MAMMOMAT Novation
SP

Replacements of Parts
Component

Detector cable CBL-00173

10410630

© Siemens 2008
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: SPB7-250.841.04.01.02 English


Replaces: n.a. Doc. Gen. Date: 02.08
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 2 of 18 Siemens


02.08 CS PS SP
Table of Contents 3
0Table of Contents

1 _______ General Information______________________________________________ 4

Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required aids and tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 _______ Replacing the Detector cable ______________________________________ 6

General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Replacing the detector cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Final Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Test exposures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Disposal information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

3 _______ Changes to Previous Version _____________________________________ 17

Siemens SPB7-250.841.04.01.02 Page 3 of 18 MAMMOMAT Novation


02.08 CS PS SP
4 General Information
1-
Notes
1General Information

Only qualified and system-trained service staff is allowed to perform the MAMMOMAT
Novation system installation, service, maintenance, and quality assurance. Ensure that the
most recent version of the service documentation is available.
To avoid any risk of injury to persons and/or damage to the system read and observe the
General Safety Notes (TD00-000.860.01.xx.02).

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 4 of 18 Siemens


02.08 CS PS SP
General Information 5

Requirements 1.1

Required documents 0

Document Print No.


Replacement of Parts, Detector cable CBL-00173 SPB7-250.841.04.01.02
General safety information TD00-000.860.01.xx.02
System Startup, with WH AWS SPB7-250.815.xx.xx.02

In addition to the technical documentation, the


• Instructions for Use and the
• Quality Control manual are needed.

Required material 0

Tab. 1 Spare part detector cable:

Quant. Material No. Name


1 104 106 30 Detector cable CBL-00173

Required aids and tools 0

Standard tools

Tool Order No.


Standard installation and service tools n.a.

Siemens SPB7-250.841.04.01.02 Page 5 of 18 MAMMOMAT Novation


02.08 CS PS SP
6 Replacing the Detector cable
2-
General information
2Replacing the Detector cable

General information 0

Under unfavorable conditions different communication problems may be observed within


the MAMMOMAT Novation system, in the small network, consisting of the FFDM detector,
the Brick, the WH AWS and the generator (via the D750 Master circuit board).
These instructions describe how to replace the detector cable CBL-00173 in the MAMMO-
MAT Novation to eliminate communication problems caused by this cable.

NOTE See also the “Contribution to the efforts to eliminate some


communication problems” in Speed Info ID: SP_76/07.

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 6 of 18 Siemens


02.08 CS PS SP
Replacing the Detector cable 7

Replacing the detector cable 2.1

Removing the rear covers of the detector


• To remove the lower rear cover of the object table assembly, unscrew the four M2
screws on the circumference (1/Fig. 1 / p. 7) and the six M3 screws on the bottom, see
(2/Fig. 1 / p. 7).

NOTE For Novation S:


Before you remove the upper rear cover, the two small face-
plates which covers the inserting holes of the magnification
table rails have to be removed. Therefore, pull out the face-
plates, see (Fig. 2 / p. 7). Caution: It will require a substantial
effort.

Fig. 2: Faceplate2
Fig. 1: Back covers_

• Remove the upper rear cover of the object table assembly. The cover fits tightly and
may be slightly jammed by the bellows; remove it carefully by slowly lifting the front.
• Remove the metal cover from above the detector main cable connector. This cover is
fixed with two Allen screws M3 x; remove them, see (1/Fig. 3 / p. 8).

Siemens SPB7-250.841.04.01.02 Page 7 of 18 MAMMOMAT Novation


02.08 CS PS SP
8 Replacing the Detector cable

Fig. 3: Remove cable cover Fig. 4: Detector cable 2

Disconnecting the detector cable


• Unscrew the screws that secure the plug and socket of the detector cable connector to
each other and disconnect them (1/Fig. 4 / p. 8). The shape clearly indicates how to
plug this connector back in.

Fig. 5: Detector cable 3 Fig. 6: Detector cable 4

• Push the connector of the detector main cable out under the right longitudinal rail see
(1/Fig. 5 / p. 8) and (1/Fig. 6 / p. 8).

Open the detector housing


• Switch ON the system and turn the swivel arm to 180° (Fig. 7 / p. 9). Switch OFF the
system.

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 8 of 18 Siemens


02.08 CS PS SP
Replacing the Detector cable 9

Fig. 7: Turn swivel arm Fig. 8: Detector base plate

• Unscrew the six screws (1/Fig. 8 / p. 9) of the base plate of the detector housing.
• Carefully lift up the base plate. In order to this, the flying wing lever (2/Fig. 8 / p. 9) has
to be raised a little.
• Place the base plate onto the bottom of the detector, as shown in (1/Fig. 9 / p. 9). Pay
attention to the cables, they shouldn’t get damaged !
Unplug the connectors
• Unplug the connectors: J3 (1/Fig. 10 / p. 9), J6A (2/Fig. 10 / p. 9), J7 (3/Fig. 10 / p. 9),
J6 (4/Fig. 10 / p. 9), J5 (5/Fig. 10 / p. 9) und J3 (6/Fig. 10 / p. 9).

Fig. 9: Detector cable 1 Fig. 10: Detector boards

Removing the detector cable


• Remove the cable strap (7/Fig. 10 / p. 9).
• Remove the cable holder by unscrewing the screw shown in (1/Fig. 11 / p. 10).

