CALIBRATION REPORT

CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)


TAG. NO. : PHT – DTI – 01 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1426
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403035
LEAST COUNT : 1 °C
LOCATION : Pre - Heater Oil Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 101 1 101 1
4 200 200 201 1 201 1
5 300 300 302 2 302 2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – DTI – 02 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1427
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403038
LEAST COUNT : 1 °C
LOCATION : ATFE Oil Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – DTI – 03 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1428
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403037
LEAST COUNT : 1 °C
LOCATION : ATFE Oil Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
`` IDENTIFICATION RANGE CAL.VALID UPTO
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – DTI – 04 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1429
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403036
LEAST COUNT : 1 °C
LOCATION : Cooling Water Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 298 -2 298 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20
CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – DTI – 05 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1430
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : 180301
LEAST COUNT : 1 °C
LOCATION : Cooling Water Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20
CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTI – 06 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1431
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : IBR Boiler - Steam Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.2 0.2 50.2 0.2
3 100.0 100.0 100.4 0.4 100.4 0.4
4 150.0 150.0 150.5 0.5 150.5 0.5
5 200.0 200.0 199.9 -0.1 199.9 -0.1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTI – 07 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1432
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : IBR Boiler - Water Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 99.7 -0.3 99.7 -0.3
4 150.0 150.0 150.1 0.1 150.1 0.1
5 200.0 200.0 199.8 -0.2 199.8 -0.2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTI – 11 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1433
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : IBR Boiler - Stack Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 08 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1434
TYPE : Digital Temp. Controller
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Boiler 02 Hytherm Return Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 09 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1435
TYPE : Digital Temp. Controller
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Boiler 02 Hytherm Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 300 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 10 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1436
TYPE : Digital Temp. Controller
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Boiler 02 Hytherm Fuel Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 49 -1 49 -1
3 100 100 99 -1 99 -1
4 200 200 199 -1 199 -1
5 300 300 298 -2 298 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – DTI – 12 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1437
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – DTI – 13 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1438
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – DTI – 14 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1439
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – DTI – 15 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1440
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – DTI – 16 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1441
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 302 2 302 2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – DTI – 17 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1442
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 302 2 302 2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 18 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1443
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 99 -1 99 -1
4 200 200 200 0 200 0
5 300 300 300 0 300 0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 19 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1444
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 99 -1 99 -1
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 20 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1445
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – DTI – 21 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1446
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : DR 4 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 99.7 -0.3 99.7 -0.3
4 150.0 150.0 149.6 -0.4 149.6 -0.4
5 200.0 200.0 199.4 -0.6 199.4 -0.6
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – DTI – 22 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1447
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 4 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – DTI – 23 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1448
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20
CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : NT – DTI – 24 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1449
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 5 Nitril Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20
CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : NT – DTI – 25 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1450
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Nitril Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 26 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1451
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 1 Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 27 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1452
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 1 Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 49 -1 49 -1
3 100 100 99 -1 99 -1
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 28 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1453
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 2 Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 101 1 101 1
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 29 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1454
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 2 Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 30 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1455
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 50.2 0.2 50.2 0.2
3 100.0 100.0 100.1 0.1 100.1 0.1
4 200.0 200.0 199.9 -0.1 199.9 -0.1
5 300.0 300.0 299.5 -0.5 299.5 -0.5
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 31 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1456
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 99.9 -0.1 99.9 -0.1
4 200.0 200.0 200.3 0.3 200.3 0.3
5 300.0 300.0 300.8 0.8 300.8 0.8
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 32 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1457
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 100.5 0.5 100.5 0.5
4 200.0 200.0 200.5 0.5 200.5 0.5
5 300.0 300.0 301.0 1.0 301.0 1.0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 33 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1458
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.1 0.1 100.1 0.1
4 200.0 200.0 199.8 -0.2 199.8 -0.2
5 300.0 300.0 299.7 -0.3 299.7 -0.3
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 34 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1459
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 99.8 -0.2 99.8 -0.2
4 200.0 200.0 199.6 -0.4 199.6 -0.4
5 300.0 300.0 299.4 -0.6 299.4 -0.6
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 35 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1460
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.1 0.1 100.1 0.1
4 200.0 200.0 200.4 0.4 200.4 0.4
5 300.0 300.0 300.7 0.7 300.7 0.7
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – DTC – 36 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1461
TYPE : Digital Temp. Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 1 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 50.2 0.2 50.2 0.2
3 100.0 100.0 100.1 0.1 100.1 0.1
4 200.0 200.0 199.9 -0.1 199.9 -0.1
5 300.0 300.0 299.5 -0.5 299.5 -0.5
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – DTC – 37 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1462
TYPE : Digital Temp. Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 2 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 99.9 -0.1 99.9 -0.1
4 200.0 200.0 200.3 0.3 200.3 0.3
5 300.0 300.0 300.8 0.8 300.8 0.8
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 – DTC – 38 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1463
TYPE : Digital Temp. Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 3 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 100.5 0.5 100.5 0.5
4 200.0 200.0 200.5 0.5 200.5 0.5
5 300.0 300.0 301.0 1.0 301.0 1.0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – DTI – 40 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1464
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : Brine Tank Temp.
ACCP. CRI. : + 2 °C
RANGE : -25 to 100 °C OPERATING RANGE : amb. to - 5°C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 -25.0 -25.0 -25.0 0.0 -25.0 0.0
2 0.0 0.0 0.0 0.0 0.0 0.0
3 25.0 25.0 25.1 0.1 25.1 0.1
4 50.0 50.0 50.2 0.2 50.2 0.2
5 100.0 100.0 100.5 0.5 100.5 0.5
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – DTI – 41 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1465
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : Chilling Tank Temp.
ACCP. CRI. : + 2 °C
RANGE : -25 to 100 °C OPERATING RANGE : amb. to - 5°C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 -25.0 -25.0 -24.9 0.1 -24.9 0.1
2 0.0 0.0 0.1 0.1 0.1 0.1
3 25.0 25.0 25.1 0.1 25.1 0.1
4 50.0 50.0 50.1 0.1 50.1 0.1
5 100.0 100.0 99.9 -0.1 99.9 -0.1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 42 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1466
TYPE : Digital Temp. Indicator
MAKE & MODEL :- Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 43 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1467
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 44 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1468
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20
CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 45 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1469
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 298 -2 298 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20
CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 46 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1470
TYPE : Digital Temp. Indicator
MAKE & MODEL :- Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 47 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1471
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vent Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 48 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1472
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Cooling Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 298 -2 298 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 49 CAL. DATE : 02 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1473
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Cooling Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 21 / 04 / 2020
CC / ECL / 0112 / 19 – 20

CALIBRATION VALID UPTO : 01 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PHT – TE – 01 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1467
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Pre - Heater Oil Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.35 -0.05
3 100 138.50 138.39 -0.11
4 150 157.31 157.23 -0.08
5 200 175.84 175.78 -0.06
6 300 212.02 211.98 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – TE – 02 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1468
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : ATFE Oil Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – TE – 03 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1469
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : ATFE Oil Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.44 0.04
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.98 0.14
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – TE – 04 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1470
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Cooling Water Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – TE – 05 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1471
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Cooling Water Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.42 0.02
3 100 138.50 138.58 0.08
4 150 157.31 157.44 0.13
5 200 175.84 176.01 0.17
6 300 212.02 212.23 0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 06 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1472
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Steam Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.49 0.09
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.27 0.25
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 07 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1473
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Water Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 176.01 0.17
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 08 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1474
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Boiler 02 Hytherm Return Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.46 0.06
3 100 138.50 138.63 0.13
4 150 157.31 157.42 0.11
5 200 175.84 176.07 0.23
6 300 212.02 212.29 0.27
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 09 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1475
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Boiler 02 Hytherm Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 300 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.99 0.15
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 10 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1476
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Boiler 02 Hytherm Fuel Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.61 0.11
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.26 0.24
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 11 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1477
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Stack Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.47 0.07
3 100 138.50 138.58 0.08
4 150 157.31 157.39 0.08
5 200 175.84 175.97 0.13
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – TE – 12 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1478
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – TE – 13 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1479
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – TE – 14 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1480
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19
CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – TE – 15 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1481
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – TE – 16 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1482
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.40 0.09
5 200 175.84 175.96 0.12
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – TE – 17 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1483
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.44 0.04
3 100 138.50 138.59 0.09
4 150 157.31 157.46 0.15
5 200 175.84 176.02 0.18
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 18 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1484
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.48 0.08
3 100 138.50 138.61 0.11
4 150 157.31 157.46 0.15
5 200 175.84 176.05 0.21
6 300 212.02 212.25 0.23
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 19 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1485
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 176.04 0.20
6 300 212.02 212.16 0.14
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 20 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1486
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.45 0.05
3 100 138.50 138.60 0.10
4 150 157.31 157.41 0.10
5 200 175.84 176.05 0.21
6 300 212.02 212.26 0.24
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – TE – 21 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1487
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 4 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.47 0.07
3 100 138.50 138.55 0.05
4 150 157.31 157.41 0.10
5 200 175.84 175.98 0.14
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – TE – 22 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1488
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 4 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.45 0.05
3 100 138.50 138.64 0.14
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.27 0.25
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – TE – 23 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1489
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 4 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.47 0.07
3 100 138.50 138.58 0.08
4 150 157.31 157.39 0.08
5 200 175.84 175.97 0.13
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : NT – TE – 24 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1490
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Nitril Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
Sr.
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.38 -0.02
3 100 138.50 138.42 -0.08
4 150 157.31 157.20 -0.11
5 200 175.84 175.68 -0.16
6 300 212.02 211.84 -0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : NT – TE – 25 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1491
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Nitril Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.43 0.03
3 100 138.50 138.55 0.05
4 150 157.31 157.42 0.11
5 200 175.84 176.01 0.17
6 300 212.02 212.22 0.20
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 26 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1492
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 1 Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.42 0.11
5 200 175.84 175.98 0.14
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 27 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1493
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 1 Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.36 -0.04
3 100 138.50 138.44 -0.06
4 150 157.31 157.24 -0.07
5 200 175.84 175.75 -0.09
6 300 212.02 211.89 -0.13
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 28 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1494
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 2 Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.41 0.10
5 200 175.84 175.98 0.14
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 29 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1495
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 2 Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 30 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1496
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 31 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1497
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.42 0.11
5 200 175.84 176.00 0.16
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 32 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1498
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.47 0.07
3 100 138.50 138.60 0.10
4 150 157.31 157.45 0.14
5 200 175.84 176.01 0.17
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 33 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1499
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.73 0.06
2 50 119.40 119.48 0.08
3 100 138.50 138.61 0.11
4 150 157.31 157.46 0.15
5 200 175.84 176.03 0.19
6 300 212.02 212.22 0.20
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 34 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1500
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.43 0.03
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.99 0.15
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19
CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 35 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1501
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.42 0.11
5 200 175.84 175.95 0.11
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – TE – 36 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1502
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 1 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.62 -0.05
2 50 119.40 119.32 -0.08
3 100 138.50 138.38 -0.12
4 150 157.31 157.17 -0.14
5 200 175.84 175.67 -0.17
6 300 212.02 211.74 -0.28
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – TE – 37 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1503
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 2 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.99 0.15
6 300 212.02 212.14 0.12
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 – TE – 38 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1504
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 3 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.47 0.07
3 100 138.50 138.58 0.08
4 150 157.31 157.37 0.06
5 200 175.84 175.92 0.08
6 300 212.02 212.05 0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – TE – 40 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1505
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Brine Tank Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.31 -0.09
3 100 138.50 138.40 -0.10
4 150 157.31 157.19 -0.12
5 200 175.84 175.70 -0.14
6 300 212.02 211.81 -0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – TE – 41 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1506
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Chilling Tank Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.47 0.07
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.25 0.23
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 42 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1507
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.31 -0.09
3 100 138.50 138.40 -0.10
4 150 157.31 157.19 -0.12
5 200 175.84 175.70 -0.14
6 300 212.02 211.81 -0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 43 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1508
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.35 -0.05
3 100 138.50 138.39 -0.11
4 150 157.31 157.23 -0.08
5 200 175.84 175.78 -0.06
6 300 212.02 211.98 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 44 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1509
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 45 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1510
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 46 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1511
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 47 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1512
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vent Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.42 0.02
3 100 138.50 138.58 0.08
4 150 157.31 157.44 0.13
5 200 175.84 176.01 0.17
6 300 212.02 212.23 0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 48 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1513
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Cooling Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.49 0.09
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.27 0.25
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 49 CAL. DATE : 03 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1514
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Cooling Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 176.01 0.17
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 °C 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3806
2 Digital Multimeter 0 to 2 KW 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 02 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 03 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1515
TYPE : Pressure Gauge
MAKE & MODEL : Fiebig Searial No. : -
LEAST COUNT : 0.2 kg/cm2
LOCATION : Boiler 2 Hytherm Inlet
ACCP. CRI. : + 2 °C
RANGE : 0 to 10.6 kg/cm2 OPERATING RANGE : 0. to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.05 0.05 10.05 0.05
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 04 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1516
TYPE : Pressure Gauge
MAKE & MODEL : Mahaveer Searial No. : -
LEAST COUNT : 0.2 kg/cm2
LOCATION : Boiler 2 Hytherm Outlet
ACCP. CRI. : + 2 °C
RANGE : 0 to 10.6 kg/cm2 OPERATING RANGE : 0. to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.02 0.02 2.02 0.02
3 4.00 4.00 4.03 0.03 4.03 0.03
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.03 0.03 8.03 0.03
6 10.00 10.00 10.02 0.02 10.02 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 05 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1517
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160024
LEAST COUNT : 1 kg/cm2
LOCATION : Boiler 2 Furnace Oil Press.
ACCP. CRI. : + 2 °C
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0. to 16 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 0.01 10.01 0.01
3 20.00 20.00 20.03 0.03 20.03 0.03
4 30.00 30.00 30.02 0.02 30.02 0.02
5 40.00 40.00 40.04 0.04 40.04 0.04

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – PG – 07 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1518
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : ST - 01 - P9 - 95
LEAST COUNT : 0.1 kg/cm2
LOCATION : ATFE Feed Pump
ACCP. CRI. : + 2 °C
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0. to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.01 0.01 6.01 0.01
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.01 0.01 10.01 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ACN – PG – 09 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1519
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160046
LEAST COUNT : 0.1 kg/cm2
LOCATION : ACN Tank Pump
ACCP. CRI. .
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0. to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 7.00 7.00 7.02 0.02 7.02 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 10 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1520
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18012432
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.11 0.11 100.11 0.11
3 200.00 200.00 200.15 0.15 200.15 0.15
4 300.00 300.00 300.24 0.24 300.24 0.24
5 400.00 400.00 400.28 0.28 400.28 0.28

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 11 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1521
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18040805
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 99.97 -0.03 99.97 -0.03
3 200.00 200.00 199.95 -0.05 199.95 -0.05
4 300.00 300.00 299.56 -0.44 299.56 -0.44
5 400.00 400.00 398.48 -1.52 398.48 -1.52

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 12 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1522
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1606348788
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 0.01 10.01 0.01
3 20.00 20.00 20.02 0.02 20.02 0.02
4 40.00 40.00 40.01 0.01 40.01 0.01
5 60.00 60.00 60.03 0.03 60.03 0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 13 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1523
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18040804
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.07 0.07 100.07 0.07
3 200.00 200.00 200.11 0.11 200.11 0.11
4 300.00 300.00 300.22 0.22 300.22 0.22
5 400.00 400.00 400.27 0.27 400.27 0.27

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 14 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1524
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1511317435
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 0.01 10.01 0.01
3 20.00 20.00 20.02 0.02 20.02 0.02
4 40.00 40.00 40.04 0.04 40.04 0.04
5 60.00 60.00 60.09 0.09 60.09 0.09

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 15 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1525
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1511317432
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 0.01 10.01 0.01
3 20.00 20.00 20.02 0.02 20.02 0.02
4 40.00 40.00 40.05 0.05 40.05 0.05
5 60.00 60.00 60.04 0.04 60.04 0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 16 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1520
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 19043282
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 2
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.05 0.05 100.05 0.05
3 200.00 200.00 200.12 0.12 200.12 0.12
4 300.00 300.00 300.21 0.21 300.21 0.21
5 400.00 400.00 400.36 0.36 400.36 0.36

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 17 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1521
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18107913
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 2
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.13 0.13 100.13 0.13
3 200.00 200.00 200.34 0.34 200.34 0.34
4 300.00 300.00 300.54 0.54 300.54 0.54
5 400.00 400.00 400.87 0.87 400.87 0.87

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 18 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1522
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1901529795
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.03 0.03 10.03 0.03
3 20.00 20.00 20.05 0.05 20.05 0.05
4 40.00 40.00 40.09 0.09 40.09 0.09
5 60.00 60.00 60.15 0.15 60.15 0.15

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 19 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1523
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 19043286
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 2
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.13 0.13 100.13 0.13
3 200.00 200.00 200.34 0.34 200.34 0.34
4 300.00 300.00 300.54 0.54 300.54 0.54
5 400.00 400.00 400.87 0.87 400.87 0.87

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 20 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1524
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 17092791
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.05 -0.05 20.05 -0.05
4 40.00 40.00 40.09 -0.09 40.09 -0.09
5 60.00 60.00 60.12 -0.12 60.12 -0.12

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 21 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1525
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1801529792
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.02 -0.02 20.02 -0.02
4 40.00 40.00 40.05 -0.05 40.05 -0.05
5 60.00 60.00 60.08 -0.08 60.08 -0.08

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : AH – PG – 22 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1526
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 01322
LEAST COUNT : 0.1 kg/cm2
LOCATION : Air Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 23 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1527
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950017
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -30" to 10 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 2.00 2.00 2.00 0.00 2.00 0.00
4 4.00 4.00 4.01 -0.01 4.01 -0.01
5 8.00 8.00 8.01 -0.01 8.01 -0.01
6 10.00 10.00 10.02 -0.02 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – PG – 24 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1528
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950016
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -30" to 10 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 2.00 2.00 2.01 -0.01 2.01 -0.01
4 4.00 4.00 4.00 0.00 4.00 0.00
5 8.00 8.00 7.99 0.01 7.99 0.01
6 10.00 10.00 9.99 0.01 9.99 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – PG – 25 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1529
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 400210
LEAST COUNT : 1 kg/cm2
LOCATION : HR 1 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 -0.01 10.01 -0.01
3 20.00 20.00 20.02 -0.02 20.02 -0.02
4 30.00 30.00 30.05 -0.05 30.05 -0.05
5 40.00 40.00 40.04 -0.04 40.04 -0.04

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – PG – 26 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1530
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 400209
LEAST COUNT : 1 kg/cm2
LOCATION : HR 2 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.04 -0.04 10.04 -0.04
3 20.00 20.00 20.05 -0.05 20.05 -0.05
4 30.00 30.00 30.09 -0.09 30.09 -0.09
5 40.00 40.00 40.07 -0.07 40.07 -0.07

