INSTRUCTION MANUAL Dear Customer, Functional description Errors and warnings Cooling lamp: The instrument waits for

ription Errors and warnings Cooling lamp: The instrument waits for the
Thank you for choosing a Hanna product. This manual will provide you lamp to cool down.
with the necessary information for the correct use of the instrument. Please ON ZERO READING
HI 96771 read it carefully before using the meter. If you need additional technical
Battery low: The battery must be replaced
information, do not hesitate to e-mail us at tech@hannainst.com. Light High: There is too much light to perform
soon.
Free Chlorine & a measurement. Please check the preparation of
the zero cuvette.
Preliminary examination
Chlorine Ultra High Range Please examine this product carefully. Make sure that the instrument is
Dead battery: This indicates that the battery
Light Low: There is not enough light to perform is dead and must be replaced. Once this
not damaged. If any damage occured during shipment, please notify
ISM your Dealer.
a measurement. Please check the preparation of indication is displayed, normal operation of the
instrument will be interrupted. Change the
the zero cuvette.
Each HI 96771 Ion Selective Meter is supplied complete with: battery and restart the meter.
• Two Sample Cuvettes and Caps Measurement procedure
No Light: The instrument cannot adjust the
• 9V Battery 1• Turn the meter on by pressing ON/OFF.
light level. Please check that the sample does Measurement
• Instruction Manual 1. RANGE/GLP/ key: press to change the parameter. Press and hold 2• When the beeper sounds briefly, the LCD
not contain any debris.
Note: save all packing material until you are sure that the instrument for three seconds to enter GLP mode. In calibration mode press to edit 2 displays dashes and “P1” (Chlorine Low
works correctly. Any defective item must be returned in its original the date and time. Range) or “P2” (Chlorine Ultra High Range)
ON SAMPLE READING
packing. 2. CAL CHECK key: press to perform the validation of the meter, or the meter is ready. The code that appears
press and hold for three seconds to enter calibration mode. Inverted cuvettes: The sample and the zero on the secondary display is the one of the
For more details about spare parts and accessories 3. ZERO/CFM key: press to zero the meter prior to measurement, to cuvette are inverted. last selected parameter. If necessary, press
or RANGE/GLP/ to change parameter.
see “Accessories”. confirm edited values or to confirm factory calibration restore.
The blinking “ZERO” indicates that the
4. READ/ /TIMER key: In measurement mode, press to make a
instrument needs to be zeroed first.
measurement, or press and hold for three seconds to start a Zero: A zero reading was not taken. Follow the
pre-programmed countdown prior to measurement. In GLP mode instructions of the measurement procedure for 3• Fill the cuvette with 10 mL of unreacted
Technical specifications: press to view the next screen. zeroing the meter. sample, up to the mark, and replace the cap.
Range 0.00 to 5.00 mg/L for Chlorine LR 4• Place the cuvette into the cuvette holder
5. ON/OFF key: to turn the meter on and off.
and ensure that the notch on the cap is
0 to 500 mg/L for Chlorine UHR 6. Liquid Crystal Display (LCD) Under range: A blinking “0.00” indicates that 3 positioned securely into the groove.
Resolution 0.01 mg/L from 0.00-3.50 mg/L; 0.10 mg/L above 7. Cuvette alignment indicator the sample absorbs less light than the zero reference. 5• Press ZERO/CFM and the lamp, cuvette
3.50 mg/L for Chlorine LR 8. Cuvette holder Check the procedure and make sure you use the and detector icons will appear on the
same cuvette for reference (zero) and measurement. 4 display, depending on the measurement
1 mg/L from 0 to 200 mg/L; 10 mg/L above
200 mg/L for Chlorine UHR phase.
6• After a few seconds the display will show
Accuracy ±0.03 mg/L ±3% of reading at 25°C for Chlorine LR DISPLAY ELEMENTS DESCRIPTION “-0.0-”. The meter is now zeroed and
Over Range: A flashing value of the maximum
±3 mg/L ±3% of reading at 25°C for Chlorine UHR 5-6 ready for measurement.
concentration indicates an over range condition.