Siemens SPB7-250.841.04.01.02 Page 9 of 18 MAMMOMAT Novation


02.08 CS PS SP
10 Replacing the Detector cable

Fig. 11: Cable holder 1 Fig. 12: Remove cable holder

• Remove the cable holder with the ferrite cores (1/Fig. 12 / p. 10).

Fig. 13: Remove old detector cable Fig. 14: New detector cable

• Pull the old detector cable downwards through the cable outlet, see (1/Fig. 13 / p. 10).

Insertion of the new detector cable


• Take the new detector cable (Fig. 14 / p. 10) and draw it in, the opposite way.

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 10 of 18 Siemens


02.08 CS PS SP
Replacing the Detector cable 11

Fig. 15: Detector cable 4


• The cable length on the plug side should be 13.5 cm, measured as of the edge of the
cable outlet, see (1/Fig. 15 / p. 11).
• Mount the cable holder with the ferrite cores (1/Fig. 16 / p. 11) and fix it with the screw
(1/Fig. 17 / p. 11).

Fig. 16: Cable holder 4 Fig. 17: New detector cable 2

• Bend the detector cable around the cable holder, as shown in (2/Fig. 17 / p. 11).
• Position the cable along the edge of the board (Fig. 18 / p. 12).

Plug in the connectors


• Connect the connectors: J3 (1/Fig. 18 / p. 12), J6A (2/Fig. 18 / p. 12), J7
(5/Fig. 18 / p. 12), J6 (6/Fig. 18 / p. 12), J3 (3/Fig. 18 / p. 12) und J5 (4/Fig. 18 / p. 12).

Siemens SPB7-250.841.04.01.02 Page 11 of 18 MAMMOMAT Novation


02.08 CS PS SP
12 Replacing the Detector cable

Fig. 18: Detector cable 2 Fig. 19: Cable strap

• Use a cable strap to fix the cable of the connector J6/ J7, see (1/Fig. 19 / p. 12).

Closing the detector housing


• For remounting the base plate of the detector, the flying wing lever has to be raised a lit-
tle, see (Fig. 20 / p. 12).

Fig. 20: Flying wing lever

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 12 of 18 Siemens


02.08 CS PS SP
Replacing the Detector cable 13

Fig. 21: Turn swivel arm Fig. 22: Detector base plate

• Remount the base plate of the detector housing, see (Fig. 21 / p. 13).
• Screwing on the detector cover with the six screws, see (1/Fig. 22 / p. 13).

Connecting the detector cable


• Switch ON the system and turn the swivel arm back to 0°.
• Switch OFF the system.

Fig. 23: Detector cable 4 Fig. 24: Detector cable 3

• Insert the detector cable (1/Fig. 23 / p. 13) underneath the grid holder
(1/Fig. 24 / p. 13).
• Place the Distance plate (if available) between the two genders (plug and receptacle) of
the connectors X608, see (Fig. 25 / p. 14) and connect the detector cable to the cable
coming out of the wing (2/Fig. 26 / p. 14).

Siemens SPB7-250.841.04.01.02 Page 13 of 18 MAMMOMAT Novation


02.08 CS PS SP
14 Replacing the Detector cable

Fig. 26: Detector assembly

Fig. 25: Distance piece

• Turn the actuating screw until there is no gap between the distance plate and the plug.
• Check that the pins on the connectors haven’t been pushed back.

Mounting the covers

Fig. 28: Back covers_


Fig. 27: Detector cable cover

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 14 of 18 Siemens


02.08 CS PS SP
Replacing the Detector cable 15

• Mount the new metal cover over the detector cable using 2 screws, see
(1/Fig. 27 / p. 14). Make sure that no cables are pinched.
• Mount the bottom wing cover with 6 screws (2/Fig. 28 / p. 14).
• Mount the top wing cover with 4 screws (1/Fig. 28 / p. 14). The cover fits tightly and may
be slightly jammed by the bellows; place it carefully by slowly lifting the front.

NOTE For the Novation S: Insert the two faceplates into the mag
table holes, see (Fig. 29 / p. 15) and push them in
(Fig. 30 / p. 15). Caution: Because of tight fitting, the inser-
tion will require some force.

Fig. 29: Faceplate out Fig. 30: Faceplate in

Siemens SPB7-250.841.04.01.02 Page 15 of 18 MAMMOMAT Novation


02.08 CS PS SP
16 Replacing the Detector cable

Final Work Steps 2.2

Test exposures 0

Test exposures
• Switch ON the system and perform some test exposures to make sure the system
works fine.

The test exposures - customary after any interference with the system by the CSE, should
be performed as an evidence of the fact that the system is being handed over from the CSE
to the operator good in working conditions.

Disposal information 0

The old detector cable should be disposed off in a manner complying withthe rules of envi-
ronmental protection and waste disposal.

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 16 of 18 Siemens


02.08 CS PS SP
Changes to Previous Version 17
3Changes to Previous Version 3-
Version 01 (initial version)

Chapter Changes
All n.a.

Siemens SPB7-250.841.04.01.02 Page 17 of 18 MAMMOMAT Novation


02.08 CS PS SP
18 Changes to Previous Version

MAMMOMAT Novation SPB7-250.841.04.01.02 Page 18 of 18 Siemens


02.08 CS PS SP