Professional Documents
Culture Documents
11 - Magnetic Resonance Imaging - Introduction To The Science of Medical Imaging - R. Nick Bryan
11 - Magnetic Resonance Imaging - Introduction To The Science of Medical Imaging - R. Nick Bryan
http://ebooks.cambridge.org/
Chapter
Sixty years after the first demonstration of nuclear water) become aligned with the field so as to create a
magnetic resonance in condensed phase, and over macroscopic magnetic moment, also called magnet-
three decades after the first cross-sectional image was ization, M. By applying a second, circularly polarized
published, magnetic resonance imaging (MRI) has radiofrequency (RF) magnetic field, denoted B1, acting
without doubt evolved into the richest and most ver- perpendicular to the static field, M is deflected into a
satile biomedical imaging technique today. Initially a plane transverse to the static field, precessing at a
mainly anatomical and morphological imaging tool, frequency ω0 ¼ γB0 , called Larmor frequency, and
MRI has, during the past decade, evolved into a func- inducing a voltage in a suitably positioned receive
tional and physiological imaging modality with a coil (which can but need not be identical to the trans-
wide spectrum of applications covering virtually all mit coil that generates the B1 RF field). At a typical
organ systems. Today, MRI is a mainstay of diagnostic field strength of a magnetic resonance imager of
imaging, playing a critically important role for patient 15 000 Gauss, the resonance frequency is 64 MHz.
management and treatment response monitoring. We can describe the motion of the magnetization
Even though the physics of MRI is well understood, vector in terms of the phenomenological Bloch
getting a grasp of the method can be challenging to equation:
the uninitiated. This brief chapter seeks to introduce
the MRI novice to the fundamentals of spin excitation dM Mxy Mz M0
¼ γM B (7:1)
and detection, detection sensitivity, spatial encoding dt T2 T1
and image reconstruction, and resolution, and to pro-
vide an understanding of the basic contrast mechan- where γ is the gyromagnetic ratio, B the effective
isms. For an in-depth treatment of theory, physics, magnetic field, M0 the equilibrium magnetization
and engineering aspects, the reader is referred to the and T1 and T2 the relaxation times for the longitu-
many excellent texts [1–3]. Applications to specific dinal and transverse component, respectively, of the
organ systems are covered in other chapters of this magnetization.
book. Equation 7.1 forms a set of coupled differential
equations, which can easily be solved for typical ini-
Magnetic resonance signal tial conditions by anyone familiar with just the fun-
Unlike transmission techniques such as computed damentals of vector calculus. Let us assume that the
tomography (CT), where the attenuation of a beam magnetization has been perturbed from equilibrium
of radiation is measured and an image reconstructed (M0) by an RF pulse, thereby rotating it into the
from multiple angular projections, the nuclear mag- transverse plane, and further, that this perturbation
netic resonance (NMR) phenomenon exploits the occurred in a short time relative to the relaxation
magnetic properties of atomic nuclei, typically pro- times, so that relaxation during the pulse can be
tons. When placed in a strong uniform magnetic neglected.
field B0 (also referred to as the static or polarizing Following the rules of vector multiplication and
field), the nuclear magnetic moments µ of a sample considering further that for the static field (the only
(such as a portion of mammalian tissue containing magnetic field present following brief action of the
Introduction to the Science of Medical Imaging, ed. R. Nick Bryan. Published by Cambridge University Press. © Cambridge University
Press 2010.
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:57 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Chapter 7: Magnetic resonance imaging
_ x ¼ ω0 My Mx =T2
M
_ y ¼ ω0 Mx My =T2
M (7:2a–c)
_ z ¼ ðMz M0 Þ=T1
M
We now see that only Equations 7.2a and 7.2b are t/T1
0 1 2 3 4 5
coupled. At this stage it is useful to introduce complex
Figure 7.2. Solution to Equation 7.2c showing temporal evolution
notation for the transverse magnetization as of longitudinal magnetization for initial condition Mz(0) = 0.
