Professional Documents
Culture Documents
Assay Summary: ADVIA Centaur ADVIA Centaur XP
Assay Summary: ADVIA Centaur ADVIA Centaur XP
ADVIA Centaur ® XP
Immunoassay Systems
AFP
Assay Summary
Sample Type Serum, amniotic fluid
Sample Volume 10 µL
Calibrator D
Sensitivity and Assay Range 1.3 – 1000 ng/mL (1.08 – 830 IU/mL)
Contents
REF Contents Number of Tests
03305838 5 ReadyPack® primary reagent packs containing ADVIA Centaur® 500
(110764) AFP Lite Reagent and Solid Phase
ADVIA Centaur AFP Master Curve card
or
03974780 1 ReadyPack primary reagent pack containing ADVIA Centaur 100
(110763) AFP Lite Reagent and Solid Phase
ADVIA Centaur AFP Master Curve card
Intended Use
For in vitro diagnostic use in the quantitative determination of alpha-fetoprotein (AFP) in the
following:
• human serum and in amniotic fluid from specimens obtained at 15 to 20 weeks gestation,
as an aid in detecting open neural tube defects (NTDs) when used in conjunction with
ultrasonography and amniography testing,
• human serum, as an aid in managing non-seminomatous testicular cancer when used in
conjunction with physical examination, histology/pathology, and other clinical evaluation
procedures, using the ADVIA Centaur and ADVIA Centaur XP systems.
WARNING: The concentration of AFP in a given specimen, as determined by assays from different
manufacturers, can vary due to differences in assay methods and reagent specificity. The results
reported by the laboratory to the physician must include the identity of the AFP assay used.
Values obtained with different AFP assay methods cannot be used interchangeably. Before
changing assay methods, the laboratory must do the following:
• For prenatal testing, the laboratory must establish a range of normal values for the new
assay based on normal serum and amniotic fluid from pregnant women with a confirmed
gestational age.
• For cancer management, the laboratory must perform additional testing to confirm baseline
values for patients being serially monitored.
United States federal law restricts this device to sale and distribution by or on the order of a
physician, or to a clinical laboratory; and use is restricted to, by, or on the order of a physician.
Use AFP results only as part of the overall clinical evaluation of a patient. Do not use AFP
results as the only criterion for diagnosis. (Refer to Warnings, Summary and Explanation of the
Test, and Limitations.)
Warnings
Elevated MSAFP (maternal serum AFP) levels may indicate open NTD, but are not used to
diagnose the defect without additional testing. In addition, elevated MSAFP levels may
indicate other forms of fetal distress or malformation, which include placental malformations,
ventral wall defects, fetal kidney dysfunction, and fetal death. MSAFP levels may also be
elevated in certain benign and malignant conditions not related to pregnancy. These conditions
include hepatitis, cirrhosis, ataxia telangiectasia, primary hepatocellular carcinoma, and
certain germ cell cancers. Furthermore, incorrect estimation of gestational age can result in
either under- or over-estimation of open NTD risk. Therefore, AFP testing requires accurate
gestational dating for reliable risk assessment for open NTDs. Confirmatory procedures such
as ultrasonography, amniography, amniotic fluid acetylcholinesterase, and amniotic fluid AFP
must be used in conjunction with MSAFP testing for accurate NTD risk assessment.
When using AFP in the evaluation of fetal defects, laboratories must establish their own
median values for each gestational week. Absolute AFP values may vary for each lab
depending on the demographics of its population including race and maternal weight.
Collect maternal serum specimens for NTD testing before amniocentesis. Refer to Special
Precautions for detailed information.
The ADVIA Centaur AFP assay is not a screening test for cancer and must never be used as
such. AFP testing is a safe and effective supplement to patient care when used as part of the
overall management strategy for patients undergoing treatment for non-seminomatous
testicular cancer or for patients being monitored after therapy is complete.
Do not interpret serum AFP as absolute evidence of the presence of malignant disease. At time
of presentation, patients with confirmed non-seminomatous testicular cancer may have serum
AFP concentrations within the range observed in healthy individuals. Since elevated AFP
levels are often found in patients with other malignant and non-malignant conditions, the
physician should rule out all other conditions associated with elevated AFP levels prior to the
use of the ADVIA Centaur AFP values in non-seminomatous testicular cancer management.
