Professional Documents
Culture Documents
Final Report
Final Report
A report on
SUBMITTED BY
To optimise the design and materials of a hip implant, different experiments are
performed on the strength, wear, fixation and friction of that implant. For this reason, one
must calculate the contact forces in the hip joint. Such information is also needed for giving
guidelines to physiotherapists how to treat a patient after treatment. The movement in the hip
joint has to be known when implant wear is tested or the load directions relative to the pelvis
are calculated from the forces acting at the femur. The experiment was performed on four
patients during the most frequent and day to day activities and an average was calculated. The
related data is taken from the HIP 98 disc.
2. Procedure
Measurements were taken in four patients during frequent activities of daily living.
The measured hip contact forces served to check the validity of calculated results from
inverse dynamics approach. The in vivo data was collected using instrumented hip implants
featuring telemetric sensors and the kinematic data was acquired through marker tracking.
3. Co-ordinate Systems
Two kinds of instrumented total hip implants with telemetric data transmission were
used to measure the hip contact forces with an accuracy of 1% at a rate of approximately 200
Hz. The hip contact force with the magnitude F and the components −𝐹𝑥 , −𝐹𝑦 𝑎𝑛𝑑 − 𝐹𝑧 were
measured in the ‘femur coordinate system’ X, Y and Z (Fig. 1). It is transmitted by the
acetabular cup to the implant head. The angles of inclination of F in three planes are denoted
as 𝐴𝑥 , 𝐴𝑦 𝑎𝑛𝑑 𝐴𝑧 . The force F causes an ‘implant moment’ M around the intersection point
NS of shaft and neck axes of the implant. The moment components 𝑀𝑥 , 𝑀𝑦 ′ 𝑎𝑛𝑑 𝑀𝑧 ′ act
clockwise around the axes X, Y’ and Z’ of the ‘implant system’ which is rotated by the angle
S relative to the femur coordinate system. Important for the implant fixation is the torsional
moment 𝑀𝑡 = −𝑀𝑧 ′ in the transverse plane which rotates the implant inwards around the
shaft axis. The other two components of M are of minor importance; they depend on the
definition of the point NS around which M is calculated.
Figure 1: Coordinate system for measured hip contact forces.
Figure 2: Joint centres, reference points and coordinate systems for gait
4. Activities
The experiments were conducted during different daily activities which were thought
to be most strenuous for the hip joint. These include slow walking, normal walking, fast
walking, and climbing upstairs, down stairs, standing up, sitting down, standing on 2-1-2 legs
and bending the knee.
Table (1): Investigated activities:-
During all activities, time dependent kinematic and kinetic data were gathered:
5. Data Presentation
The primary objective of this study is to find out the actual forces encountered by the
femur head to use them in design of implants. Thus, the data obtained for each patient was
averaged by a new method to get the representative curves for a fictional average patient
which can then be used in design.
Describes the spatial and temporal components of motion (position, velocity, and
acceleration) with no consideration of the forces causing the motion.
(a) Video:
(i) Taking a lot of photographs very quickly (i.e. video) allows you to
better see how posture varies throughout a movement. 60 frames per second is
adequate for most movements. To quantify movement, we need to identify
landmarks on the body.
(ii) We can measure movement of body segments and landmarks from
frames of the video after it is recorded.
(iii) A more accurate way to quantify movement is to record the movement
of body landmarks directly. This is commonly called motion capture. Motion
capture systems detect special dots of light, and record motion of the dots as a
measure of body motion.
Passive marker systems, such as the Vicon system, emit light from LEDs
placed in rings around cameras; the cameras detect light reflected from markers
placed on the subject. The markers cannot be individually identified the subject does
not have to wear wires etc. A schematic is shown below.
Software quantifies the location of the markers. This position data can then be
used to calculate various kinematic parameters.
Figure 3: As example of software analysis through motion capture system.
One stride lasts from heel contact of one limb to the subsequent heel contact of
the same limb. As running speed increases, the fraction of stride time taken up by
stance phase decreases. Stride length and stride frequency do not contribute equally to
increases in running speed.
Below 7 m/s:
(i) Stride length and stride frequency increase approximately linearly with
speed
(ii) Increases in speed are mainly achieved by increasing stride length
Above 7 m/s:
8. Inverse Dynamics:
The inverse dynamics approach was used to calculate the reaction forces and
moments on joints (ankle, knee and hip) using the velocity and acceleration (linear and
angular) found from gait analysis. The reaction at ankle was found first followed by knee and
hip joint.