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 – PG – 27 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1531
TYPE : Pressure Gauge
MAKE & MODEL : H Guru Searial No. : 400209
LEAST COUNT : 0.5 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to 21 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.02 -0.02 5.02 -0.02
3 10.00 10.00 10.05 -0.05 10.05 -0.05
4 15.00 15.00 15.08 -0.08 15.08 -0.08
5 21.00 21.00 20.94 0.06 20.94 0.06

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FW – PG – 29 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1532
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 150153
LEAST COUNT : 0.1 kg/cm2
LOCATION : Fire Water Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 16 kg/cm2 OPERATING RANGE : 0 to 7 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.01 -0.01 4.01 -0.01
3 8.00 8.00 8.02 -0.02 8.02 -0.02
4 12.00 12.00 12.04 -0.04 12.04 -0.04
5 16.00 16.00 16.05 -0.05 16.05 -0.05

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : WP – PG – 30 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1533
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 05349
LEAST COUNT : 0.1 kg/cm2
LOCATION : MIDC Water Pump Pressure
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.03 -0.03 8.03 -0.03
6 10.00 10.00 10.04 -0.04 10.04 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 32 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1534
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT : 5 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 kg/cm2 OPERATING RANGE : 0 to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 49.96 0.04 49.96 0.04
3 100.00 100.00 99.94 0.06 99.94 0.06
4 150.00 150.00 149.91 0.09 149.91 0.09
5 200.00 200.00 199.88 0.12 199.88 0.12
6 250.00 250.00 249.81 0.19 249.81 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 33 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1535
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 5 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 kg/cm2 OPERATING RANGE : 0 to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.08 -0.08 50.08 -0.08
3 100.00 100.00 100.12 -0.12 100.12 -0.12
4 150.00 150.00 150.13 -0.13 150.13 -0.13
5 200.00 200.00 200.15 -0.15 200.15 -0.15
6 250.00 250.00 250.20 -0.20 250.20 -0.20
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 34 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1536
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.05 -0.05 10.05 -0.05
3 20.00 20.00 20.07 -0.07 20.07 -0.07
4 30.00 30.00 30.10 -0.10 30.10 -0.10
5 40.00 40.00 40.08 -0.08 40.08 -0.08

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 35 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1537
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 5 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 kg/cm2 OPERATING RANGE : 0 to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.03 -0.03 50.03 -0.03
3 100.00 100.00 100.07 -0.07 100.07 -0.07
4 150.00 150.00 150.12 -0.12 150.12 -0.12
5 200.00 200.00 200.14 -0.14 200.14 -0.14
6 250.00 250.00 250.17 -0.17 250.17 -0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 36 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1538
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : 100337
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 9.98 0.02 9.98 0.02
3 20.00 20.00 19.99 0.01 19.99 0.01
4 30.00 30.00 30.01 -0.01 30.01 -0.01
5 40.00 40.00 40.02 -0.02 40.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 37 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1539
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 100206
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.04 -0.04 10.04 -0.04
3 20.00 20.00 20.06 -0.06 20.06 -0.06
4 30.00 30.00 30.06 -0.06 30.06 -0.06
5 40.00 40.00 40.09 -0.09 40.09 -0.09

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 38 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1540
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 17126056
LEAST COUNT : 5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 bar OPERATING RANGE : 0 to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.04 -0.04 50.04 -0.04
3 100.00 100.00 100.09 -0.09 100.09 -0.09
4 150.00 150.00 150.14 -0.14 150.14 -0.14
5 200.00 200.00 200.24 -0.24 200.24 -0.24
6 250.00 250.00 250.33 -0.33 250.33 -0.33
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 39 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1541
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 17126059
LEAST COUNT : 5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 49.98 0.02 49.98 0.02
3 100.00 100.00 99.96 0.04 99.96 0.04
4 150.00 150.00 149.98 0.02 149.98 0.02
5 200.00 200.00 200.02 -0.02 200.02 -0.02
6 250.00 250.00 250.07 -0.07 250.07 -0.07
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 02 / 01 / 2020
2018 - 19 / CFC / MUM / 3805

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 40 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1542
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18054079
LEAST COUNT : 0.5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 25 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
Under Test
Sr. INPUT
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.01 -0.01 5.01 -0.01
3 10.00 10.00 10.02 -0.02 10.02 -0.02
4 15.00 15.00 15.02 -0.02 15.02 -0.02
5 20.00 20.00 20.04 -0.04 20.04 -0.04
6 25.00 25.00 25.03 -0.03 25.03 -0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 41 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1543
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1608372636
LEAST COUNT : 0.5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 16 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.01 -0.01 4.01 -0.01
3 8.00 8.00 8.01 -0.01 8.01 -0.01
4 12.00 12.00 12.02 -0.02 12.02 -0.02
5 16.00 16.00 16.02 -0.02 16.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – PG – 42 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1544
MAKE & MODEL FGB Searial No. : D6056
LEAST COUNT : 0.2 kg/cm2
LOCATION : CEP - CAN Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 28 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 4.98 0.02 4.98 0.02
3 10.00 10.00 9.98 0.02 9.98 0.02
4 15.00 15.00 14.96 0.04 14.96 0.04
5 20.00 20.00 19.95 0.05 19.95 0.05
6 28.00 28.00 27.94 0.06 27.94 0.06

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – PG – 43 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1545
MAKE & MODEL : Baumer Searial No. : P10 - 283
LEAST COUNT : 1 kg/cm2
LOCATION : CEP - MEOH Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.03 -0.03 20.03 -0.03
4 30.00 30.00 30.03 -0.03 30.03 -0.03
5 40.00 40.00 40.02 -0.02 40.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 44 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1546
MAKE & MODEL : Baumer Searial No. : 160007
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR2 Filter Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.03 -0.03 10.03 -0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 45 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1547
MAKE & MODEL : Spencer Searial No. : 185385
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower - Pump 1
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 5.99 0.01 5.99 0.01
5 8.00 8.00 7.98 0.02 7.98 0.02
6 10.00 10.00 9.98 0.02 9.98 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 46 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1548
MAKE & MODEL : Baumer Searial No. : 160045
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower - Pump 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.00 0.00 6.00 0.00
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 47 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1549
MAKE & MODEL : Baumer Searial No. : 03268
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR 1 - Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.02 -0.02 2.02 -0.02
3 4.00 4.00 4.03 -0.03 4.03 -0.03
4 6.00 6.00 6.04 -0.04 6.04 -0.04
5 8.00 8.00 8.05 -0.05 8.05 -0.05
6 10.00 10.00 10.05 -0.05 10.05 -0.05
*
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 49 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1550
MAKE & MODEL : itec Searial No. : 100208
LEAST COUNT : 0.5 kg/cm2
LOCATION : BA - Candle Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.05 -0.05 10.05 -0.05
3 20.00 20.00 20.05 -0.05 20.05 -0.05
4 30.00 30.00 30.08 -0.08 30.08 -0.08
5 40.00 40.00 40.11 -0.11 40.11 -0.11

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 48 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1551
MAKE & MODEL : Baumer Searial No. : 03288
LEAST COUNT : 0.1 kg/cm2
LOCATION : Big Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.03 -0.03 8.03 -0.03
6 10.00 10.00 10.04 -0.04 10.04 -0.04

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 50 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1552
MAKE & MODEL : Baumer Searial No. : 03278
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR 3 - Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 1.98 0.02 1.98 0.02
3 4.00 4.00 3.97 0.03 3.97 0.03
4 6.00 6.00 5.99 0.01 5.99 0.01
5 8.00 8.00 8.01 -0.01 8.01 -0.01
6 10.00 10.00 9.98 0.02 9.98 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 51 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1553
MAKE & MODEL : Baumer Searial No. : 01340
LEAST COUNT : 0.1 kg/cm2
LOCATION : ST - 25 Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 52 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1554
MAKE & MODEL : Baumer Searial No. : 03285
LEAST COUNT : 0.1 kg/cm2
LOCATION : ST - 25
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.02 -0.02 4.02 -0.02
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.03 -0.03 10.03 -0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – PG – 55 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1555
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00072
LEAST COUNT : 0.2 kg/cm2
LOCATION : DR5 CAN
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – PG – 56 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1556
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950015
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 5 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -30" to 10 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 2.00 2.00 2.00 0.00 2.00 0.00
4 4.00 4.00 4.01 -0.01 4.01 -0.01
5 8.00 8.00 8.01 -0.01 8.01 -0.01
6 10.00 10.00 10.02 -0.02 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – PG – 57 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1557
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00072
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 5 Vent
ACC. CRI. : + 2 % FSD
RANGE : 0 to 16 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.02 -0.02 4.02 -0.02
3 8.00 8.00 8.02 -0.02 8.02 -0.02
4 12.00 12.00 12.03 -0.03 12.03 -0.03
6 16.00 16.00 16.02 -0.02 16.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – PG – 58 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1558
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 01323
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 5 Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.01 -0.01 7.01 -0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – PG – 59 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1559
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 032824
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR 1 Steam Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 7 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.02 -0.02 2.02 -0.02
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – PG – 60 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1560
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00081
LEAST COUNT : 0.1 kg/cm2
LOCATION : Condensor Cooling Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – PG – 61 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1561
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00078
LEAST COUNT : 0.1 kg/cm2
LOCATION : Condensor Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.99 0.01 5.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – PG – 62 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1562
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00080
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 Cooling Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 0.01 5.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – PG – 63 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1563
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00073
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.00 0.00 6.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 64 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1564
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00076
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Cooling Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 65 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1565
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00075
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 66 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1566
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00255
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower Makeup Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 67 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1567
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 05379
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower Makeup Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.02 -0.02 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 68 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1568
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 05362
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Chiller Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.02 -0.02 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 69 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1569
TYPE : Pressure Gauge
MAKE & MODEL : Fiebig Searial No. : -
LEAST COUNT : 0.2 kg/cm2
LOCATION : Chilling Plant - Chiller Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10.6 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.03 -0.03 10.03 -0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 70 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1570
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00257
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Cooling Water Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 71 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1571
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 01447
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Cooling Water Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 8.00 8.00 8.01 -0.01 8.01 -0.01
6 10.00 10.00 10.01 -0.01 10.01 -0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 72 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1572
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160013
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Brine Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.03 -0.03 10.03 -0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 73 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1573
TYPE : Pressure Gauge
MAKE & MODEL : Spencer Searial No. : 185385
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Chiller Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.01 -0.01 10.01 -0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HT – PG – 74 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1574
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160024
LEAST COUNT : 0.1 kg/cm2
LOCATION : Holding Tank
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 75 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1575
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 01332
LEAST COUNT : 0.1 kg/cm2
LOCATION : BA - Candle Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 0.01 5.99 0.01
5 7.00 7.00 6.99 0.01 6.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1576
TYPE : Vaccum Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 1
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 02 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1577
TYPE : Vaccum Gauge
MAKE & MODEL : LA Searial No. : 130054
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 03 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1578
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00046
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 04 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1579
TYPE : Vaccum Gauge
MAKE & MODEL : LA Searial No. : -
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 5
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – VG – 05 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1580
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00033
LEAST COUNT :10 mmHg
LOCATION : DR 5 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – VG – 06 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1581
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00043
LEAST COUNT :10 mmHg
LOCATION : DR 1 & DR 2 Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
Under Test
Sr. INPUT
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – VG – 07 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1582
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 501678
LEAST COUNT :10 mmHg
LOCATION : DR 4 Vacuum Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – VG – 08 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1583
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 501720
LEAST COUNT :10 mmHg
LOCATION : DR 3 Vacuum Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VR – VG – 09 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1584
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00036
LEAST COUNT :10 mmHg
LOCATION : Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VR – VG – 10 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1585
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT :10 mmHg
LOCATION : Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 11 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1586
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 01942
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 12 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1587
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00037
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 13 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1588
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. :
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 14 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1589
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00038
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PS – 05 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1590
TYPE : Pressure Switch
MAKE & MODEL : Orian Searial No. : -
SEARIAL NO. :-
LOCATION : Boiler 02 Hytherm Low Press.
ACC. CRI. : + 2 % FSD
RANGE : 0 to 8 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Sr. INPUT SET SET RESET
No. Kg/cm2 Kg/cm2 Kg/cm2 Kg/cm2

·          3.00 3.00 3.06 4.58

REMARKS : Contacts Checked O. K.


a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1

CALIBRATION VALID UPTO : 03 / 04 / 2020


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PS – 06 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1591
TYPE : Pressure Switch
MAKE & MODEL : Orian Searial No. : -
SEARIAL NO. :-
LOCATION : Boiler 03 Steam Seperator
ACC. CRI. : + 2 % FSD
RANGE : 0 to 15 kg/cm2 OPERATING RANGE : 0 to 10 kg/cm2
AMB : 25 + 2 ° C
Sr. INPUT SET SET RESET
No. Kg/cm2 Kg/cm2 Kg/cm2 Kg/cm2

·          10.00 10.00 10.02 8.04

REMARKS : Contacts Checked O. K.


a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 - PT - 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1592
TYPE : Pressure Transmitter
MAKE & MODEL : Honeywell Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 1 Reactor
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
kg/cm2 mA Output Output
1 0.00 4.00 3.99 -0.01 3.99 -0.01
2 7.50 8.00 7.99 -0.01 7.99 -0.01
3 15.00 12.00 11.99 -0.01 11.99 -0.01
4 22.50 16.00 15.99 -0.01 15.99 -0.01
5 30.00 20.00 19.98 -0.02 19.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Multimeter 0 to 20 mA 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 - PT - 02 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1593
TYPE : Pressure Transmitter
MAKE & MODEL : Honeywell Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 2 Reactor
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
kg/cm2 mA Output Output
1 0.00 4.00 4.00 0.00 4.00 0.00
2 7.50 8.00 7.99 -0.01 7.99 -0.01
3 15.00 12.00 11.99 -0.01 11.99 -0.01
4 22.50 16.00 15.99 -0.01 15.99 -0.01
5 30.00 20.00 20.01 0.01 20.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Multimeter 0 to 20 mA 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 - PT - 03 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1594
TYPE : Pressure Transmitter
MAKE & MODEL : Honeywell Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
kg/cm2 mA Output Output
1 0.00 4.00 4.01 0.01 4.01 0.01
2 7.50 8.00 8.02 0.02 8.02 0.02
3 15.00 12.00 12.01 0.01 12.01 0.01
4 22.50 16.00 16.00 0.00 16.00 0.00
5 30.00 20.00 19.99 -0.01 19.99 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Multimeter 0 to 20 mA 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 - VT - 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1595
TYPE : Vacuum Transmitter
MAKE & MODEL : Baumer Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Receiver
ACC. CRI. : + 0.25 % FSD
RANGE : -1 to 1.5 bar OPERATING RANGE : -1 to 1.5 bar
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
bar mA Output Output
1 -0.800 4.00 3.99 -0.01 3.99 -0.01
2 -0.375 8.00 7.99 -0.01 7.99 -0.01
3 0.250 12.00 12.00 0.00 12.00 0.00
4 0.875 16.00 16.01 0.01 16.01 0.01
5 1.500 20.00 20.01 0.01 20.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 30 / 08 / 2020
2019 - 20 / CFC / / 1340 / 1
2 Digital Multimeter 0 to 20 mA 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LT - 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1596
TYPE : Level Transmitter
MAKE & MODEL : Honeywell ST700 Searial No. : 499162
LEAST COUNT : 0.1 mmWc
LOCATION : ACN Tank
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 5600 mmWC OPERATING RANGE : 0 to 5600 mmWC
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mmWC mA Output Output
1 0 4.00 4.00 0.00 4.00 0.00
2 1400 8.00 8.00 0.00 8.00 0.00
3 2800 12.00 12.01 0.01 12.01 0.01
4 4200 16.00 16.00 0.00 16.00 0.00
5 5600 20.00 20.01 0.01 20.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Presuure Meter +/-10000 mmWC 13 / 10 / 2019
2018 - 19 / CFC / MUM / 2670
2 Digital Multimeter 0 to 20 mA 20 / 01 / 2020
CC / ECL / 2000 / 18 – 19

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LI - 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1597
TYPE : Level Indicator
MAKE & MODEL : Newtech Searial No. : -
LEAST COUNT : 1 cm
LOCATION : ACN Tank
ACC. CRI. :+1%
RANGE : 0 to 700 cm OPERATING RANGE : 0 to 700 cm
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA cm Output Output
1 4.00 0 0 0 0 0
2 8.00 175 175 0 175 0
3 12.00 350 351 1 351 1
4 16.00 525 527 2 527 2
5 20.00 700 701 1 701 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 21 / 07 / 2020
CC / ECL / 0753 / 19 – 20

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 - PIC - 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1598
TYPE : Pressure Indicator Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 1 Reactor
ACC. CRI. :+1%
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.00 0.00 0.00 0.00 0.00
2 8.00 7.50 7.52 0.02 7.52 0.02
3 12.00 15.00 15.03 0.03 15.03 0.03
4 16.00 22.50 22.52 0.02 22.52 0.02
5 20.00 30.00 30.03 0.03 30.03 0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 21 / 07 / 2020
CC / ECL / 0753 / 19 – 20

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 - PIC - 02 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1599
TYPE : Pressure Indicator Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 2 Reactor
ACC. CRI. :+1%
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.00 0.00 0.00 0.00 0.00
2 8.00 7.50 7.49 -0.01 7.49 -0.01
3 12.00 15.00 14.99 -0.01 14.99 -0.01
4 16.00 22.50 22.49 -0.01 22.49 -0.01
5 20.00 30.00 29.99 -0.01 29.99 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 21 / 07 / 2020
CC / ECL / 0753 / 19 – 20

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 - PIC - 03 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1600
TYPE : Pressure Indicator Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. :+1%
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.00 0.00 0.00 0.00 0.00
2 8.00 7.50 7.50 0.00 7.50 0.00
3 12.00 15.00 14.99 -0.01 14.99 -0.01
4 16.00 22.50 22.49 -0.01 22.49 -0.01
5 20.00 30.00 29.98 -0.02 29.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 21 / 07 / 2020
CC / ECL / 0753 / 19 – 20

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 - VI - 01 CAL. DATE : 04 / 10 / 2019
CERT. NO. : CR / 19 – 20 / 1601
TYPE : Vacuum Indicator
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 1 mmHg
LOCATION : HR 3 Reactor
ACC. CRI. :+1%
RANGE : -760 to 1140 mmHg OPERATING RANGE : -760 to 1140 mmHg
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA mmHg Output Output
1 4.00 -760 -760 0 -760 0
2 8.00 -285 -286 -1 -286 -1
3 12.00 190 191 1 191 1
4 16.00 665 666 1 666 1
5 20.00 1140 1139 -1 1139 -1

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 21 / 07 / 2020
CC / ECL / 0753 / 19 – 20

CALIBRATION VALID UPTO : 03 / 04 / 2020

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DTI – 01 CAL. DATE : 02 / 01 / 2018
CERT. NO. : CR / 17 – 18 / 1541
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Reboiler - Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Temp.
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 300 0 300 0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 29 / 03 / 2018
CC / ECL / 2833 / 16 – 17

