Professional Documents
Culture Documents
com/
Clinical Biochemistry
A multicentre study comparing two methods for serum free light chain analysis
RJ Lock, R Saleem, EG Roberts, MJ Wallage, TJ Pesce, A Rowbottom, SJ Cooper, ED McEvoy, JL Taylor and S Basu
Ann Clin Biochem 2013 50: 255
DOI: 10.1177/0004563212473447
Published by:
http://www.sagepublications.com
On behalf of:
Association for Clinical Biochemistry and Laboratory Medicine
Additional services and information for Annals of Clinical Biochemistry can be found at:
Subscriptions: http://acb.sagepub.com/subscriptions
Reprints: http://www.sagepub.com/journalsReprints.nav
Permissions: http://www.sagepub.com/journalsPermissions.nav
What is This?
Abstract
Background: Serum free light chain analysis is now well established in the investigation of monoclonal gammopathies. In the
UK there has, until recently, been a single supplier of kits for such analysis. Recently, a second method using monoclonal
antisera was introduced. We have compared the performance of these two kits in four routine laboratories.
Method: Samples submitted for routine analysis (327 samples, 258 [79%] from patients with B-cell lymphoproliferative
disease) for serum free light chains were tested by both technologies (Freelite, Binding Site and N Latex FLC, Siemens),
according to the manufacturers’ instructions.
Results: Qualitative data were available by both methods on 313 samples for serum free kappa chains and 324 samples for
lambda free light chains. We found poor correspondence of 81% for kappa and 74% for lambda. Five percent of samples were
significantly discordant in these assays.
Conclusions: These assays perform very differently in clinical practice. They cannot be used interchangeably, especially if
monitoring patient responses to therapy.
Note that some samples were not included in the regression or bias analysis. This was either through restrictions of sample volume available or where indeterminate
results arose (either less than detection limit or extremely high). MGUS, monoclonal gammopathy of undetermined significance
Table 2 Reproducibility data based on 10 replicates except N latex FLC between-batch where only three replicates were possible
Kappa serum free light chains Lambda serum free light chains
Coefficient of Coefficient of
Mean (SD) (mg/L) variation (%) Mean (SD) (mg/L) variation (%)
Within-batch
Freelite 11.17 (1.09) 9.76 9.06 (0.69) 7.64
N Latex FLC 13.29 (0.29) 2.17 13.39 (0.25) 1.84
Between-batch
Freelite 16.96 (1.21) 7.12 27.42 (1.92) 7.02
Freelite 30.37 (1.75) 5.76 54.36 (3.49) 6.4
N Latex FLC 13.93 (0.71) 5.09 12.20 (0.44) 3.57
N Latex FLC 36.57 (2.34) 6.39 35.50 (2.00) 5.63
4
Log kappa (N Latex FLC) (log mg/L)
3 Identity
Deming fit
(0.33 + 0.81x)
2
95% CI bands
0
0 1 2 3 4 5
Log kappa (Freelite) (log mg/L)
Identity
2 Deming fit
(0.42 + 0.78x)
95% CI bands
0
0 1 2 3 4
Log lambda (Freelite) log mg/L
Figure 1 Panels a and b show the Deming regression curves for kappa and lambda serum free light chains respectively. For kappa y ¼ 0.33 þ 0.81x and for
lambda y ¼ 0.42 þ 0.78x
1
Difference (log kappa [Siemens] – log kappa [BS])
Bias (0.0189706763)
0.5
95% limits of agreement
(–0.6101632518 to 0.6481046045)
–0.5
–1
–1.5
–2
–2.5
–1 0 1 2 3 4
Mean of All
1
Bias (0.1086558587)
0.5
95% limits of agreement
(–0.6358375873 to 0.8531493048)
0
[Freelite])
–0.5
–1
–1.5
–2
–2.5
–0.5 0.5 1.5 2.5 3.5
Mean of All
Figure 2 Panels a and b show the Altman –Bland plots for kappa and lambda serum free light chains, respectively. Those samples which were most discrepant
(samples with high mean [.log 1.5; .31.6 mg/L] and outside of the 95% limits of agreement) are detailed in Table 3
Ins, insufficient sample for analysis; eGFR, estimated glomerular filtration rate; beta region, unable to quantify owing to interference from other beta region proteins; TS, too small to quantify. Note three samples were
available for patient 4 and two samples for patient 10
Lock et al. Serum free light chain methods comparison
................................................................................................................................................