Conversely, low concentrations of AFP are not necessarily indicative of absence of disease,
particularly post-surgery or after chemotherapy. Testicular tumors that are histologically
categorized as pure seminoma do not synthesize AFP. The ADVIA Centaur AFP assay, as a
useful adjunct in cancer management, is intended for the evaluation of non-seminomatous
testicular cancer, or mixed tumors with non-seminomatous elements, but not for pure
seminoma. Additionally, several histologic subtypes of non-seminoma either do not synthesize
AFP (choriocarcinoma) or do so unpredictably (teratoma). Therefore, AFP levels should be
used concurrently with other diagnostic and clinical patient information.
Optional Reagents
REF Description Contents
07948423 ADVIA Centaur Multi-Diluent 2 2 ReadyPack ancillary reagent packs
(110314) containing 10 mL/pack
04855629 Multi-Diluent 2 50 mL/vial
(672260)
672428 AFP Master Curve Material 7 x 1 mL
Prenatal Testing
During pregnancy, maternal serum AFP (MSAFP) levels rise through the third trimester.
Elevated or depressed AFP levels may indicate fetal problems. Elevated MSAFP levels during
the second trimester of pregnancy are often associated with one of the most common types of
birth defects, open neural tube defects (NTDs).6-8 A number of studies9-13 have confirmed the
utility of AFP testing to detect NTDs during the second trimester of pregnancy. In addition to
AFP testing, maternal factors such as race, weight, age, diabetes, and family history must be
considered when assessing the open NTD risk.14,15 Final determination of open NTD depends
on information provided by confirmatory testing since conditions other than open NTDs, such
as cirrhosis, hepatitis, certain types of cancer, and other fetal malformations (ventral wall
defects,16 defective kidneys,17 and others), may also cause elevated MSAFP levels.14,15
Such testing includes amniotic fluid AFP (AFAFP), acetylcholinesterase, amniography,
and ultrasonography. Depressed MSAFP levels have been reported in other conditions.
Cancer Management
Interest in AFP as a tumor marker originated with a report by Abelev in 1963.18 Tatarinov
provided the first evidence linking elevated serum AFP concentrations to primary cancer of
the liver.19 Since then, investigators have demonstrated elevated serum AFP levels in
hepatocellular cancer,20-22 malignant germ cell tumors of the ovary and testis,23,24 and
teratocarcinoma of the testis.25 Although at a very low rate of incidence, increased circulating
AFP concentrations may also occur in serum specimens from patients with gastrointestinal,
pancreatic, and pulmonary cancers.26
The most important application of AFP testing in cancer management is for testicular cancer.
Although not present in pure seminoma,27 elevated serum AFP is closely associated with
non-seminomatous testicular cancer.28-30 The measurement of AFP in serum, in conjunction
with serum hCG, is an established regimen for monitoring patients with non-seminomatous
testicular cancer.31-34 In addition, monitoring the rate of AFP clearance from serum after
treatment is an indicator of the effectiveness of therapy.35,36 Conversely, the growth rate
of progressive cancer can be monitored by serially measuring serum AFP concentration
over time.37
Serial serum AFP testing is a useful adjunctive test for managing non-seminomatous
testicular cancer.
Assay Principle
The ADVIA Centaur AFP assay is a two-site sandwich immunoassay using direct
chemiluminometric technology, which uses constant amounts of two antibodies. The first
antibody, in the Lite Reagent, is an affinity purified polyclonal rabbit anti-AFP antibody
labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse
anti-AFP antibody covalently coupled to paramagnetic particles.
The system automatically performs the following steps:
• dispenses 10 μL of sample into a cuvette
• dispenses 50 μL of Lite Reagent and 250 μL of Solid Phase and incubates for
7.5 minutes at 37°C
• separates, aspirates, and washes the cuvettes with reagent water38
• dispenses 300 μL each of Acid Reagent and Base Reagent to initiate the
chemiluminescent reaction
• reports results according to the selected option, as described in the system operating
instructions or in the online help system
A direct relationship exists between the amount of AFP present in the patient sample and the
amount of relative light units (RLUs) detected by the system.