CALIBRATION VALID UPTO : 01 / 07 / 2018

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DTI – 02 CAL. DATE : 02 / 01 / 2018
CERT. NO. : CR / 17 – 18 / 1542
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Reboiler - Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Temp.
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 29 / 03 / 2018
CC / ECL / 2833 / 16 – 17

CALIBRATION VALID UPTO : 01 / 07 / 2018

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DTI – 03 CAL. DATE : 02 / 01 / 2018
CERT. NO. : CR / 17 – 18 / 1543
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Reboiler - Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Temp.
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 29 / 03 / 2018
CC / ECL / 2833 / 16 – 17

CALIBRATION VALID UPTO : 01 / 07 / 2018

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : Spare CAL. DATE : 02 / 01 / 2018
CERT. NO. : CR / 17 – 18 / 1544
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Spare
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Temp.
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 29 / 03 / 2018
CC / ECL / 2833 / 16 – 17

CALIBRATION VALID UPTO : 01 / 07 / 2018

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : TE – 01 CAL. DATE : 23 / 10 / 2017
CERT. NO. : CR / 17 – 18 / 1545
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Reboiler - Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Ohm
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 40 to 600 °C 30 / 12 / 2018
2017 - 18 / CFC / MUM / 862
2 Digital Multimeter 0 to 2 KW 22 / 01 / 2018
CC / ECL / 2270 / 16 – 17

CALIBRATION VALID UPTO : 22 / 04 / 2018


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : TE – 02 CAL. DATE : 23 / 10 / 2017
CERT. NO. : CR / 17 – 18 / 1546
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Reboiler - Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Ohm
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.40 0.09
5 200 175.84 175.96 0.12
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 40 to 600 °C 30 / 12 / 2018
2017 - 18 / CFC / MUM / 862
2 Digital Multimeter 0 to 2 KW 22 / 01 / 2018
CC / ECL / 2270 / 16 – 17

CALIBRATION VALID UPTO : 22 / 04 / 2018


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : TE – 03 CAL. DATE : 23 / 10 / 2017
CERT. NO. : CR / 17 – 18 / 1547
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Reboiler - Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Ohm
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.44 0.04
3 100 138.50 138.59 0.09
4 150 157.31 157.46 0.15
5 200 175.84 176.02 0.18
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 40 to 600 °C 30 / 12 / 2018
2017 - 18 / CFC / MUM / 862
2 Digital Multimeter 0 to 2 KW 22 / 01 / 2018
CC / ECL / 2270 / 16 – 17
CALIBRATION VALID UPTO : 22 / 04 / 2018

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : Spare CAL. DATE : 23 / 10 / 2017
CERT. NO. : CR / 17 – 18 / 1548
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Spare
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C INPUT: Temp. OUTPUT: Ohm
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.48 0.08
3 100 138.50 138.61 0.11
4 150 157.31 157.46 0.15
5 200 175.84 176.05 0.21
6 300 212.02 212.25 0.23
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 40 to 600 °C 30 / 12 / 2018
2017 - 18 / CFC / MUM / 862
2 Digital Multimeter 0 to 2 KW 22 / 01 / 2018
CC / ECL / 2270 / 16 – 17
CALIBRATION VALID UPTO : 22 / 04 / 2018

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DTC – 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5735
TYPE : Digital Temp. Controller
MAKE & MODEL : PPI Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : Q. C. - Hot Air Oven (QC.HO - 01)
SP. ACC. : ± 2 °C
RANGE : 0 to 200 °C INPUT: Temp. OUTPUT: Temp.
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.0 0.0 50.0 0.0
3 100.0 100.0 99.9 -0.1 99.9 -0.1
4 150.0 150.0 149.8 -0.2 149.8 -0.2
5 200.0 200.0 199.7 -0.3 199.7 -0.3
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DTC – 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5736
TYPE : Digital Temp. Controller
MAKE & MODEL : Multispan Searial No. :
LEAST COUNT : 1 °C
LOCATION : Q. C. - COD Digestor (QC. CD - 01)
SP. ACC. : ± 2 °C
RANGE : 0 to 200 °C INPUT: Temp. OUTPUT: Temp.
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 150 150 149 -1 149 -1
5 200 200 198 -2 198 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : TE – 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5737
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Q. C. - Hot Air Oven (QC.HO - 01)
SP. ACC. : ± 2 °C
RANGE : 0 to 200 °C INPUT: Temp. OUTPUT: Ohm
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.49 0.09
3 100 138.50 138.65 0.15
4 150 157.31 157.56 0.25
5 200 175.84 176.18 0.34

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 200 Ohm 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : TE – 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5738
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Q. C. - COD Digestor (QC. CD - 01)
SP. ACC. : ± 2 °C
RANGE : 0 to 200 °C INPUT: Temp. OUTPUT: Ohm
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.41 0.10
5 200 175.84 175.97 0.13

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 200 Ohm 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5739
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 10 kg/cm2
LOCATION : Q. C. - Hydrogen Cylinder
SP. ACC. : + 2 % FSD
RANGE : 0 to 280 kg/cm2 OPERATING RANGE : 0 to 180 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.11 0.11 50.11 0.11
3 100.00 100.00 100.26 0.26 100.26 0.26
4 150.00 150.00 150.39 0.39 150.39 0.39
5 200.00 200.00 200.42 0.42 200.42 0.42
6 280.00 280.00 280.57 0.57 280.57 0.57
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5740
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.2 kg/cm2
LOCATION : Q. C. - Hydrogen Cylinder
SP. ACC. : + 2 % FSD
RANGE : 0 to 16 kg/cm2 OPERATING RANGE : 0 to 10 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.02 0.02 4.02 0.02
3 8.00 8.00 8.03 0.03 8.03 0.03
4 12.00 12.00 12.05 0.05 12.05 0.05
5 16.00 16.00 16.06 0.06 16.06 0.06

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 03 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 20 – 21 / 5741
TYPE : Pressure Gauge
MAKE & MODEL : Messer Searial No. : -
LEAST COUNT : 10 bar
LOCATION : Q. C. - Nitrogen Cylinder
SP. ACC. : + 2 % FSD
RANGE : 0 to 315 bar OPERATING RANGE : 0 to 180 bar
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.08 0.08 50.08 0.08
3 100.00 100.00 100.17 0.17 100.17 0.17
4 150.00 150.00 150.18 0.18 150.18 0.18
5 200.00 200.00 200.24 0.24 200.24 0.24
6 300.00 300.00 300.39 0.39 300.39 0.39
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 04 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5742
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.5 bar
LOCATION : Q. C. - Nitrogen Cylinder
SP. ACC. : + 2 % FSD
RANGE : 0 to 16 bar OPERATING RANGE : 0 to 10 bar
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.00 0.00 4.00 0.00
3 8.00 8.00 7.99 -0.01 7.99 -0.01
4 12.00 12.00 11.98 -0.02 11.98 -0.02
5 16.00 16.00 15.97 -0.03 15.97 -0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 05 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5743
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : -
LEAST COUNT : 10 kg/cm2
LOCATION : Q. C. - Air Cylinder
SP. ACC. : + 2 % FSD
RANGE : 0 to 280 kg/cm2 OPERATING RANGE : 0 to 180 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 49.96 -0.04 49.96 -0.04
3 100.00 100.00 99.92 -0.08 99.92 -0.08
4 150.00 150.00 149.90 -0.10 149.90 -0.10
5 200.00 200.00 199.85 -0.15 199.85 -0.15
6 280.00 280.00 279.81 -0.19 279.81 -0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 06 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5744
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : -
LEAST COUNT : 0.2 kg/cm2
LOCATION : Q. C. - Air Cylinder
SP. ACC. : + 2 % FSD
RANGE : 0 to 16 kg/cm2 OPERATING RANGE : 0 to 10 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 3.99 -0.01 3.99 -0.01
3 8.00 8.00 7.98 -0.02 7.98 -0.02
4 12.00 12.00 11.97 -0.03 11.97 -0.03
5 16.00 16.00 15.95 -0.05 15.95 -0.05

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 07 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5745
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Hydrogen Gas Purifier
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.04 0.04 10.04 0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 08 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5746
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Hydrogen Gas Purifier
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 8.00 8.00 8.01 0.01 8.01 0.01
6 10.00 10.00 10.02 0.02 10.02 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 09 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5747
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Nitrogen Gas Purifier
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 -0.01 5.99 -0.01
5 8.00 8.00 7.99 -0.01 7.99 -0.01
6 10.00 10.00 9.98 -0.02 9.98 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 10 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5748
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Nitrogen Gas Purifier
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.04 0.04 8.04 0.04
6 10.00 10.00 10.03 0.03 10.03 0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 11 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5749
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Air Gas Purifier
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.01 0.01 10.01 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 12 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5750
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Air Gas Purifier
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 -0.01 3.99 -0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.01 0.01 8.01 0.01
6 10.00 10.00 10.01 0.01 10.01 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 13 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5751
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Hydrogen Gas Distribution Pannel
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.00 0.00 6.00 0.00
5 8.00 8.00 7.99 -0.01 7.99 -0.01
6 10.00 10.00 9.97 -0.03 9.97 -0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PG – 14 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5752
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : Q. C. - Nitrogen Gas Distribution Pannel
SP. ACC. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.00 0.00 6.00 0.00
5 8.00 8.00 10.01 2.01 10.01 2.01
6 10.00 10.00 10.02 0.02 10.02 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) ( Indo Amines Ltd)


Ms. INDO AMINES LTD, DOMBIVLI.(A 39/40)
LIST OF CALIBRATION instruments

Digital Temp. Indicator / Controller


Sr. No. TAG No. Location Range Cal. Date Due Date
1 PHT – DTI – 01 Pre - Heater Oil Inlet Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
2 ATFE – DTI – 02 ATFE Oil Outlet Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
3 ATFE – DTI – 03 ATFE Oil Inlet Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
4 CW – DTI – 04 Cooling Water Inlet Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
5 CW – DTI – 05 Cooling Water Outlet Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
6 BO – DTI – 06 IBR Boiler - Steam Temp. 0 to 200 °C 11 / 01 / 2022 10 / 07 / 2022
7 BO – DTI – 07 IBR Boiler - Water Temp. 0 to 200 °C 11 / 01 / 2022 10 / 07 / 2022
8 BO – DTC – 08 Boiler 02 Hytherm Return Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
9 BO – DTC – 09 Boiler 02 Hytherm Outlet Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
10 BO – DTC – 10 Boiler 02 Hytherm Fuel Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
11 BO – DTI – 11 IBR Boiler - Stack Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
12 DR 1 – DTI – 12 DR 1 Reactor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
13 DR 1 – DTI – 13 DR 1 Vapour Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
14 DR 1 – DTI – 14 DR 1 Reflux Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
15 DR 2 – DTI – 15 DR 2 Reactor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
16 DR 2 – DTI – 16 DR 2 Vapour Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
17 DR 2 – DTI – 17 DR 2 Reflux Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
18 DR 3 – DTI – 18 DR 3 Reactor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
19 DR 3 – DTI – 19 DR 3 Vapour Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
20 DR 3 – DTI – 20 DR 3 Reflux Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
21 DR 4 – DTI – 21 DR 4 Reactor Temp. 0 to 200 °C 11 / 01 / 2022 10 / 07 / 2022
22 DR 4 – DTI – 22 DR 4 Vapour Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
23 DR 4 – DTI – 23 DR 4 Reflux Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
24 DR 3 – DTI – 24 DR 3 Reactor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
25 DR 5 – DTI – 25 DR 5 Reactor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
26 CC – DTI – 26 Condensor Cooling DR 1 Inlet 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
27 CC – DTI – 27 Condensor Cooling DR 1 Outlet 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
28 CC – DTI – 28 Condensor Cooling DR 2 Inlet 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
29 CC – DTI – 29 Condensor Cooling DR 2 Outlet 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
30 DR 5 – DTI – 30 DR - 05 CEP 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
31 DR 5 – DTI – 31 DR - 05 CEP 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
32 DR 5 – DTI – 32 DR - 05 CEP 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
33 DR 5 – DTI – 33 DR - 05 CEP 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
34 DR 5 – DTI – 34 DR - 05 CEP 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
35 DR 5 – DTI – 35 DR - 05 CEP 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
36 HR 1 – DTC – 36 HR - 01 Reactor 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
37 HR 2 – DTC – 37 HR - 02 Reactor 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
38 HR 3 – DTC – 38 HR - 03 Reactor 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
39 VC – DTI – 39 Vapour Condensor Cooling Temp0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
40 BT – DTI – 40 Brine Tank Inlet -25 to 100 °C 11 / 01 / 2022 10 / 07 / 2022
41 BT – DTI – 41 Chilling Tank Inlet -25 to 100 °C 11 / 01 / 2022 10 / 07 / 2022
42 DR 5 – DTI – 42 DR 5 Reactor 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
43 DR 5 – DTI – 43 DR 5 Vapour 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
44 DR 5 – DTI – 44 DR 5 Vapour 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
45 DR 5 – DTI – 45 DR 5 Vapour 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
46 DR 5 – DTI – 46 DR 5 Reflux 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
47 DR 5 – DTI – 47 DR 5 Vent 0 to 200 °C 11 / 01 / 2022 10 / 07 / 2022
48 DR 5 – DTI – 48 DR 5 Cooling Inlet 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
49 DR 5 – DTI – 49 DR 5 Cooling Outlet 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
50 H 1 – DTI – 50 Column Top Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
51 H 6 – DTI – 51 Column Bottom Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
52 H 8 – DTI – 52 Preheater Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
53 CEP – DTI – 53 CEP Reactor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
54 CEP – DTI – 54 CEP Condensor Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
55 HR 2 – DTI – 55 HR - 02 Vent Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
56 BO – DTI – 56 IBR Boiler Stck Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
57 BO – DTI – 57 IBR Boiler Fuel Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
58 BO – DTI – 58 IBR Boiler Steam Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
59 HR 2 – DTI – 59 HR - 02 Vent Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022
60 HR 2 – DTI – 60 HR - 02 Vent Temp. 0 to 300 °C 11 / 01 / 2022 10 / 07 / 2022

Temp. Sensor
Sr. No. TAG No. Location Range Cal. Date Due Date
61 PHT – TE – 01 Pre - Heater Oil Inlet Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
62 ATFE – TE – 02 ATFE Oil Outlet Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
63 ATFE – TE – 03 ATFE Oil Inlet Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
64 CW – TE – 04 Cooling Water Inlet Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
65 CW – TE – 05 Cooling Water Outlet Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
66 BO – TE – 06 Boiler 01 Steam Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
67 BO – TE – 07 Boiler 01 Fuel Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
68 BO – TE – 08 Boiler 02 Hytherm Return Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
69 BO – TE – 09 Boiler 02 Hytherm Outlet Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
70 BO – TE – 10 Boiler 02 Hytherm Fuel Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
71 BO – TE – 11 Boiler 03 Steam Seperator Temp0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
72 DR 1 – TE – 12 DR 1 Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
73 DR 1 – TE – 13 DR 1 Vapour Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
74 DR 1 – TE – 14 DR 1 Reflux Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
75 DR 2 – TE – 15 DR 2 Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
76 DR 2 – TE – 16 DR 2 Vapour Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
77 DR 2 – TE – 17 DR 2 Reflux Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
78 DR 3 – TE – 18 DR 3 Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
79 DR 3 – TE – 19 DR 3 Vapour Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
80 DR 3 – TE – 20 DR 3 Reflux Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
81 DR 4 – TE – 21 DR 4 Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
82 DR 4 – TE – 22 DR 4 Vapour Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
83 DR 4 – TE – 23 DR 4 Reflux Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
84 DR 3 – TE – 24 DR 3 Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
85 DR 5 – TE – 25 DR 5 Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
86 CC – TE – 26 Condensor Cooling DR 1 Inlet 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
87 CC – TE – 27 Condensor Cooling DR 1 Outlet 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
88 CC – TE – 28 Condensor Cooling DR 2 Inlet 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
89 CC – TE – 29 Condensor Cooling DR 2 Outlet 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
90 DR 5 – TE – 30 DR - 05 CEP 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
91 DR 5 – TE – 31 DR - 05 CEP 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
92 DR 5 – TE – 32 DR - 05 CEP 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
93 DR 5 – TE – 33 DR - 05 CEP 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
94 DR 5 – TE – 34 DR - 05 CEP 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
95 DR 5 – TE – 35 DR - 05 CEP 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
96 HR 1 – TE – 36 HR - 01 Reactor 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
97 HR 2 – TE – 37 HR - 02 Reactor 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
98 HR 3 – TE – 38 HR - 03 Reactor 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
99 VP – TE – 39 Vapour Condensor Cooling Temp1 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
100 BT – TE – 40 Brine Tank Inlet -25 to 100 °C 12 / 01 / 2022 11 / 07 / 2022
101 BT – TE – 41 Chilling Tank Inlet -25 to 100 °C 12 / 01 / 2022 11 / 07 / 2022
102 DR 5 – TE – 42 DR 5 Reactor 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
103 DR 5 – TE – 43 DR 5 Vapour 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
104 DR 5 – TE – 44 DR 5 Vapour 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
105 DR 5 – TE – 45 DR 5 Vapour 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
106 DR 5 – TE – 46 DR 5 Reflux 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
107 DR 5 – TE – 47 DR 5 Vent 0 to 200 °C 12 / 01 / 2022 11 / 07 / 2022
108 DR 5 – TE – 48 DR 5 Cooling Inlet 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
109 DR 5 – TE – 49 DR 5 Cooling Outlet 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
110 HR 1 – TE – 50 Column Top Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
111 HR 6 – TE – 51 Column Bottom Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
112 HR 8 – TE – 52 Preheater Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
113 CEP – TE – 53 CEP Reactor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
114 CEP – TE – 54 CEP Condensor Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
115 HR 2 – TE – 55 HR - 02 Vent Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
116 BO – TE – 56 IBR Boiler Stck Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
117 BO – TE – 57 IBR Boiler Fuel Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
118 BO – TE – 58 IBR Boiler Steam Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
119 HR 2 – TE – 59 HR - 02 Vent Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022
120 HR 2 – TE – 60 HR - 02 Vent Temp. 0 to 300 °C 12 / 01 / 2022 11 / 07 / 2022