Typical EMC Dev. ±0.01 mg/L for Chlorine LR The concentration of the sample is beyond the 7• Remove the cuvette.
±1 mg/L for Chlorine UHR programmed range: dilute the sample and re- 8• Add the content of the packets for the
specific test reagents:
run the test.
Light Source Tungsten lamp Chlorine LR: 1 packet of HI 93701-0
Light Detector Silicon Photocell with narrow band interference DURING CALIBRATION PROCEDURE Chlorine UHR: 1 packet of HI 95771A-0
and 1 packet of HI 95771B-0.
filter @ 525 nm
Standard Low: The standard reading is less 9• Replace the cap and shake gently for
Method Adaptation of Standard Methods for the Examination than expected. 20 seconds.
of Water and Wastewater, 20th edition, 4500-Cl. 10•Replace the cuvette into the cuvette holder
Environment 0 to 50°C (32 to 122°F); 1. The measuring scheme (lamp, cuvette, detector), appears during and ensure that the notch on the cap is
different phases of zero or reading measurement 8 positioned securely into the groove.
max 95% RH non-condensing Standard High: The standard reading is higher
2. Error messages and warnings than expected. 11•Press and hold READ/ /TIMER for three
Battery Type 1 x 9 volt 3. The battery icon indicates the charge state of the battery seconds. The display will show the
Auto-Shut off After 10' of non-use in measurement mode; after 4. The hourglass appears when an internal check is in progress countdown prior to measurement. The
1 hour of non-use in calibration mode; with last 9 beeper is playing a beep at the end of
5. Status messages OTHER ERRORS AND WARNINGS
countdown period.
reading reminder. 6. The chronometer appears when the reaction timer is running
Cap error: Appears when external light Alternatively, wait for 1 minute and just
Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7”) 7. The month, day and date icons appear when a date is displayed enters in the analysis cell. Assure that the 10 press READ/ /TIMER.
8. Four digit main display cuvette cap is present. In both cases the lamp, cuvette and
Weight 360 g (12.7 oz.).
9. Measuring units detector icons will appear on the display,
10. Four digit secondary display depending on the measurement phase.
3”
12•The instrument directly displays concentration 9• At the end of the measurement the display or dirt and that they are inserted correctly. FACTORY CALIBRATION RESTORE Factory Accessories
11 9 13-14
in mg/L of free chlorine and the range will show the validation standard value. 12•Then the date of last calibration (e.g.: Calibration
It is possible to delete the calibration and Restore REAGENT SETS
number on the LCD. The reading should be within specifications “01.08.2009”) appears on the display, or
as reported on the CAL CHECK™ or restore factory calibration. HI 93701-01 Reagents for 100 Free Chlorine tests
INTERFERENCES FOR CHLORINE LR “01.01.2009” if the factory calibration
• Positive error: Bromine, Oxidized Manganese and Standard Certificate. If the value is found was selected before. In both cases the year 1• Press and hold for three seconds 2 HI 93701-03 Reagents for 300 Free Chlorine tests
Chromium, Chlorine dioxide, Ozone and Iodine. out of specifications, please check that the number is blinking, ready for date input. RANGE/GLP/ to enter GLP mode . HI 95771-01 Reagents for 100 Chlorine Ultra High Range tests
• Alkalinity above 250 mg/L CaCO3 or acidity or cuvettes are free of fingerprints, oil or dirt 13•Press RANGE/GLP/ to edit the desired HI 95771-03 Reagents for 300 Chlorine Ultra High Range tests
and repeat validation. If results are still 2• Press READ/ /TIMER to enter in the
above 150 mg/L CaCO3 will not reliably year (2009-2099). If the key is kept pressed, factory calibration restore screen. The OTHER ACCESSORIES
develop the full amount of color or it may 12 found out of specifications then recalibrate
the year number is automatically increased. instrument asks for confirmation of user HI 96771-11 CAL CHECK™ Standard Cuvettes (1 set)
rapidly fade. To resolve this, neutralize the the instrument.