Reagents
Store the reagents upright at 2° to 8°C.
Mix all primary reagent packs by hand before loading them onto the system. Visually inspect the bottom of the
reagent pack to ensure that all particles are dispersed and resuspended. For detailed information about
preparing the reagents for use, see the system operator’s guide.
CAUTION: This device contains material of animal origin and should be handled as a potential
carrier and transmitter of disease.
For In Vitro Diagnostic Use.
Loading Reagents
Ensure that the system has sufficient primary and ancillary reagent packs. For detailed
information about preparing the system, refer to the system operating instructions or to the
online help system.
Mix all primary reagent packs by hand before loading them onto the system. Visually inspect
the bottom of the reagent pack to ensure that all particles are dispersed and resuspended. For
detailed information about preparing the reagents for use, see the system operator’s guide.
Load the ReadyPack reagent packs in the primary reagent area using the arrows as a placement
guide. The system automatically mixes the primary reagent packs to maintain homogeneous
suspension of the reagents. For detailed information about loading reagents, refer to the
system operating instructions or to the online help system.
If automatic dilution of a sample is required, load ADVIA Centaur Multi-Diluent 2 in the
ancillary reagent entry.
Quality Control
Follow government regulations or accreditation requirements for quality control frequency.
For detailed information about entering quality control values, refer to the system operating
instructions or to the online help system.
To monitor system performance and chart trends, as a minimum requirement, two levels of
quality control material should be assayed on each day that samples are analyzed. Quality
control samples should also be assayed when performing a two-point calibration. Treat all
quality control samples the same as patient samples.
Siemens Healthcare Diagnostics recommends the use of commercially available quality
control materials with at least 2 levels (low and high). A satisfactory level of performance is
achieved when the analyte values obtained are within the Acceptable Control Range for the
system or within your range, as determined by an appropriate internal laboratory quality
control scheme.
If the quality control results do not fall within the Expected Values or within the laboratory’s
established values, do not report results. Take the following actions:
• Verify that the materials are not expired.
• Verify that required maintenance was performed.
• Verify that the assay was performed according to the instructions for use.
• Rerun the assay with fresh quality control samples.
• If necessary, contact your local technical support provider or distributor for assistance.
Sample Volume
This assay requires 10 μL of sample for a single determination. This volume does not include
the unusable volume in the sample container or the additional volume required when
performing duplicates or other tests on the same sample. For detailed information about
determining the minimum required volume, refer to Sample Volume Requirements in the
ADVIA Centaur Reference Manual.
NOTE: The sample volume required to perform onboard dilution differs from the sample
volume required to perform a single determination. Refer to the following information for the
sample volume required to perform onboard dilutions:
Dilution Sample Volume (µL)
1:10, 1:20, 1:100, 1:200 20
Assay Procedure
For detailed procedural information, refer to the system operating instructions or to the online
help system.
Procedural Notes
Calculations
For detailed information about how the system calculates results, refer to the system operating
instructions or to the online help system.
The system reports AFP results in ng/mL (common units) or IU/mL (SI units), depending
on the units defined when setting up the assay. The conversion formula is
1 ng/mL = 0.83 IU/mL.43
Based on a molecular weight of 70,000 daltons, 1 ng = 0.0143 nmol.
Dilutions
• Serum samples and amniotic fluid samples with AFP levels greater than 1000 ng/mL
(830 IU/mL) must be diluted and retested to obtain accurate results.
• For an automatic or manual dilution, the final AFP concentration in the diluted sample
must be ≥ 15 ng/mL (12.5 IU/mL).
• Patient samples can be automatically diluted by the system or prepared manually.
• For automatic dilutions, ensure that ADVIA Centaur Multi-Diluent 2 is loaded and set the
system parameters as follows:
Dilution point: ≤ 1000 ng/mL (830 IU/mL)
Dilution factor, serum samples: 10, 20, 100, 200
Dilution factor, amniotic fluid samples: 100
For detailed information about automatic dilutions, refer to the system operating
instructions or to the online help system.