9. Biomechanical Models:
Transducers have been developed that can be implanted surgically to measure the
force exerted by a muscle at the tendon. However, such techniques have application only in
animal experiments.
The process by which the reaction forces and muscle moments are calculated is called
link segment modelling.
(i) Gravitational Forces: The forces of gravity act downward through the
CoMs of each segment and are equal to the magnitude of the mass times
acceleration due to gravity (normally 9.8 𝑚/𝑠 2 ).
(ii) Ground Reaction or External Forces: Any external forces must be
measured by a force transducer. Such
forces are distributed over an area of
the body (such as the ground reaction
forces under the area of the foot). In
order to represent such forces as
vectors, they must be considered to
act at a point that is usually called the
centre of pressure (COP).
(iii) Muscle and Ligament Forces.
The net effect of muscle activity at a
joint can be calculated in terms of net
muscle moments. If a contraction is
taking place at a given joint, the
analysis yields only the net effect of
both agonist and antagonistic muscles.
Also, any friction effects at the joints
or within the muscle cannot be
separated from this net value. Increased friction merely reduces the effective
“muscle” moment; the muscle contractile elements themselves are actually
creating moments higher than that analysed at the tendon. At the extreme
range of movement of any joint, passive structures such as ligaments come
into play to contain the range. The moments generated by these tissues will
add to or subtract from those generated by the muscles.
Each body segment acts independently under the influence of reaction forces and
muscle moments, which act at either end, plus the forces due to gravity. Consider the planar
movement of a segment in which the kinematics, anthropometrics, and reaction forces at the
distal end are known.
(a) Known
𝑎𝑥 , 𝑎𝑦 = acceleration of segment COM
θ = angle of segment in plane of movement
α = angular acceleration of segment in plane of movement
𝑅𝑥𝑑 , 𝑅𝑦𝑑 = reaction forces acting at distal end of segment,
𝑀𝑑 = net muscle moment acting at distal joint, 28
(b) Unknown
𝑅𝑥𝑝 , 𝑅𝑦𝑝 = reaction forces acting at proximal joint
𝑀𝑝 = net muscle moment acting on segment at proximal joint
(c) Equations
∑Fx = m 𝑎𝑥
𝑅𝑥𝑝 − 𝑅𝑥𝑑 = m 𝑎𝑥
∑Fy = m 𝑎𝑦
𝑅𝑦𝑝 − 𝑅𝑦𝑑 − 𝑚𝑔 = m 𝑎𝑦
In a static situation, a person is standing on one foot on a force plate. The ground
reaction force is found to act X cm anterior to the ankle joint. Note that convention has the
ground reaction force 𝑅𝑦1 always acting upward. We also show the horizontal reaction force
𝑅𝑥1 to be acting in the positive direction (to the right). If this force actually acts to the left, it
will be recorded as a negative number. Calculate the joint reaction forces and net muscle
moment at the ankle. 𝑅𝑦1 = body weight. If the foot is in swing phase, the ground reaction is
zero.
∑Fx = m 𝑎𝑥
𝑅𝑥2 + 𝑅𝑥1 = m 𝑎𝑥 = 0
∑Fy = m 𝑎𝑦
𝑅𝑦2 + 𝑅𝑦1 − 𝑚𝑔 = m 𝑎𝑦
→ 𝑅𝑦2 = −𝑅𝑦1 + 𝑚𝑔 + m 𝑎𝑦
HIP 98 Database
Hip joint loading: Contact forces + Muscles forces + Gait data + Activities
Patients included:
H. Sonke (HSR)
P. Fichtner (PFL)
K. Walker (KWR)
Boehme (IBL)
Activities performed:
Walking
Slow
Normal
Fast
Climbing
Up stairs
Down stairs
Standing up (one leg)
Standing up (both leg)
Sitting down
Knee bend
Measuring data can be seen in HIP 98 database e.g. Normal walking.
In force window, three forces Fx , Fy and Fz and also resultant force F.
These forces are shown by different colour e.g. in graph between force (%BW) and gait
cycle.
Besides force graph, you can also see force magnitude and force direction in three
different planes (sagittal, frontal and transverse) and also its numerical data. Gait pattern can
also be analysed in these three planes with the help of animation command and you can also
vary (increase or decrease) frame time interval.