Pressure Gauge
Sr. No. TAG No. Location Range Cal. Date Due Date
121 BO – PG – 03 Boiler 02 Hytherm Inlet 0 to 16 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
122 BO – PG – 04 Boiler 02 Hytherm Outlet 0 to 10.6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
123 BO – PG – 05 Boiler 02 Furnace Oil 0 to 40 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
124 FL – PG – 06 Filter Pump 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
125 ATFE – PG – 07 ATFE Feed Pump 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
126 ATFE – PG – 08 ATFE Recieve Pump 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
127 EJ – PG – 09 Vacuum Ejector 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
128 PRV1 – PG – 10 PRV 1 0 to 400 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
129 PRV1 – PG – 11 PRV 1 0 to 400 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
130 PRV1 – PG – 12 PRV 1 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
131 PRV1 – PG – 13 PRV 1 0 to 400 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
132 PRV1 – PG – 14 PRV 1 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
133 PRV1 – PG – 15 PRV 1 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
134 PRV2 – PG – 16 PRV 2 0 to 400 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
135 PRV2 – PG – 17 PRV 2 0 to 400 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
136 PRV2 – PG – 18 PRV 2 0 to 28 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
137 PRV2 – PG – 19 PRV 2 0 to 400 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
138 PRV2 – PG – 20 PRV 2 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
139 PRV2 – PG – 21 PRV 2 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
140 DR 1 – PG – 23 DR 1 Reactor -30"Hg to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
141 DR 2 – PG – 24 DR 2 Reactor -30"Hg to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
142 HR 1 – PG – 25 HR 1 Reactor 0 to 42 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
143 HR 2 – PG – 26 HR 2 Reactor 0 to 42 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
144 HR 3 – PG – 27 HR 3 Reactor 0 to 21 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
145 FW – PG – 29 Fire Water Pump 0 to 40 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
146 WP – PG – 30 MIDC Water Pump 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
147 PRV 3 – PG – 32 PRV 3 0 to 250 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
148 PRV 3 – PG – 33 PRV 3 0 to 250 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
149 PRV 3 – PG – 34 PRV 3 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
150 PRV 3 – PG – 35 PRV 3 0 to 250 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
151 PRV 3 – PG – 36 PRV 3 0 to 100 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
152 PRV 3 – PG – 37 PRV 3 0 to 60 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
153 N2 PRV – PG – 38 N2 PRV 0 to 250 bar 13 / 01 / 2022 12 / 07 / 2022
154 N2 PRV – PG – 39 N2 PRV 0 to 250 bar 13 / 01 / 2022 12 / 07 / 2022
155 N2 PRV – PG – 40 N2 PRV 0 to 25 bar 13 / 01 / 2022 12 / 07 / 2022
156 N2 PRV – PG – 41 N2 PRV 0 to 16 bar 13 / 01 / 2022 12 / 07 / 2022
157 CEP – PG – 42 CEP ACN Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
158 CEP – PG – 43 CEP MEOH Pump 0 to 28 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
159 HR2 – PG – 44 HR 2 Filter Pump 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
160 CT – PG – 45 Cooling Tower Pump No. 1 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
161 CT – PG – 46 Cooling Tower Pump No. 2 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
162 FL – PG – 47 HR 1 Filter 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
163 DR 1 – PG – 48 DR 1 Steam Line 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
164 FL – PG – 49 Big Filter 0 to 42 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
165 FL – PG – 50 HR 3 Filter 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
166 ST – PG – 51 ST .Tank No. 22 Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
167 ST – PG – 52 ST .Tank No. 26 Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
168 DR 5 – PG – 55 DR 5 Circulation Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
169 DR 5 – PG – 56 DR 5 Reactor -30"Hg to 4 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
170 DR 5 – PG – 57 DR 5 Steam Header 0 to 25 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
171 DR 5 – PG – 58 DR 5 Cooling Outlet 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
172 DR 1 – PG – 59 HR 1 Steam Stack 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
173 CC – PG – 60 Condensor Cooling Inlet 0 to 6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
174 CC – PG – 61 Condensor Cooling Outlet 0 to 6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
175 DR 2 – PG – 62 DR 2 Cooling Inlet 0 to 6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
176 DR 2 – PG – 63 DR 2 Cooling Outlet 0 to 6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
177 DR 1 – PG – 64 DR 1 Cooling Inlet 0 to 6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
178 DR 1 – PG – 65 DR 1 Cooling Outlet 0 to 6 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
179 FL – PG – 66 BA Candle Filter 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
180 CT – PG – 67 Cooling Tower Header 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
181 CP – PG – 68 Chilling Plant Chiller Inlet 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
182 CP – PG – 69 Chilling Plant Chiller Outlet 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
183 CP – PG – 70 Chilling Plant CW Inlet 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
184 CP – PG – 71 Chilling Plant CW Outlet 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
185 CP – PG – 72 Chilling Plant Brine Pump 0 to 4 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
186 CP – PG – 73 Chilling Plant Chiller Pump 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
187 HT – PG – 74 Holding Tank 1 -30"Hg to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
188 HT – PG – 75 Holding Tank 2 -30"Hg to 4.2 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
189 ST – PG – 76 ST .Tank No. 21 Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
190 ST – PG – 77 ST .Tank No. 23 Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
191 ST – PG – 78 ST .Tank No. 1 Pump 0 to 7 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
192 BO – PG – 79 IBR Boiler Main Steam Header 0 to 10 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
193 BO – PG – 80 IBR Boiler Main Steam 0 to 21 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
194 BO – PG – 81 IBR Boiler F. O. 0 to 40 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
195 BO – PG – 82 IBR Boiler F. O. 0 to 40 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
196 BO – PG – 83 IBR Boiler Air Blower 0 to 1000 mmWC 13 / 01 / 2022 12 / 07 / 2022
197 BO – PG – 84 IBR Boiler Water Pump 1 0 to 25 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
198 BO – PG – 85 IBR Boiler Water Pump 2 0 to 42 kg/cm2 13 / 01 / 2022 12 / 07 / 2022

Pressure Switch
Sr. No. TAG No. Location Range Cal. Date Due Date
199 BO - PS - 05 Boiler 02 Steam High Press. 0 to 8 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
200 BO - PS - 06 Boiler 02 Steam Low Press. 0 to 8 kg/cm2 13 / 01 / 2022 12 / 07 / 2022

Vaccum Gauge
Sr. No. TAG No. Location Range Cal. Date Due Date
201 VP - VG - 01 Vacuum Pump 1 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
202 VP - VG - 02 Vacuum Pump 2 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
203 VP - VG - 03 Vacuum Pump 3 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
204 VR - VG - 04 Vacuum Reciever 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
205 DR 4 - VG - 05 DR 4 Reactor -30"Hg to 4.2 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
206 DR 1 - VG - 06 DR 1 & 2 Vacuum Receiver 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
207 DR 4 - VG - 07 DR 4 Vacuum Header 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
208 DR 3 - VG - 08 DR 3 Vacuum Header 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
209 VR - VG - 09 Vacuum Receiver 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
210 VR - VG - 10 Vacuum Receiver 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
211 DR 5 - VG - 11 DR 5 Couloum 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
212 DR 5 - VG - 12 DR 5 Circulation Pump 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
213 DR 5 - VG - 13 DR 5 Couloum 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
214 DR 5 - VG - 14 DR 5 Couloum 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022
215 DR 3 - VG - 15 DR 3 Vacuum Reactor -30"Hg to 4 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
216 DR 5 - VG - 16 DR 5 Receiver 0 to -760 mmHG 13 / 01 / 2022 12 / 07 / 2022

Level Transmitter / Indicator


Sr. No. TAG No. Location Range Cal. Date Due Date
217 ST - LT - 01 Storage Tank - 22 0 to 5600 mmWc 13 / 01 / 2022 12 / 07 / 2022
218 ST - LT - 02 Storage Tank - 16 0 to 3120 mmWc 13 / 01 / 2022 12 / 07 / 2022
219 ST - LT - 03 Storage Tank - 15 0 to 4640 mmWc 13 / 01 / 2022 12 / 07 / 2022
220 ST - LT - 04 Storage Tank - 25 0 to 9000 mmWc 13 / 01 / 2022 12 / 07 / 2022
221 ST - LI - 01 Storage Tank - 22 0 to 700 cm 13 / 01 / 2022 12 / 07 / 2022
222 ST - LI - 02 Storage Tank - 16 0 to 390 cm 13 / 01 / 2022 12 / 07 / 2022
223 ST - LI - 03 Storage Tank - 15 0 to 580 cm 13 / 01 / 2022 12 / 07 / 2022
224 ST - LI - 04 Storage Tank - 25 0 to 1125 cm 13 / 01 / 2022 12 / 07 / 2022

VACUUM Transmitter / Indicator


Sr. No. TAG No. Location Range Cal. Date Due Date
225 DR 2 - VT - 01 DR 2 Vacuum Receiver (- 1 to 1.5 bar) 13 / 01 / 2022 12 / 07 / 2022
226 DR 2 - VI - 01 DR 2 Vacuum Receiver -760 to 1140 mmHg 13 / 01 / 2022 12 / 07 / 2022

Pressure Transmitter / Indicator


Sr. No. TAG No. Location Range Cal. Date Due Date
227 HR 1 - PT - 01 HR 1 Reactor 0 to 30 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
228 HR 2 - PT - 02 HR 2 Reactor 0 to 30 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
229 HR 3 - PT - 03 HR 3 Reactor 0 to 30 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
230 BO - PT - 04 IBR Main Steam Header 0 to 15 bar 13 / 01 / 2022 12 / 07 / 2022
231 HR 1 - PI - 01 HR 1 Reactor 0 to 30 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
232 HR 2 - PI - 02 HR 2 Reactor 0 to 30 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
233 HR 3 - PI - 03 HR 3 Reactor 0 to 30 kg/cm2 13 / 01 / 2022 12 / 07 / 2022
234 BO - PIC - 04 IBR Main Steam Header 0 to 15 kg/cm2 13 / 01 / 2022 12 / 07 / 2022

Ms. INDO AMINES LTD, DOMBIVLI.(A 39/40)


R & D Lab

LIST OF CALIBRATION instruments


Sr. No. TAG No. Location Range Cal. Date Due Date
1 DTC – 01 Q. C. - Hot Air Oven (QC.HO - 010 to 200 °C 14 / 01 / 2022 13 / 07 / 2022
2 DTC – 02 Q. C. - COD Digestor (QC. CD - 0 to 200 °C 14 / 01 / 2022 13 / 07 / 2022
3 TE – 01 Q. C. - Hot Air Oven (QC.HO - 010 to 200 °C 14 / 01 / 2022 13 / 07 / 2022
4 TE – 02 Q. C. - COD Digestor (QC. CD - 1 to 200 °C 14 / 01 / 2022 13 / 07 / 2022
5 PG – 01 Q. C. - Hydrogen Cylinder 0 to 280 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
6 PG – 02 Q. C. - Hydrogen Cylinder 0 to 16 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
7 PG – 03 Q. C. - Nitrogen Cylinder 0 to 315 bar 14 / 01 / 2022 13 / 07 / 2022
8 PG – 04 Q. C. - Nitrogen Cylinder 0 to 16 bar 14 / 01 / 2022 13 / 07 / 2022
9 PG – 05 Q. C. - Air Cylinder 0 to 280 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
10 PG – 06 Q. C. - Air Cylinder 0 to 16 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
11 PG – 07 Q. C. - Hydrogen Gas Purifier 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
12 PG – 08 Q. C. - Hydrogen Gas Purifier 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
13 PG – 09 Q. C. - Nitrogen Gas Purifier 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
14 PG – 10 Q. C. - Nitrogen Gas Purifier 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
15 PG – 11 Q. C. - Air Gas Purifier 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
16 PG – 12 Q. C. - Air Gas Purifier 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
17 PG – 13 Q. C. - H2 Distribution Pannel 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
18 PG – 14 Q. C. - N2 Distribution Pannel 0 to 10 kg/cm2 14 / 01 / 2022 13 / 07 / 2022
CALIBRATION REPORT

