Professional Documents
Culture Documents
Bio 541L-001
Determining the pKa of P-Nitrophenol Using the Absorbance of Buffer Solutions with Differing
pH Values
Claffey 2
Purpose
The purpose of the experiment was to determine the pKa of P-Nitrophenol (PNP) using 8 buffer
solutions, each with a different pH, and a spectrophotometer to measure absorbance resulting in
Background
To find the pKa of P-nitrophenol buffer solutions are needed. A buffer solution is a
solution where the pH is constant, or has little movement, when introduced to acids3. All buffer
solutions have a buffer range. The buffer range is the amount of variance in the pH of the
solution without becoming acidic or basic4. The pKa of a compound is directly related to the pH
of the compound5. The pKa and pH can be related to each other through the Hendersen-
Hasselbalch equation5. The Hendersen-Hasselbalch equation is pH= pKa + log ([A]/[HA])6. The
pKa can be found looking at the steepest slope of the curve between pH and absorbance6. At the
The dissociation reaction is important because demonstrates what certain chemical’s effect on
Beer’s law is denoted by the equation A = elc and is useful to understand the amount of
energy of absorbed (A) using the molar attenuation coefficient (e), the path length (l) and the
solution concentration (c) being used8. To determine the absorbance of the solution a
spectrophotometer can be used7. The absorbance indicates how much solute is in the solution7.
Absorbance of different wavelengths can be found using a spectrophotometer1. After setting the
values) can be placed in the spectrophotometer and the absorbance values can be measured1.
Therefore, the spectrophotometer can help determine the pH/pKa of a solution such as P-
nitrophenol1.
To begin the experiment, use 8 unknown buffer solutions1. Tear 8 separate pieces of pH
paper and place them on a paper towel1. Label the pieces of pH paper 1-81. Using a 20 µl
micropipette, place one drop of each buffer solution on one piece of pH paper1. Immediately
compare the color of the pH paper to the pH color chart1. Record each of the buffer solutions’
pH1.
To prepare the samples, label 8 test tubes 5.0, 6.0, 7.0, 7.5, 8.0, 8.5, 9.0, 10.01. Using a
pipettor, place 1mL of P-Nitrophenol into each of the 8 test tubes1. Place 3mL of each buffer
solution into the glass test tube with the corresponding pH1. After each buffer solution is added
to its corresponding test tube discard of the pipettor tip1. Take 10 cuvettes and label them B
(blank), S (sample), 5.0, 6.0, 7.0, 7.5, 8.0, 8.5, 9.0, 10.01. In the “B” cuvette add 2.0 mL of pH 10
buffer solution1. In the “S” cuvette add 0.5mL of 0.1 P-nitrophenol and 1.5mL of the buffer pH
10.0 solution1. In the following 8 cuvettes place 2.0mL of corresponding buffer solution into the
machine set the wavelength to 360nm using the button “go to lambda” and using the keys up and
down keys1. When done, press the “OK” button and place the B-cuvette in the first position and
the S-cuvette in the fourth position1. Shut the lid, press “zero” to blank1. Pull the handle to place
the S-sample into the correct position and record the absorbance observed1. Using the
wavelengths 360nm-430nm (every 10nm), find the max wavelength1. Use all 8 cuvettes and
label the wavelength for each of the cuvettes1. Be sure to zero the spectrophotometer after each
cuvette’s wavelength is recorded1. Read the results and determine the maximum wavelength1.
To determine the pKa of P-nitrophenol, use the spectrophotometer again1. Set the
machine to the maximum wavelength found in the previous step1. Remove the S-cuvette and the
B-cuvette1. Record absorbance found for each of the 8 cuvettes1. Read the results and determine
the pKa for P-nitrophenol (PNP)1. Using excel create a graph of absorbance vs. wavelength and a
Results
Table 1
Table one includes data from 10nm increments of wavelengths between 360nm to 430nm. The
Table 2
Table 2 includes data from 8 different pH buffer solutions at 5.0, 6.0, 7.0, 7.5, 8.0, 8.5, 9.0, 10.0
Figure 1
0.3
Wavelength=
0.25
400nm
0.2
0.15
0.1
0.05
0
350 360 370 380 390 400 410 420 430 440
Wavelength (nm)
Figure 1 indicates the maximum wavelength of the 8 buffer solutions. The figure shows the 8
wavelengths of 360nm, 370nm, 380nm, 390nm, 400nm, 410nm, 420nm, 430nm vs. the observed
Figure 2
Claffey 6
Absorbance vs. pH
0.5
0.4
pKa= 7.4
Absorbance (AU)
0.3
0.2
0.1
0
0 2 4 6 8 10 12
-0.1
pH
Figure 2 indicates the pKa of P-nitrophenol. The pH of the 8 buffer solutions at 5.0, 6.0, 7.0, 7.5,
8.0, 8.5, 9.0, 10.0 vs. the observed absorbance values. The pKa of p-nitrophenol was found to be
7.4.
Discussion
The maximum wavelength of the graph was found by comparing the absorbance vs. the
wavelength1. The maximum wavelength had the highest absorbance value1. Table 1 showed the
380nm, 390nm, 400nm, 410nm, 420nm, 430nm. Figure 1 was created using the data from table
1. Figure 1 showed a bell-shaped graph in which the maximum wavelength was easily
identifiable. In figure 1, the maximum wavelength in the experiment was found to be 400nm.
The spectrophotometer was set to the maximum wavelength and the cuvettes containing
the 8 buffer solutions with pH of 5.0, 6.0, 7.0, 7.5, 8.0, 8.5, 9.0, 10.0 were be placed in the
spectrophotometer1. Table 2 showed the data collected after setting the spectrophotometer to the
maximum wavelength, 400nm, and measuring the absorbance of pH values 5.0, 6.0, 7.0, 7.5, 8.0,
8.5, 9.0, 10.0. Figure 2 was created off of the data found in table 2. In figure 2, the graph showed
Claffey 7
the absorbance vs. the pH (pH=pKa6). The pKa was found at the steepest part of the slope of the
graph.
The pH of P-nitrophenol was found to be around 7.4. Although there is a range of pH for
P-nitrophenol, it was estimated to be around 7.079. The estimated value of 7.4 was relatively
close but slightly off. This could be due to experimental error. A possible experimental error
could be not having perfect measurements due to some solution being left in the pipettor tips.
Another potential error was when estimating the pKa using figure 2. The steepest part of the
slope was determined to be the pKa of p-nitrophenol (PNP)6. The steepest part of the slope was
hard to slope was difficult to get precisely accurate. This could have caused the experimental
pKa to be higher or lower which could have led to the pKa being closer to the expected value of
7.079.
Claffey 8
References
Nitrophenol
https://www.chemguide.co.uk/physical/acidbaseeqia/buffers.html
https://courses.lumenlearning.com/introchem/chapter/buffer-range-and-capacity/
5. Reijenga J, van Hoof A, van Loon A, Teunissen B. Development of Methods for the
Determination of pKa Values. Analytical chemistry insights. 2013 Aug 8 [accessed 2021
8]. https://hmdb.ca/metabolites/HMDB0001232
Claffey 9
Appendix