New QC Lot Setup for DxFlex

I) Kindly download the new target file using the following steps:
1. Go on the link:
https://www.beckmancoulter.com/wsrportal/wsrportal/page/softwareDownloadSearch

2. Select the following options in the column


-

3. Click on “Search”
4. Search the lot no of your interest. Eg:

5. Select on “Download” as shown below.

6. Click “Save”
7. Please save it in the DxFlex system on the desktop (Note: The file wont open if double
clicked)
II) Installation of New QC Lot in Target library:

1. Open “Cytexpert for DxFlex” software


2. Click on “QC and standardization”
3. Click on “Start QC”
4. Click on “Settings”
5. Click on “Target Library”
6. At the bottom of the window opened click on “Import”

7. Open the location where target file is saved. Select the file and click “Open”.
8. Click “Initialize” to run QC. But while selecting the lot no select the new lot. Run your
QC. In the first run it may ask you to save. Save and rerun the QC.
9. If you get stuck or have doubts contact your respective application specialist.