Professional Documents
Culture Documents
Nanomaterials: Characterization
5.1 Goals
• Overview of characterization techniques.
Read:
• P. S. Hale et al., J. Chem. Educ. 82 (5), 775 (2005). Growth kinetics and modeling of
ZnO nanoparticles.
61
62 CHAPTER 5. NANOMATERIALS: CHARACTERIZATION
Figure 5.1: A table-top XRD set-up and results for a w-ZnS nanowire.
can be obtained. On the other hand, single crystals might not be readily available and
orientation of the crystal is not straightforward.
One disadvantage of XRD is the low intensity of diffracted beam for low-Z materials.
of an inelastic process whereby the energy of the incoming light is changed. The others are
elastic processes where the intensity is changed. Typical penetration depth is of the order
of 50 nm. Optical spectroscopy is attractive for materials characterization because it is fast,
nondestructive and of high resolution.
I
A = − log( ) = a × b × c, (5.2)
I0
Figure 5.2: Schematic of UV-vis spectrophotometer, an apparatus and a sample data on ZnO
nanoparticles. Absorption spectra for a colloid aged at 650 C. The spectra were recorded at
0, 1, 3, 5, 11, 16, 30, 60, 90, and 120 min after immersion in the aging water bath [9].
64 CHAPTER 5. NANOMATERIALS: CHARACTERIZATION
Figure 5.3: Absorption data from bulk cubic ZnS and w-ZnS nanowire.
A typical accelerating voltage is 100 kV for which the electrons have mean free paths
of the order of a few tens of nm for light elements and a few hundreds of nm for heavy
elements. These would be the ideal film thicknesses since much thinner films would lead
to little scattering and much thicker ones would lead to too many scattering of the same
electron resulting in a blurred image of low resolution.
The imaging mode can be controlled by the use of an aperture. If most of the unscattered
electron is allowed through, the resulting image is called a bright-field image. If specific
scattered beams are selected, the image is known as a dark-field image. Examples are shown
in Fig. 5.7. In addition to forming images, a TEM can be used for chemical analysis and
melting-point determination.
If the TEM is operated in scanning mode, it is known as a scanning electron microscope
or SEM. Electrons scattered from the sample are collected on a CRT to form the image. The
resolution is a few nm and magnification is from ∼ 10 to 500,000 times. One such SEM is
shown in Fig. 5.8.
66 CHAPTER 5. NANOMATERIALS: CHARACTERIZATION
then move creating a current. This current is known as a tunneling current whose magnitude
(as for all tunneling currents) is very sensitive to the surface–tip separation (the region in
between effectively acting as a barrier):
I ≈ V e−2κd , (5.3)
where κ is the wave number of the electrons. The current should also depend on the density
of electron states. A constant current would correspond to a constant altitude of the tip with
respect to the surface (Fig. 5.10). Hence a constant-current scan of the two-dimensional plane
reveals the surface structure. The tip motion can then be converted into a grayscale image.
The other mode of operation of an STM is in constant height mode (Fig. 5.10). Note that,
classically, the system consists of an open circuit, hence there should be no current.
The extreme sensitivity to the sample–probe separation translates into a very high pre-
cision in the vertical height (∼ 0.01 nm), with the separation being typically around 1 nm.
Indeed, the piezoelectric mechanism has extension coefficients of ∼ Å/V, resulting in very
accurate vertical movement of the tip for low voltages (typically, mV for metals and V for
semiconductors due to the latter’s band gap). The lateral resolution, on the other hand, is
determined by the tip size which can be down to a single atom and the best resolution is
∼ 0.1 nm. Prior to the STM, traditional microscopes have been based on light and electrons.
70 CHAPTER 5. NANOMATERIALS: CHARACTERIZATION
The resolution of the latter types of microscopes is limited by diffraction effects; basically,
the resolution ∼ wavelength λ/2. An example of an STM image is shown in Fig 5.11.
Example
λvisible ∼ 500 nm.
λe (1 keV) = hp = √2mE
h
∼ 0.1 nm.
While the electron microscope can display atomic resolution, the high energy required
means that the primary electron beam penetrates deep into the material and is not sensitive
to surface features. Both of the above types of microscopes require focussing elements. A
major new feature of the STM are the moving parts (controlled by piezoelectric materials)
and the shielding of the latter from external vibrations (using electromagnetic damping via
eddy currents).
A limitation of the STM is the requirement for a conducting sample. A variation on
the STM for insulators is the atomic-force microscope (AFM) which operates on the atomic
force between the sample atoms and the probe atoms (in contrast to the tunneling current
for the STM).
STM chemistry
The STM has also been used to carry out chemical reactions on surfaces. This relies on a
tip–sample interaction. The force can be either attractive or repulsive.
Thus Hla et al. [1] induced all the steps of the Ullmann reaction with the STM tip
5.7. ATOMIC FORCE MICROSCOPY 71
(Fig. 5.12); this has the potential to lead towards single-molecule engineering. The reaction
is
2C6 H5 I + 2Cu → C12 H12 + 2CuI.
Since no electric current is involved, the tip/sample does not have to be metallic. There
are two modes of operation: contact mode whereby the sample-tip distance is so small that
the important force is the core-core repulsive one, and noncontact mode where the force is
the van der Waals one. AFM’s can achieve a resolution of 10 pm.
Examples of AFM images are shown in Fig. 5.14.
5.7. ATOMIC FORCE MICROSCOPY 73
Figure 5.14: Patterns formed using AFM in polymers, 80 nm deep. From Nature Materials
2, 468472 (2003).