 1994,
‘Test method and results
‘Three types of phones were used, GSM, CT2 and DECT. The tested equipment were in
normal use when possible, otherwise some kind of patient simulator was used.
‘The phone was held initially at a distance of 1 m and then moved towards the equipment
until zero distance. This was done from several directions. Tests were also performed
close to cable connections and the like,
‘The test results were as below.
DECT: 131 devices were tested and 19 reacted in some way.
‘Three medical devices reacted when the distance to the phone was more than 10 cm:
— HPLC analyser reacted at 50 om
— Hearing aid reacted at 50 cm (see 2.5)
— Acoustic equipment for locating veins reacted at 3m
‘One device reacted at close distance in a way that could be indirectly dangerous
(judgement in the report):
= Analyser/osmometer giving incorrect reading.
Other systems: 101 out of 205 were disturbed by GSM telephones and 2 out of 106 by
cT2.2.3 A Study to Examine the interaction between Cordless PABX handsets and
Hospital Electronic Equipment
‘The investigation was performed by ECS Ld for the National Health Service, UK [41
Objective
The objective was to identify possible operational electromagnetic incompatibility
between cordless PBX equipment and medical equipment commonly found in NHS
hospitals.
A second objective was to see if any medical equipment could affect the operation of the
handsets. This was not seen.
‘Test method and result
Five transmitters were used in the tests, two different DECT handsets, CT2, TACS and a
3 W VHF portable radio, ‘The number of tested medical equipment was 26,
‘The hand portables were characterised on an open area test site without ground plane as a
separate part of the test, A difference in output power from two DECT manufacturers
were noted (150 and 237 mW respectively), Both were within the allowed tolerance,
‘The immunity tests were carried out by setting up various pieces of hospital equipment
and testing their susceptibility by moving the handsets progressively closer to the
equipment until they touched the EUT or they caused a discernible effect. Particular
attention was paid to the front panel area of each EUT and any other area where the BUT
enclosure might not form good RF screen,
‘The rest results were as below.
DECT: 26 equipment were tested and 2 were found to be susceptible
— Electrocardiograph, trace affected
— Electrocardiograph, vibrating noise but no change of trace
‘One of the two equipment reacted when the distance to the phone was more than 10 cm:
~ Hlectrocardiograph at 0.30 m and 0.37 m for the two DECT phones. This reaction was
not deemed to be directly dangerous.
Other systems: 1 of 26 were susceptible for CT2, 2 of 26 for TACS and 20 of 26 for
VHF.
2.4 Electromagnetic compatibility of Medical Devices with Mobile
Communication
‘The investigation was performed by the Medical Devices Agency, UK [10].10
Objective
The objective was to try (o answer the two questions:
- Does mobile communication equipment present a real risk when operated in the vicinity
of medical devices ?
= If $0 what actions should be taken to minimise the risks 7
‘Test method and result
‘Several types of transmittors were used including DECT, other cellular and cordless.
systems and also several types of handheld transmitters, These transmitters are described
by model name in the report but figures regarding output power and frequency are not
reported. The number of tested devices was large, in total 178, but only 31 of these were
tested together with DECT handsets,
For some of handsets including DECT the field strength was measured on an open area
test site and described in antenna diagrams. Somewhat higher values are measured in this
report compared to other measurements,
‘The immunity tests were carried out by setting up the medical devices operating "within
‘normal operating parameters”. Sometimes the equipment was connected to a patient
during the tests, The handsets were move at a steady rate of 1 cm/s over all accessible
surfaces, Any observable effects were noted. The test was repeated at 0.5 and 1 m
distance and if any effect still was observed the largest distance where the effect
continued to occur was noted.
‘Some statistical results are shown for DECT but they are based on only one or two
observed effects.
‘The test results were as below:
DECT: 31 devices were tested and 1 was found to be susceptible:
= CO,-0,-meter showed minor interference to 0,-trace at 0.5 m.
Other systems: A large number of effects are reported for systems with higher power and
lower frequency.
2.5 Other reports
Several other investigations exist but some are not published. Other reports are published
‘but do not include DECT as one of several tested systems. Several investigations of these
‘two types have been reviewed. Their results are often difficult to directly apply to the use
of DECT but the main impression is that the summarised results below are comparable
also to these reports,
A number of reports deal with hearing aids and pacemakers, This type of equipment is
not included because their general use are not specific for the health care environment.