• Manually dilute the patient samples when patient results exceed the linearity of the assay
using automatic dilution, or when laboratory protocol requires manual dilution.
• Use Multi-Diluent 2 to manually dilute patient samples, and then load the diluted sample
in the sample rack, replacing the undiluted sample.
• Ensure that results are mathematically corrected for dilution. If a dilution factor is entered
when scheduling the test, the system automatically calculates the result.
Disposal
Dispose of hazardous and biologically contaminated materials according to the practices of
your institution. Discard all materials in a safe and acceptable manner and in compliance with
all federal, state, and local requirements.
Expected Results
AFP Values in Benign and Malignant Disease
The expected results for the ACS:180® AFP assay were previously established. Data was
obtained on 1858 serum samples as shown in the following table:
Distribution of AFP (ng/mL)
Sample Category N 0−8.0 8.1−20.0 20.1−500.0 500.1−1000.0 > 1000.0
Apparently Healthy Subjects 793 780 12 1 0 0
males 397 389 7 1 0 0
females 396 391 5 0 0 0
Malignant Diseases 717 513 64 88 11 41
Testicular Cancer
seminoma 41 37 3 1 0 0
non-seminoma 204 105 19 56 5 19
Liver Cancer
primary 80 29 11 20 4 16
secondary 93 79 8 5 0 1
Other Cancer
gastrointestinal 64 54 8 2 0 0
genitourinary 40 37 3 0 0 0
ovarian 78 73 5 0 0 0
pancreatic 18 16 1 1 0 0
other 99 83 6 3 2 5
Benign Diseases 348 316 18 8 1 5
cirrhosis 60 48 4 2 1 5
hepatitis 64 51 8 5 0 0
other 224 217 6 1 0 0
In this study, 98.4% of the apparently healthy subjects had AFP values less than 8.1 ng/mL.
These results were confirmed for the ADVIA Centaur AFP assay by analyzing serum samples
in the range of 1.3 to 943.6 ng/mL (1.1 to 783.2 IU/mL). Refer to Method Comparison.
Method Comparison
For 498 serum samples in the range of 1.3 to 943.6 ng/mL (1.1 to 783.2 IU/mL), the
relationship of the ADVIA Centaur AFP assay to the ACS:180 AFP assay is described
by the following equation:
ADVIA Centaur AFP = 1.05 (ACS:180 AFP) – 0.3 ng/mL
Correlation coefficient (r) = 0.99
For 355 amniotic fluid samples in the range of 0.3 to 86.9 μg/mL, the relationship of the
ADVIA Centaur AFP assay to to the ACS:180 AFP assay is described by the following
equation:
ADVIA Centaur AFP = 0.94 (ACS:180 AFP) + 0.1 μg/mL
Correlation coefficient (r) = 0.99
Dilution Recovery
Six serum samples in the range of 336.3 to 594.7 ng/mL (279.1 to 493.6 IU/mL) were diluted
1:2, 1:4, 1:8, 1:16, and 1:32 with Multi-Diluent 2 and assayed for recovery and parallelism.
The recoveries ranged from 78.1% to 117.2% with a mean of 100.5%.