Mxy ðtÞ ¼ Mx þ iMy : We can then express Equations
7.2a and 7.2b as follows: M _ xy ¼ Mxy ðiω0 1=T2 Þ: By
imposing the initial condition Mxy ð0Þ ¼ M0 ; this Larmor frequency and its amplitude decays exponen-
equation is readily integrated to yield: tially with a time constant equal to the transverse
relaxation time, as shown graphically in Figure 7.1.
Mxy ¼ M0 eiω0 t et=T2 (7:3) The detected signal is called free induction decay (FID).
Equation 7.2c, expressing the temporal evolution
The real and imaginary components of Equation 7.3 of the longitudinal magnetization (which cannot
are thus a damped cosine and sine, respectively, with directly be detected), can be integrated to yield:
frequency ω0 and decay time constant T2. Thus, the
magnetization following an RF pulse oscillates at the Mz ðtÞ ¼ M0 þ ðMz ð0Þ M0 Þet=T1 (7:4)
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:57 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Section 2: Biomedical images: signals to pictures
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:58 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Chapter 7: Magnetic resonance imaging
S(X, Y )
(a) S(kx, ky) (b)
y
ky
kx
Figure 7.4. (a) Real part of the k-space signal for a point source of spin density located at equal distance from x- and y-axes, shown in (b).
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:58 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Section 2: Biomedical images: signals to pictures
6X6
12X12
64X64
256X256
164
Figure 7.6. Effect of increasing k-space coverage for image reconstruction: (left) k-space, (right) image space. (a) and (b) show schematically k-
space samples used for reconstruction, based on center 6 × 6 and 12 × 12 array; (c,d) left: actual k-space plots for 64 × 64 and 256 × 256 data
arrays. (Images courtesy Dr. Hsiao-Wen Chung, National Taiwan University, Taiwan.)
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:58 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Chapter 7: Magnetic resonance imaging
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:58 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Section 2: Biomedical images: signals to pictures
Figure 7.8. (Top) Plot of EPI signal intensity of time-series of images obtained at 1.5 second intervals from a patient with a left MCA territory
infarct. (Bottom) Images representing successive time-frames 17–31 showing dip in signal intensity as the bolus traverses the brain for the
right but not the left infracted region. (Data courtesy of Dr. R. Wolf, Department of Radiology, University of Pennsylvania.)
of a bolus of Gd-DTPA causing transient signal reduc- conditions. At the time of perturbation, the nuclear
tion due to the agent’s paramagnetism, along with a plot magnetic moments typically have not fully recovered
of signal intensity. from the effect of the previous pulse sequence cycle’s
RF pulses, nor is the signal usually detected immedi-
Signal-to-noise, contrast, and image ately after its creation.
Typically, a spin-echo is detected as a means to
resolution alleviate spin coherence losses from static field in-
The single most distinguishing feature of MRI (com- homogeneity. The spin-echo signal amplitude for an
pared, for example, to x-ray-based modalities such as RF pulse sequence π/2-τ-π-τ repeated every TR
CT), is its extraordinarily large innate contrast, seconds is approximately given by:
which, for two soft tissues, can be on the order of
several hundred percent. In x-ray imaging, contrast Sðt ¼ 2τÞ ð1 eTR =T1 ÞeTE =T2 (7:15)
is a consequence of differences in the attenuation
coefficients for two adjacent soft-tissue structures The above is a good approximation as long as TE << TR
and is on the order of a few percent at best. If it relied and T2 << TR, in which case there is no residual trans-
only on the differences in proton spin density, ρ, of verse magnetization across successive pulse sequence
various tissues, MRI contrast would be on the order cycles. In Equation 7.15 ρ is the voxel spin density and
of 5–20% at best, whereas in reality contrast between echo time TE = 2τ. Thus, the contrast between adjacent
two regions, A and B, defined as: structures can be optimized by solving:
can be several hundred percent. The reason for this for Ti = TR or TE. In practice, however, this is not
discrepancy is the circumstance that the MR signal particularly useful, since the relaxation times are not
166
commonly is acquired under non-equilibrium known a priori and a multitude of tissue interfaces
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:59 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Chapter 7: Magnetic resonance imaging
exist. Empirically, it is known that tissues differ in CSF, appears hyperintense relative to brain tissue,
at least one of the intrinsic quantities, T1, T2, or ρ. It whose water protons have considerably shorter
therefore suffices to acquire images in such a manner T2. These relationships are readily apparent from
that contrast is sensitive to one particular parameter. Equation 7.15.