However, a single type of hearing aid was included in the VIFKA [1] report. Hearing
aids are often susceptible to high frequency transmitters with amplitude modulation in
the acoustic frequency range (ike 100 Hz. for DECT).u
2.6 Summary of the investigations
‘The above investigations can be summarised as below. A relatively large number of
medical equipment (11 %) are susceptible when the handset is in direct contact with
medical equipment. However, nearly all of the shown reaction are of a type that has not
‘been deemed to be dangerous. Further, nearly all reactions disappear when the handset is
‘moved to a distance of 0.5 m, It should be noted that 67 % of the devices were included
in the VIFKA [1] report,
No of devices
Devices tested (with DECT) 194
‘Susceptible devices 22
‘The judgemens in the reports of
degree of impact on patient,
Have no direct impact on patient 21
Could have impact on patient 1
Largest distance for susceptibility
Susceptible at 0 - 0.1m 7
Susceptible at 0.3 - 0.5 m
Susceptible at 3 m 1
Distance for effect with possible
impact on patient
Possible impact < 0.1m) i
Possible impact > 0.1 m 012
3 Published recommendations
‘There arc recommendations ayailable from several sources. Some of them are briefly
described below,
3.1 Socialstyrelsen
(Swedish National Board of Health and Welfare, Competent Authority for
medical devices)
Socialstyrelsen has an existing recommendation [9] regarding use of
cellular telephones, cordless telephones etc. It is at the moment being
revised and will in the new version [10] especially address the use of the
DECT-system as well as risk analysis and immunity requirements on
medical devices. The revised recommendation corresponds in essential
parts to the statements made in this report. Socialstyrelsen has had a
representative in the group.
3.2 VIFKA
‘The VIFKA research [1] was primarily carried out to write recommendations, The
recommendations are mainly aimed at visitors and patients using pocket phones. The use
is not banned but it should be based on knowledge about the involved risks. The
recommendations require that information should be given to personnel, patients and
visitors and describe involved hazard and measutes to prevent them. A risk analysis
should be carried out at the institute to find areas where the risk may be higher.
For pocket phones there are some precise rules like that the phones should be kept at a
distance of at least 1.5 m from medical equipment, If this is not possible the phone has to
be switched off. In certain areas the use should be banned.
‘New medical equipment shall comply with the European EMC-regulations,
A separate part describes the use of CT2 and DECT together with a local PBX, It is
recommended to check whether interference may occur. No details are given but itis,
instead advised that recommendations should be written together with the exchange
supplier. However, it is pointed out that the phone should be supplied with a warning text
telling that the phones cannot cause interference if used according to the rules of the
institute. This is to inform patients and visitors.
3,3 Telia Mobitel
Ina final discussion, written by the person representing Telia Mobitel, some conclusions
are discussed but no detailed recommendations are given. It is only stated that measures
like using signs etc, should be taken to minimise risks and that recommendations should
be based on the local electrical environment [3].13
3.4 NHS
Inthe conclusion it is stated that the effects are few [4] .
‘The following brief recommendations that are applicable to DECT are given:
= Ask for EMC-specifications when purchasing health care equipment.
— Establish safe working distances with simple EMC-tests before taking into use,
3.5 ANSI draft on-site test method.
‘The document [5] gives detailed recommendations on how in situ tests may be performed
on existing equipment in health care facilities. General recommendations regarding EMC
are also given. In an appendix some guidelines for the use of hand-held transmitters are
given. Itis noted that the use of portable transmitters may need to be restricted. Specific
daa like a recommended separation distance between the transmitter and the equipment
are based on known immunity of the equipment,
3.6 MDA.
‘The reports have detailed recommendations for the use of several types of systems
including e.q. GSM but only the following short recommendation regarding DECT:
These systems are very unlikely to cause EMI under most circumstances, and need not
to be restricted.”4
4, Recommendations for the introduction and use of DECT
Introduction
First it must be stated that itis the opinion of the project group that DECT may be used
in health care environments. However, its use must not be totally without control. The
following discussion show a way to control the introduction and use of DECT.
It should be observed that the possibility of a dangerous malfunction of a device is
limited even if these rules are not observed.
In appendix D the recommendations are written in a form directly useful for hospitals,
General use of DECT handsets
"The project group has agreed on the following rules for use of DECT handsets:
— The DECT handsets can as a general rule be used in all areas of the health care
environment.
~ A safety distance between the handset and any medical device of 0.5 m has to be
observed.
— Aspecific waming for placing the handset in direct contact with any electronic device
should be given.
= There may exist areas were specific precautions has to be observed.
‘Comments
‘The test data from nearly 200 pieces of medical equipment tested together with DECT
handsets show few reactions if the handset is kept not to close to the tested device. A
single reaction occured when the handset was in close connection with the device that
was deemed to be indirectly dangerous for the patient, The result may be unexpected if
the generated field strength is compared to normal test limits since the field strength
close to a DECT handset is quite high. In appendix A the frequency behaviour for
susceptible equipment is discussed and itis explained why as a general rule quite high
immunity can be expected for the DECT frequencies around 1.9 GHz. This has also been
shown by the practical tests.
‘The recommendations above are based on the referenced practical tests and on the
discussion about increased immunity at higher frequencies, The practical use of DECT.
‘phones has also been taken into consideration. A safety distance of more than 0.5.
implies that the user cannot wear the handset while operating medical equipment,
However, if the distance of 0.5 m shall be observed there must be rules on how the
handset is worn, It cannot be placed in any pocket. It is therefore important that all
personnel using DECT have a knowledge about not only the rules but also their
background,
Risk analysis
Since rules seldom are observed to 100 % and something like 10 % of all medical
equipment is susceptible for a direct contact with an operating handset there may be
situations where the involved risk are too high. Equipment may have known or observed
very Jow immunity, or impact of an equipment failure is not acceptable etc, In these cases
the group recommend a risk analysis and exceptions from the general rule.