MODIFICATION INSTRUCTIONS
for the
Vitros 250 Chemistry System

Service Code: 3332


Modification No. 20
Type 2

Purpose: To provide the following after installation of a new INCUBATOR COVER and TIP LOCATOR
COVER:
• Easy installation and removal of the INCUBATOR COVER.
• Easy access to the INCUBATOR.
• Improved reliability of the 250 Chemistry System by eliminating false errors caused by
INTERLOCK SWITCHES.
This Modification is also a prerequisite to installation of the IMMUNO•WASH METERING
ASSEMBLY.

Important
Only qualified personnel should install this modification.
Service Effects: Reduces the number of parts in the service publication.
Special Requirements: None.
Serial Numbers: 25010101 - 25010944.
Installation Time: 2.3 hours.
Special Tools: None.
Parts Status: The Modification Kit is available at the Central Warehouse.
Parts Requirements: See the Parts List on Page 2.

© Ortho-Clinical Diagnostics, Inc.


PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Ortho-Clinical Diagnostics, Inc. (OCD) prior to publication.
No patent license is granted by this information.
OCD reserves the right to change this information without notice, and makes no warranty,
express or implied, with respect to this information. OCD shall not be liable for any loss or
damage, including consequential or special damages, resulting from any use of this information,
even if loss or damage is caused by the negligence or other fault of OCD.

This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Parts List:
Modification Kit 1C2514 includes:

Part No. Description Quantity


682553 SPLICE CLOSED END CRIMP 2
1C2096 INCUBATOR COVER ASSEMBLY 1
1C2188 TIP LOCATOR COVER ASSEMBLY 1
999182 VIEW PORT COVER ASSEMBLY 1
999178 BRACKET ASSEMBLY 1
1C2172 LABEL - ESD, ELECTROMETER 1
356877 THUMBSCREW 1
999125 WEAR PAD, INCUBATOR 3
1C3408 JUMPER 1
1C2190 Packing List 1
1C2189 Modification Instructions 1

2 30MAY96 – 1C2189
SCREW (3) [1] From the main menu, touch “SHUTDOWN
ANALYZER”.
[2] At the prompt, “This function will require a
reboot. Do you wish to continue?”, enter [Y] and
press [øEnter].
[3] When “The ANALYZER has been shutdown,
Reset or power down is required.” is displayed,
de-energize the 250 Chemistry System.
[4] Remove and discard the INCUBATOR COVER.
[5] Remove:
• 3 SCREWS
• 3 GUIDE POSTS
• 2 SCREWS
• magnetic STRIKE PLATE
Note
Do not remove the SHIM under the STRIKE PLATE.
SCREW (2) INCUBATOR COVER
STRIKE PLATE,
not visible GUIDE POST (3)
E012_0740GCA
E012_0740GA

[6] Remove the 2 SCREWS and the VIEW•PORT


COVER.
[7] Install the new VIEW•PORT COVER 999182.

VIEW PORT
COVER-
999182

E012_4721ACA
SCREW (2) E012_4721AA

30MAY96 – 1C2189 3
MODIFICATION INSTRUCTIONS/TYPE 3

[8] Remove the 2 PANELS.

PANEL (2)

E012_0803ACF
E012_0803AA

TORX SCREW [9] Remove:


• TORX SCREW
WASHER • WASHER
SPRING
• SPRING
BEARING INCUBATOR DISK
• BEARING
• BELT
• INCUBATOR DISK
[10] Remove and discard the 3 WEAR PADS.
[11] Install the 3 new WEAR PADS 999125.
[12] Install:
• INCUBATOR DISK
• BEARING
• SPRING
• WASHER
BELT
• TORX SCREW
• BELT

WEAR PAD (3) -


999125

E012_5528CCA
E012_5528CA

4 30MAY96 – 1C2189
[13] Does the 250 Chemistry System have the INTERLOCK SWITCH indicated in the graphic?

Yes No

Advance to Step 14. Advance to Step 21.

CONNECTOR (4)

INTERLOCK SWITCH (2)


BRACKET ASSEMBLY

SCREW (2)

SCREW (4) Do not remove LATCH PLATE E012_4474BCA


LATCH (2) not visible E012_4474BA

[14] Disconnect the 4 wires from the 2 INTERLOCK SWITCHES.