For example a “T2-weighted” image would be The parameters T1, T2, and ρ are called “intrinsic”
acquired with TE ~ T2 and TR >> T1 and, similarly, a because they are endogenous and, in general, not
“T1-weighted” image with TR < T1 and TE << T2. In a modifiable except when a relaxation reagent is
T1-weighted image, the signal of protons with “long administered. Most common among the latter are
T1” would be attenuated relative to that of protons chelates of gadolinium-III that shorten T1 of tissue
with shorter T1. A case in point is cerebrospinal fluid water protons (see, for example, [17] and Chapter 9).
(CSF), which has a T1 approximately three times Figure 7.9 shows a pair of images obtained from
longer than protons in gray or white matter, and a patient with brain pathology, illustrating the char-
thus appears hypointense. Similarly, in a T2-weighted acteristic pulse sequence parameter-dependent
image, the signal from protons with long T2, such contrast.
as those in malignant brain tumors or ventricular T1 relaxation times gradually increase with
increasing field strength [18]. For this reason, when
T1 relaxation times are reported, the field strength
Table 7.1. T1 and T2 in various mammalian tissues at 1.5 T must be indicated. At 1.5 tesla field strength, T1 in
Tissue T1 (ms) T2 (ms)
soft tissues ranges from about 500 milliseconds to
1 second (Table 7.1). An exception is adipose tissue
Muscle
(e.g., fat in the scalp, Figure 7.10, or bone marrow and
skeletal 870 47
subcutaneous tissue, Figure 7.6), where the signal
cardiac 870 57 results from fatty-acid glyceride protons, which
Liver 490 43 have much shorter T1. Water in white matter of the
Kidney 560 58 corpus callosum has a shorter T1 than water in gray
Spleen 780 62 matter of the brain (Figure 7.10). This difference is a
Adipose 260 84 consequence of the lower water content of white
Brain
relative to gray matter.
Another important quantity that affects the per-
Gray matter 920 101
ceived contrast is the signal-to-noise ratio (SNR). The
White matter 790 92
smallest signal-producing element is the voxel (volume
T1 was calculated using an empirical relationship for the field element), which is equal to the product of pixel area
dependence [18]; T2 values are from Table 12 of Reference [18]. and slice thickness. Therefore, for a given hardware
configuration (magnetic field strength, RF coil, and
167
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:59 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Section 2: Biomedical images: signals to pictures
receive system) the relative signal strength is propor- CNRAB ¼ CAB =eff (7:18)
tional to voxel volume, and thus represents the numer-
ator in the expression for SNR. The noise has a It is noted, for example, that small anatomic detail
complicated dependence on various parameters, such as the stalk of the pituitary gland is better visual-
which include field strength, size and design of the ized in the image of Figure 7.10(b) in spite of the fact
radiofrequency receive coil, and the volume of tissue that the two images have equal contrast.
“seen” by the coil. We will return to this aspect below. We have previously seen that resolution is deter-
For the time being it suffices to note that SNR can be mined by the area of k-space covered (Equation 7.12
increased by summing of the signal through repeated and Figure 7.6). Ordinarily, the pixel size is regarded as
data collection. This is possible because of the stochas- the figure of merit for resolution. Suppose we increase
tic nature of the noise, whose amplitude increases only the number of samples from 256 to 512, which can, for
as the square root of the number of signal samples example, be achieved by doubling the frequency-
taken, whereas the signal increases linearly. Signal encoding sampling time. In this case the k-space area
averaging occurs at several levels during the k-space is doubled, thus pixel size halved, and resolution is
sampling process. therefore improved by a factor of two. But now con-
From the foregoing we obtain a simple formula for sider the situation where the pixel size is halved in
SNR per voxel of volume ΔV: image space (for example by linear interpolation).