M/s. INDO AMINES LTD

A - 39/40, DOMBIVLI
CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. (A 39/40)
TAG. NO. : PHT – DTI – 01 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5501
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403035
LEAST COUNT : 1 °C
LOCATION : Pre - Heater Oil Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – DTI – 02 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5502
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403038
LEAST COUNT : 1 °C
LOCATION : ATFE Oil Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 101 1 101 1
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – DTI – 03 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5503
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403037
LEAST COUNT : 1 °C
LOCATION : ATFE Oil Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
`` IDENTIFICATION RANGE CAL.VALID UPTO
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – DTI – 04 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5504
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : 1403036
LEAST COUNT : 1 °C
LOCATION : Cooling Water Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 99 -1 99 -1
4 200 200 199 -1 199 -1
5 300 300 298 -2 298 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – DTI – 05 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5505
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : 180301
LEAST COUNT : 1 °C
LOCATION : Cooling Water Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTI – 06 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5506
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : IBR Boiler - Steam Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTI – 07 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5507
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : IBR Boiler - Water Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.3 0.3 100.3 0.3
4 150.0 150.0 149.9 -0.1 149.9 -0.1
5 200.0 200.0 199.7 -0.3 199.7 -0.3
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTI – 11 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5508
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : IBR Boiler - Stack Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 08 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5509
TYPE : Digital Temp. Controller
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Boiler 02 Hytherm Return Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 09 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5510
TYPE : Digital Temp. Controller
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Boiler 02 Hytherm Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 300 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 10 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5511
TYPE : Digital Temp. Controller
MAKE & MODEL : itherm Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Boiler 02 Hytherm Fuel Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – DTI – 12 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5512
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – DTI – 13 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5513
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – DTI – 14 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5514
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – DTI – 15 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5515
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 300 0 300 0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – DTI – 16 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5516
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – DTI – 17 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5517
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 49 -1 49 -1
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 18 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5518
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 19 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5519
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 20 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5520
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – DTI – 21 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5521
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : DR 4 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 99.7 -0.3 99.7 -0.3
4 150.0 150.0 149.5 -0.5 149.5 -0.5
5 200.0 200.0 199.5 -0.5 199.5 -0.5
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – DTI – 22 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5522
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 4 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – DTI – 23 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5523
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – DTI – 24 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5524
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 25 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5525
TYPE : Digital Temp. Indicator
MAKE & MODEL : Kwik Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 5 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 298 -2 298 -2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 26 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5526
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 1 Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 99 -1 99 -1
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 27 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5527
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 1 Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
INPUT Desired Output
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 28 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5528
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 2 Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – DTI – 29 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5529
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Condensor Cooling DR 2 Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 30 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5530
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.8 -0.2 49.8 -0.2
3 100.0 100.0 999.0 899.0 999.0 899.0
4 200.0 200.0 200.1 0.1 200.1 0.1
5 300.0 300.0 300.2 0.2 300.2 0.2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 31 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5531
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.2 0.2 100.2 0.2
4 200.0 200.0 199.8 -0.2 199.8 -0.2
5 300.0 300.0 299.6 -0.4 299.6 -0.4
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 32 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5532
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 99.8 -0.2 99.8 -0.2
4 200.0 200.0 199.7 -0.3 199.7 -0.3
5 300.0 300.0 299.8 -0.2 299.8 -0.2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 33 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5533
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.8 -0.2 49.8 -0.2
3 100.0 100.0 99.7 -0.3 99.7 -0.3
4 200.0 200.0 199.6 -0.4 199.6 -0.4
5 300.0 300.0 299.5 -0.5 299.5 -0.5
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 34 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5534
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 -0.1 -0.1 -0.1 -0.1
2 50.0 50.0 49.8 -0.2 49.8 -0.2
3 100.0 100.0 99.9 -0.1 99.9 -0.1
4 200.0 200.0 200.2 0.2 200.2 0.2
5 300.0 300.0 300.3 0.3 300.3 0.3
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – DTI – 35 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5535
TYPE : Digital Temp. Indicator
MAKE & MODEL : Nippon Searial No. : 0918D160
LEAST COUNT : 0.1 °C
LOCATION : DR 5 CEP Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.3 0.3 100.3 0.3
4 200.0 200.0 200.4 0.4 200.4 0.4
5 300.0 300.0 300.5 0.5 300.5 0.5
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – DTC – 36 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5536
TYPE : Digital Temp. Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 1 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 50.2 0.2 50.2 0.2
3 100.0 100.0 100.4 0.4 100.4 0.4
4 200.0 200.0 200.5 0.5 200.5 0.5
5 300.0 300.0 300.4 0.4 300.4 0.4
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – DTC – 37 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5537
TYPE : Digital Temp. Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 2 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 50.2 0.2 50.2 0.2
3 100.0 100.0 100.3 0.3 100.3 0.3
4 200.0 200.0 200.5 0.5 200.5 0.5
5 300.0 300.0 300.7 0.7 300.7 0.7
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 – DTC – 38 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5538
TYPE : Digital Temp. Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 3 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.3 0.3 100.3 0.3
4 200.0 200.0 200.5 0.5 200.5 0.5
5 300.0 300.0 300.4 0.4 300.4 0.4
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – DTI – 40 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5539
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : Brine Tank Temp.
ACCP. CRI. : + 2 °C
RANGE : -25 to 100 °C OPERATING RANGE : amb. to - 5°C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 -25.0 -25.0 -25.1 -0.1 -25.1 -0.1
2 0.0 0.0 0.0 0.0 0.0 0.0
3 25.0 25.0 25.0 0.0 25.0 0.0
4 50.0 50.0 50.1 0.1 50.1 0.1
5 100.0 100.0 100.1 0.1 100.1 0.1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – DTI – 41 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5540
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : Chilling Tank Temp.
ACCP. CRI. : + 2 °C
RANGE : -25 to 100 °C OPERATING RANGE : amb. to - 5°C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 -25.0 -25.0 -24.9 0.1 -24.9 0.1
2 0.0 0.0 0.0 0.0 0.0 0.0
3 25.0 25.0 25.1 0.1 25.1 0.1
4 50.0 50.0 49.9 -0.1 49.9 -0.1
5 100.0 100.0 100.2 0.2 100.2 0.2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTC – 42 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5541
TYPE : Digital Temp. Controller
MAKE & MODEL : itherm Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 43 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5542
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 44 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5543
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 45 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5544
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 101 1 101 1
4 200 200 201 1 201 1
5 300 300 302 2 302 2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 46 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5545
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 47 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5546
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : DR5 Vent Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.0 0.0 0.0 0.0
2 50.0 50.0 49.9 -0.1 49.9 -0.1
3 100.0 100.0 99.7 -0.3 99.7 -0.3
4 150.0 150.0 150.3 0.3 150.3 0.3
5 200.0 200.0 199.9 -0.1 199.9 -0.1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 48 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5547
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Cooling Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – DTI – 49 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5548
TYPE : Digital Temp. Indicator
MAKE & MODEL : Mini Electronics Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR5 Cooling Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : H 1 – DTI – 50 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5549
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : H 1 Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : H 6 – DTI – 51 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5550
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : H 6 Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : H 8 – DTI – 52 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5551
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : H 8 Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – DTI – 53 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5552
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : CEP Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – DTI – 54 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5553
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : CEP Condensor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VC – DTI – 39 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5554
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : Vapour Condensor Cooling Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – DTI – 55 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5555
TYPE : Digital Temp. Indicator
MAKE & MODEL : Range Technology Searial No. : -
LEAST COUNT : 1 °C
LOCATION : HR 2 Vent Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 56 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5556
TYPE : Digital Temp. Controller
MAKE & MODEL : Selec Searial No. : -
LEAST COUNT : 1 °C
LOCATION : IBR Boiler Stck Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 57 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5557
TYPE : Digital Temp. Controller
MAKE & MODEL : Selec Searial No. : -
LEAST COUNT : 1 °C
LOCATION : IBR Boiler Fuel Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – DTC – 58 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5558
TYPE : Digital Temp. Controller
MAKE & MODEL : Selec Searial No. : -
LEAST COUNT : 1 °C
LOCATION : IBR Boiler Steam Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – DTI – 59 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5559
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : HR 2 Vent Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – DTI – 60 CAL. DATE : 11 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5560
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : HR 2 Vent Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 0.1 0.1 0.1 0.1
2 50.0 50.0 50.1 0.1 50.1 0.1
3 100.0 100.0 100.3 0.3 100.3 0.3
4 150.0 150.0 149.7 -0.3 149.7 -0.3
5 200.0 200.0 199.3 -0.7 199.3 -0.7
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 10 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PHT – TE – 01 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5561
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Pre - Heater Oil Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – TE – 02 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5562
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : ATFE Oil Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – TE – 03 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5563
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : ATFE Oil Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.42 0.11
5 200 175.84 176.00 0.16
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – TE – 04 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5564
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Cooling Water Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.47 0.07
3 100 138.50 138.60 0.10
4 150 157.31 157.45 0.14
5 200 175.84 176.01 0.17
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CW – TE – 05 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5565
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Cooling Water Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.73 0.06
2 50 119.40 119.48 0.08
3 100 138.50 138.61 0.11
4 150 157.31 157.46 0.15
5 200 175.84 176.03 0.19
6 300 212.02 212.22 0.20
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 06 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5566
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Steam Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.43 0.03
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.99 0.15
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 07 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5567
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Water Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
Sr.
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.42 0.11
5 200 175.84 175.95 0.11
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 08 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5568
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Boiler 02 Hytherm Return Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 250 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.62 -0.05
2 50 119.40 119.32 -0.08
3 100 138.50 138.38 -0.12
4 150 157.31 157.17 -0.14
5 200 175.84 175.67 -0.17
6 300 212.02 211.74 -0.28
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 09 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5569
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Boiler 02 Hytherm Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 300 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.99 0.15
6 300 212.02 212.14 0.12
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 10 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5570
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Boiler 02 Hytherm Fuel Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.47 0.07
3 100 138.50 138.58 0.08
4 150 157.31 157.37 0.06
5 200 175.84 175.92 0.08
6 300 212.02 212.05 0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 11 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5571
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Stack Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.31 -0.09
3 100 138.50 138.40 -0.10
4 150 157.31 157.19 -0.12
5 200 175.84 175.70 -0.14
6 300 212.02 211.81 -0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – TE – 12 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5572
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.47 0.07
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.25 0.23
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – TE – 13 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5573
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.31 -0.09
3 100 138.50 138.40 -0.10
4 150 157.31 157.19 -0.12
5 200 175.84 175.70 -0.14
6 300 212.02 211.81 -0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – TE – 14 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5574
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.35 -0.05
3 100 138.50 138.39 -0.11
4 150 157.31 157.23 -0.08
5 200 175.84 175.78 -0.06
6 300 212.02 211.98 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – TE – 15 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5575
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – TE – 16 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5576
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – TE – 17 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5577
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 18 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5578
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.42 0.02
3 100 138.50 138.58 0.08
4 150 157.31 157.44 0.13
5 200 175.84 176.01 0.17
6 300 212.02 212.23 0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 19 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5579
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.49 0.09
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.27 0.25
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 20 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5580
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 176.01 0.17
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – TE – 21 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5581
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 4 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – TE – 22 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5582
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 4 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – TE – 23 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5583
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 4 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.40 0.09
5 200 175.84 175.96 0.12
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – TE – 24 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5584
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 3 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.44 0.04
3 100 138.50 138.59 0.09
4 150 157.31 157.46 0.15
5 200 175.84 176.02 0.18
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 25 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5585
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.48 0.08
3 100 138.50 138.61 0.11
4 150 157.31 157.46 0.15
5 200 175.84 176.05 0.21
6 300 212.02 212.25 0.23
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 26 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5586
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 1 Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 176.04 0.20
6 300 212.02 212.16 0.14
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 27 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5587
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 1 Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.45 0.05
3 100 138.50 138.60 0.10
4 150 157.31 157.41 0.10
5 200 175.84 176.05 0.21
6 300 212.02 212.26 0.24
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1
CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 28 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5588
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 2 Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.35 -0.05
3 100 138.50 138.39 -0.11
4 150 157.31 157.23 -0.08
5 200 175.84 175.78 -0.06
6 300 212.02 211.98 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1
CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – TE – 29 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5589
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Condensor Cooling DR 2 Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 30 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5590
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.44 0.04
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.98 0.14
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 31 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5591
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 32 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5592
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.42 0.02
3 100 138.50 138.58 0.08
4 150 157.31 157.44 0.13
5 200 175.84 176.01 0.17
6 300 212.02 212.23 0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 33 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5593
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.49 0.09
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.27 0.25
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 34 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5594
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 176.01 0.17
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – TE – 35 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5595
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 5 CEP Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.46 0.06
3 100 138.50 138.63 0.13
4 150 157.31 157.42 0.11
5 200 175.84 176.07 0.23
6 300 212.02 212.29 0.27
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – TE – 36 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5596
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 1 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.41 0.10
5 200 175.84 175.99 0.15
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – TE – 37 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5597
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 2 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.61 0.11
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.26 0.24
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 – TE – 38 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5598
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 3 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.47 0.07
3 100 138.50 138.58 0.08
4 150 157.31 157.39 0.08
5 200 175.84 175.97 0.13
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – TE – 40 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5599
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Brine Tank Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BT – TE – 41 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5600
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Chilling Tank Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 42 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5601
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 43 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5602
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 44 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5603
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.40 0.09
5 200 175.84 175.96 0.12
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 45 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5604
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vapour Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.44 0.04
3 100 138.50 138.59 0.09
4 150 157.31 157.46 0.15
5 200 175.84 176.02 0.18
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 46 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5605
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Reflux Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.38 -0.02
3 100 138.50 138.42 -0.08
4 150 157.31 157.20 -0.11
5 200 175.84 175.68 -0.16
6 300 212.02 211.84 -0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 47 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5606
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Vent Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.43 0.03
3 100 138.50 138.55 0.05
4 150 157.31 157.42 0.11
5 200 175.84 176.01 0.17
6 300 212.02 212.22 0.20
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 48 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5607
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Cooling Inlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
Sr.
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.42 0.11
5 200 175.84 175.98 0.14
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – TE – 49 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5608
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR5 Cooling Outlet Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to -5 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.36 -0.04
3 100 138.50 138.44 -0.06
4 150 157.31 157.24 -0.07
5 200 175.84 175.75 -0.09
6 300 212.02 211.89 -0.13
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : H 1 – TE – 50 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5609
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : H 1 Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.41 0.10
5 200 175.84 175.98 0.14
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : H 6 – TE – 51 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5610
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : H 6 Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : H 8 – TE – 52 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5611
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : H 8 Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – TE – 53 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5612
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : CEP Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.42 0.11
5 200 175.84 176.00 0.16
6 300 212.02 212.17 0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – TE – 54 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5613
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : CEP Condnsor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.72 0.05
2 50 119.40 119.47 0.07
3 100 138.50 138.60 0.10
4 150 157.31 157.45 0.14
5 200 175.84 176.01 0.17
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VC – TE – 39 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5614
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : Vapour Condnsor Cooling Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.38 -0.02
3 100 138.50 138.42 -0.08
4 150 157.31 157.20 -0.11
5 200 175.84 175.68 -0.16
6 300 212.02 211.84 -0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – TE – 55 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5615
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 2 Vent Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.43 0.03
3 100 138.50 138.55 0.05
4 150 157.31 157.42 0.11
5 200 175.84 176.01 0.17
6 300 212.02 212.22 0.20
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 56 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5616
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Stack Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.42 0.11
5 200 175.84 175.98 0.14
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 57 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5617
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Fuel Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.36 -0.04
3 100 138.50 138.44 -0.06
4 150 157.31 157.24 -0.07
5 200 175.84 175.75 -0.09
6 300 212.02 211.89 -0.13
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – TE – 58 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5618
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : IBR Boiler Steam Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.46 0.06
3 100 138.50 138.58 0.08
4 150 157.31 157.41 0.10
5 200 175.84 175.98 0.14
6 300 212.02 212.19 0.17
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – TE – 59 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5619
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 2 Vent Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – TE – 60 CAL. DATE : 12 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5620
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : HR 2 Vent Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 200 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 11 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 03 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5621
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 654821
LEAST COUNT : 0.2 kg/cm2
LOCATION : Boiler 2 Hytherm Inlet
ACCP. CRI. : + 2 °C
RANGE : 0 to 16 kg/cm2 OPERATING RANGE : 0. to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.02 0.02 4.02 0.02
3 8.00 8.00 8.03 0.03 8.03 0.03
4 12.00 12.00 12.05 0.05 12.05 0.05
5 16.00 16.00 16.07 0.07 16.07 0.07

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 04 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5622
TYPE : Pressure Gauge
MAKE & MODEL : Mahaveer Searial No. : -
LEAST COUNT : 0.2 kg/cm2
LOCATION : Boiler 2 Hytherm Outlet
ACCP. CRI. : + 2 °C
RANGE : 0 to 10.6 kg/cm2 OPERATING RANGE : 0. to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 5.99 -0.01 5.99 -0.01
5 8.00 8.00 8.01 0.01 8.01 0.01
6 10.00 10.00 10.02 0.02 10.02 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 05 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5623
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160024
LEAST COUNT : 1 kg/cm2
LOCATION : Boiler 2 Furnace Oil Press.
ACCP. CRI. : + 2 °C
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0. to 16 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.05 0.05 10.05 0.05
3 20.00 20.00 20.07 0.07 20.07 0.07
4 30.00 30.00 30.09 0.09 30.09 0.09
5 40.00 40.00 40.13 0.13 40.13 0.13

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR – PG – 06 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5624
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 860065
LEAST COUNT : 0.5 kg/cm2
LOCATION : HR 3 Filter Pump
ACCP. CRI. : + 2 °C
RANGE : 0 to 28 kg/cm2 OPERATING RANGE : 0. to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 4.99 -0.01 4.99 -0.01
3 10.00 10.00 9.98 -0.02 9.98 -0.02
4 15.00 15.00 15.03 0.03 15.03 0.03
5 20.00 20.00 20.04 0.04 20.04 0.04
6 28.00 28.00 28.11 0.11 28.11 0.11
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – PG – 07 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5625
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : ST - 01 - P9 - 95
LEAST COUNT : 0.1 kg/cm2
LOCATION : ATFE Feed Pump
ACCP. CRI. : + 2 °C
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0. to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.01 0.01 6.01 0.01
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.02 0.02 10.02 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ATFE – PG – 08 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5626
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1310201036
LEAST COUNT : 0.2 kg/cm2
LOCATION : ATFE Recieve Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 INPUT: Pressure OUTPUT: Analog.
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.02 -0.02 4.02 -0.02
4 6.00 6.00 6.03 -0.03 6.03 -0.03
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.04 -0.04 10.04 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : EJ – PG – 09 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5627
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 254058
LEAST COUNT : 0.1 kg/cm2
LOCATION : Vacuum Ejector
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0. to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 7.00 7.00 7.02 0.02 7.02 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 10 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5628
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18012432
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.15 0.15 100.15 0.15
3 200.00 200.00 200.16 0.16 200.16 0.16
4 300.00 300.00 300.19 0.19 300.19 0.19
5 400.00 400.00 400.24 0.24 400.24 0.24

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 11 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5629
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18040805
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 99.98 -0.02 99.98 -0.02
3 200.00 200.00 199.92 -0.08 199.92 -0.08
4 300.00 300.00 299.90 -0.10 299.90 -0.10
5 400.00 400.00 399.82 -0.18 399.82 -0.18

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 12 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5630
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1606348788
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 9.98 -0.02 9.98 -0.02
3 20.00 20.00 20.01 0.01 20.01 0.01
4 40.00 40.00 40.04 0.04 40.04 0.04
5 60.00 60.00 60.06 0.06 60.06 0.06

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 13 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5631
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18040804
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.05 0.05 100.05 0.05
3 200.00 200.00 200.12 0.12 200.12 0.12
4 300.00 300.00 300.20 0.20 300.20 0.20
5 400.00 400.00 400.28 0.28 400.28 0.28

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 14 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5632
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1511317435
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.03 0.03 10.03 0.03
3 20.00 20.00 20.06 0.06 20.06 0.06
4 40.00 40.00 40.05 0.05 40.05 0.05
5 60.00 60.00 60.08 0.08 60.08 0.08

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 1 – PG – 15 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5633
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1511317432
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 9.97 -0.03 9.97 -0.03
3 20.00 20.00 19.95 -0.05 19.95 -0.05
4 40.00 40.00 39.92 -0.08 39.92 -0.08
5 60.00 60.00 59.88 -0.12 59.88 -0.12

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 16 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5634
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 19043282
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 2
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.18 0.18 100.18 0.18
3 200.00 200.00 200.27 0.27 200.27 0.27
4 300.00 300.00 300.33 0.33 300.33 0.33
5 400.00 400.00 400.54 0.54 400.54 0.54

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 17 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5635
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18107913
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 2
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.12 0.12 100.12 0.12
3 200.00 200.00 200.32 0.32 200.32 0.32
4 300.00 300.00 300.43 0.43 300.43 0.43
5 400.00 400.00 400.54 0.54 400.54 0.54

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 18 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5636
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 529798
LEAST COUNT : 0.5 kg/cm2
LOCATION : PRV 1
ACCP. CRI. : + 2 °C
RANGE : 0 to 28 kg/cm2 OPERATING RANGE : 0. to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.01 0.01 5.01 0.01
3 10.00 10.00 10.02 0.02 10.02 0.02
4 20.00 20.00 20.02 0.02 20.02 0.02
5 28.00 28.00 28.03 0.03 28.03 0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 19 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5637
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 19043286
LEAST COUNT : 10 kg/cm2
LOCATION : PRV 2
ACCP. CRI. : + 2 °C
RANGE : 0 to 400 kg/cm2 OPERATING RANGE : 0. to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 100.00 100.00 100.14 0.14 100.14 0.14
3 200.00 200.00 200.17 0.17 200.17 0.17
4 300.00 300.00 300.19 0.19 300.19 0.19
5 400.00 400.00 400.22 0.22 400.22 0.22

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 20 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5638
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 17092791
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 -0.01 10.01 -0.01
3 20.00 20.00 20.03 -0.03 20.03 -0.03
4 40.00 40.00 40.05 -0.05 40.05 -0.05
5 60.00 60.00 60.03 -0.03 60.03 -0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 2 – PG – 21 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5639
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1801529792
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.05 -0.05 20.05 -0.05
4 40.00 40.00 40.06 -0.06 40.06 -0.06
5 60.00 60.00 60.09 -0.09 60.09 -0.09
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 23 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5640
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950017
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -30" to 10 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 2.00 2.00 1.99 0.01 1.99 0.01
4 4.00 4.00 3.99 0.01 3.99 0.01
5 8.00 8.00 7.99 0.01 7.99 0.01
6 10.00 10.00 9.98 0.02 9.98 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – PG – 24 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5641
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950016
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -30" to 10 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 2.00 2.00 2.01 -0.01 2.01 -0.01
4 4.00 4.00 4.02 -0.02 4.02 -0.02
5 8.00 8.00 8.01 -0.01 8.01 -0.01
6 10.00 10.00 10.01 -0.01 10.01 -0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 – PG – 25 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5642
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 400210
LEAST COUNT : 1 kg/cm2
LOCATION : HR 1 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.04 -0.04 20.04 -0.04
4 30.00 30.00 30.06 -0.06 30.06 -0.06
5 40.00 40.00 40.07 -0.07 40.07 -0.07

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – PG – 26 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5643
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 400209
LEAST COUNT : 1 kg/cm2
LOCATION : HR 2 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.03 -0.03 10.03 -0.03
3 20.00 20.00 20.05 -0.05 20.05 -0.05
4 30.00 30.00 30.04 -0.04 30.04 -0.04
5 40.00 40.00 40.02 -0.02 40.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 – PG – 27 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5644
TYPE : Pressure Gauge
MAKE & MODEL : H Guru Searial No. : 400209
LEAST COUNT : 0.5 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. : + 2 % FSD
RANGE : 0 to 21 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.02 -0.02 5.02 -0.02
3 10.00 10.00 10.03 -0.03 10.03 -0.03
4 15.00 15.00 15.03 -0.03 15.03 -0.03
5 21.00 21.00 21.04 -0.04 21.04 -0.04

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FW – PG – 29 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5645
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 650014
LEAST COUNT : 1 kg/cm2
LOCATION : Fire Water Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0 to 7 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.03 -0.03 10.03 -0.03
3 20.00 20.00 20.04 -0.04 20.04 -0.04
4 30.00 30.00 30.08 -0.08 30.08 -0.08
5 40.00 40.00 40.15 -0.15 40.15 -0.15

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : WP – PG – 30 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5646
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160008
LEAST COUNT : 0.1 kg/cm2
LOCATION : MIDC Water Pump Pressure
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
Under Test
Sr. INPUT
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.00 0.00
4 6.00 6.00 6.02 -0.02 6.01 -0.01
5 8.00 8.00 8.02 -0.02 8.01 -0.01
6 10.00 10.00 10.03 -0.03 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 32 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5647
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT : 5 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 kg/cm2 OPERATING RANGE : 0 to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.06 -0.06 50.06 -0.06
3 100.00 100.00 100.09 -0.09 100.09 -0.09
4 150.00 150.00 150.15 -0.15 150.15 -0.15
5 200.00 200.00 200.21 -0.21 200.21 -0.21
6 250.00 250.00 250.27 -0.27 250.27 -0.27
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 33 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5648
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT : 5 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 kg/cm2 OPERATING RANGE : 0 to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.02 -0.02 50.02 -0.02
3 100.00 100.00 100.06 -0.06 100.06 -0.06
4 150.00 150.00 150.08 -0.08 150.08 -0.08
5 200.00 200.00 200.09 -0.09 200.09 -0.09
6 250.00 250.00 250.11 -0.11 250.11 -0.11
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 34 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5649
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.03 -0.03 20.03 -0.03
4 40.00 40.00 40.02 -0.02 40.02 -0.02
5 60.00 60.00 60.03 -0.03 60.03 -0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 35 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5650
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : 01864
LEAST COUNT : 5 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 kg/cm2 OPERATING RANGE : 0 to 25 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.05 -0.05 50.05 -0.05
3 100.00 100.00 100.08 -0.08 100.08 -0.08
4 150.00 150.00 150.09 -0.09 150.09 -0.09
5 200.00 200.00 200.11 -0.11 200.11 -0.11
6 250.00 250.00 250.14 -0.14 250.14 -0.14
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 36 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5651
TYPE : Pressure Gauge
MAKE & MODEL : Waaree Searial No. : 110167
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 100 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 20.00 20.00 20.03 -0.03 20.03 -0.03
3 40.00 40.00 40.05 -0.05 40.05 -0.05
4 60.00 60.00 60.09 -0.09 60.09 -0.09
5 80.00 80.00 80.15 -0.15 80.15 -0.15
6 100.00 100.00 100.24 -0.24 100.24 -0.24
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : PRV 3 – PG – 37 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5652
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT : 1 kg/cm2
LOCATION : PRV 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to 60 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 9.98 0.02 9.98 0.02
3 20.00 20.00 19.97 0.03 19.97 0.03
4 40.00 40.00 39.95 0.05 39.95 0.05
5 60.00 60.00 59.88 0.12 59.88 0.12