Observed Expected Observed Expected
Sample Dilution (ng/mL) (ng/mL) (IU/mL) (IU/mL) Recovery %
1 — 594.7 493.6
1:2 286.6 297.4 237.9 246.8 96.4
1:4 147.7 148.7 122.6 123.4 99.3
1:8 78.3 74.3 65.0 61.7 105.4
1:16 39.3 37.2 32.6 30.9 105.6
1:32 20.0 18.6 16.6 15.4 107.5
Mean 102.8
2 — 472.1 391.9
1:2 239.0 236.1 198.4 195.9 101.2
1:4 123.9 118.0 102.8 98.0 105.0
1:8 63.7 59.0 52.9 49.0 108.0
1:16 33.1 29.5 27.5 24.5 112.2
1:32 16.9 14.8 14.1 12.2 114.2
Mean 108.1
3 — 405.3 336.4
1:2 203.2 202.7 168.7 168.2 100.2
1:4 103.3 101.3 85.7 84.1 102.0
1:8 53.3 50.7 44.2 42.1 105.0
1:16 28.1 25.3 23.3 21.0 111.1
1:32 14.5 12.7 12.0 10.5 114.2
Mean 106.5
4 — 388.1 322.1
1:2 191.4 194.1 158.8 161.1 98.6
1:4 99.2 97.0 82.3 80.5 102.3
1:8 52.0 48.5 43.2 40.3 107.2
1:16 27.9 24.3 23.1 20.1 114.8
1:32 14.2 12.1 11.8 10.1 117.4
Mean 108.1
5 — 346.3 287.4
1:2 161.0 173.2 133.7 143.7 93.0
1:4 78.1 86.6 64.8 71.9 90.2
1:8 38.7 43.3 32.1 35.9 89.4
1:16 20.7 21.6 17.2 18.0 95.8
1:32 11.1 10.8 9.2 9.0 102.8
Mean 94.2
6 — 336.3 279.2
1:2 146.5 168.2 121.6 139.6 87.1
1:4 67.8 84.1 56.3 69.8 80.6
1:8 32.8 42.0 27.2 34.9 78.1
1:16 17.3 21.0 14.4 17.4 82.4
1:32 9.1 10.5 7.5 8.7 86.7
Mean 83.0
Mean 100.5
Spiking Recovery
Known amounts of AFP ranging from 21.2 to 342.4 ng/mL (17.6 to 284.2 IU/mL) were added
to five patient samples with endogenous AFP levels between 35.6 and 47.4 ng/mL (29.6 to
39.3 IU/mL). When compared to expected values, the measured (recovered) levels of AFP
averaged 99.0% with a range of 91.9% to 109.2%.
Amount Added Observed Amount Added Observed
Sample (ng/mL) (ng/mL) (IU/mL) (IU/mL) Recovery %
1 — 35.6 — 29.5
21.2 56.4 17.6 46.8 98.1
89.7 118.3 74.5 98.2 92.2
155.1 197.6 128.7 164.0 104.4
342.4 387.7 284.2 321.8 102.8
Mean 99.4
2 — 39.2 — 32.5
21.2 58.8 17.6 48.8 92.5
89.7 122.0 74.5 101.3 92.3
155.1 199.6 128.7 165.7 103.4
342.4 377.5 284.2 313.3 98.8
Mean 96.7
3 — 39.9 — 33.1
21.2 61.4 17.6 51.0 101.4
89.7 125.0 74.5 103.8 94.9
155.1 209.3 128.7 173.7 109.2
342.4 386.1 284.2 320.5 101.1
Mean 101.7
4 — 47.4 — 39.3
21.2 69.3 17.6 57.5 103.3
89.7 136.3 74.5 113.1 99.1
155.1 205.1 128.7 170.2 101.7
342.4 387.8 284.2 321.9 99.4
Mean 100.9
5 — 41.5 — 34.4
21.2 62.2 17.6 51.6 97.6
89.7 133.3 74.5 110.6 102.3
155.1 185.1 128.7 153.6 92.6
342.4 356.2 284.2 295.6 91.9
Mean 96.1
Mean 99.0
Precision
Seven samples were assayed 3 times, in ≥ 4 runs, on ≥ 2 systems, (n = 113 for each sample),
over a period of 2 days. The following results were obtained:
Mean (ng/mL) Mean (IU/mL) Within-run % CV Run-to-run % CV Total % CV
16.5 13.7 3.6 4.4 5.7
21.9 18.2 4.0 3.8 6.1
37.3 31.0 3.3 3.9 5.7
67.1 55.7 2.7 3.4 4.9
173.9 144.3 2.8 3.7 5.0
499.9 414.9 3.1 2.0 5.0
732.0 607.6 3.4 2.0 5.5
Evaluating Results
The following is recommended when you observe poor reproducibility of AFP values at low
levels or if you are not satisfied with assay performance:
• Ensure that the assay reagent and calibrator lot numbers and expiration dates match those
entered in the system.
• Ensure that the calibrators, quality control materials, and assay reagents were prepared
according to the recommended procedures.