While this operation may result in a smoother image,
pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
SNR ¼ C DV Nx Ny Nav the resolution, in terms of the observer’s ability to
pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
¼ C Dx Dy dz Nx Ny Nav (7:17) discern small structures, is not enhanced. Similarly, if
we were to double k-space without actually acquiring
the high-frequency data (by zero filling those loca-
In Equation 7.17 Dx ¼ FOVx =Nx ; Dy ¼ FOVy =Ny are
tions), we would, after Fourier reconstruction, obtain
the pixel dimensions given by the field of view (FOV)
an image at half the pixel size. Even though this image
and the number of data samples, Nx and Ny, along x
would again be more pleasing to the eye, its effective
and y coordinates. The quantity dz is the thickness of
resolution would not be improved. Figure 7.11 dis-
the slab selected by the slice-selective RF pulse and is
plays an axial image through the orbit, showing
the spin density weighted by relaxation effects deter-
exquisite anatomic detail and achieved by sampling
mined by relaxation times T1 and T2 and the pulse
kx- and ky-space from − 21.3 to + 21.3 cm−1.
sequence timing parameters. Finally, Nav is the num-
ber of signal averages. Figure 7.10 shows two images of
the human brain obtained from the same anatomic MRI hardware
location but differing in SNR. The hardware comprising an MRI system differs sig-
Lastly, the noise (essentially the standard deviation nificantly from that of other imaging modalities
of the signal, σeff) affects the effective contrast per- such as CT or PET, discussed elsewhere in this book.
168 ceived by the observer. The contrast-to-noise ratio A simplified functional block diagram of a typical
(CNR) is simply the ratio of contrast divided by σeff: MR imaging apparatus is given in Figure 7.12. The
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:59 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Chapter 7: Magnetic resonance imaging
heart of the system is the magnet, typically a super- induces a RF voltage in the receive coil (which may
conducting system, to create the polarizing field. be identical by serving for both transmission and
Another critical subsystem is the RF transmit/receive reception). The ensuing signal is then amplified
assembly. It consists of a transmitter and power ampli- in the receiver and digitized. Further unique to the
fier for generating the RF voltage to be fed into the MR imaging device is the gradient subsystem,
RF coil thereby creating the B1 field for spin excitation. which permits generation of the previously defined
The resulting transverse magnetization, in turn, time-dependent gradient fields needed for spatial
encoding. Both transmit/receive and gradient subsys-
tem are under the control of a data acquisition pro-
cessor, which ties in to the main computer. For more
detail the reader is referred to the many excellent texts
(see, for example, [1]).
169
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:01:59 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Section 2: Biomedical images: signals to pictures
S ω2
SNR ¼ / pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi (7:19)
N aω1=2 þ bω2
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:02:00 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016
Chapter 7: Magnetic resonance imaging
organ metabolism. The quantum step in detection 13. Sorensen AG, Buonanno FS, Gonzalez RG, et al.
sensitivity brought about by technologies for polar- Hyperacute stroke: evaluation with combined multisection
izing spins several orders of magnitude above that diffusion-weighted and hemodynamically weighted
achieved by thermal polarization is likely to revolu- echo-planar MR imaging. Radiology 1996; 199: 391–401.
tionize the field and lead to MR rivaling methods 14. Conturo TE, Lori NF, Cull TS, et al. Tracking neuronal
such as positron emission tomography. fiber pathways in the living human brain. Proc Natl
Acad Sci USA 1999; 96: 10422–7.
References 15. Schmitt F, Stehling MK, Turner R. Echo-Planar Imaging:
Theory, Technique and Applications. Berlin: Springer,
1. Vlaardingerbroek MT, den Boer JA, eds. Magnetic 1998.
Resonance Imaging: Theory and Practice. Berlin:
Springer, 1996. 16. Sorensen AG, Tievsky AL, Ostergaard L, Weisskoff RM,
Rosen BR. Contrast agents in functional MR imaging.