259
260 Annals of Clinical Biochemistry Volume 50 May 2013
................................................................................................................................................
Table 4 Correspondence data for kappa and lambda free light chain the absolute values for FLC that are recommended as risk
analysis stratification markers in all B-cell malignancies). This
N Latex FLC would include the value of .500 mg/L used to identify
Low Normal High patients who potentially have acute renal failure secondary
(a) Lambda FLC
to multiple myeloma.10 In addition, international guidelines
Freelite Low 37 15 0 recommend a concentration of .100 mg/L to accurately
Normal 19 85 52 monitor patients, this is particularly useful for AL
High 0 2 125 amyloid and oligosecretory/non-secretory multiple
Correspondence 74% myeloma patients. This could have considerable impact on
(b) Kappa FLC the clinical interpretation of results found to be disparate
Freelite Low 4 11 0 between the two assays.
Normal 9 79 18 We found 17 samples (5.2% of all samples) from 14
High 0 23 173
Correspondence 81%
patients where the clonal light chains were poorly detected
by the monoclonal antisera (Table 3), in one case with a
The categories Low, Normal and High are defined by the manufacturers’ highly misleading normal ratio (Table 3, patient 10) or,
reference ranges. Kappa statistic for lambda FLC was 0.58 (interpreted as
more worryingly, not detected at all (Table 3, patient 2).
‘fair’) and for kappa FLC 0.62 (interpreted as ‘moderate’)
More significant outliers were found for lambda than
kappa. It is interesting to note that Pretorius et al. 3 described
essential to determine whether this system, based on the use three patients with lymphoproliferative disease in whom
of monoclonal, rather than polyclonal, antisera is equally normal FLC ratios were seen in the N Latex FLC assay
able to detect and quantify monoclonal free light chains in (1 AL amyloid patient, 1 lambda light chain myeloma
serum. Thus far, we know of only two papers comparing patient and 1 chronic myelomoncytic leukaemia patient
the use of these two assays on routine clinical samples.2,3 with a positive kappa Bence Jones protein). All three were
Our major concern was that there was a theoretical possi- detected as having an abnormal FLC ratio by Freelite.
bility that, with some monoclonal free light chains, the use Similarly, Hoedemakers et al. 2 noted one patient with
of monoclonal antisera with restricted epitope specificity lambda light chain myeloma with a normal ratio by N
might not produce sufficiently large immune complexes to Latex FLC that was abnormal by Freelite.2 So, although
allow detection by light scattering in the nephelometric the numbers are small, and therefore not conclusive, both
systems used for analysis. In contrast to Hoedemakers these and our own study support our contention that
et al.,2 our study included a large percentage of samples some clones might be missed by the more limited epitope
from patients with monoclonal gammopathies. The percen- specificity expected of monoclonal antisera.
tage of samples from monoclonal patients is unclear in the While in general terms the two assays perform similarly
study of Pretorius et al.,3 but we note that many potential in the detection of serum free light chains, we believe that
cases may have been excluded by their selection criteria there are sufficient data here to suggest the assays are not
(samples ,50 mg/L on Freelite were excluded). The three equivalent. We would stress that this last point is very
studies are thus not directly comparable. Nevertheless, important. As there is poor correlation, the current inter-
some themes emerge. national guidelines cannot be cross applied to the Siemens
Correspondence between the two assays was quite low assay. More data are required to establish equivalent FLC
(kappa 81%, lambda 74%), certainly lower than in the concentrations, as quantified by the N Latex FLC assay,
selected data of Pretorius et al. 3 as indicated by the lower for the stratification and management of patients with
kappa statistic, but in many samples this might be expected B-cell lymphoproliferative disorders.