• Ensure that the recommended sample collection and handling procedures were followed.
• Ensure that the recommended system cleaning procedures were followed.
• Ensure that Type II reagent water was used when operating the system.38
• Visually check the probe and tubing for obstructions, leaks, and deformities such as
pinched or crimped tubing.
• Take further corrective action following established laboratory procedures.
• Calibrate the system using new assay reagents, calibrators, and quality control samples.
• Contact Siemens for technical assistance.
Technical Assistance
For customer support, please contact your local technical support provider or distributor.
www.siemens.com/diagnostics
References
1. Ruoslahti E, Seppala M. Studies of carcinofetal proteins: physical and chemical properties of human α-
fetoprotein. Int J Cancer 1971;7:218–25.
2. Bergstrand CG, Czar B. Demonstration of a new protein fraction in serum from the human fetus. Scand J Clin
& Lab Invest 1956;8:174.
3. Ruoslahti E, Engvall E, Kessler MJ. Chemical properties of alpha-fetoprotein. In: Herberman RB, McIntire
KR, editors. Immunodiagnosis of cancer. NY: Marcel Dekker, 1979. p.101–17.
4. Morinaga T, Sakai M, Wegmann T, Tamaoki T. Primary structures of human α fetoprotein and its mRNA. Proc
Nat’l Acad Sci 1983;80:4604–8.
5. Gitlin D, Perricelli A, Gitlin G. Synthesis of α-fetoprotein by liver, yolk sac and gastrointestinal tract of the
human conceptus. Cancer Res 1972;32:979–82.
6. Harris R, Jennison RF, Barson AJ, Laurence KM, Ruoslahti E, Seppala M. Comparison of amniotic-fluid and
maternal serum alpha-fetoprotein levels in the early antenatal diagnosis of spina bifida and anencephaly.
Lancet 1974;i(855):428.
7. Brock DJH, Bolton AE, Monaghan JM. Prenatal diagnosis of anencephaly through maternal serum alpha
fetoprotein measurement. Lancet 1973;ii(835):923–4.
8. Wald NJ, Brock DJH, Bonnar J. Prenatal diagnosis of spina bifida and anencephaly by maternal serum alpha-
fetoprotein measurement, a controlled study. Lancet 1974;i:765–7.
9. Maternal serum alpha-fetoprotein measurement in antenatal screening for anencephaly and spina bifida in
early pregnancy. Report of the UK collaborative study on alpha-fetoprotein in relation to neural tube defects.
Lancet 1977;i:1323–32.
10. Second report of the UK collaborative study on alpha-fetoprotein in relation to neural-tube defects. Lancet
1979;ii:652–62.
11. Fourth report of the UK collaborative study on alpha-fetoprotein in relation to neural tube defects. J Epidemiol
Community Health 1982;36:87–95.
12. Johnson AM, Palomaki GE, Haddow JE. Maternal serum α fetoprotein levels in pregnancies among black and
white women with fetal open spina bifida; a US collaborative study. J Obstet Gynecol 1990;162:328–31.
13. Brock DJH. The prenatal diagnosis of neural tube defects. Obstet Gynocol Surv 1976;31(1):32–40.
14. Knight GJ. Maternal serum α-fetoprotein screening. In: Hommes FA, editor. Techniques in diagnostic human
biochemical genetics; a laboratory manual. NY: Wiley-Liss, Inc, 1991. p.491–518.
15. Burton BK. Elevated maternal serum alpha-fetoprotein (MSAFP); interpretation and follow-up. Clin Obstet
Gynecol 1988;31(1):293–305.
16. Palomaki GE, Hill LE, Knight GJ, et al. Second trimester maternal serum alpha-fetoprotein levels in
pregnancies associated with gastroschisis and omphalocele. Obstet & Gynecol 1988;71(6 Pt i):906–09.
17. Seppala M, Rapola J, Huttunen NP, et al. Congenital nephrotic syndrome: prenatal diagnosis and genetic
counseling by estimation of amniotic fluid and maternal serum alpha-fetoprotein. Lancet 1976;ii:123–24.