[15] Remove the 2 SCREWS.
[16] Remove and discard the 2 INTERLOCK SWITCHES and BRACKET ASSEMBLY.
[17] Cut:
• the 4 CONNECTORS and discard them
• 1 cm of insulation from the end of each of the 4 wires
[18] Use the 2 SPLICE CLOSED END CRIMPS 682553 to terminate the wires in pairs by color:
• red with black
• green with white
[19] From the back of the 250 Chemistry System, push the disconnected CABLE behind the ISO BOARD. The
CABLE will remain disconnected and inactive.
[20] Remove:
• 4 SCREWS
• 2 LATCHES

30MAY96 – 1C2189 5
MODIFICATION INSTRUCTIONS/TYPE 3

[21] Remove the 2 SCREWS.


ESD LABEL
1C2172 [22] Remove and discard the BRACKET ASSEMBLY
INTERLOCK
from the INTERLOCK SWITCH.
SWITCH
[23] Install:
• new BRACKET ASSEMBLY 999178
• 2 SCREWS, tighten partially
[24] Apply the ESD LABEL 1C2172 to the cover of
the ELECTROMETER.

BRACKET
ASSEMBLY-
SCREW (2) 999178 E012_5812ACB
E012_5812AA

Detach 1

TIP LOCATOR
COVER-
1C2188 INCUBATOR COVER-
1C2096

E012_5538BCA
E012_5538BA

[25] Install the new INCUBATOR COVER 1C2096. If necessary, adjust the BRACKET ASSEMBLY for the
INTERLOCK SWITCH and tighten the 2 SCREWS.
[26] Install new TIP LOCATOR COVER 1C2188.

6 30MAY96 – 1C2189
Detach 2

FAN
ASSEMBLY

CONNECTOR
1B1J/P1

SCREW (3)
E012_0010BCB
E012_0010BA

[27] Disconnect CONNECTOR 1B1J/P1.


[28] Loosen the 3 SCREWS.
[29] Remove the FAN ASSEMBLY.
[30] Disconnect all the CONNECTORS on the
POWER POWER CONTROL and DISTRIBUTION
DISTRIBUTION BOARD.
BOARD [31] Remove the POWER DISTRIBUTION BOARD.

E012_1768ACB
E012_1768AC

30MAY96 – 1C2189 7
MODIFICATION INSTRUCTIONS/TYPE 3
Detach 3

3 3 3

5 5 5

1 6 1 6 1 6

POWER CONTROL JUMPER


and DISTRIBUTION BOARD
E012_1773HCA
E012_1773HC

[32] Use a SOLDERING IRON to install the JUMPER


3A3J/P6 1C3408 on the POWER DISTRIBUTION
BOARD as indicated in the illustration.
3A3J/P3 [33] For both PIN 10 (red wire) and PIN 22 (black
wire) on 3A3P1:
3A3J/P2
(a) Cut the wires close to the CONNECTOR.
3A3J/P7
(b) Cut the wires close to the CABLE
3A3J/P5 insulation.
TERMINAL Note
BLOCK 1
Removing a length of the wire will prevent a short
3A3J/P4 circuit in the wires.
[34] Connect all the CONNECTORS and install the
3A3J/P1
POWER DISTRIBUTION BOARD.

TERMINAL E012_1768ACA
[35] Install the FAN ASSEMBLY.
E012_1768AC
BLOCK 2 [36] Tighten the 3 SCREWS.
[37] Connect CONNECTOR 1B1J/P1.

8 30MAY96 – 1C2189
Detach 4

SLIDE SUPPLY ACCESS DOOR

WASHER

SCREW

E012_0730DCB
E012_0730DA

[38] Remove and discard from the SLIDE SUPPLY


ACCESS DOOR:
1 2 3 4 5 6 7 8 9 10 11 • SCREW
12 13 14 15 16 17 18 19 20 21 • WASHER
22 23 24 25 26 27 28 29 30 31 32 [39] Install the THUMBSCREW 356877.
33 34 35 36 37 38 39 40 41 42
[40] Circle 20 on the MODIFICATION LABEL for
43 44 45 46 47 48 49 50 51 52 53 the SLIDE PROCESSOR.
54 55 56 57 58 59 60 61 62 63 [41] Return the 250 Chemistry System to the normal
64 65 66 67 68 69 70 71 72 73 74 operating condition.
75 76 77 78 79 80 81 82 83 84 [42] Provide feedback through the Call Management
85 86 87 88 89 90 91 92 93 94 95 System:
96 97 98 99 A1 A2 A3 A4 A5 A6
Service Code 3332
A7 A8 A9 B1 B2 B3 B4 B5 B6 B7 B8
Modification No. M20
B9 C1 C2 C3 C4 C5 C6 C7 C8 C9
Part Kit No. 1C2514
E012_0034AC

30MAY96 – 1C2189 9
Vitros is a trademark.

1c2189.fm+

Printed In USA Ortho-Clinical Diagnostics, Inc. ■ 100 Indigo Creek Drive ■ Rochester, N.Y. 14626