2. Haacke EM, Brown RW, Thompson MR, Venkatesan R. J Magn Reson Imaging 1997; 7: 47–55.
Magnetic Resonance Imaging: Physical Principles and
Sequence Design. New York, NY: Wiley–Liss, 1999. 17. Runge VM, Kirsch JE, Wells JW, et al. Magnetic resonance
contrast agents in neuroimaging: new agents and
3. Liang ZP, Lauterbur PC. Principles of Magnetic applications. Neuroimaging Clin N Am 1994; 4: 175–83.
Resonance Imaging. IEEE Press Series in Biomedical
Engineering. New York, NY: IEEE, 2000. 18. Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM.
A review of normal tissue hydrogen NMR relaxation
4. Lauterbur PC. Image formation by induced local times and relaxation mechanisms from 1–100 MHz:
interactions: examples employing nuclear magnetic dependence on tissue type, NMR frequency,
resonance. Nature 1973; 242: 190. temperature, species, excision, and age. Med Phys
5. Kumar A, Welti D, Ernst R. NMR Fourier 1984; 11: 425–48.
zeugmatography. J Magn Res 1975; 18: 69–83. 19. Hoult DI, Lauterbur PC. The sensitivity of the
6. Edelstein WA, Hutchison JMS, Johnson G, Redpath T. zeugmatographic experiment involving human subjects.
Spin warp NMR imaging and applications to human J Magn Reson 1979; 34: 425–33.
whole-body imaging. Phys Med Biol 1980; 25: 751–6. 20. Edelstein WA, Glover GH, Hardy CJ, Redington RW.
7. Bydder GM, Steiner RE, Young IR, et al. Clinical NMR The intrinsic signal-to-noise ratio in NMR imaging.
imaging of the brain: 140 cases. AJR Am J Roentgenol Magn Reson Med 1986; 3: 604–18.
1982; 139: 215–36. 21. Hardy CJ, Bottomley PA, Roemer PB, Redington R.
8. Haase A, Frahm J, Matthaei D, Hänicke W, Merboldt Rapid 31P spectroscopy on a 4 T whole system. Magn
KD. FLASH Imaging: rapid NMR imaging using low flip Reson Med 1988; 8: 104–9.
angle pulses. J Magn Res 1986; 67: 258–66. 22. Norris DG. High field human imaging. J Magn Reson
9. Wehrli FW. Fast-Scan Magnetic Resonance: Principles Imaging 2003; 18: 519–29.
and Applications. New York, NY: Raven, 1990. 23. Robitaille PM, Abduljalil AM, Kangarlu A, et al. Human
10. Haacke EM, Tkach J. Fast MR imaging: techniques and magnetic resonance imaging at 8 T. NMR Biomed
clinical applications. AJR Am J Roentgenol 1990; 155: 1998; 11: 263–5.
951–64. 24. Robitaille PM, Abduljalil AM, Kangarlu A. Ultra
11. Mansfield P. Multiplanar image formation using NMR high resolution imaging of the human head at 8 tesla:
spin echoes. J Phys C 1977; 10: L55–8. 2K × 2K for Y2K. J Comput Assist Tomogr 2000;
12. Detre JA, Alsop DC. Perfusion magnetic resonance 24: 2–8.
imaging with continuous arterial spin labeling: methods 25. Yacoub E, Shmuel A, Pfeuffer J, et al. Imaging brain
and clinical applications in the central nervous system. function in humans at 7 tesla. Magn Reson Med 2001;
Eur J Radiol 1999; 30: 115–24. 45: 588–94.
171
Downloaded from Cambridge Books Online by IP 85.122.26.89 on Wed Mar 16 13:02:00 GMT 2016.
http://dx.doi.org/10.1017/CBO9780511994685.008
Cambridge Books Online © Cambridge University Press, 2016