18. Abelev GI. Study of the antigenic structure of tumors. Acta Unio Internationalis Contra Cancrum 1963;
19:80–2.
19. Tatarinov YS. Detection of an embryo-specific alpha-globulin in the blood sera of patients with primary liver
tumor. Vopr Med Khim 1964;10:90–1.
20. Abelev GI. Alpha-fetoprotein in ontogenesis and its association with malignant tumors. Adv Cancer Res
1971;14:295–358.
21. Chen Ding-Shinn, Sung, Juei-Low. Serum alpha fetoprotein in hepatocellular carcinoma. Cancer
1977;40:779–83.
22. McIntire KR, Vogel CL, Princeler GL, et al. Serum α-fetoprotein as a biochemical marker for hepatocellular
carcinoma. Cancer Res 1972;32:1941–6.
23. Kawai M, Furuhashi Y, Kano T, et al. α fetoprotein in malignant germ cell tumors of the ovary. Gynecol Oncol
1990;39:160–6.
24. Javadpour N. Serum and cellular biologic tumor markers in patients with urologic cancer. Hum Pathol
1979;10(5):557–68.
25. Masopust J, et al. Occurrence of fetoprotein in patients with neoplasms and non-neoplastic diseases. Int J
Cancer 1968;3:364–73.
26. Waldmann TA, McIntire KR. The use of a radio-immunoassay for alpha-fetoprotein in the diagnosis of
malignancy. Cancer 1974;34(4 Sup):1510–5.
27. Javadpour N, McIntire KR, Waldmann TA. Human chorionic gonadotropin and alpha-fetoprotein in sera and
tumor cells of patients with testicular seminoma. Cancer 1978;42:2768–72.
28. Lange PH, McIntire KR, Waldmann TA, et al. Serum alpha-fetoprotein and human chorionic gonadotropin in
the diagnosis and management of nonseminomatous germ-cell testicular cancer. N Engl J Med
1976;295(22):1237–40.
29. Javadpour N, McIntire KR, Waldmann TA, et al. The role of the radioimmunoassay of serum alpha-fetoprotein
and human chorionic gonadotropin in the intensive chemotherapy and surgery of metastatic testicular tumors.
J Urol 1978;119:759–62.
30. Kohn J, Orr AH, McElwain TJ, et al. Serum alpha1-fetoprotein in patients with testicular tumors. Lancet
1976;ii:433–6.
31. Perlin E, Engeler JE, Edson M, et al. The value of serial measurement of both human chorionic gonadotropin
and alpha-fetoprotein for monitoring germinal cell tumors. Cancer 1976;37:215–9.
32. Scardino PT, Cox HD, Waldmann TA, et al. The value of serum tumor markers in the staging and prognosis of
germ cell tumors of the testis. J Urol 1977;118:994–9.
33. Javadpour N. The role of biologic tumor markers in testicular cancer. Cancer 1980;45:1755–61.
34. Mason MD. Tumour markers. In:Horwich A, editor. Testicular cancer investigation and management.
Baltimore: Williams and Wilkins, 1991. p.35–50.
35. Toner GC, Geller NL, Tam C, et al. Serum tumor marker half-life during chemotherapy allows early prediction
of complete response and survival in non-seminomatous germ cell tumors. Proc Am Soc Clin Oncol
1990;9:133.
36. Kirkpatrick AM, Kirkpatrick KA. Clearance-corrected differencing and other analytic techniques useful in the
interpretation of serum AFP values. In: Kirkpatrick AM, et al, editors. Alpha-fetoprotein: laboratory
procedures and clinical applications. NY: Masson, 1981. p.135–48.
37. Price P, Hogan SJ, Horwich A. The growth rate of metastatic non-seminomatous germ cell testicular tumors
measured by marker production doubling time-I. Theoretical basis and practical application. Eur J Cancer
1990;26(4):450–3.
ADVIA Centaur, ReadyPack, and ACS:180 are trademarks of Siemens Healthcare Diagnostics.
© 2008 Siemens Healthcare Diagnostics Inc. All rights reserved.
US Pats 4,745,181; 4,918,192; 5,110,932; 5,609,822; 5,788,928