to have little clinical impact (e.g. normal by one assay and
low by the other). Note however that this only applies
if the sensitivities of the assays for detecting disease state
DECLARATIONS
serum samples are the same, for which there are as yet no
conclusive data. There were, however, some worrying Competing interests: None.
results, as shown in Table 3. In some the major concern Funding: None.
is that patients transferring between centres could not be Ethical approval: Not applicable.
monitored effectively if different methods were used in Guarantor: RJL.
the different centres. Contributorship: EGR, MJW, TJP, RS, EM and JT performed
It should be noted that it is unclear as to which result is the analyses. EGR, MJW, SC, RS, TJP, EM and JT helped
the more accurate in many of these patients. There are no locate, source and compile data. RJL analysed the data.
international standards for the assays. There are well docu- All authors contributed to the planning, writing and
mented problems of non-linearity and antigen excess detec- approval of final draft.
tion. Providing that the assay used is consistent and able Acknowledgements: The authors would like to thank both
to detect all clonal free light chains then the numeric the Binding Site and Siemens for the support provided to
differences are academic. However, many of the recommend- allow this study to take place. Our thanks go to Francis
ations in the literature are based on experience with the Smith, Gloucester Royal Hospital for supplying some of
Freelite assay and in the light of our data may not be the data for Table 3 and to Jayne Parkes, Haematology,
directly transferable to the Siemens assay (for example, New Cross for invaluable liaison and sourcing some of
the data. N Latex FLC kits were kindly donated free of 5 CLSI. Method comparison and bias estimation using patient samples;
charge by Siemens. Approved guideline – second edition (interim revision). CLSI document
EP09-A2-IR. Wayne, PA, USA: Clinical Laboratory Standards Institute,
2010
6 Katzmann JA, Clark RJ, Abraham RS, et al. Serum reference intervals and
REFERENCES diagnostic ranges for free k and free l immunoglobulin light chains:
relative sensitivity for detection of monoclonal light chains. Clin Chem
1 Dispenzieri A, Kyle R, Merlini G, et al. International Myeloma 2002;48:1437 – 44
Working Group guidelines for serum-free light chain analysis in multiple 7 Lock RJ, Unsworth DJ. Immunoglobulins and Immunoglobulin
myeloma and related disorders. Leukemia 2009;23:215 –24 subclasses in the elderly. Ann Clin Biochem 2003;40:143 –8
2 Hoedemakers RM, Pruijt JF, Hol S, et al. Clinical comparison 8 Robson EJD, Taylor J, Beardsmore C, Basu S, Mead G, Lovatt T. Utility of
of new monoclonal antibody-based nephelometric assays for free serum free light chain analysis when screening for lymphoproliferative
light chain kappa and lambda to polyclonal antibody-based disorders. Lab Med 2009;40:325 –9
assays and immunofixation electrophoresis. Clin Chem Lab Med 9 Palladini GDA, Gertz MA, Kumar S, et al. Validation of the criteria of
2011;50:489 –95 response to treatment in AL amyloidosis. Blood 2010;11:1364a
3 Pretorius CJ, Klingberg S, Tate J, Wilgen U, Ungerer JP. Evaluation of the 10 Leung N, Gertz MA, Zeldenrust SR, et al. Improvement of cast
N Latex FLC free light chain assay on the Siemens BN analyser: nephropathy with plasma exchange depends on the diagnosis and on
precision, agreement, linearity and variation between reagent lots. Ann reduction of serum free light chains. Kidney Int 2008;73:1282 –8
Clin Biochem 2012;49:450 – 5
4 UK NEQAS. Immunology, Immunochemistry and Allergy Annual
Report 2010. See http://www.kpmd.co.uk/immqas/downloads/
Annual_Report_2010_V1.1_FINAL.pdf (last checked 26 July 2012) (Accepted 17 November 2012)