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 38 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5653
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 17126056
LEAST COUNT : 5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 bar OPERATING RANGE : 0 to 150 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.06 -0.06 50.06 -0.06
3 100.00 100.00 100.09 -0.09 100.09 -0.09
4 150.00 150.00 150.13 -0.13 150.13 -0.13
5 200.00 200.00 200.16 -0.16 200.16 -0.16
6 250.00 250.00 250.22 -0.22 250.22 -0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 39 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5654
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 17126059
LEAST COUNT : 5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 250 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 50.00 50.00 50.05 -0.05 50.05 -0.05
3 100.00 100.00 100.07 -0.07 100.07 -0.07
4 150.00 150.00 150.08 -0.08 150.08 -0.08
5 200.00 200.00 200.14 -0.14 200.14 -0.14
6 250.00 250.00 250.15 -0.15 250.15 -0.15
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 400 bar 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 4

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 40 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5655
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 18054079
LEAST COUNT : 0.5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 25 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.00 0.00 5.00 0.00
3 10.00 10.00 10.01 -0.01 10.01 -0.01
4 15.00 15.00 15.02 -0.02 15.02 -0.02
5 20.00 20.00 20.03 -0.03 20.03 -0.03
6 25.00 25.00 25.04 -0.04 25.04 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : N2 PRV – PG – 41 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5656
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : 1608372636
LEAST COUNT : 0.5 bar
LOCATION : N2 PRV
ACC. CRI. : + 2 % FSD
RANGE : 0 to 16 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 4.00 4.00 4.00 0.00 4.00 0.00
3 8.00 8.00 8.01 -0.01 8.01 -0.01
4 12.00 12.00 12.02 -0.02 12.02 -0.02
5 16.00 16.00 16.01 -0.01 16.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – PG – 42 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5657
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160024
LEAST COUNT : 0.1 kg/cm2
LOCATION : CEP - CAN Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CEP – PG – 43 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5658
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 511465
LEAST COUNT : 0.5 kg/cm2
LOCATION : CEP - MEOH Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 28 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 4.02 0.98 5.01 -0.01
3 10.00 10.00 8.97 1.03 10.01 -0.01
4 15.00 15.00 14.01 0.99 15.02 -0.02
5 20.00 20.00 18.95 1.05 20.04 -0.04
6 28.00 28.00 26.94 1.06 28.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 – PG – 44 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5659
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 750072
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR2 Filter Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.02 -0.02 4.02 -0.02
4 6.00 6.00 6.03 -0.03 6.03 -0.03
5 8.00 8.00 8.03 -0.03 8.03 -0.03
6 10.00 10.00 10.05 -0.05 10.05 -0.05
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 45 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5660
TYPE : Pressure Gauge
MAKE & MODEL : Spencer Searial No. : 185385
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower - Pump 1
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.02 -0.02 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 46 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5661
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160045
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower - Pump 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.99 0.01 5.99 0.01
5 7.00 7.00 6.98 0.02 6.98 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 47 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5662
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00435
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR 1 - Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 0.01 5.99 0.01
5 7.00 7.00 6.98 0.02 6.98 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 49 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5663
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 100208
LEAST COUNT : 0.5 kg/cm2
LOCATION : BA - Candle Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.02 -0.02 10.02 -0.02
3 20.00 20.00 20.04 -0.04 20.04 -0.04
4 30.00 30.00 30.03 -0.03 30.03 -0.03
5 40.00 40.00 40.05 -0.05 40.05 -0.05

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 48 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5664
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 01341
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 - Steam Line
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 6.01 -2.01 6.01 -2.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 50 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5665
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 03278
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR 3 - Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.02 -0.02 10.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 51 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5666
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 080160
LEAST COUNT : 0.1 kg/cm2
LOCATION : ST .Tank No. 22 Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.98 0.02 3.98 0.02
4 6.00 6.00 5.97 0.03 5.97 0.03
5 7.00 7.00 6.94 0.06 6.94 0.06

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 52 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5667
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 080120
LEAST COUNT : 0.1 kg/cm2
LOCATION : ST .Tank No. 26 Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.98 0.02 5.98 0.02
5 7.00 7.00 6.98 0.02 6.98 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – PG – 55 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5668
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 280035
LEAST COUNT : 0.2 kg/cm2
LOCATION : DR5 Circulation Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – PG – 56 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5669
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 800081
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 5 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -30" to 4 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 1.00 1.00 1.00 0.00 1.00 0.00
4 2.00 2.00 2.00 0.00 2.00 0.00
5 3.00 3.00 3.01 -0.01 3.01 -0.01
6 4.00 4.00 3.99 0.01 3.99 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – PG – 57 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5670
TYPE : Pressure Gauge
MAKE & MODEL : Nisan Searial No. : -
LEAST COUNT : 01 kg/cm2
LOCATION : DR 5 Steam Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to 25 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.01 -0.01 5.01 -0.01
3 10.00 10.00 10.01 -0.01 10.01 -0.01
4 15.00 15.00 15.03 -0.03 15.03 -0.03
5 20.00 20.00 20.05 -0.05 20.05 -0.05
6 25.00 25.00 25.04 -0.04 25.04 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 5 – PG – 58 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5671
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 01323
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 5 Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 6.99 0.01 6.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 59 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5672
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 032824
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Steam Stack
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 7 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – PG – 60 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5673
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00081
LEAST COUNT : 0.1 kg/cm2
LOCATION : Condensor Cooling Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 0.01 5.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CC – PG – 61 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5674
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00078
LEAST COUNT : 0.1 kg/cm2
LOCATION : Condensor Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.99 0.01 5.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – PG – 62 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5675
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00080
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 Cooling Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 – PG – 63 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5676
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00073
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.99 0.01 5.99 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 64 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5677
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00076
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Cooling Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – PG – 65 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5678
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00075
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 Cooling Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 6 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 0.01 5.99 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : FL – PG – 66 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5679
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 00079
LEAST COUNT : 0.1 kg/cm2
LOCATION : BA Candle Filter
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 -0.01 2.01 -0.01
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CT – PG – 67 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5680
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 210928
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 0.01 5.99 0.01
5 8.00 8.00 7.98 0.02 7.98 0.02
6 10.00 10.00 9.98 0.02 9.98 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 68 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5681
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 70087
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Chiller Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 69 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5682
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 70214
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Chiller Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 70 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5683
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 70189
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Cooling Water Inlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.98 0.02 5.98 0.02
5 7.00 7.00 6.98 0.02 6.98 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 71 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5684
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 01447
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Cooling Water Outlet
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 7.00 7.00 7.01 -0.01 7.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 72 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5685
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 160013
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Brine Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 4 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 1.00 1.00 1.00 0.00 1.00 0.00
3 2.00 2.00 2.00 0.00 2.00 0.00
4 3.00 3.00 3.00 0.00 3.00 0.00
5 4.00 4.00 4.01 -0.01 4.01 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : CP – PG – 73 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5686
TYPE : Pressure Gauge
MAKE & MODEL : Spencer Searial No. : 185385
LEAST COUNT : 0.1 kg/cm2
LOCATION : Chilling Plant - Chiller Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 3 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
Under Test
Sr. INPUT
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.02 -0.02 6.02 -0.02
5 8.00 8.00 8.02 -0.02 8.02 -0.02
6 10.00 10.00 10.03 -0.03 10.03 -0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HT – PG – 74 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5687
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950015
LEAST COUNT : 0.1 kg/cm2
LOCATION : Holding Tank 1
ACC. CRI. : + 2 % FSD
RANGE : -30" to 10 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 2.00 2.00 1.99 0.01 1.99 0.01
4 4.00 4.00 3.99 0.01 3.99 0.01
5 8.00 8.00 7.99 0.01 7.99 0.01
6 10.00 10.00 9.98 0.02 9.98 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HT – PG – 75 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5688
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 01332
LEAST COUNT : 0.1 kg/cm2
LOCATION : BA - Candle Filter
ACC. CRI. : + 2 % FSD
RANGE : -30" to 4.2 kg/cm2 OPERATING RANGE : -30" to 1 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
"Hg / Kg/cm2 "Hg / Kg/cm2 Output Output
1 -25.00 -25.00 -25.00 0.00 -25.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 1.00 1.00 0.98 0.02 0.98 0.02
4 2.00 2.00 1.99 0.01 1.99 0.01
5 4.00 4.00 3.97 0.03 3.97 0.03

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 76 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5689
TYPE : Pressure Gauge
MAKE & MODEL : Hi Tech Searial No. : 0044
LEAST COUNT : 0.1 kg/cm2
LOCATION : Storage Tank No. 21 Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 77 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5690
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 080125
LEAST COUNT : 0.1 kg/cm2
LOCATION : Storage Tank No. 23 Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 -0.01 4.01 -0.01
4 6.00 6.00 6.01 -0.01 6.01 -0.01
5 7.00 7.00 7.02 -0.02 7.02 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST – PG – 78 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5691
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 670126
LEAST COUNT : 0.1 kg/cm2
LOCATION : Storage Tank No. 01 Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 2 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.98 0.02 5.98 0.02
5 7.00 7.00 6.96 0.04 6.96 0.04

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 79 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5692
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : IBR Boiler Main Steam Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 8 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 0.01 3.99 0.01
4 6.00 6.00 5.99 0.01 5.99 0.01
5 8.00 8.00 7.98 0.02 7.98 0.02
6 10.00 10.00 9.98 0.02 9.98 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 80 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5693
TYPE : Pressure Gauge
MAKE & MODEL : Poiner Searial No. : -
LEAST COUNT : 0.5 kg/cm2
LOCATION : IBR Boiler Main Steam
ACC. CRI. : + 2 % FSD
RANGE : 0 to 21 kg/cm2 OPERATING RANGE : 0 to 8 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.01 -0.01 5.01 -0.01
3 10.00 10.00 10.02 -0.02 10.02 -0.02
4 15.00 15.00 15.04 -0.04 15.04 -0.04
5 20.00 20.00 20.05 -0.05 20.05 -0.05
6 21.00 21.00 21.06 -0.06 21.06 -0.06
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 81 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5694
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT : 0.5 kg/cm2
LOCATION : IBR Boiler F. O. Burner
ACC. CRI. : + 2 % FSD
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.01 -0.01 10.01 -0.01
3 20.00 20.00 20.04 -0.04 20.04 -0.04
4 30.00 30.00 30.05 -0.05 30.05 -0.05
5 40.00 40.00 40.07 -0.07 40.07 -0.07

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 82 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5695
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT : 0.5 kg/cm2
LOCATION : IBR Boiler F. O. Burner
ACC. CRI. : + 2 % FSD
RANGE : 0 to 40 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 9.99 0.01 9.99 0.01
3 20.00 20.00 19.98 0.02 19.98 0.02
4 30.00 30.00 29.96 0.04 29.96 0.04
5 40.00 40.00 39.95 0.05 39.95 0.05

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 83 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5696
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT : 10 mmWC
LOCATION : IBR Boiler F. O. Burner
ACC. CRI. : + 2 % FSD
RANGE : 0 to 1000 mmWC OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmWC mmWC Output Output
1 0 0 0 0 0 0
2 200 200 200 0 200 0
3 400 400 400 0 400 0
4 600 600 600 0 600 0
5 800 800 800 0 800 0
6 1000 1000 1000 0 1000 0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Presuure Meter +/-10000 mmWC 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 3

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 84 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5697
TYPE : Pressure Gauge
MAKE & MODEL : General Searial No. : -
LEAST COUNT : 0.5 kg/cm2
LOCATION : IBR Boiler F. O. Water Pump No. 1
ACC. CRI. : + 2 % FSD
RANGE : 0 to 25 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.01 -0.01 5.01 -0.01
3 10.00 10.00 10.02 -0.02 10.02 -0.02
4 15.00 15.00 15.03 -0.03 15.03 -0.03
5 20.00 20.00 20.04 -0.04 20.04 -0.04
6 25.00 25.00 25.05 -0.05 25.05 -0.05
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PG – 85 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5698
TYPE : Pressure Gauge
MAKE & MODEL : Fiebig Searial No. : -
LEAST COUNT : 1 kg/cm2
LOCATION : IBR Boiler F. O. Water Pump No. 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to 42 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
Kg/cm2 Kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 10.00 10.00 10.03 -0.03 10.03 -0.03
3 20.00 20.00 20.06 -0.06 20.06 -0.06
4 30.00 30.00 30.08 -0.08 30.08 -0.08
5 40.00 40.00 40.12 -0.12 40.12 -0.12

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PS – 05 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5699
TYPE : Pressure Switch
MAKE & MODEL : Orian Searial No. : -
SEARIAL NO. :-
LOCATION : Boiler 02 Hytherm Low Press.
ACC. CRI. : + 2 % FSD
RANGE : 0 to 8 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Sr. INPUT SET SET RESET
No. Kg/cm2 Kg/cm2 Kg/cm2 Kg/cm2

·          3.00 3.00 3.05 4.56

REMARKS : Contacts Checked O. K.


a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO – PS – 06 CAL. DATE : 13 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5700
TYPE : Pressure Switch
MAKE & MODEL : Orian Searial No. : -
SEARIAL NO. :-
LOCATION : Boiler 03 Steam Seperator
ACC. CRI. : + 2 % FSD
RANGE : 0 to 15 kg/cm2 OPERATING RANGE : 0 to 10 kg/cm2
AMB : 25 + 2 ° C
Sr. INPUT SET SET RESET
No. Kg/cm2 Kg/cm2 Kg/cm2 Kg/cm2

·          10.00 10.00 10.03 8.05

REMARKS : Contacts Checked O. K.


a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 12 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5701
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 01946
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 1
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 03 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5702
TYPE : Vaccum Gauge
MAKE & MODEL : LA Searial No. : 668
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 2
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VP – VG – 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5703
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. : 666
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 3
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VR – VG – 04 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5704
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. : 669
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump 4
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – VG – 05 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5705
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. : 01946
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 4 Reactor
ACC. CRI. : + 2 % FSD
RANGE : -760 to 4.2 Kg/cm2 OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg/Kg mmHg/Kg Output Output
1 -700.00 -700.00 -700.00 0.00 -700.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 1.00 1.00 1.00 0.00 1.00 0.00
4 2.00 2.00 2.00 0.00 2.00 0.00
5 3.00 3.00 3.01 -0.01 3.01 -0.01
6 4.00 4.00 4.02 -0.02 4.02 -0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 1 – VG – 06 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5706
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00037
LEAST COUNT :10 mmHg
LOCATION : DR 1 & DR 2 Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 4 – VG – 07 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5707
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : 501678
LEAST COUNT :10 mmHg
LOCATION : DR 4 Vacuum Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 3 – VG – 08 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5708
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : 501720
LEAST COUNT :10 mmHg
LOCATION : DR 3 Vacuum Header
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VR – VG – 09 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5709
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. : -
LEAST COUNT :10 mmHg
LOCATION : Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : VR – VG – 10 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5710
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 1943
LEAST COUNT :10 mmHg
LOCATION : Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 11 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5711
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 00043
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 12 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5712
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : 650018
LEAST COUNT :10 mmHg
LOCATION : DR 5 Circulation Pump Line
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 13 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5713
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. :
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 14 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5714
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. : 00038
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR3 – VG – 15 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5715
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : 800085
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR5 Vacuum Reactor
ACC. CRI. : + 2 % FSD
RANGE : -760 to 4.2 Kg/cm2 OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg/Kg mmHg/Kg Output Output
1 -700.00 -700.00 -700.00 0.00 -700.00 0.00
2 0.00 0.00 0.00 0.00 0.00 0.00
3 1.00 1.00 1.00 0.00 1.00 0.00
4 2.00 2.00 2.00 0.00 2.00 0.00
5 3.00 3.00 3.00 0.00 3.00 0.00
6 4.00 4.00 3.99 0.01 3.99 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR5 – VG – 16 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5716
TYPE : Vaccum Gauge
MAKE & MODEL : Hi Tech Searial No. : -
LEAST COUNT :10 mmHg
LOCATION : DR5 Couloum
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Under Test
Sr. INPUT Before Adjustment After Adjustment
Output
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 - PT - 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5717
TYPE : Pressure Transmitter
MAKE & MODEL : Honeywell Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 1 Reactor
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
kg/cm2 mA Output Output
1 0.00 4.00 3.99 -0.01 3.99 -0.01
2 7.50 8.00 7.99 -0.01 7.99 -0.01
3 15.00 12.00 12.00 0.00 12.00 0.00
4 22.50 16.00 16.01 0.01 16.01 0.01
5 30.00 20.00 19.99 -0.01 19.99 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 - PT - 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5718
TYPE : Pressure Transmitter
MAKE & MODEL : Honeywell Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 2 Reactor
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
kg/cm2 mA Output Output
1 0.00 4.00 3.99 -0.01 3.99 -0.01
2 7.50 8.00 7.99 -0.01 7.99 -0.01
3 15.00 12.00 12.00 0.00 12.00 0.00
4 22.50 16.00 15.99 -0.01 15.99 -0.01
5 30.00 20.00 19.98 -0.02 19.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 - PT - 03 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5719
TYPE : Pressure Transmitter
MAKE & MODEL : Honeywell Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
kg/cm2 mA Output Output
1 0.00 4.00 4.01 0.01 4.01 0.01
2 7.50 8.00 8.01 0.01 8.01 0.01
3 15.00 12.00 12.01 0.01 12.01 0.01
4 22.50 16.00 16.01 0.01 16.01 0.01
5 30.00 20.00 20.02 0.02 20.02 0.02
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO - PT - 04 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5720
TYPE : Pressure Transmitter
MAKE & MODEL : Nishko Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : IBR Main Steam Header
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 16 bar OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
bar bar Output Output
1 0.00 4.00 3.99 -0.01 3.99 -0.01
2 4.00 8.00 7.99 -0.01 7.99 -0.01
3 8.00 12.00 12.00 0.00 12.00 0.00
4 12.00 16.00 16.00 0.00 16.00 0.00
5 16.00 20.00 19.99 -0.01 19.99 -0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 - VT - 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5721
TYPE : Vacuum Transmitter
MAKE & MODEL : Baumer Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 Receiver
ACC. CRI. : + 0.25 % FSD
RANGE : -1 to 1.5 bar OPERATING RANGE : -1 to 1.5 bar
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
bar mA Output Output
1 -0.800 6.28 6.30 0.02 6.30 0.02
2 -0.375 8.00 8.00 0.00 8.00 0.00
3 0.250 12.00 11.99 -0.01 11.99 -0.01
4 0.875 16.00 15.99 -0.01 15.99 -0.01
5 1.500 20.00 19.99 -0.01 19.99 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LT - 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5722
TYPE : Level Transmitter
MAKE & MODEL : Honeywell ST700 Searial No. : 499162
LEAST COUNT : 0.1 mmWc
LOCATION : ACN Tank
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 5600 mmWC OPERATING RANGE : 0 to 5600 mmWC
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mmWC mA Output Output
1 0 4.00 3.99 -0.01 3.99 -0.01
2 1400 8.00 8.01 0.01 8.01 0.01
3 2800 12.00 12.01 0.01 12.01 0.01
4 4200 16.00 16.01 0.01 16.01 0.01
5 5600 20.00 19.99 -0.01 19.99 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Presuure Meter +/-10000 mmWC 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 3
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LT - 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5723
TYPE : Level Transmitter
MAKE & MODEL : Honeywell ST700 Searial No. : 499162
LEAST COUNT : 0.1 mmWc
LOCATION : ACN Tank
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 3120 mmWC OPERATING RANGE : 0 to 3120 mmWC
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mmWC mA Output Output
1 0 4.00 3.99 -0.01 3.99 -0.01
2 780 8.00 8.00 0.00 8.00 0.00
3 1560 12.00 12.00 0.00 12.00 0.00
4 2340 16.00 15.99 -0.01 15.99 -0.01
5 3120 20.00 19.98 -0.02 19.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Presuure Meter +/-10000 mmWC 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 3
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LT - 03 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5724
TYPE : Level Transmitter
MAKE & MODEL : Honeywell ST700 Searial No. : 499162
LEAST COUNT : 0.1 mmWc
LOCATION : ACN Tank
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 4640 mmWC OPERATING RANGE : 0 to 4640 mmWC
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mmWC mA Output Output
1 0 4.00 3.99 -0.01 3.99 -0.01
2 1160 8.00 8.00 0.00 8.00 0.00
3 2320 12.00 12.00 0.00 12.00 0.00
4 3480 16.00 16.01 0.01 16.01 0.01
5 4640 20.00 20.02 0.02 20.02 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Presuure Meter +/-10000 mmWC 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 3
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LT - 04 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5725
TYPE : Level Transmitter
MAKE & MODEL : Honeywell ST700 Searial No. : 499162
LEAST COUNT : 0.1 mmWc
LOCATION : ACN Tank
ACC. CRI. : + 0.25 % FSD
RANGE : 0 to 9000 mmWC OPERATING RANGE : 0 to 9000 mmWC
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mmWC mA Output Output
1 0 4.00 3.99 -0.01 3.99 -0.01
2 2250 8.00 8.00 0.00 8.00 0.00
3 4500 12.00 11.99 -0.01 11.99 -0.01
4 6750 16.00 15.98 -0.02 15.98 -0.02
5 9000 20.00 19.98 -0.02 19.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Presuure Meter +/-10000 mmWC 08 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 3
2 Digital Multimeter 0 to 20 mA 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LI - 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5726
TYPE : Level Indicator
MAKE & MODEL : Newtech Searial No. : -
LEAST COUNT : 1 cm
LOCATION : ACN Tank
ACC. CRI. :+1%
RANGE : 0 to 700 cm OPERATING RANGE : 0 to 700 cm
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
Sr.
No. Actual Actual
Deviation Deviation
mA cm Output Output
1 4.00 0 0 0 0 0
2 8.00 175 175 0 175 0
3 12.00 350 351 1 351 1
4 16.00 525 526 1 526 1
5 20.00 700 703 3 703 3