14•When the correct year has been set, press 15-16 calibration delete. HI 721310 9V battery (10 pcs)
sample with diluted HCl or NaOH. CALIBRATION Calibration ZERO/CFM or READ/ /TIMER to confirm. 3• Press ZERO/CFM to restore the factory 3-4 HI 731318 Cloth for wiping cuvettes (4 pcs)
• In case of water with hardness greater than
Note: It is possible to interrupt the calibration Now the display will show the month blinking. HI 731331 Glass cuvettes (4 pcs)
500 mg/L CaCO3, shake the sample for or calibration or press RANGE/GLP/ again
procedure at any time by pressing 4 HI 731335 Caps for cuvettes (4 pcs)
approximately 2 minutes after adding the or 15•Press RANGE/GLP/ to edit the desired to abort factory calibration restore.
powder reagent. CAL CHECK or ON/OFF keys. HI 93703-50 Cuvette cleaning solution (230 mL).
month (01-12). If the key is kept pressed, 4• The instrument briefly indicates “donE”
INTERFERENCES FOR CHLORINE UHR When calibrating, only the selected range the month number is automatically increased. upon restoration of factory calibration prior
• Positive error: Bromine, Oxidized Manganese is affected.
16•When the correct month has been set, press to returning to measurement mode.
Warranty
and Chromium, Chlorine dioxide, Ozone and 1• Turn the meter on by pressing ON/OFF. ZERO/CFM or READ/ /TIMER to confirm.
Iodine. HI 96771 is warranted for two years against defects in workmanship and
2• When the beeper sounds briefly and the Now the display will show the day blinking. materials when used for its intended purpose and maintained according
Validation and Calibration procedures LCD displays dashes, the meter is ready. 5 17 to the instructions.
3• To change the range, simply press
17•Press RANGE/GLP/ to edit the desired Battery management This warranty is limited to repair or replacement free of charge.
Warning: do not validate or calibrate the instrument with standard day (01-31). If the key is kept pressed,
solutions other than the Hanna CAL CHECK™ Standards, otherwise RANGE/GLP/ . 18 To save battery, the instrument shuts down after 10 minutes of non-use Damages due to accident, misuse, tampering or lack of prescribed
the day number is automatically increased. maintenance are not covered.
erroneous results will be obtained. 4• Press and hold CAL CHECK for three in measurement mode and after 1 hour of non-use in calibration mode.
For accurate validation and calibration results, please perform tests at seconds to enter calibration mode. The 6-7 Note: It is possible to change the editing from If a valid measurement was displayed before If service is required, contact your dealer. If under warranty, report the
day to year and to month by pressing 19 model number, date of purchase, serial number and the nature of the
room temperature (18 to 25°C; 64.5 to 77.0°F). display will show “CAL” during calibration auto shut off, the value is displayed when the
procedure. The blinking “ZERO” asks for READ/ /TIMER. failure. If the repair is not covered by the warranty, you will be notified
Use the Hanna CAL CHECK™ cuvettes (see instrument is switched on. The blinking “ZERO” of the charges incurred.
instrument zeroing. means that a new zero has to be performed.
“Accessories”) to validate or calibrate instruments. 18• Press ZERO/CFM to save the calibration If the instrument is to be returned to Hanna Instruments, first obtain a
5• Place the CAL CHECK™ Standard date. One fresh battery lasts for around 750 measurements, depending on the Returned Goods Authorization Number from the Customer Service Department
VALIDATION HI96771-11 Cuvette A into the cuvette light level. and then send it with shipment costs prepaid. When shipping any
holder and ensure that the notch on the 19•The instrument displays “Stor” for one 20
Note: The validation is performed only for the Validation second and the calibration is saved. The remaining battery capacity is evaluated at the instrument startup instrument, make sure it is properly packaged for complete protection.
cap is positioned securely into the groove. To validate your warranty, fill out and return the enclosed warranty card
selected parameter. For full validation of and after each measurement.