Scales for graphs:
Time and vector diagrams
Automatic [Scaling range define automatically]
Fixed standard [Scaling range as per standard data]
Choose [We give scaling arbitrary]
Activities performed:
Walking
• Slow
• Normal
• Fast
Climbing
• Upstairs
• Down stairs
Standing up (one leg)
Standing up (both leg)
Sitting down
Knee bend
Data available:
Measured load components
Measurement of 6 load components are reported here in the coordinates system of the femur.
Implant is rotated about 3 axes.
The general pattern and the magnitudes of the ground reaction forces were similar for all
trials. The ground reaction forces were characterized by a dominant, vertically directed
component. The relative variability of the vertical peak forces for a single patient
For both walking and stair climbing, the general characteristics of the resultant inter-
segmental moments at the hip were similar for the different trials for each patient. The largest
moment was always the flexion–extension moment.The average relative variability in the
flexion extension moment for all patients was
a. 19% during walking, and
b. 11% during stair climbing
Observations
Ground reaction forces : used to indicate the reproducibility of the gait pattern.
Large deviations in the pattern and magnitude would have shown variations in the
patients’ performance of activities, e.g. in different trials of walking.
The small variations (observed) suggest that the patients walked in a quite similar and
reproducible way.
Skin movement errors : affect the location of bony landmarks derived from the
marker positions
Longitudinal scaling of the bones was unaffected by errors in the spatial position of
the landmarks.
Thigh revealed erroneous axial rotation data, as the femoral condyles did not match
the tibia.
To remove this, the 3-D model allowed correction of the rotation of the affected limb by
controlling the congruency of the femoral condyles and the tibial plateau.
Since the participants of this study underwent THA, it may not be expected that the
loading conditions in these patients are identical to those in healthy subjects. Nevertheless,
the approach presented here represents the only methodology to achieve a validation with
respect to hip contact forces.
Inverse dynamics calculation is an iterative process starting from the ankle joint, the
largest errors due to error propagation and error accumulation were most likely to occur at the
most proximal joint in the model, the hip joint.
Ground reaction forces showed an intra-individual variation of 4%,
The flexion-extension moments at the hip varied by as much as 19%.
The actual three-dimensional volumetric structures and curved pathways of the
muscles had to be simplified and modeled as straight lines.
This explain why the correlation between hip contact forces measured in vivo and
calculated hip contact force components in the transverse plane was not as good as for the
axial component.
Conclusions
The musculo-skeletal model of the lower extremity presented in this study allowed
prediction of proximal femoral loading for walking and stair climbing in four THA patients
The calculated hip contact forces and those measured in vivo during these every day
activities were similar. However, a varying degree of conformity between the individual force
components was found.
The component acting along an idealized femoral midline showed best agreement,
While the results for the significantly smaller forces in the transverse plane were less
accurate.
The cycle-to-cycle validation revealed that absolute peak loads differed by an average
of only 12% during walking and 14% during stair climbing.
In order to predict musculo-skeletal loading conditions, two issues seem to be
important.
Availability of a suitable measuring procedure to validate the prediction
o the in vivo measured hip contact forces
Patient individual models should be used to approximate the loading conditions in
each individual case.
o The biomechanical model used in the present study was adapted to the individual
anatomy and prosthesis configuration.
Model presented here can be considered to be a means to :
to determine mechanical boundary conditions for
o implant loading,
o bone remodeling, and
o Fracture healing.
Investigating the implant–bone load sharing and primary stability of implants under
loading conditions.
Analysis of exercises usually carried out during rehabilitation can be of help in
identifying activities with possibly hazardous musculo-skeletal loading and thus improve the
outcome of rehabilitation.
REFERENCE
[1] Bergmann, G., Deuretzbacher, G., Heller, M., Graichen, F., Rohlmann, A., Strauss, J.,
& Duda, G. N. (2001). Hip contact forces and gait patterns from routine
activities. Journal of biomechanics, 34(7), 859-871.
[2] Heller, M. O., Bergmann, G., Deuretzbacher, G., Dürselen, L., Pohl, M., Claes, L., ...
& Duda, G. N. (2001). Musculo-skeletal loading conditions at the hip during walking
and stair climbing. Journal of biomechanics, 34(7), 883-893.
[3] https://orthoload.com