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LI - 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5727
TYPE : Level Indicator
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.1 cm
LOCATION : Tank No. 16
ACC. CRI. :+1%
RANGE : 0 to 390 cm OPERATING RANGE : 0 to 390 cm
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
INPUT Desired Output
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA cm Output Output
1 4.00 0.0 0.0 0.0 0.0 0.0
2 8.00 97.5 97.6 0.1 97.6 0.1
3 12.00 195.0 195.2 0.2 195.2 0.2
4 16.00 292.5 292.7 0.2 292.7 0.2
5 20.00 390.0 390.2 0.2 390.2 0.2

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LI - 03 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5728
TYPE : Level Indicator
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 1 cm
LOCATION : Tank No. 15
ACC. CRI. :+1%
RANGE : 0 to 580 cm OPERATING RANGE : 0 to 580 cm
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA cm Output Output
1 4.00 0 0 0 0 0
2 8.00 145 146 1 146 1
3 12.00 290 291 1 291 1
4 16.00 435 437 2 437 2
5 20.00 580 582 2 582 2

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : ST - LI - 04 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5729
TYPE : Level Indicator
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 1 cm
LOCATION : Tank No. 25
ACC. CRI. :+1%
RANGE : 0 to 1125 cm OPERATING RANGE : 0 to 1125 cm
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA cm Output Output
1 4.00 0 0 0 0 0
2 8.00 281 281 0 281 0
3 12.00 563 562 -1 562 -1
4 16.00 844 846 2 846 2
5 20.00 1125 1126 1 1126 1

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 1 - PIC - 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5730
TYPE : Pressure Indicator Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 1 Reactor
ACC. CRI. :+1%
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.00 0.00 0.00 0.00 0.00
2 8.00 7.50 7.51 0.01 7.51 0.01
3 12.00 15.00 15.01 0.01 15.01 0.01
4 16.00 22.50 22.52 0.02 22.52 0.02
5 20.00 30.00 30.01 0.01 30.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 2 - PIC - 02 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5731
TYPE : Pressure Indicator Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 2 Reactor
ACC. CRI. :+1%
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.00 0.00 0.00 0.00 0.00
2 8.00 7.50 7.50 0.00 7.50 0.00
3 12.00 15.00 14.99 -0.01 14.99 -0.01
4 16.00 22.50 22.49 -0.01 22.49 -0.01
5 20.00 30.00 29.99 -0.01 29.99 -0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : HR 3 - PIC - 03 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5732
TYPE : Pressure Indicator Controller
MAKE & MODEL : Udian Searial No. : -
LEAST COUNT : 0.01 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. :+1%
RANGE : 0 to 30 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.00 0.00 0.00 0.00 0.00
2 8.00 7.50 7.50 0.00 7.50 0.00
3 12.00 15.00 14.99 -0.01 14.99 -0.01
4 16.00 22.50 22.49 -0.01 22.49 -0.01
5 20.00 30.00 29.98 -0.02 29.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : BO - PIC - 04 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5733
TYPE : Pressure Indicator Controller
MAKE & MODEL : Selec Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : HR 3 Reactor
ACC. CRI. :+1%
RANGE : 0 to 16 kg/cm2 OPERATING RANGE : 0 to 20 kg/cm2
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA kg/cm2 Output Output
1 4.00 0.0 0.0 0.0 0.0 0.0
2 8.00 4.0 4.0 0.0 4.0 0.0
3 12.00 8.0 8.0 0.0 8.0 0.0
4 16.00 12.0 12.0 0.0 12.0 0.0
5 20.00 16.0 15.9 -0.1 15.9 -0.1

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI.(A 39/40)
TAG. NO. : DR 2 - VI - 01 CAL. DATE : 14 / 01 / 2022
CERT. NO. : CR / 21 – 22 / 5734
TYPE : Vacuum Indicator
MAKE & MODEL : Udian Searial No. :-
LEAST COUNT : 1 mmHg
LOCATION : HR 3 Reactor
ACC. CRI. :+1%
RANGE : -760 to 1140 mmHg OPERATING RANGE : -760 to 1140 mmHg
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
mA mmHg Output Output
1 4.00 -760 -758 2 -758 2
2 8.00 -285 -286 -1 -286 -1
3 12.00 190 191 1 191 1
4 16.00 665 667 2 667 2
5 20.00 1140 1142 2 1142 2

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Calibrator 0 to 20 mA 02 / 09 / 2022
2021 - 22 / CFC / MUM / 250 / 1

CALIBRATION VALID UPTO : 13 / 07 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


Dear Sir,

List of instruments for Calibration A 39/40 Plant(Sigma)


Sr. No. Description Qty. Unit Rate Amount
1 Digital Temp. Indicator 60 150 9000
2 Temp. Element (Sensor) 60 125 7500
3 Pressure Gauge 100 75 7500
4 Vacuum Gauge 16 75 1200
5 Pressure Switch 2 250 500
6 Level Transmitter 4 400 1600
7 Level Indicator 4 250 1000
8 Vacuum Transmitter 1 400 400
9 Vacuum Indicator 1 250 250
10 Pressure Transmitter 4 400 1600
11 Pressure Indicator 4 250 1000

Total 31550
G.S. T. 18 % 5679
37229

Thanking you

Best regards

Mr. VISHAL DHAGE

M/s:     NEWTECH INSTRUMENT SERVICES


            13, Virat Dham CHS,
            Opp. Adarsh Vidyalay,
            Kulgoan, Badlapur (E)
            Dist. - Thane 421503
Email : nis_cal@yahoo.co.in
            newtech.instrument@rediffmail.com
CELL: 9272159539 / 9960738338

Dear Sir,

Aditional List of instruments for Calibration A 39/40 Plant(Sigma)


Sr. No. Description Qty. Unit Rate Amount
1 Pressure Gauge 18 75 1350
2 Pressure Transmitter 1 400 400
3 Pressure Indicator 1 250 250
4 Vacuum Gauge 16 75 1200
Total 3200
G.S. T. 18 % 576
3776

Thanking you

Best regards

Mr. VISHAL DHAGE

M/s:     NEWTECH INSTRUMENT SERVICES


            13, Virat Dham CHS,
            Opp. Adarsh Vidyalay,
            Kulgoan, Badlapur (E)
            Dist. - Thane 421503
Email : nis_cal@yahoo.co.in
            newtech.instrument@rediffmail.com
CELL: 9272159539 / 9960738338

Dear Sir,

List of instruments for Calibration A 43 Plant


Sr. No. Description Qty. Unit Rate Amount
1 Digital Temp. Indicator 10 150 1500
2 Temp. Element (Sensor) 10 125 1250
3 Pressure Gauge 15 75 1125
4 Vacuum Gauge 5 75 375

Total 4250
G.S. T. 18 % 765
5015

Thanking you

Best regards

Mr. VISHAL DHAGE


M/s:     NEWTECH INSTRUMENT SERVICES
            13, Virat Dham CHS,
            Opp. Adarsh Vidyalay,
            Kulgoan, Badlapur (E)
            Dist. - Thane 421503
Email : nis_cal@yahoo.co.in
            newtech.instrument@rediffmail.com
CELL: 9272159539 / 9960738338
A 39/40 Plant(Sigma)
62 150 9300
62 125 7750
92 75 6900
16 75 1200
2 250 500
4 400 1600
4 250 1000
1 400 400
1 250 250
4 400 1600
4 250 1000

252
Total 31500
G.S. T. 18 % 5670
37170
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PHT – DTI – 01 TAG NO. ATFE – DTI – 02 TAG NO. ATFE – DTI – 03
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CW – DTI – 04 TAG NO. CW – DTI – 05 TAG NO. BO – DTI – 06
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – DTI – 07 TAG NO. BO – DTC – 08 TAG NO. BO – DTC – 09
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – DTC – 10 TAG NO. BO – DTI – 11 TAG NO. DR 1 – DTI – 12
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 – DTI – 13 TAG NO. DR 1 – DTI – 14 TAG NO. DR 2 – DTI – 15
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 2 – DTI – 16 TAG NO. DR 2 – DTI – 17 TAG NO. DR 3 – DTI – 18
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 3 – DTI – 19 TAG NO. DR 3 – DTI – 20 TAG NO. DR 4 – DTI – 21
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 4 – DTI – 22 TAG NO. DR 4 – DTI – 23 TAG NO. DR3 – DTI – 24
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR5 – DTI – 25 TAG NO. CC – DTI – 26 TAG NO. CC – DTI – 27
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CC – DTI – 28 TAG NO. CC – DTI – 29 TAG NO. DR 5 – DTI – 30
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – DTI – 31 TAG NO. DR 5 – DTI – 32 TAG NO. DR 5 – DTI – 33
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – DTI – 34 TAG NO. DR 5 – DTI – 35 TAG NO. HR 1 – DTC – 36
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. HR 2 – DTC – 37 TAG NO. HR 3 – DTC – 38 TAG NO. BT – DTI – 40
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BT – DTI – 41 TAG NO. DR 5 – DTI – 42 TAG NO. DR 5 – DTI – 43
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – DTI – 44 TAG NO. DR 5 – DTI – 45 TAG NO. DR 5 – DTI – 46
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Sign. Sign. Sign.
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – DTI – 47 TAG NO. DR 5 – DTI – 48 TAG NO. DR 5 – DTI – 49
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. H 1 – DTI – 50 TAG NO. H 6 – DTI – 51 TAG NO. H 8 – DTI – 52
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CEP – DTI – 53 TAG NO. CEP – DTI – 54 TAG NO. HR 2 – DTI – 55
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – DTC – 56 TAG NO. BO – DTC – 57 TAG NO. VC - DTI - 39
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – DTC – 58 TAG NO. HR 2 – DTI – 59 TAG NO. HR 2 – DTI – 60
Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2 Cal. Date: 1 1 0 1 2 2
Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign. Due Date: 1 0 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PHT – TE – 01 TAG NO. ATFE – TE – 02 TAG NO. ATFE – TE – 03
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CW – TE – 04 TAG NO. CW – TE – 05 TAG NO. BO – TE – 06
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – TE – 07 TAG NO. BO – TE – 08 TAG NO. BO – TE – 09
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – TE – 10 TAG NO. BO – TE – 11 TAG NO. DR 1 – TE – 12
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 – TE – 13 TAG NO. DR 1 – TE – 14 TAG NO. DR 2 – TE – 15
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 2 – TE – 16 TAG NO. DR 2 – TE – 17 TAG NO. DR 3 – TE – 18
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 3 – TE – 19 TAG NO. DR 3 – TE – 20 TAG NO. DR 4 – TE – 21
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 4 – TE – 22 TAG NO. DR 4 – TE – 23 TAG NO. DR3 – TE – 24
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR5 – TE – 25 TAG NO. CC – TE – 26 TAG NO. CC – TE – 27
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CC – TE – 28 TAG NO. CC – TE – 29 TAG NO. DR 5 – TE – 30
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Sign. Sign. Sign.
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – TE – 31 TAG NO. DR 5 – TE – 32 TAG NO. DR 5 – TE – 33
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – TE – 34 TAG NO. DR 5 – TE – 35 TAG NO. HR 1 – TE – 36
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. HR 2 – TE – 37 TAG NO. HR 3 – TE – 38 TAG NO. BT – TE – 40
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BT – TE – 41 TAG NO. DR 5 – TE – 42 TAG NO. DR 5 – TE – 43
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – TE – 44 TAG NO. DR 5 – TE – 45 TAG NO. DR 5 – TE – 46
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – TE – 47 TAG NO. DR 5 – TE – 48 TAG NO. DR 5 – TE – 49
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. H 1 – TE – 50 TAG NO. H 6 – TE – 51 TAG NO. H 6 – TE – 52
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CEP – TE – 53 TAG NO. CEP – TE – 54 TAG NO. HR 2 – TE – 55
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – TE – 56 TAG NO. BO – TE – 57 TAG NO. VC – TE – 39
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – TE – 58 TAG NO. HR 2 – TE – 59 TAG NO. HR 2 – TE – 60
Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2 Cal. Date: 1 2 0 1 2 2
Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign. Due Date: 1 1 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – PG – 03 TAG NO. BO – PG – 04 TAG NO. BO – PG – 05
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ATFE – PG – 07 TAG NO. EJ – PG – 09 TAG NO. PRV1 – PG – 10
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PRV1 – PG – 11 TAG NO. PRV1 – PG – 12 TAG NO. PRV1 – PG – 13
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PRV1 – PG – 14 TAG NO. PRV1 – PG – 15 TAG NO. PRV2 – PG – 16
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PRV2 – PG – 17 TAG NO. PRV2 – PG – 18 TAG NO. PRV2 – PG – 19
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Sign. Sign. Sign.
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PRV2 – PG – 20 TAG NO. PRV2 – PG – 21 TAG NO. AH – PG – 22
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 – PG – 23 TAG NO. DR 2 – PG – 24 TAG NO. HR 1 – PG – 25
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. HR 2 – PG – 26 TAG NO. HR 3 – PG – 27 TAG NO. FW – PG – 29
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. WP – PG – 30 TAG NO. PRV 3 – PG – 32 TAG NO. PRV 3 – PG – 33
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PRV 3 – PG – 34 TAG NO. PRV 3 – PG – 35 TAG NO. PRV 3 – PG – 36
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. PRV 3 – PG – 37 TAG NO. N2 PRV – PG – 38 TAG NO. N2 PRV – PG – 39
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. N2 PRV – PG – 40 TAG NO. N2 PRV – PG – 41 TAG NO. CEP – PG – 42
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CEP – PG – 43 TAG NO. FL – PG – 06 TAG NO. HR2 – PG – 44
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CT – PG – 45 TAG NO. CT – PG – 46 TAG NO. FL – PG – 47
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 – PG – 48 TAG NO. FL – PG – 49 TAG NO. FL – PG – 50
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ST – PG – 51 TAG NO. ST – PG – 52 TAG NO. DR 5 – PG – 55
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 – PG – 57 TAG NO. DR 5 – PG – 58 TAG NO. DR 5 – PG – 56
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 – PG – 59 TAG NO. CC – PG – 60 TAG NO. CC – PG – 61
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 2 – PG – 62 TAG NO. DR 2 – PG – 63 TAG NO. DR 1 – PG – 64
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 – PG – 65 TAG NO. CP – PG – 66 TAG NO. CT – PG – 67
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CP – PG – 68 TAG NO. CP – PG – 69 TAG NO. CP – PG – 70
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. CP – PG – 71 TAG NO. CP – PG – 72 TAG NO. CP – PG – 73
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. HT – PG – 74 TAG NO. FL – PG – 75 TAG NO. BO - PS - 05
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO - PS - 06 TAG NO. ATFE – PG – 08 TAG NO. ST – PG – 76
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ST – PG – 77 TAG NO. ST – PG – 78 TAG NO. BO – PG – 79
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – PG – 80 TAG NO. BO – PG – 81 TAG NO. BO – PG – 82
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO – PG – 83 TAG NO. BO – PG – 84 TAG NO. BO – PG – 85
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. VP - VG - 01 TAG NO. VP - VG - 02 TAG NO. VP - VG - 03
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ST - LT - 01 TAG NO. ST - LI - 01 TAG NO. DR 2 - VT - 01
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 2 - VI - 01 TAG NO. HR 1 - PT - 01 TAG NO. HR 2 - PT - 02
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. HR 3 - PT - 03 TAG NO. HR 1 - PI - 01 TAG NO. HR 2 - PI - 02
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. HR 3 - PI - 03 TAG NO. VP - VG - 04 TAG NO. DR 4 - VG - 05
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 1 - VG - 06 TAG NO. DR 4 - VG - 07 TAG NO. DR 3 - VG - 08
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. VR - VG - 09 TAG NO. VR - VG - 10 TAG NO. DR 5 - VG - 11
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 5 - VG - 12 TAG NO. DR 5 - VG - 13 TAG NO. DR 5 - VG - 14
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. DR 3 - VG - 15 TAG NO. DR 5 - VG - 16 TAG NO. ST - LT - 02
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ST - LT - 03 TAG NO. ST - LT - 04 TAG NO. ST - LI - 02
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ST - LI - 03 TAG NO. ST - LI - 04 TAG NO. ST - LT - 04
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. ST - LT - 03 TAG NO. ST - LT - 04 TAG NO. BO - PT - 04
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. BO - PIC - 04 TAG NO. TAG NO.
Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2 Cal. Date: 1 3 0 1 2 2
Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign. Due Date: 1 2 0 7 2 2 Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES NEWTECH INSTRUMENT SERVICES
Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539 Mob. : 9423344342 / 9272159539
TAG NO. TAG NO. TAG NO.
Cal. Date: Cal. Date: Cal. Date:
Due Date: Sign. Due Date: Sign. Due Date: Sign.
CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 01 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7401
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. :
LEAST COUNT : 1 °C
LOCATION : DR 1 - Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 02 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7402
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 101 1 101 1
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 03 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7403
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Vent Trap Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
`` IDENTIFICATION RANGE CAL.VALID UPTO
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 04 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7404
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Vent Trap Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 101 1 101 1
4 200 200 201 1 201 1
5 300 300 302 2 302 2
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 05 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7405
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1
CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 06 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7406
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Productive Condensor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 99 -1 99 -1
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 07 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7407
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Product Sub Cooler Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR1 – DTI – 08 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7408
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 1 - Product Sub Cooler Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 201 1 201 1
5 300 300 301 1 301 1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 2 – DTI – 09 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7409
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 0.1 °C
LOCATION : DR 2 - Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0.0 0.0 -0.2 -0.2 -0.2 -0.2
2 50.0 50.0 49.8 -0.2 49.8 -0.2
3 100.0 100.0 99.7 -0.3 99.7 -0.3
4 150.0 150.0 149.5 -0.5 149.5 -0.5
5 200.0 200.0 199.6 -0.4 199.6 -0.4
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR2 – DTI – 10 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7410
TYPE : Digital Temp. Indicator
MAKE & MODEL : Creative Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 2 - Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 199 -1 199 -1
5 300 300 299 -1 299 -1
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR3 – DTI – 11 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7411
TYPE : Digital Temp. Indicator
MAKE & MODEL : Plastech Searial No. : -
LEAST COUNT : 1 °C
LOCATION : DR 3 - Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Before Adjustment After Adjustment
Sr. INPUT Desired Output
No. Actual Actual
Deviation Deviation
°C °C Output Output
1 0 0 0 0 0 0
2 50 50 50 0 50 0
3 100 100 100 0 100 0
4 200 200 200 0 200 0
5 300 300 300 0 300 0
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 RTD Calibrator - 100 to 400 °C 18 / 06 / 2022
2021 - 22 / CFC / MUM / 91 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 01 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7412
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 - Reactor Temp.
ACC. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.47 0.07
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.25 0.23
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2000 KOhm 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 02 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7413
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 - Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.31 -0.09
3 100 138.50 138.40 -0.10
4 150 157.31 157.19 -0.12
5 200 175.84 175.70 -0.14
6 300 212.02 211.81 -0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 03 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7414
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 - Vent Trap Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.66 -0.01
2 50 119.40 119.35 -0.05
3 100 138.50 138.39 -0.11
4 150 157.31 157.23 -0.08
5 200 175.84 175.78 -0.06
6 300 212.02 211.98 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 04 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7415
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 - Vent Trap Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 05 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7416
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 - Reflux Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 50 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.70 0.03
2 50 119.40 119.45 0.05
3 100 138.50 138.57 0.07
4 150 157.31 157.43 0.12
5 200 175.84 175.97 0.13
6 300 212.02 212.20 0.18
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 06 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7417
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. :-
LEAST COUNT :-
LOCATION : DR 1 - Productive Condensor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.69 0.02
2 50 119.40 119.47 0.07
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – TE – 07 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7418
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. :-
LEAST COUNT :-
LOCATION : DR 1 - Product Sub Cooler Inlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.65 -0.02
2 50 119.40 119.42 0.02
3 100 138.50 138.58 0.08
4 150 157.31 157.44 0.13
5 200 175.84 176.01 0.17
6 300 212.02 212.23 0.21
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : BO – TE – 08 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7419
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 1 - Product Sub Cooler Outlet Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.49 0.09
3 100 138.50 138.62 0.12
4 150 157.31 157.48 0.17
5 200 175.84 176.05 0.21
6 300 212.02 212.27 0.25
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 2 – TE – 09 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7420
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 - Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.67 0.00
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 176.01 0.17
6 300 212.02 212.21 0.19
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 2 – TE – 10 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7421
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 - Vapour Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 300 °C OPERATING RANGE : amb. to 100 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.71 0.04
2 50 119.40 119.46 0.06
3 100 138.50 138.59 0.09
4 150 157.31 157.45 0.14
5 200 175.84 175.99 0.15
6 300 212.02 212.24 0.22
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 3 – TE – 11 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7422
TYPE : RTD Sensor (PT – 100 )
MAKE & MODEL :- Searial No. : -
LEAST COUNT :-
LOCATION : DR 2 - Reactor Temp.
ACCP. CRI. : + 2 °C
RANGE : 0 to 200 °C OPERATING RANGE : amb. to 150 °C
AMB : 25 + 2 ° C
Sr. INPUT Desired Output Actual Output Deviation
No. °C W W W
1 30 111.67 111.68 0.01
2 50 119.40 119.45 0.05
3 100 138.50 138.59 0.09
4 150 157.31 157.43 0.12
5 200 175.84 175.98 0.14
6 300 212.02 212.18 0.16
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Thermometer with Sensor - 50 to 600 ° C 10 / 01 / 2023
2021 - 22 / CFC / MUM / 511 / 2
2 Digital Multimeter 0 to 2 KW 30 / 03 / 2022
2020 - 21 / CFC / MUM / 648 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022


CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – PG – 01 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 21 – 22 / 7423
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 950058
LEAST COUNT : 0.05 kg/cm2
LOCATION : DR 1 - Reactor Pressure
ACCP. CRI. : + 2 % FSD
RANGE : -760 mmHg to 1 kg/cm2 OPERATING RANGE : -760mmHg to 0.5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
mmHg/ kg/cm2 mmHg/ kg/cm2 Output Output
1 -600 -600 -600 0 -600 0
2 -400 -400 -400 0 -400 0
3 0.00 0.00 0.00 0.00 0.00 0.00
4 0.50 0.50 0.50 0.00 0.50 0.00
5 1.00 1.00 1.01 0.01 1.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – PG – 02 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7424
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 670082
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 - Steam Line Pressure
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 6 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 7.00 7.00 7.01 0.01 7.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – PG – 03 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7425
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 580114
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 - Cooling Water Inlet Pressure
ACCP. CRI. : + 2 °C
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 5.99 -0.01 5.99 -0.01
5 7.00 7.00 6.98 -0.02 6.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 1 – PG – 04 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7426
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 880087
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 1 - Cooling Water Outlet Pressure
ACCP. CRI. : + 2 °C
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 7.00 7.00 7.02 0.02 7.02 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 2 – PG – 05 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7427
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 800088
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 2 - Reactor Pressure
ACCP. CRI. : + 2 % FSD
RANGE : -760 mmHg to 4 kg/cm2 OPERATING RANGE : -760mmHg to 0.5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
mmHg/ kg/cm2 mmHg/ kg/cm2 Output Output
1 -600 -600 -600 0 -600 0
2 -400 -400 -400 0 -400 0
3 0.00 0.00 0.00 0.00 0.00 0.00
4 2.00 2.00 2.00 0.00 2.00 0.00
5 4.00 4.00 4.00 0.00 4.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : AC – PG – 07 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7428
TYPE : Pressure Gauge
MAKE & MODEL : MAC Searial No. : -
LEAST COUNT : 0.5 kg/cm2
LOCATION : Air Compressor
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 21 kg/cm2 OPERATING RANGE : 0 to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 6.00 6.00 6.01 0.01 6.01 0.01
3 12.00 12.00 12.05 0.05 12.05 0.05
4 18.00 18.00 18.07 0.07 18.07 0.07
5 21.00 21.00 21.15 0.15 21.15 0.15

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : SCR – PG – 08 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7429
TYPE : Pressure Gauge
MAKE & MODEL : FGB Searial No. : D 6058
LEAST COUNT : 0.2 kg/cm2
LOCATION : Scrubber Pump
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 28 kg/cm2 OPERATING RANGE : 0 to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 5.00 5.00 5.03 0.03 5.03 0.03
3 10.00 10.00 10.05 0.05 10.05 0.05
4 15.00 15.00 15.06 0.06 15.06 0.06
5 20.00 20.00 20.08 0.08 20.08 0.08
6 28.00 28.00 28.14 0.14 28.14 0.14
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : CT – PG – 09 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7430
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 25428
LEAST COUNT : 0.1 kg/cm2
LOCATION : Cooling Tower
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 7.00 7.00 7.01 0.01 7.01 0.01

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR 3 – PG – 11 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7431
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 670003
LEAST COUNT : 0.1 kg/cm2
LOCATION : DR 3 - Reactor Pressure
ACCP. CRI. : + 2 % FSD
RANGE : -760 mmHg to 4.2 kg/cm2 OPERATING RANGE : -760mmHg to 0.5 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
mmHg/ kg/cm2 mmHg/ kg/cm2 Output Output
1 -600 -600 -600 0 -600 0
2 -400 -400 -400 0 -400 0
3 0.00 0.00 0.00 0.00 0.00 0.00
4 2.00 2.00 2.01 0.01 2.01 0.01
5 4.00 4.00 4.02 0.02 4.02 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : AC – PG – 12 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7432
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 03286
LEAST COUNT : 0.1 kg/cm2
LOCATION : Air Compressor
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 7 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
Output Output
Sr.
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.03 0.03 10.03 0.03
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 13 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7433
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : 160006
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 Cooling Condensor
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 7 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.00 0.00 4.00 0.00
4 6.00 6.00 6.01 0.01 6.01 0.01
5 8.00 8.00 8.02 0.02 8.02 0.02
6 10.00 10.00 10.01 0.01 10.01 0.01
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 14 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7434
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 200387
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 Surg Tank
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 3.99 -0.01 3.99 -0.01
4 6.00 6.00 5.98 -0.02 5.98 -0.02
5 8.00 8.00 7.97 -0.03 7.97 -0.03
6 10.00 10.00 9.96 -0.04 9.96 -0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 15 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7435
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 Cloth Filter
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.02 0.02 4.02 0.02
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.03 0.03 8.03 0.03
6 10.00 10.00 10.04 0.04 10.04 0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 16 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7436
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 Cloth Filter
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.02 0.02 4.02 0.02
4 6.00 6.00 6.04 0.04 6.04 0.04
5 8.00 8.00 8.05 0.05 8.05 0.05
6 10.00 10.00 10.07 0.07 10.07 0.07
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 17 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7437
TYPE : Pressure Gauge
MAKE & MODEL : Baumer Searial No. : -
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 Air Reciever
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 10 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.01 0.01 2.01 0.01
3 4.00 4.00 4.02 0.02 4.02 0.02
4 6.00 6.00 6.02 0.02 6.02 0.02
5 8.00 8.00 8.03 0.03 8.03 0.03
6 10.00 10.00 10.04 0.04 10.04 0.04
REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 18 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7438
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 050937
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 PSA 1
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.02 0.02 6.02 0.02
5 7.00 7.00 7.02 0.02 7.02 0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PG – 19 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7439
TYPE : Pressure Gauge
MAKE & MODEL : itec Searial No. : 730053
LEAST COUNT : 0.1 kg/cm2
LOCATION : N2 PSA 2
ACCP. CRI. : + 2 % FSD
RANGE : 0 to 7 kg/cm2 OPERATING RANGE : 0 to 4 kg/cm2
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
kg/cm2 kg/cm2 Output Output
1 0.00 0.00 0.00 0.00 0.00 0.00
2 2.00 2.00 2.00 0.00 2.00 0.00
3 4.00 4.00 4.01 0.01 4.01 0.01
4 6.00 6.00 6.01 0.01 6.01 0.01
5 7.00 7.00 6.98 -0.02 6.98 -0.02

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : NP – PS – 01 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7440
TYPE : Pressure Switch
MAKE & MODEL : Indfos Searial No. : 7911
SEARIAL NO. :-
LOCATION : Nitrogen Plant - Surge Tank Pressure
ACC. CRI. : + 2 % FSD
RANGE : 1 to 7 kg/cm2 OPERATING RANGE : 2 kg/cm2
AMB : 25 + 2 ° C
Sr. INPUT SET SET RESET
No. Kg/cm2 Kg/cm2 Kg/cm2 Kg/cm2

·          1.50 1.50 1.53 1.28

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY


(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : DR – VG – 01 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7441
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : 970100
LEAST COUNT :10 mmHg
LOCATION : Vacuum Reciever
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1

CALIBRATED BY
CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : VP – VG – 02 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7442
TYPE : Vaccum Gauge
MAKE & MODEL : Baumer Searial No. : 01840
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATED BY

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)

CALIBRATION REPORT
CLIENT : INDO AMINES LTD, DOMBIVLI. A - 43
TAG. NO. : VP – VG – 03 CAL. DATE : 04 / 03 / 2022
CERT. NO. : CR / 19 – 20 / 7443
TYPE : Vaccum Gauge
MAKE & MODEL : itec Searial No. : 030735
LEAST COUNT :10 mmHg
LOCATION : Vacuum Pump
ACC. CRI. : + 2 % FSD
RANGE : 0 to -760 mmHg OPERATING RANGE : 0 to -760 mmHg
AMB : 25 + 2 ° C
Master Output
Sr. INPUT Under Test Output Before Adjustment After Adjustment
No. Actual Actual
Deviation Deviation
mmHg mmHg Output Output
1 0 0.00 0.00 0.00 0.00 0.00
2 -200 -200.00 -200.00 0.00 -200.00 0.00
3 -400 -400.00 -400.00 0.00 -400.00 0.00
4 -600 -600.00 -600.00 0.00 -600.00 0.00
5 -700 -700.00 -700.00 0.00 -700.00 0.00

REMARKS : O. K
a) Calibration results reported in the certificate(s) are valid at the
time of & under the stated condition of measurement.
b) Error observed is within required acceptance criteria, hence
instrument is under calibration for given calibrated range.
MASTER INSTRUMENT USED
Sr.
IDENTIFICATION RANGE CAL.VALID UPTO
No.
1 Digital Pressure Gauge -1 to 40 bar 17 / 09 / 2022
2021 - 22 / CFC / MUM / 279 / 1
CALIBRATION VALID UPTO : 03 / 09 / 2022

CALIBRATED BY REVIEWED & CHECKED BY

(Newtech Instrument Services) (Indo Amines Ltd)


Ms. INDO AMINES LTD, DOMBIVLI.(A 43)
LIST OF CALIBRATION instruments

Digital Temp. Indicator


Sr. No. TAG No. Location Range Cal. Date Due Date
1 DR 1 – DTI – 01 DR 1 - Reactor Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
2 DR 1 – DTI – 02 DR 1 Vapour Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
3 DR 1 – DTI – 03 DR 1 - Vent Trap Inlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
4 DR 1 – DTI – 04 DR 1 - Vent Trap Outlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
5 DR 1 – DTI – 05 DR 1 Reflux Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
6 DR 1 – DTI – 06 DR 1 - Productive Condensor Temp. 0 to 200 °C 04 / 03 / 2022 03 / 09 / 2022
7 DR 1 – DTI – 07 DR 1 - Product Sub Cooler Inlet Temp. 0 to 200 °C 04 / 03 / 2022 03 / 09 / 2022
8 DR 1 – DTI – 08 DR 1 - Product Sub Cooler Outlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
9 DR 2 – DTI – 09 DR 2 Reactor Temp. 0 to 200 °C 04 / 03 / 2022 03 / 09 / 2022
10 DR 2 – DTI – 10 DR 2 Vapour Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
11 DR 3 – DTI – 11 DR 3 Reactor Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022

Temp. Sensor
Sr. No. TAG No. Location Range Cal. Date Due Date
12 DR 1 – TE – 01 DR 1 - Reactor Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
13 DR 1 – TE – 02 DR 1 Vapour Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
14 DR 1 – TE – 03 DR 1 - Vent Trap Inlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
15 DR 1 – TE – 04 DR 1 - Vent Trap Outlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
16 DR 1 – TE – 05 DR 1 Reflux Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
17 DR 1 – TE – 06 DR 1 - Productive Condensor Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
18 DR 1 – TE – 07 DR 1 - Product Sub Cooler Inlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
19 DR 1 – TE – 08 DR 1 - Product Sub Cooler Outlet Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
20 DR 2 – TE – 09 DR 2 Reactor Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
21 DR 2 – TE – 10 DR 2 Vapour Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022
22 DR 3 – TE – 11 DR 3 Reactor Temp. 0 to 300 °C 04 / 03 / 2022 03 / 09 / 2022

Pressure Gauge
Sr. No. TAG No. Location Range Cal. Date Due Date
23 DR 1 – PG – 01 DR 1 Reactor -760 mmHg to 1 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
24 DR 1 – PG – 02 DR 1 - Steam Line 0 to 7 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
25 DR 1 – PG – 03 DR 1 - Cooling Water Inlet 0 to 7 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
26 DR 1 – PG – 04 DR 1 - Cooling Water Outlet 0 to 7 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
27 DR 2 – PG – 05 DR 2 - Reactor -760 mmHg to 4 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
28 AC – PG – 07 Air Compressor 0 to 21 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
29 SCR – PG – 08 Scrubber Pump 0 to 28 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
30 CT – PG – 09 Cooling Tower 0 to 7 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
31 DR 3 – PG – 11 DR 3 - Reactor -760 mmHg to 4.2 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
32 AC – PG – 12 Air Compressor 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
33 NP - PG - 13 N2 Cooling Condensor 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
34 NP - PG - 14 N2 Surge Tank 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
35 NP - PG - 15 Cloth Filter 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
36 NP - PG - 16 Cloth Filter 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
37 NP - PG - 17 N2 Air Reciever 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
38 NP - PG - 18 N2 PSA 1 0 to 7 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
39 NP - PG - 19 N2 PSA 2 0 to 7 kg/cm2 04 / 03 / 2022 03 / 09 / 2022
40 NP – PS – 01 N2 Surge Tank 0 to 10 kg/cm2 04 / 03 / 2022 03 / 09 / 2022

Vaccum Gauge
Sr. No. TAG No. Location Range Cal. Date Due Date
41 DR 1 - VG - 01 DR 1 & 2 Vacuum Receiver 0 to -760 mmHG 04 / 03 / 2022 03 / 09 / 2022
42 VP - VG - 02 Vacuum Pump 0 to -760 mmHG 04 / 03 / 2022 03 / 09 / 2022
43 VP - VG - 03 Vacuum Pump 0 to -760 mmHG 04 / 03 / 2022 03 / 09 / 2022

You might also like