3 6• Press ZERO/CFM and the lamp, cuvette 20•The instrument will return automatically within 14 days from the date of purchase.
the instrument, the following procedure The instrument displays a battery indicator with three levels as follows:
and detector icons will appear on the display, to the measurement mode by displaying Recommendations for Users
must be performed for each parameter. depending on the measurement phase. 9 • 3 lines for 100 % capacity
dashes on the LCD. Before using these products, make sure that they are entirely suitable for your specific application and for
1• Turn the meter on by pressing ON/OFF. 7• After a few seconds the display will show • 2 lines for 66 % capacity the environment in which they are used.
2• When the beeper sounds briefly and the “-0.0-”. The meter is now zeroed and Operation of these instruments may cause unacceptable interferences to other electronic equipments, this
• 1 line for 33 % capacity
LCD displays dashes, the meter is ready. 4-5 ready for calibration. The blinking “READ” GLP requiring the operator to take all necessary steps to correct interferences.
• Battery icon blinking if the capacity is under 10 %. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
3• Place the CAL CHECK™ Standard asks for reading calibration standard. 10-11 performance.
In the GLP mode, the last calibration date can be verified and the factory If the battery is empty and accurate measurements can’t be taken any
HI96771-11 Cuvette A into the cuvette 8• Remove the cuvette. calibration can be restored.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
Last more, the instrument shows “dEAd bAtt” and turns off. safety do not use or store the instrument in hazardous environments.
holder and ensure that the notch on the 9• Place the specific CAL CHECKTM Standard
cap is positioned securely into the groove. Calibration To restart the instrument, the battery must be replaced with a fresh one.
HI96771-11 Cuvette B into the cuvette
4• Press ZERO/CFM and the lamp, cuvette holder, for and ensure that the notch on Date To replace the instrument’s battery, follow the steps: Hanna Instruments reserves the right to modify the
and detector icons will appear on the display, the cap is positioned securely into the groove. LAST CALIBRATION DATE • Turn the instrument off by pressing ON/OFF. design, construction and appearance of its products
depending on the measurement phase. 10•Press READ/ /TIMER and the lamp, 1 • Turn the instrument upside down and remove the battery cover by without advance notice.
1• Press RANGE/GLP/ to enter GLP mode.
5• After a few seconds the display will show cuvette and detector icons will appear on turning it counterclockwise.
The calibration month and day will appear
“-0.0-”. The meter is now zeroed and the display, depending on the measurement For additional information, contact your
ready for validation. 7 on the main display and the year on the
phase. dealer or the nearest
secondary display.
6• Remove the cuvette. 11•The instrument will show for three seconds Hanna Customer Service Center.
7• Place the specific CAL CHECKTM Standard the CAL CHECK™ standard value. 2• If no calibration was performed, the factory To find the Hanna Office in your area,
HI96771-11 Cuvette B into the cuvette Note: If the display shows “STD HIGH”, the calibration message, “F.CAL” will appear on
2 visit our web site
holder, for and ensure that the notch on 8 standard value was too high. If the the main display and the instrument returns
the cap is positioned securely into the groove. 12

IST96771
display shows “STD LOW”, the standard to measurement mode after three seconds.
8• Press CAL CHECK key and the lamp, value was too low. Verify that both • Extract the battery from its location and replace it with a fresh one.
cuvette and detector icons together with • Insert back the battery cover and turn it clockwise to close.

11/10
CAL CHECK™ Standard HI96771-11
“CAL CHECK” will appear on the display,
Cuvettes, A and B are free from fingerprints
depending on the measurement phase.

You might also like